Received on 07 May 2019; received in revised form, 09 September 2019; accepted, 13 November 2019; published 01 March 2020
RESULTS AND DISCUSSION: The temperatures at injector and detector were set
Optimization of Conditions: Dichloromethane at 200 ºC and 260 ºC. Using the above stated
was selected as a diluent as it did not show optimized conditions, epichlorohydrin was eluted
interference peak with epichlorohydrin peak. Gas from the column at a retention time of 11.312 min
chromatography column CP-Volamine (length 60 Fig. 2.
m × identification 0.32 mm, film thickness 5.0 µm)
yielded a good peak shape. Nitrogen was used as Validation: The proposed gas chromatography
carrier gas to get a steady and constant baseline. with flame ionization detection method was
Column flow rate, split ratio and injection volume validated using the guidelines of ICH and USP 20,
21
were set at 2 ml/min, 1:1 and 4 µl, respectively as .
these set values gave adequate sensitivity and
Specificity: To evaluate the method's specificity,
excellent precision.
dichloromethane blank, epichlorohydrin standard
solution (20.8 ppm), linezolid sample (100 mg/ml)
and epichlorohydrin spiked linezolid sample (20
ppm) preparations were subjected to gas
chromatography with flame ionization detection.
Chromatograms were collected.
FIG. 3C: LINEZOLID SAMPLE (100 mg/ml) FIG. 3D: EPICHLOROHYDRIN (20 ppm)
CHROMATOGRAM SPIKED LINEZOLID SAMPLE
Precision: Precision is the degree of repeatability The method’s precision was evaluated by
under standard conditions for an analytical method. calculating the relative standard deviation of six
replicate peak area determinations by injecting Limits of Detection (LOD) and Quantitation
freshly prepared 20 ppm solutions of (LOQ): Standard epichlorohydrin solutions were
epichlorohydrin separately on the same day. The % separately injected. The detection limit and
RSD value was 1.18%. The low % RSD values via quantification limit was then determined at the
peak areas confirm the method’s good precision. lowest concentration with a signal to noise
proportion of 3 and 10, respectively, under the
TABLE 1: PRECISION STUDY RESULTS FOR
provided experimental conditions. The limits of
EPICHLOROHYDRIN
Injection no. Epichlorohydrin peak area detection and quantification values for
1 65.35 epichlorohydrin were found as 2.03 ppm and 5.12
2 63.86 ppm, respectively. The values show the satisfactory
3 63.31
4 64.77 sensitivity of the method for detecting and
5 64.76 monitoring epichlorohydrin. Epichlorohydrin
6 64.98 chromatograms at LOD and LOQ concentrations
Average value 64.5
% RSD 1.18 are illustrated in Fig. 4A and 4B.
Linearity: Linearity was evaluated by series of more than or equal to 80% and should not be more
injections of six standards with concentrations than 120%. The recovery of epichlorohydrin at four
ranging from LOQ to 150% (5.12 ppm, 10 ppm, 16 specification levels was assessed Table 3. The
ppm, 20 ppm, 24 ppm, and 30 ppm) of the expected recovery data show method’s accuracy.
concentration range. The peak area of each
concentration was determined using the described Robustness: Method’s robustness was evaluated
gas chromatography conditions. The response was by making minor but intentional variations in
proportionate to epichlorohydrin concentration. column, injector, and detector temperatures. The
The calibration curve across the range of 5.12 to effect of variations has been studied on the peak
31.3 ppm was drawn between peak areas versus area of epichlorohydrin. The results are shown in
epichlorohydrin concentration. The slope, intercept Table 4. The percentage of RSD values of the peak
and correlation coefficient values were derived area found to be inside the acceptance criteria in all
from linear least square regression treatment Table varied conditions.
2. The method’s good linearity was proven by the TABLE 2: LINEARITY AND LEAST SQUARE REGRESSION
correlation coefficient viz. 0.9986 for TREATMENT RESULTS FOR EPICHLOROHYDRIN
epichlorohydrin. Level (%) Concentration (ppm) Area of peak
LOQ 5.12 16.6
Accuracy: Method’s accuracy was evaluated by 50 10 31.3
standard method of addition. A known amount of 80 16 54.1
100 20 66.9
epichlorohydrin spiked at four specification levels 120 24 78.5
(LOQ, 50%, 100%, and 150%) to the linezolid 150 30 103.9
sample solution. The spiked solutions were Correlation coefficient 0.9986
analyzed in triplicate by described gas Slope 3.476
chromatography conditions. The recovery must be Y-Intercept -2.3504