Professional Documents
Culture Documents
STANHOPE-SETA
THE QUEEN’S AWARDS FOR ENTERPRISE
INTERNATIONAL TRADE 2012
INNOVATION 2014
INNOVATION 2019
Regulatory conformance
This instrument conforms to the following directives and standards:
• 2006/95/EC – EC Low voltage directive.
• 2014/30/EU – Electromagnetic compatibility directive.
• 2011/65/EU – Restriction of the use of certain hazardous substances (RoHS) in electrical
and electronic equipment directive.
• BS EN 61010-1:2010 – Safety requirements for electrical equipment for measurement,
control, and laboratory use. General requirements.
• BS EN 61010-2-010:2014 – Safety requirements for electrical equipment for measurement,
control, and laboratory use. Particular requirements for laboratory equipment for the heating
of materials.
• BS EN 61326-1:2013 – Electrical equipment for measurement, control and laboratory use.
EMC requirements. General requirements.
This instrument is CE tested and marked.
Stanhope-Seta operates in conformance with waste electrical and electronic equipment (WEEE)
directives, (registration WEE/AD0054TQ) and will provide advice on disposal of Seta instruments on
request.
Quality assurance
All parts and sub-assemblies have been checked against test procedures and specifications before
final assembly.
This instrument has been fully tested, validated and calibrated by Stanhope-Seta in accordance with
an ISO9001 quality assurance system.
This instrument has been subjected to PAT (Portable Appliance Test) tests for electrical safety.
A quality assurance certificate is supplied with this instrument.
2
Revision: 2.0
21 November 2019
Publication: 34300H2B
Equipment identification
The model number and serial number of the Multiflash Universal Base Unit are marked on an identity
plate mounted on the back of the base unit. The power supply voltage and frequency are also marked
on the identity plate.
The model number and serial number of the Cleveland Open Cup Test Module are marked on an
identity plate on the back of the module.
The instrument displays the software version and date, and the circuit board serial number when you
start the unit. Quote this information in any technical query, or when ordering accessories and spare
parts.
Notation
This document uses the following notation:
Safety
Read this manual before installing or operating this instrument.
Do not install or operate this instrument unless you are trained to do so.
Always wear safety glasses, protective clothing and gloves when operating this
instrument.
Use this instrument only as described in this manual. The intrinsic protection of
the unit may be impaired if it is not used as described.
Comply with any applicable local and national health and safety regulations
when installing or operating this instrument.
Comply with any applicable local and national health and safety regulations
when storing, handling and disposing of samples.
This instrument is used to heat and test materials that may be flammable, toxic
or give off vapours and fumes. Use vapour extraction facilities and eliminate
ignition sources.
3
Revision: 2.0
21 November 2019
Publication: 34300H2B
Do not touch surfaces that become hot during operation. Hot parts may reach
temperatures in excess of 600°C. Gas and electric ignitors reach temperatures
far in excess of test temperatures.
This instrument includes gas and electric ignitors designed to ignite flammable
materials. Keep all such materials away from the instrument unless they are
being tested. Follow the guidelines below for Handling flammable liquids.
Do not modify the instrument in any way as this may result in injury or damage
to the equipment and will invalidate the warranty.
You must only carry out the repairs described in this handbook. A
Stanhope-Seta representative must carry out any other servicing and repairs.
Always isolate the electrical power supply before moving or maintaining the
unit. Failure to do so may result in death or serious injury.
Always disconnect the gas and compressed air supplies before moving the unit
or when carrying out certain maintenance tasks identified in this handbook.
Failure to do so may result in death or serious injury.
Take care when using this instrument. It has moving parts, which could trap
your fingers.
The instrument is suitable for indoor use only. Check that the environmental
conditions of the laboratory are within the limits given in section 2.2.1.
Use only the accessories and spares that are designed for use with the
instrument. Refer to section 9 for a list of compatible accessories and spares.
You are responsible for the safety of any external system to which the
instrument is connected. External systems may include but are not limited to
instrument air systems, water supplies, computer networks and laboratory
information systems (LIMS).
4
Revision: 2.0
21 November 2019
Publication: 34300H2B
5
Revision: 2.0
21 November 2019
Publication: 34300H2B
Contents
Regulatory conformance ......................................................................... 2
Quality assurance ................................................................................... 2
Scope of this manual .............................................................................. 2
Equipment identification .......................................................................... 3
Notation .................................................................................................. 3
Safety ..................................................................................................... 3
Handling flammable liquids ..................................................................... 5
Contents ................................................................................................. 6
1 Overview.......................................................................................... 10
1.1 Technical specification ................................................................................ 10
1.2 Conformance to test methods ..................................................................... 11
1.3 Product description ..................................................................................... 11
1.3.1 Introduction .....................................................................................................11
1.3.2 Instrument layout ............................................................................................12
1.3.3 Operating modes ............................................................................................17
1.3.4 Test types .......................................................................................................17
1.3.5 Safety devices ................................................................................................18
2 Installation ....................................................................................... 19
2.1 Unpacking ................................................................................................... 19
2.2 Location ...................................................................................................... 19
2.2.1 Operating conditions .......................................................................................19
2.3 Utilities ........................................................................................................ 20
2.3.1 Power supply ..................................................................................................20
2.3.2 Gas supply ......................................................................................................20
2.3.3 Compressed air supply ...................................................................................20
2.4 Initial set up ................................................................................................. 21
2.4.1 After transportation or storage ........................................................................21
2.4.2 Levelling the instrument ..................................................................................21
2.4.3 Mounting the optional draught screen .............................................................22
2.4.4 Connecting the cables ....................................................................................22
2.4.5 Connecting a printer or computer ....................................................................23
2.4.6 Setting the voltage ..........................................................................................25
2.4.7 Connecting the power supply ..........................................................................25
6
Revision: 2.0
21 November 2019
Publication: 34300H2B
7
Revision: 2.0
21 November 2019
Publication: 34300H2B
9
Revision: 2.0
21 November 2019
Publication: 34300H2B
1 Overview
1.1 Technical specification
Multiflash Cleveland Open Cup Flash Point Tester
Power consumption 1 kW
Current 9 A (max)
IP Rating IP11
Weight 22 kg
10
Revision: 2.0
21 November 2019
Publication: 34300H2B
11
Revision: 2.0
21 November 2019
Publication: 34300H2B
13
10 12 11 10
12
Revision: 2.0
21 November 2019
Publication: 34300H2B
1 2 3 4 5 6 7
10
11
17 16 15 14 13 12
13
Revision: 2.0
21 November 2019
Publication: 34300H2B
1 2
14
Revision: 2.0
21 November 2019
Publication: 34300H2B
15
Revision: 2.0
21 November 2019
Publication: 34300H2B
1 2
12
11
4
5
10
9 8
16
Revision: 2.0
21 November 2019
Publication: 34300H2B
1 Snuffer
17
Revision: 2.0
21 November 2019
Publication: 34300H2B
18
Revision: 2.0
21 November 2019
Publication: 34300H2B
2 Installation
2.1 Unpacking
• Check the condition of the packaging and photograph any damage.
• Check the Shock Watch, if fitted, and photograph if the indicator is red.
• Retain all packaging for future use in shipping or for long-term storage of the instrument.
• Check the instrument visually for damage, particularly if the packaging is damaged.
Photograph any areas of concern.
• Check that the operating voltage and frequency marked on the label on the rear of
instrument match the local power supply.
• Check the contents against the packing list enclosed.
• Contact Stanhope-Seta or our agents at the earliest opportunity to report any damage,
shortfall or problems with compatibility with the local power supply.
NOTE – The instrument may show signs of use. This is due to pre-delivery calibration
and testing.
2.2 Location
Locate the instrument:
• On a flat, non-flammable, level surface that can support the weight of the instrument and
any accessories.
• Where the fans in the side and rear of the unit are clear from obstruction.
• Where the mains plug and switch are easily accessible during use.
• In a draft-free, well-ventilated environment with fume extraction facilities.
• In a well-lit area with a stable temperature.
2.2.1 Operating conditions
The instrument is designed to operate in the following conditions:
19
Revision: 2.0
21 November 2019
Publication: 34300H2B
2.3 Utilities
2.3.1 Power supply
Only connect this instrument to a power supply with a safety earth (ground)
terminal.
Set the voltage selector switch to the correct voltage for your power supply
before switching on the instrument. Failure to do so may damage the
instrument and invalidate your warranty.
Only use the power cable supplied with the instrument. Use of any other power
cable could damage the instrument.
The Cleveland Flash Point Tester has a switched power supply and supports the following power
supplies:
• 110 – 120 Vac, 50/60 Hz
• 220 – 240 Vac, 50/60 Hz
The unit has a voltage selector switch located on the rear of the unit as shown in Figure 2 on page 13.
You must set the switch to the correct voltage for your power supply before switching on the unit.
The label on the rear of the instrument shows the unit’s operating voltage and frequency.
2.3.2 Gas supply
The gas supply pressure must not exceed 3 kPa gauge (0.44 psi gauge).
The instrument has both gas and electric hot wire ignitors. You must connect the instrument to an
external gas supply in order to use the gas ignitor.
The gas supply pressure must not exceed 3 kPa gauge (0.44 psi gauge). As the gas is only required
intermittently, the gas supply will be held against a closed valve for long periods.
2.3.3 Compressed air supply
The compressed air supply pressure must not exceed 670 kPa gauge
(100 psi gauge).
The compressed air system must be able to supply intermittent users, as the air
inlet valve will be closed for long periods.
The flash point tester is available with an optional cooling module that is usually factory fitted. If the
cooling module is fitted, you can connect a compressed air supply to cool the hotplate more quickly
between tests.
The compressed air supply pressure must not exceed 670 kPa gauge (100 psi gauge). As the
additional cooling is only required intermittently, the air supply will be held against a closed valve for
long periods.
20
Revision: 2.0
21 November 2019
Publication: 34300H2B
Leave the instrument to stand for two hours at room temperature before using
it, to allow any condensation to evaporate. Failure to do so may damage the
instrument.
TOOLS – You need a spirit level such as the one listed in section 9.1, to complete this
task.
1. Place the instrument in the location you have chosen for it.
21
Revision: 2.0
21 November 2019
Publication: 34300H2B
22
Revision: 2.0
21 November 2019
Publication: 34300H2B
23
Revision: 2.0
21 November 2019
Publication: 34300H2B
NOTE – You need an RS232 cable such as the one listed in section 9.1, to complete
this task.
24
Revision: 2.0
21 November 2019
Publication: 34300H2B
Set the voltage selector switch to the correct voltage for your power supply
before switching on the instrument. Failure to do so may damage the instrument
and invalidate your warranty.
Only connect this instrument to a power supply with a safety earth (ground)
terminal.
Only use the power cable supplied with the instrument. Use of any other power
cable could damage the instrument.
Set the voltage selector switch to the correct voltage for your power supply by
following the instructions in section 2.4.6 before switching on the instrument.
25
Revision: 2.0
21 November 2019
Publication: 34300H2B
NOTE – An electromagnet holds the snuffer in position when you retract it. You must
therefore switch on the instrument before retracting the snuffer.
Always switch off the instrument and isolate the electrical power supply before
installing the ignitor.
3. Check that the shape of the hot wire is as shown in Figure 7. If necessary, carefully bend
the hot wire into the correct shape.
27
Revision: 2.0
21 November 2019
Publication: 34300H2B
6. Adjust the position of the ignitor by following the instructions in section 2.4.10.1.
NOTE – You need to retract the snuffer before you can adjust the ignitor.
NOTE – You need the two sweep-arm gauges supplied with the instrument to complete
this task.
28
Revision: 2.0
21 November 2019
Publication: 34300H2B
NOTE – The gauges are different thicknesses and are marked with a part number and a
description.
Ignitor
Sweep-arm
gauge
Sample
cup
29
Revision: 2.0
21 November 2019
Publication: 34300H2B
NOTE – You need the probe-height gauge supplied with the instrument to complete this
task.
30
Revision: 2.0
21 November 2019
Publication: 34300H2B
31
Revision: 2.0
21 November 2019
Publication: 34300H2B
TOOLS – 13 mm open-ended
spanner.
32
Revision: 2.0
21 November 2019
Publication: 34300H2B
5.6. Repeat steps 4 and 5 until the safety PRT is in the correct position.
2.4.12 Installing the sample PRT and the flash and fire detector
To install the sample PRT and the flash and fire detector:
33
Revision: 2.0
21 November 2019
Publication: 34300H2B
34
Revision: 2.0
21 November 2019
Publication: 34300H2B
35
Revision: 2.0
21 November 2019
Publication: 34300H2B
36
Revision: 2.0
21 November 2019
Publication: 34300H2B
13. Adjust the height of the sample PRT and the flash and fire detector by following the
instructions in sections 2.4.12.1 and 2.4.12.2.
NOTE – You need the probe-height gauge supplied with the instrument to complete this
task.
4. Tighten the probe block handle to hold the sample PRT in position.
5. Leave the probe-height gauge beneath the PRT to keep it in place, and adjust the position
of the flash and fire detector by following the instructions in section 2.4.12.2.
37
Revision: 2.0
21 November 2019
Publication: 34300H2B
NOTE – You can leave the probe-height gauge beneath the sample PRT to keep it in
place while you adjust the flash and fire detector. For clarity, the images below do not
show the probe-height gauge.
NOTE – You need one of the sweep-arm gauges and the probe-height gauge supplied
with the instrument to complete this task.
1. Use the probe-height gauge to keep the sample PRT in place while you adjust the position
of the flash and fire detector. For clarity, the images below do not show the probe-height
gauge.
NOTE – For clarity, the images in this section do not show the probe-height gauge.
7. Tighten the probe block handle to hold the flash and fire detector in position.
38
Revision: 2.0
21 November 2019
Publication: 34300H2B
The gas supply pressure must not exceed 3 kPa gauge (0.44 psi gauge).
The gas system must be able to supply intermittent users, as the gas inlet valve
will be closed for long periods.
NOTE – You need the gas adaptor supplied with the instrument and a length of hose
with a 5 mm internal diameter (ID) to complete this task.
39
Revision: 2.0
21 November 2019
Publication: 34300H2B
2.4.14 Connecting the compressed air supply to the optional cooling module
If your instrument is fitted with the optional cooling module, you need to connect a compressed air
supply to the instrument before you can use the additional cooling.
To connect the compressed air supply:
The compressed air supply pressure must not exceed 670 kPa gauge
(100 psi gauge).
The compressed air system must be able to supply intermittent users, as the air
inlet valve will be closed for long periods.
40
Revision: 2.0
21 November 2019
Publication: 34300H2B
3 User interface
The Cleveland Multiflash Flash Point Tester has a menu-driven user interface. You can navigate
through the interface using the control dial, the edit button and the select button .
3.1 Navigation
Figure 8 shows the Home screen. Use the control dial to move the cursor up or down. The cursor
changes depending on the options available. If the cursor displays, press the button to select the
option to the right of the cursor. If the cursor displays, press and hold the button, while turning
the control dial to edit the parameter or value to the right or above the cursor.
41
Revision: 2.0
21 November 2019
Publication: 34300H2B
42
Revision: 2.0
21 November 2019
Publication: 34300H2B
43
Revision: 2.0
21 November 2019
Publication: 34300H2B
4 Operation
4.1 Preparing the sample
To prepare the sample:
Prepare the sample in accordance with the test method. Failure to do so will affect
the accuracy of the test.
44
Revision: 2.0
21 November 2019
Publication: 34300H2B
45
Revision: 2.0
21 November 2019
Publication: 34300H2B
46
Revision: 2.0
21 November 2019
Publication: 34300H2B
47
Revision: 2.0
21 November 2019
Publication: 34300H2B
48
Revision: 2.0
21 November 2019
Publication: 34300H2B
49
Revision: 2.0
21 November 2019
Publication: 34300H2B
NOTE – There is a 4 mm
circle engraved on each side
of the sample cup to help you
set the correct test flame size.
NOTE – If the test flame is too large or flares suddenly, the safety PRT may
trip the instrument. If this happens, close the gas control valve slightly, press
the reset button on the rear of the instrument, retract the snuffer and repeat
steps 4.2 to 4.7.
50
Revision: 2.0
21 November 2019
Publication: 34300H2B
51
Revision: 2.0
21 November 2019
Publication: 34300H2B
52
Revision: 2.0
21 November 2019
Publication: 34300H2B
53
Revision: 2.0
21 November 2019
Publication: 34300H2B
54
Revision: 2.0
21 November 2019
Publication: 34300H2B
55
Revision: 2.0
21 November 2019
Publication: 34300H2B
56
Revision: 2.0
21 November 2019
Publication: 34300H2B
57
Revision: 2.0
21 November 2019
Publication: 34300H2B
You must clean the sample cup, sample PRT, flash and fire detector, and probe
block before a test. Failure to do so will affect the accuracy of the test.
NOTE – You need a suitable solvent, a lint-free cloth and the bench stand supplied
with the instrument to complete this task.
58
Revision: 2.0
21 November 2019
Publication: 34300H2B
6. Clean and dry the sample cup in accordance with the test method.
NOTE – You will need to heat bitumen and very viscous samples before cleaning.
7. Wipe the sample PRT, flash and fire detector, and probe block with a suitable solvent to
remove sample residues.
8. Wait for the sample PRT, flash and fire detector, and probe block to dry completely before
loading a sample.
59
Revision: 2.0
21 November 2019
Publication: 34300H2B
NOTE – If testing asphalts, you need Hussami kit 34301-0 listed in section 9.1 to
complete this task.
1. Clean the sample cup and probes by following the instructions in section 4.2.4.
60
Revision: 2.0
21 November 2019
Publication: 34300H2B
4.7. Go to step 6.
61
Revision: 2.0
21 November 2019
Publication: 34300H2B
62
Revision: 2.0
21 November 2019
Publication: 34300H2B
63
Revision: 2.0
21 November 2019
Publication: 34300H2B
2. Set the terminal emulation software to data capture mode and enter a file name for the
exported data.
Take care when using this instrument as hot parts may reach temperatures in
excess of 600°C depending on the test temperature. Gas and electric ignitors
reach temperatures far in excess of test temperatures.
You must prepare the sample and instrument by following the steps in
sections 4.1 and 4.2. Failure to do so will affect the accuracy of the results.
64
Revision: 2.0
21 November 2019
Publication: 34300H2B
65
Revision: 2.0
21 November 2019
Publication: 34300H2B
66
Revision: 2.0
21 November 2019
Publication: 34300H2B
67
Revision: 2.0
21 November 2019
Publication: 34300H2B
68
Revision: 2.0
21 November 2019
Publication: 34300H2B
69
Revision: 2.0
21 November 2019
Publication: 34300H2B
70
Revision: 2.0
21 November 2019
Publication: 34300H2B
NOTE – You need to install a terminal emulator on the connected computer in order to
export test data to the computer. Refer to section 2.4.5.1 for more information.
2. Set the terminal emulation software to data capture mode and enter a file name for the
exported data.
71
Revision: 2.0
21 November 2019
Publication: 34300H2B
72
Revision: 2.0
21 November 2019
Publication: 34300H2B
73
Revision: 2.0
21 November 2019
Publication: 34300H2B
74
Revision: 2.0
21 November 2019
Publication: 34300H2B
75
Revision: 2.0
21 November 2019
Publication: 34300H2B
76
Revision: 2.0
21 November 2019
Publication: 34300H2B
Do not change configuration settings other than those described in this manual
unless instructed to do so by Stanhope-Seta.
77
Revision: 2.0
21 November 2019
Publication: 34300H2B
78
Revision: 2.0
21 November 2019
Publication: 34300H2B
79
Revision: 2.0
21 November 2019
Publication: 34300H2B
80
Revision: 2.0
21 November 2019
Publication: 34300H2B
81
Revision: 2.0
21 November 2019
Publication: 34300H2B
82
Revision: 2.0
21 November 2019
Publication: 34300H2B
Calibration Frequency
The barometer must be at the same elevation as the instrument. Using a reading
taken at a different elevation will affect the accuracy of the results.
Many aneroid barometers are pre-corrected to give readings at sea level. Do not
use this type of barometer to calibrate the instrument.
83
Revision: 2.0
21 November 2019
Publication: 34300H2B
The barometer must be at the same elevation as the instrument. Using a reading
taken at a different elevation will affect the accuracy of the results.
Many aneroid barometers are pre-corrected to give readings at sea level. Do not
use this type of barometer to calibrate the instrument.
84
Revision: 2.0
21 November 2019
Publication: 34300H2B
85
Revision: 2.0
21 November 2019
Publication: 34300H2B
86
Revision: 2.0
21 November 2019
Publication: 34300H2B
87
Revision: 2.0
21 November 2019
Publication: 34300H2B
Take care when calibrating this instrument as hot parts may reach temperatures in
excess of 55°C.
88
Revision: 2.0
21 November 2019
Publication: 34300H2B
89
Revision: 2.0
21 November 2019
Publication: 34300H2B
90
Revision: 2.0
21 November 2019
Publication: 34300H2B
91
Revision: 2.0
21 November 2019
Publication: 34300H2B
Do not touch the temperature probes or temperature well on the Fluke calibrator
as you may burn yourself.
18. Wait an additional 20 minutes for the calibrator and temperature probes to equilibrate.
92
Revision: 2.0
21 November 2019
Publication: 34300H2B
93
Revision: 2.0
21 November 2019
Publication: 34300H2B
94
Revision: 2.0
21 November 2019
Publication: 34300H2B
95
Revision: 2.0
21 November 2019
Publication: 34300H2B
30. Re-install the sample PRT into the probe block by following the instructions in
section 2.4.12.
31. Adjust the positions of the sample PRT, and flash and fire detector by following the
instructions in sections 2.4.12.1 and 2.4.12.2.
32. Verify the performance of the instrument by running a test using a certified reference
material such as one of those listed in section 9.1.1.
Take care when calibrating this instrument as hot parts may reach temperatures in
excess of 55°C.
2. Attach the sample PRT to the PRT extension cable suppled with the instrument.
96
Revision: 2.0
21 November 2019
Publication: 34300H2B
97
Revision: 2.0
21 November 2019
Publication: 34300H2B
8. Insert the sample PRT into the temperature-controlled bath. Make sure that that water
covers the probe up to the first bend.
9. Insert the probe of the digital thermometer into the water bath.
98
Revision: 2.0
21 November 2019
Publication: 34300H2B
15. Wait an additional 20 minutes for temperature probes and bath to equilibrate.
99
Revision: 2.0
21 November 2019
Publication: 34300H2B
100
Revision: 2.0
21 November 2019
Publication: 34300H2B
101
Revision: 2.0
21 November 2019
Publication: 34300H2B
27. Re-install the sample PRT into the probe block by following the instructions in
section 2.4.12.
102
Revision: 2.0
21 November 2019
Publication: 34300H2B
28. Adjust the positions of the sample PRT, and flash and fire detector by following the
instructions in sections 2.4.12.1 and 2.4.12.2.
29. Verify the performance of the instrument by running a test using a certified reference
material such as one of those listed in section 9.1.1.
5.2 Verification
Verify the performance of the Cleveland Flash Point Tester at least every 12 months using a certified
reference material (CRM) such as the ones listed in section 9.1.1. Refer to the test method for further
details.
Carry out more frequent performance tests using secondary working standards (SWS). Refer to the
test method for further details.
103
Revision: 2.0
21 November 2019
Publication: 34300H2B
6 Troubleshooting
Fault Cause Resolution
Display does not light No power to the Check that the power supply is
up when you switch on instrument connected, available and switched on.
the instrument.
Fuse has blown Check fuse and replace if necessary.
Refer to section 7.6.9 for details of how
to change the fuse.
Ignitor does not glow Ignitor short circuit due 1. Remove the ignitor by following the
to carbon deposits instructions in section 7.6.5.1.
between sweep arm
2. Clean the taper on the end of the
and ignitor
sweep arm and the inside of the
ignitor by following the instructions in
section 7.6.5.3.
3. Refit the ignitor by following the
instructions in section 7.6.5.4.
Gas flame does not Gas supply not 1. Connect the gas supply by following
light but ignitor glows connected or not the instructions in section 2.4.13.
switched on
2. Switch on the gas supply.
104
Revision: 2.0
21 November 2019
Publication: 34300H2B
Gas flame is unstable Gas ignitor is blocked Remove the ignitor, clean it and refit it
or is difficult to set at with carbon deposits by following the instructions in
4 mm section 7.6.5.
Sweep arm jams or Build-up of gum and Check the sweep arm movement and
does not move sample residues in the clean the guide slot by following the
smoothly guide slot instructions in sections 7.6.6.1 and
7.6.6.2.
Sweep arm height Build-up of gum and Clean the guide slot by following the
cannot be adjusted sample residues in the instructions in section 7.6.6.2.
guide slot
Air cooling does not Air supply is not Check that the air supply is connected,
operate at the end of a connected available and switched on.
test.
Faulty air valve Contact Stanhope-Seta.
Unexpected sample Faulty sample PRT 1. Check the sample PRT by following
temperature the instructions in section 7.6.1.1.
Replace the PRT if it is faulty by
following the instructions in sections
7.6.1.2 and 7.6.1.3.
2. If the problem persists, contact
Stanhope-Seta.
105
Revision: 2.0
21 November 2019
Publication: 34300H2B
Displayed sample Instrument is carrying Wait while the instrument completes the
temperature does not out pre-test checks pre-test checks.
change
Faulty sample PRT 1. Check the sample PRT by following
the instructions in section 7.6.1.1.
Replace the PRT if it is faulty by
following the instructions in sections
7.6.1.2 and 7.6.1.3.
2. If the problem persists, contact
Stanhope-Seta.
Samples consistently Ignitor requires Check the position of the ignitor relative
do not flash when adjustment to the sample pot by following the
expected but ignitor and instructions in section 2.4.10.
sweep arm are
operating correctly NOTE – The ignitor position
is different for gas and
electric ignition.
Faulty flash and fire 1. Check the flash and fire detector by
detector following the instructions in
section 7.6.2.1. Replace the flash
and fire detector if it is faulty by
following the instructions in
sections 7.6.2.2 and 7.6.2.3.
2. If the problem persists, contact
Stanhope-Seta.
Instrument constantly Faulty flash and fire 1. Check the flash and fire detector by
shows that sample has detector following the instructions in
flashed section 7.6.2.1. Replace the flash
and fire detector if it is faulty by
following the instructions in
sections 7.6.2.2 and 7.6.2.3.
2. If the problem persists, contact
Stanhope-Seta.
106
Revision: 2.0
21 November 2019
Publication: 34300H2B
Snuffer does not deploy Faulty safety PRT 1. Check the safety PRT by following
automatically when a the instructions in section 7.6.4.1.
sustained fire occurs Replace the PRT if it is faulty by
following the instructions in
sections 7.6.4.2 and 7.6.4.3.
2. If the problem persists, contact
Stanhope-Seta.
Snuffer does not stay in Instrument is switched Switch on the instrument and wait until
position when retracted off or is initialising the Home screen displays before
retracting the snuffer.
ARM ALREADY You tried to move the Move sweep arm to the right. See
LEFT sweep arm to the left section 7.6.6.1.
but it is already in this
position.
ARM ALREADY You tried to move the Move sweep arm to the left. See
RIGHT sweep arm to the section 7.6.6.1.
right but it is already
in this position.
ARM SNUFFER Snuffer is over the Retract the snuffer plate by following
PLATE sample cup the instructions in section 2.4.9.
FIT PLATE PRT Hotplate PRT is not Connect the hotplate PRT. See
connected section 2.4.4, step 2.
107
Revision: 2.0
21 November 2019
Publication: 34300H2B
PLATE PRT HHHH Hotplate PRT has a 1. Check the hotplate PRT by
short circuit or is following the instructions in
faulty section 7.6.3.1. Replace the PRT if
it is faulty by following the
PLATE PRT LLLL Hotplate PRT is open instructions in sections 7.6.3.2 and
circuit or faulty 7.6.3.3.
2. If the problem persists, contact
Stanhope-Seta.
PLATE TEMP TOO Hotplate temperature Wait for the hotplate to cool down to
HIGH is more than 100°C less than 100°C
above the sample
temperature
RESET OVER-TEMP Safety PRT has Press the reset button on the rear of
initiated a trip the instrument. See Figure 2 on
page 13.
SAMPLE PRT HHHH Sample PRT is short 1. Check the sample PRT by following
circuit or faulty the instructions in section 7.6.1.
Replace the sample PRT if it is
SAMPLE PRT LLLL Sample PRT is open faulty by following the instructions
circuit or faulty in sections 7.6.1.2 and 7.6.1.3.
2. If the problem persists, contact
Stanhope-Seta.
SAMPLE TEMP TOO Sample temperature Wait for the sample to cool down
HIGH is higher than the first before starting a test.
sweep temperature
002 SAMPLE Sample temperature Wait for the sample to cool down
EXCEEDED H is >600°C before starting a test.
108
Revision: 2.0
21 November 2019
Publication: 34300H2B
006 ERR ION OC Faulty flash and fire 1. Check the flash and fire
detector (open circuit) detector by following the
instructions in section 7.6.2.1.
Replace the detector if it is
faulty by following the
instructions in sections 7.6.2.2
and 7.6.2.3.
2. If the problem persists,
contact Stanhope-Seta.
010 ERR GLOW Hot-wire ignitor is not 1. Check the hot-wire ignitor by
OC fitted, is faulty or following the instructions in
requires cleaning section 7.6.5. Replace the
ignitor if it is faulty by following
011 ERR GLOW Hot-wire ignitor has a the instructions in section
SC short circuit 7.6.5.4.
2. If the problem persists,
contact Stanhope-Seta.
012 ERR SWEEP1 Sweep arm has not Check the sweep arm movement
TM reached limit and clean the guide slot by
switch 1. following the instructions in
section 7.6.6.
013 ERR SWEEP2 Sweep arm has not
TM reached limit
switch 2.
020 ERR SAFETY Safety PRT tripped 1. Clean the sample cup, probes
CUT due to an and the underside of the
over-temperature trip snuffer.
or fire
2. Press the reset button on the
rear of the instrument. See
Figure 2 on page 13.
109
Revision: 2.0
21 November 2019
Publication: 34300H2B
024 ERR PRE Sample flashed 1. Wait for the instrument to cool
TEST during the pre-test down.
sweep
2. Lower the expected flash
point and retest the sample.
026 ERR TM RMP Sample PRT is not Place the probe block on the
LMT submerged in the mounting bracket and check that
sample the sample PRT is in the correct
position by following the
instructions in section 2.4.12.1.
110
Revision: 2.0
21 November 2019
Publication: 34300H2B
7 Maintenance
You may invalidate your warranty if you do not follow the maintenance
procedures provided in this handbook.
111
Revision: 2.0
21 November 2019
Publication: 34300H2B
1. Check the power lead for signs of wear and replace if necessary.
2. Check the condition of the gas supply hose and replace if necessary.
3. Check the condition of the air supply hose and replace if necessary.
112
Revision: 2.0
21 November 2019
Publication: 34300H2B
• Pins 1 and 5
3 1
• Pins 1 and 3 8
• Pins 4 and 5
7 6
• Pins 4 and 3
4. If any of the measurements taken in steps 1 and 3, are significantly outside the given
limits, remove the existing sample PRT by following the instructions in section 7.6.1.2
and fit a new PRT by following the instructions in section 2.4.12.
113
Revision: 2.0
21 November 2019
Publication: 34300H2B
114
Revision: 2.0
21 November 2019
Publication: 34300H2B
115
Revision: 2.0
21 November 2019
Publication: 34300H2B
4. If any of the measurements taken in steps 1 to 3, are significantly outside the given
limits, remove the existing flash and fire detector by following the instructions in section
7.6.2.2 and fit a new detector by following the instructions in section 2.4.12.
116
Revision: 2.0
21 November 2019
Publication: 34300H2B
117
Revision: 2.0
21 November 2019
Publication: 34300H2B
check the PRT for faults and replace it if necessary by following the instructions in sections 7.6.3.2
and 7.6.3.3.
7.6.3.1 Checking the hotplate PRT
To check the hotplate PRT:
• Pins 1 and 5
3 1
• Pins 1 and 3 8
• Pins 4 and 5
7 6
• Pins 4 and 3
4. If any of the measurements taken in steps 1 and 3, are significantly outside the given
limits, remove the existing hotplate PRT by following the instructions in section 7.6.3.2
and fit a new PRT by following the instructions in section 7.6.3.3.
118
Revision: 2.0
21 November 2019
Publication: 34300H2B
Always switch off the instrument and isolate the electrical power supply before
removing the hotplate PRT. Failure to do so may result in death or serious
injury.
Always disconnect the gas supply before removing the sweep arm assembly.
Failure to do so may result in death or serious injury.
If your instrument is fitted with the optional cooling module, always disconnect
the compressed air supply to the instrument before removing the hotplate PRT.
Failure to do so may result in serious injury.
119
Revision: 2.0
21 November 2019
Publication: 34300H2B
120
Revision: 2.0
21 November 2019
Publication: 34300H2B
121
Revision: 2.0
21 November 2019
Publication: 34300H2B
122
Revision: 2.0
21 November 2019
Publication: 34300H2B
1. Remove the existing hotplate PRT by following the instructions in section 7.6.3.2.
123
Revision: 2.0
21 November 2019
Publication: 34300H2B
124
Revision: 2.0
21 November 2019
Publication: 34300H2B
7.3. Go to step 8.
125
Revision: 2.0
21 November 2019
Publication: 34300H2B
126
Revision: 2.0
21 November 2019
Publication: 34300H2B
16. Reconnect the gas supply by following the instructions in section 2.4.13.
127
Revision: 2.0
21 November 2019
Publication: 34300H2B
4. If any of the measurements taken in steps 1 to 3, are significantly outside the given
limits, remove the existing safety PRT by following the instructions in section 7.6.4.2 and
fit a new PRT by following the instructions in section 7.6.4.3.
128
Revision: 2.0
21 November 2019
Publication: 34300H2B
129
Revision: 2.0
21 November 2019
Publication: 34300H2B
130
Revision: 2.0
21 November 2019
Publication: 34300H2B
9. Adjust the position of the safety PRT by following the instructions in section 2.4.11.
You must use the gauge supplied to correctly position the safety PRT after
replacing it. The snuffer may not deploy to extinguish a fire if you fail to do so.
If the processes described in this section do not resolve the issue, you may need replace the sweep
arm assembly or adjust the ignitor parameters. Contact Stanhope-Seta for advice. Sections 7.6.6.3
and 7.6.6.4 explain how to replace the sweep arm assembly.
7.6.5.1 Removing the ignitor
To remove the ignitor:
Always allow the ignitor to cool down completely before removing it. You could
burn yourself if you fail to do so.
Always switch off the instrument and isolate the electrical power supply before
removing the ignitor.
Handle the ignitor with care. The hot-wire coil is fragile and easily damaged.
1. Wait for the ignitor to cool down completely before attempting to remove it.
131
Revision: 2.0
21 November 2019
Publication: 34300H2B
Gas jet
132
Revision: 2.0
21 November 2019
Publication: 34300H2B
133
Revision: 2.0
21 November 2019
Publication: 34300H2B
6. Refit the cleaned ignitor or fit a new one by following the instructions in section 7.6.5.4.
1. Fit the ignitor to the end of the sweep arm by following the instructions in
section 2.4.10.
134
Revision: 2.0
21 November 2019
Publication: 34300H2B
135
Revision: 2.0
21 November 2019
Publication: 34300H2B
7. If the sweep arm did not move smoothly or did not move at all, clean the guide slot by
following the instructions in section 7.6.6.2.
Always switch off the instrument and isolate the electrical power supply before
cleaning the guide slot. Failure to do so may result in death or serious injury.
Do not lubricate the guide slot or sweep arm mechanism. Doing so will increase
the rate of accumulation of dirt and deposits.
136
Revision: 2.0
21 November 2019
Publication: 34300H2B
NOTE – You need a spray can of contact cleaner fitted with an extension tube to
complete this task.
8. Wait for the contact cleaner to evaporate before switching on the instrument.
10. Check if the sweep arm moves freely by following the instructions in section 7.6.6.1.
137
Revision: 2.0
21 November 2019
Publication: 34300H2B
Always switch off the instrument and isolate the electrical power supply before
removing the sweep arm assembly. Failure to do so may result in death or serious
injury.
Always disconnect the gas supply before removing the sweep arm assembly.
Failure to do so may result in death or serious injury.
If your instrument is fitted with the optional cooling module, always disconnect
the compressed air supply to the instrument before removing the sweep arm
assembly. Failure to do so may result in serious injury.
Do not lubricate the guide slot or sweep arm mechanism. Doing so will increase
the rate of accumulation of dirt and deposits.
138
Revision: 2.0
21 November 2019
Publication: 34300H2B
6. Remove the ignitor by following the instructions in section 7.6.5.1. Store it somewhere
where it will not be damaged.
139
Revision: 2.0
21 November 2019
Publication: 34300H2B
140
Revision: 2.0
21 November 2019
Publication: 34300H2B
141
Revision: 2.0
21 November 2019
Publication: 34300H2B
23. Fit a new sweep arm by following the instructions in section 7.6.6.4.
1. Remove the existing sweep arm assembly by following the instructions in section 7.6.6.3.
142
Revision: 2.0
21 November 2019
Publication: 34300H2B
6. Fit and adjust the ignitor by following the instructions in section 2.4.10.
143
Revision: 2.0
21 November 2019
Publication: 34300H2B
TOOLS – 8 mm spanner.
144
Revision: 2.0
21 November 2019
Publication: 34300H2B
The following steps require you to switch on the instrument with the enclosure
cover removed. Do not touch any wires or electrical connections inside the
sweep-arm enclosure. Doing so may result in death or serious injury.
17. Check that the sweep arm moves correctly by following the instructions in
section 7.6.6.1.
145
Revision: 2.0
21 November 2019
Publication: 34300H2B
24. Reconnect the gas supply by following the instructions in section 2.4.13.
146
Revision: 2.0
21 November 2019
Publication: 34300H2B
Vent the air trapped in the air supply hose before disconnecting the hose. Failure
to do so may result in injury.
147
Revision: 2.0
21 November 2019
Publication: 34300H2B
1. Switch off the gas supply to the instrument before disconnecting it.
NOTE – As the gas supply pressure is very low, you do not need to vent the contents of
the supply hose before disconnecting it.
Always switch off the instrument and isolate the electrical power supply before
changing the fuse. Failure to do so may result in death or serious injury.
148
Revision: 2.0
21 November 2019
Publication: 34300H2B
6. Fit a new fuse into the fuse holder. Refer to section 1.1 for the complete fuse
specification.
149
Revision: 2.0
21 November 2019
Publication: 34300H2B
150
Revision: 2.0
21 November 2019
Publication: 34300H2B
151
Revision: 2.0
21 November 2019
Publication: 34300H2B
99880-0 Flashcheck
Secondary working standard for verifying instrument
performance in accordance with ASTM D92. Nominal flash
point of 86.8°C. Certified to ISO 17025 by a UKAS accredited
laboratory. Suppled with a works certificate.
Pack of 3 × 80-ml bottles.
9.2 Spares
Part No. Thumbnail Description
34100-004 Ignitor
Coaxial gas/electric ignitor.
152
Revision: 2.0
21 November 2019
Publication: 34300H2B
153
Revision: 2.0
21 November 2019
Publication: 34300H2B
154
Revision: 2.0
21 November 2019