Technical Information

TI 244T/02/en
RTD Temperature Sensor
60018846
omnigrad M TR 44, TR 45

Hygienic RTD assembly


With or without thermowell and M.I. replaceable inset
PCP (4...20 mA), HART® or Profibus-PA® electronics

The temperature sensors Omnigrad M • Small probe diameter or reduced tip for
type TR 44 and TR 45 are resistance ther- fast response time (TR 44)
mometers especially designed for hygi- • Tapered or reduced tip and thermocon-
enic applications (food, pharmaceutical ductive paste for fast response time
and fine chemicals industry). (TR 45)
They consists of a measurement probe • Surface finishing down to Ra < 0.4 µm,
with (TR 45) or without (TR 44) a protec- with or without electropolishing
tion well, and a housing, which may con- • Stainless steel, aluminum or plastic
tain the transmitter for the conversion of housing, all of them easily cleanable
the measured variable. and IP65 at minimum (IP68 available)
Due to their modular design, TR 44 and • Replaceable mineral insulated insert
TR 45 fit to all the hygienic industrial proc- (TR 45); it, installed inside the thermo-
esses. well, avoids plant shut down during the
substitution or the verification of the
instrument.
• PCP (4...20 mA, with enhanced accu-
racy), HART® and Profibus-PA® 2-wire
transmitters
74-01 • Pt 100 sensing element with class A
Features and Benefits accuracy (DIN EN 60751)
• 3-A® and EHEDG certification • Double Pt 100, for redundancy or vali-
• SS 316L/1.4435 for "wetted" parts (BN 2 dation purposes
compliance on request) • ATEX EEx ia certification (TR 45)
• Most common hygienic process con- • Material certification (3.1.B, ...)
nections as standard; other on request • EA calibration certificate
• Customized immersion length • Ferrite content determination
• Sensor matching calibration
omnigrad M TR 44, TR 45

Areas of application
• Food industry: milk, beer, fruit juice, syrup, chocolate, oils/fats, powders, auxiliary services,
storage tanks/silos, CIP/SIP systems.
• Biotechnology industry: fermenters, auxiliary services, CIP/SIP systems.
• Pharmaceutical industry: fluids, acids, purified water, auxiliary services, CIP/SIP systems.
• Fine chemicals industry: cosmetics, auxiliary services, CIP/SIP systems.

Function and system design


Measuring principle In RTD (Resistance Temperature Detector) thermometers the sensing element consists of an elec-
trical resistance with value of 100 Ω at 0°C (called Pt 100, in compliance with standard DIN EN
60751), which increases at higher temperatures according to a coefficient characteristic of the
resistor material (Platinum).

Overall design The temperature sensors Omnigrad M TR 44 and TR 45 are constituted by a measurement probe,
with (TR 45) or without (TR 44) a protection well, and a housing (head), which may contain a trans-
mitter or the terminals for the electrical connection.
In TR 45 the probe is a sensing insert, placed inside the thermowell and spring loaded on the bot-
tom of it, so as to improve the heat transfer.
The sensing element (Pt 100), like in TR 44, is positioned close to the tip of the probe.
The thermowell (TR 45) is obtained from a 9 mm diameter pipe. The final (sensing) part can be
straight, tapered, that is with diameter smoothly decreased by means of a hammering procedure,
or reduced (stepped).
For TR 44, the sheath of the probe is obtained
from a 8 mm or from a 6 mm pipe. In the case of
8 mm, the final (sensing) part is always with
reduced (stepped) diameter. In the case of 6
mm, the probe pipe is always straight.
The probe (TR 44) or thermowell (TR 45) can be
installed on the plant (pipe or vessel) by means
of a hygienic process connection, which can be
chosen among the most common types (see
d
E

section "System components"). d


E

For both TR 44 and TR 45 the construction, exe-


cuted according to 3-A® and EHEDG design cri-
teria, allows the sensor to withstand the stresses R=3.2
R=3.2
caused by CIP (Cleaning In Place) and SIP (Ster-
L

ilization In Place) processes.


The electrical structure of the instruments com-
L

plies always with the rules of DIN EN 60751


standard.
The housing can be of different types and mate-
D
rials (plastic, painted aluminium, stainless steel). D
Its fitting to the rest of the probe and the gland for
the cable entry, assure an IP (Ingress Protection)
Figure 1. TR 44 (left) and TR 45 (right).
grade of 65 at minimum.

Materia Wetted parts in SS 316L/1.4435.

Weight From 0.5 to 2.5 kg for standard options.

2 Endress+Hauser
omnigrad M TR 44, TR 45

Electronics
The wanted type of output signal can be obtained by choosing the right head-mounted transmit-
ter.
Endress + Hauser provides state-of-the-art transmitters (iTemp® series) built in 2-wire technology
and with 4…20 mA, HART® or Profibus-PA® output signal. All the transmitters are easily program-
mable with a personal computer, by means of the public domain software ReadWin® 2000 (for
4…20 mA and HART® transmitters) or the software Commuwin II (Profibus-PA® transmitters).
The HART® transmitters can be programmed also by means of the handheld operating module
DXR 275 (Universal HART® Communicator).
A PCP (4…20 mA) model (TMT 180) is available with enhanced accuracy, particularly advisable
for hygienic applications.
In the case of Profibus-PA® transmitters, E+H suggests the use of Profibus dedicated connectors:
the Weidmüller type (Pg 13.5 M12) is provided as standard option.
For detailed information about transmitters, please refer to the specific concerned documentation
(see TI codes at the end of this document).
If a head-mounted transmitter is not employed, the sensor probe can be connected through the
terminal block to a remote converter (i.e. DIN rail transmitter).

Performances
Operating conditions Ambient temperature (housing without head-mounted transmitter)
• metal housings -40÷130°C
• plastic housings -40÷85°C

Ambient temperature (housing with head-mounted transmitter) -40÷85°C


Ambient temperature (housing with display) -20÷70°C

Process temperature
The measurement range (see below) can be limited by the gasket employed in the process con-
nection.

Max process pressure


• TR 44, TR 45 5 MPa (50 bar) at 20°C
• TR 44, TR 45 3.3 MPa (33 bar) at 250°C
• TR 45 2.4 MPa (24 bar) at 400°C
Lower maximum pressures can be due to the process connection (i.e. clamp, …).

Max flow velocity


The highest flow velocity tolerated by the sensor stem diminishes with increasing lengths of the
well/probe exposed to the fluid stream.

Shock and vibration resistance


According to DIN EN 60751 2.8 g peak / 10÷500 Hz

Accuracy Probe max error


• cl. A 3σ = 0.15+0.0020ItI 50…250°C
3σ = 0.30+0.0050ItI 250…400°C

• cl. 1/3 DIN B 3σ = 0.10+0.0017ItI 0…100°C


3σ = 0.15+0.0020ItI -50...0 / 100...250°C
(ItI=absolute value of temperature in °C) 3σ = 0.30+0.0050ItI 250…400°C

Endress+Hauser 3
omnigrad M TR 44, TR 45

Transmitter max error


See specific documentation (codes at the end of this
document).

Display max error 0.1% FSR + 1 digit

The "4 wires" configuration assures no additional


errors in case of long connecting cables (without
head-mounted transmitter); with the "4 wires" config-
1/3 DIN B (°C)
uration, generally speaking, there is a higher guar-
antee of accuracy.

Measurement range • TR 44 -50…250°C


• TR 45 -50…400°C

Response time Test in water at 0.4 m/s (according to DIN EN 60751; 23 to 33°C step change):

Sensor Type Response time Without thermo- With thermo-


conductive paste conductive paste
TR 44 t50 5s -
t90 13 s -
TR 45 straight tip t50 20 s 10 s
t90 60 s 30 s
TR 45 tapered tip t50 12 s 5s
t90 40 s 12 s
TR 45 reduced tip t50 8s 3.5 s
t90 23 s 8s

Insulation Insulation resistance between terminals and probe sheath more than 100 MΩ at 25°C
(according to DIN EN 60751, test voltage 250 V) more than 10 MΩ at 300°C

MTBF It is heavily influenced by the operating conditions (high pressure, vibrations, shocks, thermal
shocks, corrosion, …) of the sensors.
For a normal use, a calculation made according to MIL-HDBK-217F, gives as result a value higher
than 100 years in the case without built-in electronics, and more than 20 years with a head-
mounted transmitter (35 years with TMT 181).

Self heating Negligible when E+H transmitters are employed.

Installation
Omnigrad M TR 44 and TR 45 can be mounted on the wall of pipes or vessels.
The counterparts for the process connections and the respective gaskets or sealing rings, are
generally not supplied with the sensors, and are on the customer's responsibility (EHEDG and 3-
A® requirements must be fulfilled).
The only exception are the G1" and the Ingold connections, for which it is possible to provide the
sensor with the adapter to be welded on the plant.
Furthermore the Ingold connection and the G1" Liquiphant M type counterpart are supplied with
the necessary sealing o-rings.
As a general rule, the sensors should be installed in a manner which does not adversely affect
their cleanability.

4 Endress+Hauser
omnigrad M TR 44, TR 45

In case of ATEX certified components (transmitter, insert), please refer to the specific documen-
tation (see the code at the end of this document).
The immersion length may have an influence on the accuracy of the sensors: if it is too low, an
error might be generated in the measured temperature, due to the lower temperatures of the proc-
ess fluid near the walls, and to the heat transfer which occurs through the sensor stem.
The incidence of such error can be not negligible if the difference between the process and the
ambient temperature is big.
To avoid this source of inaccuracy, the immersion length (L) should be, if possible, at least 90÷120
mm. In pipes with little diameter, the axis line of the duct must be reached, and even slightly
exceeded, by the tip of the probe (see figure 3).
Insulation on the outer part of the sensor reduces the effect of the low immersion.
Another solution could be a tilted installation (see figure 2).
Attention should be paid in the choice of the measurement point in case of two-phase flows, which
may cause fluctuations in the detected temperature value.
For installation of the sensors in small pipes, suitable solutions are shown in figure 3.

Insulation

d
h<d/2

Figure 2. General installation solutions.

In hygienic applications, a good, strict, installation rule consists in not to leave any dead spaces
along the run of the process fluid. For this aim, the flush-mounting is allowed by the connections
Varivent, G1" Liquiphant M type (+ purpose built adapter) and Ingold (+ purpose built adapter).
Also the clamp flanges can partly fulfil this requirement, if the Tri-Clamp® components 7IMPS or
TL7IWWMS (Instrument Tees) are used (see the related figure 4).

Varivent D=50 mm Varivent D=31 mm Microclamp


for pipes DN25 for pipes DN10/15 or miniclamp

Figure 3. Installation solutions in small pipes.

Endress+Hauser 5
omnigrad M TR 44, TR 45

For other connection arrangements, the scheme illustrated in fig. 2 should be observed [h <=
d/2].
For weld-in connections, care should be taken by the user in the execution of the welding on the
process side (suitable weld material, welding radius > 3 mm, absence of pits, folds, crevices, …).
Purely threaded and metal-to-metal joints are dis-
couraged by some hygienic design standards (i.e.
3-A® Standard 74-01 and document 8 from EHEDG
respectively). This is the reason why E+H Tempera-
ture division doesn’t suggest those solutions in "san-
itary" applications. Moreover often metal-to-metal
couplings can be used properly only one time.
Concerning corrosion, the basic material of the wet-
ted parts (SS 316L/1.4435) is capable to tolerate
common corrosive media up to high temperatures.
The gaskets supplied with the Ingold process con-
nection and with the G1" Liquiphant M type weld-in
adapter, besides CIP and SIP-proof, resist to a wide
variety of aggressive substances. For particular
applications, please ask to the E+H service.
In case of disassembling of the sensors, new gas-
kets equal to the originals and specific torques must
be employed for the re-assembling procedure, to
assure the stated IP (Ingress Protection) grade of the
enclosures.
When the ambient has a high humidity rate and the
process is at low temperature, a plastic housing can
be an effective choice (i.e. model TA20B) to avoid Figure 4. Installation with the Tri-Clamp®
problems due to condensation. component 7IMPS.

System components
Housing The housing which contains the electric terminals or the transmitter, can be of different types and
materials (plastic, painted aluminium, stainless steel). Its fitting to the rest of the probe and the
gland for the cable entry, assure an IP grade of 65 at minimum (see also the figure below).
All the supplied heads have internal geometry according to DIN 43729 standard (form B), and
thermometer connection M24x1.5.
Head type TA20A is the basic E+H aluminium housing for temperature sensors, provided in the
E+H corporate colours, without supplementary charges in the version IP66/IP67.
Head TA20B is a polyamide housing, black or white, sometimes defined with the code BBK in the
Temperature market.
In TA21E a screw cap is employed, joined to the head body by a chain.
Head type TA20D (aluminium), called also BUZH, is able to contain a terminal block and a trans-
mitter at the same time, or two transmitters.
Head TA20J is the stainless steel housing with E+H corporate design, and can be provided also
with a LCD display (4 digits), operating with 4…20 mA transmitters.
TA20R is the head suggested by E+H Temperature division for hygienic applications, due to its
material (stainless steel) and its "clean" design.
TA20W (BUS type) is an aluminium round-shaped grey head, with a clip for the cap closure.

6 Endress+Hauser
omnigrad M TR 44, TR 45

Husing Type IP66 IP67 Husing Type IP66 IP67


TA20A TA20J
110 108

YES YES YES YES

98
80
TA21E TA20R
Ø55
E8
0

// YES YES YES


95

79
TA20D TA20W
140 140
113

YES // YES //

85
Figure 5. Housings.

Head transmitter The available head-mounted transmitters are (see also the section "Electronics"):
• TMT 180 PCP 4…20 mA
• TMT 181 PCP 4…20 mA
• TMT 182 Smart HART®
• TMT 184 Profibus-PA®

TMT 180 and TMT 181 are PC Programmable transmitters. TMT 180 can be supplied in a version
with a better accuracy (0.1°C vs. 0.2°C) in the bandwidth -50…250°C; a model with a fixed meas-
urement range (defined by the user in the order phase) is available too.
TMT 182 output consists of 4…20 mA and HART® superposed signals.
For TMT 184, with Profibus-PA® output signal, the communication address can be set via software
or by means of a mechanical dip-switch: the user may ask for the desired configuration in the
order phase
.

Figure 6. TMT 181.

Extension neck The extension neck (part between the process connection and the housing), built from stainless
steel, is constituted by a tube with standard diameter of 15 mm and length 82 mm.
For the Ingold process connection, this dimension of 82 mm is not sufficient, and a greater length
must be specified by the user through the "8" digit in the product order code.

Endress+Hauser 7
omnigrad M TR 44, TR 45

Process connection It can be chosen among the types:

Microclamp (TR 44) SMS 1147/8

ISO 2852 clamp DIN 11864-1-A

Tri-Clamp® DIN 11851

Varivent® Weld-in

G1" Liquiphant M type Ingold

The connections are supplied in several sizes. Others (i.e. Neumo, APV) are available on request.
The process connection is continuously welded on the probe or on the well in such a way to obtain
a welding radius of 3.2 mm at minimum, between the lower surface of the connection and the sen-
sor stem (requirement of EHEDG and 3-A® standards).
The Ingold connection is supplied with the sealing o-ring mounted on, whose material (silicone)
is in compliance with FDA CFR Title 21, § 177.2600 (max temperature 230°C).
Concerning the available welding adapters, please see the paragraph "Accessories".
The Varivent® connections must be used with dedicated Tuchenhagen® in-line components or
tank adapters; for Varivent® flanges with small nominal diameter, the maximum immersion lengths
for usual applications are listed in the following table (see also figure 3)

TR 44 (6 mm probe, straight tip) TR 45 (reduced tip)


Varivent® Nominal Diameter DN 10/15 DN 25 DN 10/15 DN 25

Suggested Immersion Length (L) 30÷50 mm 15 mm 15 mm 15 mm


(low viscosity fluids) (special option)
15 mm
(high viscosity fluids)

Please be aware of the fact that in the past the Varivent® connection DN25 has been used also for
pipes with nominal diameter DN32.

Probe In TR 45 the measuring probe is constituted by a mineral insulated insert positioned inside the
thermowell.
For its replacement, the insert length (IL) must be chosen depending on the immersion length (L)
of the thermowell; in case of need for a spare part, please refer to the following table:

8 Endress+Hauser
omnigrad M TR 44, TR 45

Sensor Probe Tip Insert Insert Diam. Ext. Neck Insert Length
TR 45 straight TET 100 6 mm 82 IL = L+92
TR 45 red./tapered TET 105 3 mm 82 IL = L+92
TR 45 straight TET 100 6 mm E IL = L+E+10
TR 45 red./tapered TET 105 3 mm E IL = L+E+10

In TR 44 the protection well and the sensing part cannot be separated.


The immersion length is available in some standard values, or it can be choosen "customized"
within a range (please see the product structures in the last pages of this document).
For special versions of TR 45, with immersion shorter than 30 mm, the diameter of the thermowell
should be 5.3 mm, as in the reduced tip.
On request the starting material for the wetted parts can be supplied in compliance with Basler
Norm 2 (BN2), which imposes a limited ferrite content, in order to obtain an improved corrosion
resistance. For some sensor’s configurations the compliance with the requirements of BN2 can
be assured also after the welding and machining operations, that is on the finished product.
The surface roughness (Ra) of the wetted parts is supplied down to 0.4 µm. A roughness below
0.4÷0.5 µm is not demonstrated to have advantages for hygienic purposes.
Electropolishing consists in an electrolytic treatment of the metal surface, which leads to a clean-
ing, levelling and passivation of it.
In TR 45 a thermically conductive compound can be requested inside the thermowell, to improve
the heat transfer between the well and the internal sensing insert. The compound shouldn’t be
requested for immersion lengths longer than 300 mm, and used over 200°C.
26~
~

M24x1.5

d
d
E

d
E

121.0
L

D
Ø D

Figure 7. Functional components in TR 45 (straight tip).

1xPt100

Red
White
Red
65

Red
3 wires
White
20

Red
Ø6.6 Ø5.3

Figure 8. Standard wiring (3-wires) and tip designs ("tapered" on the left, "reduced" on the right).

Endress+Hauser 9
omnigrad M TR 44, TR 45

Certifications
Certifications from exter- • EHEDG Type 1 certification (TNO report n. V3912). EHEDG accepted process connections are:
nal bodies Varivent, IDF type ISO 2853, APV (acc. DIN 11864), APV Inline, DIN 11851 (only in combination
with EHEDG certified gasket from Siersema), Suedmo, Naue, Neumo.
• 3-A® Authorization no. 1144 for the declaration of compliance with standard 74-01.
• ATEX certificate KEMA 01 ATEX1169 X (1 GD IIC EEx ia T1÷T6) for TR 45.

Manufacturer’s certifica- The 3.1.B material certificate, according to standard EN 10204, is selectable directly from the sale
tions structure of the product. Other kinds of material certifications can be requested separately.
The "standard" one is a simplified and cost effective version of the certificate, in which the docu-
mentation about the origin of the employed materials refers to each sensor. The "labelled" version
has in addition a specific marking relating to the wetted parts of the sensor, and assures the
placement in the archives of the relevant data by means of the serial number of the thermometer.
Concerning tests and calibration, the "inspection test report" consists in a declaration of conform-
ity to the essential points of standard DIN EN 60751.
The "Factory calibration" is carried out in the E+H EA (European Accreditation) accredited labo-
ratory for temperature calibrations, following an internal procedure. A calibration performed
according to an EA accredited procedure (SIT calibration) can be requested separately.
The calibration is executed on the thermometer insert, when that is present in the sensor (TR 45).
For TR 44 a minimum immersion length (90 mm) is needed to assure a correct calibration.
With the Sensor Matching Calibration, a preliminary Factory calibration of the sensor is executed,
and subsequently the output curve of the transmitter is adjusted in order to minimize the meas-
urement error. In this way some of the errors of the sensor (probe or "loop" probe+transmitter) are
compensated by means of the transmitter.

Maintenance Omnigrad M thermometers don't need a specific maintenance. For the models supplied with
process connections including seals, the integrity of the sealing ring should be checked regularly
and it must be substituted when necessary.
In case of ATEX certified components (transmitter, insert), please refer to the specific documen-
tation (see the code at the end of this document).

Delivery time For small quantities (about 10 units) and standard options, from 5 to 20 days depending on the
requested configuration.

Accessories
G1" Liquiphant M type weld-in adapter
for flush-mounting of "EA" process connection 41
seal: silicone o-ring complying with FDA CFR Title 21, § 177.2600
max temperature: 230°C
29.6

material: SS 316L/1.4435; weight: 0.13 kg


provided with leakage detection port
mat. nr.: 60017886 60

"Ingold" D=25 mm, L=50 mm weld-in adapter


for flush-mounting of "NB" process connection G 1 1/4”

material: SS 316L/1.4435; weight: 0.32 kg.


provided with leakage detection port
mat. nr.: 60017887
50

Set of gaskets (no. 5 o-rings) for G1" Liquiphant M type weld-in adapter
seal: silicone o-ring complying with FDA CFR Title 21, § 177.2600
max temperature: 230°C
mat. nr.: 60018912
Set of gaskets (no. 5 o-rings) for "Ingold" process connection
seal: silicone o-ring complying with FDA CFR Title 21, § 177.2600
max temperature: 230°C
mat. nr.: 60018911
TW 45: thermowell for TR 45 See TI 252T/02/en

10 Endress+Hauser
omnigrad M TR 44, TR 45

Ordering information
TR44 - RTD temperature
sensor Process connection
BM Microclamp flange DN25 mm for pipe DN8/18
BA ISO2852 clamp flange DN12/21.3 (miniclamp)
BB ISO2852 clamp flange DN25/38
BC ISO2852 clamp flange DN40/51
BF Tri-Clamp® flange 1" and 1 1/2"
BH Tri-Clamp® flange 2"
CD Dairy connection DIN11851 DN25
CF Dairy connection DIN11851 DN40
CG Dairy connection DIN11851 DN50
CH Aseptic connection DIN11864-1-A DN25
CJ Aseptic connection DIN11864-1-A DN40
DA Welding conn. cylindrical D30xL40 mm
DB Welding conn. cyl./sph., D30xL40 mm
EA G1" for Liquiphant M type weld-in adapter (weld-in adapter 60017886; not included)
FA Varivent® D=68 mm for pipes DN32/125
FB Varivent® D=50 mm for pipes DN25
FC Varivent® D=31 mm for pipes DN10/15
JD SMS DN25 (1")
NB "Ingold" connection D=25 mm, L=50mm (weld-in adapter 60017887; not included)
YY Special version

Neck dimensions: E (75-250 mm) and d (material stainless steel)


5 82 mm neck length E 15mm diameter d
6 .... mm neck length E to specify, 15mm diam d
9 .... mm special neck length E and diameter d

Immersion length L (15-250 mm) (Up to 50mm the pipe diameter must be 6mm)
G 30 mm immersion length L, only reduced tip
A 50 mm immersion length L
B 90 mm immersion length L
F 120 mm immersion length L,
D 160 mm immersion length L
E 220 mm immersion length L
X ..... mm immersion length L, to specify
Y ..... mm special immersion, length L

Pipe diameter D, material and finishing of wetted parts, price per 100 mm of L
(Pipe D=8mm is always with reduced tip)
A 6 mm = D SS 316L/1.4435, Ra<=0.8 µm
B 6 mm = D SS 316L/1.4435, Ra<=0.4 µm
C 6 mm = D SS 316L/1.4435, Ra<=0.4 µm, electropolishing
E 8 mm = D SS 316L/1.4435, Ra<=0.8 µm
F 8 mm = D SS 316L/1.4435, Ra<=0.4 µm
G 8 mm = D SS 316L/1.4435, Ra<=0.4 µm, electropolishing
Y ...mm Special version

Tip design
S Straight tip, for pipe D=6mm
R Reduced tip (stepped), for pipe D=8mm
Y Special version

Terminal type
2 Flying leads
3 Ceramic terminal block
4 HR fiber terminal block

RTD type and wiring diagram


H 1 Pt100 class A, 3-wire
L 2 Pt100 class A, 3-wire
M 1 Pt100 class A, 4-wire
P 1 Pt100 1/3 DIN B, 3-wire
Q 2 Pt100 1/3 DIN B, 3-wire
R 1 Pt100 1/3 DIN B, 4-wire
Y Special version

Endress+Hauser 11
omnigrad M TR 44, TR 45

Housing material, cable entry, IP grade


A TA20A Alu., M20x1.5 conduit IP66/IP67
3 TA20A Alu., Pg16, IP66/IP68
4 TA20A Aluminium, Profibus® connector, IP66
B TA20B Polyamide white, Pg16, IP65
C TA20B Polyamide black, Pg16, IP65
E TA20E Aluminium, screw cap., M20x1.5 IP67
D TA20D Aluminium, high lid, Pg16, IP66
5 TA20D Aluminium, high lid, Profibus® connector IP66
J TA20J SS316L, M20x1.5, IP66/IP67
K TA20J SS316L, M20x1.5, + display IP66/IP67
M TA20J SS316L, Profibus® connector IP66/IP67
R TA20R SS316L, screw cap , M20x1.5, IP66/IP67
S TA20R SS316L, screw cap , Profibus® connector IP66/IP67
W TA20W Aluminium, round lid, clip, Pg16 IP66
Y Special version

Built-in head transmitter


See the subsequent table

Certificates
0 Certificates not required
B 3.1.B EN10204, standard for wetted parts
G 3.1.B EN10204, labelled for wetted parts
D 3.1.B EN10204, standard + roughness
H 3.1.B EN10204, labelled + roughness
L 3.1.B EN10204, labelled + ferrite content
J 3.1.B, labelled+roughness+ferrite content

Tests and calibration (Minim. immersion


length for calibrations is L=90mm; the loop cali-
bration is meant on the display when this is
present)
0 Tests and calibration not required
1 Inspection test report (TZC135-A), sensor
2 Inspection test report (TZC135-D), loop
A Factory calibration, single RTD, 0-100°C
B Factory calibr., loop single RTD, 0-100°C
C Factory calibration, double RTD, 0-100°C
E Factory calibr., single RTD, 0-100-150°C
F Factory. calibr., loop single RTD, 0-100-150°C
G Factory calibr., double RTD, 0-100-150°C
S Sensor matching calibration, single RTD,
0-100-150°C
(sensor-to-transmitter matching calibration)
T Sensor matching calibration, loop single RTD,
0-100-150°C
(sensor-to-transmitter matching calibration)

Additional options
0 Additional options not required

TR 44- Complete order code

12 Endress+Hauser
omnigrad M TR 44, TR 45

Ordering information
TR45 - RTD temperature
sensor Process connection
BA ISO2852 clamp flange DN12/21.3 (miniclamp)
BB ISO2852 clamp flange DN25/38
BC ISO2852 clamp flange DN40/51
BF Tri-Clamp® flange 1" and 1 1/2"
BH Tri-Clamp® flange 2"
CD Dairy connection DIN11851 DN25
CF Dairy connection DIN11851 DN40
CG Dairy connection DIN11851 DN50
CH Aseptic connection DIN11864-1-A DN25
CJ Aseptic connection DIN11864-1-A DN40
DA Welding conn. cylindrical D30xL40 mm
DB Welding conn. cyl./sph., D30xL40 mm
EA G1" for Liquiphant M type weld-in adapter (weld-in adapter 60017886; not included)
FA Varivent® D=68 mm for pipes DN32/125
FB Varivent® D=50 mm for pipes DN25
FC Varivent® D=31 mm for pipes DN10/15
JD SMS DN25 (1")
NB "Ingold" connection D=25 mm, L=50mm (weld-in adapter 60017887; not included)
YY Special version

Neck dimensions: E (75-250 mm) and d (material stainless steel)


5 82 mm neck length E 15mm diameter d
6 .... mm neck length E to specify, 15mm diam d
9 .... mm special neck length E and diameter d

Immersion length L (15-250 mm) (Up to 50 mm the pipe diameter must be 6 mm)
G 30 mm immersion length L, only reduced tip
A 50 mm immersion length L
B 90 mm immersion length L
F 120 mm immersion length L,
D 160 mm immersion length L
E 220 mm immersion length L
X ..... mm immersion length L, to specify
Y ..... mm special immersion, length L

Pipe diameter D, material and finishing of wetted parts, price per 100 mm of L
1 9 mm = D SS 316L/1.4435, Ra<=0.8 µm
3 9 mm = D SS 316L/1.4435, Ra<=0.4 µm
4 9 mm = D SS 316L/1.4435, Ra<=0.4 µm, electropolishing
9 ...mm Special version

Tip design
S Straight tip
p Straight tip + thermoconductive paste
R Reduced tip (stepped), only for L>=30mm
Q Reduced, for L>=30mm + thermocond. paste
T Tapered tip, only for L>=65 mm
U Tapered, for L>=65 + thermocond. paste
Y Special version

Terminal type
2 Flying leads
3 Ceramic terminal block
4 HR fiber terminal block

RTD type and wiring diagram


H 1 Pt100 class A, 3-wire
L 2 Pt100 class A, 3-wire
M 1 Pt100 class A, 4-wire
P 1 Pt100 1/3 DIN B, 3-wire
Q 2 Pt100 1/3 DIN B, 3-wire
R 1 Pt100 1/3 DIN B, 4-wire
Y Special version

Endress+Hauser 13
omnigrad M TR 44, TR 45

Housing material, cable entry, IP grade


A TA20A Alu., M20x1.5 conduit IP66/IP67
3 TA20A Alu., Pg16, IP66/IP68
4 TA20A Aluminium, Profibus® connector, IP66
B TA20B Polyamide white, Pg16, IP65
C TA20B Polyamide black, Pg16, IP65
E TA20E Aluminium, screw cap., M20x1.5 IP67
D TA20D Aluminium, high lid, Pg16, IP66
5 TA20D Aluminium, high lid, Profibus® connector IP66
J TA20J SS316L, M20x1.5, IP66/IP67
K TA20J SS316L, M20x1.5, display , IP66/IP67
M TA20J SS316L, Profibus® connector IP66/IP67
R TA20R SS316L, screw cap , M20x1.5, IP66/IP67
S TA20R SS316L, screw cap , Profibus® connector IP66/IP67
W TA20W Aluminium, round lid, clip, Pg16 IP66
Y Special version

Built-in head transmitter


See the subsequent table

Certificates
0 Certificates not required
B 3.1.B EN10204, standard for wetted parts
G 3.1.B EN10204, labelled for wetted parts
D 3.1.B EN10204, standard + roughness
H 3.1.B EN10204, labelled + roughness
L 3.1.B EN10204, labelled + ferrite content
J 3.1.B, labelled+roughness+ferrite content

Tests and calibration (Minim. immersion


length for calibrations is L=90mm; the loop cali-
bration is meant on the display when this is
present)\
0 Tests and calibration not required
1 Inspection test report (TZC135-A), sensor
2 Inspection test report (TZC135-D), loop
A Factory calibration, single RTD, 0-100°C
B Factory calibr., loop single RTD,0-100°C
C Factory calibration, double RTD, 0-100°C
E Factory calibr., single RTD, 0-100-150°C
F Factory. calibr., loop single RTD, 0-100-150°C
G Factory calibr., double RTD, 0-100-150°C
S Sensor matching calibration, single RTD,
0-100-150°C
(sensor-to-transmitter matching calibration)
T Sensor matching calibration, loop single RTD,
0-100-150°C
(sensor-to-transmitter matching calibration)

Additional options
A Atex EEx-i certified inset
0 Additional options not required

TR 45- Complete order code

14 Endress+Hauser
omnigrad M TR 44, TR 45

Built-in head transmitter


0 Without built-in transmitter
1 Built-in trasm.THT1, separate position
2 Fixed range TMT180-A21, from...to...°C - accuracy 0.2K, span limit: -200...650°C
3 Fixed range TMT180-A22, from...to...°C - accuracy 0.1K, span limit: -50...250°C
4 Programmable TMT180-A11, from...to...°C - accuracy 0.2K, span limit: -200...650°C
5 Programmable TMT180-A12, from...to...°C - accuracy 0.1K, span limit: -50...250°C
P PCP transmitter 2wire TMT181-A - isolated, adjusted from ... to ... °C
Q PCP ATEX transmitter 2wire TMT181-B - isolated, adjusted from ... to ... °C
R HART® transmitter 2wire TMT182-A - isolated, adjusted from ... to ... °C
S Profibus-PA® transmitter, 2wire, TMT184-A
T HART® ATEX transm. 2wire TMT182-B isolated, adjusted ...to ... °C
V Profibus-PA® ATEX transm. 2wire TMT184-B
9 Special version

Supplementary documentation
❑ RTD thermometers Omnigrad TST - General information TI 088T/02/en
❑ Terminal housings - Omnigrad TA 20 TI 072T/02/en
❑ Temperature head transmitter iTEMP® Pt TMT 180 TI 088R/09/en
❑ Temperature head transmitter iTEMP® PCP TMT 181 TI 070R/09/en
❑ Temperature head transmitter iTEMP® HART® TMT 182 TI 078R/09/en
❑ Temperature head transmitter iTEMP® PA TMT 184 TI 079R/09/en
❑ Pt 100 inset - Omnigrad TET 100 TI 071T/02/en
❑ Pt 100 inset - Omnigrad TET 105 TI 103T/02/en
❑ Thermowell for temperature sensor - Omnigrad M TW 45 TI 252T/02/en
❑ Safety instructions for use in hazardous areas XA 003T/02/z1
❑ E+H Thermolab - Calibration certificates for industrial thermometers and working standards.
RTD’s and thermocouples TI 236T/02/en
❑ Application Note on Fermentation

Endress+Hauser 15
Subject to modification

Endress+Hauser
Gmbh+Co.
Instruments International
P.O. Box 2222
D-79574 Weil am Rhein
Germany

Tel. (07621) 975-02


Tx 773926
Fax (07621) 975 345
http://www.endress.com
info@ii.endress.com

TI 244T/02/en/12.01