You are on page 1of 6

Journal of Orthopaedic & Sports Physical Therapy

Official Publication of the Orthopaedic and Sports Physical Therapy Sections of the American Physical Therapy Association

Effect of Transducer Velocity on


Intramuscular Temperature During a
1-MHz Ultrasound Treatment
Stephanie L. Weaver, LAT, ATC 1
Timothy J. Demchak, PhD, LAT, ATC 2
Marcus B. Stone, PhD, LAT, ATC 3
Jody B. Brucker, PhD, LAT, ATC 4
Phillip O. Burr, PA, MS, ATC 5

M
Study Design: A 3 × 2 repeated-measures design was used. The independent variables were ost textbooks rec-
transducer velocity (2-3 cm/s, 4-5 cm/s, and 7-8 cm/s) and time (pretreatment and posttreatment). ommend the fol-
Objective: To determine if transducer velocity of a 1-MHz ultrasound treatment affects intramus- lowing parameters
cular tissue temperature. to vigorously heat
Background: Most authors advocate ultrasound transducer velocities of 2 to 4 cm/s within an area deep muscle with
of 2 to 3 times the effective radiating area or 2 times the size of the transducer head. However, a
ultrasound: frequency, 1 MHz;
much faster rate of application (approximately 7-8 cm/s) is often observed in clinical settings.
intensity, 1.5 W/cm2; duration,
Methods and Measures: Eleven healthy screened volunteers (9 males, 2 females; mean ± SD age,
22.6 ± 1.7 years; mean ± SD height, 175.7 ± 13.7 cm; mean ± SD body mass, 82.5 ± 19.5 kg)
10 minutes; treatment area, twice
were randomly assigned to a treatment order with all conditions administered during a single the size of the transducer; velo-
testing session. Each transducer velocity condition was administered for 10 minutes, using 1-MHz city of the transducer, 2 to 4
ultrasound with a 100% continuous duty cycle at an intensity of 1.5 W/cm2 over an area twice cm/s.6-8,11,23,29 Although support
the size of the transducer head. After the first treatment, the 2 remaining subsequent velocity for the frequency, intensity, dura-
conditions were administered after the intramuscular temperature returned to within ± 0.3°C of the tion, and treatment area can be
initial pretreatment temperature for 5 minutes. The dependent variable was left triceps surae found,3,5,9,12,16,21,28 information on
muscle temperature measured at 3 cm below one half the measured skinfold thickness. the effect of transducer velocity on
Results: Temperature increase across the 3 velocities was within 0.4°C (F2,20 = 0.07, P = .93). increases in tissue temperature is
Posttreatment values (mean ± SD) ranged from 42.7°C ± 2.3°C for the slowest velocity to 43.1°C absent in the literature. This void
± 1.4°C for the fastest velocity. Temperature increase was significant for time (F1,10 = 155.68,
is apparent when tracing refer-
P ⬍ .00001), increasing from 37.8°C ± 0.8°C pretreatment to 42.9°C ± 1.9°C after treatment.
ences from several textbooks and
Conclusion: Very similar intramuscular temperature increases can be observed among ultrasound
treatments (10-minute duration, 1-MHz frequency, 100% continuous duty cycle, 1.5 W/cm2
published articles, which provide
intensity, within an area twice the size of the transducer head), with transducer velocities of 2 to no data to support the suggested
3, 4 to 5, and 7 to 8 cm/s. J Orthop Sports Phys Ther 2006;36(5):320-325. doi:10.2519/ transducer velocity. 1-5,8,9,10,12,16-
18,20-22,24,26-28
jospt.2006.2157 High beam nonuni-
formity ratio (BNR) is typically
Key Words: calf heating, piezoelectric modality, therapeutic modality
cited as a reason to move the
transducer head in this fashion to
1
Marketing Representative, OrthoIndy, Indianapolis, IN. decrease the risk of causing pain
2
Assistant Professor, Athletic Training Department, Indiana State University, Terre Haute, IN.
3
Director of Orthopedic Trauma Research, OrthoIndy, Indianapolis, IN. to the patient, while maximizing
4
Assistant Professor, Athletic Training Department, Indiana State University, Terre Haute, IN. thermal effects. 6-8,23,29 Today,
5
Student, Athletic Training Department, Indiana State University, Terre Haute, IN. manufacturers are producing
Address correspondence to Timothy J. Demchak, C-10 Arena Building, Athletic Training Department,
Indiana State University, Terre Haute, IN 47809. E-mail: athdemchi@sugw.indstate.edu transducers that contain crystals
This study was funded by the Indiana State University Office of Sponsored Programs. Informed consent with lower BNRs, which may allow
and approval for research using human subjects were obtained from the Health and Human Performance for transducers to be moved at
Institutional Review Committee of Indiana State University. We affirm that we have no financial affiliation
(including research funding) or involvement with any commercial organization that has a direct financial slower rates, while still protecting
interest in any matter included in this manuscript. the patient.

320 Journal of Orthopaedic & Sports Physical Therapy


Therapeutic modality textbooks6,11,23,25,29 often a Latin square and 2 subjects went through each of
state that moving the transducer rapidly at greater the 6 possible treatment orders. During the entire
than 4 cm/s will result in decreasing the total testing procedure, all subjects remained prone. A
amount of energy absorbed per unit area and lead to surgical pen mark was made on the posterior aspect
decreased heating benefit. However, as previously of the medial triceps surae muscle at the greatest
stated, there is no empirical evidence of this claim. girth to identify the middle of the treatment area. A
Moreover, we have observed the application of higher mean of 3 consecutive vertical skinfold measurements
transducer velocities (7-8 cm/s) in several clinical at this site was calculated to estimate superficial tissue
settings and believe that many clinicians use higher thickness. The desired depth for the thermocouple
transducer velocities than recommended. below the treatment area was 3 cm plus half the value
The purpose of this study was to examine the effect of the measured superficial tissue thickness. The
of 3 transducer velocity conditions (2-3, 4-5, and 7-8
insertion site was located and marked by laying the
cm/s) on intramuscular tissue temperature during a
carpenter’s square flush against the medial triceps
standardized ultrasound application. We hypothesized
surae muscle so that the 90° angle was 5.08 cm
that there would be no differences in temperature
(length of the needle) from the middle of the
increases among the velocity conditions. This is based
on the theory that an equal amount of acoustic treatment area and measuring down from the right
energy will be transmitted into the tissue from a angle of the carpenter’s square 3 cm, plus half the
standardized ultrasound treatment regardless of the value of the measured superficial tissue thickness
transducer velocity. There is evidence to suggest that (Figure 1).
ultrasound treatments with equivalent spatial tempo- The treatment area, insertion site, and the sur-
ral averages result in equal increases in tempera- rounding areas were shaved and cleansed with
ture.14 Betadine (The Purdue Frederick Company, Norwalk
CT). Prior to insertion, the thermocouple (type
TX-23-21; Columbus Instruments, Columbus, OH)
METHODS
was threaded into a 5.08-cm 18-gauge reusable hypo-
A 3 × 2, repeated-measures design was used. The dermic needle (Popper & Sons, Inc, New Hyde Park,
independent variables were transducer velocity (2-3, NY) so that the tip of the thermocouple was within
4-5, and 7-8 cm/s) and time (pretreatment and the beveled tip of the needle. The needle and
posttreatment). The dependent variable was left tri- thermocouple were both fully inserted (5.08 cm)
ceps surae intramuscular tissue temperature at a parallel to the carpenter square so that the thermo-
depth of 3 cm below one half the measured skinfold couple was in the middle of the treatment area
thickness. This depth was chosen to compare our (Figure 2). The thermocouple was stabilized while
results to other ultrasound studies. the needle was extracted and then was secured to the
leg with clear adhesive tape to prevent it from
Subjects moving.
The thermocouple was connected to a portable
Eleven healthy (9 males, 2 females; mean ± SD age, data acquisition device (Datalogger, MSS-3000; Com-
22.6 ± 1.7 years; mean ± SD height, 175.7 ± 13.7 cm; mtest Instruments Ltd, Christchurch, New Zealand),
mean ± SD body mass, 82.5 ± 19.5 kg) of 12 subjects which measured and recorded the temperature at the
who were recruited and volunteered to participate tip of the thermocouple. Intramuscular temperature
completed this study. Data from 1 subject were was recorded every second starting at the time of
RESEARCH
discarded due to equipment malfunction. Subjects thermocouple insertion, and continued until the last
were excluded if they reported vascular or neurologi- treatment was complete. The pretreatment intramus-
cal disorders, recent injuries, surgeries in the 6 cular temperature prior to the first treatment was
months prior to testing, or allergies to penicillin, recorded after the temperature remained unchanged
Keflex, or latex. The study was approved by the (± 0.1°C) for 5 minutes, which took approximately 20
Health and Human Performance Institutional Review to 30 minutes after thermocouple insertion. This
Committee of Indiana State University and each temperature was used as the pretreatment tempera-
subject consented to participate prior to data collec- ture for the rest of the study.
REPORT

tion. To make sure the sound head was channeled


overtop an area equal to 2 widths of the transducer
Procedures head, a clear (Lucite) rectangular-shaped template
cut to twice the size of the transducer head (Figure
Each subject reported to the laboratory dressed in 3) was placed on the skin. This ensured that the
shorts and a T-shirt. Upon arrival the subjects com- center of the transducer traveled directly over the
pleted the informed consent and were screened. The location of the tip of the thermocouple and that the
subjects were assigned to a treatment order based on transducer did not consistently remain over any area

J Orthop Sports Phys Ther • Volume 36 • Number 5 • May 2006 321


skin or air during the treatment. Ultrasound gel was
Middle of applied as needed throughout each of the conditions.
treatment 5.08 cm
Following each condition, the treatment area and
thermocouple insertion site were observed for any
indication of adverse reaction. At the conclusion of
3 cm + the study, the thermocouple was removed, the sub-
superficial tissue ject’s leg was cleansed with Betadine and a sterile
thickness bandage was applied to the thermocouple insertion
site. The subjects were given instructions on proper
care of the wound and were instructed to seek care at
Insertion site the Student Health Center if any concerns arose.
Thermocouples were disinfected by soaking them in
CidexPlus 3.4% glutaraldehyde solution (Johnson &
Johnson Company, Irvine, CA) for at least 20 minutes
prior to use, and the reusable hypodermic needles
were sterilized by a regimen of soaking in CidexPlus
FIGURE 1. Determination of thermocouple insertion site. Middle of and being autoclaved.
treatment area is 5.08 cm horizontal from insertion site. The
insertion site is 3 cm plus half the measured skinfold thickness.
Statistical Analysis
Descriptive statistics at the pretreatment and post-
treatment times for each transducer velocity were
calculated. Pretreatment temperatures across the 3
conditions were analyzed using a 1-way repeated-
measures ANOVA to confirm that the tissue tempera-
ture returned to the initial pretreatment temperature
prior to beginning the next condition.
A 2 × 3 repeated-measures ANOVA was used to
detect velocity and time differences, and the Scheffe
multiple-comparison test was used for post hoc test-
ing. The Geisser-Greenhouse method was used to
adjust for the violation of covariance matrix circular-
ity. Significance level was set at P ⬍ .05 prior to
testing. The NCSS 2000 software (Number Cruncher
FIGURE 2. Thermocouple insertion is made parallel to the carpenter Statistical Systems, Kaysville, UT) was used for all data
square. summarization and analyses.
within the treatment site. Room temperature was
controlled by a wall-mounted thermostat. RESULTS
Each subject received all 3 velocity conditions (2-3,
4-5, and 7-8 cm/s). The transducer velocity was Descriptive statistics for each transducer velocity
controlled using a metronome. The distance the pretreatment and posttreatment are presented in
transducer traveled each beat was 2.4 cm (Figure 4). Table 1. The results of the 1-way ANOVA indicated
The metronome was set at 60, 88, and 192 beats per no statistically significant difference across the 3
minute for the 2- to 3-, 3- to 4-, and 7- to 8-cm/s pretreatment temperature values (F2,26 = 2.66, P =
conditions, respectively. For all conditions, ultrasound .10). Mean intramuscular temperature prior to each
(Omnisound 3000C, Accelerated Care Plus, Topeka, condition varied from 37.7°C to 37.9°C.
KS) was delivered with a 5-cm2-area transducer, with Results of the 2-way ANOVA indicated a significant
an effective radiating area of 5.0 cm2 and a BNR of main effect for time (F1,10 = 155.68, P ⬍ .00001), with
2.1:1 for 10 minutes at a frequency of 1 MHz, with a the posttreatment values (42.9°C ± 1.9°C) being
100% (continuous) duty cycle and intensity of 1.5 greater than pretreatment values (37.8°C ± 0.8°C).
W/cm2. After the initial condition, the subsequent No significant main effect for velocity (F2,20 = 0.07,
conditions did not start until the intramuscular tem- P = .93) and no significant interaction were identified
perature returned to within 0.3°C of the pretreat- (F2,20 = 0.33, P = .72).
ment temperature and remained unchanged A clinically relevant temperature change of 2°C
(± 0.1°C) for 5 minutes. Aquasonic Clear ultrasound (moderate heating), as the difference needed for
gel (Parker Laboratories, Inc, Fairfield, NJ), kept at influencing the efficacy of the modality, was se-
room temperature, was applied to the treatment area lected.15 Using Tukey-Kramer simultaneous confi-
to prevent direct contact of the transducer with the dence intervals (CIs) for multiple comparisons for all

322 J Orthop Sports Phys Ther • Volume 36 • Number 5 • May 2006


Lucite
form

Treatment
area

Center of
treatment area

2.4 cm Transducer
head

FIGURE 3. Ultrasound template. Line depicts distance (2.4 cm) that the transducer was moved each beat of the metronome.

pairs, the mean difference and 95% CIs for the ultrasound at lower velocities (2-4 cm/s) is that
difference in temperatures between the 4- to 5-cm/s ‘‘Rapid movement of the sound head causes the
velocity and the other velocities at the end of the therapist to slip into treating a larger area, thus the
study are reported in Table 2. Because the 95% CIs desired temperatures may not be attained.’’25 In fact,
on the differences from the 4- to 5-cm/s tempera- diminished thermal effects of ultrasound have been
tures did not include the clinically chosen 2°C, the reported using larger treatment areas.15 However, it is
lack of effect of transducer speed on temperature not known if the higher transducer velocity would
increases is not attributed to a lack of statistical power increase the treatment area if a template were not in
(type II error). place. Further research investigating the effect of
transducer velocity on treatment area is warranted.
DISCUSSION Another variable affecting the efficacy of thermal
Our data reveal that the heating effect was similar ultrasound is the difference among transducer energy
among conditions regardless of the transducer veloc- output. Differences in temperature change have been
ity. This result supports our hypothesis that there reported with different ultrasound transducers, de-
would be no differences in temperature increases spite the same treatment parameters. Two studies
among the velocity conditions. The increases in tissue directly comparing ultrasound manufacturers re-
temperature that we reported for transducer rates of ported difference in overall temperature increases
2 to 3 and 3 to 4 cm/s were similar to those reported and the rate of increase, despite similar BNR, be-
in other studies using similar param- tween the transducers.18,22 Differences in overall tem-
eters;3,9,10,12,13,26,28 therefore, we believe our data to perature increases are also evident when comparing
be accurate with the number of subjects tested and studies that used different ultrasound transducers,
the absence of a type II error. Gallo et al14 reported
RESEARCH
no differences in temperature increase between 2
groups with equivalent energy transfer into the body
for 3-MHz ultrasound treatments.
The amount of temperature increase has been used
to quantify the effect of an ultrasound treatment.10 In
this study, the average temperature increase of 5.1°C
would be considered vigorous heating (defined as a
temperature increase greater than 4°C) and be associ-
REPORT

ated with accelerating metabolic rate, increasing


blood flow, increasing tissue extensibility, decreasing
pain and muscle spasm, and inhibition of sympathetic
activity.10,29
We recognize that the parameters used in this study
were very well controlled and that there may be other
considerations when applying ultrasound clinically. FIGURE 4. Ultrasound performed using a template twice the size of
First, one common belief for administering the transducer head.

J Orthop Sports Phys Ther • Volume 36 • Number 5 • May 2006 323


2. Bishop S, Draper DO, Knight KL, Feland JB, Eggett D.
TABLE 1. Summary of temperature (mean ± SD °C) by trans- Human tissue-temperature rise during ultrasound treat-
ducer velocity and time (n = 11). ments with the aquaflex gel pad. J Athl Train.
Time* 2-3 cm/s 4-5 cm/s 7-8 cm/s 2004;39:126-131.
3. Burr PO, Demchak TJ, Cordova ML, Ingersoll CD, Stone
Pretreatment 37.9 ± 1.0 37.8 ± 0.8 37.7 ± 0.8 MB. Effects of altering intensity during a 10 minute 1
Posttreatment 42.7 ± 2.3 43.0 ± 2.0 43.1 ± 1.4 MHz ultrasound treatment on triceps surae temperature
elevation. J Sport Rehabil. 2004;13:275-286.
* Statistical analysis using a 2 × 3 repeated-measures ANOVA
indicate a main effect for time P ⬍ .00001. 4. Byl NN, McKenzie A, Wong T, West J, Hunt TK.
Incisional wound healing: a controlled study of low and
high dose ultrasound. J Orthop Sports Phys Ther.
1993;18:619-628.
TABLE 2. Summary of Tukey-Kramer simultaneous confidence 5. Chan AK, Myrer JW, Measom GJ, Draper DO. Tempera-
intervals for the multiple comparisons to 4 to 5 cm/s (°C) (n = ture Changes in Human Patellar Tendon in Response to
22, df = 20, SEM = 1.8, coefficient of variation [CV] = 3.6). Therapeutic Ultrasound. J Athl Train. 1998;33:130-135.
Group Mean 6. Denegar CR. Therapeutic Modalities for Athletic Inju-
Subtracted Difference Lower 95% Upper 95% ries. Champaign, IL: Human Kinetics; 2000:164.
7. Draper DO. Guidelines to enhance therapeutic
2-3 cm/s 0.1 –0.9 1.2 ultrasound treatment outcomes. Athl Ther Today.
7-8 cm/s 0.0 –1.0 1.0 1998;3:7-11.
8. Draper DO. Ten mistakes commonly made with
ultrasound use: current research sheds light on myths.
but the same treatment parameters.10,21,26 Another Athl Train Sports Health Care Perspect. 1996;2:95-107.
possibility is that there are differences in rate of 9. Draper DO, Castel JC, Castel D. Rate of temperature
temperature increase between ultrasound transducers increase in human muscle during 1 MHz and 3 MHz
continuous ultrasound. J Orthop Sports Phys Ther.
from the same manufacturer.12,28 Recently, a new 1995;22:142-150.
technique for analyzing the ultrasound beam profile 10. Draper DO, Harris ST, Schulthies S, Durrant E, Knight
(Schlieren analysis) has found difference in the KL, Ricard M. Hot-pack and 1-MHz ultrasound treat-
profiles of ultrasound transducers from the same ments have an additive effect on muscle temperature
manufacturer as well as between different manufac- increase. J Athl Train. 1998;33:21-24.
11. Draper DO, Prentice WE. Therapeutic Modalities for
turers.19 It is still unknown if the differences in beam Allied Health Professionals. New York, NY: McGraw-
profile translate into differences in rate or overall Hill Health Professions Division; 1998.
tissue temperature change. 12. Draper DO, Schulthies S, Sorvisto P, Hautala AM.
Temperature changes in deep muscles of humans during
ice and ultrasound therapies: an in vivo study. J Orthop
CONCLUSION Sports Phys Ther. 1995;21:153-157.
13. Draper DO, Sunderland S, Kirkendall DT, Ricard M. A
The results of this study demonstrate that when
comparison of temperature rise in human calf muscles
using the same ultrasound settings between treat- following applications of underwater and topical gel
ments for the same ultrasound machine, the velocity ultrasound. J Orthop Sports Phys Ther. 1993;17:247-
of the transducer will not affect the overall tempera- 251.
ture change. This should be verified by using differ- 14. Gallo JA, Draper DO, Brody LT, Fellingham GW. A
comparison of human muscle temperature increases
ent ultrasound machines from different
during 3-MHz continuous and pulsed ultrasound with
manufacturers. Clinically this means that if the appli- equivalent temporal average intensities. J Orthop Sports
cation of ultrasound remains within a set treatment Phys Ther. 2004;34:395-401.
area (twice the size of the transducer), it does not 15. Garrett CL, Draper DO, Knight KL. Heat distribution in
matter how fast the transducer is moved within the the lower leg from pulsed short-wave diathermy and
ultrasound treatments. J Athl Train. 2000;35:50-55.
treatment area.
16. Hayes BT, Merrick MA, Sandrey MA, Cordova ML.
Three-MHz ultrasound heats deeper into the tissues
AKNOWLEDGEMENTS than originally theorized. J Athl Train. 2004;39:230-
234.
This study was funded by the Indiana State Univer- 17. Holcomb WR. The effects of superficial heating before
sity Office of Sponsored Programs. We would also like 1-MHz ultrasound on tissue temperature. J Sport Rehab.
to thank Rudolph Medical Supplies for the use of the 2003;12:95-103.
18. Holcomb WR, Joyce CJ. A comparison of temperature
Omnisound 3000T Ultrasound Unit.
increases produced by 2 commonly used ultrasound
units. J Athl Train. 2003;38:24-27.
19. Johns LD, Straub SJ, LeDet EG. Ultrasound beam
profiling: comparative analysis of 4 new ultrasound
REFERENCES heads at both 1 and 3.3 MHz shows variability within a
manufacturer. J Athl Train. 2004;39:S-6.
1. Ashton DF, Draper DO, Myrer JW. Temperature Rise in 20. Lehmann JF, Stonebridge JB, deLateur BJ, Warren CG,
Human Muscle During Ultrasound Treatments Using Halar E. Temperatures in human thighs after hot pack
Flex-All as a Coupling Agent. J Athl Train. 1998;33:136- treatment followed by ultrasound. Arch Phys Med
140. Rehabil. 1978;59:472-475.

324 J Orthop Sports Phys Ther • Volume 36 • Number 5 • May 2006


21. Leonard J, Merrick MA, Ingersoll CD, Cordova ML. A 26. Rimington SJ, Draper DO, Durrant E, Fellingham G.
comparison of intramuscular temperatures during 10- Temperature changes during therapeutic ultrasound in
minute 1-MHz ultrasound treatments at different intensi- the precooled human gastrocnemius muscle. J Athl
ties. J Sport Rehab. 2004;13:244-254. Train. 1994;29:325-327.
22. Merrick MA, Bernard KD, Devor ST, Williams MJ. 27. Romani WA, Perrin DH, Dussault RG, Ball DW, Kahler
Identical 3-MHz ultrasound treatments with different DM. Identification of tibial stress fractures using thera-
devices produce different intramuscular temperatures. peutic continuous ultrasound. J Orthop Sports Phys
J Orthop Sports Phys Ther. 2003;33:379-385. Ther. 2000;30:444-452.
23. Michlovitz SL. Thermal Agents in Rehabilitation. Phila- 28. Rose S, Draper DO, Schulthies SS, Durrant E. The
delphia, PA: F.A. Davis Company; 1996:200.
stretching window part two: rate of thermal decay in
24. Myrer JW, Measom GJ, Fellingham GW. Intramuscular
deep muscle following 1-MHz ultrasound. J Athl Train.
temperature rises with topical analgesics used as cou-
pling agents during therapeutic ultrasound. J Athl Train. 1996;31:139-143.
2001;36:20-25. 29. Starkey C. Therapeutic Modalities. Philadelphia, PA:
25. Prentice W. Therapeutic Ultrasound. New York: F.A. Davis Company; 1999:269-304.
McGraw-Hill; 1998.

RESEARCH
REPORT

J Orthop Sports Phys Ther • Volume 36 • Number 5 • May 2006 325

You might also like