Operating
Instructions
for PMH5000
Respiratory
Gas Humidifier

1275
Technical data 1275

Apparatus type protection class I


Verify compliance of mains voltage against operating voltage
indicated on the nameplate before starting operation.

Operating voltage 220 V ~ +10% -15% or


230 V ~ +10% -15%
Mains frequency 50/60 Hz
Power consumption 160 VA max. (heating wires off)
190 VA max. (with one heating wire)
220 VA max. (with two heating wires)
Current consumption max. 1.0 A at 220 V
max. 0.9 A at 230 V
Heating, respiratory hoses 22 V ~, 60 W, 50/60 Hz
Respiratory gas flow Recommended respiratory gas flow
3 to 60 liters/min for adults
Humidity More than 33 mg/liter respiratory gas
at a flow rate of 3 to 60 liters/min
Dimensions Height: 142 mm
Width: 140 mm
Depth: 190 mm
Weight ca. 3.0 kg
Ambient temperature In operation +10°C bis +40°C
In storage - 20°C bis +60°C

Legend
Symbols – Respiratory Gas Humidifier PMH5000

! ! ! ! IPX
! !1
IPX 1IPX 1 HOT IPX
HOT1
IPX IPX
1 HOT
HOT 1
O I /IiO
I / OI / HOT HOT
I /iOI / O
/O ii i i
Type B Attention, Protected Caution! On / Off Attention,
observe against Temperature may observe
documentation! dripping water be higher than 85°C documentation!


Displays
01. Measured temperature “actual temperature”.
Displayed is the close-to-the-patient
temperature (changeover switch not pressed,
see function button 06) temperature
of humidifier chamber (changeover switch
pressed, see 06).
02. Set temperature “setpoint temperature”.
Displayed is the desired close-to-the-patient
temperature (changeover switch not pressed,
see 06) or the desired temperature of the
humidifier chamber (changeover switch
pressed, see 06).
03. Actual humidity situation
too dry – optimum – too humid.
04. When LED lights up, it is an indication
that selected setpoint temperature is
too low for intubated patients.

Function buttons
05. Changing of setpoint temperature
06. Changeover switch: from close-to-the-patient
temperature to humidifier chamber temperature
07. Alarm silent button:
Pressing short: audible alarm is
Switched off for 120 seconds
Pressing long: Switching therapy pause
on or off (display 01. “OFF”)
08. Heating wire On/Off with pilot lamp

Alarm indications
Temperature too high: left indicator lights up
Temperature too low: right indicator lights up
09. Close-to-patient temperature too high / low
10. Chamber temperature too high / low
11. Temperature sensor is missing or defect
12. Heating wire is missing or defect
3
Connections right

ON / OFF switch
Retainer basic unit

Connections left

TEMP: Temperature probe


INSP: Heating wire “inspiratory”
EXP: Heating wire “expiratory”
POWER: Mains connection cable

4
Setting up and starting operation
of respiratory gas humidifier PMH5000
1 Respiratory gas humidifier must be Plug in both temperature probe
clean and dry. sensors into the openings provided at
2 Fasten respiratory gas humidifier the heating wire adapter above the
horizontally and safe, e.g. by means of humidifier chamber and at the end of
the existing retainer. Respiratory gas the patient hose system.
humidifier must be mounted below 9 Check all connections for proper
patient‘s level! and tight fit.
3 Attach humidifier chamber onto 10 Switch on therapy apparatus first,
respiratory gas humidifier and lock then respiratory gas humidifier.
till you can hear it (Figure 1). Before switching on respiratory
4 Fill humidifier chamber at the most up gas humidifier, make sure that
to maximum level with sterile water. respiratory gas flows freely up to
5 Connect therapy apparatus with inlet the outlet.
pipe-end of humidifier chamber 11 Preset close-to-patient temperature
(short hose). at the respiratory gas humidifier:
6 In general, the operating instructions Press + or – and wait until the
for each respective humidifier chamber displayed setpoint temperature
shall be observed. changes. Press + or – repeatedly if
7 Connect plugs for heaters at the patient necessary (Figure 3).
hoses with the respective connection 12 Preset humidifier chamber temperature
cables (Figure 2). Make sure not to at the respiratory gas humidifier:
interchange inspiratory and expiratory Holding changeover switch depressed
side! Disposable hose systems are (Figure 4), simultaneously press + or –.
connected via an additional electric Press + or – repeatedly if necessary.
adapter (“Dispo adapter”). 13 The respiratory gas humidifier
8 Connect temperature measuring probe requires a warm-up period of 20 - 30
with the connection cable provided minutes until the preset temperature
at the respiratory gas humidifier is reached.
(Figure 2).

5
Figure 1 Figure 2 Figure 3 Figure 4
PMH5000 Respiratory Gas Humidifier
The PMH5000 warms and humidifies respiratory gas that is supplied to the patient.
The PMH5000 is applicable in combination with respirators, bi-level and CPAP devices
for mask respiration as well as for intubated patients.

General notes

Starting operation This apparatus must only be


Humidifiers are subject to specific safety implemented by authorized and instructed
precautions regarding electro-magnetic specialized staff according to the
compatibility (EMC) and need to be respective legal provisions in force.
installed and put in operation according
to the EMC instructions included in the Determining and/or changing of
technical descriptions. patient settings must be effected by a
physician or by persons commissioned
Interference by HF by a physician.
Portable and mobile high frequency
Communication devices may influence Notes in these operating instructions
the humidifier. For details please refer do not replace any medical or official
to the technical description. instructions of patients care. The “best
solution” decreed by a physicians
Operation community or applied by a health care
Respiratory gas humidifiers and institution shall take priority over these
humidifier chambers are not inhaler operating instructions.
devices and therefore, not suitable for
the application of medicaments!

Also when heating wires are switched


on, condensation water may form inside
the patient hose system during operation.

6
Cleaning Disturbances
Do not sterilize basic unit or immerse it In case of disturbances, the respiratory
in water! Do not sterilize temperature gas humidifier must be disconnected from
measuring probe! Basic unit and the mains and the authorized repair
temperature measuring probe can be service shall be notified.
cleaned with mild a cleaning agent such
as, for example, WILAsil. Disinfection is Accessories
achieved by wiping using an aldehyde- For operating the PMH5000 at least the
free, alcohol-based quick-disinfection following will additionally be needed:
preparation. For detailed instructions on One or two respiratory hose(s)
cleaning and care of the apparatus please coordinated with the length of the heating
refer to the technical description. wire, e.g. 120 cm (Order No. 270.146)

Maintenance A humidifier chamber, e.g. PMC500


Any intrusion on the apparatus – reusable (Order No. 550.183)
also maintenance, tasting, calibration –
must only be performed by trained A wide selection of respiratory hoses
service technicians. For details please and hose systems for one-time use as
refer to the technical description. well as for multiple application you‘ll find
in the WILAmed product catalog.

Waste disposal
When disposing basic device and/or
accessories, the respective regulations
in force shall be observed.

7
Warnings
Danger of electric shock! Danger of burns! Heating plate can
Operate respiratory gas humidifier on reach temperatures of more than 85°C.
an outlet with properly functioning Do not touch heating plate while in
ground wire only! operation!

Potential explosion risk! Temperature of respiratory gas


Do not use near flammable anesthetics! administered may exceed 41°C if
Make sure no anesthesia agent temperature sensor is improperly
dissociation takes place! installed and connected.

Always position respiratory gas Make sure therapy apparatus


humidifier below therapy apparatus and functions properly and therapy
below patient‘s level! parameters are preset correctly
before connecting patient hose
Operating instructions for system with the patient.
accessories must be observed!
Before connecting patient hose
Any use of humidifier chambers, system with the patient, respiratory gas
hose systems, and accessories that were must flow through the humidifier chamber.
not released by WILAmed may derogate
the operational safety of the system. Danger of burns! Do not cover up
respiratory hose system with insulating
Water for filling humidifiers chambers materials (e.g. cloths, blankets). Hose
should not be warmer than 37°C! system must not touch the patient.

Fill humidifier chamber up to the A minimum flow of respiratory gas


maximum filling mark at the most. inside the hose system is necessary
When filling level is exceeded, water at all times to avoid overheating.
from the humidifier chamber could swash
over to the patient hose system.

8
The temperature of respiratory gas
administered must be checked in
regular intervals. When temperature of
respiratory gas administered is too high,
burns of the patient‘s respiratory tract
may occur.

When administering of respiratory


gas is interrupted, switch off apparatus
or – when reasonable – switch it to
“therapy pause” mode.

Pull out mains plug before any


maintenance or inspection work and
before any checking because of alarms
at the respiratory gas humidifier.

When apparatus is operated outside


the specified ambient temperature range,
its humidifying capacity may be impaired.

9
Alarms and indications
Indication Condition Possible cause Measures
Alarm – respiratory Respiratory gas 1) Sudden change Mute alarm and
gas temperature too warmer than 41°C of flow rate observe, whether
high or Respiratory gas at temperature
least more than 2°C goes down to a
above setpoint value permissible value
2) Temperature Remove hose system
sensor defect and exchange
temperature sensor

3) Respiratory gas Remove hose


humidifier defect system and notify
repair service

Alarm – respiratory Respiratory gas 1) Sudden change Mute alarm and


gas temperature too temperature at least of flow rate observe, whether
low more than 4°C below temperature
setpoint value increases to a
permissible value
2) Temperature Remove hose system
sensor defect and exchange
temperature sensor

3) Respiratory gas Remove hose


humidifier defect system and notify
repair service

4) Humidifier Insert humidifier


chamber incorrectly chamber correctly
inserted

10
Indication Condition Possible cause Measures
Alarm – temperature Temperature of 1) Sudden change Mute alarm and
of humidifier chamber humidifier chamber: of flow rate observe, whether
too high - for 20 minutes 4°C temperature
above setpoint goes down to a
value permissible value
- at least 10°C 2) Temperature Remove hose system
above setpoint sensor defect and exchange
value temperature sensor

3) Respiratory gas Remove hose


humidifier defect system and notify
repair service

Alarm – temperature Temperature of 1) Sudden change Mute alarm and


of humidifier chamber humidifier chamber: of flow rate observe, whether
too low - for 20 minutes 4°C temperature
below setpoint value increases to a
- at least 10°C permissible value
below setpoint value 2) Temperature Remove hose system
sensor defect and exchange
temperature sensor

3) Respiratory gas Remove hose


humidifier defect system and notify
repair service

4) Humidifier Insert humidifier


chamber incorrectly chamber correctly
inserted

11
Indication Condition Possible cause Measures
Alarm – heating wire Fault in wiring of 1) Heating wire Connect heating wire
heating wire not connected
2) Heating wire Replace heating
defect wire
3) Leads to heating Notify repair service
wire defect
4) Fuse defect Notify repair service
Alarm – temperature Temperature sensor 1) Temperature Connect temperature
sensor interrupted or faulty sensor not connected sensor, switch
respiratory gas
humidifier off and
on again
2) Temperature Replace temperature
sensor defect sensor
Temperature display Error in starting the 1) Temperature Replace temperature
shows “EEE” apparatus sensor defect sensor.
2) Heating plate Notify repair service
defect
Temperature display Automatic switch- Reason for alarm was Switch off apparatus,
shows “000” off in case of alarm not eliminated within eliminate reason
for more than 10 10 minutes for alarm, switch on
minutes apparatus again

12
Indication Condition Possible cause Measures
Temperature display Therapy pause Therapy pause is Press mute alarm
shows “OFF” preset at the device button for more than
3 seconds to switch
off therapy pause

No indications at all Apparatus does not 1) Heating plate is Remove hose system
work or cannot be overheated and and notify repair
switched on temperature limit service
switch has switched
off
2) Fuses defect Remove hose
system and notify
repair service
3) Respiratory gas Remove hose
humidifier defect system and notify
repair service

13