Professional Documents
Culture Documents
Illustration 1 g02451337
Typical rod linkage
(1) Rod end
(2) Nut
(3) Rod
(4) Body position sensor
(5) Center of the rod ends
The calibration procedure for body position sensor (4) must be performed when one or more of the following conditions occur:
Body position sensor (4) is located near the pivot pin for the body. A lever is attached between the sensor and the body. When the body is raised or
lowered, the lever rotates the sensor. This action changes the Pulse Width Modulation (PWM) signal that is sent to the "Chassis" ECM.
If the expected PWM signal does not occur, then the "Chassis" ECM generates an event code. The code is logged toVIMSTM.
Note: The recommended duty cycle output is within the range of 14 to 19 percent.
The following occurs when all the conditions are met for 12 seconds:
The calibration value will be the average of the values that were recorded during that 12 seconds.
Expected Values
When the bed is fully lowered, the "Chassis" ECM expects a value that is greater than 1 percent PWM.
If the expected PWM value is not obtained, then an FMI 04 will be logged. If this event occurs, then perform the "Rod Adjustment" in the "Body
Position Sensor - Adjust" section.
Tolerance Range
The standard range for the calibration limits is plus or minus 2 percent of the last calibration point. If the tolerance range is not exceeded, then the
lower calibration point will not change.
If the tolerance range is exceeded, then the upper calibration point will default to a 50 percent duty cycle until the upper calibration is performed.
The lower calibration point will change to the average value over the last 12 seconds.
This change occurs because there has been a mechanical change in the system that measures the output of the sensor. For example, a change
could occur under any of the following conditions:
Any of these conditions could affect the duty cycle of the full lower calibration.
The current full up duty cycle must be greater than the previous full up duty cycle.
Note: When the bed has reached the maximum height, the operator should move the hoist lever from the RAISE position to the HOLD position.
Five seconds after the hoist lever is in position the calibration will begin.
The following occurs when all the conditions are met for 500 milliseconds:
The calibration value will be the average of the values that were recorded during the 500 milliseconds.
If the current upper calibration value is less than the previous upper calibration value, then the value will not change.
Expected Values
When the bed is fully raised, the Chassis ECM expects a value of less than 96 percent PWM.
If the expected PWM value is not obtained, then an FMI 03 will be logged. If this event occurs, then perform the "Rod Adjustment" in the "Body
Position Sensor - Adjust" section.
Known Issues
The following issues could prevent calibration of the position sensor. The issues could cause sensor diagnostics to be logged.
If the body pins are worn, perform the "Rod Adjustment" in the "Body Position Sensor - Adjust". Verify that the angle of the body is less than 7
degrees when the body is fully lowered.
1. Adjust the "Hoist Lower Valve Adjustment Status" value with Cat® Electronic Technician (Cat® ET).
3. Continue to adjust the "Hoist Lower Valve Adjustment Status" value until the desired snubbing is obtained.
3. Twist the rod (3) to get a distance of 360 ± 3 mm (14.2 ± 0.1 inch) between the centers of the rod ends (5).
Angle Verification
Verify that the angle of the body is within tolerance when any of the following components are replaced:
A Sensor
A Body
An ECM
Note: If the angle is not within tolerance, then the "Chassis" ECM cannot correctly calibrate the position sensor.
Lower the body to the frame rails. If the value is less than 7 degrees, the software in the "Chassis" ECM calibrates the sensor. The calibration is
performed automatically by the "Chassis" ECM when the body is raised and lowered. If the value is not less than 7 degrees, then perform the "Rod
Adjustment" in the "Body Position Sensor - Adjust" section.
Illustration 2 g02451016
(6) Rod end
(7) Nut
(8) Rod
(9) Body position sensor
(10) Centers of the rod ends
Body position sensor (9) is located on the frame near the left pivot pin for the body. Rod (8) is connected between the body position sensor (9) and
the body. When the body is raised, rod (8) rotates body position sensor (9). This action changes the PWM signal that is sent to the "Chassis" ECM.
The adjustment of the rod between the sensor and the body is important. See the below "Body Position Sensor - Adjust" for the proper adjustment
procedure.
The "Chassis" ECM is constantly checking for the need to calibrate body position sensor (9). Body position sensor (9) is calibrated by the "Chassis"
ECM when all of the following conditions exist for 30 seconds:
The hoist control has been placed in the HOLD position before being placed in the FLOAT position.
The duty cycle from the position sensor for the body is stable.
The duty cycle from the position sensor for the body is within the specified range for the LOWER position.
The signal from the body position sensor is also used for the following items:
Signal a new load count. (The hoist control must be in the RAISE position for 10 seconds.)
Illuminate the dash lamp that indicates that the body is not against the frame.
Body position sensor (9) is used to limit the top gear while the body is UP. The "Chassis" ECM comes from the factory with the top gear value that
is set to FIRST gear. The transmission will not shift past FIRST gear until the body is DOWN. The top gear value is configurable and can be
changed using Cat ET). The top gear value can be changed to SECOND gear or THIRD gear.
The signal from body position sensor (9) is also used to control the SNUB position of the hoist control valve. The "Chassis" ECM will signal the
proportioning solenoids to move the hoist control valve to the SNUB position when the body nears the frame. The speed in FLOAT is reduced in
the SNUB position to prevent the body from contacting the frame too hard.
Known Issues
The following issues could prevent calibration of the position sensor. The issues could cause sensor diagnostics to be logged.
If the body pins are worn, perform the "Body Position Sensor - Adjust" procedure. Verify that the angle of the body is less than 7 degrees when the
body is fully lowered.
3. Continue to adjust the "HLVAS" value until the desired snubbing is obtained.
The distance between the centers of the rod ends (10) must be 423 ± 3 mm (16.7 ± 0.1 inch).
Rod Adjustment
Perform the following to adjust the rod:
3. Adjust rod end (6) to get a distance of 423 ± 3 mm (16.7 ± 0.1 inch) between the centers of the rod ends.
PSP-00043039
2021/05/21
13:21:50-05:00
i03881814
© 2021 Caterpillar Inc.
Caterpillar:
Confidential Green