I- INTRODUCTION :

The ISOSAMPLE 5100 JAC has been designed to sample oils from
upstream crude-oils to refined products according to the requirements
of the most usual standards :ISO 3171 ; ASTM D4177 ; IP Petroleum.
II- BENEFITS :

A main benefit of ISOSAMPLE 5100 JAC is to


integrate the complete line of features required to keep
sample integrity e.g. :
- Large bore By-pass Loop Isokinetic sampling 1” OD
with large opening size and 2.4 m/s constant velocity
(as per API recommendations)
- or
- True Isokinetic sampling as option with By-pass
Loop controlled identical to the main line velocity .
- Accurate extractor grab size repeatability assured
by the three stage positive displacement principle of
the JISKOOT AUTOMATION cell Model 210
(pneumatic) or Model 710 (electric ).
- Sample extracted from Process Line is directly
stored in the Sample Receiver Chamber integrated
to the Sampler .
- Sample Receiver Chamber is equipped with an
high-shear mechanical mixer complying with ISO
3171 chapter 12.2.2 operated at 3000 rev/mn to
prevent stratification of heavier fractions and to
produce water droplets below 50 µm but not less
than 1 µm during the sample transfer to the 1 liter
portable receiver
- Sample Receiver Chamber is a floating piston
cylinder to suit with split phase crude-oil ( Constant
Pressure Operation ) .
- A feed-back to the Control Unit indicates the
amount of sample contained inside of Sample
Receiver and a local indication of filling is given by
the piston shaft position .
- An other feed-back free contact is issued to verify
II- that pulses have been transferred to the sample
solenoid valve .
1- Sample receiver - Sampling software is in compliance with Service
2- Portable receiver Factor Requirement (ISO 3171 )
3- Control Unit -
III - OPERATING PRINCIPLE :

The ISOSAMPLE 5000 is a Cell Sampler fed


by a 5 m3/h pump (API) or at main line velocity.
At the Sampling start-up, the upper chamber of
sampler (hydraulic chamber) is fully filled with
air ( pneumatic system ) or hydraulic oil
electric system ) and the piston is in down
position .
For each grab operated by the 210 or 710
sample extractor , a calibrated volume of
sample (usually 1cc) is transferred in the
Sample Receiver Chamber .
As the cell extractor operates within a unique
three-stage positive displacement action , the
sampler is absolutely insensitive to the sample
process pressure, viscosity , density and vapor
pressure .
Sample in Receiver Chamber pressurised at
the process pressure is prevented to vaporising
even if the vapor pressure at ambient
temperature exceeds barometric pressure .
For remote monitoring of sample contained
inside of Receiver ,the position of the floating
piston in the cylinder is monitored by a
resistive displacement level transmitter tied to
the piston shaft . No calibration is required
within the life of this component .
At the end of sampling , the stirrer is operated
and a flow trough transportation cylinder of
1liter ( approx. 3 Kg) is filled for lab.
Analysis .
Sampling processing is operated by pilot pulses
issued by separate Control Unit for safe or
flammable area .
Grabs are taken in proportion to flow or to time
.
A permanent comparison between the volume
collected in the Receiver and the batch volume
actually monitored on line is achieved ,
therefore confirmation that sampler extracts a
uniform grabs in flow or time proportional
manner is made .
At the end of sampling , compliance with Service Factor
Requirement (ISO 3171 ) is checked and reported by
alarm contact .
All equipment is explosion-proof certified to ATEX II 2G
EEx d II BT4 ( as option material Class IIC certified can
be supplied) .

IV : TECHNICAL SPECIFICATIONS :
SAMPLER PROBE :
- Sampler applications : Crude Oil , Refined Oil ,
Chemical Products .
- Wetted materials : Stainless Steel 316L with PTFE
seals.
- Operating pressure : from 16 barsg up to 50 barsg (
725 psig ) for electric type ; up to 100 barsg ( 1450
psig ) for pneumatic type .
- Operating temperature : -20° to 100 °C ( 4F to 212 F)
- Max. viscosity : 5 000 CST ( electric ) or 8 000 CST (
pneumatic ).
- Mounting flange inlet/outlet : 1" Class
150/300/600#RF
- Sample Grab Size : usually 1cc.
- Grab Size Repeatability : better than +/- 2%
(pneumatic ) or +/- 5% ( electric ).
- Grab rate : up to 100 grabs/mn ( pneumatic ) or 50
grabs/mn ( electric ).
- Power supply : 115/230 VAC - 50/60 Hz- 200 W
- Instrument Air supply : 2 to 8 barg ( 29 to 116 psig)
- Protection rating : CENELEC EExde II B T3

V - CONTROL UNIT “ SCAMEX “ :

- Providing full automatic or manual control plus


reporting and logging as specified above the
SCAMEX interfaces to the user Station Control
System by RS-485 data-link.
- As option a MODBUS communication protocol can
be supplied .
- Micro-processed equipment
- Inlets :
- 4-20 MA from Customer Process Line flowmeter
- Resistance though EExi isolator interface from
Receiver level transmitter
- Outlet : 24 V cc to Sampler Grab Actuator
- SPDT Alarm contact .
- Read-out & key-board in front door
SAMPLE RECEIVER - Version : Ex-proof ATEX II 2 G EEx d II CT4
FLOATING PISTON CYLINDER - Power supply : 115/230 VAC-50 /60 Hz.
VI – SAMPLING CABINET :

All the system of RECEIVER with isolation


valves , manifold for connection of
transportation cylinder and CONTROL UNIT
“SCAM ” is enclosed in a rugged cabinet in
Stainless Steel 20/10 mm sheet and IP 55 rating
to accept location in harsh environment.

The front door features laminated safety


windows for control unit display view.

Electrical radiator ATEX II 2G EEx d II B T4 is


provided to make the system suitable from – 20
to + 50 °C.

When required and as option , two receivers


auto-switched are provided in the cabinet .