You are on page 1of 5

User’s Manual/BodyPart parameter file converter

Other commands (described in Keyword.dat)


Settings
Entry Name Entry description Default Note
[Data format]
PartPramName Display name of the imaging method button [String]

TrayPosS Display position of the tray type button [uint] (1-16) 0:not displayed

TrayPosT Display position of the category type button [uint] (1-28) 0: setting prohibited.
[UINT]
Described by 3 digits
ID number of the imaging method key made up from the Body Position of the imaging Editing prohibited because the system sets
ProcedurelID
Used at automatic pressing. part tab No.2 digit + method button it automatically.
imaging method position
2 digits
ID number for communicating with The ID usually contains 10 alphanumerical
RisProtclID [char]
CCRHIS.DLL characters
The ID usually contains 10 alphanumerical
ID number for communicating with
GenProtclID [char] characters
CCRGEN.DLL
Multiple entries are possible.
XGeneratorID The ID of the X-ray generator. [uint] 1 Editing prohibited.

English Edition Rev.03 Aug, 2004 -1- All Rights Reserved.


(C)Canon Inc.
User’s Manual/BodyPart parameter file converter

Settings
Entry Name Entry description Default Note
[Data format]
X-ray tube gearing parameters
Use the system settings from the left (1) or not
use (0). If not using, the button-specific
parameters listed after this parameter will become
valid.
X-ray channel to use
Aperture channel
Aperture parameter ID
Aperture enable (1), disable (0) Default generated based on Ignored in the sample.
XTParam [String]
Aperture width (mm) the detected sensor type. New in V4.0
Aperture height (mm)
The ID of the offset specifying the vertical
difference between the aperture-center and the
sensor-center.
Move the aperture-center (1) based on the offset,
or not to move it (0)
Offset quantity (mm)
Example: XTParam=1,1,1,0,0,0,0,0,0,0
[unsigned char]
0: not set
gcFocalSize X-ray tube focus size 1: small
2: large
3: medium
[unsigned char]
gcBodysize Physique of the patient
0: not set 5-11
[unsigned char]
gcAEC AEC enable/disable
0: disable 1: enable
[unsigned char]
0: not used
gcPtLeft Left photo timer when gcAEC=1
1:Off
2:On
[unsigned char]
0: not used
gcPtCenter Central photo timer when gcAEC=1
1:Off
2:On

English Edition Rev.03 Aug, 2004 -2- All Rights Reserved.


(C)Canon Inc.
User’s Manual/BodyPart parameter file converter

Settings
Entry Name Entry description Default Note
[Data format]
[unsigned char]
0: not used
gcPtRight Right photo timer when gcAEC=1
1:Off
2:On
[unsigned char]
gcDensity Film density when gcAEC=1 0: meaningless
5 - 11:-3 - +3
Grid parameters
The string specifies bUseSystem, uiGridMode,
bWithDesiredGrid, uiRecommendGridID
separated by commas.
BUseSystem: Use the system value?
1: use
0: don’t use
uiGridMode: grid mode standing, table: 1,0,1,ffff
0: moving grid mode bucky:
1: fixed grid mode 1,0,0,0
GridParam [String]
2: no grid cassette
3: All possible mode (discontinued as of 1,2,0,0
12/07/2000)
4: forced mode
bWithDesiredGrid: Display warning?
0: don’t display
1: display
uiRecommendGridID: recommended grid ID
(hexadecimal display)
Example: 1,0,1,ffff

BodyPartCode DICOM CR body part code [uint]

ViewPosCode DICOM CR view position code [uint]


[ValueTableInfo]
kV [double]
X-ray tube voltage when kV_TableEnable=0

English Edition Rev.03 Aug, 2004 -3- All Rights Reserved.


(C)Canon Inc.
User’s Manual/BodyPart parameter file converter

Settings
Entry Name Entry description Default Note
[Data format]
X-ray tube current when
mA [double]
[ValueTableInfo]mA_TableEnable=0
cm Imaging distance [double]
Irradiation time when Can be set to be automatically specified
msec [double]
[ValueTableInfo]msec_TableEnable=0 based on the mAs value
mAs value when [ValueTableInfo] Can be set to be automatically specified
mAs [double]
mAs_TableEnable=0 based on the msec value
Bucky msec
BuckyMSec [double] Editing prohibited in the case of fixed grid
(Only valid in the case of moving grid)
Memo Series description for the button [String] No
ptr_ori Editing prohibited. 8
Rex value check during imaging
Setting the REX value checking during imaging
From the left
Use the system settings (1), don't use (0)
RexInfo 1,0,50,0,3000
Lower limit check enable (1), disable (0)
Lower limit
Upper limit check enable (1), disable (0)
Upper limit
1: Display
DisplayLaterality Annotation setting for laterality
0: Don’t display
Laterality Laterality [uint]
1: Display
DisplayMark Annotation setting for mark
0: Don’t display
MarkID Mark ID [String]
LinkOrient Flag for linking the mark and the orientation
vpt_total Number of fixed ROIs (up to 5) [uint]
1: use
vpt_UseSysInfo Use system fixed ROI
0: don’t use
843,518,188,312,0,1
[uint]:Top,
Virtual phototimer (fixed ROI) when 1656,518,188,312,0,1
vpt_info%1 %1 [uint]:Left, 1: enable / 0: disable
vpt_UseSysInfo=0 1250,1188,188,312,0,1
ROI number (0-4) [uint]:Width, 1: display / 0: don’t display
0,0,0,0,0,0
[uint]:Height,
0,0,0,0,0,0

English Edition Rev.03 Aug, 2004 -4- All Rights Reserved.


(C)Canon Inc.
User’s Manual/BodyPart parameter file converter

Settings
Entry Name Entry description Default Note
[Data format]
vpt_sys_info%1 %1 Usage condition of each phototimer when
ROI number (0-4) vpt_UseSysInfo=1
X-ray monitor ON/OFF 1: enable
XrayMonitor 1
(Display warning / don’t display warning) 0: disable
IpString Image processing parameters See the attachment
Not yet used.
IpMiscString See the attachment
(Second image processing parameters planned.)
Use of annotation system settings according to 1: use
AnnoUseSystemset 1
body parts 0: don’t use
1: use
AnnoEnable%1 Usage condition of annotation 0
0: don’t use
AnnoCtrl%1 Position, text alignment, font of the annotation [key]=[data];
PsFilmSizeId ID specifying the film size
PsPrintId Printer ID
PsPrinterName Printer nickname
PsOrientation Film orientation
Using or not using the system settings for
PsUseSysInfo
handling large irradiation fields.
1: use
PsMultiFormatReq Multi-format output condition
0: don’t use
PsMultiFormatLR Left-right division number [uint]
PsMultiFormatTB Top-bottom division number [uint]
PsMargin Margin ratio
PsHandlingBit Handling setting
Fixed reduction mode Life-size print mode
PsPrintMode Print mode
(MINIFY_MODE) (LIFESIZE_MODE)
PsMnifyReq Fixed reduced print reduction ratio int

English Edition Rev.03 Aug, 2004 -5- All Rights Reserved.


(C)Canon Inc.

You might also like