Professional Documents
Culture Documents
Cit0121 Supp MDT Combined
Cit0121 Supp MDT Combined
Sponsored by Medtronic
Funding for this supplement provided by Medtronic
T
ranscatheter aortic valve replacement (TAVR) with the Traditional deployment of the CoreValve and subsequently
self-expanding, supra-annular CoreValve and Evolut R Evolut bioprosthesis had relied upon obtaining a three-cusp
bioprostheses (Medtronic) has been associated with coplanar view and then adjusting the gantry angle, generally
the need for permanent pacemaker implantation (PPI) caudal, to eliminate parallax from the capsule marker
in 15.7% to 28.6% of patients within 30 days after the band. Subsequently, Piazza and colleagues demonstrated
procedure.1-9 The frequent requirement for PPI has been one that the three-cusp technique with marker adjustment
of the main challenges influencing an operator’s willingness distorted the native anatomy on fluoroscopy.10 Using the
to perform TAVR with this bioprosthesis. We present a novel three-cusp technique, the operator often deployed the valve
method that describes a series of procedural steps designed without a clear representation of the location of the native
to provide a controlled implantation depth and minimize annulus relative to the conduction system. The location of
the interaction of the bioprosthesis with the atrioventricular the conduction system was obscured by the radiographic
(AV) conduction system, potentially reducing the foreshortening of the left ventricular outflow tract in
subsequent clinical sequelae of conduction disturbances. noncusp overlap views.
Figure 1. Cusp overlap view. The left panel illustrates the cusp overlap position with the NCC (yellow), RCC (green), and left
coronary sinus (red). The right panel shows the volume rendering of the cusp overlap view showing the commissure of the non- and
right coronary sinus (courtesy James Harvey, MD).
A B C
Figure 2. Location of the AV node and basal annular plane. The blue AV node is shown best in the cusp overlap view (A) with an
increased distance from the AV node to the basal annular plane. The three-cusp coplanar view generally requires caudal angulation
to remove the parallax from the marker band and the AV does appear closer to the basal annular plane due to foreshortening of the
LVOT (B). The LAO view shows the AV node with overlaps the RCC and NCC and isolates the LCC (C). The LVOT is also foreshortened
in this view.
To develop a strategy that minimizes the risk of unforeshortening of the LVOT, there is also a greater distance
interaction of the conduction system with the bioprosthesis, between the NCC insertion and the compact AV node
a better understanding of the location of the conduction (Figure 2). This fluoroscopic view may lead to a more precise
system relative to the aortic annulus basal plane was 3-mm implantation depth, thereby minimizing the risk of
needed. The noncoronary cusp (NCC), the most inferiorly interaction with the conduction system.
oriented cusp in the left ventricular outflow tract (LVOT), In addition to the cusp overlap view, several procedural
is isolated from the left coronary cusp (LCC) and the right steps have been embedded in this technique to further
coronary cusp (RCC) using the cusp overlap technique. reduce interaction with the conduction system during
Below the non-right commissure, the membranous septum TAVR deployment. We now focus on more of a “top-down”
houses the conduction system. More ventricular to the deployment of the THV, starting with the catheter marker
membranous septum, which may be of variable length, these band positioned at the midportion of the pigtail in the
conduction fibers become more superficial in the muscular NCC (Figure 3). With retraction of the nitinol capsule, the
septum. If the transcatheter heart valve (THV) is deployed inflow of the prosthesis advances across the annulus and
below the membranous septum, there is a greater chance is positioned at 3 mm below the annulus. This maneuver
of interaction with the conduction system. Jilaihawi et al avoids traumatic advancement of the bioprosthesis into the
described the variation in PPI rates depending on the length ventricle with inflow flaring deeper within the left ventricle,
of the membranous septum and the depth of the THV. Their resulting in subsequent maneuvers to move it more aortic
series suggested that when a THV is deployed below the and interacting with the muscular septum and conduction
membranous septum, the risk of PPI was increased.4 system. We also use a stiffer, double-curved, Lunderquist wire
(Cook Medical) in most cases to maintain a wire position
CUSP OVERLAP METHOD in the non-right commissure and begin the prosthesis
Our method demonstrates that by isolating the NCC deployment along the posterior aspect of the annular plane.
and overlapping the NCC/RCC commissure along the basal The stiffer wire may result in more symmetrical deployment
annular plane, the implantation view can be optimized and is especially valuable when deploying larger-sized THVs.
during THV deployment. This view can be easily identified Using the cusp overlap view to maintain a reference to the
during computed tomography (CT) reconstruction (3Mensio, native annular plane, the marker band on the THV delivery
Pie Medical Imaging) and contribute to preprocedural case catheter does tend to lose parallax when approaching
planning (Figure 1). The cusp overlap view unforeshortens the valve plane. The loss of parallax of the marker band
and elongates the LVOT and accentuates the NCC/ is the result of the delivery catheter following the stiff left
RCC commissure in the center of the fluoroscopic view, ventricular wire that is generally positioned in the NCC/RCC
creating an implant that references the conduction commissure. This approach may lead to more confidence
system and separates it from the annular plane. Given the in the initial positioning of the THV in relation to the
A B
Figure 3. Location of the marker band at initial deployment. The marker band is placed at the mid position of the pigtail catheter
that is positioned at the lowest portion of the noncoronary sinus (A). With retraction of the nitinol capsule, the inflow of the Evolut
bioprosthesis advances to a 3-mm position below the annulus (B).
insertion of the NCC and a better assessment at the point single-center experience utilizing cusp overlap found that
of no-recapture. We also favor sufficient pacing during the only one (1.5%) in 65 patients required permanent pacing.
deployment to minimize cardiac output and the occurrence As the highest enrolling center in the Evolut low-risk trial, we
of premature ventricular contraction burden, allowing for a attribute this difference to the comprehensive use of the cusp
stable bioprosthesis deployment. Finally, once we are at 80% overlap technique at our institution.
deployment, we rotate the gantry to an LAO projection to We have also analyzed our experience in 134 patients
visualize the LCC and ensure that the inflow is not supra- without antecedent PPI who underwent placement of
annular. We aim for an implantation depth of 3 mm, and the 34-mm Evolut R THV, which has been shown to be a
no deeper than 5 mm, below the NCC to reduce our risk of predictor of PPI,11 independent of membranous septum
conduction disturbance. We occasionally aim for shallower length.4 After preprocedural computed tomography (CT)
deployment in patients at high risk for conduction system planning to determine the optimal cusp overlap projection
abnormality but recommend recapture for bioprosthesis angle, we were able to achieve the projected cusp overlap
positions < 1 or > 5 mm within the ventricle. Once final gantry view on CT reconstruction in 88% of patients, and
positioning is confirmed, we retract the left ventricular wire, a near cusp overlap projection in the remaining patients.
centralize the nose cone, and slowly release the delivery In this series, we obtained an effective 30-day PPI rate of
catheter from the bioprosthesis by the release of the frame 5.2%.11 In addition, the new-onset left bundle branch block
paddles. We are cautious to avoid interaction of the delivery rate was 10.9%, below contemporary rates previously
catheter and the bioprosthesis as the delivery catheter is published.2 When compared to previous studies (Figure 5),4-9
retracted into the aorta. we believe that the cusp overlap view contributed to the
lower rate of PPI rates in our series. The Optimize PRO study
THE UPMC PINNACLE EXPERIENCE WITH (NCT04091048) will provide additional insights into the safety
CUSP OVERLAP and efficacy of this method in a broader array of patients at
After starting with routine use of the cusp overlap multiple centers.
technique in October 2015, we have observed a significant
decline in the post-TAVR PPI rates at our institution, especially PROCTORING EXPERIENCE IN ADOPTION OF
with the Medtronic CoreValve and Evolut platforms. In the CUSP OVERLAP
Medtronic Low Risk trial, the rate of PPI 30-days post-TAVR Prior to the adoption of the cusp overlap technique as the
was 17.4%,2 with wide site-specific variability (Figure 4). Our standard for Medtronic self-expanding THV deployment, we
Figure 4. Funnel plot of site-level variability of 30-day post-TAVR PPI rate in the Medtronic Low Risk trial. Each dot represents an
enrolling center and UPMC Pinnacle had the highest enrollment with a PPI rate lower than 2 confidence intervals, using the cusp
overlap technique for all TAVR deployments.
were able to teach the cusp overlap technique successfully to experienced a reduction in the 2019 TAVR mean length
several international centers, including those in the United of stay (LOS) to 1.23 days after 386 TAVR procedures, at
States. A focused didactic experience involving several Latin least partially due to a reduction in the need for new PPI
American centers was initiated in July 2018.12 There were implantation. More specifically, an analysis of 567 TAVR
eight formal encounters with Latin American physicians, cases performed at our institution between July 2015 and
involving didactics through programmatic lecture, proctored June 2018 included 34 patients who required PPI during
and presented cases, as well as case observation at our their hospital stay. The average cost of PPI was $4,415, but
institution. We analyzed the results of these experiences the patients who went onto receive PPI had a significantly
extending to October 2019.12 Fourteen implanting physicians longer LOS than those who did not (4.49 vs 1.75 days),
from seven countries performed consecutive procedures resulting in a modified contribution margin differential of
on 114 patients. Each physician implanted only 22.6 ± $12,654 per case. Although the reasons for this reduction
10.9 THVs the previous year, with a lifetime experience of in LOS differential were multifactorial, a primary reason
129 ± 110 THVs. Of the 114 patients, 105 (92%) did not was related to the reduced postprocedural conduction
have a prior PPI. In this series, the in-hospital rate of new disturbances and the need for a temporary or permanent
PPI post-THV was 5.7%. During the 30-day follow-up that pacemaker. For example, patients who developed
included 85 patients, no additional patients required a new temporary conduction disturbances requiring temporary
permanent pacemaker. These results suggest that the use pacing were not placed on the “fast track” protocol
of the cusp overlap in lower volume operators resulted in and early ambulation. In those patients who required
reduced rates of PPI and may be useful to potentially shorten permanent pacing, usually on postprocedural day 1, an
the learning curve for this self-expanding bioprosthesis. additional 24-hour period of bed rest was needed, leading
to slower periprocedural recovery and increased resource
OPTIMIZING POSTPROCEDURAL MANAGEMENT utilization to facilitate suitable disposition.
Our institution has focused on optimizing
postprocedural management over the past several years, LIMITATIONS AND MITIGATIONS
which has included an effort to reduce resource utilization There are three potential limitations for the cusp overlap
during the hospitalization period. Our institution method that will be informed by additional single and
Figure 5. Rate of PPI in the UPMC Pinnacle 34-mm Evolut R series. Using the cusp overlap technique, the lowest reported rate of
30-day PPI with the 34-mm Evolut R was achieved.4-9
multicenter studies, including the Optimize PRO study. One A third limitation is the use of a stiff Lunderquist wire.
concern relates to higher implantation of the self-expanding Our technique supports the use of the double-curved
bioprosthesis with its 13- to 14-mm skirt height that raises Lunderquist wire in most ventricular anatomies. We believe
concerns for coronary access. Preprocedural CT planning that the stiffer wire provides several advantages, including
can mitigate this risk by assessing the heights of the coronary biasing the valve toward the posterior aspect of the annular
arteries and placing the bioprosthesis at a position to allow plane, eliminating parallax in the marker band, increasing
coronary access after TAVR. efficiency, and providing a more predictable deployment.
Another potential limitation of the cusp overlap Like all ventricular wires, the possibility of ventricular
technique is the operator’s anxiety that a shallow TAVR perforation is always a consideration. We deploy the
implantation may lead to a higher rate of valve embolism Lunderquist through a pigtail catheter established toward
(“pop-outs”), potentially resulting in increased procedural the apex of the left ventricle. After unsheathing the wire,
complexity and patient morbidity. In our experience, we there is no forward manipulation, thereby reducing the
have not found this to be the case, and required the use of a possibility of trauma.
second THV in only one (0.5%) of approximately 200 cases.
In fact, the cusp overlap technique may facilitate a better INNOVATION AND IMPACT WITHOUT DEVICE
understanding of the true THV depth relative to the NCC, ITERATIONS
and importantly, the target 3-mm implantation depth in Our description of the novel cusp overlap technique
relation to the NCC is standard for use for all THV platforms. includes both optimization of the CT image reconstruction
Because the LVOT is elongated in this particular view, our to define an ideal implant angle, and simplification of the
assessment of the depth assessment is likely more accurate. procedure to minimize the interaction of the transcatheter
In a non-cusp overlap view, the LVOT is foreshortened, bioprosthesis and the conduction system. The simplified
leading to the false perception that THV is positioned cusp overlap technique has the potential to enhance clinical
shallower relative to the NCC insertion. Another reason to outcomes by focusing on procedural modifications rather
position the THV at a 3-mm implantation depth is that if than to rely on new technology or expanded infrastructure
postdilation is indicated, there is a buffer for the potential for clinical improvements. In our institution, the cusp
shortening of the THV during postdilatation. (Continued on page 19)
A
s transcatheter aortic valve replacement (TAVR) is Ochiai et al studied coronary positioning with
now approved in lower-risk, younger patients, lifetime computed tomography (CT) after implantation of
management of this population is becoming more 66 Evolut R/Evolut PRO valves from December 2015
important.1-5 In surgical aortic valve replacement to November 2017. Coronary reaccess was considered
(SAVR), after excision of the native valve leaflets, unfavorable if the ostium ended up below the
commissural alignment of the new prosthesis is guaranteed. 13/14-mm skirt (depending on valve size implanted) or if
This has been demonstrated by Fuchs et al who studied the transcatheter heart valve (THV) commissure landed
SAVR implantation and found that all commissures were near the ostium. Results showed that the skirt eclipsed
within 30° of native commissures.6 Cardiac surgeons never the left coronary artery (LCA) in 12.8% of cases and the
intentionally rotate the valve off its native axis nor do they right coronary artery (RCA) in 3% of cases. The reason for
compromise the depth of annular sutures. no immediate hemodynamic collapse was that there was,
Because the prevalence of significant coronary artery on average, 4.2 to 4.8 mm of horizontal distance between
disease is 40% to 75% in patients with severe aortic the coronary ostia and the skirt blocking them. Coronary
stenosis undergoing TAVR, one must consider the need access was also obscured when the commissural triangle
for coronary reaccess when placing a TAVR valve.7,8 In one (up to 26 mm at the posts) blocked the LCA in 22.7%
study of 779 patients undergoing TAVR, 4.7% had acute of cases and the RCA in 21.2%.8 This led to the inability
coronary syndrome within 1 year after TAVR and 10% did to selectively engage the coronaries during coronary
during a median follow-up of 25 months. These patients angiography or percutaneous coronary intervention
were on average younger than the total population.9,10 Our after TAVR.
group has studied optimal commissural alignment with the Similarly, Abdelghani et al looked at 101 patients
cusp overlap technique using self-expanding Evolut valves from October 2015 to June 2019 who had Evolut R/PRO
(Medtronic) to decrease likelihood of future coronary
access challenges.
C-tab
COMMISSURAL ALIGNMENT FOR CORONARY
REACCESS
Barriers to coronary reaccess after TAVR include the
native aortic valve leaflets, the stent frame (whose distal "Hat" Marker
end extends above the sinotubular junction), and the
commissures if not properly aligned.11 The likelihood of the
frame interfering with coronary reaccess is highest when
there is a small angle to the commissure and short inflow Figure 1. C-tab loaded 90° clockwise from the “Hat” marker in
to ostium distance.10 the Evolut TAVR system.
A B
C
Figure 4. “Hat” marker in the OC
position (A). Commissural alignment
confirmed on post-TAVR CT (B).
Straightforward coronary access with
commissural alignment (C).
is to optimize overlap without needing another arterial marker was OC as seen in the left anterior oblique (LAO)
puncture, one pigtail can be placed in the right coronary 14°/caudal (CAU) 22° image (Figure 4A). The valve was
cusp and another pigtail, a 0.035-inch J wire or an 0.018- implanted in the three-cusp view. Commissural alignment
inch nitinol wire in the left coronary cusp, which is then was confirmed on post-TAVR CT (Figure 4B). One week
removed once the cusp overlap view is confirmed.13 later, the patient returned with anginal symptoms.
With improved visualization of the nadir of the Secondary to commissural alignment, coronary access was
non-coronary cusp using the cusp overlap technique, straightforward and no coronary obstruction was identified
permanent pacemaker implantation (PPI) can be reduced. in this case (Figure 4C).
A B
Figure 5. Three-cusp and cusp-overlap views. “Hat” marker at the CF position (A). C-tab at the inner curve with ideal
commissural alignment (B).
1. Buzzatti N, Romano V, De Backer O, et al. Coronary access after repeated transcatheter aortic valve implantation: a glimpse
view was in the CF position on cusp-overlap and the into the future. JACC Cardiovasc Imaging. 2020;13:508-515. doi: 10.1016/j.jcmg.2019.06.025
C-tab landed at the inner curve, with ideal commissural 2. Rogers T, Greenspun BC, Weissman G, et al. Feasibility of coronary access and aortic valve reintervention in low-risk TAVR
alignment after deployment (Figure 5A, B). Having the patients. JACC Cardiovasc Interv. 2020;13:726-735. doi: 10.1016/j.jcin.2020.01.202
3. Alexis SL, Zaid S, Sengupta A, et al. TAVR aortic root orientation: implications for future coronary access and redo TAVR. Ann
“Hat” marker at the OC position in the three-cusp view or Cardiothorac Surg. 2020;9:502-504. doi: 10.21037/ACS-2020-AV-10
CF in the RAO cusp-overlap view allows for optimal results. 4. Khan M, Senguttuvan NB, Krishnamoorthy P, et al. Coronary angiography and percutaneous coronary intervention after
transcatheter aortic valve replacement with Medtronic self-expanding prosthesis: Insights from correlations with computer
tomography. Int J Cardiol. 2020;317:18-24. doi: 10.1016/j.ijcard.2020.05.065
CONCLUSION 5. Yudi MB, Sharma SK, Tang GH, et al. Coronary angiography and percutaneous coronary intervention after transcatheter aortic
Although the self-expanding Evolut valve has tall valve replacement. J Am Coll Cardiol. 2018;71:1360-1378. doi: 10.1016/j.jacc.2018.01.057
6. Fuchs A, Kofoed KF, Yoon SH, et al. Commissural alignment of bioprosthetic aortic valve and native aortic valve following
commissural posts, we have found a valve positioning surgical and transcatheter aortic valve replacement and its impact on valvular function and coronary filling. JACC Cardiovasc
methodology that lowers risk of coronary obstruction Interv. 2018;11:1733-1743. doi: 10.1016/j.jcin.2018.05.043
7. Goel SS, Ige M, Tuzcu EM, et al. Severe aortic stenosis and coronary artery disease- implications for management in
while reducing the risk of interaction with the conduction the transcatheter aortic valve replacement era: a comprehensive review. J Am Coll Cardiol. 2013;62:1-10. doi: 10.1016/j.
system. This is increasingly important in the lifetime jacc.2013.01.096
management of lower-risk and younger patients. By 8. Ochiai T, Chakravarty T, Yoon SH, et al. Coronary access after TAVR. JACC Cardiovasc Interv. 2020;13:693-705. doi: 10.1016/j.
jcin.2020.01.216
inserting the flush port away from the operator, we 9. Vilalta V, Asmarats L, Ferreira-Neto AN, et al. Incidence, clinical characteristics, and impact of acute coronary syndrome
are able to control the “Hat” marker in the OC/CF following transcatheter aortic valve replacement. JACC Cardiovasc Interv. 2018;11:2523-2533. doi: 10.1016/j.jcin.2018.09.001
10. Abdelghani M, Landt M, Traboulsi H, et al. Coronary access after TAVR with a self-expanding bioprosthesis: insights from
position > 90% of the time to facilitate future coronary computed tomography. JACC Cardiovasc Interv. 2020;13:709-722. doi: 10.1016/j.jcin.2020.01.229
engagement. Using the cusp overlap technique, we bring 11. Tang GHL, Kaneko T, Cavalcante JL. Predicting the feasibility of post-TAVR coronary access and redo TAVR: more unknowns
precision to valve deployment to achieve the target than knowns. JACC Cardiovasc Interv. 2020;13:736-738. doi: 10.1016/j.jcin.2020.01.222
12. Tang GHL, Zaid S, Fuchs A, et al. Alignment of Transcatheter Aortic-Valve Neo-Commissures (ALIGN TAVR): impact on final
implantation depth, providing physicians the tools to valve orientation and coronary artery overlap. JACC Cardiovasc Interv. 2020;13:1030-1042. doi: 10.1016/j.jcin.2020.02.005
reduce the risk of TAV interaction with the conduction 13. Tang GHL, Zaid S, Michev I, et al. “Cusp-overlap” view simplifies fluoroscopy-guided implantation of self-expanding valve in
system. n transcatheter aortic valve replacement. JACC Cardiovasc Interv. 2018;11:1663-1665. doi: 10.1016/j.jcin.2018.03.018
T
o minimise the risk of interaction with the conduction the left and noncoronary cusps. These angulations have the
system, precise sizing, positioning, and expansion of the effect of elevating the right coronary cusp (RCC) above the
Evolut valve (Medtronic) are required. An ideal valve remaining left and noncoronary cusps and so, if the device is
position would include a position just below the annular deployed below the still visible non-coronary cusp (NCC) and
plane, and coaxial alignment with the left ventricular left coronary cusp (LCC), the operator can then be confident
outflow tract (LVOT). This would ensure the maximum that the device will deploy below all three cusps, thereby
expansion diameter within the annulus to minimise the risk minimizing the risk of a pop out.
of paravalvular leak, and a high implantation depth would However, the problem with this approach has been
reduce the chance of compression of the conduction system, uncertainty regarding the true depth of the valve using
and the subsequent need for a permanent pacemaker. off-annular plane projections. A potential solution was first
However, in practice, upward displacement of the valve can proposed by Piazza et al, who realized that the annulus
occur during release, due to systolic ejection, valve shortening remained in-plane in a range of views with different relative
during expansion, and melon seeding when the LVOT is cusp positions, described by an S curve, and that the device
narrow—and if the valve is placed too high at the outset, a plane could similarly be imaged from different angiographic
‘pop-out’ can occur. Furthermore, coaxial alignment may be angles, described by a second S curve. From transfemoral
difficult to achieve due to the differing axes of the ascending access, Piazza et al described how these two S curves intersect
aorta (adopted by the valve) and the LVOT (of which the in an RAO caudal view in most cases. This view, being the
annulus is its proximal cross-section). unique projection in which both the device and annulus are
In practice, the best results are therefore achieved by in-plane, allows the true device depth below the annulus to be
minimising the depth of the device below the noncoronary assessed.1
cusp (the deepest cusp), whilst allowing for a small degree Determining the S curve of the annulus is easily
of upward movement during expansion. This requires accomplished using computed tomography (CT) imaging
precise positioning, as well as control of the device during software, such as 3mensio (Pie Medical Imaging BV) at the
deployment. time of procedural planning. However, determining the
S curve of the device requires image analysis at the time of
JUDGING THE PROSTHESIS DEPTH the procedure and is not routinely available. This led to initial
Medtronic's previous guidance for placement of the Evolut difficulties adopting this technique.
valve has been to first align the lowest points of the three However, operators realized that the solution could be
cusps to create an annular plane three-cusp view. From estimated by simply choosing an eccentric RAO caudal angle
transfemoral access, the device is often poorly aligned in this from the annular S curve generated from the CT data and
view, meaning that different sides of the expanded valve will that this view not only aligned the device and annular planes
occupy different depths with respect to a line joining the base but frequently also overlapped the LCCs and RCCs along the
of the cusps (the annular plane). It is, therefore, not possible annular plane, leading to the term cusp overlap view (COV).
in this view to be confident that the expanded valve will be Medtronic has now adopted use of this view, in addition to
below all three cusps. Medtronic’s guidance has therefore other procedural tweaks, to create the current procedural
been to add either caudal angulation (common in the guidance.
United Kingdom) or left anterior oblique (LAO) angulation
(common in the United States) to this three-cusp view to MANCHESTER EXPERIENCE
achieve a device plane. Using these device plane projections, When we transitioned from a Sapien-only centre (Edwards
the operator can adjust the device position so that it is below Lifesciences) to include the Evolut R valve in 2015, we had
A B C
Figure 2. Valve ring level with the annulus (red line) in three-
cusp view but valve not in plane (valve ring oval shape) (A).
Following LAO angulation, valve in plane but now looks high
Figure 1. Single-centre experience (N = 487). Trend toward relative to the pigtail and a line joining NCC and LCC (B). Valve
lower pacemaker (P = .13) and left bundle branch block (LBBB) at optimal height (blue line) 3 mm below the annular plane (red
rates using Evolut R in comparison to Sapien 3 valves. line) (C).
been used to only implanting in the three-cusp annular plane caudal or LAO angulation was applied (Figure 3). We've
view (as per Edwards Lifesciences instructions for use). We, calculated that each 10° of caudal or LAO angulation off the
therefore, felt uncomfortable coming off the annular plane
and were attracted to ways to visualise both the valve and
annulus in plane. After experimentation with RAO caudal MANCHESTER CUSP OVERLAP TECHNIQUE
annular projections predicted by the CT analysis, we similarly
adopted the cusp overlap view as a convenient place to start. PROCEDURAL STEPS
We were pleased by how often the cusp overlap technique
1. Image in three-cusp view to verify S curve angles
aligned both the annulus and device and were struck by the
from CT. Apply any difference from CT angle to all
low rate of pacemaker implantation, trending to be lower than
subsequent views.
our Sapien experience (N = 487; Evolut, 4.6%; Sapien 3, 8%
pacemaker at 30 days) and the low rates of perivalvular leak 2. Switch to COV. With the pigtail at the base of the
(Figure 1).2,3 NCC, slowly unsheath the valve (up to a point before
In the lab, we have found that the COV has been a very annular contact) while keeping the base of the valve
approximately 1 mm horizontally below the base of
workable view with both RAO and caudal angulations < 30°
the pigtail. This will be approximately 3 mm below the
in roughly 80% of cases. In the remaining cases, we have found
annular plane, defined as the line joining the base of the
that a less extreme view derived from the CT annular S curve
NCC and LCC.
in between the three-cusp and cusp overlap views works well
with good device alignment. 3. We then deploy the valve to 80% in the cusp overlap
Furthermore, as the COV puts the lowest point of the view, under rapid pacing to reduce cardiac output
NCC lateral to the valve, rather than behind the valve (as in and minimize the likelihood of movement. During
the three-cusp view and modifications) visualisation of its deployment aim to keep the valve 1 mm horizontally
below the base of the pigtail held at the base of the
position is further simplified (Figure 2). With the pigtail firmly
NCC. Check valve depth below a line joining the NCC
at the base of this cusp, the device can be easily adjusted
and LCC at 80%—aiming for 2-4 mm below this line.
relative to the base of the pigtail. We have also been confident Consider recapture if < 1 or > 5 mm.
that when the valve is below the NCC in the COV it will
be below all three cusps in almost all cases. As Medtronic 4. Assess expansion and depth below the LCC in this view.
now recommends, we aim to place the valve at a target 5. Check in the LAO cranial view to judge tension on
implantation depth of 3 mm, with a range of 1 to 5 mm; the valve before release, with the aim of a mid aortic
shallower implantations may lead to pop-outs during position. Check device below LCC, risk of coronary
postdilation, whereas deeper implants increase the pacemaker obstruction, degree of aortic regurgitation in this view.
rates. This cusp overlap technique has allowed us to greatly Check degree of valve expansion in both views.
simplify deployment (Figure 2). 6. Do not release unless a good position is achieved.
The effect of caudal and LAO angulation off the annular Recapture and reposition the valve, if necessary, to
plane has been further examined in a modeling study.3 This obtain an optimal position.
study showed that the valve was elevated above the annular
plane in a similar way with respect to the RCC whenever
T
ranscatheter aortic valve replacement (TAVR) is a crest of the interventricular septum within the interleaflet
safe and effective alternative to surgical aortic valve triangle separating the NCC and RCC. Kawashima et al
replacement for the treatment of symptomatic described three variants of bundle of His anatomy: (1) the
severe aortic valve stenosis across all surgical bundle courses within the right half of the MS in 50%
risk categories.1-7 In 2007, when TAVR was first of cases, (2) within the left half of the MS in 30%, and
commercialized in Europe, such a statement would have (3) coursing on to the MS just under the endocardium in
not seemed possible for a nascent technology that was 20%.11 The latter two anatomic variances would increase
plagued by high rates of three important complications: the risk of new conduction disturbances during TAVR.
postprocedural paravalvular leak, vascular complications,
and new permanent pacemaker implantation (PPI). A
Progressive iterations of transcatheter heart valve (THV)
design, preprocedural planning, and implantation
technique have seen these complications become less
frequent. Higher rates of new PPI have been the most
challenging issues to address due to the anchoring
mechanism of most THV technology (radial force) and
the proximity of the implantation site to the cardiac
conduction system. Current rates of new PPI range from
2% to 36% depending on the THV design and the patient
population studied.8 Although the clinical impact of new
PPI after TAVR has been somewhat controversial, meta-
analyses have suggested an increased risk of all-cause
mortality at 1 year in patients receiving a new permanent
pacemaker.9 Certainly, the requirement for PPI is B
associated with longer hospital stay and increased costs.10
The cusp overlap implantation technique has the
potential to lower the risk of interaction with the
conduction system by providing a more accurate
assessment of THV depth. In this article, we review the
rationale and practicalities of using this technique.
Figure 2. Double S-curve and chamber anatomy. The double-S curve gives a fluoroscopic projection where the aortic annulus is
aligned and the delivery catheter parallax is removed. This fluoroscopic view corresponds to the point where the aortic annulus
S-curve (yellow curve) and the delivery catheter S-curve (blue curve) cross and is usually obtained in an RAO-caudal view. The LAO-
cranial view corresponds to a four-chamber view where the LVOT is foreshortened, while the RAO-caudal view corresponds to a
3-chamber view where the LVOT is elongated. AA, ascending aorta; CAU, caudal; CRA, cranial; LAO, left anterior oblique; LA, left
atrium; LV, left ventricle; LVOT, left ventricular outflow tract; RAO, right anterior oblique.
The depth at which a THV is implanted in the left S-CURVES, DOUBLE S-CURVE, AND CHAMBER
ventricular outflow tract (LVOT) has been consistently ANATOMY
associated with the requirement for new PPI for both Device implantation is optimally performed in an
balloon- and self-expandable THVs.12,13 This is axiomatic angiographic projection that is perpendicular to the
because the deeper the THV is implanted the greater the virtual plane of the aortic annulus and has parallax
risk of contact between the THV and the conduction removed from the delivery system. In other words, it is
tissue. Such interaction induces either direct or indirect recommended to implant in an angiographic view where
(localized edema, hematoma, etc.) injury to the conduction there is no foreshortening of either the patient’s anatomy
tissue. A shallower THV implantation in the LVOT is or the delivery system. Traditionally, with self-expandable
therefore among the simplest ways to reduce new PPI devices that have a marker band at the distal tip of the
rates. It is, however, important to implant the THV at a delivery capsule, it was uncommon to have both the
sufficient depth to avoid pop-up of the valve into the aorta anatomy and device in plane together, and hence, there
and to ensure anchoring and sealing. Each valve has specific were two options: (1) implant in an angiographic view
guidance regarding implantation depth; for example, the where the anatomy was in the correct imaging plane
recommended Evolut TAV (Medtronic) implantation and accept that the device was not perfectly aligned, or
depth is 3 mm depth below the aortic annulus. Because (2) implant in an angiographic view where the device is
a few millimeters in implantation depth can make a big in-plane and accept that the anatomy was not perfectly
difference for PPI rates, accurate imaging of the aortic aligned. Neither of these implantation techniques is ideal
root during valve implantation is crucial to achieving the because the presence of parallax in either the anatomy
correct implantation depth and a low PPI rate. Recently, or the device renders the relationship between them
Jilaihawi et al described the Minimizing Depth According (implant depth) unclear.
to the membranous Septum (MIDAS) approach for To address this problem, Piazza and Thériault-Lauzier
the deployment of self-expandable THVs, aiming for an developed the double S-curve implantation technique
implantation depth of less than the MS length, which in using bespoke multislice CT (MSCT) software (FluoroCT).
their experience reduced the incidence of new conduction Using this software program, the operator could construct
disturbances and requirement of new PPI.14 the S-curve of the annular plane in the usual way and then
A B C D E F
G H I J K L
Figure 3. Cusp overlap technique. Traditional three-cusp view where all three cusps are visible, generally corresponding to an LAO-
CRA view (A-C). In this view, the NCC (yellow dot) is on the left side, the RCC (green dot) is in the middle, and the LCC (red dot) is on
the right side. THV positioning, deployment, and final assessment under a 3-cusp view (D-F). New 2-cusp view where the LCC and
the RCC are overlapped and the NCC is isolated, generally corresponding to an RAO-CAU view (G-I). In this view, the NCC (yellow
dot) remains on the left side while the RCC (green dot) and the LCC (red dot) are overlapped on the right side. THV positioning,
deployment, and final assessment under a 2-cusp view (J-L). CAU, caudal; CRA, cranial; LAO, left anterior oblique; LCC, left coronary
cusp; NCC, non-coronary cusp; RAO, right anterior oblique; RCC, right coronary cusp; THV, transcatheter heart valve.
construct the S-curve of the delivery catheter once it was projection. If the LAO-cranial view is used for THV
sitting in the aortic root of the patient during the TAVR implantation, the delivery catheter appears to be closer
procedure. The software generates the delivery catheter to the anatomy (represented by the NCC pigtail) than
S-curve once a left anterior oblique (LAO) and a right it actually is, and therefore, the depth of implantation
anterior oblique (RAO) projection (> 20º separation) with appears to be less than it actually is. This does not
the delivery catheter in-plane are input into the system occur if the RAO-caudal projection is used for THV
(Figure 2). The junction of the aortic annulus S-curve implantation.17
and the delivery catheter S-curve gives the angiographic Although the double-S technique was used successfully
projection where both the anatomy and delivery catheter in several centers, the requirement for dedicated MSCT
are simultaneously in-plane. Thus, in this angiographic view, software and the time required to generate the S-curve
the distance between the base of the NCC (pigtail catheter) of the delivery catheter in the LVOT during TAVR meant
and the bottom of the THV (the implant depth) can be that widespread adoption of the double-S implantation
trusted to be accurate because there is no foreshortening technique did not occur.
of either structure. Importantly, in 90% of cases, the
junction of these two S-curves (hence the title of double CUSP OVERLAP TECHNIQUE
S-curve technique) was in an RAO-caudal projection, with Because the double-S curve implantation view occurs
the remaining cases being either an anteroposterior-caudal in the RAO-caudal projection in 90% of cases, a “modified
or a shallow LAO-caudal projection.15,16 double-S” can be achieved by simply selecting an RAO-
Further understanding of the fluoroscopic anatomy of caudal implantation view from the annular S-curve and
the heart, described as chamber anatomy, reveals why then adjusting this angle slightly during the procedure (if
the RAO-caudal implantation view removes parallax required) to ensure that the marker band at the base of the
from both the anatomy and the delivery catheter: in delivery catheter is in-plane.
this projection, the heart is imaged in a three-chamber More recently, Gada and colleagues recognized that
view (echocardiographic nomenclature) with the LVOT in the RAO-caudal implant view, the RCC and the left
elongated (no foreshortening). On the other hand, in an coronary cusp (LCC) are superimposed on MSCT, leaving
LAO-cranial view, the heart is imaged in a four-chamber the NCC isolated. This two-cusp view, or cusp overlap
view with the LVOT foreshortened (Figure 2). Thus, if view, is distinct from the traditional three-cusp view that is
a catheter is placed in the LVOT, it is foreshortened in found usually in an LAO-cranial projection (Figure 3). This
the LAO-cranial projection but not in the RAO-caudal important observation affords TAVR operators the ability
9. Faroux L, Chen S, Muntané-Carol G, et al. Clinical impact of conduction disturbances in transcatheter aortic valve (Continued from page 7)
replacement recipients: a systematic review and meta-analysis. Eur Heart J. 2020;41:2771-2781. doi: 10.1093/
eurheartj/ehz924
overlap technique has been associated with reproducibly
10. Fadahunsi OO, Olowoyeye A, Ukaigwe A, et al. Incidence, predictors, and outcomes of permanent pacemaker low PPI rates and improved clinical outcomes through
implantation following transcatheter aortic valve replacement: analysis from the U.S. Society of Thoracic what we believe is a rational, evidence-based, and intuitive
Surgeons/American College of Cardiology TVT registry. JACC Cardiovasc Interv. 2016;9:2189-2199. doi: 10.1016/j.
jcin.2016.07.026 approach that is supported by sound principles in anatomy
11. Kawashima T, Sato F. Visualizing anatomical evidences on atrioventricular conduction system for TAVI. Int J Cardiol. and pathophysiology. n
2014;174:1-6. doi: 10.1016/j.ijcard.2014.04.003
12. De Torres-Alba F, Kaleschke G, Diller GP, et al. Changes in the pacemaker rate after transition from Edwards SAPIEN
XT to SAPIEN 3 transcatheter aortic valve implantation: the critical role of valve implantation height. JACC Cardiovasc 1. Reardon MJ, Van Mieghem NM, Popma JJ, et al. Surgical or transcatheter aortic-valve replacement in intermediate-risk
patients. N Engl J Med. 2017;376:1321-1331. doi: 10.1056/NEJMoa1700456
Interv. 2016;9:805-813. doi: 10.1016/j.jcin.2015.12.023
2. Popma JJ, Deeb GM, Yakubov SJ, et al. Transcatheter aortic-valve replacement with a self-expanding valve in low-risk
13. Gomes B, Geis NA, Chorianopoulos E, et al. Improvements of procedural results with a new-generation
patients. N Engl J Med. 2019;380:1706-1715. doi: 10.1056/NEJMoa1816885
self-expanding transfemoral aortic valve prosthesis in comparison to the old-generation device. J Interv Cardiol.
3. Adams DH, Popma JJ, Reardon MJ, et al. Transcatheter aortic-valve replacement with a self-expanding prosthesis. N Engl J
2017;30:72-78. doi: 10.1111/joic.12356
Med. 2014;370:1790-1798. doi: 10.1056/NEJMoa1400590
14. Jilaihawi H, Zhao Z, Du R, et al. Minimizing permanent pacemaker following repositionable self-expanding 4. Jilaihawi H, Zhao Z, Du R, et al. Minimizing permanent pacemaker following repositionable self-expanding transcatheter
transcatheter aortic valve replacement. J Am Coll Cardiovasc Interv. 2019;12:1796-1807. doi: 10.1016/j. aortic valve replacement. JACC Cardiovasc Interv. 2019;12:1796-1807. doi: 10.1016/j.jcin.2019.05.056
jcin.2019.05.056 5. Eitan A, Witt J, Stripling J, et al. Performance of the Evolut-R 34 mm versus Sapien-3 29 mm in transcatheter aortic valve
15. Thériault-Lauzier P, Andalib A, Martucci G, et al. Fluoroscopic anatomy of left-sided heart structures for replacement patients with larger annuli: early outcome results of Evolut-R 34 mm as compared with Sapien-3 29 mm in
transcatheter interventions: insight from multislice computed tomography. JACC Cardiovasc Interv. 2014;7:947-957. patients with annuli >/=26 mm. Catheter Cardiovasc Interv. 2018;92:1374-1379. doi: 10.1002/ccd.27588
doi: 10.1016/j.jcin.2014.06.002 6. Tang GHL, Reardon MJ, Kodali SK, et al. Comparison of clinical and echocardiographic outcomes after transcatheter aortic
16. Thériault-Lauzier P, Spaziano M, Vaquerizo B, et al. Computed tomography for structural heart disease and valve implantation with 31-mm CoreValve versus 34-mm Evolut R bioprostheses from the STS/ACC TVT Registry. Am J Cardiol.
interventions. Interv Cardiol. 2015;10:149-154. doi: 10.15420/ICR.2015.10.03.149 2019;124:1091-1098. doi: 10.1016/j.amjcard.2019.07.010
17. Piazza N, Mylotte D, Theriault Lauzier P. Fluoroscopic "heart chamber" anatomy - the case for imaging modality- 7. Dowling C, Firoozi S, Doyle N, et al. Initial experience of a large, self-expanding, and fully recapturable transcatheter aortic
independent terminology. EuroIntervention. 2016;18;12:Y9-Y15. doi: 10.4244/EIJV12SYA3 valve: the UK & Ireland Implanters’ registry. Catheter Cardiovasc Interv. 2019;93:751-757. doi: 10.1002/ccd.27934
18. Tang GHL, Zaid S, Michev I, et al. "Cusp-overlap" view simplifies fluoroscopy-guided implantation of self- 8. Kuhn C, Frerker C, Meyer AK, et al. Transcatheter aortic valve implantation with the 34 mm self-expanding CoreValve
expanding valve in transcatheter aortic valve replacement. JACC Cardiovasc Interv. 2018;11:1663-1665. doi: 10.1016/j. Evolut R: initial experience in 101 patients from a multicentre registry. EuroIntervention. 2018;14:e301-e305. doi: 10.4244/
jcin.2018.03.018 EIJ-D-17-01153
9. Harnath A, Gomes B, Herwig V, et al. First experience with the 34 mm self-expanding Evolut R in a multicentre registry.
EuroIntervention. 2018;14:e298-e300. doi: 10.4244/EIJ-D-18-00137
10. Piazza N, Mylotte D, Theriault Lauzier P. Fluoroscopic “heart chamber” anatomy - the case for imaging modality-
Xavier Armario, MD independent terminology. EuroIntervention. 2016;12:Y9-Y15. doi: 10.4244/EIJV12SYA3
11. Gada H, Vora A, Siddique S, et al. TCT CONNECT-473 Reduction of rates of permanent pacemaker implantation with 34-mm
Department of Cardiology Evolut R using cusp overlap technique. J Am Coll Cardiol. 2020;76 (17 suppl S):B196.
Galway University Hospital 12. Gada H, Vora A, Millan-Iturbe O, et al. TCT CONNECT-473 Reproducibility of cusp overlap technique to reduce permanent
pacemaker implantation with Evolut: the Latin American experience. J Am Coll Cardiol. 2020;76:B202.
Galway, Ireland
Disclosures: None.
UC202111562 EN
20 SUPPLEMENT TO CARDIAC INTERVENTIONS TODAY JANUARY/FEBRUARY 2021 VOL. 15, NO. 1