Beijing Aeonmed Co., Ltd. (Aeonmed for short) holds the copyrights to this manual, which is
non-public, published, and reserves the rights to keep it as a secure document. Refer to this
manual when operating, maintaining and repairing products only. Anyone other than Aeonmed
may not make it known to others.
Proprietary materials protected by the copyright law are included in this manual. Any section of it
cannot be reproduced, copied, or translated into other languages without any prior written
approval from Aeonmed who reserves the copyright.
Everything written in this manual is considered to be correct. Aeonmed is not legally responsible
for any mistakes printed within and any damages caused by incorrect installation and operation.
Aeonmed does not supply privileges endowed by the patent law to any other parties. Aeonmed is
not legally responsible for the results caused by patent law breaking and any rights of the third
party violating.
Refer to this manual before the product is used. The manual includes operating procedures which
must be performed with cautiously, operations that may result in non-normal working conditions
and the dangers which may damage equipment or cause bodily harm. Aeonmed is not
responsible for the security, reliability and function of the equipments in case that the dangers,
damages and non-normal phenomenon mentioned in this manual happen. Free repairs for these
malfunctions will not be provided by Aeonmed.
Aeonmed has the rights to replace any content in this manual without notice.
Manufacturer Responsibility:
Aeonmed is responsible for the security; reliability and function of the equipments when to
following conditions are adhered to:
I
Aeon8800A Service Manual Statement
CAUTION: Malfunctioning equipment may become invalid and cause bodily injury if a
The paid theoretical framework diagram will be supplied according to customer requirements by
Aeonmed, plus calibrating method and other information to help the customer, under the
assistance of qualified technicians, repair the equipment parts where can be done by customer
himself based on the stipulation by Aeonmed.
Warranty:
Manufacturing techniques and materials:
For a period of one year from the date of original delivery, the components and assemblies of this
product is warranted to be free from defects manufacturing techniques and materials, provided
that the same is properly operated under the conditions of normal use and regular maintenance.
The warranty period for other parts is three months. Expendable parts are not included. Our
obligation under the above warranties is limited to repairing free of charge.
Free Obligations:
Our obligation under the above warranties does not include the freight and other fees;
Aeonmed is not responsible for any direct, indirect or final product broken and delay
which result from improper use, alteration by using the assemblies unratified and
maintenance by anyone other than Aeonmed;
Improper use;
II
Aeon8800A Service Manual Statement
This product is not operated correctly in accordance with the manual instruction. The
power supply used or operating environment does not follow the requirements in this
manual.
Return
Follow the steps below in case that the product needs to be returned to Aeonmed:
1. Obtain the rights of return
Contact our customer service by informing them the number and type of the product. The number
is marked on the surface of the product. Return is unacceptable if the number cannot be identified.
Enclose a statement of the number, type and the reason of return as well.
2. Transportation charges
Transportation and insurance charges must be prepaid by the user for transporting the product to
Aeonmed for repairing. (Customers charges are added with regard to the products sold to
non-Chinese mainland users).
Use Notice
Welcome to use our products.
To use this product correctly and effectively, the user must read this user manual carefully before
using this product.
When the user is using this product, this user manual must be understood totally and observe
strictly.
This product is only applied to the use this user manual described.
Only trained and professional service persons are allowed to repair and examine this product.
If there is any case happened in the process of use, please call us and intense service will be
provided.
If there is any change in product specification, another informs will not be given.
III
Aeon8800A Service Manual Statement
P.C.: 100070
Tel: +86-10-83681616
Fax: +86-10-63718989
E-mail: service@aeonmed.com
IV
Aeon8800A Service Manual Table of Contents
Table of Contents
1. Product Description.................................................................................................................................. 1
V
Aeon8800A Service Manual Table of Contents
3. Checkout ................................................................................................................................................ 77
3.3.2. Check...................................................................................................................................... 88
3.3.2.1. Leak Test ........................................................................................................................................... 88
VI
Aeon8800A Service Manual Table of Contents
VII
Aeon8800A Service Manual Table of Contents
6.5.5. Disassembling the Display from Main Control Unit ................................................................ 212
VIII
Aeon8800A Service Manual Table of Contents
IX
Aeon8800A Service Manual Product Description
1. Product Description
1
Aeon8800A Service Manual Product Description
1
16
15
2 14
13
3
12
4 11
10
5
9
6 8
2
Aeon8800A Service Manual Product Description
3
Aeon8800A Service Manual Product Description
4
Aeon8800A Service Manual Product Description
2 AC Power Assy
5
Aeon8800A Service Manual Product Description
6
Aeon8800A Service Manual Product Description
1.5. Symbols
The following table provides descriptions of symbols that are used on the device and/or within this
manual.
Dangerous voltage
Earth (ground)
7
Aeon8800A Service Manual Product Description
Equipotentiality
Date of manufacture
Manufacturer
Catalogue number
Serial number
Lock
Unlock
Insp. Flow
Exp. Flow
8
Aeon8800A Service Manual Product Description
O2 flush O2+
Fuse
Recycle
9
Aeon8800A Service Manual Product Description
Alarm icon
10
Aeon8800A Service Manual Product Description
1.6. Abbreviations
Abbreviation Definition
A
AGSS Anesthetic gas scavenging system
APL Adjustable pressure limiting
B
BDU Basic data unit
C
C Compliance
C·G·O Common gas outlet
cmH2O Centimeters of Water
CO2 Carbon oxide
D
Des. Desflurane
E
EEPROM Electrically erasable programmable read only memory
Enf. Enflurane
ETCO2 End-expiratory CO2 concentration
F
FiO2 Fraction of Inspired Oxygen
Flow-t Flow-time wave
Flow Flow
Freq Frequency
FreqMIN Minimum Frequency in PS mode
G
GUI Graphic user interface
H
Hal. Halothane
I
I:E Inspiratory to Expiratory Ratio
INSCO2 Inspiratory CO2 concentration
Iso. Isoflurane
IPPV Intermittent positive pressure ventilation
L
11
Aeon8800A Service Manual Product Description
M
MAC Minimum Alveolar Concentration
Manual Manual ventilation
PMEAN Mean pressure
mL Mill Liter
MRI Magnetic resonance imaging
MV Minute volume
N
N2O Nitro oxide
O
O2 Oxygen
P
Paw Airway pressure
Paw-t Pressure-time wave
Pb Plumbum
PLAT Plat pressure
PCV Pressure Control Ventilation
PEEP Positive End Expiratory Pressure
Pmax Maximum pressure
PMEAN mean pressure
PPEAK Peak pressure
PS Pressure Support Ventilation
PTARGET Target Pressure
S
SIMV Synchronized Intermittent Mandatory Ventilation
Sev. Sevoflurane
T
TINSP Inspiratory Time
TP Inspiratory Pause Time
Trigger Flow trigger
TSLOPE Inspiratory Slope Time
12
Aeon8800A Service Manual Product Description
V
VT Yidal volume
W
WDT Watch dog timer
ΔP Differential Pressure
13
Aeon8800A Service Manual Operation Theory
2. Operation Theory
Pneumatic system plays the role of supplying O2, N2O, AIR and anesthetic agent to breathing
system, ensuring their pressures are within normal range respectively, so that normal operation of
anesthesia machine and patient’s safety can be guaranteed.
14
Aeon8800A Service Manual Operation Theory
15
Aeon8800A Service Manual Operation Theory
16
Aeon8800A Service Manual Operation Theory
17
Aeon8800A Service Manual Operation Theory
18
Aeon8800A Service Manual Operation Theory
NOTE:Between tube 38 and tube 39, there is a gas resistance which is shaded by arrow.
19
Aeon8800A Service Manual Operation Theory
Pipeline gas supply system consists of gas inlet module and pressure relief valve. Gas inlet
module has three inlet ports for O2, Air and N2O. The pressure regulator can decrease the inlet
gas pressure to meet the need of patients.
20
Aeon8800A Service Manual Operation Theory
21
Aeon8800A Service Manual Operation Theory
Cylinder gas inlet assembly consists of YOKE assembly and pressure regulators. The pressure
regulator can decrease the high pressure gas from cylinder, and then the release gas goes into
pressure relief valve.
NOTE: N2O pressure regulator Assembly includes the two nuts and the subpart marked with (1).
Air and O2 pressure regulator Assembly are the same as N2O pressure regulator Assembly.
22
Aeon8800A Service Manual Operation Theory
Electronic flowmeter, the mechanical device used to adjust flow control. The flow setting is
displayed to the left of the adjustment knob and also below the corresponding flow tube on the
display. The Total Flow Tube will display the sum of O2, N2O, and air flow. The flow controller is
equipped with a control valve for the ratio of O2 and N2O, so as to ensure that theO2 concentration
is no less than 21% when the user is using O2 and N2O at the same time. With no O2 supply
pressure, N2O is cut off. The effective range of the flow controller is 0~15L/min and the accuracy
is ±10%. When the value is out of the range, the flowmeter readings are for reference only.
23
Aeon8800A Service Manual Operation Theory
24
Aeon8800A Service Manual Operation Theory
NOTE: When replacing Part1 and Part 2, please refer to Diagram1; when replacing Part4, Part6
and Part7, please refer to Diagram 2.
NOTE: Don’t leave filtration and sealing gasket behind when replacing flow sensor.
25
Aeon8800A Service Manual Operation Theory
As shown in figure, vaporizer manifold can be used to support anesthetic vaporizer. When
vaporizer manifold is installed with anesthetic vaporizer, the anesthetic gas in it can be input to
anesthesia system through vaporizer manifold. When anesthetic vaporizer is unloaded, vaporizer
manifold can be used as gas flow channel.
Part Number: 122005534
1. Gas inlet port, through which gas goes into the vaporizer.
2. Gas outlet port, through which the fresh gas flow to the breathing system.
26
Aeon8800A Service Manual Operation Theory
As shown follow figure, by pressing the button, high flow of 35L/min~50L/min can be provided to
fresh gas outlet. When connected to Central Gas Supply System or Back-up Cylinder, even if the
machine is power off, press the “O2+” button, the high flow O2 could be sent to breathing system
or breathing bag.
O2 flush valve
1 Tube connector connected to the oxygen pressure regulator, when O2flush button is pressed,
O2 flows into the connector (1) from oxygen pressure regulator
2 Tube connector connected to C.G.O; when O2 flush button is pressed, O2 flows from
connector (1) to connector (2), and finally to the patient.
27
Aeon8800A Service Manual Operation Theory
Auxiliary O2 flow-meter
Auxiliary O2 flow-meter is fixed on the machine with 2screws.there is rotary knob, with which user
can adjust the flow of O2.
2. Auxiliary drive gas output port
When the range of inlet gas pressure is 280—600kpa,the flow of drive gas output is more than
90L/min.
WARNING: The requirement flow of the device connected to the auxiliary drive gas outlet
interface cannot exceed 90L/min.
28
Aeon8800A Service Manual Operation Theory
The breathing system is integrated into a compact aluminum block, directly connecting with
patient’s respiratory tract to help complete breathing process. The main function is to store
anesthesia gas, oxygen and air, discharge waste gas, absorbed carbon dioxide. Also, this block is
heated to body temperature to prevent condensing of humidified gases within the breathing
system, thus improving airway climatization for the patient's re-breathing of respiratory gases.
29
Aeon8800A Service Manual Operation Theory
30
Aeon8800A Service Manual Operation Theory
122005513
Expiratory
valve
122005512
Inspiratory
valve
31
Aeon8800A Service Manual Operation Theory
32
Aeon8800A Service Manual Operation Theory
33
Aeon8800A Service Manual Operation Theory
34
Aeon8800A Service Manual Operation Theory
Aeon8800A machine is divided into two parts, the upper part and down part. The upper parts of
machine provide the operation, alarm and system management functions. The down parts of
machine realize the functions of signal collection, signal process and critical controlling (mode
algorithm) to improve the system integration and stabilization grade, and realize the function of
power management and flowmeter electrical displayed through using MCU.
First, the gas flow and pressure signal convert to electric signal via the flow and pressure sensor.
After the magnification and regulation circuit, the electronic signal converts to digital signal
through the AD conversion circuit. The controller manages the electrical element to realize the
ventilation function. Aeon8800A anesthesia machine is a multi-processor system, the down part
of machine can complete the controlling function while the upper part of machine can realize the
operator-machine intercommunication function; the controller on the communication board can
deal with the keyboard input data and manage the power supply, also can monitor the working
state of host through the communicating data.
35
Aeon8800A Service Manual Operation Theory
Line Power
Power Cord AC Fuse
filter Supply
Power Board
LVDS
12V VGA Board
Touch control
Board Power 5V PC104 5V
Single
Power
COM1 COM1 with me
COM3 with
UI PC104 COM3 Information
systems
LCD
COM4 COM4 with CO2
Communication Flowmeter
Board Connect Board Board
USB USB Reservation
Communication
with Power Board
Internet
Internet
Reservation
Alarm driver
CCFL POWER 12V
CCFL
Single
and Controller
Power
Information
systems
interface
CO2
Insulation Insulation
Interface
interface Board
Speaker
interface Main controller Board O2 O2 Sensor
Sensor
36
Aeon8800A Service Manual Operation Theory
Input AC Touching
panel
LCD
J3
Connect W33
J1 W7 W28
Gas Supply W17 Key Board
Check Board W27
W8 PC104
J4 J3
Touch
J5 JJ1
W5 W32 Control
Switcher Board
Encode
LJ3
Fuse 250V/10A Breathing W34 W6 W25
J6 W35
System W29
W23
AC Indicate Top Light W26 W30
J2
J3
Socket- LED
Isolate Board Speaker
Fuse 250V/2A J4 W22 Top Light
outlet W16 J11 J1 J7 Switch
LED Switcher LED
J13J12 P6 W21 EFM Display Board
Socket- J6 J8
Fuse 250V/2A J1 P4 P1 P4
outlet W19
W11
J3
ConnectA W31
J9
Socket- J2 J5 Flow Sensor
Fuse 250V/2A
outlet
W10 W12 LJ4
Socket- W20 W24 O2
J12
Fuse 250V/2A
J4
Sensor
outlet
J2
J4
J13
Filter Ambient
J5
W9 Board
Pressure
P2
J1
Board
J1
J3 J2 J1
W4
W15 W14
LJ2 W13
LJ1 W2 W3
W1 Switcher Battery Solenoid Woodward
PEEP
Power Group Valve Valve
37
Aeon8800A Service Manual Operation Theory
Part
NO. Part Name Connector Ⅰ Connector Ⅱ
Number
Switching Power Supply input Switching Power
W1 122006831 LJ1
cable Supply
Switching Power
Switching Power Supply output PSB
W2 122006832 Supply output
cable (PSB) J2
connector
J12 on B side of
W5 122006825 Host switch cable Host switch SW1
Interface Board
Interface Board –Breathing J13 on B side of
W6 122006835 LJ3
system heating cable Interface Board
LED
W7 210003473 LED-PC104 cable PC104
LCD G1211L04
J4 on
W8 210003474 Screen backlight cable Communication Screen
Board
J3 on Interface
W9 122006821 PSB-Interface Board cable P2 on PSB
Board
PSB- Interface Board J2 Interface
W10 122006822 J4 on PSB
communication cable Board
Main control
J9 on A side of
W11 122006823 board(MCB)-Interface board J4 on MCB
Interface Board
power supply cable
MCB-Interface board
W12 122006824 J5 on MCB J3 on Interface
communication cable
Solenoid Valve assembly J8 on MCB
W13 122001791 Solenoid Valve
(AMISCO24V) PE to valve
J1 on MCB
W14 240000129 Proportional valve 1350-5083 Inspiratory valve
38
Aeon8800A Service Manual Operation Theory
J11 on B side of
W16 122006826 LED indicator assembly LED signal light
Interface Board
J14 on B side of
W17 122006828 GSPB cable GSPB J1
Interface Board
39
Aeon8800A Service Manual Operation Theory
Speaker
W30 122006833 speaker assembly ISB J3
(8ohm/0.5W)
Flow sensors
SFM7050 O2、
W31 122006847 SFM7050 flow sensor cable EFM board
SFM7050 AIR、
SFM7050 N2O
Communication
W32 122002061 Encoder assembly Encoder
board J3
Breathing system
W34 122001624 Host- breathing system cable LJ3
heating LJ6
Communication board -PC104 Communication
W35 122002194 PC104 J5
power supply cable board J5
NOTE:
GSPB Gas Supply Pressure Board
40
Aeon8800A Service Manual Operation Theory
2 1
J1
P2
J4
J2
3 4
41
Aeon8800A Service Manual Operation Theory
42
Aeon8800A Service Manual Operation Theory
43
Aeon8800A Service Manual Operation Theory
T19 T7 T16
T26
T3
T8
T1 T5
T10
T4
T2
T6
T18
T21
T22
T14
T24
T12
44
Aeon8800A Service Manual Operation Theory
45
Aeon8800A Service Manual Operation Theory
1
J5 J4
J1
2、3、4
J12
J2
J3
J13
J7
46
Aeon8800A Service Manual Operation Theory
47
Aeon8800A Service Manual Operation Theory
48
Aeon8800A Service Manual Operation Theory
Pin Description
1 +5V
2 GND
3 Turbine signal
49
Aeon8800A Service Manual Operation Theory
T4 T3 T10 T6 T9
T2
T8
T1
T5
T12
T11
T7
50
Aeon8800A Service Manual Operation Theory
51
Aeon8800A Service Manual Operation Theory
52
Aeon8800A Service Manual Operation Theory
J11
J13
J12
J1
P6
J14
J10
J7
J6
J8
J4
53
Aeon8800A Service Manual Operation Theory
J3
J2
J5
J3
Pin Description
1 GND
2 LED voltage
54
Aeon8800A Service Manual Operation Theory
Pin Description
1 RETURN
2 DQ
3 OUT
Pin Description
1 GND
2 NC
3 ON
Pin Description
1 +12V
2 GND
3 +5V
4 GND
5 EFM_RX
6 EFM_TX
55
Aeon8800A Service Manual Operation Theory
Pin Description
1 Flow meter signal
2 +10V
3 GND
4 NC
Pin Description
1 PRESSURE1
2 +12V
3 GND
4 +12V
5 S_IN
Pin Description
1 +24V
2 GND
Pin Description
1 +24V
2 ON/OFF
56
Aeon8800A Service Manual Operation Theory
Pin Description
1 +24V
2 Top light
Pin Description
1 Top light
2 OFF
3 Low
4 High
57
Aeon8800A Service Manual Operation Theory
Pin Description
1 +3.6V
2 GND
3 P12V
4 GND
5 A12V
6 GND
7 +5V
8 GND
9 RETURN
10 POUT
11 +24V
12 PWBUS
13 GND
14 GND
Pin Description
1 /PWON
2 RS232 TXD
3 P3.6V_Enable
4 GND
5 TSN(temperature data sample)
6 RS232 RXD
7 GND
8 P12V_Enable
9 PRESSURE_AD2(No Function)
J5: Communication Connector
Connector type: D-SUB, DCH-15P
58
Aeon8800A Service Manual Operation Theory
Pin Description
1 GND
2 BDU_UI_RS232_RXD
3 EFM_RX
4 GND
5 P3.6V_Enable
6 Flow meter signal
7 NC
8 S_IN(No Function)
9 BDU_UI_RS232_TXD
10 GND
11 EFM_RX
12 P12V_Enable
13 GND
14 GND
15 PRESSURE1
Pin Description
1 +3.6V
2 GND
3 GND
4 PWBUS
5 P12V
6 GND
7 GND
8 GND
9 A12V
10 +5V
59
Aeon8800A Service Manual Operation Theory
J3 J4 1
J1
J6
COM1
COM2
COM3
J5 JJ1 COM4 J7
60
Aeon8800A Service Manual Operation Theory
Pin Description
1 +5V
2 GND
3 GND
4 +12V
61
Aeon8800A Service Manual Operation Theory
Pin Description
1 RX-
2 TX-
3 USB+5V
4 PC_D-1
5 OUTA
6 GND
7 COM4RXD
8 COM3RXD
9 RX+
10 TX+
11 PC_D+1
12 OUTB
13 +5V
14 COM4TXD
15 COM3TXD
62
Aeon8800A Service Manual Operation Theory
Pin Description
1 TXD
2 RXD
3 GND
4 +5V
5 NC
Pin Description
1 +12V
2 +12V
3 GND
4 GND
5 ON/OFF
6 DIM
63
Aeon8800A Service Manual Operation Theory
Pin Description
1 GND
2 TSP
3 GND
4 TSB
5 TSA
6 ECV
64
Aeon8800A Service Manual Operation Theory
T1 T4 T3
T2
65
Aeon8800A Service Manual Operation Theory
J1
Pin Description
1 GND
2 Ambient Pressure Signal
3 +5V
66
Aeon8800A Service Manual Operation Theory
J1
Pin Description
1 Gas Supply Pressure Signal
2 A12V
3 NC
4 NC
67
Aeon8800A Service Manual Operation Theory
Pin Description
1 RETURN
2 TSN
3 POUT
4 PE
68
Aeon8800A Service Manual Operation Theory
69
Aeon8800A Service Manual Operation Theory
70
Aeon8800A Service Manual Operation Theory
71
Aeon8800A Service Manual Operation Theory
72
Aeon8800A Service Manual Operation Theory
2.3.2. IP valve
IP valve fixed to the machine with 4 screws on it, shown as Fig. The main function is controlling
PEEP valve, peak pressure and plateau pressure. The PEEP valve regulates the pressure at
which the exhaust valve opens, therefore the exhaust valve opens only when the pressure
exceeds the set PEEP pressure.
Fig. IP Valve
73
Aeon8800A Service Manual Operation Theory
74
Aeon8800A Service Manual Operation Theory
75
Aeon8800A Service Manual Operation Theory
76
Aeon8800A Service Manual Checkout
3. Checkout
WARNING:Make sure that the breathing system is in good connection and with no damage.
CAUTION: Perform the electrical safety inspection as the last step after completing a repair or
after routine maintenance. Perform this inspection with all covers, panels, and
screws installed.
1. Unplug the power cables from the convenience receptacles at the rear of Aeon8800A.
2. Plug Aeon8800A into a safety analyzer.
3. Connect the case ground lead of the analyzer to the U-blade ground of one of the
convenience receptacles. Perform the following tests with the case grounded:
Normal polarity
Normal polarity with open neutral
4. Perform the following tests with the case ungrounded:
Normal polarity
Normal polarity with open neutral
Reverse polarity
5. Verify that the maximum leakage current does not exceed 300μA (0.3mA).
77
Aeon8800A Service Manual Checkout
WARNING: Make sure the assembly of patient breathing tube and controlling setting is correct
before ventilating.
Fig.3-1
Fig.3-2
78
Aeon8800A Service Manual Checkout
Fig.3-3
2) Select and confirms “Skip”, the
“Manual Test” interface appears.
Fig.3-4
79
Aeon8800A Service Manual Checkout
Fig.3-5
Select “Continue”, the system
enter into the Automatic Test
interface.
“Automatic Test…”appears in the
user interface.
Fig.3-6
80
Aeon8800A Service Manual Checkout
Fig.3-7
2. If the bellows is not full, the
message “Push “O2+” flush to
completely fill the Bellows” appears on
the user interface.
1) Select “Continue” to perform
Manual Test again.
2) Select “Skip” to proceed to Manual
Test.
Fig.3-8
81
Aeon8800A Service Manual Checkout
Fig.3-10
82
Aeon8800A Service Manual Checkout
Fig.3-12
5. If the leakage is at the range of
200—400ml/min, the interface
shows as the picture on the right.
Fig.3-13
83
Aeon8800A Service Manual Checkout
Fig.3-14
84
Aeon8800A Service Manual Checkout
Fig.3-15
Select “Skip”, the system will enter
into normal interface.
Fig.3-16
85
Aeon8800A Service Manual Checkout
Fig.3-18
3. If the Y-piece port and manual
bag arm connector are not occluded,
the interface after test is shown as
the picture on the right.
86
Aeon8800A Service Manual Checkout
Fig.3-20
5. If no operation is done during the
first two 60s count down after the
Manual test interface appears, the
interface displays shown as the
picture on the right.
Fig.3-21
6.Manual test success
If the manual test succeeds, the
interface displays as shown on the
right.
The system will enter into Standby
mode automatically after 5s count
time.
Fig.3-22
87
Aeon8800A Service Manual Checkout
3.3.2. Check
In standby mode, when the “Check” button is selected, the system test will be performed including
Leak test and Compliance Test.
Fig.3-24
88
Aeon8800A Service Manual Checkout
Fig.3-25
Fig.3-26
89
Aeon8800A Service Manual Checkout
Fig.3-28
Select “Start” to perform Compliance Test
after completing the procedures on the
screen.
Fig.3-29
90
Aeon8800A Service Manual Checkout
Fig.3-30
Fig.3-31
91
Aeon8800A Service Manual Checkout
1. Attach the digital flow meter to the common fresh gas outlet.
2. Verify that the O2 flush flow is between 35~50L/min when pressing the O2 flush valve.
1. Remove the N2O line pressure hose form the line pressure inlet on Aeon8800A.
2. Mount a full N2O cylinder to the rear panel yoke.
3. Open the N2O cylinder until its pressure gauge indicates cylinder pressure.
4. Close the N2O cylinder.
5. The N2O cylinder pressure gauge should not drop more than 10% of its initial pressure over 1
minute.
1. Remove the O2 line pressure hose form the line pressure inlet on Aeon8800A.
2. Mount a full O2 cylinder to the rear panel yoke.
3. Open the O2 cylinder until its pressure gauge indicates cylinder pressure.
4. Close the O2 cylinder.
5. The O2 cylinder pressure gauge should not drop more than 10% of its initial pressure over 1
minute.
92
Aeon8800A Service Manual Checkout
1. Remove the AIR line pressure hose form the line pressure inlet on Aeon8800A.
2. Mount a full AIR cylinder to the rear panel yoke.
3. Open the AIR cylinder until its pressure gauge indicates cylinder pressure.
4. Close the AIR cylinder.
5. The AIR cylinder pressure gauge should not drop more than 10% of its initial pressure over 1
minute.
93
Aeon8800A Service Manual Checkout
94
Aeon8800A Service Manual Checkout
3.6.2. Drive Gas Pressure Loss Alarm, N2O Cut off Valve Test
3.7. Vaporizers
1. Attach two vaporizers to the vaporizer mounting manifold and lock them in place.
2. Rotate either of the vaporizer dial to 3%.
3. Verify that the other vaporizer dial cannot be rotated to a setting.
4. Set both vaporizer dials to 0.
5. Rotate the other vaporizer dial to 3%.
6. Verify that the first vaporizer dial cannot be rotated.
7. Rotate both vaporizer dials to T and remove both vaporizers.
8. Verify that the locking spring is intact.
9. Reconnect both vaporizers to the vaporizer mounting manifold.
CAUTION Verify the system has passed the leak test without the vaporizers prior to this test.
95
Aeon8800A Service Manual Checkout
1. Ensure that the gas pressure for O2, N2O, and AIR are at 50±10psi.
2. Power ON Aeon8800A.
3. Perform the start up tests per the on-screen instructions. Ensure successful completion.
4. Attach a breathing circuit and test lung to the Y-fitting of the breathing circuit.
5. Set the APL valve to 15cmH2O.
6. Set the AIR flow to 5L/min using the flow control valve.
7. Squeeze the breathing bag once every 10 seconds to inflate and deflate the test lung to
approximately 20cmH2O of pressure.
8. Verify the inflation and deflation of test lung.
96
Aeon8800A Service Manual Checkout
1. While in the manual ventilation mode, stop ventilating the test lung.
2. Verify that the following APNEA alarm signals activates at approximately 60 seconds from the
last bag compression.
APNEA appears on the screen.
An alarm tone sounds.
97
Aeon8800A Service Manual Checkout
98
Aeon8800A Service Manual Checkout
CAUTION: Limit the volume in the test lung to provide sufficient airway pressure to satisfy the
Low Peak Pressure alarm. Or reduce the Peak Pressure alarm limit to a lower
value to prevent the alarm when using an adult test lung.
1. Attach a breathing circuit and breathing bag.
2. Attach an adult test lung to the Y-fitting of the breathing circuit.
3. Attach a vent tester between the expiratory port and expiratory hose.
4. Set the O2 flow to 2L/min and set the N2O and AIR flow rates to minimum flow.
5. Set the ventilation controls to:
99
Aeon8800A Service Manual Checkout
1. While the ventilator is running, disconnect the expiratory limb from the expiratory port on the
Breathing System.
2. Verify the following airway pressure disconnect alarm signals activate:
APNEA message appears on the screen.
Low Airway Pressure message appears on the screen.
An alarm tone sounds.
100
Aeon8800A Service Manual Checkout
101
Aeon8800A Service Manual Checkout
3.9.1. Set Up
1. Ensure that the gas pressure for O2, N2O and AIR are at 50±10psi.
2. Power ON Aeon8800A.
3. Perform the Startup Tests per the on-screen instructions. Ensure successful completion.
4. Attach a breathing circuit and breathing bag.
5. Attach an adult test lung to the Y-fitting of the breathing circuit.
6. Set the O2 flow to 2L/min and set the N2O and AIR flow rates to minimum flow.
7. Set the ventilator controls to:
102
Aeon8800A Service Manual Checkout
103
Aeon8800A Service Manual Checkout
1. Set the MV Low alarm limit set point to the highest value.
2. Verify the following alarms activate:
Low MV message appears on the screen.
An alarm tone sounds.
3. Set the MV Low alarm limit to minimum setting.
4. Verify the alarm signals cease.
5. Set the MV High alarm set point to the lowest value.
6. Verify the following alarms activate:
High MV message appears on the screen.
An alarm tone sounds.
7. Set the MV High alarm limit set point to the highest value.
8. Verify that the alarm signals cease.
104
Aeon8800A Service Manual Checkout
1. Before starting this test, verify that the battery is fully charged. The battery icon in the upper
right corner of the screen is solid when the battery is fully charged.
2. Interrupt AC line voltage.
3. Verify that the following alarms activate:
An alarm tone sounds.
AC Power Failure message appears on the screen.
4. Verify that a fully charged battery operates the ventilator for a minimum of 45 minutes.
5. Plug Aeon8800A into AC line voltage.
6. Verify that the alarm signals cease.
7. Verify the presence of the battery charging icon in the upper right corner of the screen (status
bar moving from right to left).
1. Turn on the work light located on the bottom side of the top panel.
2. Verify that it lights in both on positions.
1. Verify an AIR flow of 15L/min can be obtained by connecting the auxiliary AIR hose to the
pressure source and opening the flow meter.
2. Verify an O2 flow of 15L/min can be obtained by connecting the auxiliary O2 hose to the
pressure source an opening the flow meter.
105
Aeon8800A Service Manual Checkout
106
Aeon8800A Service Manual Configuration menu
4. Configuration menu
This section describes the Service level function of the anesthesia machine.
The Configuration menu has three parts which are password protected, including “Vent Defaults”,
“Alarm Defaults” and “Service Mode”.
Click on “ ”menu on the right column of the screen, the correspond sub-menu appears
on the screen
Fig.4-1
Fig.4-2
107
Aeon8800A Service Manual Configuration menu
108
Aeon8800A Service Manual Configuration menu
Fig.4-5
Enter the password “2020” and click on
“ ”, the interface shown as the picture
on the right appears
If enter wrong password, click on “ ”
to clear the entered number.
109
Aeon8800A Service Manual Configuration menu
Fig.4-7
Enter the right password and click on
“ ”, the interface shown as the picture
on the right appears.
Fig.4-8
110
Aeon8800A Service Manual Configuration menu
4.3.1. Calibration
The calibration submenu including Flow Valve, Flow Sensor, Pressure Sensor, PEEP Valve
and Cal Data. Select “Calibration” submenu, the screen displays as bellow picture.
Fig.4-9
111
Aeon8800A Service Manual Configuration menu
Fig.4-11
112
Aeon8800A Service Manual Configuration menu
Fig.4-12
Fig.4-13
Fig.4-14
113
Aeon8800A Service Manual Configuration menu
Fig.4-15
114
Aeon8800A Service Manual Configuration menu
Remove Bellows,
Restore the cover
Flow Regulator
Fig.4-16 Inspiratory flow sensor calibration diagram
NOTE: Before connecting TSI to anesthesia machine, please set “Gas” of TSI to “Oxygen”, and
set “Flow Units” of TSI to “Volumetric L/min”. The TSI window have been set is shown as
following picture.
115
Aeon8800A Service Manual Configuration menu
Fig.4-18
Select “start”, the screen shown as the
picture on the right appears.
116
Aeon8800A Service Manual Configuration menu
Fig.4-20
Select “Start”, the screen shown as the
picture on the right appears.
117
Aeon8800A Service Manual Configuration menu
Fig.4-24
If all the calibrations of 0 , 3 , 12 ,
30L/min success, the total calibration is
successful and the interface shown as
the picture on the right will appear.
118
Aeon8800A Service Manual Configuration menu
119
Aeon8800A Service Manual Configuration menu
Remove Bellows,
And then restore
the cover
Tube connector
opens to AIR
NOTE: Before connecting TSI to anesthesia machine, please set “Gas” of TSI to “Oxygen”, and
set “Flow Units” of TSI to “Volumetric L/min”. The TSI window have been set is shown as
following picture.
120
Aeon8800A Service Manual Configuration menu
Fig.4-29
121
Aeon8800A Service Manual Configuration menu
Fig.4-31
Select “Start”, the screen shown as the
picture on the right appears.
122
Aeon8800A Service Manual Configuration menu
Fig.4-35
Fig.4-36
123
Aeon8800A Service Manual Configuration menu
124
Aeon8800A Service Manual Configuration menu
Fig.4-38
Fig.4-39
125
Aeon8800A Service Manual Configuration menu
Fig.4-40
Fig.4-41
Fig.4-42
126
Aeon8800A Service Manual Configuration menu
Fig.4-43
Fig.4-44
127
Aeon8800A Service Manual Configuration menu
Fig.4-45
Fig.4-46
Fig.4-47
128
Aeon8800A Service Manual Configuration menu
4.3.2. SW Update
129
Aeon8800A Service Manual Configuration menu
Fig.4-52
130
Aeon8800A Service Manual Configuration menu
Fig.4-53
131
Aeon8800A Service Manual Configuration menu
function.
function. Fig.4-54
132
Aeon8800A Service Manual Configuration menu
Fig.4-57
133
Aeon8800A Service Manual Configuration menu
selected.
Fig.4-58
Fig.4-59
134
Aeon8800A Service Manual Configuration menu
4.3.5. Setup
Fig.4-61
Fig.4-62
135
Aeon8800A Service Manual Configuration menu
136
Aeon8800A Service Manual Troubleshooting
5. Troubleshooting
137
Aeon8800A Service Manual Troubleshooting
138
Aeon8800A Service Manual Troubleshooting
2.GUI、BDU and Keyboard are incompatible. 2. Update software version. Enter SERVICE manual to check
out the real software version. If the version is 00.00, turn the
power supply switch to OFF, and power on again.
3. The data wire is disconnected or 3. Check whether the data wire (signal wire at backside of
damaged. display) is disconnected or is damaged.
4. MCB is defective. 4. Check whether D10, D11 and D12 on MCB are flashing and
Software
Startup test D8 is lighting, if not, there is something wrong with MCB.
Version FAIL
Replace the MCB and perform each item of calibration and
startup test again.
5. The keyboard board is defective. 5. Check whether keyboard indicator is flashing when
keyboard software is not suited to machine, There is
something wrong with keyboard if indicator is not flashing.
Replace the user interface assembly.
6. PC104 board is defective. 6. If all the software communication fail while keyboard and
MCB works well; Replace PC104 board with the compatible
software version.
139
Aeon8800A Service Manual Troubleshooting
Vent/Manual 3. MCB, interface board power supply cable 3. Check whether the connection of cable is correct with a
Startup test
Valve Failure failure multimeter. Or measure the voltage of MCB yellow cable to
earth. If the voltage is not beyond 22-25V, the cable is
defective.
4. MCB is defective 4. Check the connection cable and replace MCB. Then
perform each item of the calibration and startup tests again.
140
Aeon8800A Service Manual Troubleshooting
1. The connection wire of the MCB is 1. Check whether the connection wire of the MCB is
disconnected or is damaged. disconnected or damaged.
Pressure
Startup test
Sensor FAIL
2. The connection of the MCB is defective. 2. Check the power supply. Replace the MCB.
3. The tubing kinked or occluded. 3. Check the tubing for kinks and remove any occlusions.
1. The O2 concentration is below 16%. 1. Increase O2 concentration.
141
Aeon8800A Service Manual Troubleshooting
3. The MCB is defective. 3. Replace the MCB and then perform each item of the
PEEP Valve
Startup test calibration and startup test again.
FAIL
4. The PEEP valve calibration fails. 4. Clean the check valves inside the Breathing System, and
then perform the flow sensor calibration and flow valve
calibration.
5. The PEEP valve is blocked. 5. Clean or replace the PEEP valve.
1. The connection wire of the inspiration 1. Check whether the inspiration valve s disconnected or
valve is disconnected or damaged. damaged.
Inspiration 2. The inspiration valve is defective. 2. Replace the inspiration valve.
Startup test
Valve FAIL
3. The MCB is defective. 3. Replace the MCB and then perform each item of the
calibration and startup test again.
142
Aeon8800A Service Manual Troubleshooting
4. MCB is defective 4. Replace MCB and then perform each item of the calibration
and startup test again.
1. The connection wire to the speaker is 1. Check whether the connection wire of the speaker is
disconnected or damaged. disconnected or damaged.
2. The speaker assembly is not connected. 2. The communication board is not pretty welded, weld the
Alarm
Startup test communication board again.
Speaker FAIL
4. The loop of the speaker is turned off. 4. Replace the speaker.
5. The communication board is defective. 5. Replace the communication board, and confirm the software
version is compatible with the new one.
143
Aeon8800A Service Manual Troubleshooting
pressure mode.
Startup test
displayed on
the pressure
gauge does not
stay within
25~35cmH2O.
144
Aeon8800A Service Manual Troubleshooting
145
Aeon8800A Service Manual Troubleshooting
146
Aeon8800A Service Manual Troubleshooting
147
Aeon8800A Service Manual Troubleshooting
148
Aeon8800A Service Manual Troubleshooting
149
Aeon8800A Service Manual Troubleshooting
The Vt reading 9. Recalibrate the flow sensor again. 9. The calibration of pressure sensor is inaccurate.
is out of Runtime
tolerance. 10. Recalibrate the inspiratory valve again. 10. The calibration of inspiratory valve is inaccurate.
O2flush is
Runtime 11. Replace MCB. 11. MCB is defective.(No flow rate wave)
defective.
150
Aeon8800A Service Manual Troubleshooting
The power switch 1. The power switch or the connection cable is 1. Replace the power switch or the connection
on the main unit is All defective. cable.
non-functional. 2. The communication board is defective. 2. Replace the communication board.
All or some of the 1. The key board is defective. 1. Replace the key board.
keys are All 2. The communication board or COM 1 connection
2. Check and replace the communication board.
non-functional. cable is defective.
1. The encoder is defective. 1. Replace the encoder.
The encoder is
All 2. The communication board or COM 1 connection
non-responsive. 2. Check and replace the communication board.
cable is defective.
1. The work light circuit board or the cable is 1. Replace the work light circuit board or check
All or some of the defective. the wire connection.
work lights are All 2. The flow meter drive backlight is overloaded, which
2. Decrease the flow meter backlight overload.
non-functional. leads to trigger protection.
3. The work light switch is defective. 3. Replace the work light switch.
1. The power supply to the light is defective. 1. Replace the power supply to the light.
The flow meter 2. The foil of light is defective. 2. Replace the foil of light.
light is All 3. The inverter is defective. 3. Replace the inverter.
non-functional. 4. Replace the power board or replace the
4. The flow meter light is still light after power off.
communication board.
151
Aeon8800A Service Manual Troubleshooting
Calibration fails. All 1. The flow valve calibration does not pass. 1. Power off and power on, and then calibrate again,
152
Aeon8800A Service Manual Troubleshooting
153
Aeon8800A Service Manual Troubleshooting
154
Aeon8800A Service Manual Troubleshooting
155
Aeon8800A Service Manual Troubleshooting
156
Aeon8800A Service Manual Troubleshooting
157
Aeon8800A Service Manual Troubleshooting
158
Aeon8800A Service Manual Troubleshooting
3.MCB is defective,
If the voltage is not within the range of 22-25V,
MCB-Interface Board cable is defective, replace
MCB-Interface Board cable
159
Aeon8800A Service Manual Troubleshooting
1. PSB is defective. valve with a multimeter. If there is a good connectivity, the expiratory
valve is defective, replace expiratory valve.
2. MCB is defective.
3. Remove MCB-Interface Board signal cable from the machine.
Expiratory valve Power on, and measure the voltage between T7 and T16 of PSB.
Startup test 3. Expiratory valve connector is
fails
loose. If the voltage is not within the range of 3-4.5V, PSB is defective,
replace PSB.
4. Expiratory valve is defective.
If the voltage is within the range of 3-4.5V, measure the voltage
between T54 and T55 of MCB, If the voltage is not within the range of
3-4.5V, MCB-Interface Board power supply cable is defective.
Replace MCB-Interface Board power supply cable
160
Aeon8800A Service Manual Troubleshooting
2. MCB is defective. If the voltage is not within the range of 11-13V, PSB is defective,
Inspiratory valve replace PSB.
Startup test
fails 3. Inspiratory valve connector is
loose. If the voltage is within the range of 11-13V, measure the voltage
between T56 and Ground of MCB, If the voltage is not within the
range of 11-13V, MCB-Interface Board power supply cable is
defective. Replace MCB-Interface Board power supply cable
161
Aeon8800A Service Manual Troubleshooting
162
Aeon8800A Service Manual Troubleshooting
1 to Ground: 11-13V;
2 to Ground: 11-13V
3 and 4 have connectivity to ground.
1. LED is defective. 5 to Ground: 4-5V;
6 to Ground: 2-5V;
2. Backlight is defective.
One of above condition can’t be satisfied, communication board is
defective, replace the communication board.
Black Screen All 3. PC104 is defective
163
Aeon8800A Service Manual Troubleshooting
164
Aeon8800A Service Manual Troubleshooting
165
Aeon8800A Service Manual Troubleshooting
Problem 2 Something wrong with Startup when using the battery supply only
Phenomenon
Disconnect AC power supply to use the battery only, and rotate the switch to “ON” to find that the
machine can’t startup.
Action:
1. Rotate main switch to “OFF”, Disconnect the AC power.
2. Remove the lower rear panel of the machine.
3. Disconnect the cable connected to the battery.
4. Test the voltage between the middle pin and upper pin; then the voltage between the lower
and the middle pin; Shown as the following picture.
If both the two voltage are over 24V, that’s to say PSB is with a malfunction.
166
Aeon8800A Service Manual Troubleshooting
If one or two voltage is less than 24V, that’s to say cable connected to the battery is with a
malfunction.
Phenomenon:
The automatic leak test fails during the startup self-test.
Action:
1. Observe the airway pressure gauge value during Automatic Leak Test.
The normal airway pressure values changing during Automatic Leak Test:
First, the airway pressure ascends to 80cmH2O, and then quickly descends to 50cmH2O, and
then descends to 30cmH2O slowly and stays at 30cmH2O for a while.
2. If the airway pressure ascends slowly or even doesn’t ascend, the flow valve needs
calibration.
3. If the airway pressure descends from 30cmH2O without staying, that to say the machine has
a leakage.
Phenomenon:
The flow valve calibration in the Service Mode fails.
Action:
1. Make sure the gas supply pressure is normal.
2. Make sure flowmeter is fully closed; the drive gas tube, CGO, absorber canister O2 sensor
and airway pressure gauge are all well connected or installed.
3. Make sure the bellows and Insp. /Exp. valve disks are removed and the bellows cover and
Insp./Exp. Valve covers are installed.
4. Seal the Expiratory port.
167
Aeon8800A Service Manual Troubleshooting
If the flow valve calibration failure is displayed quickly after performing the calibration, at the same
time, the threshold voltage and the maximum flow are very small, recalibrate the flow sensor.
If the flow valve calibration failure is after a while, and the maximum flow rate is bellow 90L/min,
replace the pressure regulator.
Phenomenon:
It needs 5 periods to get to the set value after setting Vt value, and the Vt always changes in the
ventilating process.
Action:
Enter into Service Mode and recalibrate the flow valve.
Phenomenon:
High O2 supply pressure over 0.5Mpa in hospital may lead to the unsteady tidal volume.
Action;
Recalibrate flow valve if the O2 supply pressure is over 0.5Mpa.
168
Aeon8800A Service Manual Troubleshooting
Phenomenon:
The machine has one or some leakage, the leakage place should be found.
Action:
1. First do the manual leak test.
Set to “STANDBY” mode.
Seal Insp. /Exp. Ports and bag arm port.
Rotate the flowmeter knob to make the airway pressure up to 30cmH2O.
Close the flowmeter, observe the airway pressure value to find if the value descends or
not.
2. Second do the automatic leak test.
Set to “STANDBY” mode, open the “Check” sub-menu and to perform the automatic leak test.
If leakage is found both in manual and automatic leak test progress:
Check absorber canister, O2 sensor interface, airway pressure gauge and Insp. /Exp.
Ports valve covers.
If leakage is only found in manual leak test process
Check the APL valve and connector of the bag arm.
If leakage is only found in automatic leak test process
Check the bellows, bellows cover, U-shape sealing ring and drive gas tube.
169
Aeon8800A Service Manual Troubleshooting
Phenomenon:
In the progress of automatic leak test during startup, the bellows falls rather than stays at the top
of bellows cover after the bellows is full filled with gas when pressing “O2+”.
The bellows falls abnormally by accident during normal ventilation, at the same time the leak test
result is normal.
Action:
The phenomenon above indicates that there is something wrong with the bellows check valve.
NOTE: The four screws are in the four deep holes. Don’t remove the breathing system from
machine when disassembly breathing system. First, loose the screws and then remove them
when disassembly.
3. Open the bellows fixing plate shown as the following pictures.
170
Aeon8800A Service Manual Troubleshooting
NOTE: Daub a little silica gel on the O-ring surface and push into the slot softly when installing
bellows check valve.
Install the bellows fixing plate
1. Put the expiratory valve into the bellows fixing board.
2. Make sure there are 5 O-rings and every surface of O-rings is glazed without any damage.
3. Put all the screws into the breathing system holes and the tighten them.
171
Aeon8800A Service Manual Troubleshooting
Expiratory valve
4. Connect drive gas tube, breathing system heating cable and corrugated hose.
5. Install the bellows and bellows cover.
6. Perform the leak test.
Phenomenon:
Absorber canister ascend-descend assembly got stuck and can’t descend.
Action:
Absorber canister lock assembly needs to be replaced.
1. Open the lock assembly and remove absorber canister.
2. Disconnect corrugated hose and chain, and then remove ascend-descend assembly, shown
as the following pictures.
3. Loose and remove the two M4 screws shown as the following picture and replace the lock
assembly.
172
Aeon8800A Service Manual Troubleshooting
Phenomenon:
The loose breathing system lock assembly leads to the leakage from absorber canister.
Action:
Check the two screws on the suspender and two screws on the lock assembly to make sure
whether the screws are loose or not, if so, tighten the screws and check the leakage; if not
regulate the lock assembly again.
1. Remove the absorber canister, corrugated hose and ascend-descend assembly
2. Remove the two M4 screws on the lock assembly and then remove the lock assembly.
3. Loosen the screws on the suspender, rotate to regulate the suspender, make the length of the
suspender 284mm~284.5mm, and then tighten the screws.
173
Aeon8800A Service Manual Troubleshooting
4. Reinstall the lock assembly, ascend-descend assembly, absorber canister and corrugated
hose.
5. Perform leakage test.
174
Aeon8800A Service Manual Maintenance Information
6. Maintenance Information
Warning: Whenever using anesthetic gases, nitrous oxide, oxygen, or any hospital gas
Caution: This device uses high pressure compressed gas. When attaching or
disconnecting backup gas cylinders, the user should always turn the cylinder
valves slowly. Use Aeon8800A flow meters to bleed down the pressure,
watching the cylinder gauge indicate the depleting cylinder pressure, before
disconnecting the cylinder from the yoke. Always open and close cylinder
valves fully.
Caution: This device operates using compressed gas at high pressures from the hospital
central supply. When connecting gas supply lines attach the hose connection to
the machine before connecting the quick disconnect fitting to the hospital source.
Disconnect the supply hose from the hospital source connection prior to
disconnecting it from Aeon8800A gas connection fittings.
175
Aeon8800A Service Manual Maintenance Information
Screen assembly
Touch Screen
5 122012063 1 Yes
Assy
LCD Controlled
8 210004112 1 As needed Yes
NLB150XG01L-01 component
Communication Controlled
9 122006839 1 Yes
board component
10 122005425 Screen Controller 1 Yes
E-flowmeter Assy.
176
Aeon8800A Service Manual Maintenance Information
Auxiliary
12 130007947 Flowmeter 1 Yes
Penetrable Board
Auxiliary
13 122003737 1 Yes
Flowmeter Assy
N2O Input
14 122003754 1 Yes
Integrated Assy
N2O Input
15 130007975 1 Yes
Connector
AIR Input
16 122003753 1 Yes
Integrated Assy
AIR Input
17 130007974 1 Yes
Connector
O2 Input Integrated
18 122003752 1 Yes
Assy
19 130007973 O2 Input Connector 1 Yes
Gas Supply
20 122005527 1 Yes
Pressure Board
Drive Gas
21 122003757 Pressure-Relief 1 Yes
Valve Assy
Single-Output
22 122003756 Pressure-Relief 1 Yes
Valve Assy
Double-Output
23 122003755 Pressure-Relief 2 Yes
Valve Assy
24 130003930 Filter 3 3 years
177
Aeon8800A Service Manual Maintenance Information
High Pressure
25 122003071 N2O Gauge 1 Yes
Assembly
High Pressure AIR
26 122003070 1 Yes
Gauge Assembly
High Pressure O2
27 122003069 1 Yes
Gauge Assembly
Pressure Gauge
28 130008025 6 Yes
Bracket
Low Pressure N2O
29 122005495 1 Yes
Gauge Assy
Low Pressure AIR
30 122005496 1 Yes
Gauge Assy
Low Pressure O2
31 122005497 1 Yes
Gauge Assy
Knob assembly
Switch Assembly
Outlet module
Outlet Module
36 122003760 1 Yes
Assy
Interface board-
breathing system
37 122006835 1 Yes
heating cable of
outputting
178
Aeon8800A Service Manual Maintenance Information
O2 Sensor assembly
N2O pressure
240000542 regulator 1
Assembly
Air pressure
240000543 regulator 1
Assembly
O2 pressure
240000541 regulator 1
Assembly
179
Aeon8800A Service Manual Maintenance Information
Pneumatic Box
49 122006806 1 Yes
Assy
50 122005725 IP Valve Assy 1 Yes
Breathing System
180
Aeon8800A Service Manual Maintenance Information
Assy
Expiratory Valve
68 130007763 1 Yes
Cover
69 122001737 Switch Assy 1 Yes
Expiratory Valve
70 122005518 1 Yes
Assy
Bellow Check
71 122005519 1 Yes
Valve Assy
72 130004091 U-Type Gasket 1 3 Years Yes
181
Aeon8800A Service Manual Maintenance Information
valve Assy
Used in
inspiratory
O- shape silica gel
85 304000108 1 Yes valve Assy、
sealing t 42X1.8
expiratory
valve Assy
Other components
Used in
O-shape silica gel
86 304000044 1 Year Yes vaporizer
sealing14×2.65
mount Assy
87 210003051 LED Signal Light 1 Yes
92 130011897 Panel(E)
Controlled
93 122006843 Interface board
component
94 122003762 Spotlight Assy 1 Yes
Auxiliary Work
95 122003761 1 Yes
Table Assy
O2 Sensor Output
96 122003741 1 Yes
Assy
Five core shielded
210003052 cableLM.BA105.A 1
BE
Integrated Silica
97 130010364 1 3 Years Yes
Tube
Host-breathing
98 122001624 1 Yes
system Cable
99 122005534 Vaporizer 1 Yes
182
Aeon8800A Service Manual Maintenance Information
Mount(two kettles)
Cylinder Bracket
100 130003864 3 Yes
Gasket
Silica gel screwed
101 230000318 1 3 Years Yes
conduitφ17
102 220000280 YOKE Gasket 3 1 Year Yes
183
Aeon8800A Service Manual Maintenance Information
184
Aeon8800A Service Manual Maintenance Information
185
Aeon8800A Service Manual Maintenance Information
WARNING: During installation, make sure the polarity and the plug-in connections are correct
WARNING: Make sure not to kink or squeeze any hoses when reassembling units to avoid the
Before disconnecting pneumatic fittings, bleed all gas pressure from the machine.
1. Close all cylinder valves and disconnect all pipeline supplies from the source.
2. Set the main switch to ON.
3. Ensure that all cylinder and pipeline pressures read zero.
4. Establish a flow for the affected gas to bleed down the pressure.
5. Set the system to Standby mode.
186
Aeon8800A Service Manual Maintenance Information
Figure 6-1
Disassembly the lower rear panel
1. Remove the two fixing screws near battery box with screwdriver.
2. Then remove the other ten screws around the panel.
3. Remove the lower rear panel from the machine with care.
187
Aeon8800A Service Manual Maintenance Information
Figure 6-3
1. Rotate the screwdriver anticlockwise and then remove the fixing screws on the panel.
2. Open the YOKE panel.
188
Aeon8800A Service Manual Maintenance Information
WARNING: During installation, make sure the polarity and the plug-in connections are correct
WARNING: Make sure not to kink or squeeze any hoses when reassembling units to avoid the
1. Remove the screws on the power supply panel. (Right side of above picture)
2. Open the power supply panel Shown as the following picture. (Left side)
3. Loose four screws on the box and remove the cover. Shown as the following picture.(Right
side)
4. Repair or replace any parts in the box as need.
Note: It is only necessary to remove the AC Input Panel when replacing the flow sensor
189
Aeon8800A Service Manual Maintenance Information
Gas inlet integration module includes O2 inlet integration module, AIR inlet integration module and
N2O inlet integration module.
Disassembly of O2 inlet integration module is an example of the replacement of gas inlet
integration module.
190
Aeon8800A Service Manual Maintenance Information
NOTE: The detail tubes connection and cables connection refer to chapter 2.1.2 and chapter
2.2.2.
Warning: After replacing the filter, check and verify the sealing ability of gas inlet integration
module.
Warning: Daub proper glue to the screw when installing the filter to avoid the connector loosing
Figure 6-7
2. Remove the gas connector from the module
and put on a safe place.
3. Pull out the O-ring sealing gasket (1) and Figure 6-8
filter (2). Replace with a new filter. 1 O-ring sealing gasket
2 Filter
3 Fixing screw
191
Aeon8800A Service Manual Maintenance Information
Reassembly:
Figure 6-9
2. Daub proper glue to the screw surface.
Figure 6-10
3. Align the gas connector to other parts of the
module and secure the gas connector with two
fixing screws.
192
Aeon8800A Service Manual Maintenance Information
193
Aeon8800A Service Manual Maintenance Information
W19
W31
194
Aeon8800A Service Manual Maintenance Information
Figure 6-15
a: O2 flow sensor
b: Air flow sensor
c: N2O flow sensor
Disassembly:
E-flowmeter assembly
1. Disconnect gas tubes connected to gas regulators.
2. Disconnect gas tube connected to vaporizer.
3. Disconnect W19 connected to Interface Board for gas supply and communication.
4. Loose and remove four screws secured E-flowmeter assembly to machine on the four corners.
5. Remove E-flowmeter assembly from anesthesia machine.
Gas flow sensors
1. Disconnect all the gas tubes connected to gas flow sensors.
2. Disconnect W31 connected to EFM.
3. Loose and remove four screws secured gas flow sensors to E-flowmeter assembly.
4. Remove gas flow sensors from E-flowmeter assembly.
O2/N2O ratio-controlled valve
1. Disconnect gas tubes connected to O2/N2O ratio-controlled valve.
195
Aeon8800A Service Manual Maintenance Information
2. Loose and remove two screws on the manifold that fixing O2/N2O ratio-controlled valve to
E-flowmeter assembly.
3. Remove O2/N2O ratio-controlled valve from E-flowmeter assembly.
.
Assembly:
1. Replace the trouble parts as needed.
2. Record the serial number of the new parts.
3. Assembly the E-flowmeter assembly in reverse order of disassembly procedure
4. Secure E-flowmeter assembly to the machine.
5. Connect all the gas tubes and cable to E-flowmeter assembly.
196
Aeon8800A Service Manual Maintenance Information
W13
W18
W15
W20
W14
W11 W12
197
Aeon8800A Service Manual Maintenance Information
Interface Board
PSB
198
Aeon8800A Service Manual Maintenance Information
W2
W10
W4
W9
199
Aeon8800A Service Manual Maintenance Information
W1
Figure 6-19
W2
Figure 6-20
Disassembly and Assembly:
1. Remove 2 screws fixing Switching Power Supply to machine frame.
2. Disconnect W2 connected to PSB
2. Disconnect W1 from connected to Interface Board A side.
3. Disconnect W2 from J2 connected to LJ1
6. Replace the new Switching Power Supply
7. Reassembly the new one in reverse order of disassembly procedure.
200
Aeon8800A Service Manual Maintenance Information
WARNING: During installation, make sure the polarity and the plug-in connections are correct
WARNING: Make sure not to kink or squeeze any hoses when reassembling units to avoid the
WARNING: In case of the risk from the wrong connection of tube, pay more attention when
201
Aeon8800A Service Manual Maintenance Information
Inspiratory W18
Valve
Solenoid Valve
Figure 6-22
202
Aeon8800A Service Manual Maintenance Information
8. Record the new version and serial number of new Gas Module
Assembly:
1. Fix the Ambient Pressure Sensor Board to the Gas Module with two screws
2. Fix Gas Module with two screws to the machine
3. Connect W12 of Insp.-Valve、W13of Solenoid Valve cable to JJ1, JJ2 interface on MCB
4. Connect one end of W15 to J6 on SBD, connect another end of W15 to Ambient Pressure
Sensor Board
5. Connect two tubes to connectors on the left side of Gas Module
6. Connect two tubes which has bigger diameter to connectors on the right side of Gas Module.
7. Connect four tubes to connectors on the front side of Gas Module.
CAUTION: The screws fixing proportion valve have plastic gasket, please pay more attention
CAUTION: Line up the proportion valve and holes in the module when reinstalling the
203
Aeon8800A Service Manual Maintenance Information
Figure 6-25
3. Replace and the new proportion valve
4. Record the version and serial number of the new proportion valve.
5. Put the hexagon screws into the proportion valve holes.
6. Put the plastic gasket on the screws.
7. Line up the proportion valve and the module.
8. Rotate the screws counter clockwise with screwdriver and fix proportion valve.
9. Check the gas module sealing ability, refer to chapter 4.3.1.
204
Aeon8800A Service Manual Maintenance Information
CAUTION: Do not forge the sealing gasket when reassembling the solenoid valve to insure the
sealing ability.
Replacement:
205
Aeon8800A Service Manual Maintenance Information
chapter 6.5.4.3.
Figure 6-28
3. Remove the fixing screws of the solenoid valve and the remove the valve from the module.
4. Replace the solenoid valve.
5. Install solenoid valve use the reverse procedure of disassembly
6. Check the gas module sealing ability, refer to chapter 4.3.1.
206
Aeon8800A Service Manual Maintenance Information
CAUTION: Check the sealing gasket when replacing the safety valve. If the sealing gasket is
207
Aeon8800A Service Manual Maintenance Information
CAUTION: Replace APB directly when the ambient pressure sensor doesn’t work normally.
CAUTION: The screws fixing APB have plastic gasket, please pay more attention when
208
Aeon8800A Service Manual Maintenance Information
209
Aeon8800A Service Manual Maintenance Information
Warning: Rotate main switch to “OFF” to close the anesthesia machine when replacing the
battery.
Caution: The disposal of used batteries must be conducted in accordance with local
regulation requirements.
Caution: The battery has the same specifications with Aeon8800A Anesthesia machine is
recommended
Step 1 Step 2
Step 3 Step 4
210
Aeon8800A Service Manual Maintenance Information
Step 5
Figure 6-34、35、36、37、38
Replacement:
1. Disconnect the AC power supply;
2. Set the main switch to “OFF”
3. Remove four screws fixing the battery assembly to machine frame.
4. Pull the battery assembly out.
5. Disconnect battery cable connected to LJ2.
6. Remove the two screws on the securing plate and then remove the securing plate.
7. Pull out the old battery and disposal of it in accordance with local regulation requirements.
8. Replace the old battery with a new one.
9. Reinstall the new battery in the reverse order of disassembly.
211
Aeon8800A Service Manual Maintenance Information
1 Screw
2 Spring
lock
212
Aeon8800A Service Manual Maintenance Information
Four screws
Two screws
a. Remove the four screws in Figure 6-40 and remove the mounting plate;
b. Remove the two screws in Figure 6-41;
c. Pry open the back-shell of display with a flat blade screwdriver at the areas circled in Figure
6-42.Be careful not to damage the display housing.
d. Open the back-shell slowly and remove speaker by removing the two screws in the
Figure 6-43, then remove the back-shell.
Screws
PCB mask
Figure 6-44
213
Aeon8800A Service Manual Maintenance Information
e. Remove the eight screws, and then remove the PCB mask, Figure6-44;
f. Figure6-45, shows the components in the display that can be easily removed.
Figure 6-45
6 1G memory card
Test: When replacing any of the above components, you need to check if the screen is working
normally.
214
Aeon8800A Service Manual Maintenance Information
6.5.7. Spotlight
Figure 6-46
Replacement:
1. Disconnect the AC power supply;
2. Set the main switch to “OFF”;
3. Remove the upper rear panel;
4. Remove the two screws (circled).
5. Disconnect the cables as marked in the above figure.
6. Replace the spotlight with a new one.
7. Reinstall the new spotlight with the reverse procedure of disassembly.
215
Aeon8800A Service Manual How to Get Help
Figure 6-47、48、49
Disassembly:
1. Disconnect the AC power supply;
2. Set the main switch to “OFF”;
3. Remove the screws (circled) around the coping and then remove the coping.
4. Remove the four screws (circled) which fix the top light board on the top plate.
5. Replace the top light backlight board.
6. Record the version and serial number of the new top light board.
7. Reassembly and reinstall the top light board with the reverse procedure of disassembly.
216
Aeon8800A Service Manual Maintenance Information
Figure 6-50、51
1. Remove the upper rear panel of machine.
2. Open the gas inlet integration module panel.
3. Pull out the two pipelines as the direction the arrows marked.
4. Remove the four screws (circled) with the socket wrench.
5. Pull out the vaporizer mount.
6. Replace the vaporizer mount.
7. Reassembly the vaporizer manifold with the reverse procedure of disassembly.
217
Aeon8800A Service Manual How to Get Help
218
This manual No.: 130014954
Directive 93/42/EEC
concerning Medical Devices
Edition 00.00