This document has been designed to be a guideline from version 3.1 of BA400 User Software
BA400 EUG v 2.0 en
First Steps
1. Check levels of Washing Solution ( 25 mL AC16434 / 5L distilled water ) and High Concentration Waste . Bear in mind to confirm
your actions by pressing ‘the bottle’ button when you have filled up the container. This button resets the scales.
2. You must go to Tools → Change reaction rotor (only in ‘standby’ mode) in order to swap it by another new one. Press ‘the up
arrow’ button to release the washing station from the reaction rotor cover. Press ‘the gear wheel’ button in order to come back
at the initial conditions. The program will execute a centering and adjusting of optical reading.
3. Check all the covers are correctly closed (Reaction and reagent rotor covers are critical !)
4. Connection with the PC is possible when the beep sounds. Bear in mind to press ‘the circled double arrow’ button to restore
communications or when you have entered the user software before to switch on the analyser.
5. Run a conditioning (the ‘blue drop’ button) from the menu Tools daily and every time you have filled up the Washing Solution
container.
6. Product reference (5412/5408) is necessary to perform manual pre dilutions of urine samples in a 1:10 ratio for ISE tests.
7. Automatic ISE washing cycle is required every time you have run at least 50 ISE preparations. It’s recommended you request an
Automatic washing cycle by yourself (Tools → ISE Utilities → General → Wash) if software does not request it by itself.
This document has been designed to be a guideline from version 3.1 of BA400 User Software
BA400 EUG v 2.0 en
Usual Routine Workflow
This document has been designed to be a guideline from version 3.1 of BA400 User Software
BA400 EUG v 2.0 en
Calibrators Programming
Name
Lot
Programming Calibrators Calibrator by test - sample
Expiry
Nº
This document has been designed to be a guideline from version 3.1 of BA400 User Software
BA400 EUG v 2.0 en
Controls Programming
Name
Level*
Programming Controls Lot number Tests
Sample type
Expiry
Tests
Insert minimum and maximum margins for every test
Selection
This document has been designed to be a guideline from version 3.1 of BA400 User Software
BA400 EUG v 2.0 en
MANUAL working mode
Tests
Select all these tests to
run a blank *
or
You are able to reorder the tests and their availability by
clicking from Programming→ Tests
This document has been designed to be a guideline from version 3.1 of BA400 User Software
BA400 EUG v 2.0 en
MANUAL working mode
Calibrations request
Tests
Select all the tests to
run
or
This document has been designed to be a guideline from version 3.1 of BA400 User Software
BA400 EUG v 2.0 en
MANUAL working mode
Controls request
Tests
Select all the tests to
run
or
This document has been designed to be a guideline from version 3.1 of BA400 User Software
BA400 EUG v 2.0 en
Storing a work session list
This document has been designed to be a guideline from version 3.1 of BA400 User Software
BA400 EUG v 2.0 en
MANUAL working mode
Sample request
or
This document has been designed to be a guideline from version 3.1 of BA400 User Software
BA400 EUG v 2.0 en
AUTOMATIC Working mode (connected to LIS)
This document has been designed to be a guideline from version 3.1 of BA400 User Software
BA400 EUG v 2.0 en
The instrument repeat automatically the analysis when the result is over limit of
linearity, below limit of detection or, in case of kinetics, when there is an alarm for
Substrate Depletion.
This document has been designed to be a guideline from version 3.1 of BA400 User Software
BA400 EUG v 2.0 en
ISE Utilities
Tools ISE Utilities Installation/ Activation: Execute all the steps in the order of appearance (only
when the module is installed for the first time)
This document has been designed to be a guideline from version 3.1 of BA400 User Software
BA400 EUG v 2.0 en
Replace reagent pack: Execute all the steps in order of appearance in order
to swap the reagent pack by another new one.
Trigger button
Replace electrodes: Execute all the steps in order of appearance in order to
swap one or more than one electrode. This menu must be followed in case
you may subtract one electrode for maintenance purposes.
Replace pump tubes: Execute all the steps in order of appearance in order
to change external pumps tubing.
ISE Alarms: All pending requirements appear when this function is triggered.
In case all is OK message ISE Ready appears.
This document has been designed to be a guideline from version 3.1 of BA400 User Software
BA400 EUG v 2.0 en