Professional Documents
Culture Documents
Foreword..........................................................................................................................................3
1. Revisions....................................................................................................................................4
2. Legal Information....................................................................................................................5
2. Processing Documents........................................................................................................9
3. Review of Contracts............................................................................................................10
5. Continuous Improvement..................................................................................................12
7. Certificates..............................................................................................................................16
Technical Requirements....................................................................................................21
1. Personnel..................................................................................................................................22
3. Laboratory Equipment........................................................................................................26
4. Pre-analytical Procedures.................................................................................................57
5. Analytical Procedures.........................................................................................................58
7. Results Report........................................................................................................................65
Appendices..................................................................................................................................67
1. Glossary of Terms.................................................................................................................68
2. References...............................................................................................................................78
1. Revisions....................................................................................................................................4
2. Legal Information....................................................................................................................5
2.1. Declaration of Conformity...........................................................................................................5
2.2. Notice of Liability........................................................................................................................5
2.3. Trademarks.................................................................................................................................5
2.4. Graphics......................................................................................................................................5
2.5. Document Symbols.....................................................................................................................6
2.6. Copyright ® 2011 by HORIBA ABX SAS....................................................................................6
2.7. Document Intended Use.............................................................................................................6
1. Revisions
All information included in this document is current as of the date of creation of this version. Changes
that may occur will be available on www.horiba.com.
To update a paper document, please contact your local HORIBA Medical representative.
2. Legal Information
This instrument responds to the Standards and Directives named in the Declaration of Conformity.
Latest version of the EC Declaration of Conformity for this instrument is available on www.horiba.com.
The information in this manual is distributed on an "As Is" basis, without warranty. While every
precaution has been taken in the preparation of this manual, HORIBA Medical will not assume any
liability to any persons or entities with respect to loss or damage, caused or alleged to be caused
directly or indirectly by not following the instructions contained in this manual, or by using the
computer software and hardware products described herein in a manner inconsistent with our
product labelling.
2.3. Trademarks
2.4. Graphics
All graphics including screens and printouts, photographs are for illustration purposes only and are
not contractual.
To alert the operator of potentially hazardous conditions, symbols described in this chapter are
provided wherever necessary throughout the manual.
Emphasizes information that must be followed to avoid hazard to either the operator or the
environment, or both.
Emphasizes information that must be followed to avoid possible damage to the instrument
or erroneous test results.
Emphasizes information that can be helpful to the operator before, during or after a
specific operational function.
All rights reserved. No part of this publication may be reproduced or transmitted in any form or by any
means, electronic, mechanical, photocopying, recording, or otherwise, without the prior written
permission of HORIBA Medical.
This document is intended for clinical laboratories, to help them implementing their accreditation
process and to provide them with the necessary references and reference documents based on the
standard ISO 15189 : 2007 requirements. The paragraphs of this standard specified hereafter are
associated to the response given by HORIBA Medical, in order to facilitate the accreditation process.
This document is applicable to both ABX Micros ES 60 and ABX Micros 60.
Performance data is also available in this document.
2. Processing Documents........................................................................................................9
3. Review of Contracts............................................................................................................10
5. Continuous Improvement..................................................................................................12
7. Certificates..............................................................................................................................16
2. Processing Documents
3. Review of Contracts
All contracts (maintenance, remote assistance policy, suppliers charter) signed with HORIBA Medical
must be retained.
ABX Micros ES 60, its reagents, controls and calibrators comply with the standards and directives
that are included in each declaration of conformity.
Latest version of the CE Declaration of Conformity for this instrument is available online at
www.horiba.com.
The quality management system of HORIBA Medical is certified with respect to the standards
ISO 13485 : 2003 and ISO 9001 : 2008.
5. Continuous Improvement
These records may include but are not limited to the following:
■ a- request forms (including the patient chart or medical record only if used as the request form);
■ b- examination results and reports;
■ c- instrument printouts;
■ d- examination procedures;
■ e- elaboratory work-books or sheets;
■ f- accession records;
■ g- calibration functions and conversion factors;
■ h- quality control records;
■ i- complaints and action taken;
■ j- records of internal and external audits;
■ k- external quality assessment records/interlaboratory comparisons;
■ l- quality improvement records;
■ m- instrument maintenance records, including internal and external calibration records;
■ n- lot documentation, certificates of supplies, package inserts;
■ o- incident/accident records and action taken;
■ p- staff training and competency records.
§ c, g, h, m: Refer to the ABX Micros ES 60 user manual: in Setup section for the printing conditions,
and in Maintenance and Troubleshooting section for the log entries printouts (calibration, reagents,
etc).
§ k: HORIBA Medical proposes a Quality Control Program for the inter-laboratory comparison of Daily
Internal Control results via its website.
§ n: HORIBA Medical Latest versions of these documents are available online at www.horiba.com.
§ p: Following each of the training courses undertaken in accordance with the quality control
procedures of HORIBA Medical, a certificate of attendance and a certificate of training will be
awarded.
■ j- nonconformities;
■ k- monitoring of turnaround time;
■ l- results of continuous improvement processes;
■ m- evaluation of suppliers.
Shorter intervals between reviews should be adopted when a quality management system is being
established. This will allow early action to be taken in response to those areas identified as requiring
amendment of the quality management system or other practices.
Address:
HORIBA Medical
Parc Euromédecine - Rue du Caducée
B.P. 7290
34184 MONTPELLIER Cedex 4 - FRANCE
Phone: +33 (0)4 67 14 15 16
Fax: +33 (0)4 67 14 15 17
Website: http://www.horiba.com/medical/
Company Managers
President Bertrand de Castelnau
email: katia.krouk@horiba.com
Tel. +33 (0)4 67 14 15 38 (secretariat)
Reagents/Consumables Information
The reagents are listed in an annual catalogue available in paper format and on the website.
The reagents are all CE marked in conformity with the 98/79/CE directive.
HORIBA Medical provides the reagents notices and the security information sheets of the reagents
and blood controls and calibrators in a multi-language electronic version.
These documents are available on the CD-Rom provided with the analyser.
These documents are also available on www.horiba.com
7. Certificates
1. Personnel..................................................................................................................................22
3. Laboratory Equipment........................................................................................................26
3.1. Calibration.................................................................................................................................26
3.2. Maintenance..............................................................................................................................26
3.3. Energy Consumption................................................................................................................26
3.4. Waste Disposal.........................................................................................................................27
3.5. Performance..............................................................................................................................27
3.6. Labeling.....................................................................................................................................54
3.7. Individual Machine Data Sheets................................................................................................55
3.8. Manuals and Instructions for Use.............................................................................................55
3.9. Disposal of Chemical Products and Biological Materials.........................................................56
3.10. Use of Computers for the Analysis Data.................................................................................56
4. Pre-analytical Procedures.................................................................................................57
4.1. Specimen for Sampling.............................................................................................................57
5. Analytical Procedures.........................................................................................................58
7. Results Report........................................................................................................................65
1. Personnel
To meet this requirement, HORIBA Medical has a training center that enables laboratory staff to
receive appropriate training for the correct use of the machines. This structure makes it possible to
undergo user training, on the client’s premises; these training courses are carried out at the time of
installation. Also, complete training courses (operation, result interpretation, maintenance) are
available in the HORIBA Medical training centers. All of these training courses have specific programs,
documentation, and registrations.
Biological risk
■ Bleach:
Bleach is essentially marketed in 2 forms:
■ concentrated bleach in 250 mL sachets: 9.6% active chlorine
■ ready-to-use bleach in containers: 2.6% active chlorine.
Bottles or containers of ready-to-use 2.6% active chlorine solution keep for at least one year after
opening.
Sachets of 9.6% active chlorine solution have limited stability over time and must be diluted within 2.5
to 3 months following the manufacturing date shown on the packaging.
■ Concentration and contact time: (Use of diluted bleach)
0.5% of active chlorine inactivates CTAs (Conventional Transmissible Agents) through contact of
at least 15 minutes (dilution of 1/5 using a 2.6% concentration or dilution of 1/20 using a 9.6%
concentration).
In practice, allow for overnight contact time so that the chlorine ions are able to disperse upon
contact with proteins in the treated environment.
Chemical risk
Decree 2002-540 of the environmental code (Decree no. 2007-1467).
Chemical risks must be evaluated based on laboratory activities, the types of instruments and
reagents used and the frequency of tests.
Consult safety data sheets for the reagents used to identify dangerous substances and calculate the
final concentrations of those substances for comparison with the limits set in the environmental code
(Art R 541-7, Art R 541-8 and appendices 1 and 2, Art R 541-9, Art R 541-10).
Conclusion: following elimination of biological risk and in the absence of chemical risk,
the effluent may be emptied into the sink.
In the event of residual chemical risk, the user must take the necessary measures to
dispose of this effluent
Reference:
■ Circular DGS/SD5C/DHOS/E2/DRT/CT1/CT2 no. 2004-382 dated July 30, 2004, regarding
precautions to be taken in anatomy and cytology departments, autopsy rooms, mortuary facilities
and "NCTA specialist" biological laboratories concerning the risk of transmission of conventional
transmissible agents (CTA) and non-conventional transmissible agents (NCTA).
■ Circular DGS/5C/DHOS/E2 no. 2001-138 dated March 14, 2001, regarding precautions to be
taken during treatment aimed at reducing the risks of transmission of non-conventional
transmissible agents.
■ Laboratory Biosafety Manual, Second Edition (revised), World Heath Organization. Geneva, 2003.
■ Notice from the SFHH (Société Française d'Hygiène Hospitalière – The French Society for Hospital
Hygiene) regarding the use of bleach in healthcare establishments. June 2006.
3. Laboratory Equipment
3.1. Calibration
The controls and calibrators associated with the ABX Micros ES 60, supplied by HORIBA Medical,
comply with the requirements of the standard ISO 17511 : 2003, relating to the metrological
traceability of values attributed to calibration agents and control equipment (in vitro diagnostic
medical devices) for the measurement of quantities in biological samples. This makes it possible to
guarantee that the results obtained by the ABX Micros ES 60, duly controlled and calibrated, are
sufficiently exact to enable a clinically correct interpretation and comparability over time and space.
3.2. Maintenance
Refer to Specifications > Technical Specifications chapter of the user manual (RAB237).
For waste disposal, refer to Specifications > Waste Handling Procedure and to Introduction >
Environmental Protection chapters of the user manual (RAB237).
3.5. Performance
3.5.1. Parameters
The techniques used for counting, measuring, and detecting the various parameters are explained in
the Description and Technology section of the user manual (RAB237).
* PDW and PCT have not been established as indications for use in United States for this
instrument. Their use should be restricted to Research Use Only (RUO). Not for use in
diagnostic procedure.
Mean CV obtained with several batches of control blood (ABX Minotrol 16):
These results are based on the study on precision by replicate runs of control bloods, according to
the protocol detailed in the CLSI EP15-A2 document.
ABX Micros ES 60 OT
ABX Micros ES 60 CT
Parameters Level CV
LYM% 24.74% 5.86%
MON% 6.35% 10.12%
GRA% 68.13% 4.09%
Repeatability:
Trueness is defined as ‘the closeness of agreement between the average value obtained from a large
series of test results and an accepted reference value’ (ISO 3534- 1-3.12) 1 . Accuracy, on the other
hand, is defined as the ‘closeness of the agreement between the result of a measurement and a true
value of the measurand’ (VIM 93-3.5 ) 2 .
In ISO/CD 17511 3 , the concept of ‘accuracy of measurement’ relates both to trueness of
measurement and precision of measurement. The IVD-Directive 4 , on the other hand, refers to
specific analytical performance characteristics and uses the term 'accuracy' and 'trueness'
synonymously. Büttner 5 (see figure below) has presented a classification of analytical performance
characteristics and techniques for their evaluation. There is an important difference between
‘trueness’ and ‘accuracy’.
The definition of ‘uncertainty of measurement’ from the same source (VIM, 1993)2 is also widely
defined as a ‘parameter, associated with the result of a measurement, that characterises the
dispersion of the values that could reasonably be attributed to the measurand’. All laboratory tests are
subject to uncertainties inherent in the test or errors arising during performance.
ERROR
Bias calculation
Trueness, quantified by the bias, was calculated by comparing the mean obtained (m) in
reproducibility with respect to the expected target value (V), assimilated to the true value of the tested
sample. For the white blood cell differential count, the bias was calculated with respect to the value
obtained using a manual count on 200 cells.
1 Statistics-Vocabulary and symbols-Part 1: Probability and general statistical terms. ISO 3534-1, Geneva: International
calibrators and Control Materials, ISO/CD 17511, Geneva: International Organization for Standardization (ISO), 2000.
4 Directive 98/79/EC of the European Parliament and of the Council of 27 October 1998 on in vitro diagnostic medical devices
The bias was calculated using data obtained on a selection of analyzers. The indicated bias limit is
calculated from the control charts to two standard deviations.
(m - V)
Bias in % = ______ x 100
V
Bias on the blood counts undertaken with several batches of ABX Minotrol 16:
ABX Micros ES 60 OT
ABX Micros ES 60 CT
Calculation of uncertainty
The uncertainty on blood results is calculated from two components:
■ The component U1 due to the reliability approached by the reproducibility standard deviation. This
component is obtained from the values obtained with control bloods with the exception of the
differential white blood cell count for which the calculation was undertaken on human blood.
U1 = S repro
■ The component of uncertainty due to trueness, U2
Uncertainty #
U2 = _____________
√3
Uncertainty # = (m - V) : where m is the mean obtained, and V is the expected target value.
■ The combined uncertainty Uc is calculated using the following formula:
_____
Uc = √ U12 +U22
■ Expanded uncertainty = 2 * Uc
The capability of an analytical system is the measurement of the relationship between its true
performance and the requested performance. In the “Six Sigma” approach, the capability indicator
represents the difference between the requested performance (total allowable error TEa) and the
mean expressed in numbers of standard deviations, hence the formula:
(TEa - bias %)
Sigma = _______
CV
■ CV: Coefficient of Variation
■ TEa: total allowable error is that given by Ricos & al. 6 for a risk of 5 %
6 DESIRABLE SPECIFICATIONS FOR TOTAL ERROR, IMPRECISION, AND BIAS, DERIVED FROM BIOLOGIC VARIATION by Ricos
C, Alvarez V, Cava F, Garcia-Lario JV, Hernandez A, Jimenez CV, Minchinela J, Perich C, Simon M. "Current databases on biologic
variation: pros, cons, and progress." Scand J Clin Lab Invest 1999;59:491-500 (revised in 2008).
Graphs of capability
DIF parameters TEa Ricos & al. Level Bias % Uncertainty % Sigma
14 13.8 % 2.34 % 3.68
LYM% 22.4 % 31 11.4 % 1.93 % 5.70
50 18.3 % 2.51 % 1.62
3.0 126.7 % 7.39 % -14.97
MON% 16% 7.5 1.4 % 5.42 % 2.69
14.0 41.8 % 4.47 % -5.78
DIF parameters TEa Ricos & al. Level Bias % Uncertainty % Sigma
39.5 18.6 % 4.72 % 1.97
GRA% 27.9 % 61.2 6.1 % 1.37 % 15.98
80.5 0.6 % 0.62 % 43.83
Refer to the Specifications section for parameters definitions and to the Description and Technology
section of the user manual (RAB237) for the measurement techniques and the specificities of the
parameters of ABX Micros ES 60.
The device composed of the ABX Micros ES 60 is a system enabling the in vitro diagnostic analysis of
whole blood samples. Refer to the Specifications > Limitations chapter of the user manual (RAB237)
for the types of samples authorized for the ABX Micros ES 60.
Parameter Linearity limits Visible Range Sample Tested range Error limits
Each obtained value must be within the range (defined by the Error limits) regarding the expected
value.
AMI validation is the process of confirming that the assay system will correctly recover the quantity of
the measurand over the AMI.
9
y=x
8 R2 = 1
7
Obtained result G/L
0
0 1 2 3 4 5 6 7 8 9
Expected result G/L
The more appropriate analytical term is recovery, as established by serial dilutions of a known high
concentration of cells. The goal is to establish the AMI, which may be a subset of the CRI, as follows:
■ The AMI is the range of analyte values that a method can directly measure on the specimen
without any dilution, concentration, or other pretreatment not part of the usual assay process.
■ The Clinically Reportable Interval (CRI) is the range of analyte values that a method can
measure, allowing for specimen dilution, concentration, or other pretreatment used to extend the
direct AMI. A manufacturer establishes the AMI through its validation testing whereas the CRI is
the end-user clinical laboratory’s responsibility.
White Blood Cell Concentration, Red Blood Cell Concentration, Hemoglobin, Hematocrit and
Platelet Concentration.
Commercially available "linearity kits" are assayed and evaluated. These data may be used to
represent the AMI based on commercial kit material processed per manufacturers’ directions, but do
not necessarily represent performance with actual fresh blood clinical specimens.
Contamination was assessed by running a sample with high results for the following parameters, three
times in a row (H1, H2, H3) followed by a sample with low results, also 3 consecutive times (L1, L2,
L3).
(L1 - L3)
Carry-over (%) = ________ x 100
(H3 - L3)
ABX Micros ES 60 OT
ABX Micros ES 60 CT
Results obtained are within the specifications. There is no significant carry-over effect observed.
3.5.11. Stability
Reagent stability
Refer to the reagent leaflets for stability (Storage and Stability chapter). Latest version of documents
online at www.horiba.com.
3.5.12. Robustness
The robustness of an analysis method is the resistance to any variation in its results when minor
modifications occur regarding experimental conditions described in the procedure.
The inter-laboratory comparison (via Quality Control Program - QCP) gives a monthly overview of the
influence of factors such as: changing instrument, changing reagents, maintenance operations,
operator, geographic place, calibration, and temperature, humidity and pressure conditions.
Robustness of the blood count parameters can thereby be confirmed.
Results transmitted to QCP in October 2009 on control blood batches allowed the drawing up of the
following tables.
Precision of the peer-groups allows to detemine the robustness of the tests.
Normal level:
High level:
Unlysed red blood cells: in certain cases of membrane resistance, partial lysis of red blood cells may
be observed. These unlysed red blood cells may cause an erroneously high white blood cell count.
These unlysed red cells can be detected on the WBC curve via an L1 alarm or in the form of an
elevated base line of the lateral ascending section of the lymphocyte population.
Multiple myeloma: the precipitation of immunoglobulins in patients with multiple myeloma may give
elevated WBC counts.
Hemolysis: hemolyzed specimens contain an erythrocyte stroma, which may cause elevated white
blood cell counts.
Platelet agglutination: the accumulation of platelets may cause an elevated white blood cell count.
Platelet agglutination triggers the alarms L1.
Leukemia: leukemia can cause fragility of the leukocytes and subsequent destruction of these cells
during the count, thus resulting in an abnormally low white blood cell count. An abnormally low
leukocyte count may also be seen in patients with chronic lymphoblastic leukemia due to the
presence of abnormally small lymphocytes, which may not be counted by the analyzer.
Chemotherapy: cytotoxins and immunosuppressants may weaken the leukocyte membranes and
result in a low leukocyte count.
Cryoglobulins: the increased levels of cryoglobulins that may be associated with various conditions
(myeloma, carcinoma, leukemia, macroglobulinema, lymphoproliferative disorders, metastatic tumors,
autoimmune disorders, infections, aneurysms, pregnancy, thromboembolic phenomena, diabetes,
etc.), may cause an increase in the leukocyte, erythrocyte, and platelets counts and the hemoglobin
concentration. The samples should be warmed to 37°C (99°F) in a water bath for 30 minutes and then
re-run immediately afterwards (using the analyzer or a manual method).
Macrothrombocytes: in excessive numbers, they may affect the leukocyte count by increasing the
number of leukocytes counted.
Erythroblasts: high concentration of erythroblasts may increase the leukocyte count.
The red blood cell dilution contains all of the elements found in the blood: erythrocytes, leukocytes,
and platelets. During the erythrocyte count, the platelets are not counted as they are smaller than the
defined minimum threshold. In very rare cases with an extremely high leukocyte count, the erythrocyte
count may be increased. It should be corrected, especially if the latter is very low in comparison with
the leukocyte count.
Agglutinated red blood cells: these may cause a falsely low RBC count. Blood samples containing
agglutinated red blood cells can be identified by abnormal MCH and MCHC values and the
examination of a stained blood smear.
Cold agglutinins: IgM, which are elevated in Cold Agglutinin Disease, may lower erythrocyte and
platelet counts and increase the MCV. The samples should be warmed to 37°C (99°F) in a water bath
for 30 minutes and then re-run immediately afterwards (using the analyzer or a manual method).
Turbidity of the blood sample: several physiological and/or therapeutic factors may produce falsely
elevated hemoglobin results. To obtain accurate results in blood samples with increased turbidity,
determine the cause of the turbidity and follow the appropriate method below:
■ An elevated leukocyte count: a very high leukocyte count will cause excessive diffusion of the
light. In such cases, the reference methods (manual) should be used. The diluted sample should
be centrifuged, and the supernatant fluid measured with a spectrophotometer.
■ Elevated lipemia: elevated lipemia levels makes the plasma look milky. This phenomenon can be
seen with hyperlipidemia, hyperproteinemia (as in gammopathies) and hyperbilirubinemia.
Accurate hemoglobin measurement can be achieved by using reference (manual) methods and a
plasma blank.
Increased turbidity: this phenomenon can be seen with red blood cells that are resistant to lysis. It
causes a falsely elevated HGB concentration, but can be detected due to abnormal MCHC and MCH
values and to an increase in the base line of the ascending section of the WBC curve. An erroneous
hemoglobin concentration also causes erroneous MCH and MCHC values.
Fetal blood: the mixing of fetal and maternal bloods may produce a falsely elevated hemoglobin
value.
Red blood cells agglutination: can cause an inaccurate HCT value. Red blood cell agglutination may
be detected by observing abnormal MCV and MCH values, and by examining a stained blood smear.
In such cases, manual methods may be required to obtain an accurate hematocrit value.
Red blood cell agglutination: can cause an inaccurate MCV value. Red blood cell agglutination may
be detected by observing abnormal MCH and MCHC values, and by examining a stained blood
smear.
Excessive numbers of large platelets: and/or the presence of an excessively high WBC count may
interfere with the accurate determination of the MCV value. In such cases, careful examination of a
stained blood smear may reveal the error.
The interferences cited for HGB and RBC affect the MCH and may cause inaccurate results.
The interferences cited for HGB and the HCT affect the MCHC and may cause inaccurate results.
The interferences cited for RBC and MCV affect the RDW and may cause inaccurate results.
Red blood cell agglutination: this phenomenon may cause a falsely low erythrocyte count and an
erroneous RDW. In the blood samples, red blood cell agglutination may be detected by observing
abnormal MCH and MCHC values, and by examining a stained blood smear.
Nutritional deficiency or blood transfusion: these phenomena may cause elevated RDW results due
to iron, vitamin B12, or folate deficiencies. It is also possible to observe an elevated RDW from the
bimodal distribution of red blood cells from transfused blood.
Very small erythrocytes (microcytes): the presence of erythrocyte fragments (schistocytes), and
WBC fragments may interfere with the platelet count giving falsely elevated values.
Red blood cell agglutination: may trap the platelets and cause a falsely low platelet count. Red
blood cell agglutination may be detected by observing abnormal MCH and MCHC values, and by
examining a stained blood smear.
Excessive numbers of Macro platelets: this phenomenon may cause a falsely low platelet count
due to the fact that these macro platelets exceed the upper threshold defined for platelets and are
therefore not counted as platelets.
Chemotherapy: cytotoxins and immunosuppressants may weaken these cells and result in a falsely
low count. Manual methods may be necessary to obtain the platelet count.
Hemolysis: hemolyzed samples contain a red blood cell stroma which may affect the platelet count.
Citrate blood - Blood anti-coagulated with citrate may contain platelet aggregates which could
decrease the platelet count.
RBC Inclusions: including Howell-Jolly bodies, Heinz bodies, siderotic and basophilic granules, etc.,
may cause falsely elevated platelet counts.
Platelet agglutination: the accumulation of platelets may cause a low platelet count. The sample
should be repeated and drawn into a sodium citrate anticoagulant tube to rule out the anticoagulant
as a cause of aggregation and run again to determine the platelet count alone. The final platelet count
should be corrected, making allowance for the dilution caused by the sodium citrate. Platelet
agglutination triggers the alarms L1.
Elevated lipids and/or cholesterol: may interfere with correct platelet counting. From patients
undergoing parenteral nutrition with intralipids brought, it is noted an over-estimate of the platelet
counting which can mask a thrombopenia.
Elevated bilirubine: may interfere with correct platelet counting. From patients with severe hepatic
disorder, liver transplant… It is noted an over-estimate of the platelet counting which can mask a
thrombopenia.
Parenteral nutrition: Interference in PLT result may occur for samples from patients undergoing
parenteral nutrition with injection of lipid emulsion.
Macro platelets: their volume exceeds the upper threshold defined for platelets and they are not
therefore included in the calculation of the mean platelet volume by the analyzer. The MPV value may
be falsely lowered.
Very small erythrocytes (microcytes) or presence of red blood cell fragments (schistocytes) and
white blood cell fragments may interfere with the accurate determination of the mean platelet
volume.
Red blood cell agglutination: may trap the platelets causing an incorrect MPV. Red blood cell
agglutination may be detected by observing abnormal MCH and MCHC values, and by examining a
stained blood smear.
Chemotherapy: may also affect platelet volume.
Blood samples collected in EDTA will not maintain a stable Mean Platelet Volume.
Platelets collected in EDTA swell with time and temperature.
The presence of erythroblasts and erythrocytes that are resistant to lysis may cause an inaccurate
lymphocyte count. Limitations to the leukocyte count also apply to the determination of the number
(absolute value) and percentage of lymphocytes.
The presence of large lymphocytes, atypical lymphocytes, lymphoblasts, and excessive numbers of
basophils may cause an inaccurate monocyte count. Limitations to the leukocyte count also apply to
the determination of the number (absolute value) and percentage of monocytes.
This study is based on known interferences on the predicate device (ABX Micros 60) and on the CLSI
EP7-A2 – Interference testing in clinical chemistry.
Interferent test concentration
The paired interference testing is conducted at the highest concentrations that a laboratory would
expect to observe among patient specimen.
Potential Interfering substances:
■ Hemolysis
■ Icterus (total bilirubin)
■ Urea
■ Lipemia
Acceptance Criteria
Evaluating the effect of potentially interfering substances added to the sample of interest
The smallest significant differences between test and control is:
%
WBC RBC HGB HCT MCV MCH MCHC RDW PLT MPV % LYM % GRA
MON
Relative
7% 5% 4% 6% 2% 4% 4.3% 9% 22% 9% 1.3% 30% 7%
%
_ _
Compute the observed interference effect, dobs = Interference = (X test - X control)
A one sided test is performed. The 95 % confidence interval for the interference effect is calculated. If
the point estimate dobs is less than or equal to the cut-off value, dc, conclude the bias caused by the
substance is less than dmax.
Urea
On all parameters, no significant interference can be observed with urea for a concentration of
29.5 mmol/L, either on low, normal or high level.
Bilirubin
On all parameters, no significant interference can be observed with bilirubin for a concentration of
150 µmol/L, either on low, normal or high level.
Intra Lipid
On all parameters, no significant interference can be observed with Intra Lipid for a concentration of
3 mmol/L, either on low, normal or high level.
Hemolysis
On all parameters, no significant interference can be observed with Hemolysis. The WBC value is
slightly increased but still into the acceptance criteria.
The reference values vary according to the population and / or the region. Each laboratory is strongly
advised to establish its own set of normal ranges according to the local population.
WBC (103/mm3):
Male Female
Age
# of individuals Reference values # of individuals Reference values
4 - 10 1 880 4.30 - 11.90 1 755 4.40 - 11.70
10 - 18 1 800 4.10 - 10.80 1 620 4.20 - 11.60
18 - 65 3 317 4.10 - 10.50 2 709 6.30 - 10.50
RBC (106/mm3):
Male Female
Age
# of individuals Reference values # of individuals Reference values
4 - 10 1 877 4.20 - 5.30 1 752 4.10 - 5.30
10 - 14 1 167 4.30 - 5.50 168 4.20 - 5.30
14 - 18 1 627 4.50 - 5.70 1 548 4.20 - 5.20
18 - 25 1 562 4.60 - 5.80 1 576 4.10 - 5.20
25 - 45 2 209 4.50 - 5.70 232 4.10 - 5.20
45 - 65 1 509 4.40 - 5.60 186 4.10 - 5.40
HGB (g/dL):
Male Female
Age
# of individuals Reference values # of individuals Reference values
4 - 10 1 878 11.6 - 14.8 1 754 11.6 - 14.9
10 - 14 1 170 12.5 - 15.8 170 12.4 - 15.3
14 - 18 1 631 13.5 - 17.4 1 550 12.6 - 15.6
18 - 25 1 571 14.1 - 17.8 1 574 12.4 - 15.7
25 - 35 1 480 14.1 - 17.6 1 175 11.9 - 15.6
Male Female
Age
# of individuals Reference values # of individuals Reference values
35 - 45 1 744 14.0 - 17.5 1 566 11.6 - 15.8
45 - 55 1 341 13.7 - 17.4 1 291 11.5 - 15.8
55 - 65 1 173 13.6 - 17.6 186 12.7 - 15.9
MCV (fL):
Male Female
Age
# of individuals Reference values # of individuals Reference values
4 - 10 1 858 76 - 88 1 740 76 - 89
10 - 14 1 158 78 - 90 148 79 - 91
14 - 18 1 624 80 - 95 1 544 80 - 95
18 - 25 1 571 81 - 97 1 571 81 - 96
25 - 35 1 487 83 - 97 1 178 80 - 97
35 - 45 1 744 83 - 98 1 569 77 - 98
45 - 55 1 340 84 - 98 1 294 76 - 97
55 - 65 1 172 84 - 100 186 81 - 99
PLT (103/mm3):
Male Female
Age
# of individuals Reference values # of individuals Reference values
4 - 10 7 215 203 - 484 6 726 210 - 482
10 - 14 5 211 185 - 419 4 882 186 - 422
14 - 18 4 163 165 - 385 4 068 175 - 390
18 - 25 2 730 150 - 349 3 934 164 - 384
25 - 45 12 273 153 - 355 13 601 161 - 386
45 - 55 3 515 152 - 368 3 760 168 - 399
55 - 65 1 367 151 - 365 1 480 171 - 386
> 65 260 143 - 361 306 154 - 386
Male Female
Age
# of individuals Reference values # of individuals Reference values
4 - 10 1 880 1.30 - 8.70 1 755 1.40 - 9.00
10 - 18 1 800 1.40 - 8.10 1 620 1.60 - 8.40
18 - 65 3 317 1.70 - 8.00 2 709 1.70 - 8.20
Male Female
Age
# of individuals Reference values # of individuals Reference values
4 - 10 1 880 0 - 1.20 1 755 0 - 1.10
10 - 18 1 800 0 - 1.90 1 620 0 - 1.90
18 - 65 3 317 0 - 1.70 2 709 0 - 1.70
Male Female
Age
# of individuals Reference values # of individuals Reference values
4 - 10 1 880 0 - 0.40 1 755 0 - 0.40
10 - 18 1 800 0 - 0.30 1 620 0 - 0.30
18 - 65 3 317 0 - 0.30 2 709 0 - 0.30
Lymphocytes (103/mm3):
Male Female
Age
# of individuals Reference values # of individuals Reference values
4 - 10 1 880 1.90 - 7.10 1 755 1.00 - 7.10
10 - 18 1 800 1.80 - 6.10 1 620 1.80 - 6.10
18 - 65 3 317 1.70 - 5.20 2 709 1.70 - 5.30
Monocytes (103/mm3):
Male Female
Age
# of individuals Reference values # of individuals Reference values
4 - 10 1 880 0.02 - 1.30 1 755 0.02 - 1.20
10 - 18 1 800 0.03 - 1.30 1 620 0.02 - 1.20
18 - 65 3 317 0.04 - 1.30 2 709 0.02 - 1.10
References:
AIDE MEMOIRE D’HEMATOLOGIE, 1998. Prof. C. Sultan / M. Gouault - Helman / M. Imbert, Service
Central d’Hématologie de l’Hôpital Henri Mondor, Faculté de médecine de Créteil (Paris XII).
Lecture critique de l’hémogramme : valeurs seuils à reconnaître comme probablement pathologiques
et principales variations non pathologiques. ANAES : Agence Nationale d’Accréditation et d’Evaluation
en Santé.
The correlation of the following parameters of the analyzer has been proven by regression analysis
comparing the analyzer with two recognized comparative analyzers, on 100 samples of whole blood
from patients, within the normal operating ranges of the analyzer (according to EP09-A2 evaluation
protocol):
ABX Micros ES 60 OT
R2 (comparison of means)
Parameter Accuracy requirements
Versus ABX Micros 60 Versus ABX Pentra 80
WBC 0.9971 0.9947 > 0.95
RBC 0.9915 0.9970 > 0.95
HGB 0.9951 0.9981 > 0.95
HCT 0.9909 0.9942 > 0.95
MCV 0.9884 0.9909 > 0.95
PLT 0.9826 0.9868 > 0.95
LYM% 0.9881 - > 0.95
MON% 0.9205 - > 0.95
R2 (comparison of means)
Parameter Accuracy requirements
Versus ABX Micros 60 Versus ABX Pentra 80
GRA% 0.9898 - > 0.95
ABX Micros ES 60 CT
R2 (comparison of means)
Parameter Accuracy requirements
Versus ABX Micros 60 Versus ABX Pentra 80
WBC 0.9963 0.9952 > 0.95
RBC 0.9873 0.9928 > 0.95
HGB 0.9879 0.9919 > 0.95
HCT 0.9858 0.9849 > 0.95
MCV 0.9647 0.9662 > 0.95
PLT 0.9860 0.9863 > 0.95
LYM% 0.9890 - > 0.95
MON% 0.8680 - > 0.95
GRA% 0.9885 - > 0.95
3.6. Labeling
4. Pre-analytical Procedures
■ b- procedures for:
■ 1- preparation of the patient (e.g. instructions to caregivers and phlebotomists),
■ 2- identification of primary sample,
■ 3- primary sample collection (e.g. phlebotomy, skin puncture, blood, urine and other body
fluids), with descriptions of the primary sample containers and any necessary additives;
■ c- instructions for:
■ 1- completion of request form or electronic request,
■ 2- type and amount of the primary sample to be collected,
■ 3- special timing of collection, if required,
■ 4- any special handling needs between time of collection and time received by the
laboratory (transport requirements, refrigeration, warming, immediate delivery, etc.),
■ 5- labelling of primary samples,
■ 6- clinical information (e.g. history of administration of drugs),
■ 7- positive identification, in detail, of the patient from whom a primary sample is collected,
■ 8- recording the identity of the person collecting the primary sample,
■ 9- safe disposal of materials used in the collection;
■ d- instructions for:
■ 1- storage of examined samples,
■ 2- time limits for requesting additional examinations,
■ 3- additional examinations,
■ 4- repeat examination due to analytical failure or further examinations of same primary
sample.
Refer to the Specifications > Limitations chapter of the user manual (RAB237) of the
ABX Micros ES 60.
5. Analytical Procedures
Calibration
The controls and calibrators associated with the ABX Micros ES 60, supplied by HORIBA Medical,
comply with the requirements of the standard ISO 17511 : 2003, relating to the metrological
traceability of values attributed to calibration agents and control equipment (in vitro diagnostic
medical devices) for the measurement of quantities in biological samples. This makes it possible to
guarantee that the results obtained by the ABX Micros ES 60, duly controlled and calibrated, are
sufficiently exact to enable a clinically correct interpretation and comparability over time and space.
HORIBA Medical Hematology Control and Calibrator values, manufactured by R&D Systems, Inc are
traceable to standard reference methods.
Hematology analyzers in R&D Systems’ Quality Assurance Laboratory are whole blood calibrated to
values obtained using the following standard reference methods. Whole blood samples drawn from
normal, healthy donors are collected in EDTA anticoagulant and analyzed within six hours of
collection.
The White Blood Cell (WBC) and Red Blood Cell (RBC) are analyzed on a Coulter Counter Z series
instrument. All counts are corrected for coincidence.
Hemoglobin is measured using the Clinical Laboratory Standards Institute (CLSI) recommended
reagent for the hemoglobincyanide (cyanmethemoglobin) method. Readings are made at 540 nm in a
colorimeter/spectrophotometer calibrated according to CLSI H15-A3 and ICSH recommendations 7 .
The hematocrit (packed cell volume) is measured using plain glass microhematocrit tubes (not
coated with anticoagulant) centrifuged for 5 minutes in a microhematocrit centrifuge according to the
CLSI H7-A3 document 8 . No correction is made for trapped plasma.
Platelets are assayed using a hemocytometer and phase contrast optics.
All brands and products are trademarks or registered trademarks of their respective owners.
7 National Committee for Clinical Laboratory Standards (now Clinical Laboratory Standards Institute.) Reference and Selected
Procedures for the Quantitative Determination of Hemoglobin in Blood: Approved Standard-Third Edition. NCCLS document H15-
A3. Wayne, PA: NCCLS, 2000.
8 National Committee for Clinical Laboratory Standards (now Clinical Laboratory Standards Institute.) Procedure for Determining
Packed Cell Volume by the Microhematocrit Method: Approved Standard, NCCLS document H7-A3. NCCLS, Wayne, PA: NCCLS,
2001.
Primary reference
measurement
procedure
ICSH reference
method for the
enumeration of
RBC and WBC*
Manufacturer's
working
calibrator
Fresh human blood
Manufacturer's
standing measurement
procedure
Manufacturer's
product
calibrator
ABX Minocal
End user's
routine measurement
procedure
Routine sample
EDTA whole blood
Result
RBC and WBC count
* The White Blood Cell (WBC) and Red Blood Cell (RBC) are analyzed on a Coulter Counter Z series
instrument. All counts are corrected for coincidence.
Primary reference
measurement
procedure
ICSH reference
method for the
measurement of
Hemoglobin*
Manufacturer's
working
calibrator
Fresh human blood
Manufacturer's
standing measurement
procedure
Manufacturer's
product
calibrator
ABX Minocal
End user's
routine measurement
procedure
Routine sample
EDTA whole blood
Result
Hemoglobin
* Hemoglobin is measured using the Clinical Laboratory Standards Institute (CLSI) recommended
reagent for the hemoglobincyanide (cyanmethemoglobin) method(1). Readings are made at 540 nm in
a colorimeter/spectrophotometer calibrated according to CLSI H15-A3 and ICSH recommendations 9 .
9 National Committee for Clinical Laboratory Standards (now Clinical Laboratory Standards Institute.) Reference and Selected
Procedures for the Quantitative Determination of Hemoglobin in Blood: Approved Standard-Third Edition. NCCLS document H15-
A3. Wayne, PA: NCCLS, 2000.
Primary reference
measurement
procedure
ICSH reference
method for the
measurement of
Hematocrit*
Manufacturer's
working
calibrator
Fresh human blood
Manufacturer's
standing measurement
procedure
Manufacturer's
product
calibrator
ABX Minocal
End user's
routine measurement
procedure
Routine sample
EDTA whole blood
Result
Hematocrit
* The hematocrit (packed cell volume) is measured using plain glass microhematocrit tubes (not
coated with anticoagulant) centrifuged for five minutes in a microhematocrit centrifuge according to
the CLSI H7-A3 document 10 . No correction is made for trapped plasma.
10 National Committee for Clinical Laboratory Standards (now Clinical Laboratory Standards Institute.) Procedure for Determining
Packed Cell Volume by the Microhematocrit Method: Approved Standard, NCCLS document H7-A3. NCCLS, Wayne, PA: NCCLS,
2001.
Primary reference
measurement
procedure
ICSH reference
method for the
enumeration of
Platelets*
Manufacturer's
working
calibrator
Fresh human blood
Manufacturer's
standing measurement
procedure
Manufacturer's
product
calibrator
ABX Minocal
End user's
routine measurement
procedure
Routine sample
EDTA whole blood
Result
PLT count
7. Results Report
1. Glossary of Terms.................................................................................................................68
2. References...............................................................................................................................78
2.1. Regulatory References..............................................................................................................78
2.2. General Standardization References........................................................................................78
2.3. COFRAC-EA Documentation....................................................................................................79
2.4. Method Validation.....................................................................................................................79
2.5. Websites...................................................................................................................................80
1. Glossary of Terms
Accuracy
Ability of the instrument to agree with a predetermined reference value at any point within the
operating range; closeness of a result to the true (accepted) value.
Analyte
Component, substance, material to be measured in a possibly complex environment.
Analytical sensitivity
In compliance with the Common Technical Specifications (CTS), the «analytical sensitivity» refers to
the limit of detection, i.e. the smallest quantity of target marker that can be detected with precision.
Analytical specificity
The capacity of the method to determine only the target marker.
Background count
Measure of the amount of electrical or particle interference.
Calibration
Set of operations to establish, under specified conditions, the relationship between the values of the
quantity indicated by a measuring instrument or a measurement system or the values represented by
a materialized measurement or by a reference material, and the corresponding values of the quantity
given by standards.
Calibration factors
These are correction factors that the system uses to fine-tune instrument accuracy.
Calibrator
A (reference) material (e.g., solution, suspension) or device of known quantitative/qualitative
characteristics (e.g., concentration, activity, intensity, reactivity) used to calibrate, graduate, or adjust
a measurement procedure or to compare the response obtained with the response of a test
specimen/sample (NCCLS H38-P)
Carry-over
Amount of blood cells remaining in diluent following the cycling of a blood sample (in percent).
Cell control
Preparation made of human blood with stabilized cells and surrogate material used for daily
instrument quality control.
Characterization (evaluation)
Study To test the analysis protocol (analytical procedure) to determine the values of performance
criteria, which are determined beforehand. This study can be undertaken on an intra-laboratory (by
the supplier or manufacturer) or inter-laboratory basis.
COFRAC
«Comité français d'accréditation» (French Accreditation Committee)
Mission: To attest that the accredited bodies are competent and impartial, to obtain on an
international level the acceptance of their services and the recognition of expertise by laboratories,
and inspection and certification bodies.
Accreditation bodies affiliated to the multilateral agreement, ILAC (International Laboratory
Accreditation Cooperation):
Contaminant (Effect)
Undesirable effect, resulting from contamination. Most commonly, this is the effect exerted by a
serum on that which follows or precedes it. It may also arise from contaminating effects between
reagents.
Control
Substance used for monitoring the performance of an analytical process or instrument.
Correction
Value that is algebraically added to the raw result of a measurement to compensate for a systematic
error.
■ the correction is equal to the opposite of the estimated systematic error
■ since the systematic error cannot be precisely known, the compensation cannot be complete.
Correlation coefficient
Quotient of the covariance of two characteristics by the product of their standard-deviations. It
expresses the possible relationship between two variables that are known to be independent. Its value
must only be tested in comparison with zero according to a chosen risk. It is usually of no interest in
technical comparisons.
Coverage factor
Numerical factor used as a multiplier of the combined standard uncertainty to obtain the expanded
uncertainty.
Default setting
Original factory setting.
Deviation
Value minus its reference value.
Drift
Slow variation over time of a metrological characteristic of a measuring instrument.
Error
Result of a measurement minus a true value of the measurand (Bias).
Exactitude (Precision)
Closeness of the agreement between the result of a measurement and the true value of the
measurand.
Expanded uncertainty
Quantity defining an interval about the result of a measurement that may be expected to encompass a
large fraction of the distribution of values that could reasonably be attributed to the measurand.
■ The fraction may be viewed as the probability or level of confidence of the interval.
■ The association of a specific level of confidence with the interval defined by the expanded
uncertainty requires explicit or implicit assumptions regarding the probability distribution
characterized by the measurement result and its combined standard uncertainty. The level of
confidence that may be attributed to this interval can be known only to the extent to which such
assumptions may be justified.
■ The expanded uncertainty is sometimes known as the overall uncertainty.
Femtoliter (fL)
One quadrillionth (10-15) of a liter.
Grading
Material positioning of each marker (possibly of certain principal markers only) of a measurement
instrument according to the value of the measurand.
L.I.S.
Laboratory Information System
Lot number
Manufacturer’s code that identifies a batch of product (either reagents, controls or calibrators).
Matrix
Environment in which the analyte is found.
Mean, m
The sum of observations divided by their number. Unless otherwise indicated, the term “mean”
designates the arithmetic value.
Measurand
Specific quantity subjected to measurement.
Measurement
A series of operations whose aim is to determine a value of a quantity.
Noise
Corresponds to random variations of the measurement signal for a given level. It is measured by the
standard deviation of a series of at least 30 measurements of the signal, at the level in question.
Operating range
Range of results over which the instrument displays, prints and transmits data.
Parameter
Component of blood that the instrument measures and reports.
Performance criteria
Parameters characterizing the analytical procedure (linearity, repeatability, trueness, etc.)
Platelet concentrate
Labile blood product, composed of platelets, produced by blood bank centers and intended for
transfusion.
Random error
Result of a measurement minus the mean of an infinite number of measurements of the same
measurand, undertaken under conditions of repeatability.
■ the random error is equal to the error minus the systematic error.
■ since one can only perform a finite number of measurements, it is only possible to determine an
estimation of the random error.
Reagent blank
Corresponds to the signal resulting from the reagent(s) used during an assay or a measurement of
catalytic activity. The sample is replaced by an equal volume of an appropriate solvent.
Reference values
Results obtained for a given component in a reference population whose individuals are exempt from
disease or treatments that may alter their values. The reference values may vary, notably according to
the geographic origin, sex, and age of individuals. They are usually expressed as a function of lower
and upper limits that have been determined via statistical studies. They may be established by the
biologist, according to the analytical techniques used, or possibly verified when data from scientific
publications is used. The expression «reference value» is preferable to «usual value» or «normal
value».
Reliability (Precision)
Aptitude of a measuring instrument to give very similar indications during the repeated application of
the same measurand under the same measurement conditions.
Repeatability
Closeness of the agreement between the results of successive measurements of the same
measurand, measurements undertaken entirely in the same conditions of measurement.
Reproducibilty
Closeness of the agreement between the results of measurements of the same measurand,
measurements undertaken under a variety of measurement conditions.
Result of a measurement
Value attributed to a measurand, obtained by measurement.
Robustness
The “robustness” of an analysis technique implies its capacity to be unaffected by small, but
deliberate variations in parameters of the method, and it gives an indication about its reliability under
normal conditions of use.
Sensitivity (diagnostic)
The probability that a device gives a positive result in the presence of a target marker.
Sensitivity of a technique
Relationship between the variation of the measured signal to the concentration unit of the analyte
being studied.
Shutdown cycle
Cleans the instrument’s fluidic lines and apertures to help prevent residue build-up.
Specificity (diagnostic)
The probability that a device gives a negative result in the absence of the target marker.
Specimen
To avoid any confusion with the term sample (in the following context: group of individuals from a
population), it is preferable to use the term specimen to designate a biological sample (blood
specimen, urine specimen, etc.)
Specimen blank
Signal resulting from certain properties of the environment in which the analyte is found. Ideally, it
results from a signal measurement undertaken under the conditions of the reaction on the sample that
does not contain the analyte, or on the sample following elimination or inactivation of the analyte.
Standard
Materialized measurement, measuring apparatus, reference material or measurement system
designed to define, undertake, store, or reproduce a unit or one or several values of a quantity to
serve as a reference.
Standard uncertainty
Uncertainty of the result of a measurement expressed as a standard deviation.
Startup cycle
Ensures that the instrument is ready to run; includes performing a background test.
Systematic error
Mean that would be obtained from an infinite number of measurements of a same measurand,
undertaken under the conditions of repeatability, minus a true value of the measurand.
■ the systematic error is equal to the error minus the random error.
■ as with the true value, the systematic error and its causes cannot be fully known.
■ for a measuring instrument, see "trueness error".
Traceability (VIM)
Property of the result of a measurement or a standard such that it can be linked to determined
references, generally national or international standards, via the intermediary of an uninterrupted chain
of comparisons, all with determined uncertainties.
■ This concept is often expressed by the adjective «traceable».
■ The uninterrupted chain of comparisons is known as the chain of standardization or chain of
linkage of standards.
■ The manner in which the linkage to the standards is performed is known as the link to standards.
Traceability (ISO 9000 : 2000): ability to find the history, implementation, or location of that which
is being examined.
Trueness
Aptitude of a measuring instrument to give results that are exempt from systematic error.
Uncertainty
Parameter associated with the result of a measurand that characterizes the dispersion of values that
could reasonably by attributed to the measurand.
Validation technique
A validation technique is a technique that, following exhaustive study (literary and experimental
research), is designated for use as a reference of exactitude, for comparative assays or titrations of
control sera. It is chosen from amongst the reference techniques or selected by national or
international companies, or by consensus. It has acceptable and recognized levels of precision and
practicability. It should be described in detail, tested in several laboratories, and include the definition
of the equipment, controls required on the apparatus, the description of the various reagents and
specifications, the conditions of storage and use of the reagents, the operating procedure, the
calibration or standardization method, the nature of the calibrators, and the control method.
Whole blood
Non-diluted blood (blood and anticoagulant only).
2. References
■ Arrêté du 26 novembre 1999 relatif à la bonne exécution des analyses de biologie médicale.
(GBEA) J.O. Numéro 287 du 11 décembre 1999, page 18441 (NOR : MESP9923609A)
■ Directive 98/79/CE du Parlement européen et du Conseil, du 27 octobre 1998, relative aux
dispositifs médicaux de diagnostic in vitro (http://www.lne.fr/publications/directives/98-79.pdf)
■ Décret n° 2004-108 du 4 février 2004 relatif aux dispositifs médicaux de diagnostic in vitro et
modifiant le code de la santé publique (deuxième partie : Décrets en Conseil d'Etat): J.O. du 6
février 2004, page 2577, NOR: SANP0324626D
■ Essential Criteria for Quality Systems of Medical Laboratories (Eur J Clin Chem Clin Biochem
1997; 35(2): 123-132 © 1997) Walter de Gruyter Berlin New York
■ CLIA (Clinical Laboratory Improvement Amendments) regulation (http://www.westgard.com/
cliafinalrule4.htm)
■ The Fitness for Purpose of Analytical Methods: A Laboratory Guide to Method Validation and
Related Topics. Eurachem Guides (1998).
■ A WHO guide to good manufacturing practice requirements - Part 2: Validation. World Health
Organisation (1997).
■ Analyses de biologie médicale : spécification et normes d'acceptabilité à l'usage de la validation
des techniques. A. Vassault, D. Grafmeyer, J. de Graeve, R. Cohen, A. Beaudonnet, J. Bienvenu.
Ann Biol Clin 1999, 57 : 685-95.
■ AIDE MEMOIRE D’HEMATOLOGIE, 1998. Prof. C. Sultan / M. Gouault - Helman / M. Imbert,
Service Central d’Hématologie de l’Hôpital Henri Mondor, Faculté de médecine de Créteil (Paris
XII).
■ Protocole de validation de techniques (Protocole Valtec) - Document B. Ann. Biol. Clin 44, 686-
745. Vassault A., Grafmeyer D., Naudin C., Dumont G., Bailly M., Henny J., Gerhardt MF., Georges
P. et les membres de la commission de Validation de techniques de la SFBC (1986).
■ Dictionnaire des termes à l'usage de la validation des techniques, Commission validation de
technique SFBC, Ann Biol Clin 1986, 44: 679-85.
■ Guidelines for the Evaluation of blood cell analyzers including those used for differential leukocyte
and reticulocyte counting and cell marker applications. International Council for Standardization in
Hematology; Clin. Lab. Haemat. 1994, 16, 157-174.
■ Guidelines for the evaluation of analyzers in clinical chemistry ECCLS Document 3 n°2. Haeckel R,
Busch EW, Jennings RD, Truchaud A (1986) Beuth Verlag Berlin.
■ Proposed quality specifications for the acceptability of analytical systems for clinical chemistry.
Fraser CG, Hyltoft Petersen P, Ricos C, Haeckel R. Eur J Clin Chem Clin Biochem 1992, 30 : 311-
317.
■ Evaluation de l'accord entre trois automates d'hématologie. F. Hennouchi, S. Anselme Martin, V.
Chanteperdrix, N. Roux Bouisson, B. Polack, P. Mossuz. Ann Biol Clin 2002, 60 : 351-5.
■ Lecture critique de l'hémogramme : valeurs seuils à reconnaître comme probablement
pathologiques et principales variations non pathologiques – ANAES – http://www.has-sante.fr/
portail/jcms/c_271914
■ Age-based Reference Intervals from data is the third National Health and Nutrition Examination
Survey (NHANES III) - http://www.dgrhoads.com/refint/index.htm
■ DESIRABLE SPECIFICATIONS FOR TOTAL ERROR, IMPRECISION, AND BIAS, DERIVED FROM
BIOLOGIC VARIATION by Ricos C, Alvarez V, Cava F, Garcia-Lario JV, Hernandez A, Jimenez CV,
Minchinela J, Perich C, Simon M. "Current databases on biologic variation: pros, cons and
progress." Scand J Clin Lab Invest 1999;59:491-500 (mis à jour en 2008) - http://
www.westgard.com/biodatabase1.htm
■ Defining, Establishing, and Verifying Reference Intervals in the Clinical Laboratory; Approved
Guideline ; CLSI Document C28-A3 ((ISBN 1-56238-682-4)), 2008.
■ Assessment of the Clinical Accuracy of Laboratory Tests Using Receiver Operating Characteristic
(ROC) Plots; Approved Guideline, CLSI document GP10 (ISBN 1-56238-285-3) 1995.
■ Calibration and Quality Control of Automated Hematology Analyzers; CLSI Document H38-P (ISBN
1-56238-398-1), 1999.
■ Evaluation of Precision Performance of Clinical Chemistry Devices; Approved Guideline, CLSI
document EP5-A2 (ISBN 1-56238-542-9) 2004.
■ Validation, Verification, and Quality Assurance of Automated Hematology Anaysers; Approved
Standard ; CLSI Document H26-A2 (ISBN 1-56238-728-6), 2010.
■ Protocols for Determination of Limits of Detection and Limits of Quantification; Approved
Guideline, CLSI document EP17-A I(SBN 1-56238-551-8) 2004.
■ Method Comparison and Bias Estimation Using Patient Samples; Approved Guideline ; CLSI
Document EP9-A2 ((ISBN 1-56238-472-4)), 2002.
■ Reference Leukocyte (WBC) Differential count (Proportional) and Evaluation of Instrumental
Methods; Approved Guideline ; CLSI Document H20-A2 ((ISBN 1-56238-628-X)), 2007.
■ Procedures for the Handling and Processing of Blood Specimens; Approved Guideline; NCCLS
Document H18-A4 (ISBN 1-56238-724-3), 2010.
■ User protocol for Evaluation of Qualitative Test Performance; Approved Guideline - Second
Edition, Approved Guideline ; CLSI Document EP12-A2 ((ISBN 1-56238-654-9)), 2008.
■ Evaluation of the Linearity of Quantitative measurement procedures: a statistical Approach;
Approved Guideline, NCCLS document EP6-A I(SBN 1-56238-498-8) 2003.
■ Interference Testing in Clinical Chemistry: Approved Guideline ; NCCLS Document EP7-A2 ((ISBN
1-56238-584-4)), 2005.
■ ICSH guidance (Clin. Lab. Haemat.1994, 16, 157-174).
■ H20-A2 : Reference Leukocyte (WBC) Differential Count (Proportional) and Evaluation of
Instrument Methods, January 2007.
2.5. Websites
Q
Q.H.S.E. policy, 16
QCP, 32
Quality Management System, 8
Quality of Analytical Procedures, 59
Quality Records, 13
R
RBC
Inclusions, 43
Red blood cells agglutination, 43
Reference Values, 46
References
COFRAC-EA Documentation, 79
General Standardization, 78
Regulatory, 78
Websites, 80
Repeatability, 29
Reproducibility, 28
Results Report, 65
Review of Contracts, 10
Robustness, 39
S
Sigma Calculation, 34
Small erythrocytes, 44
Specimen for Sampling, 57
Stability, 39
Storage Conditions, 23
Supplies, 11
Symbols definition
Caution, 6
Danger, 6
Nota, 6
T
Technical Records, 13
Technical Requirements, 0
Turbidity, 42
U
Uncertainty Calculation, 32
Unlysed RBC, 41
Use of Computers, 56
W
Waste Disposal, 27
Westgard, 34