You are on page 1of 35

OPERATION MANUAL

AURO REF / KERATOMETER URK-700

Revision 1.0 B UNICOS Co., Ltd. www.e-unicos.com

Before use this instrument, be sure to read this manual collectively.


IMPORTANT NOTICE

This product may malfunction due to electromagnetic waves caused by


portable personal telephones, transceivers, radio-controlled toys, etc.
Be sure to avoid having objects such as, which affect this product,
brought near the product.

It should be used under the supervision of medical staff of hospital

The information in this publication has been carefully checked and is


believed to be entirely accurate at the time of publication. UNICOS
assumes no responsibility, however, for possible errors or omissions, or
for any consequences resulting from the use of the information contained
herein.

UNICOS reserves the right to make changes in its products or product


specifications at any time and without prior notice, and is not required to
update this documentation to reflect such changes.

UNICOS Co., Ltd.


210-4, Maero-dong, Seo-gu, Daejeon, 302-020, Korea
All rights reserved.
Under copyright laws, this manual may not be copied, in whole or in part,
without the prior written consent of UNICOS.
SAFETY INFORMATION

Accessory equipment connected to the analog and digital interfaces must be


certificated according to the respective IEC/EN standards (e.g. IEC/EN 60950
for data processing equipment and IEC/EN 60601-1 for medical equipments).
Furthermore all configurations shall comply with the system standard EN
60601-1-2:2001. Everybody who connects additional equipment to the signal
input part or signal output part configures a medical system, and is therefore
responsible that the system complies with the requirements of the system
standard EN 60601-1-1:2001. If in doubt, consult the technical service
department or your local representative.

For U.S.A
Do not make any changes or modifications to the equipment unless otherwise
specified in the manual.
If such changes or modifications should be made, you could be required to stop
operation of the equipment.
For EU Countries
The following mark, the name & address of the EU Representative shows
compliance of the instrument with Directive 93/42/EEC.

EU Representative:

MailShop GmbH Augenoptik

Ziegeleistr. 18-24, 75417 Muhlacker, Germany

- ⅱ-
Symbols marked on the Instrument
Symbol Description

TYPE B EQUIPMENT

Protective earth (ground)

~ Alternating current

! Attention, consult ACCOMPANYING DOCUMENTS

O Off (power: disconnect to the mains)

I On (power: connection to the mains)

General Safety Information


If you see any warnings or cautions printed on the warning labels, follow
the safety instructions in this manual. Ignoring such cautions or warnings
while handling the product may result in injury or accident. Be sure to
read and fully understand the manual before using this product.
Keep this manual in easy-to-access place.

Caution Sign Description


! WARNING This indicates a potentially hazardous situation which could
results in death or serious injury to you or others.
This indicates hazardous situations which may result in
! CAUTION minor injury to you or others, or may result in machine
damage.

NOTE This is used to emphasize essential information. Be sure to


read this information to avoid incorrect operation.

- ⅲ-
Only operate the instrument with the power supply indicated
! WARNING on the rating plate. Otherwise, it may result in fire or
electric shock.

! WARNING Be sure to turn OFF the power switch before connecting or


disconnecting the cables. Also, do not handle them with wet
hands. Otherwise, you may get an electric shock that may
result in death or serious injury.

! WARNING Never disassemble or modify this instrument because it may


result in fire or electric shock. Also, since this instrument
incorporates high-voltage parts and other hazardous parts,
touching them may cause death or serious injury.

Should any of the following occur, immediately turn OFF the


! WARNING power switch, unplug the power cable from the AC outlet,
and contact the dealer or the agent who/where you purchase
this instrument.
 When there is smoke, strange odor or abnormal sound.
 When liquid has been spilled into the instrument or a
metal object has entered through an opening.
 When the product has been dropped or its housing
damaged.

! CAUTION This instrument is shipped with a grounding type power


cable. To reduce the risk of electric shock, always plug the
cable into a grounded power outlet.

Ensure that the examinee has not placed his/her hand or


! CAUTION fingers under the chin rest. Otherwise, hand or fingers may
be hurt.

! CAUTION Wipe the forehead rest with ethanol or glutaraldehyde


solution to disinfect it each time a different examinee uses
it, in order to prevent infection.
! CAUTION Change the chin rest paper each time the examinee changes
in order to keep the chin rest clean.

Do not place your hand or fingers between the stage and


! CAUTION base. Also ensure that the examinee does not place his/her
hand or fingers there either. Otherwise, hand or fingers may
be hurt.

- ⅳ -

URK OPERATION MANUAL

Contents
IMPORTANT NOTICE············································ ⅰ
SAFETY INFORMATION········································ ⅱ
1. Features / 1
2. Notes for Using the Instrument / 2
3. Description / 3
3.1 Main Unit / 3
3.2 Operation Buttons / 5
4. Practicing by Model Eye / 6
5. Measurement / 8
5.1 Continuous Keratometry and Refractometry / 8
5.2 Refraktometry / 11
5.3 Keratometry / 142
5.4 Contact Lens Base Curve Measurement / 17
5.5 Corneal diameter measurement / 19
6. Other Measurement Modes / 22
6.1 SETUP Mode / 22
6.2 External Video Display / 24
7. Self Inspection and Maintenance / 25
7.1 Before calling a Service Person / 25
7.2 Replacement of Printing Paper / 26
7.3 Replacement of Chinrest Paper / 26
7.4 Cleaning / 26
7.5 When moving the Instrument / 26
7.6 Service Information / 26
8. Specifications / 28
9. Components / 29

1. Features
1. Various Measurements Supported.
Not only the usual refractometry and keratometry, but also c
orneal diameter and base curve of contact lens can be measu
red with this one instrument. Thus, measurements of eye and
prescriptions for glasses and contact lenses can be made m
ore efficiently.

2. Wide Dioptric Measurement Range


Because the URK-700 covers a wide measurement range, from

-25D to +22D, even an examinee with strong myopia can be mea


sured.

3. More accurate Measurement.


The foggy method of the eye fixation target makes examinee’s
eye comfortable and enables to get more accurate measurement
data.

4. Easy Connection with other Equipments


This instrument is designed to connect other equipments such
as VCR or TV monitor.
- 1 -

2. Notes for Using the Instrument


Caution Sign Description
Only operate the instrument with the power supply indicated on th
! WARNING e rating plate. Otherwise, it may result in fire or electric shock.
Be sure to turn OFF the power switch before connecting or discon

! WARNING necting the cables. Also, do not handle them with wet hands. Othe
rwise, you may get an electric shock that may result in death or s
erious injury.
Never disassemble or modify this instrument because it may result
! WARNING in fire or electric shock. Also, since this instrument incorporates
high-voltage parts and other hazardous parts, touching them may c
ause death or serious injury.
Ensure that the examinee has not placed his/her hand or fingers u
! CAUTION
nder the chin rest. Otherwise, hand or fingers may be hurt.
1. Do not hit or drop the instrument. The instrument may be dam
aged when it receives a strong impact. The impact may cause
damage the function of this instrument. Please to handle with c
are.
2. An exposure to the direct sunlight or too bright indoor lights m
ay influence on the result of accurate measurement.
3. To connect this with other relevant equipments, consult with th
e dealer about the way to make it.
4. A sudden heating of the room in cold areas will cause condens
ation on the protective glass in the monitor screen and on opti
cal parts inside the instrument. In this case, just wait until con
densation disappears before performing measurement.
5. Keep the objective glass of the examinee side clean. If it was
stained, it may cause on ERROR or inaccurate measurements.
6. In case there is smoke, strange odor or noise on working disco
nnect the power supply and consult the dealer.
7. Don’t use organic solution such as alcohol, thinner, benzene, et
c. to clean the surface of this instrument. It may damage the in
strument.
8. When moving this URK-700, fix the stage by using clamping bo
lt and stage holding knob, always keep power off, and then lift
the bottom of the unit with both hands.
9. If you leave URK-700 without using for certain period, disconn
ect the power supply and protect the unit with dust cover.
- 2 -

3. Description.
3.1 Main Unit.

Name Function

① LCD Monitor Monitor that displays Measurement.

② User Key There are function keys.

③ Measurement button Press this button for measurement.

④ Stage Clamping Bolt Makes the system stage fixed.

⑤ Printer Print the measured result.


- 3 -

Name Function

① Head Rest Place the examinee’s forehead against this rest.


② Measurement windo Window for the examinee to look at for measurem
w ent

③ Chin Rest Place the examinee’s chin on the rest.

④ Stage Holding Knob Holds the movement of stage.

⑤ EXT Video Connect outside Video

⑥ RS-232 JACK Connect outside PC


⑦ Power IN LET Connector for the power supply code.

Protects instrument from the excess electric powe


⑧ Fuse:TA2L 250V
r.

⑨ Power Switch Switch for turning power ON and OFF.

- 4 -
3.2 Operation Buttons

Title Technical skill

MODE To change measurement mode

CLEAR To erase measured data on the screen

MENU To change and set up the function

PRINT To print the data of measurement


To measure the patient’s eye with cataract or IOL(intra-ocular l
IOL
ens)

- 5 -

4. Practicing by Model Eye


Practice measurement by using the Test Model Eye before doing the measu
rement of patient’s eye.

(1) Turn ON the power


Turn ON the power switch on the lower left side.
(2) Set the model eye on
Remove the chin rest paper and align the holes on the base of the model ey
e with the holes on the chin rest. And insert pins.
(3) Release stage lock
Fully turn the bolt in the underneath cover in counter clockwise direction. A
nd turn the stage holding bolt in counter clockwise direction to release th
e stage lock.
(4) Adjust height of Test Model Eye.
Adjust the height of the chin rest by rotating Height Adjustment knob so tha
t The Test Mode Eye is aligned with the Height Adjustment mark on the he
adrest.
(5) Enter K/R or REF Mode
If neither K/R nor REF is displayed on the monitor press MODE button until
either one is displayed.
(6) Adjust the position and focus on the Model Eye.
Looking at the monitor incline the operation joystick toward the Model Eye
until a bright dot appears near the inner alignment mark. Place the bright do
t in the center of inner alignment mark. Focus on the Model Eye by inclinin
g the operation joystick forward and backward so that KERATO RING image
is displayed clearly on the monitor, at the time adjusted the bar would be a
ppeared above the inner alignment mark.
(7) Measurement
 Manual measurement
A. Adjust the position and focus on the Mode Eye like the procedure 6
.
B. Press the measurement button. If measurement is failed with messag
e like RETRY repeat the procedure A. and press the measurement b
utton again.
C. Check whether diopter value is measured or not. Diopter value is re
corded in the bottom area of the Model Eye.

- 6 -
 Automatic measurement
A. To initialize program press MENU button.
B. After User setting is displayed select Auto Start by cursor and meas
urement will start automatically. Then press EXIT button, menu will
be saved and measurement mode appear.
C. When the bright dot enters the inner alignment mark and Model Eye
is in focus properly measurement starts automatically.
- 7 -

5. Measurement
5.1 Continuous Keratometry and Refractometry ( K/R Mode)
In the K/R Mode refractometry is performed after keratometry.
- 8 -
(1) Enter the K/R Mode
A. Press MODE button until “R/K” is displayed on the top of the displ
ay.
(2) Adjust height of examinee’s eye
A. Have the examinee sit and place his/her chin and forehead against
the chin rest and forehead rest.
B. Adjust the height of the chin by rotating the height adjustment kno
b so that the eye of the examinee is aligned with the height adjust
ment mark on the headrest.
(3) Perform Focusing
A. Slide the operation joystick to the left so that the right eye of the
examinee is displayed on the monitor.
B. Ask the examinee to look at the red balloon in the center of the s
cene (eye fixation target).
C. Looking at the monitor, check that image of the KERATO RING is
not obscured by the upper eyelid, if it is covering the ring, instruct
the examinee to keep his/her eye opened wide until measurement
ends.
D. Looking at the monitor, incline the operation joystick to right or lef
t, and turn the operation joystick so the pupil is centered with the
inner alignment mark and focus on the KERATO RING, at the time
adjusted the bar would be appeared above the inner alignment mar
k.
(4) Measurement
A. Press the Measurement button
B. The newest measured result will be displayed on the monitor.
C. According to the 6.1 SETUP Mode you can select and change VD,
CYL points.
(5) Repeated measurement
A. Measurement could be repeatedly performed if necessary.
B. The newest result will be displayed on the monitor whenever meas
ured.
C. CAUTION: before starting the new measurement you have to press
CLEAR button to remove previous data.
(6) Measure the other eye.
A. Slide the stage to the right side and measure the left eye.
B. After both eyes are measured, pupil distance (PD) will be displayed
on the monitor.

- 9 -
(7) Print
A. Press the PRINT button .
B. The selected data in the 6.1SETUP Mode will be printed.

Ensure that the examinee has not placed his/her hand or fi


! CAUTION ngers under the chin rest. Otherwise, hand or fingers may
be hurt.
At the time the instrument would be moved from right to le
! CAUTION ft or from left to right of examinee’ s eye, pull the instrum
ent maximum to examiner’s side. Otherwise, face may be h
urt.

NOTES:Safety information about infrared light radiation for measurement


As light sources for measurement, this instrument adapted two type of infrared
LED. It was chosen to satisfy the amount of energy exiting from the instrument
never exceed the limit value recommended by the international standard ISO
15004. This condition is satisfied even when the instrument is operating at
maximum light intensity and maximum aperture! (Maximum intensity is the
highest brightness the instrument is capable of delivering, including the highest
brightness achievable if overvoltage is provided)
Detailed radiation information at normal usage of this instrument is like bellows.
① Keratometry measurement
Radiation output: below 0.03 mW/cm2
Limit by ISO15504: 100 mW/cm2
② Refractometry measurement
Radiation output: below 0.15 mW/cm2
Limit by ISO15504: 100 mW/cm2

- 10 -
5.2 Refractometry (REF Mode)
Only refractometry can be performed in the REF mode.
- 11 -
(1) Enter the mode.
A.Press the MODE button until “REF” appear on the top of display
(2) Adjust height of examinee’s eye
A. Have the examinee sit and place his/her chin and forehead against
the chin rest and forehead rest.
B. Adjust the height of the chin by rotating the height adjustment kno
b so that the eye of the examinee is aligned with the height adjust
ment mark on the face rest.
(3) Perform Focusing
A. Slide the operation joystick to the left so that the right eye of the
examinee is displayed on the monitor.
B. Ask the examinee to look at the red balloon in the center of the s
cene (eye fixation target).
C. Looking at the monitor, check that image of the KERATO RING is
not obscured by the upper eyelid, if it is covering the ring, instruct
the examinee to keep his/her eye opened widely until measuremen
t ends.
D. Looking at the monitor, incline the operation joystick to right or lef
t, and turn the operation joystick so the pupil is centered with the
inner alignment mark and focus on the KERATO RING, at the time
adjusted the bar would be appeared above the inner alignment mar
k.
(4) Measurement
A. Press the Measurement button
B. The newest measured result will be displayed on the monitor.
C. According to the 6.1 SETUP Mode you can select and change VD,
CYL points.
(5) Repeated measurement
A. Measurement could be repeatedly performed if necessary.
B. The newest result will be displayed on the monitor whenever meas
ured.
C. CAUTION: before starting the new measurement you have to press
CLEAR button to remove previous data.
(6) Measure the other eye.
A. Slide the stage to the right side and measure the left eye.
B. After both eyes are measured, pupil distance (PD) will be displayed
on the monitor.

- 12 -
(7) Print
A. Press the PRINT button .
B. The selected data in the 6.1SETUP Mode will be printed.
Ensure that the examinee has not placed his/her hand or fi
! CAUTION ngers under the chin rest. Otherwise, hand or fingers may
be hurt.
At the time the instrument would be moved from right to le
! CAUTION ft or from left to right of examinee’ s eye, pull the instrum
ent maximum to examiner’s side. Otherwise, face may be h
urt.
- 13 -
5.3 Keratometry (KER Mode)
Only the radius of curvature of the cornea can be measured in KER Mode.
- 14 -
(1) Enter the KER Mode.
A. Press the MODE button until “KER” is displayed in the top of the displa
y
(2) Adjust height of examinee’s eye
A. Have the examinee sit and place his/her chin and forehead against the
chin rest and forehead rest.
B. Adjust the height of the chin by rotating the height adjustment knob so
that the eye of the examinee is aligned with the height adjustment mar
k on the face rest.
(3) Perform Focusing
A. Slide the operation joystick to the left so that the right eye of the exa
minee is displayed on the monitor.
B. Ask the examinee to look at the red balloon in the center of the scene
( eye fixation target ).
C. Looking at the monitor, check that image of the KERATO RING is not o
bscured by the upper eyelid, if it is covering the ring, instruct the exa
minee to keep his/her eye opened widely until measurement ends.
D. Looking at the monitor, incline the operation joystick to right or left, an
d turn the operation joystick so the pupil is concentric with the inner al
ignment mark and focus on the KERATO RING, at the time adjusted the
bar would be appeared above the inner alignment mark.
(4) Measurement
A. Press the Measurement button
B. The newest measured result will be displayed on the monitor.
C. According to the 6.1 SETUP Mode you can select and change VD, CYL,
KERATO points.
(5) Repeated measurement
A. Measurement could be repeatedly performed if necessary.
B. The newest result will be displayed on the monitor whenever measured.
C. CAUTION: before starting the new measurement you have to press CLE
AR button to remove previous data.
(6) Measure the other eye.
A. Slide the stage to the right side and measure the left eye.
B. After both eyes are measured, pupil distance (PD) will be displayed on
the monitor.
- 15 -
(7) Print
A. Press the PRINT button .
The selected data in the 6.1SETUP Mode will be printed.
Ensure that the examinee has not placed his/her hand or fing
! CAUTION ers under the chin rest. Otherwise, hand or fingers may be h
urt.
At the time the instrument would be moved from right to left
! CAUTION or from left to right of examinee’ s eye, pull the instrument
maximum to examiner’s side. Otherwise, face may be hurt.
- 16 -
5.4 Contact Lens Base Curve Measurement (CLBC mode)
Base curve of hard contact lens can be measured in the CLBC mode.

(1) Enter the Mode.


A. Press the MODE button until “CLBC MODE” is displayed in the top of t
he display
(2) Attach Contact Lens.
A. Put water in the concave section of contact lens holder at backside of t
he TEST Model Eye and place the contact lens on the holder with the
concave surface facing upward. Take care that the contact lens not dro
p.
(3) Attach the test Model Eye.
A. Remove the chin rest paper. Fix the Model Eye attached with contact l
ens using fixation pin.
- 17 -
(4) Perform Focusing
A. Looking at the monitor, incline the operation joystick to right or left, an
d turn the operation joystick so the pupil is centered with the inner alig
nment red mark.
B. Moving operation joystick straight and back fix up KERATO RING clearl
y, at the time adjusted the bar would be appeared above the inner align
ment mark.
(5) Measurement
A. Press the Measurement button
B. The newest measured result will be displayed on the monitor.
C. According to the 6.1 SETUP Mode you can select and change VD, CYL,
KERATO points.
(6) Repeated measurement
A. Measurement could be repeatedly performed if necessary.
B. The newest result will be displayed on the monitor whenever measured.
C. CAUTION: before starting the new measurement you have to press CLE
AR button to remove previous data.
(7) Print
A. Press the PRINT button .
The selected data in the 6.1SETUP Mode will be printed.
- 18 –
5.5 Corneal diameter Measurement (SIZE Mode)
Diameter of cornea can be measured in the SIZE mode.
- 19 -
(1) Enter the Mode
A. Press MODE button until “SIZE MODE” is displayed in the top of t
he display.
(2) Adjust height of examinee’s eye
A. Have the examinee sit and place his/her chin and forehead against
the chin rest and forehead rest.
- 20 -
B. Adjust the height of the chin by rotating the height adjustment kno
b so that the eye of the examinee is aligned with the height adjust
ment mark on the headrest.
(3) Perform Focusing
A. Slide the operation joystick to the left so that the right eye of the
examinee is displayed on the monitor.
B. Ask the examinee to look at the red balloon in the center of the s
cene ( eye fixation target ).
C. Looking at the monitor, check that image of the KERATO RING is
not obscured by the upper eyelid, if it is covering the ring, instruct
the examinee to keep his/her eye opened widely until measuremen
t ends.
D. Looking at the monitor, incline the operation joystick to right or lef
t, and turn the operation joystick so the pupil is centered with the
inner alignment mark and focus on the KERATO RING.
(4) Measurement
A. Press the Measurement button
B. To move the left/right bar by pressing the →/← button.
C. The newest measured result will be displayed on the monitor.
D. According to the 6.1 SETUP Mode you can select and change VD,
CYL , KERATO points.
(5) Repeated measurement
A. Measurement could be repeatedly performed if necessary.
B. The newest result will be displayed on the monitor whenever meas
ured.
C. CAUTION: before starting the new measurement you have to press
CLEAR button to remove previous data.
(6) Measure the other eye.
A. Slide the stage to the right side and measure the left eye.
(7) Print
A. Press the PRINT button .
B. The selected data in the 6.1SETUP Mode will be printed.
Ensure that the examinee has not placed his/her hand or finge
! CAUTION rs under the chin rest. Otherwise, hand or fingers may be hur
t.
At the time the instrument would be moved from right to left
! CAUTION or from left to right of examinee’ s eye, pull the instrument m
aximum to examiner’s side. Otherwise, face may be hurt.

- 21 –

6. Other Measurement Mode


6.1 SETUP MODE
Change all the settings about measurements, printouts.
- 22 -
(1) STEP
To change measurement data units.
Select 0.25/0.12 pressing button →/←
(2) VD
To select Vertex Distance (VD)
Select 0/10/12/13.5/15 pressing button →/←
(3) CYL
To select Cylinder value and form
Select -/+/+- pressing button →/←
(4) SPH SHIFT
To adjust spherical shift value
Select shift value pressing button -/+
(5) AUTO START
To select Manual or Auto measurement
Select Manual/Auto pressing button →/←.
If select AP, print the measurement results automatically after automatic m
easuring both eyes.
(6) AUTO REPEAT
To select repeated measurements.
Select 1/3/5/7 pressing button →/←.
(7) MONITOR BRIGHT
Monitor brightness control (“1” is the brightest)
Select 1/2/3/4 pressing button →/←
(8) KEY SOUND
To control sound volume (“ON” is the highest)
Select OFF/MID/ON pressing button →/←
(9) PRINT
Change printout forms.
select ALL/IMG/ECONO/OFF pressing button →/←.
If select IMG, would be printed out an image as emmetropia, hyperopia an
d myopia
(10) PRINT NO.
Printout Numbering.
Select ON/OFF/RESET pressing button →/←.

- 23 –
(11) DATA PREVIEW
Press PRINT button after DATA ON to display the latest measurements of
KER mode or REF mode.
select ON/OFF pressing button →/←.
(12) SCREEN SAVE
Anytime you don’t operate the instrument it will turn to the power saving
mode.
select OFF/3/5/10 pressing button →/←.
(13) USER MESSAGE
Enter a message to be printed out with the results of measurement using t
he internal printer in this window..
To display the user’s message select the one and press EDIT button.
Using buttons ←/↑/↓ select message and press SET button. You can edit t
he message using →,←, DEL, BAK buttons . then press SET button again.
Select END and SET button to complete message.
(14) DATE FORM
To setup order of displaying the date.
select YMD/DMY/MDY pressing button →/←.
(15) 24H MODE
to change 12 hour/24 hour time
change 12H/24H pressing button →/←.
(16) DATA
To setup year/month/day
Select year/month/day pressing button →/← and use +/- button.
(17) TIME
To setup hour/minute/second
Select hour/minute/second pressing button →/← and use +/- button.

6.2 External Video Display


In case of the necessity external video display, to connect RCA video cable bet
ween the EXT video terminal in the instrument(URK-700) and other instrument
to be able to input NTSC video signal.
And than user could display on external video instrument

- 24 -

7. Self Ins pect ion an d Maint enan ce


7. 1 Be f or e Ca l li n g a Se r v i c e Per s o n
Warning messages will be displayed on the monitor if some problems occur. It might
be operation errors or problems of the machine.
In this case, refer the following instructions.
If the function is still not salvaged or recovered, disconnect the power supply and
consult the dealer.

(1) Message When Power On.


Message Cause Remedy

FOG MOTOR FAIL Turn OFF the power switch and turn on again
after 10 seconds.
Internal error
If the message appears again, consult the
SHUT MOTOR FAIL
dealer.

(2) Message On Measuring


Message Cause Remedy

RETRY Measure after aligning the pupil and the


Alignment is improper
Alignment Mark properly.
Instruct the examinee to open his or her eyes
Eyelid or eyelashes are
wide, or lift up the eyelid lightly with your
covering the pupil.
fingers and measure again
The minimum pupil diameter that can be
measured is 2.0 mm. Although it is possible to
When the pupil is smaller than measure in the bright place, don’t expose
the Outer Alignment Mark. examinee’s eyes to the direct sunlight or too
bright indoor lights to prevent the contraction
of the pupil.
Observe the eye in SIZE Mode. If cataract is not
When the examinee has some
severe, measurement can be performed in the
illness like cataract.
IOL mode.
Examinee has IOL(intra-ocular
Measure in the IOL mode.
lens) implanted.
When the Mire Image is odd
shaped because of tears.
Instruct the examinee to open and close his o
When the Mire Image is not r her eyes several times and measure again.
clear because the cornea is
dry.

Examinee has strong irregular Impossible to measure


astigmatism or corneal disease.

- 25 –
7.2 Replacement of Printing Paper
Replace the roll of printing paper as soon as possible if the red line appears in the
paper.
Make sure to confirm kind and size of the paper(kind: thermal paper, size: 58mm in
width and 50mm in diameter).

7.3 Replacement of chinrest Paper


(1) Pull out the two pins on the chin rest.
(2) Insert the pins into the holes on the chin rest paper. More than 50 sheets of paper
can be attached.
(3) Insert the pins straight into the holes on the chin rest.

7. 4 Cle a n i n g
(1) Basically, keep this instrument clean. Don’t use volatile object, thinner or benzene,
etc.
(2) Polish each part with a dry cloth containing detergent solution.

7. 5 Whe n Mo v i n g the In s t r u m en t
(1) Turn OFF the power switch.
(2) Disconnect the power cable.
(3) Close the stage holding dial in the clockwise direction.
(4) Move this machine holding the lower part of the mains to keep horizontally.

7. 6 Se r v i c e In f o r m a t i o n

(1) Repair
If problem cannot be solved even after taking the measures indicated in section
7.1, contact UNICOS representative or distributor for repair.
Please refer to the name plate and let us have the following information:
Name of the instrument: URK-700
Serial Number: 7-digit characters indicated on the name plate
Phenomenon: In detail

K7FBA3A Serial No.

- 26 –
(2) Limit for Supplying Performance Parts for Repair
Performance part (required to maintain the functioning of the product) of this
product will be stocked for eight years after discontinuation of product, to allow
for repair
(3) Disposal of the instrument

! CAUTION This instrument incorporates a lithium battery, which may


pollute the environment if the instrument is abandoned.
Please ask a professional waste disposal company to handle
disposal or distributor before disposing of the instrument.

- 27 –

8. Specifi cat ions


Refractometry
Vertex Distance(VD) 0.0, 10.0, 12.0, 13.5, 15.0 mm
Sphere Power(SPH) -25.00 ~ +22.00 D (at the vertex distance of 12
mm (Increments selectable between 0.12 and 0.25 D)
Cylinder Power(CYL) 0.00 ~ ±10.00 D
(Increments selectable between 0.12 and 0.25 D)
Axis(AX) 1 ~ 180˚ (Increments: 1˚)
Cylinder Form -, +, MIX
Pupil Distance(PD) 10 ~ 85 mm
Minimum Pupil Diameter 2.0 mm

Keratometry
Radius of Curvature 5.0 ~ 10.2 mm (Increments: 0.01 mm)
Corneal Power 33.00 ~ 67.50 D(when cornea equivalent refractive index is 1.3375)
(Increments selectable from 0.12, 0.25 D)
Corneal astigmatism 0.00 ~ -15.00 D(Increments selectable from 0.12, 0.25 D)
Axis 1 ~ 180˚ (Increments: 1˚)

Corneal Diameter 2.0 ~ 12.0 mm (Increments: 0.1 mm)

Memory of Data 9 measured value for each right and left eye

Internal Printer Uses thermal line printer

Monitor 5.6-inch color TFT LCD monitor

Environmental requirements:
Operation Temperature: +10 to +40℃
Humidity: 30 to 85% RH
Storage and Transportation Temperature: -10 to +55℃
Humidity: 10 to 95% RH

Power supply AC100-240V, 50/60Hz

Power consumption Approximately 100VA

Size 280(W)  476(D)  475(H) mm

Weight Approximately 21kg

- 28 –

9. Components
URK-700 Main Unit ····················································································· 1

Operation Manual ················································································· 1


Power Supply Cable ····················································································· 1

Test Model Eye ······························································································ 1

Chin Rest Paper ········································································ about 100


sheets

Printing Paper ····················································································· 2 rolls

- 29 –

You might also like