Visual inspection – Check that inscriptions, information and warning signs are properly and completely
fixed.
Visual inspection – Perform inspection of battery compartment and battery connection for corrosion and
other contamination.
Rate test – (10 mA; 3.0 mV; 30 ppm) Verify that the pulse-to-pulse interval is within the specified
tolerance. 1800 – 2200 ms
Rate test – (10 mA; 3.0 mV; 70 ppm) Verify that the pulse-to-pulse interval is within the specified
tolerance. 771 – 943 ms
Rate test – (10 mA; 3.0 mV; 180 ppm) Verify that the pulse-to-pulse interval is within the specified
tolerance. 300 – 367 ms
Rapid atrial pacing – Check that the output rate is 320 ppm +/- 10%. (169 – 206 ms)
Rapid atrial pacing – Check that the rapid atrial pacing is available over the range of 80 – 800 ppm +/-
10% using the INCREASE/DECREASE keys.
Pulse width – Verify that the pulse width is 1.8 ms +/- 10% (1.62 – 1.98 ms) as measured from the half
-amplitude point on the leading edge to the half amplitude point on the trailing edge.
Output – (0.1 mA) Verify that the measured outputs are within specified tolerances. 0 – 0.11 V
Output – (5.0 mA) Verify that the measured outputs are within specified tolerances. 2.08 – 2.92 V
Output – (20.0 mA) Verify that the measured outputs are within specified tolerances. 8.45 – 11.55 V
Sensitivity – (0.5 mV) Verify that the threshold falls within the range specified for each setting. 0.2 – 0.8
mV
Sensitivity – (5 mV) Verify that the threshold falls within the range specified for each setting. 3.75 –
6.25 mV
Sensitivity – (20 mV) Verify that the threshold falls within the range specified for each setting. 15.0 –
25.0 mV
Frequency response – (40 ms (25 Hz)) Verify the sensitivity, keeping in mind any percent error to be
added from the initial 3 mV measurement. 3 mV (reference).
Frequency response – (5 ms (200 Hz)) Verify the sensitivity, keeping in mind any percent error to be
added from the initial 3 mV measurement. 46.5 mV +/- 25%.
Refractory period – Check that the signal generator amplitude is 200 mV. The 10:1 attenuator on the
output of the generator will reduce this to a 20 mV pulse at the pacemaker.
Refractory period – (input the squarewave with a 235 ms delay) Verify that the pacing rate remains at 80
ppm (750 ms).
Refractory period – (input the squarewave with a 265 ms delay) Verify that the pace-to-pace interval is
increased to approximately 1015 ms.
AC (50 Hz) – Check that the interference rejection is 2.0 mVp-p noise while sensing 3.0 mV, 40 ms sine-
squared test pulses.