You are on page 1of 8

Evita 4 edition

One of the best


keeps getting better

Emergency Care · Perioperative Care · Critical Care · Perinatal Care · Home Care Because you care
Even when there’s little room
for improvement …

European screen image

enhanced
The tradition of leadership continues To assure immediate recognition of the required
The highly dependable Evita 4 has been settings, the user-friendly display shows only the
recognized around the world as a leader in active control elements. For added convenience
ventilation performance. Now, the Evita 4 edition and efficiency, the robust touch screen combines
takes this popular ICU ventilator to yet another highly flexible operation with direct access to
level, combined with design refinements that rotary knobs and keys.
redefine user value.
Improved workflow efficiency
A more user-friendly interface The design of the patient part heating, expiration
Evita 4 edition’s ergonomic display offers block housing and flow sensor cover has been
exceptional viewing. It also speeds routine tasks optimized to reduce condensation in the patient
ranging from calibration to maintenance by system and expiration valve. As a result, Evita 4
providing ideal access to the hose system, edition extends flow sensor life and minimizes the
flow sensor and expiration valve. time clinicians spend checking and changing
The full-color screen provides a comprehensive sensors. In addition, a drawer has been integrated
customized overview of the patient’s ventilation so flow sensors, operating instructions or other
therapy, complete with configurable trends, frequently needed items are readily accessible at
loops, measured values, curves and logbook. all times.

How you benefit:


• First-class ventilator performance
• One ventilator cost-effectively
meets patient needs
• Enhancements improve workflow
and minimize training
… we still find ways
• CareArea™ Solution connectivity to make it better
streamlines data capture so staff
can focus on patient care
It takes a lot to meet all your ventilation needs …

advanced
One ventilator for every patient Evita 4 edition also offers improved Automatic
Evita 4 edition is designed to accommodate Tube Compensation (ATC™), an option that more
patients from neonate to adult. With a wide range precisely compensates for the tube-related work
of options, it allows you to handle ever-changing of breathing (WOB). This further improves
patient populations and situations with a single weaning by adapting to patient efforts, and
solution. For example, the NeoFlow™ option gives the feeling of virtual extubation.
offers flow measurement at the Y-piece for precise
volume and leak monitoring of neonates with NIV option helps recovery
accurate, responsive triggering based on patient Mask ventilation can complement the weaning
effort. process by reducing re-intubation rates or even
eliminating the need for intubation in the first
Excellent performance is standard place. Because Evita 4 edition accommodates
The Evita 4 edition allows respiratory loops to be both conventional and non-invasive ventilation,
displayed full screen, frozen and scrolled through it can help minimize the potential for infections,
with a cursor to show exact values. By pressing enhance recovery times and reduce length of
the reference key, each loop can be used as a stay compared to traditional therapies.
reference loop. In the case of external flow
sources, sensor tolerances can automatically be Due to an improved algorithm, non-invasive
increased to prevent nuisance alarms. ventilation is now even more responsive to
And every Evita 4 edition is equipped with variable leakages. Quicker leakage “learning”
powerful weaning parameters. Rapid Shallow and more responsive leakage compensation
Breathing index, Negative Inspiratory Force index further enhance the patient-ventilator interaction.
and VTASB all contribute to improved consistency
of successful weanings.

Improve gas exchange


Our “Room to Breathe” concept encourages
spontaneous breathing. PCV+/BIPAP™ – the
universal mode for ventilation and weaning in all
pressure-controlled modes – gives a patient the
freedom to breathe spontaneously at any time.
AutoFlow facilitates spontaneous breathing in all
volume-controlled modes. This optimizes weaning
by reducing the need for sedation and
invasiveness of ventilation.

… that’s why the Evita 4 edition is so complete.


More of what you look for …

A safe investment Less training required


The Evita 4 edition is based on a fundamental The continuity evident in the design of our user
guiding principle of Dräger Medical’s CareArea™ interfaces, even across different care areas,
Solutions… the concept of continuous innovation. means current Dräger Medical customers will
The device’s modular design ensures upgradeabil- already be familiar with the operation of the
ity and upward compatibility while the user inter- Evita 4 edition. Increased user productivity and
face and sensors guarantee a high degree of enhanced patient safety are two key advantages
familiarity and consistency. Add options, such as of Dräger Medical’s comprehensive and highly
ATC®, PPS, NIV and NeoFlow, as needed to integrated CareArea™ Solutions.
customize the system to specific departmental
needs. Then, easily upgrade to new features as Comprehensive support
they become available to ensure that the Evita 4 Dräger Medical provides a comprehensive
edition remains a sound investment over time. support system that ensures you and your staff
are able to optimize the use of your Evita 4
In addition, as part of Dräger Medical’s CareArea™ edition. By focusing on your needs, we provide
Solution for Critical Care, the Evita 4 edition can the tools to improve the quality of care
be integrated with leading-edge monitoring, infor- for your patients.
mation management, architectural systems and
smart accessories to help you steadily redefine
process efficiency.

Operating costs
As the world of medicine becomes increasingly
cost-conscious, the focus is shifting towards the
operating costs of medical equipment. High-
quality and long-lasting components characterize
this proven ICU ventilator platform, resulting in
lower operating costs.

proven
… to positively impact critical care
Technical data Evita 4 edition
Patient type Adults and childeren (bodyweight of at least 6.6 lbs. (3 kg))
With NeoFlow option: infants and and neonates (min. body
weight of 1.1.lbs (0.5 kg))
Ventilation settings
Ventilation mode • IPPV, IPPVAssist (CMV, CMVAssist)
• SIMV, SIMVASB (SIMV, SIMV/Psupp)
• MMV, MMVASB (MMV, MMV/Psupp)
• BIPAP1), BIPAP1)ASB, BIPAP1)Assist
(PCV+, PCV+/Psupp, PCV+Assist)
• APRV
• CPAP, CPAPASB (CPAP, CPAP/Psupp)
• ILV
• PPS (optional)*
Enhancements • AutoFlowTM – Automatic adaptation of inspiratory flow
in volume controlled modes
• ATCTM – Automatic Tube Compensation (optional)
• NIV – Mask Ventilation (optional)
Ventilation frequency (f) 0 to 100 /min, 0 to 150 /min (Neonatal)
Inspiration time (Tinsp) 0.1 to 10 s
Tidal volume (VT) (BTPS*) • 0.1 to 2.0 L (Adult)
• 0.02 to 0.3 L (Pediatric)
• 0.003 to 0.1 L (Neonatal)
Inspiratory flow • 6 to 120 L/min (Adult)
• 6 to 30 L/min (Pediatric and Neonatal)
Inspiratory pressure 0 to 80 mbar (cmH2O)
PEEP / intermittent PEEP 0 to 35 mbar (cmH2O) Europe, Middle East, Africa, Latin America,
Pressureassist (PASB) (Psupp) 0 to 80 mbar (cmH2O) Asia, Pacific:
Rise time for inspiratory pressure 0 to 2 s
O2 concentration 21 to 100 Vol.% Dräger Medical AG & Co. KG
Multi-sense Trigger Criteria Internal automatic pressure trigger, Flow, Volume Moislinger Allee 53–55
(Flow adjustable 0.3 to 15 L/min) 23542 Lübeck
Measured values displayed GERMANY
Airway pressure Peak pressure, plateau pressure, mean pressure, PEEP, Tel: +49-1805-3 72 34 37
min. pressure (0 to 99 mbar/cmH2O) +49-451-882-822
Minute volume (MV), (BTPS*) MV, MVspont, MVleak (0 to 99 L/min) Fax: +49-451-882-37 79
Tidal volume (VT), (BTPS*) Inspired VT, expired VT, VTPS (0 to 3999 mL)

90 50 818 / 03.07-6 / ls-dw / Printed in Germany / Chlorine-free - environmentally compatible / Subject to modifications / © 2007 Dräger Medical AG & Co. KG
E-mail: Business.Support@draeger.com
Breathing frequency (f) ftotal, fspon, fmand. (0 to 300/bpm)
O2 concentration (FiO2) Inspired O2 concentration (15 to 100 Vol.%)
Lung mechanics • Resistance ((0.0 to 600 mbar/L/s) (cmH2O/L/s)) USA:
• Compliance ((0.0 to 300 mL/mbar) (mL/cmH2O))
Breathing gas temperature 18° to 51°C Draeger Medical, Inc.
Waveforms Airway pressure-time, flow-time, volume-time, ... 3135 Quarry Road
Trends (8 anyone configurable) FiO2, MV, VT, f, PEEPi, R, C, etCO2, ... Telford, PA 18969
• •
Loops Paw-V, V-Flow, Flow-Paw, ... USA
Capnography (etCO2) (optional) • 0 to 100 mmHg Tel: +1-215-721-5400
CO2 production (VCO2) (optional) • 0 to 999 mL/min, STPD* Toll-free: +1-800-437-2437
Serial dead space Vds • 0 to 999 mL, BTPS*
Dead space ventilation (Vds/VT) • 0 to 99 % Fax: +1-215-723-5935
Alarms / Monitoring E-mail: info@draegermed.com
Airway pressure High / Low
Expired minute volume High / Low
Tidal volume High Canada:
Apnea alarm time 5 to 60 s
Spontaneous breath frequency High Draeger Medical Canada Inc.
Inspired O2 concentration High / Low 120 East Beaver Creek Road Suite 104
Breathing gas temperature High Richmond Hill Ontario L4B 4V1
etCO2 (optional) High / Low CANADA
Performance data Tel: +1-905-763-3702
Max. flow for pressure support Toll-free: +1-866-343-2273
and spontaneous breathing 180 L/min (adult), 60 L/min (pediatric) Fax: +1-905-763-1890
Valve response time T0...90 ≤ 5 ms E-mail: Canada.Support@draeger.com
Control principle Time cycled, volume constant, pressure-controlled
Safety relief valve 100 mbar (cmH2O)
Leakage compensation www.draeger.com
Hose system compensation
Outlet for pneumatic nebulizer
Operating data
Mains power connection 100 to 240 V, 50/60 Hz, 10 to 30 V DC (optional)
Power consumption Approx. 125 W
Gas supply operating pressure O2, air: 2.7 to 6 bar / 39 to 87 PSI
Physical Specifications Manufacturer:
Dimensions ventilator (W x H x D) 530 x 290 x 450 mm (20.9 x 11.4 x 17.7 in.) (without trolley) Dräger Medical AG & Co. KG
Diagonal screen size 10.4" TFT LCD color touch screen 23542 Lübeck, Germany
Weight basic unit Approx. 27 kg (59.5 lbs.)
Machine outputs The quality management system at
Digital output Output and reception via an RS 232 C interface Dräger Medical AG & Co. KG is
Digital output Output for independant lung ventilation (ILV) certified according to ISO 13485, ISO 9001
Digital output (optional) For output and reception via two RS 232 C interfaces and Annex II of Directive 93/42/EEC
Analog output (optional) For analog output of two measured values (Medical devices).
* PPS is not available in the US 1)
used under license
ATCTM, Trademarked by Dräger / AutoFlowTM, Trademarked by Dräger
BTPS*, Body Temperature Pressure Saturated
Measured values relating to the conditions of the patients lung, Body temperature 37°C, steam-saturated gas, ambient pressure
STPD* Standard Temperature, Pressure, Dry.
Measured values based on normal physical conditions: 0°C, 1013 hPa, dry