HDI 1000 Calibration guide

Division of Sensor Technology


Humanray Co.,Ltd
2012.05.16
Uhm Jisook
Introduction
Structure of Sensor

Scintillator
Scintillator
X-ray image convert into the visible ray through scintillator.
CMOS
CMOS
HIC Imaging Sensor

Definition of a term

Term Description
Dark image Offset, without X-ray image
Bright image Gain, with X-ray image
Signal Signal = Gain – Offset [LSB]
Value of gray scale
LSB (Least Significant Bit)
ex.) 12bit = 1~4096 LSB

2012-05-17 2
Introduction
Serial No of each sensor

Each sensor has the unique serial number and the different calibration data.
Be careful about confusing the calibration data and ini file.

Calibration Step
Step 1. Sensor setting
Step 2. Copy & paste EzSensor.ini file
Step 3. Delete the previous
calibration data.
Step 4. Execute software.
And then Calibration software appear.
Step 5. S/W Setting
Step 6. Dark frame Acquisition
Step 7. Bright frame Acquisition
Step 8. Generate Bad pixels map
Step 9. Copy & paste calibration file
Step 10. Copy & paste EzSensor.ini file

2012-05-17 3
Preparation for calibration
Step 1. Sensor setting
(1) Distance between X-ray cone and sensor : 8cm
※ Distance is depend on capability of X-ray generator. So, refer 'Recommendation on
Exposure Time' table in manual.

8cm

(2) Position the cone make horizontally with sensor.


※ Recommend that put the 4mm aluminum filter on a sensor.

2012-05-17 4
Preparation for calibration
Step 2. Copy & paste EzSensor.ini file

Serial no. : CR15HMHB4111-0001

(1) Copy ‘EzSensor.ini’ file in …


C:\EzSensor\MultiSensor\CRHMHB4111-0001

(2) Paste ‘EzSensor.ini’ file in …


C:\EzSensor

C:\EzSensor\(Serial No)\EzSensor.ini
This file has the setting information about (Serial No) sensor.

2012-05-17 5
Preparation for calibration
Step 3. Delete the previous calibration data.

(1) Delete all files (except BPMM.raw) in …


C:\EzSensor\CAL

2012-05-17 6
Start calibration software
Step 4. Execute software.
And then Calibration software appear.

C:\EzSensor\

2012-05-17 7
Basic settings for calibration
Step 5. S/W Setting

Check below items at the ‘Settings’ tab.

(1) Full Frame Size


Size 1.5: 686 x 944
Size 2.0: 744 x 1030

(2) Rotate and Flip

(3) Detector setting


[Device]에서 V-Reset setting
(Typical 1.48~2.00V)

2012-05-17 8
Dark calibration
Step 6. Dark frame Acquisition

② (1) Click Get Dark button.


(2) Check the dark frame(image) by clicking View Frames
button.

Dark image file save in the ‘C:\EzSensor\CAL’ directory.


2012-05-17 9
Bright calibration
Step 7. Bright frame Acquisition

② (1) Click Get Bright button.


(2) Exposure X-ray.
(3) Check the bright frame(image) by clicking View Frames
button.

① Bright image file save in the ‘C:\EzSensor\CAL’ directory.

X00001A.raw
Signal level of capturing image

2012-05-17 10
Bright calibration
Step 7-1. Get 5 different bright frames for generating Cal. Point.
Get 5 different bright frame that refer the left table.
(1) Click Get Bright button.
Exposure Time Number
Point Signal (2) Adjust the exposure time of X-ray generator and Exposure X-
For Vatech X-ray of frame
ray.
1 0.04 sec 400~600 5
(3) Check the signal level in the message list.
2 0.10 sec 1200~1400 5

3 0.16 sec 1900~2300 5

4 0.20 sec 2400~2700 3

5 0.25 sec 2900~3300 2

If signal level which you want to acquiring is too low or


too high, delete this file by click ‘<Purge’ button. (4) Keep the X-ray generator settings, and Exposure X-ray. Then

And then try again after adjust the exposure time of X-ray the image of a similar level is obtained.
(5) Get the frames as 'number of frame' through repeating above
generator.
step (4).

5 frames for Point 1

(6) Adjust the exposure time for each point refer (1)~(3). And get
the frames as 'number of frame' through repeating above step.

2012-05-17 11
Bright calibration
Step 7-2. Generate Cal. Point.

(1) Set ‘Tolerance’ value as ‘50’.

More than 2 different frames are averaged as 1 frame by


tolerance value.

(2) Click Generate button. And then Cal. Point files


are created.

① ②

2012-05-17 12
Bright calibration
Step 7-3. Delete raw bright frames.

(1) Click ‘<Purge’ button.


(2) Select ‘All items in the list’.
(3) Click ‘OK’.

② ③

2012-05-17 13
Bad Pixels Map calibration
Step 8. Generate Bad pixels map

(1) Set ‘Auto Bad Pixels’ Map’ options as same as left.


(2) Click Generate button. And then BPM files are created.
① (3) Click ‘OK’.

2012-05-17 14
Apply new calibration data
Step 9. Copy & paste calibration file

Serial no. : CR15HMHB4111-0001

(1) Copy all file in …


C:\EzSensor\CAL

(2)Paste in …
C:\EzSensor\MultiSensor\CRHMHB4111-0001

2012-05-17 15
Apply new calibration data
Step 10. Copy & paste EzSensor.ini file

Serial no. : CR15HMHB4111-0001


(1) Copy ‘EzSensor.ini’ file in …
C:\EzSensor

(2)Paste ‘EzSensor.ini’ file in …


C:\EzSensor\MultiSensor\CRHMHB4111-0001

2012-05-17 16