Professional Documents
Culture Documents
For safe and better use of this analyzer and to improve your work efficiency,
please read the Instructions thoroughly before start.
After reading the Instructions, keep it handy for quick reference in the future.
BC1200 63000002
Biossays 240 Plus 63000040
Biossays 240 63000043
Intellectual Property Statement
Shenzhen New Industries Biomedical Engineering Co., Ltd. owns the intellectual
properties to these products and copyright of this instructions.
All rights reserved. Any part of this operating instructions cannot be copied, modified,
translated, by anyone or organization without permission.
, , , , and are
the registered trademarks or trademarks owned by Snibe in China and other
countries.
Intended Use: It is used in conjunction with adapter reagents for qualitative and/or
quantitative analysis of the analyte in a human sample.
Compile Date:06/2018
Issued Date: 09/2018
Table of Contents
Table of Contents
NOTICES ............................................................................................................................................. 1
1.1 OVERVIEW................................................................................................................................1-1
1.2 INSTRUMENT SPECIFICATIONS .......................................................................................................1-1
1.3 SYSTEM STRUCTURE....................................................................................................................1-7
1.3.1 Appearance .......................................................................................................................1-7
1.3.1.1 Front view ............................................................................................................................... 1-7
1.3.1.2 Top View ............................................................................................................................... 1-10
1.3.2 System composition ........................................................................................................1-13
1.3.3 Composition of operating unit ........................................................................................1-15
1.4 INSTRUMENT SYMBOLS .............................................................................................................1-15
1.5 OTHER LABELS .........................................................................................................................1-25
Operating Instructions-EN I
Table of Contents
3.3.9 Parabola..........................................................................................................................3-11
3.3.10 Spline ..........................................................................................................................3-11
3.4 ISE TEST PRINCIPLES .................................................................................................................3-12
3.4.1 Operating principle .........................................................................................................3-12
3.4.2 Production principle of electrode potential ....................................................................3-12
3.4.3 Measurement methods...................................................................................................3-13
II Operating Instructions-EN
Table of Contents
IV Operating Instructions-EN
Notices
Notices
This part involves important information and regulations concerning safety and
proper usage of the analyzer.
Operating Instructions-EN 1
Notices
Purpose
Biossays series automatic biochemistry analyzers and reagents are strictly limited to
use for the purpose of professional in vitro diagnosis. To ensure the reliability of results,
please use reagents and consumables manufactured by Shenzhen New Industries
Biomedical Engineering Co., Ltd. If non-snibe reagents need to be used, please contact
our after-sales staff.
The instructions provide operating instructions for the Biossays series automatic
biochemistry analyzer. The instructions will help you to understand the structure,
operations, routine maintenance, troubleshooting, etc. of the Biossays series automatic
biochemistry analyzer. Please operate in accordance with the Instructions.
Sign introduction
Signal
Sign Description
Word
Warning indicates a situation that, if not avoided, will result
Warning in personal injury, instrument damage, data error or
infection of diseases.
Attention indicates important information that you must
Attention
pay attention to.
Safety precautions
For safe use of this system, read the Instructions carefully before operating the
analyzer. Any operation violating the safety precautions may cause personal injury or
instrument damage.
This system complies with safety requirements on electronic medical instruments. It
shall be installed and operated by qualified personnel in strict accordance with laws and
regulations.
Warning:
1) If you fail to carry out necessary maintenance as per the Instructions,
it may cause instrument failure and health risk.
2) For sake of safety and reliability, the analyzer must be installed and
maintained only by maintenance engineer authorized or approved
by our company. All spare parts used for the analyzer must be
provided or approved by our company or our agent.
2 Operating Instructions-EN
Notices
Warning:
1) During running of the analyzer, do not look directly at the light beam,
which will hurt your eyes.
2) To change light source, disconnect the main power supply of the
analyzer and wait at least for 15min until the light source cools down.
Do not touch the light source when it is still hot.
2. Moving parts
Please comply with the following precautions to prevent personal injury caused by
moving parts during running of the analyzer.
Warning:
1) NEVER touch moving parts, including sample / reagent disk, sample
/ reagent pipetting mechanism, reaction disk, stirring mechanism,
washer, etc. or get your hands or any body part into their motion
path.
2) Do not place any obstacle on the path of moving parts; or it will
cause personal injury or instrument damage.
3) Sample tubes with cap may collide with the pipetting probe, so
please make sure caps are removed from all tubes in the sample
disk.
3. Electrical hazards
In order to prevent electric shock, please comply with the following precautions.
Warning:
1) High voltage inside. Only authorized maintenance personnel can
open the rear and side covers.
2) If liquids like reagent or sample enter into the analyzer, it may cause
failure and electric shock. In such cases, switch off power
immediately and contact our technical service department.
3) To replace parts and components, disconnect power first and then
open rear and side covers.
4) Incorrect earthing may result in electric shock and instrument
damages.
5) Make sure the input voltage meets requirements of the analyzer.
6) Do not touch or carry out electrostatic discharge on parts labelled
with ESD warning sticker.
Operating Instructions-EN 3
Notices
4. Fire
Using organic solution may cause fire. Please comply with the following precautions.
Warning:
1) Do not use organic solution in test.
2) The analyzer is not explosion-proof. If you have to use organic
solution nearby, take extreme care to avoid fire or explosion.
Warning:
The laser generated by the barcode reader will shoot the retina directly
and cause the eye injury. Do not stare at the laser beam of the barcode
reader directly.
6. Biochemical hazard
Please comply with the following precautions to prevent biochemical hazards
effectively.
Warning:
1) Improper use of sample may cause infection. Do not contact
samples, mixtures and waste liquids with your hands or other parts
of your body. Please wear gloves, masks and work clothes before
operation to prevent being infected. Wear protective glasses if
necessary.
2) Use reagent and wash liquid with care and avoid contact with them.
In case of skin contact, flush with soap and water promptly. If they
enter eyes by accident, rinse with plenty of water immediately and
visit an ophthalmologist.
3) If a bit of reagent or sample spills on the analyzer, wipe it off with
soft cloth damped with alcohol; if a large amount of reagent or
sample spills on the analyzer, stop using it and contact our
authorized engineer immediately.
4) The analyzer shall be thoroughly disinfected before transportation
for long distance in order to avoid potential spreading of infection.
4 Operating Instructions-EN
Notices
Warning:
1) Some substances contained in reagent, QC liquid, calibration
solution, washing liquid and waste liquid are subject to regulations
on pollutants and their discharge. Please comply with all applicable
state and local regulations, and consult with the manufacturer or
distributor if you have any question.
2) Infectious waste fluid must be disposed properly in infectious waste
treatment facility.
Warning:
As some materials used by the analyzer are subject to regulations on
pollution, the disposal shall comply with relevant laws and regulations.
Warning:
1) Do not do what you are not supposed to do on the computer, such
as clicking unknown popups, to prevent corrupting software system
due to factors like virus invasion or misoperation. Computer virus
may spread via USB disk, Internet, etc.
2) Do not install any software and hardware other than specified by our
company, so as to prevent hampering normal running of computer
software system. During system running, do not run other software.
Operating Instructions-EN 5
Notices
1. General precautions
Before using the analyzer, it is a good start to know its applications and general
precautions. Violation of the Instructions may defeat the protection integrated in the
analyzer’s design.
Attention:
1) This product is an in vitro diagnostic medical device, and is used in
conjunction with adapter reagents for qualitative and/or quantitative
analysis of the analyte in a human sample. When making the clinical
decision based on the analysis results, combine with considering the
clinical symptoms or other test results.
2) The Instructions may be revised without prior notice. You can
consult with our customer representative if you have any query.
3) The analyzer is intended for use only by trained professional users.
4) Do not touch computer’s display, mouse or keyboard with hands
stuck with chemicals.
5) Do not fold or press drainage pipes, which may cause blockage in
pipes and overflow of waste liquid from other openings.
6) During its running, the analyzer dissipates heat through its rear part.
Make sure the work environment is well ventilated. Use ventilation
equipment if necessary. but avoid airflow blowing on the analyzer
directly, or it may affect reliability of test results.
7) Before its first use, the analyzer must be calibrated to ensure
accurate performance.
8) No air bubbles can be present in samples, reagents and wash liquid.
Failure to do prevent air bubbles may result in incorrect sample
loading and the reliability of the test results cannot be guaranteed.
Do not move or exchange the reagents before the test is finished.
9) To ensure safe operation and stable test results, do not use expired
wash liquid.
10) Start the analyzer at least 30min before its usage to ensure that the
light source system and the reaction disk temperature is stable.
11) Before testing, check whether consumables (purified water, wash
liquid, etc.) are enough for the tests.
12) Before testing samples, quality control procedures must be
conducted to ensure reliable test results.
6 Operating Instructions-EN
Notices
2. Service environment
Attention:
Please install this instrument in environment specified herein. Any
installation and usage beyond specified conditions may result in unreliable
results, or even damage to the instrument.
3. Electromagnetic compatibility
Attention:
1) BC1200,Biossays 240 Plus, Biossays 240 automatic biochemistry
analyzer complies with electromagnetic emission and immunity
requirements in IEC 61326-2-6-2012.
2) Users are responsible for ensuring the electromagnetic compatibility
environment that allows the analyzer to work properly.
3) It is suggested to assess electromagnetic environment before using
the analyzer.
Warning:
1) BC1200,Biossays 240 Plus, Biossays 240 automatic biochemistry
analyzer is designed and tested according to requirements for Class
A analyzer in IEC/CISPR 11:2010. This analyzer may cause radio
interference in household environment, and therefore protective
measures should be taken.
2) It is prohibited to use the analyzer next to a strong radiation source
(e.g. unshielded RF source) because it may interfere with normal
operation of the analyzer.
4. System maintenance
Attention:
1) Please maintain the instrument periodically as specified herein.
Improper maintenance may affect accuracy and precision of test
results, and even damage the instrument or hurt people.
2) Before maintenance, turn off power and pull off the power plug;
otherwise it may damage the instrument or hurt people.
3) Please wear gloves and work clothes before maintenance to prevent
possible infection caused by residue of patient samples on the
instrument.
Operating Instructions-EN 7
Notices
Attention:
1) Medicines, anticoagulants, preservatives and other materials
contained in sample may cause interference to some test results.
2) Store samples properly. Improper storage of sample may lead to
changed sample composition or wrong test results.
3) To prevent evaporation, please do not expose sample to open air for
a long time; otherwise the test results will be inaccurate.
4) The samples with hemolysis, lipemia or jaundice will impact the
results.
5) You must ensure no clot is contained in samples, otherwise the
sample probe will be blocked, which significantly impacts the results.
6) If reagent, calibration solution and quality control liquid are stored
improperly, it may result in inaccurate test results and deteriorated
system performance even within validity period. Usage, storage and
other matters of reagent, calibration solution and quality control
liquid shall be subject to manufacturer’s instructions.
7) Always perform calibration assays after changing reagent; otherwise
the test results will be inaccurate.
8) During testing, reagent cross contamination may influence test
results. Please contact reagent manufacturer for reagent
contamination information.
6. Data backup
Attention:
During testing, this system saves data to hard disk drive automatically. To
prevent data loss due to accidental deletion or hard disk failure, backup
test results and instrument parameters to other media, like CD-ROM, on a
regular basis.
8 Operating Instructions-EN
Chapter 1 System Overview
1.1 Overview
Reagent bottle
20mL, 70mL
capacity
Reaction solution
Reaction 150~450μL
volume
system
Reaction time 10 minutes, 22 minutes
Access to LIS/HIS
Allowed
system
Temperature 10℃~30℃
Temperature -20℃~55℃
Weight 270kg
Reagent bottle
10mL, 30mL, 50mL
capacity
Reagent storage
2°C~10°C, with semiconductor refrigeration
temperature
Sample & reagent
Liquid level detection, clot detection, anti-collision
probe
Cuvette material Optical plastics
Temperature 10°C~30°C
Working Relative humidity ≤70%
environment
Atmospheric
86.0kPa~106.0kPa
pressure
Temperature -20°C~55°C
Relative humidity ≤93%
Storage Atmospheric
50.0kPa~106.0kPa
environment pressure
Well-ventilated indoor space without strong sunlight
Others
or corrosive gases.
1.3.1 Appearance
BC1200
(1) (2)
⑴ ⑵
⑶ ⑷ ⑸
(1) Roof cover (2) Glass (3) Manufacturer logo
(4) Alarm light (5) Model mark
Biossays 240
⑴ ⑵
⑶ ⑷ ⑸
(1) Roof cover (2) Glass (3) Manufacturer logo
(4) Alarm light (5) Model mark
BC1200
⑴ ⑵ ⑶ ⑷ ⑸ ⑹
Biossays 240
⑴ ⑵ ⑶ ⑷ ⑸
BC1200
log
conversion
A/D
Interface
conversion
Microcomputer
Reaction Optical
Washer
disk system
Biossays 240
log
conversion
A/D
Interface
conversion
Microcomputer
Reaction Optical
disk system
BC1200
“Protective Earthing”
This symbol indicates protective earthing label.
It is located on grounding screw holes of the equipment cabinet baseplate, close to the AC
power input port.
"S"
This symbol indicates sample disk.
It is labeled at the bottom center of sample disk.
"R1"
This symbol indicates R1 reagent disk.
It is labeled above the "Make the Plate Cover be Closed When the Analyzer is Working"
label, and align the left side of the label.
"R2"
This symbol indicates R2 reagent disk.
It is labeled at the bottom left of R2 reagent disk, and align with the top of the "Make the
Plate Cover be Closed When the Analyzer is Working" label.
"Breaker"
This symbol indicates the type of breaker: ~220V 20A.
It is labeled above the main switch.
"Main Switch"
This symbol indicates the main switch of the instrument.
It is labeled at the bottom center of the main switch.
"Submain Switch"
This symbol indicates the submain switch of the instrument.
It is labeled at the bottom center of the submain switch.
"RS 232"
This symbol indicates the RS 232 port.
It is labeled above the RS 232 port.
“Warning Infection!”
This symbol reminds users of biological infection risk.
It is labeled at areas with biological infection risk, including:
Above the "Light Waste Liquid Outlet" label;
Bottom of "Concentrated Waste Liquid Outlet" label;
Bottom right of "Waste Liquid Outlet of Vacuum Tank" label;
Bottom right of sample disk.
"Do not Switch Off the Main Power When the Reagent is Refrigerated"
This symbol and words remind users not to switch off the main power when the reagent is
refrigerated.
It is labeled at the bottom center of the "Main Switch" label.
“Caution”
This symbol indicates cautions on user’s safety.
It is labeled at bottom right of R1 reagent disk.
“Protective Earthing”
This symbol indicates protective earthing label.
It is located on the bottom plate between the switching power supply and the filter.
“Lamp”
This symbol indicates light source lamp and that here is the position for installation of light
source lamp.
It is located right above the heat sink of lamp.
“Purified Water”
This symbol indicates purified water.
It is located on the purified water tank.
“Sensor”
This symbol and signal words indicate interfaces of various sensors.
It is located above sensor interfaces on the left of instrument enclosure.
“Liquid”
This symbol and signal words indicate liquid inlets for various liquid circuits.
It is located below liquid interface on the left of instrument enclosure.
“Circuit Breaker”
This symbol and signal words indicate circuit breaker and remind users not to switch off
the main power when the reagent is refrigerated.
It is located above the main switch.
“Caution”
This symbol indicates cautions on user’s safety.
It is labeled at Middle of cross beam on upper cover.
“Please Make the Plate Cover be Closed when the Analyzer is Working”
This symbol and words remind users to make the plate cover be closed when the analyzer
is working.
It is labeled at the following positions:
Lower right of sample & reagent disk;
Upper left of reaction disk.
“Warning Infection!”
This symbol reminds users of biological infection risk.
It is labeled at areas with biological infection risk, including:
On the panel near the sample & reagent probe wash trough;
On the panel near the stirring needle wash trough;
Above waste liquid outlets.
Label Description
Manufacturer
Serial No.
The definition of WEEE symbol below only applies to EU member states. The use of
WEEE symbol on a device indicates the device should not be disposed of as domestic
waste. Ensuring proper scrapping of the device helps avoid potential influence of
hazardous substances on environment and human health. For more information, please
contact your local distributor.
THIS WAY UP
This symbol indicates correct upright position of the transport package.
It is labeled at the upper center of 4 sides of the packing box.
FRAGILE
This symbol indicates contents of the transport package are fragile therefore it shall be
handled with care.
It is labeled at the upper center of 4 sides of the packing box.
DO NOT ROLL
This symbol indicates the transport package shall not be rolled.
It is labeled at the upper center of 4 sides of the packing box.
DO NOT STACK
This symbol indicates stacking of the transport package is not allowed and no load should
be placed on the transport package.
It is labeled at the upper center of 4 sides of the packing box.
Warning:
Failure to meet above requirements may affect the instrument’s
performance or damage it, or even cause personal injury.
To ensure normal working of the analyzer, water supply and drainage shall meet the
following requirements:
Water used by this analyzer shall be provided with professional pure water
manufacturing machine;
Electrical conductivity of purified water shall be ≤1μs/cm;
Water supply shall purify water for at least:
BC1200 25L/h
Biossays 240 Plus
5L/h
Biossays 240
Warning:
Waste liquid shall be discharged according to relevant local provisions
on disposal of medical waste.
After installation, ensure that power lines and communication lines have been
properly connected to the instrument and computer. Meanwhile, ensure all the pipes such
as pure water inlet pipe, concentrated waste liquid outlet pipe, light waste liquid outlet pipe
as well as wash liquid inlet pipe connecting correctly to the instrument without looseness.
Then power-on and start the system.
1. Power on the analyzer. Turn on its main and submain power switches.
2. Power on the computer.
3. After logging into the Windows operating system, double click the shortcut icon of the
user software to start it. After start, enter the username and password in the pop-up
login dialog to log into the user software system.
Attention:
The system administrator’s username and initial password both are
“snibe”.
4. Upon successful connection between software and instrument, the system will be
initialized automatically, and all parts will finish initialization.
5. When the initialization is completed, and there are no warnings or error messages,
confirm the status of the instrument. For related operations and requirements, see
4.2.3 Confirm instrument status.
The system is comprised of software and instrument. The former processes data
input and output as well as controls operation of the instrument, while the latter completes
all test actions.
This chapter involves test principle, analysis methods and calibration methods,
detailed as below.
I
R lg 0 CL
It
Where: R is the light absorbance rate of a solution when light passes through the
This system is equipped with multiple analysis methods for more reasonable tests on
various biochemical reactions, mainly including 1-point endpoint method, 2-point endpoint
method, rate A method, and 2-point rate method.
R R2
S+R1
RX
Time (s)
Select the first point before reaction, select the second point when reaction ends or
equilibrates, and then calculate sample concentration with difference between
absorbance of the two points. Under certain conditions, this method can reduce sample’s
disturbance to the reaction or the reaction’s color disturbance. The reaction curve is
shown as Fig. 3.2-2.
R2
R
S+R1
(RM+RM-1)/2
(RL+RL-1)/2
Time (s)
and L-1 from average absorbance of points M and M-1, i.e. absorbance; k is volume
2-point rate is also called first-order kinetics or fixed-time method, which means
reaction velocity is proportional to first power of substrate concentration within a certain
period of reaction. As the substrate is consumed gradually, the reaction goes slower and
increase (or reduction) of absorbance becomes smaller and smaller. As it takes a long
time to reach reaction equilibrium, theoretically, measuring can be done at any time.
However, due to the complex components of serum, reaction will become stable only after
some time of complicated and heterogeneous reactions. For any kind of first order
reaction, substrate concentration S at any given time t after reaction starts is:
S S 0 e kt
S S 0 e kt2 e kt1
That is, at a given interval, it is common in first order reaction that change in substrate
concentration is proportional to initial concentration of substrate and so is the increase (or
reduction) of absorbance to concentration of measured substance.
2-point rate is stricter than endpoint method. All factors (e.g. pH, temperature,
enzyme concentration, etc.) affecting reaction velocity must be constant and accurate at 2
points and be calibrated with calibration solution at the same time. The reaction curve is
shown as Fig. 3.2-3 below.
R2
R
(RM+RM-1)/2
S+R1
(RL+RL-1)/2
t
Time (s)
Rate A is also called kinetic method or zero-order kinetics, which means reaction
velocity is proportional to zero power of substrate concentration, that is to say reaction
velocity is independent of concentration of the substrate. This is a method for quantitative
analysis based on measured production velocity of reaction product or consumption
velocity of substrate during continuous monitoring of reaction process. In the whole
reaction process, reactant can produce certain product at constant velocity which leads to
uniform reduction or increase in absorbance of the measured solution under a certain
R R2
RM
S+R1
RL
t
Time (s)
R X R
T
Where, RX is the rate of change per minute in absorbance between points L and M
measured with least square method.
For assays with dual wavelengths, the actual absorbance during every measurement
cycle equals to the difference between absorbance at the primary wavelength and the
secondary wavelength.
Calibration equation: C K ( R R0 )
C1
R0 0 or R0 R1 .
K
calibration solution is 0, R0 0 .
吸
光
度
RX
R1
C1 CX 浓度
Calibration equation: R aC
R
There is only one parameter a , a .
C
It is required to provide 1 calibration solution. C is the concentration of calibration
RX
C CX Concentration
Calibration equation: R aC b
R2 R1 R2 R1
There are 2 parameters, i.e. a and b , where a , b R1 C1 .
C2 C1 C 2 C1
Absorbance
R2
RX
R1
C1 CX C2 Concentration
Calibration equation: R aC b
There are 2 parameters a and b , which can be obtained by linear least square
method in following equation:
n
n n
C R
i i i Ri n
C
i 1 i 1
a i 1 2
n
n
i 1
Ci Ci n
2
i 1
n n n
n
Ci R i
C i
Ri n n
C n
b Ri n
i 1 i 1 i 1
i 1 n
n
2 i
Ci Ci n
i 1
2
i 1 i 1
Absorbance
R3
RX
R2
R1
C1 C2 CX C3 Concentration
3.3.5 Logistic-Log 4P
1
Calibration equation: R R0 K
1 exp[ (a b ln C )]
C1 C2 C3 C4 Concentration
3.3.6 Logistic-Log 5P
1
Calibration equation: R R0 K
1 exp[ (a b ln C cC )]
C1 C2 C3 C4 C5 Concentration
3.3.7 Exponential 5P
C1 C2 C3 C4 C5 Concentration
3.3.8 Polynomial 5P
R R0 R R0 2 R R0 3
Calibration equation: ln C a b( ) c( ) d( )
100 100 100
There are 5 parameters, i.e. R0 , a , b , c and d .
It is required to provide at least 5 calibration solutions. Concentration (activity) of the
3.3.9 Parabola
Calibration equation: R aC 2 bC c
There are 3 parameters, i.e. a , b and c .
It is required to provide at least 3 calibration solutions and solve liner equation in 3
dimensions with least-square calculation method of polynominal.
3.3.10 Spline
R R0i ai C Ci bi C Ci ci C Ci
2 3
Calibration equation:
R0i ai bi c
There are 4 parameters, i.e. , , and i .
It is required to provide 2-6 calibration solutions. As it is piecewise fitting, the degree
of fitting is highest among all calibration types.
When starting analysis, the instrument puts A Std. into sampling cup with A pump and
moves A Std. to electrode pipes with sample pump to measure relative potentials
comparing with potential of reference electrode. After completion of measurement, take A
Std. away with sample pump and make sampling cup and electrode pipe empty. After
measure potential of calibration solution A, move B Std. into sampling cup with B pump
and pump B Std. again into electrode pipe with sample pump to measure its potential and
compare with potential of reference electrode. Upon completion of measurement, bring
wash liquid for A Std. with A pump and remove the liquid away with sample pump. When
sample probe takes sample and fills it into sampling cup, pump sample into electrode pipe
with sample pump to measure its potential. After measurement, bring wash liquid for A Std.
with A pump, remove the liquid away with sample pump and empty sampling cup and
electrode pipe.
E x E0 ( RT / nF ) * ln a x (1)
ax f x cx (2)
Where:
f x : activity coefficient
c x : concentration
n : charge number of given ion (i) (positive ion is positive, negative ion is negative)
E A E0 S lg f A c A (3)
E B E0 S lg f B c B (4)
E x E0 S lg f x c x (5)
f
Where, f A , f B and x are activity coefficients of standard solutions A, B and sample
solution. In practical application, adjust activity coefficient of standard solution to make
f A f B f x , and calculate actual slope of electrode with equations (3) and (4) as:
CB
S ( E B E A ) / lg( )
CA (6)
Calculate concentration
C x of sample with equations (5) and (3) as:
C x C A *10 ( Ex E A ) / S (7)
(2) Cell blank confirm Photoelectricity Confirm the measured value is within the
maintenance allowed range through cell blank inspection.
5. Confirm analysis
conditions
(1) Assay parameter Assay Confirm parameters of the assays.
confirm parameter Confirm whether such assays are calibrated,
(2) Calibration Curve fitting the calibration is in the valid period, and
parameter confirm calibration methods and parameters are to be
used.
6. Prepare reagent
(1) Position confirm Reagent info Confirm whether reagent register information is
(1) Exit the operating Click <Exit System> in the shortcut key area to
software confirm exiting the software.
(2) Shutdown Turn off the power of the instrument and the
computer.
15. Operations after Cover each reagent bottle in the reagent disk;
the end Take out calibration solution, QC solution and
samples of the sample disk.
Empty the waste liquid tanks.
Prior to startup, the following inspection measures should be carried out to ensure the
system can run normally after startup.
Warning:
For the following inspection operations, you must pay attention to
bio-infectious risks, and wear gloves and working clothes to avoid being
infected, and if necessary, protective glasses.
1. Power on Biossays series automatic biochemistry analyzer. Turn on its main and
submain power switches. Its main switch should be normally open to ensure the
running of cooling system if the reagent disk places reagents.
2. Power on the computer and the printer.
3. After you login Windows operating system, double click the shortcut icon of the
user software to start it. After the user software starts, a login dialog (shown as
Fig. 4.2-1) pops up in the screen. Enter the user and password and click <Login>
to enter the user software.
Attention:
The system administrator’s name and initial password both are “snibe”.
4. After you login the user software, the instrument will be initialized, and all parts
will finish initialization. When the reaction disk’s temperature is stable, test can
be started.
Attention:
After the components of the instrument are initialized, it takes about 20
minutes for the reaction disk temperature to stabilize, and 10 minutes for
the light source to stabilize. Thus, after the reaction disk temperature
becomes stable after startup, sample test can be started.
After the testing, the result column will show the results of all cuvettes under 16-way
wave, cuvette No. and times. A judgment will be drawn with test data whether the cuvettes
can be used for sample analysis.
The cell blank test results within ±3,500 meets requirements.
Should cell blank data meet requirements, it can be tested. If abnormal, users are not
recommended to continue sample test, or it may impact the reliability of results.
Before the testing, confirm whether assay parameters, calibration methods, normal
range and other settings of such assay are correct, such assay is calibrated and the
calibration is within the valid period. Group/Profile and cross contamination information
should be set up based on demands.
3. Group/Profile settings
Click <Setting> in the master key area to open the [Setting] window. In the window,
click <Group/Profile> to open the [Group/Profile Setting] interface. Then, Group/Profile will
be set up based on the detailed demands.
Profile means putting related assays together, such as liver function tests. Several
assays can be added by clicking the profile name, which is convenient for rapid input
during adding the sample order.
Attention:
Some assays may impact the results of other assays due to reagent
formula during analysis. The contamination degree varies with reagents.
Operators are recommended to place the reagents subject to cross
contamination separately. If not so, cross wash can be set up to reduce
cross contamination between assays.
Place the respective reagent in the reagent position of the reagent disk. Open the
cover of reagent bottle to confirm the remaining volume is sufficient for this sample test.
Attention:
1、 After registration of reagent information, please detect remaining
volume to confirm remaining volume and times before sample
analysis.
2、 Remaining times are calculated to multiply the section area of the
reagent bottle by reagent lever height measured by the reagent
probe, then divide the reagent volume. Errors of the reagent height
and section area may result in the error of remaining times. Thus,
the remaining times should be regarded as estimates which may not
be in accordance with the actual times of each reagent bottle.
When preparations prior to analysis are completed, routine sample analysis can be
made.
Click <Calibration> in the master key area to open the [Calibration] window, and
<Calibration Order> to open the [Calibration Order] window, which is shown as Fig. 4.3-1.
When you select the assays for calibration order, the right lower calibration solution
information list will show calibration solution setting information of the selected assay.
After the calibration solution and calibration method are confirmed to be correct. Click
<Add Order> to add calibration order of such assay.
If you want to delete the calibration orders, click <Calibration Result> in the
[Calibration] window to open the [Calibration Result] window. Select order information of
such assay in the calibration information list, and click <Delete> to delete the calibration
order.
4.3.2 QC order
Click <QC> in the master key area to open the [QC] window, and <QC Order> to
open the [QC Order] window, which is shown as Fig. 4.3-2.
In the [QC Order] window, input QC name, position, QC lot number, target value, SD
and interval, and select the assay for QC order. Click <Add> to complete QC information
input of such assay.
If you want to delete QC information of an assay, select such information in the left
working area of the window (it shows after selected), and click <Delete> to delete
such information.
When you select the assays for which QC order is necessary in the left working area
of the window (it shows after selected), and confirm that the QC information is correct,
and click <Add Order> to add the QC order .
If you want to delete the QC order, click <QC Result> to open the [QC Result] window.
In the upper working area of the window, select such QC order information (it shows
after selected), and click <Delete> to delete them.
Click <Work List> in the master key area to open the [Work List] window, which is
shown as Fig. 4.3-3.
Sample No. includes start and end No. They can be added by the system or by
manual. If by manual, input start No. in the first edit box, and end No. in the second. When
the system defaults that these two numbers are the same, it means only one sample;
when end No. is larger than start No., it means batch sample.
Then, the [Test List] window will show information about all sample orders.
If you want to add STAT sample orders, select the STAT button (it gets green after
selected). It means the current batch samples are STAT which will be tested in the first
priority.
If such batch samples need diluting, select the Dilution button (it gets green after
selected) to open the [Select Dilution Ratio] dialog window. In the window, select the
dilution rate for the assays to be diluted.
Click <Add Order> to complete adding batch sample orders. If the sample No. of such
batch has been existed, there will be a sign indicating that adding sample order failed.
Obtain the sample’s barcode information and its position No. in the sample disk by
barcode scanning. The barcode scanner in the sample disk will scan all sample positions
of its outer and middle rings to obtain samples’ barcode information.
After starting LIS function to connect LIS server successfully, you can get the
sample’s assays from the LIS sever through LIS application function.
Warning:
1. The samples with hemolysis, lipemia or jaundice will impact the
results.
2. You must ensure no clot is contained in samples, otherwise the
sample probe will be blocked, which significantly impacts the results.
3. Some substances of the samples, such as drugs, anticoagulants,
preservatives, will disturb the results.
4. Don’t place the samples in the opened container for a long term,
otherwise they may volatilize, which will impact the results.
5. Incorrect parameter settings will impact the results.
6. Shenzhen New Industries Biomedical Engineering Co., Ltd. doesn’t
suggest users themselves to modify and add the results, and shall
not assume any responsibilities for sequences resulting from such
actions.
2. Start test
When finishing adding orders, click <Start Test> in the shortcut area to send a test
command, and the instrument will start test.
Attention:
1. Don’t change samples when the sample probe and the sample disk
are running. It can be available when they stop.
2. When the instrument is working, contact with the sample / reagent
probe, the stirring needle and the reaction disk and other running
devices should be avoided.
3. When the instrument is working, don’t change the sample disk or
open the sample / reagent disk’s cover.
4. QC solutions and calibration solutions should be tested with
standard or trace cup.
Click <OK> to pause sampling of new test. Tests which have sample added in the
reaction disk will continue. The sample can be placed in the sample disk only after the
instrument’s sample probe stops running. At this time, it is necessary to confirm whether
the position of such sample in the sample disk is consistent with the sample order
information. After the sample is in place, click <Start Test> in the shortcut key area to
complete continuous loading operations which allow the instrument to start sampling and
continue analysis.
Attention:
1. Please do not keep the system in the Pause Sampling status for a long
time. Otherwise, certain tests may be affected to some extent.
2. In Pause Sampling status, only the sample probe suspends sample, but
the reagent probe does not stop aspirating the reagent. At the time, you
cannot put a new reagent bottle in the reagent disk or uncap the disk.
3. During Pause Sampling status, analysis of sampled assays will not stop.
Attention:
Since the sample probe and the reagent probe of the Biossays 240 Plus /
Biossays 240 analyzer are the same one, that is, the sample & reagent
probe. Please note the following information about the Biossays 240 Plus /
Biossays 240 analyzer:
1. In pause sampling status, the sample & reagent probe stops sampling of
new test while assay analysis for which R1 reagent has been added will
not stop, and the sample & reagent probe will not stop running
immediately.
2. After clicking <Pause Sampling>, if the sample & reagent probe is running,
the <Pause Sampling> button will be shown in red. At this moment, it is
not allowed to place samples and reagents in the sample & reagent disk
or open the sample & reagent disk cover.
3. In pause sampling status, samples and reagents can be placed in the
sample & reagent disk only after the instrument’s sample & reagent
probe stops running and the <Pause Sampling> button turns green.
During the testing, monitor status of sample / reagent disk and reaction disk, and the
instrument’s temperature, voltage and liquid status in real time. In case of additional
sample fault, instrument fault and other conditions, pause sampling. If major faults occur, it
is necessary to stop test.
Click <Status> in the master key area to open the [Status] window which shows the
current status of sample / reagent disk and reaction disk, and the instrument’s current
temperature, voltage and liquid status.
After the testing, the result can be observed in the [Result] window.
Result disposal includes confirmation of results and testing of rerunning samples.
Click <Result> in the master key area to open the [Result] window, which is shown as
Fig. 4.4-1. As the window shows test results of all assays, users can view, delete, rerun
and print test results, and view reaction curves, etc. They can also input sample details in
the patient information list of the [Report] window for previewing and printing reports.
In the [Result] window, click [Expt. Result] to enter the [Expt. Result] window. The left
sample list will show information about all samples. When you click the sample you want
to view, the right result list will show all assays and result information of such sample.
Click <Search Sample> to input the respective date. You can query historical result
information as well as result information of a certain sample No.
Click <Send Sample> to open [Send Sample - Sample List] dialog window. When you
select sending conditions in the window, it will send result information that matches the
conditions. After you click <OK>, the window will disappear. Thus, sending the result is
finished.
Click <Print Sample> to open [Print Sample - Sample List] dialog window. When you
select printing conditions in the window, it will print result information that matches the
conditions. After you click <OK>, the window will disappear. Thus, printing the result is
finished.
Click <Export Sample> to select export conditions. It will output result information that
matches the conditions. After you click <OK>, the [Export Sample - Sample List] dialog
window will disappear. Thus, the result is exported via Excel document.
Warning:
1. The samples with hemolysis, lipemia or jaundice will impact the
results.
2. You must ensure no clot is contained in samples, otherwise the
sample probe will be blocked, which significantly impacts the results.
3. Some substances of the samples, such as drugs, anticoagulants,
preservatives, will disturb the results.
4. Don’t place the samples in the opened container for a long term,
otherwise they may volatilize, which will impact the results.
5. Incorrect parameter settings will impact the results.
6. Shenzhen New Industries Biomedical Engineering Co., Ltd. doesn’t
suggest users themselves to modify and add the results, and shall
not assume any responsibilities for sequences resulting from such
actions.
1. Reaction curve
If you want to view the reaction curve of a certain result, click result information of
such test and <Reaction Curve> to open the [Reaction Curve] dialog window. The window
shows the reaction curve of the result, which is shown as Fig. 4.4-2.
The result information list in the below window shows assay, sample No., barcode,
cuvette No., primary wave, secondary wave, test date, result time, etc. The right Abs list
shows Abs value of all points’ primary wave and secondary wave. The reaction curve can
be expressed in three ways: primary wave-secondary wave, primary wave, secondary
wave
Click <Print> to output the reaction curve of the result, including result information,
reaction curve and Abs value of all points.
Click <Cancel> to exit the [Reaction Curve] dialog window.
Select the test date of the reports to be printed in the drop-down list box of “Test
Date:”.
When you select sample No. to be viewed in the sample information list, the left
patient information list will show details of the selected samples, and the middle result
information list will show the result information of all assays of the selected samples.
Details of the sample in the left patient information list need editing by users in order
to print a complete report. Such details include sample No., patient type, patient No.,
name, sex, age, department, sample type, character, sender, send time, operator, verifier,
diagnosis and remark. Click <Save> to finish the edition of sample patient information.
Click <Preview> to preview reports to be printed, which is shown as Fig. 4.4-4 .
In the [Preview] dialog window, click <Cancel> to exit the window. If you want to print
the report, click <Print> to output it.
If you want to print the batch test reports, click <Print> in the [Report] window (Fig.
4.4-3) to open the [Sample No. Range] dialog window, which is shown as the figure below.
Enter start and end No. of the samples to be printed to output the test reports of the
selected samples. After you click <OK>, the window will disappear. Thus, printing the
reports by batch is finished.
Attention:
1. Only the current day’s test results can be rerun.
2. Sample rerun testing must be made by clicking <Start Test> in the
shortcut area to send a rerun command.
After all tests are completed, test results are confirmed and test reports are printed,
you can exit the operating software and Windows operating system, and then turn off the
power supply of each part.
4.5.1 Shutdown
When the current day’s tests come to an end that all operations and tests are finished,
you can exit the user software and Windows operating system, and turn off the power
supply of each part in the following order:
1. Turn off the printer’s power;
2. Turn off the computer’s power;
3. Turn off the submain switch power of the instrument.
4. If there is pure water unit, turn off the pure water unit’s power.
Attention:
When its auxiliary switch power is off, the cooling system is still working.
In order to close the system, please turn off the master power switch.
In order for the next work preparations, the following should be checked:
1. Check whether each pipetting probe, the stirring needle and the washer are in
the right position, polluted and bent, and their tips have drops.
2. Cover each reagent bottle in the reagent disk.
3. Take out calibration solutions, QC solutions and samples in the sample disk.
4. Empty the waster liquid tank.
5. Check whether there are stains in the instrument’s plane part. If any, wipe out
them with clean soft cloth.
shows the connected status of user software and the instrument as follows:
Red It represents the user software could not open the port connecting
Red the instrument, and the connection between the user software and
the instrument failed.
Green It represents the user software could open the port connecting the
Red instrument, but the connection between the user software and the
instrument failed.
Green It represents the user software could open the port connecting the
Green instrument, and the connection between user software and the
instrument succeeded.
shows the connected status of user software and LIS server as follows:
Red It represents the user software could not open the port connecting
Red LIS server, and the connection between the user software and LIS
server failed.
Green It represents the user software could open the port connecting LIS
Red server, but the connection between user software and LIS server
failed.
Green It represents the user software could open the port connecting LIS
Green server, and the connection between user software and LIS server
succeeded.
The shortcut key area is in the lower user software interface, which displays
commonly used function buttons, convenient for users’ operation.
<Start Test>button. When you click this button, the user software
sends a start test command to the instrument, and the latter starts
test.
Attention:
1、 Before you click <Start Test> button, please confirm samples,
calibration solutions, QC solutions, reagents and others are in the
right position.
2、 Unless specifically specified, for example, instrument faults, it’s not
recommended to use such function. After the testing stops, all the
tests which have not finished running should void. Click <Start
Test>button to re-start all unfinished tests.
3、 When the operator takes a break, please lock the operating software
to avoid that non-users destroy or modify data.
Click <Work List> in the master key area to open the [Work List] window as shown in
the figure below. You can add sample order, view and delete sample order information,
and view test list etc.
Button Function
Add Order Click this button to complete adding single or batch sample orders.
Scan Barcode When you click this button, the barcode scanner in the sample disk
will scan all sample positions of its outer and middle rings to obtain
samples’ barcode information.
LIS Apply When you start LIS function to connect LIS server successfully,
click this button to obtain samples’ assay information from the LIS
server.
Test List Click this button to open the [Test List] window showing test list
information of sample order that matches the screening criteria,
you can delete, print, export test list information and add assay.
Delete Sample Click this button to delete the sample order information that
matches the screening criteria.
Button Function
STAT It means the current sample is STAT after selected.
Dilution The [Select Dilution Ratio] dialog window will pop up after selected.
Select the sample’s assays to be diluted and dilution rate.
During the testing, the instrument will prepare dilution sample with
dilution solution and sample which are mixed in a clean cuvette,
and pipette sample from the cuvette for loading and testing during
the sampling.
Warning:
Sample deletion will result in void tests of the sample. As data cannot be
recovered, please be careful.
Click <Result> in master key area to open [Result] window. The Button function of its
sidebar menu is described as follows:
Button Function
Expt. Result Display result and carry out search, rerun, print, send, delete on it
and check reaction curve etc.
If result is abnormal, it gets yellow. If you select the result, result
flag will be displayed in Result Flag Info window which is movable
and in the lower right corner. If result is normal, the Result Flag Info
window will be hidden.
Report You can edit patient information, preview and print report in this
window.
Statistics/Calc. This window is used to conduct statistics on results of an assay
within specified period
or recalculate results based on new calibration curve.
Attention:
1. After adding sample rerun orders, click <Start Test> button in
shortcut key area to retest the assays that are required for rerun.
2. For assay tested in other systems, it is possible to add its result and
print report with other assays in this system.
Warning:
1、 Results once deleted, are unrecoverable, please delete with care.
2、 It is not advised to modify/add results, and Shenzhen New
Industries Biomedical Engineering Co., Ltd. will assume no
responsibility for any consequences resulted therefrom.
Attention:
Concentration, once recalculated based on the latest calibration curve, is
unrecoverable, please recalculate with care.
Click <Reagent Info> button in master key area to open [Reagent Info] window, where
you can set up reagent information, scan reagent barcode information, detect remaining
reagent volume etc.
Attention:
Do not open the reagent disk cover during sample analysis to avoid
dangers or damages to the analyzer.
Click <Calibration> button in master key area to open [Calibration] window. The
Button function of its sidebar menu is described as follows:
Button Function
Calibration Order This window is used for adding calibration orders of assay.
Calibration Result This window is used to view, search, delete calibration result
and check reaction curve of the result etc.
If calibration result is abnormal, it will turn yellow and a result
flag will show in Result Flag Info window. If calibration result is
normal, the Result Flag Info Window will be hidden.
Curve Fitting This window is used to display and print relevant information of
calibration curve, such as calibration method, fitting formula,
fitting parameter and calibration curve, etc.
ISE Calibration This window is used to view ISE calibration result and carry
out point calibration, slope calibration, removing protein and
pipe washing etc.
Attention:
1. In case of changes in analysis conditions due to modification of lot
No., test parameters, light source lamp and other reasons,
recalibration is necessary.
2. Calibration results once deleted, are unrecoverable, please delete
with care.
Attention:
It is advisable to conduct reagent blank test everyday as reagent blank
test is vital for obtaining correct results and blank value is useful to check
validity of reagent, deduct reaction background and to deduct
absorbance changes resulted from reagent in rate A and 2-point rate
methods.
Click <QC> button in master key area to open [QC] window. The Button function of its
sidebar menu is described as follows:
Button Function
QC Order This window is used for adding QC order of assay.
If you set up QC interval of assay, during routine sample
analysis after selecting assay, the analyzer will automatically
insert QC assay for test in accordance with number of interval
samples.
QC Result This window is used to view, search, recalculate, delete QC
result and check reaction curve of the result etc.
If QC result is abnormal, it will turn yellow and a result flag will
show in movable Result Flag Info window.
Lot QC This window is used to check the QC Lot Graph, QC data and
QC analysis results of the QC liquid with the same lot No.
Conduct QC analysis based on Westgard Multirules.
Monthly QC This window is used to check the QC Month Graph, QC data
and QC analysis results of the QC liquid of different lot No. or
of different QC liquids.
Conduct QC analysis based on Westgard Multirules.
Warning:
QC Results once deleted, are unrecoverable, please delete with care.
after selected), and click <Recalculate> button to open [Message Confirm] dialog window,
which is shown as the figure below.
Click <OK>to recalculate concentration based on the latest calibration curve of the
assay. Concentration will be updated in the upper workspace in [QC Result] window.
Click <Cancel> to cancel recalculation.
Attention:
Concentration, once recalculated based on the latest calibration curve, is
unrecoverable, please recalculate with care.
Click <Status> in master key area to open [Status] window. The Button function of its
sidebar menu is described as follows:
Fig. 5.8-1 BC1200 [Status] menu Fig. 5.8-2 Biossays 240 Plus /
Biossays 240 [Status] menu
Button Function
Sample Disk This window is used to display the sample disk graph and its
current status. The current status of samples in the sample
disk are distinguished by color.
R1 Disk This window is used to display the reagent1 disk graph and its
current status. The current status of each reagent position is
differentiated by color.
R2 Disk This window is used to display the reagent2 disk graph and its
current status. The current status of each reagent position is
differentiated by color.
Sample&Reagent Disk This window is used to display the sample & reagent disk
graph and its current status.
The type and status of each sample position in the disk are
differentiated by color combination of the inner and outer rings.
The outer color represents type, the inner color represents
current status.
The current status of each reagent position is differentiated by
color.
Reaction Disk This window is used to display the reaction disk graph and its
current status. The current status of each cuvette is
differentiated by color.
ISE This window is used to display ISE module’s running status, A
Std. reagent kit status, B Std. reagent kit status, information
related to electrode status and provide respective functions of
replacing A Std., B Std. and Electrode, etc.
Instrument Status This window is used to display the analyzer’s current
temperature, voltage and liquid status. If respective
temperature and voltage exceeds the limited range, it is
displayed in red. The liquid status gets green in the normal
status and red in other status.
Click <Setting> in master key area to open the [Setting] window. The Button function
of its sidebar menu is described as follows:
Button Function
Assay Param. This window is used to set up base parameter, normal range,
calibration parameter, QC rule, dilution parameter and auto
dilution of assays.
Group/Profile This window is used to set up profile information. All selectable
profiles are displayed in the profile bar of the [Work List]
window.
Relevant assays are put together to create assay profile with
obvious clinical significance, for example, liver function, and
renal function, for quick addition of assays during adding
sample orders.
Calculate Assay This window is used to set up calculate assay and its normal
range.
Calculate assay means using results of two or more assays to
create a new assay with certain clinical significance by certain
calculations, for example, A/G.
Cross Contamination This window is used to set up cross contamination of reagent,
cuvette and sample.
Button Function
Cross contamination wash is a function that reduces or avoids
the contamination between assays. Due to reagents, serious
cross contamination may occur between some assays. In
order to eliminate such contamination, users are required to
isolate assays which produce cross contamination with those
which don’t produce when they preset assays. If not so, auto
wash should be added by setting contamination information
before testing contaminated assays.
ISE Assay This window is used to set up ISE assay information, normal
range and whether using ISE module.
Manual Assay This window is used to set up manual assay and its normal
range.
Manual assay refers to all assays which are not tested in the
analyzer. Users can input patient’s manual assay results which
will be printed with reports.
System Setting This window is used to set up user, hospital information, report
information, LIS, print, experiment parameter, language and
other settings.
Workload This window is used to view and print workload within the
selected period. With statistic workload, you can count
workload of send departments, senders, operators and assays
in the selected period.
Attention:
1. When sex and sample type are the same within normal range, no
inclusion or overlap is allowed for age.
2. After respective dilution rate for an assay has setted up in the
[Dilution Parameter] window, you can select it in [Calibration
Parameter], [Auto Dilution] and [Work List] windows.
3. If a test assay has selected “Use Auto Dilution”. When the assay
result exceeds set threshold, dilution test will be made automatically
at set dilution rate (testing is subject to quantity of auto dilution
rates).
4. “snibe” is the default administrator’s name. The name and authority
cannot be changed. “snibe” user’s initial password is “snibe”. Please
change it immediately.
5. After setting the department, doctor, patient type, clinical diagnosis,
remark, sample type and sample character in the [Hospital Setting]
and [Report Setting] windows, you can select the appropriate
information when editing patient information in the [Report] window.
6. After setting the unit name in the [Report Setting] windows, you can
select results’ unit information when you edit assay parameter,
calculate assay, manual assay and ISE assay.
Click “Basic Parameter” in the [Assay Param.] window to open the [Basic Parameter]
window, where you can set up base assay parameter, shown as the figure below.
Parameter Meaning
Effective Input appropriate point range according to the reagent
Photometry Points instructions. 0 means no points.
The analyzer records Abs data once per
BC1200
12s.
Biossays 240
The analyzer records Abs data once per
Plus
15s.
Biossays 240
Serum Index Test If selected (it shows after selected), all the samples which
containing the assay test will have serum index test.
Abs Limits Set up Abs check range in the input box.
If such check is not carried out, input 0 (negative reaction) or
3.3 (positive reaction).
Prozone Check Input range of prozone check.
If such check is not carried out, input -3.3 (lower) or 3.3
(upper).Select upper and lower and input respective value in
the drop-down list box.
Calculation Factor It is for relationship calibration to verify consistency of results
(Y = aX + b) from the analyzer and other instruments.
Attention:
Set up relevant parameters correctly in accordance with the reagent
instructions. Incorrect settings cannot ensure correct results.
Click “Calibration Parameter” in the [Assay Param.] window to open the [Calibration
Parameter] window where you can set up calibration parameter of an assay, shown as the
figure below.
Parameter Meaning
Cal. Interval (Day) Expired period of calibration curve.
Please set up calibration interval of the calibration curve in
order to judge whether it is within expired period.
Sensitivity Check Difference between the maximum and minimum calibration
solution contraction. If it is lower than this set value, alarm will
sound for exceeding sensitivity.
If such check is not carried out, input 0.
Discrete Check During the calibration, each calibration solution should be
tested twice and taken an average. If Abs difference of two
tests is larger than the set value of discrete check, alarm will
sound for exceeding discrete check range.
If such check is not carried out, input 3.3.
Drift Rate Check Difference between concentration absorbance calculated by
approximate expression and test absorbance in non-linear or
multi-point linear calibration curves. If it exceeds set value of
drift rate check, alarm will sound.
If such check is not carried out, please input 3.3.
Blank Level Check During the calibration, average Abs value of calibration
solution 1 in the two tests exceeds the blank level check range,
alarm will sound.
Lower in the first edit box, and upper in the second.
If such check is not carried out, please input -3.3-3.3.
Attention:
1. Set up relevant parameters correctly in accordance with the reagent
instructions. Incorrect settings cannot ensure correct results.
2. Calibration method must be followed to determine quantity of
calibration solutions. Save failed if insufficient.
Button Function
System Info This window is used to view system version information.
Initialize Click this button to send an initialization command to the
analyzer, and the latter will perform the command.
Reset Click this button to send a reset command to the analyzer, and
the latter will perform the command.
Component Maintenance Click this button to open the [Component Debugging] dialog
window. Then you can select the components for which
initialization is required in the window ( shows after selected)
and click <Send>, the software will send an initialization
command of the components to the analyzer, and the latter will
perform the command.
Button Function
Photoelectricity This window is used to conduct cell blank test, stability test,
maintenance light intensity check.
Under normal circumstances, users are recommended to do
cell blank test once a week. They should do so after replacing
cuvette. Should cell blank data meet requirements, it can
conduct sample tests.
Under normal circumstances, users are recommended to do
light intensity check once a month. They should do so after
replacing light source. Should data meet requirements, it can
conduct sample tests.
Low Level Command This window is used to send a low level command by which a
component of the analyzer is controlled to complete a certain
operation.
The analyzer must complete initialization before a low level
command is sent.
System Log This window is used to check and delete alarm and error
information. Users can take respective measures according to
alarm and error information.
In order to guarantee the system’s reliability, good running status and service life, you
are required to operate and regularly maintain the system in strict accordance with the
instructions.
If you have any problems about use and maintenance unmentioned in this Chapter,
please contact technical service department or your local dealer of Shenzhen New
Industries Biomedical Engineering Co., Ltd.
Hitergent wash
7 Reagent disk Replenish when used up
liquid
8 ISE cleaner Sample disk Replenish when used up
ISE-Reagent
9 —— Replenish when used up
pack
6.1.1 Tools
6.1.2 Water
The instrument should use purified water with conductivity ≤ 1 μs/cm during the
routine operation and system maintenance. When you use water supply device,
remember to conduct maintenance and check on a regular basis. See water device
manual for details or contact the manufacturer or retailer.
Wash liquid is used to clean parts of the analyzer. All kinds of wash liquid are
available from Shenzhen New Industries Biomedical Engineering Co., Ltd. In case of any
other alternatives, cuvette, sample / reagent probe, stirring needle and piping may fail to
be cleaned up, affecting results accuracy and precision. The company shall not assume
responsibilities for inaccuracy caused by such actions.
Start water purifier and check the conductivity which should be ≤1μs/cm; otherwise
the water purifier manual should be followed to make maintenance.
Warning:
1. Pay attention to the danger of biology infections when you are
operating. Please wear gloves and working clothes to prevent
infection, and if necessary, protective glasses. As for disposal of
waste liquid tanks, you should obey local emission standard, and
consult relevant reagent manufacturers or distributors.
2. Tank top cannot be higher than the waste liquid outlet.
3. The pipe should be inserted into the tank, without bending or
twisting; otherwise concentrated waste liquid will be leaked out.
Warning:
1. Pay attention to the danger of biology infections when you are
operating. Please wear gloves and working clothes to prevent
infection, and if necessary, protective glasses. As for disposal of
light waste liquid tanks, you should obey local emission standard,
and consult relevant reagent manufacturers or distributors.
2. Waste gauze should be disposed of according to relevant provisions.
Don’t discard carelessly.
Observe whether the washer’s needle is bent. If bent, please contact our company’s
technical service department or local dealer.
As there is dust in cooling fan surface after long-term use of the analyzer, it should be
cleaned once per 6 months.
Cooling fan cleaning:
1. Turn off the analyzer’s main power.
2. A dust collector is used to remove dust in the fan.
As there is dust in dust boot surface of cooling fan after long-term use of the BC1200
analyzer, it should be cleaned once per 6 months.
Cleaning of dust boot:
1. Hold the handle of dust boot at both sides of the fan and unplug the two dust boots
directly.
2. Remove the dust on the boot with the dirt collector.
3. Rinse the boots with the clean water.
4. After wiping with the cloth, install the dust boots back.
When water flows are found abnormal during the sample / reagent probe cleaning, it
may be caused due to probe block and should be cleaned. The sample / reagent probe
should follow disassembly, cleaning and installation orders.
If the sample / reagent probe is damaged or has bent tip, you should replace it in strict
accordance with the following steps.
1. The analysis unit is powered off.
2. Gently pull up the rocker arm of the sample / reagent probe to the top, and move it
to make the sample / reagent probe stay in a position of the sample / reagent bin
convenient for operation.
3. Hold the rocker arm’s shell with fingers, pull up and take it down.
4. Screw off the pipe connector and connection lines.
5. Disassemble the sample /reagent probe.
6. Put a new one into the shell, tighten the pipe connector and plug connection lines.
7. Install the shell in the respective position.
If the stirring needle is damaged or has bent tip, you should replace it in strict
accordance with the following steps.
1. The analysis unit is powered off.
2. Gently pull up the rocker arm of the stirring needle to the top, and move it to make
the stirring needle stay in a position of the sample/reagent bin convenient for
operation.
3. Hold the rocker arm’s shell with fingers, pull up and take it down.
4. Screw off two screws of the stirring needle with a small cross head screwdriver,
and take it down.
5. Put these two screws in a new stirring needle, align the motor shaft and tighten the
screws.
6. Install the shell in the respective position.
When a light source lamp is burn in, its light quantity will damp, reducing the
analyzer’s accuracy. You should do light intensity check. If it exceeds the range, replace it
immediately.
1. Semi-permanent cuvette:
If cell blank value is disqualified after the cuvette is cleaned, you should replace a
new cuvette.
Attention:
A new cuvette should be soaked in 2% hitergent wash liquid for 8 hours
before use. Take it out, and install in the reaction disk after a complete
clean. After installed, blank level check should be done. When light
quantity meets requirements, you can have a test.
2. Disposable cuvette:
Disposable cuvettes need to be replaced after each use.
a) Wait for the analyzer to stop and be in the "Standby" status.
b) In the "Replace Cuvettes" column on the right side of the [Reaction Disk Status]
window, select the cuvette group No. to be replaced and click the <Replace>
button.
c) Wear protective gloves, open the cover of the reaction disk, and remove the
thumbscrews fixing the cuvettes of the corresponding group No.
d) Take down the group of cuvettes.
e) Install the new cuvettes in the same positions on the reaction disk and tighten
screw by hand.
f) After confirming that the replacement of the group of cuvettes has been completed
in the pop-up dialog, the status of the group of cuvettes will be updated in the
[Reaction Disk Status] window.
g) After replacing all cuvettes, cover the reaction disk. Click the <Scan Cuvettes>
button in the [Reaction Disk Status] window to scan all the cuvettes. The
difference between the cell blank data of cuvettes and reference cell blank value
should be within ±3,500. Cuvette that meet requirements can be used for testing.
Warning:
1. Beware of the risk of biological infection and wear gloves and work
clothes during operation to prevent infection, and wear protective
glasses when necessary.
2. For the disposal of waste liquids and waste cuvettes, please comply
with local emission standards, relevant laws and regulations and
consult with related manufacturers or distributors.
BC1200:
1. Push the instrument rear panel gate open.
2. Pull out the alkaline wash liquid tank bracket.
3. Unscrew the alkaline wash liquid tank cover counterclockwise.
4. Fill the alkaline wash liquid.
5. Tighten the alkaline wash liquid tank cover.
6. Push the alkaline wash liquid tank bracket inside the instrument.
7. Close the instrument rear panel gate.
See Table 6.7-1 for parts to be cleaned regularly and components to be replaced
After the ISE test, electrodes may be polluted by protein, fat or bacteria. When it is
over, electrodes should be cleaned up for use in the next day.
When the current day’s testing is finished, place more than 250μL of ISE cleaner in
the sample disk; then click <Calibration> in the [Remove Protein] window of the
[Calibration] window to execute “Remove Protein”.
During the electrode use, the analyzer will do sample test, automatic electrode
maintenance and automatic timing calibration. If it’s turned off in these operations, it may
result in liquid residue in ISE pipe, even electrode soaking in liquid, which will affect the
analyzer’s normal use in the next time. There, it’s recommended to wash ISE pipe with
water before power-off.
1. Select “Pipe Wash” in the [ISE Calibration] window of the [Calibration] window.
2. After the wash, turn off the analyzer.
After the long-term use of electrodes, they will be polluted so as to result in inaccurate
results. Thus, the method below should be followed to have a check.
Select “Point Cal.” 10 times in the [ISE Calibration] window of the [Calibration]
window to obtain check results.
Attention:
The difference between two check values of the same electrode should
be less than 0.2.
Attention:
1. Generally, don’t open the ISE cover, or it may cause poor control
over temperature to result in incorrect results.
2. Electrode replacement and device check should be finished in 1
hour.
After the long-term use of ion-selective electrode, its potential will become small to
lead to poor response, and you should replace with a new electrode.
If slope of ISE calibration is abnormal, an alarm will sound. See Table 6.7-2.
When slope of Na+, K+, Cl-, iCa2+ or pH is abnormal up to alarm level, replace with a
new ion-selective electrode.
When slope of Na+, K+, Cl-, iCa2+ or pH electrodes is lower or unstable, replace with
new reference electrodes.
If alarm still sounds or QC test cannot to satisfy requirements with normal slope, it
demonstrates poor electrode response. In such case, it may be caused by pipe pollution
and can be resolved by cleaning.
If ISE calibration value is normal in the previous day, but sharp change of the current
day’s slope, it may result from other reasons except for electrode. Check whether there is
liquid leakage, and blockage and bubbles in the pipe.
1. Click <Replace Electrode> in the [ISE Status] window of the [Status] window, and
operate in accordance with the dialog.
2. After the electrode is replaced, the system will go through initialization and
electrode calibration because of change of electrode status.
BC1200:
1. Open the instrument analysis unit ISE cover plate.
2. Unplug the old rigid pipe and install the new one.
3. If drops occur, wipe them with gauze with water.
4. Install ISE cover plate back.
1. Click <Replace A Std.> or <Replace B Std.> in the [ISE Status] window of the
[Status] window, and operate in accordance with the dialog.
2. After the reagent is replaced, the system will go through pipe perfusion and
electrode calibration because of change of reagent status.
Chapter 7 Troubleshooting
Instrument faults and relevant solutions are listed in the table below:
Biochemical module
0.5 ±0.025
1.0 ±0.07
(CV)
ISE module
A.2.10 Accuracy
Accuracy should conform to Table 3.
A.2.11 Precision
Precision should conform to Table 3.
A.2.12 Linearity
Linearity deviation of instrument should conform to Table 3.
A.2.13 Stability
Stability of instrument should conform to Table 3.
B.2.1 Appearance
The following appearance requirements should be satisfied:
a) Appearance should be neat without crack or scratch; text and logo should be clear;
b) Painted surfaces should have uniform color without obvious color difference;
c) Moving parts of the analytical system should be steady without problems like getting
stuck, jump and significant backlash; the key group should rebound flexibly;
d) Surfaces of metal parts should be free from obvious pits and burrs;
e) Fasteners should be connected securely and reliably without looseness;
0.5 ±0.025
1.0 ±0.07
Test the nominal minimum reagent adding volume 20μL and maximum reagent
adding volume 350μL. Adding error should not exceed ±5% and coefficient of variation not
exceed 2%.
(CV)
Linearity(D)
+
K not exceed ≤1.5% 1.5~7.5 ≤3.0% ≤2.0% ≤1.5%
±3.0%
≥0.995
+
Na not exceed ≤1.5% 100.0~180.0 ≤3.0% ≤2.0% ≤1.5%
±3.0%
-
Cl not exceed ≤1.5% 80.0~160.0 ≤3.0% ≤2.0% ≤1.5%
±3.0%
not exceed
≤5.0%or
±5.0%or
2+
iCa ≤1.5% 0.50~2.50 ≤0.05 ≤3.0% ≤2.0%
±0.05
mmol/L
mmol/L
±3.0%