no Tender Specification BD - Alaris GH Plus Fresenius Agilia MC


Flow rate: It should have flow rate programmable
range at least from 0.1 to 1200 ml/hrs. This
1 YES YES
flexibility to choose flow over wide range is needed
for ICU patient clinical management
Syringe size accommodation: should work with all
2 YES YES
commonly available syringe from 5 ml-50 ml.
Only 3 occlusion setting
Occlusion setting range: should have minimum 11
available ( with preset range
3 occlusion alarm setting with selectable range from YES
setting between 50 - 900
0-1000 mmHg.
mmHG)
Profile programming and drug library: Should have
programmable delivery of up to 30 profile for easy
and quick start-up for different drugs. At least 3000 Only 19 profiles can be
4 YES
Drugs should be programmable with drug name programed
display in pump during infusion. The nurse has to
just enter weight and start infusion.
Drug Dose calculation mode/ weight based
5 setting: Drug dose calculation software should be YES YES
available to reduce calculation error.
Guardrail system/ Medication error prevention Vigilant - DRUG ERROR
software: The machine should program the drug REDUCTION software,
6 YES
with minimum and maximum drug delivery range recently had MAJOR FDA
to prevent medication error. recall
Purge Option: Programmable purge setting
7 (enabled/Disabled) should be there to give bolus YES YES
drug dosage while on flow.
Bolus: Bolus delivery rate 10-1200 ml/hr with
programmable hands on and hands free option
8 YES YES
should be there.Also should have dedicated bolus
key for instant bolus.
Event log: Minimum of 1500 events log and 200
9 YES YES
key presses log should get captured.
Hard ware safety: There should be integrated
10 piston brake within syringe drive to prevent free YES YES
flow during syringe change.
Back off safety feature: Once occlusion sensed, the
11 pump should stop infusing and piston backs off to YES YES
reduce in line pressure.
Pump Alarm feature: Alarm should be available for
the following - Occlusion drive disengaged, AC NO AC POWER FAILURE
12 YES
power failure, Internal Malfunction, Battery low, ALARM
Battery Empty, Near end of Infusion, VTBI done,
Display Screen: Should have an ergonomic and
brightly lit display which should be easy to read
13 YES SMALL SCREEN SIZE
critical data at a glance, clearly visible from a
distance in all lightning conditions.
Power supply: Should have detachable main
14 YES YES
adaptor from standalone operation.
Battery: Should have rechargeable battery pack
15 YES YES
with at least 6 Hrs. battery backup.
Connectivity and Data transfer: Should have IR
16 YES YES
port for data transfer.
Docking station (Stacker): Should be available in
17 YES YES
3,5,7,9 pumps docking facility.
Certification: European CE certification/FDA
18 YES, CE YES, CE
certificate/BIS/Should be available.