Scintillator
Scintillator
X-ray image convert into the visible ray through scintillator.
CMOS
CMOS
HIC Imaging Sensor
Definition of a term
Term Description
Dark image Offset, without X-ray image
Bright image Gain, with X-ray image
Signal Signal = Gain – Offset [LSB]
Value of gray scale
LSB (Least Significant Bit)
ex.) 12bit = 1~4096 LSB
2012-05-17 2
Introduction
Serial No of each sensor
Each sensor has the unique serial number and the different calibration data.
Be careful about confusing the calibration data and ini file.
Calibration Step
Step 1. Sensor setting
Step 2. Copy & paste EzSensor.ini file
Step 3. Delete the previous
calibration data.
Step 4. Execute software.
And then Calibration software appear.
Step 5. S/W Setting
Step 6. Dark frame Acquisition
Step 7. Bright frame Acquisition
Step 8. Generate Bad pixels map
Step 9. Copy & paste calibration file
Step 10. Copy & paste EzSensor.ini file
2012-05-17 3
Preparation for calibration
Step 1. Sensor setting
(1) Distance between X-ray cone and sensor : 8cm
※ Distance is depend on capability of X-ray generator. So, refer 'Recommendation on
Exposure Time' table in manual.
8cm
2012-05-17 4
Preparation for calibration
Step 2. Copy & paste EzSensor.ini file
C:\EzSensor\(Serial No)\EzSensor.ini
This file has the setting information about (Serial No) sensor.
2012-05-17 5
Preparation for calibration
Step 3. Delete the previous calibration data.
2012-05-17 6
Start calibration software
Step 4. Execute software.
And then Calibration software appear.
C:\EzSensor\
2012-05-17 7
Basic settings for calibration
Step 5. S/W Setting
2012-05-17 8
Dark calibration
Step 6. Dark frame Acquisition
2012-05-17 9
Bright calibration
Step 7. Bright frame Acquisition
X00001A.raw
Signal level of capturing image
2012-05-17 10
Bright calibration
Step 7-1. Get 5 different bright frames for generating Cal. Point.
Get 5 different bright frame that refer the left table.
(1) Click Get Bright button.
Exposure Time Number
Point Signal (2) Adjust the exposure time of X-ray generator and Exposure X-
For Vatech X-ray of frame
ray.
1 0.04 sec 400~600 5
(3) Check the signal level in the message list.
2 0.10 sec 1200~1400 5
And then try again after adjust the exposure time of X-ray the image of a similar level is obtained.
(5) Get the frames as 'number of frame' through repeating above
generator.
step (4).
(6) Adjust the exposure time for each point refer (1)~(3). And get
the frames as 'number of frame' through repeating above step.
2012-05-17 11
Bright calibration
Step 7-2. Generate Cal. Point.
① ②
2012-05-17 12
Bright calibration
Step 7-3. Delete raw bright frames.
② ③
2012-05-17 13
Bad Pixels Map calibration
Step 8. Generate Bad pixels map
2012-05-17 14
Apply new calibration data
Step 9. Copy & paste calibration file
(2)Paste in …
C:\EzSensor\MultiSensor\CRHMHB4111-0001
2012-05-17 15
Apply new calibration data
Step 10. Copy & paste EzSensor.ini file