Professional Documents
Culture Documents
5-372, 1996
(:opyright Q 1996 Elsevier Science Ltd for IPEMR
Printed in Great Britain. .All rights reserved
13504533(95)00066-6 1.1,50-xi3~96 $15.00 + 0.00
ELSEVIER
ABSTRACT
An apparatus for simultaneously monitoring heart and respiratory rates was developed using photoplethysmography
(PPG) and digital JUTS, and compared with conventional methods. The PPG signal. which in&&es both heart
and re.@ratory components, was measured at the earlobe with an original transmission mode photopkthysmographic-
deoice. A digital filtering technique was used to distinguish heart and respiratory s@alsfrom the PPG .signal. The
cut-osffrequency oJ the respiratoy signalJiltvr was selected automatically depending on the heart rate. Using digiial
J&ring techniques, head and re.$ratoly signals were separated at rest and during exercise. The digital signal
proces.~or was employed to realize an adaptive and readtimejiltering. The heart rate was calculated by the zenhcrosring
method and the respiratory rate from the peak interual of the filtered signal. 7i, evaluate the newly o!eveloped monilor;
an li(X; for heart rate and a transthoracir impedance plpthysmogram ,for respiraton; rate were monitored simul-
taneously: To obtain higher heart and respiratov rates, exerrisp was perf ormed on an e/ectn’cal bicycle prgomett7:
Heart and respiratov rates calculated by the new method compare to those obtained.from ECG and the tran.tfhorucic
impedanrf pleth~smogram. The maximum mar of heart and re,$iratoq rates 7m~ 10 beats/ min and 7 brealh.c/ min.
rrcpecti~~el~. Coj&<h/ 0 1996 l<lseuier Science I.td li)r IPlGWB.
Keywords: Heart rate, respiratov rate, photoplethysmography;. digital filter, digital signal processc~t-,
ewrcisr
Analogue part
Digital part
Digital signal
PPG signal - - processor -
ND Personal
ECG ___L convertor computer
Transthoracic impedance c
plethysmograph signal
or -
Nasal themistor signal
366
(16 MHz 386SX, PC-9801NS/E, NEC, Japan), a (over 100 dB), each filter has 1024 coefficients.
1%bit A/D converter board (Analog-prol, Can- These coefficients, which were stored as a file,
opus Co., Japan), and a 16-bit fixed point digital were entered into the DSP board from the PC’s
signal processor (DSP) board (ADSP-2101-50, hard disk by a control program before starting a
12.5 MIPS, Analog devices, DSP98-2 101 BPC, recording program. The heart-related wave was
Micro Science Co., Japan). The PPG signal was distinguished using filter no. 1, while the respir-
sampled at 100/s. ation-related wave was distinguished by filters nos.
2-4. The appropriate cut-off frequencies of the
filters were chosen empirically.
Digital filters
Depending on the heart rate, three low-pass fil-
Digital filters are divided into two types; infinite ters were selected automatically. To obtain stable
impulse response (IIR) and finite impulse and smooth filtering, hysteresis inclusion was
response (FIR) filters”. Input and output signals required. Reference heart rates for the filter-elec-
to both types of filter are related by the convol- tion logic are shown in Fipr~ 3. These reference
ution sum. The current output sample, y(n), is heart rates were determined empirically.
given by equation (1) for the IIR and by equation
(2) for the FIR filter:
Evaluation of heart and respiratory rates
The PPG signal was simultaneously recorded with
c b&n-k)
kl
a telemeter ECG monitor (Dyna scope DS-3100,
Fukuda Denshi Co., Japan) for the heart rate and
a transthoracic impedance plethysmograph (TR-
601T, Nihon Kohden Co., Japan) for the respirat-
(2) ory rate. Although transthoracic impedance ple-
thysmography has some problems such as moving
artefact during measurement, difficult-to-evaluate
where nk and bk are the coefficients of the filters,
qualitative values and attachment of electrodes,
and N and M are the values of uk and b,, respect-
we chose it as a reference method because this
ively. In equation (1)) y( 12) is a function of past
technique was an unconstrained and relatively
outputs as well as present and past input samples;
suitable monitor during exercise using the bicycle
the IIR is therefore a feedback system. In the FIR
ergometer if the electrodes were securely
equation y(n) is a function only of past and pre-
attached. The heart rate was calculated from the
sent input values. FIR requires larger N for sharp
filtered heart waveform of the PPG using the zero-
cut-off filters than IIR. Thus for a given amplitude
crossing method. The respiratory rate was calcu-
response specification, more processing time and
lated from the respiratory waveform of the PPG
storage are required for FIR implementation.
using the peak detection method. The peak inter-
However, we chose the FIR for the following
val of the transthordcic impedance plethysmog-
reasons: (1) FIR filters can be constructed to have
ram was counted to obtain a reference respiratory
a strictly linear phase response, so that no phase
rate. In our monitor, heart and respiratory rates
distortion is introduced into the signal by the fil-
were calculated as the median value of ten and
ter. The phase responses of IIR filters are nonlin-
five sequential counts, respectively. The respirat-
ear; (2) FIR filters realized nonrecursively, that is
ory rate obtained from the transthoracic imped-
by direct evaluation of equation (2)) are always ance plethysmogram was also calculated as the
stable. The stability of IIR filters cannot be guaran-
median value of five sequential counts. The heart
teed; (3) the effects of using a limited number of
and respiratory rates displayed on the PC
bits to implement filters such as round-off noise
obtained from PPG and references were stored
and coefficient quantization errors are much less
every 1 s (2 s, except for two subjects) on the RAM
severe in FIR than in IIR; and (4) DSP can calcu-
disk of the PC.
late rapidly enough for real-time monitoring of
heart and respiratory rates (for example, when ,li
= 1024, the calculation time using this DSP was Experimental procedure
165 ps).
To obtain suitable coefficients of the FIR, the To obtain higher heart and respiratory rates, exer-
window method’* was achieved. We could obtain cise was performed on an electrical bicycle erg-
coefficients by evaluating the inverse Fourier ometer (RE800 Ergo system, Rodby Elektronick
transform of an ideal frequency characteristics. AB, Sweden). The subjects were studied in a sit-
Then, white-noise was filtered by FIR filter using ting position on the bicycle ergometer and were
determined coefficients, and the filtered signal allowed to rest for at least 10 min before exper-
was evaluated again by Fourier transform for imental recording. The total recording time was
checking the characteristics of FIR filter. We 20 min. During the first 3 min. subjects were
obtained suitable specifications of the filters, as allowed to rest in a sitting position, and exercise
shown in Table 1. Figure 2 shows the characteristics was then performed for 10 min. The ramp load
of the filters for separating heart- and respiration- was applied increasing by 10 W per min from 30
related waves. The heart-related signal was band- to 130 W. The pedalling rate was maintained at
pass filtered, and a low-pass filter was used for res- 55 rpm. After exercise, the subject rested, sitting
piration. To obtain enough attenuation level on the bicycle for the remaining 7 min.
367
Heart and respiratq rate monitor: K. Nakajima et al.
368
Heart and resplratmy ratr monitor: K Nakajma et al.
(a> PPG
ECG
PPG
(b)
ECG
1s
Figure 4 Heart rate signals measured by PPG (upper linr) and ECG (lower line): (a) at rest; (b) at high heart rate
(4
PPG
Transthoracic
impedance
plethysmogram
(W PPG
Transthoracic
impedance
plethysmogram
Figure5 Respiratov signals measured by PPG (upper line) and transthoracic impedance plethysmography (lower line): (a) at rest; (b) at
high respiratory rate
180
Sitting Exercising Sitting
i --I------------------~-
t
g- 150 -
E
Time (min)
Figure 6 Time course of heart rate obtained by PPG waveform (solid line) and EC<; (dashed line)
369
Heart and respiratq rate monitor: K. Nakajima et al.
180 - DISCUSSION
The apparatus for monitoring simultaneous heart
g 160 -
and respiratory rates was developed using PPG
i-i and digital filtering techniques with DSP, and
Q) 140 - compared with conventional methods. Heart rates
a showed good agreement between our monitor
and the EGG monitor, while the respiratory rate
E 120 - also showed agreement between this new method
CL
and transthoracic impedance plethysmography.
2 100 - However, there were certain problems encoun-
tered in measuring respiratory rate with the new
al
3 80- monitor.
t:
The analogue filters were used to attenuate a
$ 60- higher frequency than the heart rate during
I
exercise and a lower one than the respiratory
40 I I I I I 0 I I I I I I I rate at rest. The sampling theory of A/D conver-
40 60 80 100 120 140 160 180 sion and stop-band of digital filters had a lim-
ited attenuation level. High- and low-frequency
Heart rate with ECG (beatdmin) noises were not attenuated using digital filters
Figure 7 Relationship between heart rate obtained from PPG and only.
ECG. Points and bars indicate mean and standard deviation of heart l The amplitude of the PPG signal varied
rate by PPG between subjects. In this study, the amplifier
gain was adjusted for individual subjects. The
heart signal of the PPG increased with the
The relationship between heart and respiratory increase in blood flow caused by exercise.
rates Therefore, in clinical use the monitor system
Figure 10 shows a typical example of the relation- might require an auto-gain control circuit.
ship between heart and respiratory rates at rest, During exercise, the amplitude of the synchron-
during exercise, and after exercise, omitting the ous respiratory PPG signal was smaller than the
initial 30 s of the recording in the same record as signal at rest. The motion artefact was large in
shown in Figures 6 and 8. The relationship the PPG signal during exercise. Therefore, it
between heart and respiratory rates showed a hys- was difficult to obtain the synchronous respirat-
teresis of counterclockwise rotation in five subjects ory PPG signal clearly at a high respiratory rate
(include Figure IO), clockwise rotation in four (see Figure 5).
subjects, and showed no hysteresis in two subjects.
The size and the rotary direction of hysteresis For these reasons, the error increased with the
depended on the subject. respiratory rate. In order to distinguish the respir-
20
IO - ."- 1 '....1
No.4
No.3 No.3
f No.2 No.2
0 /, I ,,I, I a I I I I I1 Is I I II
0 5 10 15 20
Time (min)
Figure 8 Time course of respiratory rate obtained by PPG waveform (solid line) and transthoracic impedance plethysmography (dashed line)
370
Heart and respiratmy ratp monitor: K. Nakajima et al.
371
Heart and respiratgr rate monitor: K. Nakajima et al.
probe for micro-vascular surgery. Microsurgeq 1984; 5: 9. Astrand PO, Rodahl K. Respiration. In Textbook of Work
105-6. Physiology, 2nd edn, Chap. 7. McGraw-Hill, New York,
4. Dorlas JC, Nijboer JA. Photoelectric plethysmography as 1977; 207-66.
a monitoring device in anaesthesia. BrJAnaesth 1985; 57: 10. NakaJima K, Tamura T, Ohta T, Miike H, oberg PA. Pho-
524-30. toplethysmographic measurement of heart and respirat-
5. Lindberg LG, Ugnell H, Oberg PA. Monitoring of respir- ory rates using digital filters. In Proc 15th Ann Znt Con/
atory and heart rates using a fiber-optic sensor. Med Biol IEEE Eng Med Biol Sot, Part 2, San Diego, CA, 1993;
Eng Comput 1992; 30: 533-7. 1006-7.
6. Vegfors M, Ugnell H, Hok B, Oberg PA, Lennmarken C. 11. Nakajima K, Tamura T, Miike H. Photoplethysmography
Experimental evaluation of two new sensors for respirat- for simultaneous monitoring of heart and respiratory
ory rate monitoring. Physiol Meas 1993; 14: 171-81. rates using digital filters. JpzJ Med Elec Biol Eng 1993; 31:
7. Sara CA, Shanks CA. The peripheral pulse monitor-a 326-33 (Abstract in English).
review of electrical plethysmography. Anuesth Zntens Cure 12. Ifeachor EC, Jervis BW. A framework for digital filter
1978; 6: 226-33. design; Finite impulse response (FIR) filter design. In
8. Farag AA, Tacker WA, Foster KS, Bourland JD, Geddes Digital Signal Processing, Dagless EL, O’Reilly J, eds, Chaps
LA. Detection of pulse and respiratory signals from the 5, 6. Addison-Wesley, Woking, 1993; 251-73.
wrist using dry electrodes. Biom Znst Tech 1994; 28: 311-4.
372