4109 9900 31.

00
2021/06/21

BMO PSA Medical Oxygen


Page 1 of 13

Concentrator Plant
BMO BeaconMedaes PSA Medical Oxygen Concentrator Plant

SPECIFICATION
Overview
The BMO series “all-in-one” Oxygen Concentrator Systems uses
Pressure Swing Adsorption technology (PSA) and is designed to
provide Pharmacopeia quality medical Oxygen with purity 93±3%
for hospitals.

Picture of BMO All-In-One Oxygen Concentrator

Main differentiating features of the system are listed in table below:

Feature Benefit
• Compact – small space needed in the plant room
• Less pressure loss thanks to shorter and optimized piping
All-in-one concept – compressed air supply, air dryer, • Fast installation
filtration, oxygen concentrator and analyzer in one
• Assembled and tested as complete plant at the factory
compact platform
• Reduced transport cost; Airfreight worthy
• Single point power and oxygen connection
• Fast ramp-up to nominal purity
Proprietary adsorbent sieve mix and cycle management • Resilience and high performance at high ambient temperature
• Long lifetime
• Energy efficiency
Variable speed drive compressor with permanent magnet • Stable pressure and oxygen purity
drive technology • Low in-rush current, possibility to power from power generator
• Compact and light weight
• Ease of operation, no skilled personnel required
Central control with 10-inch display with one-button
• Intelligent system manages plant as a whole
operation
• Data trending and alarm log
• Fast start-up (under 30 min first time commissioning)
Short ramp up to nominal purity
• Quick back online after power failure or service intervention
Connectivity • Remote monitoring and troubleshooting capability
Smart control module (optional) • Intelligent Primary, Secondary and reserve source management
Low noise level • Less disturbance to hospital personnel and patients
• Class II Medical device certified by Australian TGA.
Certified Medical device
• ISO13485 for manufacturing and sales of medical devices

BeaconMedæs • Telford Crescent, Staveley, Derbyshire S43 3PF • Phone +44 (0) 1246 474 242 • www.beaconmedaes.com
4109 9900 31.00
2021/06/21
Page 2 of 13

Introduction Startup time


The BMO series (BeaconMedaes Oxygen) “all-in-one” oxygen Time needed for the concentrator system to reach nominal purity
concentrator systems is designed to provide Pharmacopeia quality shall not exceed 30 minutes during the commissioning. After short
medical Oxygen with purity 93±3%. The concentrator uses pressure service interruption (for example for swapping power generators
swing adsorption (PSA) technology. for compressor maintenance), time to reach the nominal purity
shall be within 10 minutes.
The system shall be designed as “plug-and-play” installation and
“one-button operation”. The single concentrator control unit shall
control the complete system including air compressor and dryer, Noise level
so that it becomes single point of human-machine-interface (HMI)
and no interaction with compressor nor dryer controller is required Noise level of the system shall not exceed 85 dB(a) at 1 meter.
by hospital personnel.
The system shall include air compressor with efficiency rating Oxygen filter
IE4 or higher and turn-down capability down to 25% of nominal The system shall be equipped with oxygen approved 0.01-micron
capacity for efficient oxygen supply in times of low hospital outlet filter, conforming to class 1 according to ISO8573-1:2010.
occupancy without causing frequent load/unload or start-stop of
the compressor.
Variable speed drive air compressor shall guarantee the narrow air Remote monitoring
supply pressure band that will result in stable oxygen purity and Compressor (as the equipment that have the most wear) shall
output. include the communication module that enables real time remote
monitoring of the system. The comprehensive remote monitoring
The dryer shall be capable of producing 3-degree Pressure Dew
of the plant shall be available via Smart Control Module.
Point (PDP) for adequate quality feeding air and shall be controlled
by the concentrator’s central control unit. The filtration and dryer
shall guarantee quality of the air according to ISO 8573-1 Class 1.
Data Logging functionality
The central controller shall log air pressure, oxygen purity, oxygen
Footprint requirements pressure, alarm log as well as “patrol” check (logging at the
controller during personnel checks on the unit).
The Oxygen concentrator system shall incorporate the air supply
preparation and Oxygen concentrator in one module to make
installation as compact as possible and should not require footprint
of more than 2.7 m2 for system with flow 150-500 lpm or 1.5 m2 for Alarm
systems below 150 lpm (including air supply preparation). The system shall include the following alarms on the central
controller: High Air pressure, High Oxygen Pressure, Low Oxygen
Purity. Alarms from the compressors controller shall be transmitted
Power requirements via remote monitoring systems and shall be visible from remote
location. Low purity output shall be provided for the master alarm.
The system shall have in-rush current not more than 15% of
nominal, so that it wouldn’t require contactors oversizing and It
could be used in regional hospitals with limited power availability
or be powered by electric generators. Ensuring continuity of supply
According to ISO 7396-1 continuity of the oxygen supply shall
The concentrator shall have single point electrical connection to the be ensured with three sources of supply, which can be various
concentrator system which distributes power to air compressors, combination of the one or two concentrators, reserve manifold and
dryer and Oxygen concentrator. local cylinders. Oxygen concentrators as primary and secondary
sources of supply shall be preferred, especially for large plants,
where cylinders will only provide oxygen for limited number of
Transportability hours.
The system shall be transportable by either sea, road or airfreight.
In cases where dual oxygen production lines are installed, the
The height of the unit shall not exceed 1.85 meter. Weight of the unit
system can be equipped with the Smart Control Module. Smart
shall not exceed 1450 kg for 500 lpm system, 1100 kg for systems
Control Module shall manage multiple units, automatically switch
with flow under 300 lpm and 550 kg for systems under 130 lpm
concentrators in case of failure and rotate unit for equal wear and
(including the complete air supply part).
tear.

BeaconMedæs • Telford Crescent, Staveley, Derbyshire S43 3PF • Phone +44 (0) 1246 474 242 • www.beaconmedaes.com
4109 9900 31.00
2021/06/21
Page 3 of 13

Oxygen Standards Oxygen Concentrator Plant Sizing


PSA oxygen modules are able to produce oxygen compliant with HTM 02-01 Part A – Appendix M gives indication of the average
the United States Pharmacopeia (USP) or European Pharmacopeia oxygen consumption per bed (+/- 1.5 - 2 l/min per bed).
(EurPh) monograph for Oxygen 93. Both standards were created
explicitly to permit the use of PSA produced oxygen for hospitals. In surgery rooms (operating theatres) consumption can peak to 10
lpm per bed. For COVID hospital, average diversified consumption
has been typically assumed 10 lpm per bed.
The concentrator sizing for the specific facility remains the
Standard responsibility of the hospital personnel.

Component EurPh Oxygen 93 USP Oxygen 93

Oxygen v/v% 93+/-3% 90 to 96%

Carbon
5 ppm 5 ppm
Monoxide
Carbon
300 ppm 300 ppm
Dioxide
Water 67 ppm Not Specified

BMO plants technical data

BMO Model

Parameter Unit 150V 300V 600V 700V 900V 1000V 2500V 3500V

Oxygen Flow* m³/h 6 10 15 18 25 30 45 55


Oxygen Flow* lpm 100 166 250 300 416 500 750 916
Oxygen purity % 93+/-3%
Oxygen output
MPa (bar) 0.35~0.43 (3.5~4.3)
pressure
Power supply 3 phase 400V±10%, 50 Hz (other voltages on request)
Compressor power kW 7.5 15 18 22 30 37 55 75
Compressor type Oil injected variable speed with permanent magnet motor drive technology
Compressor Model GA7 GA15 GA18 GA22 GA30 GA37 GA55 GA75
Noise level dB(a)@1m <85
Length, mm 1620 2150 2400 3700
Dimensions, excl.
Width, mm 825 1120 1120 1500
oxygen buffer kit
Height, mm 1680 1820 1820 2200
Weight of Oxygen
kg 520 650 980 1100 1400 1450 3300 3500
buffer kit
Dimensions of the Diam./Height Ф600 /
Ф800 / 2485 Ф1000 / 3175
oxygen buffer tank. mm 2185
Weight of Oxygen
kg 200 380 380 380 380 380 800 800
buffer tank
Recommended
oxygen buffer m3 0.6 1 1 1 1 1 2 2
capacity**
Fuses required A 25 40 50 80 80 100 175 200

BeaconMedæs • Telford Crescent, Staveley, Derbyshire S43 3PF • Phone +44 (0) 1246 474 242 • www.beaconmedaes.com
4109 9900 31.00
2021/06/21
Page 4 of 13

BMO Model

Parameter Unit 150V 300 V 600 V 700 V 900 V 1000 V 2500 V 3500 V

Compressed oil
L 8 8 15 15 16 16 32 32
capacity
Oxygen Outlet
G3/8 G1/2 G3/4
connection size
Pressure Vessels
GB-150
certification
External Air buffer Integrated buffer, external air storage is not required
Minimum
Compressor air 22 % of nominal turn down without compressor unload
delivery

Startup time First use or after a long-term shut down requires 20~30 minutes

Oxygen analyzer Ultrasonic sensor, integrated, tolerance +/-0.5%

Operating
ºC 5 ºC ~40 ºC
temperature
Transport and
ºC -40 ºC ~+55 ºC
Storage temperature

Atmospheric
KPa 80~106 KPa
pressure range

Controller Central controller with 10” touch screen

Displayed value Air Pressure, Oxygen Pressure, Oxygen Purity

On-screen alarms High Air pressure, High Oxygen Pressure, Low Oxygen Purity

Data Trending All measured parameter stored and numerical and graphical format. Alarm log

Oxygen filter 0.01-micron oxygen approved, Maximum particle size class 1 acc to ISO8573-1:2010

Air filtration class ISO 8573-1 Class 1

Flow meter Optional external

Number of power
One power connection to the unit
leads to the plant
Number of
Pneumatic One connection for oxygen outlet to the buffer tank
connections

BeaconMedæs • Telford Crescent, Staveley, Derbyshire S43 3PF • Phone +44 (0) 1246 474 242 • www.beaconmedaes.com
4109 9900 31.00
2021/06/21
Page 5 of 13

BMO Model

Parameter Unit 150V 300V 600V 700V 900V 1000V 2500V 3500V

Primary/secondary
source management Optional Smart Control to manage Primary secondary sources of supply, back up, hours
and hour equalization and remote monitoring
equalization
Remote monitoring
Standard for compressor, Smart control required for plant monitoring
possibility

Medical device Class II registered in Australian Therapeutic Goods Administration,


Medical certification
ISO13485, China NMPA

* Output at 93% concentration, 20°C ambient. Correction factors apply. See Flow Factoring Tables for more details.
** If booster is included oxygen buffer capacity should be increase, see booster specification table below.

Flow Factoring at Different Temperatures and altitude Optional booster and filling ramp
Although the optimal working temperature for PSA type systems is Booster and cylinder filling ramp can be added to the scope of
20°C, the BMO can perform at wide range of operating condition. supply for two main uses in hospital context:
For extreme conditions the flow derating factors should be applied
depending on the altitude and ambient temperature as per table a. For hospitals without medical gas pipeline system, BMO
below. (Example: BMO- 1000V can deliver 500 lpm at normal plant with the booster can be used for filling cylinder which
condition. If system is used at 38°C and altitude 1500 meters, the will be moved next to a patient to be used local source
derating factor will be 0.84 and the minimum guaranteed flow will of oxygen supply. In this case the Oxygen concentrator
be 0.84 * 500 lpm = 420 lpm). plant serves as high pressure cylinder filling plant. Booster
maximum flow shall be sized close to nominal capacity of
the BMO (70-85%).
b. In a hospital where the medical gas pipeline is present,
Temperature, ºC booster can be utilized to fill cylinders that will be used as
secondary or reserve source of oxygen supply. In this case
Below 30 ºC - 40 ºC 40 ºC - 45ºC the BMO will be connected to medical gas pipeline system.
30ºC
Booster shall normally be sized to 10-30% of the BMO
Below nominal capacity, so that only spare capacity is used for
1.00 1.00 0.83
1000 booster (available apart from supply into pipeline).
1000 0.89 0.89 0.83 Note: Filling oxygen cylinders is dangerous. Cylinders shall be handled with extreme
Meters, care following all the safety precautions and local regulations.
1500 0.84 0.84 0.83
above
sea level 2000 0.78 0.78 0.78
2500 0.73 0.73 0.73
>2500 Sizing on request

BeaconMedæs • Telford Crescent, Staveley, Derbyshire S43 3PF • Phone +44 (0) 1246 474 242 • www.beaconmedaes.com
4109 9900 31.00
2021/06/21
Page 6 of 13

Technical Specification

Booster model

Parameter Unit 3/4-150 8/4-150 12/4-150 15/4-150 20/4- 25/4-150 40/4-150 50/4-
150 150
Oxygen Flow1) m³/h 3 8 12 15 20 25 40 50
Oxygen Flow 1)
lpm 50 133 200 250 333 417 667 833
Suction pressure MPa (barg) 0.3~0.4 (3~4)
Maximum
MPa (barg) 15 (150)
Discharge pressure
Number of 47l
cylinders (6.8m3) 10 28 42 52 70 88 141 176
per 24 hrs shift1)

Motor power kW 1.5 3 5.5 7.5 11 11 15 15

Power supply V@Hz 3 phase, 400V@50Hz

Weight Kg 150 310 320 450 780 960 1000 1050

Length, mm 700 1000 1050 1500

Dimensions Width, mm 650 750 750 950

Height, mm 650 1150 1020 1500

Recommended
oxygen buffer tank m3 1 2 2 3 3 3 2x3 2x3
Capacity2)
Smallest
compatible BMO 150V 300V 600V 700V 900V 1000V 2500V 3500V
model3)
Maximum inlet air
ºC 45 45 45 45 45 45 45 45
temperature

Maximum ambient
ºC 45 45 45 45 45 45 45 45
temperature

Type Oil-free Oxygen booster

Duty cycle 24/7

Cooling type Air cooled

1)
20°C ambient, 20°C inlet air, 4 bar inlet pressure.
when booster is used with PSA plants, it is recommended to oversize the oxygen buffer tank size (compared to direct supply to the
2)

hospital).
3)
the booster shall always have less capacity than the nominal capacity of the concentrator.

BeaconMedæs • Telford Crescent, Staveley, Derbyshire S43 3PF • Phone +44 (0) 1246 474 242 • www.beaconmedaes.com
4109 9900 31.00
2021/06/21
Page 7 of 13

Continuity of supply
The supply system should be designed to achieve continuity of Types, capacities and locations of primary, secondary and
supply to the terminal units in normal condition and in a single reserve sources of supply will be based on both system design
fault condition. A single fault condition is where a single means for parameters and the need for supply security, identified by a risk
protection against a safety hazard in equipment is defective or a assessment during the planning stage.
single external abnormal condition is present. Loss of supply due
to maintenance of a supply source (or a component within it) is not primary, secondary and reserve sources of supply will be based
considered a single fault condition. on both system design parameters and the need for supply
security, identified by a risk assessment during the planning
stage.

Optional Emergency reserve manifold


Automatic changeover manifold (left) or Emergency Reserve
Manifold (right) HTM02-01 to be used as secondary or reserve
source of supply. Both manifolds can come with different number
of cylinders, it is recommended to have at least 4 hours reserve
supply. Please see separate technical data sheets.

BeaconMedæs • Telford Crescent, Staveley, Derbyshire S43 3PF • Phone +44 (0) 1246 474 242 • www.beaconmedaes.com
4109 9900 31.00
2021/06/21
Page 8 of 13

Installation recommendations
Three sources of supply are recommended to ensure continuity Oxygen tanks (3) used as a buffer after the concentrator to store
of supply. The recommended installation is shown on the picture oxygen. Emergency Reserve Manifold (ERM) (6) used only as a
below. Depending on the risk assessment by the consultant or reserve (3rd source of supply). Cylinders can be supplied by a
the hospital personnel and depending on the equipment already gas company or refilled from the booster. An alternative reserve
present, some of this equipment might be excluded from the source can be implemented via individual cylinders located by
scope of supply. This is an example of BMO-1000V with 25 m3/h the patient bed in the ward areas.
booster (416 l/min), filling ramp and emergency manifold in a plant
room 6 x 6 meters and cylinder room 2 x 6 meters. It is a safety Oxygen booster compressor (7) and filling ramp (8), if included,
recommendation - high-pressure cylinder installations should not is used to increase oxygen pressure from 0.4 MPa to 15 MPa (4
be located inside the compressor room. bar to 150 bar) and to refill the cylinders. Cylinders can be used
for the reserve manifold (self back-up) or to be sent out to other
To follow the Continuity of Supply Principle, two BMO systems facilities.
should be used to supply the hospital: BMO -1000V (1- primary)
and BMO-1000V (2 - secondary). Smart Control Module (5) is used to manage up to three BMO’s, it
monitors the oxygen flow, equalizes running hours, and provides
information about the system via connectivity system.
Digital flow meter (4), - Only used if there is no Smart Control
Module installed. The oxygen filter kit is used to filter the oxygen
from particles that might occur.

Emergency
manifold and 6
cylinder ramp
Connection
to the hospital 9
8 Cylinder fillling pipeline
ramp, 20 cylinder
Flow meter
and filtration 4

BOM - 1000V
Secondary 2
5 Smart Control
module

3 Oxygen buffer tank

BOM - 1000V
Primary 1

7 Oxygen booster

BeaconMedæs • Telford Crescent, Staveley, Derbyshire S43 3PF • Phone +44 (0) 1246 474 242 • www.beaconmedaes.com
4109 9900 31.00
2021/06/21
Page 9 of 13

Appendix. I Reference part numbers

BMO Model

Models 150V 300V 600V 700V 900V 1000V 2500V 3500V

Part
4109001497 4109001499 4109001500 4109001501 4109001503 4109001504 4109001552 4109001554
Number

Oxygen Vessels (see recommended sizes in the technical data)

Capacity, m3 1 2 3

Part Numbers 4109001555 4109001556 4109001557

Optional Equipment Part Number

Flow meter 4109001558


Smart Control Module 4109001559
Stainless Steel Flexible connections from BMO to Oxygen Buffer and from Oxygen
4109001560
buffer to wall kit
Oxygen Emergency Reserve Manifold (origin UK, see separate datasheet) 4109001561

Filling ramp 6+6 cylinders BS341 4109001562

Filling ramp 10+10 cylinders BS341 4109001563

Filling ramp 15+15 cylinders BS341 4109001564

Filling ramp 20+20 cylinders BS341 4109001565

Booster Product Numbers

Models 3/4-150 8/4-150 12/4-150 15/4-150 20/4-150 25/4-150 40/4-150 50/4-150

Part
4109001752 4109001565 4109001566 4109001567 4109001568 4109001781 4109001569 4109001570
Number

BeaconMedæs • Telford Crescent, Staveley, Derbyshire S43 3PF • Phone +44 (0) 1246 474 242 • www.beaconmedaes.com
4109 9900 31.00
2021/06/21
Page 10 of 13

Appendix. II Reference pictures

BMO-1000V at an emergency COVID-19 hospital (Bangladesh)

Oxygen Purity Data trend Combined Purity, Flow and Pressure trends

Smart control Module Main BMO screen

BeaconMedæs • Telford Crescent, Staveley, Derbyshire S43 3PF • Phone +44 (0) 1246 474 242 • www.beaconmedaes.com
4109 9900 31.00
2021/06/21
Page 11 of 13

Appendix. III Dimensioned Layout example (temporary instal-


lation with small footprint and flexible connections,)
Minimum footprint layout with cylinders and manifold in one room
is shown on the picture below. In this proposal the concentrator,
oxygen buffer tank and the booster are connected with flexible
connections with screw in fittings.
NOTE: it is recommended to install high pressure cylinders and manifold in a
separate room for safety reasons. Please verify against the local regulations if it is
allowed to keep cylinders in the same room.

6
300

1500

1000 1700

5
1

4
2
5040

2400

1000
1120

1206
1000

600 1500

6700

BeaconMedæs • Telford Crescent, Staveley, Derbyshire S43 3PF • Phone +44 (0) 1246 474 242 • www.beaconmedaes.com
4109 9900 31.00
2021/06/21
Page 12 of 13

Appendix. IV Wall filtration kit

1 2 3 4 5 6 7 8 9 10

6
4 5 A
1 2 3
requirements of the List of Prohibited Substances

2
All materials supplied are in compliance with the

1 Oxygen Concentrator 1 BMO


Braided Hoses
2 2 Stainless Steel,L=2m
(Optional)
3 Oxygen Buffer Tank 1
4 Flow Meter (Optional) 1 R3/4"
E
Includes Dust Filter, Needle
5 Wall Mounted Kit 1
Valve, Non-Return Valve
6 Muffler 1 R3/4"
7 Hose 1 Rubber,L=2m
This document is property of Atlas Copco AB and shall not without our permission be altered,

8 Copper Outlet 1 DN=22mm


copied, used for manufacturing or communicated to any other person or company.

No Desc Qty Remark


Part list F
Tolerances, if not indicated, according to:

Name Confidentiality Class


Notifications: BMO Wall Filtration Kit acc. to 1102 K

1. Except the wall mounted kit Part 5, all other parts will be packed separately Material Confidential

2. It's necessary to keep all connections well sealed and clean while assembling on site Treatment
INV
CONFIDENTIAL:

3. Flow meter and braided hoses are optional to order Scale 1 : 15 Family Compare Drawing
Owner
Transferred
from

Drawn by Blank nr. Replaces G


- Drawing format Blank wt. Fini wt. 1 / 1
Kg N/A Designation Sheet
STATUS A3
Posi- Des checked. Prod checked. Approved. Date
Ed
tion
Modified from Date Intr./Appd. Parent 3D model Ed . Version 3D Preliminary 5/31/2021

BeaconMedæs • Telford Crescent, Staveley, Derbyshire S43 3PF • Phone +44 (0) 1246 474 242 • www.beaconmedaes.com
In an effort to continuously improve our 4109 9900 31.00
products, the right is reserved to change the 2021/06/21
specification of the items described herein
Page 13 of 13
at any time. Please contact us for further
information and up to date specifications.

Appendix. V Configuration guide

Component Optional/ Standard Comment

BMO concentrator contains the air supply,


BMO Main part number for oxygen concentrator filtration, PSA unit, control in one box and delivers
Pharmacopeia quality oxygen out of the box.
Always included in BMO main part number, Connections from threads to copper pipe - needed
Stub connection pipes
no separate ordering required to braze medical copper pipeline.
Includes oxygen filter, purge kit and flow limiting and
Always included in BMO main part number,
Wall kit non-return valves. Purge valve is needed to remove
no separate ordering required
oxygen from the oxygen buffer tank.
Always needed, size of the buffer tank to be selected
Oxygen Tank Shall always be ordered with separate PN depending on the concentrator flow and if booster is
used (larger tanks to be used with booster).

Flow meter Optional Not needed if smart Control Module is used.

Flexible connections from BMO to oxygen tank and


from oxygen tank to wall kit, to booster and from
booster to filling ramp(stainless steel flexibles
Flexible connections Optional
are only used as temporary solution to simplify
commissioning, recommended to be replaced by
brazed copper pipe).

Smart control Module Optional One PN for all models

Needed if booster is used. Number of cylinders and


Filling ramp Optional
connection type to be specified
Used for setting up the reserve source of supply
Oxygen Emergency
Optional from high pressure cylinders. Needed if not already
Reserve manifold
installed in the facility.
Used for cylinders filling. Size to be chosen based
Booster Optional on the concentrator capacity and number of
cylinders to be filled per day.

Cylinders Optional Cylinders to be provided by facility.

BeaconMedæs
Telford Crescent
Staveley, Derbyshire
S43 3PF, England
Tel: +44 (0) 1246 474 242
www.beaconmedaes.com