Aeon8800A Service Manual Statement

Statement
Beijing Aeonmed Co., Ltd. (Aeonmed for short) holds the copyrights to this manual, which is
non-public, published, and reserves the rights to keep it as a secure document. Refer to this
manual when operating, maintaining and repairing products only. Anyone other than Aeonmed
may not make it known to others.

Proprietary materials protected by the copyright law are included in this manual. Any section of it
cannot be reproduced, copied, or translated into other languages without any prior written
approval from Aeonmed who reserves the copyright.

Everything written in this manual is considered to be correct. Aeonmed is not legally responsible
for any mistakes printed within and any damages caused by incorrect installation and operation.
Aeonmed does not supply privileges endowed by the patent law to any other parties. Aeonmed is
not legally responsible for the results caused by patent law breaking and any rights of the third
party violating.

Refer to this manual before the product is used. The manual includes operating procedures which
must be performed with cautiously, operations that may result in non-normal working conditions
and the dangers which may damage equipment or cause bodily harm. Aeonmed is not
responsible for the security, reliability and function of the equipments in case that the dangers,
damages and non-normal phenomenon mentioned in this manual happen. Free repairs for these
malfunctions will not be provided by Aeonmed.

Aeonmed has the rights to replace any content in this manual without notice.

Manufacturer Responsibility:
Aeonmed is responsible for the security; reliability and function of the equipments when to
following conditions are adhered to:

 Installation, adjustments, mending and repairs must be performed by individuals authorized


by Aeonmed;
 Necessary electrical equipment and the working environment must be in accordance with the
national standards, professional standards and the requirements listed in this manual;
 Equipment must be used as instructed in the operating instructions.

I
Aeon8800A Service Manual Statement

CAUTION: This equipment is not for home use.

CAUTION: Malfunctioning equipment may become invalid and cause bodily injury if a

set of effective and approving repairing proposals cannot be submitted by


the institution which is responsible for using this equipment.

The paid theoretical framework diagram will be supplied according to customer requirements by
Aeonmed, plus calibrating method and other information to help the customer, under the
assistance of qualified technicians, repair the equipment parts where can be done by customer
himself based on the stipulation by Aeonmed.

Warranty:
Manufacturing techniques and materials:
For a period of one year from the date of original delivery, the components and assemblies of this
product is warranted to be free from defects manufacturing techniques and materials, provided
that the same is properly operated under the conditions of normal use and regular maintenance.
The warranty period for other parts is three months. Expendable parts are not included. Our
obligation under the above warranties is limited to repairing free of charge.

Free Obligations:
 Our obligation under the above warranties does not include the freight and other fees;

 Aeonmed is not responsible for any direct, indirect or final product broken and delay
which result from improper use, alteration by using the assemblies unratified and
maintenance by anyone other than Aeonmed;

 This warranty does not apply to the followings:

Improper use;

Machines without maintenance or machines broken;

The label of our original serial number or mark is removed or replaced;

Other manufacturers’ product.

Safety, reliability and operating condition:


Aeonmed is not responsible for the safety, reliability and operating condition of this product in
case that:

 The assemblies are disassembled, extended and readjusted

II
Aeon8800A Service Manual Statement

 This product is not operated correctly in accordance with the manual instruction. The
power supply used or operating environment does not follow the requirements in this
manual.

Return
Follow the steps below in case that the product needs to be returned to Aeonmed:
1. Obtain the rights of return
Contact our customer service by informing them the number and type of the product. The number
is marked on the surface of the product. Return is unacceptable if the number cannot be identified.
Enclose a statement of the number, type and the reason of return as well.
2. Transportation charges
Transportation and insurance charges must be prepaid by the user for transporting the product to
Aeonmed for repairing. (Customers charges are added with regard to the products sold to
non-Chinese mainland users).
Use Notice
Welcome to use our products.
To use this product correctly and effectively, the user must read this user manual carefully before
using this product.
When the user is using this product, this user manual must be understood totally and observe
strictly.
This product is only applied to the use this user manual described.
Only trained and professional service persons are allowed to repair and examine this product.
If there is any case happened in the process of use, please call us and intense service will be
provided.
If there is any change in product specification, another informs will not be given.

III
Aeon8800A Service Manual Statement

Manufacture: Beijing Aeonmed Co., Ltd.

Registered Address: 11B2, Fengtai Science Park, Beijing, PEOPLE’SREPUBLIC OF


CHINA

European Shanghai International Holding Corp. GmbH(Europ)


Representative Eiffestrasse 80,20537 Hamburg Germany
Address:

Production No.10, Chaobai St., Yanjiao Development Zone, Sanhe, Hebei


address: Province, PEOPLE’SREPUBLIC OF CHINA

P.C.: 100070

Tel: +86-10-83681616

Fax: +86-10-63718989

Web site: Http://www.aeonmed.com/

Service Hotline: 800-810-8333

E-mail: service@aeonmed.com

IV
Aeon8800A Service Manual Table of Contents

Table of Contents

1. Product Description.................................................................................................................................. 1

1.1. General Information ......................................................................................................................... 1


1.2. Anesthesia machine components ...................................................................................................... 1
1.2.1. Front View ................................................................................................................................ 2
1.3. Side View .......................................................................................................................................... 3
1.4. Back View ......................................................................................................................................... 4
1.4.1. Explosive Veiw .......................................................................................................................... 5
1.4.2. Breathing system components .................................................................................................. 6
1.5. Symbols ............................................................................................................................................ 7
1.6. Abbreviations ................................................................................................................................. 11

2. Operation Theory ................................................................................................................................... 14

2.1. Pneumatic system ........................................................................................................................... 14


2.1.1. Pneumatic Diagram................................................................................................................. 15
2.1.2. Gas tube connection diagram.................................................................................................. 17
2.1.3. Pipeline gas inlet module assembly ......................................................................................... 20
2.1.4. Cylinder gas inlet assembly (Optional) ..................................................................................... 22
2.1.5. Electronic Flowmeter .............................................................................................................. 23
2.1.6. Vaporizer Manifold ................................................................................................................. 26
2.1.7. O2 Flush Valve ......................................................................................................................... 27
2.1.8. Auxiliary O2 output device ....................................................................................................... 28
2.1.9. Breathing System .................................................................................................................... 29
2.1.9.1. Partial structure of breathing system.................................................................................................. 30

2.1.9.2. Detail Structure of Inspiratory/Expiratory Valve.................................................................................. 31

2.1.9.3. Bellows assembly............................................................................................................................... 32

2.1.9.4. Interior structure of bellows............................................................................................................... 33

2.1.9.5. Absorber Canister .............................................................................................................................. 34

2.2. Electrical system ............................................................................................................................. 35


2.2.1. Electrical components overview .............................................................................................. 36
2.2.2. Cable connection diagram ....................................................................................................... 37
2.2.3. PSB (PSB) ................................................................................................................................ 41
2.2.4. Main control board (MCB)....................................................................................................... 46
2.2.5. Interface Board ....................................................................................................................... 53
2.2.6. Communication Board ............................................................................................................ 60

V
Aeon8800A Service Manual Table of Contents

2.2.7. Ambient Pressure Board (APB) ................................................................................................ 66


2.2.8. Gas Supply Pressure Signal Board ............................................................................................ 67
2.2.9. Breathing System Heating Board ............................................................................................. 68
2.2.10. Monitor Interface Board Components Overview ..................................................................... 69
2.2.11. Touch Screen Control Board.................................................................................................... 70
2.3. Electric Box ..................................................................................................................................... 71
2.3.1. Gas module assembly ............................................................................................................. 72
2.3.2. IP valve ................................................................................................................................... 73
2.3.3. Battery Assembly .................................................................................................................... 74
2.4. AC power supply assembly .............................................................................................................. 75
2.5. Screen assembly ............................................................................................................................. 76

3. Checkout ................................................................................................................................................ 77

3.1. Inspect the system .......................................................................................................................... 77


3.2. Electrical safety tests ...................................................................................................................... 77
3.3. System test ..................................................................................................................................... 78
3.3.1. Set-up test .............................................................................................................................. 78
3.3.1.1. System Self–test ................................................................................................................................ 78

3.3.1.2. Automatic Leak Test........................................................................................................................... 80

3.3.1.3. Manual Test....................................................................................................................................... 85

3.3.2. Check...................................................................................................................................... 88
3.3.2.1. Leak Test ........................................................................................................................................... 88

3.3.2.2. Compliance Test ................................................................................................................................ 90

3.4. Gas Delivery System Tests ............................................................................................................... 92


3.4.1. O2 Flush Verification ............................................................................................................... 92
3.4.2. O2:N2O Ratio System ............................................................................................................... 92
3.5. Pneumatic Leak Tests ...................................................................................................................... 92
3.5.1. N2O Cylinder Leak Test ............................................................................................................ 92
3.5.2. O2 Cylinder Leak Test .............................................................................................................. 92
3.5.3. AIR Cylinder Leak Test ............................................................................................................. 93
3.5.4. Line Supply Check Valve Test................................................................................................... 93
3.5.5. N2O Line Pressure Leak Test .................................................................................................... 93
3.5.6. O2 Line Pressure Leak Test ...................................................................................................... 93
3.5.7. AIR Line Pressure Leak Test ..................................................................................................... 94
3.5.8. Cylinder Supply Check Valve Test ............................................................................................ 94
3.6. Breathing System Checks ................................................................................................................ 94

VI
Aeon8800A Service Manual Table of Contents

3.6.1. Internal Gas Connections Test ................................................................................................. 94


3.6.2. Drive Gas Pressure Loss Alarm, N2O Cut off Valve Test............................................................. 95
3.7. Vaporizers ...................................................................................................................................... 95
3.7.1. Vaporizer Interlock Test .......................................................................................................... 95
3.7.2. Vaporizer Leak Test ................................................................................................................. 95
3.8. Performance Verification ................................................................................................................ 96
3.8.1. STANDBY Mode Ventilation Test ............................................................................................. 96
3.8.2. Manual Mode Ventilation Test ................................................................................................ 96
3.8.3. APNEA Alarm Test ................................................................................................................... 97
3.8.4. Alarm MUTE Test .................................................................................................................... 97
3.8.5. IPPV Ventilation Mode Test..................................................................................................... 97
3.8.5.1. IPPV Adult Ventilation Mode Test....................................................................................................... 97

3.8.5.2. IPPV Child Ventilation Mode Test ....................................................................................................... 99

3.8.6. Airway Disconnect Alarm Test ............................................................................................... 100


3.8.7. PCV Adult Ventilation Mode Test .......................................................................................... 100
3.8.8. Pressure Support (PS) Ventilation Mode Test ........................................................................ 101
3.9. Alarms and Failsafe Functions ....................................................................................................... 102
3.9.1. Set Up................................................................................................................................... 102
3.9.2. Low FiO2 Alarm Test .............................................................................................................. 102
3.9.3. High FiO2 Alarm Test ............................................................................................................. 103
3.9.3.1. Peak Pressure Alarms Test ............................................................................................................... 104

3.9.3.2. Minute Volume Alarm Test .............................................................................................................. 104

3.10. Miscellaneous Tests ...................................................................................................................... 105


3.10.1. Test the Line Voltage Alarm .................................................................................................. 105
3.10.2. Wheel Brakes Test ................................................................................................................ 105
3.10.3. Work Light Test..................................................................................................................... 105
3.10.4. Auxiliary Flow Meter ............................................................................................................. 105

4. Configuration menu ............................................................................................................................. 107

4.1. Vent Defaults ................................................................................................................................ 107


4.2. Alarm Defaults .............................................................................................................................. 109
4.3. Service Mode ................................................................................................................................ 110
4.3.1. Calibration ............................................................................................................................ 111
4.3.1.1. Flow Valve ....................................................................................................................................... 112

4.3.1.2. Flow Sensor ..................................................................................................................................... 115

4.3.1.3. Pressure Sensor ............................................................................................................................... 125

VII
Aeon8800A Service Manual Table of Contents

4.3.1.4. PEEP Valve....................................................................................................................................... 127

4.3.1.5. Cal Data........................................................................................................................................... 129

4.3.2. SW Update ........................................................................................................................... 129


4.3.3. Optional Functions................................................................................................................ 132
4.3.4. Optional Language ................................................................................................................ 134
4.3.5. Setup .................................................................................................................................... 135

5. Troubleshooting ................................................................................................................................... 137

5.1. General Troubleshooting............................................................................................................... 137


5.2. General Troubleshooting Guide..................................................................................................... 164

6. Maintenance Information .................................................................................................................... 175

6.1. Maintenance introduction ............................................................................................................ 175


6.2. Maintenance Guidelines ............................................................................................................... 176
6.3. Repair backup parts list ................................................................................................................. 176
6.4. Special Tools Required .................................................................................................................. 184
6.5. Maintenance of machine .............................................................................................................. 186
6.5.1. Bleed gas pressure from the machine.................................................................................... 186
6.5.2. Disassembly the rear panels .................................................................................................. 186
6.5.3. Anesthesia Machine Upper Parts Repairing ........................................................................... 189
6.5.3.1. Power supply panel repairing ........................................................................................................... 189

6.5.3.2. Gas inlet integration module panel repairing .................................................................................... 190

6.5.3.3. Replacement of filter in gas inlet integration module ........................................................................ 191

6.5.3.4. Drive Gas Pressure Regulator Replacement ...................................................................................... 193

6.5.3.5. Replacement of E-flow meter assembly ............................................................................................ 194

6.5.4. Maintenance of electrical box ............................................................................................... 197


6.5.4.1. Replacement of MCB ....................................................................................................................... 197

6.5.4.2. Replacement of PSB (PSB) ................................................................................................................ 198

6.5.4.3. Replacement of Gas Module ............................................................................................................ 201

6.5.4.4. Replacement of proportion valve ..................................................................................................... 203

6.5.4.5. Replacement of solenoid valve ......................................................................................................... 205

6.5.4.6. Replacement of safety valve............................................................................................................. 207

6.5.4.7. Replacement of Ambient Pressure Board (APB) ................................................................................ 208

6.5.4.8. Replacement of PEEP Valve .............................................................................................................. 209

6.5.4.9. Replacement of battery ................................................................................................................... 210

6.5.5. Disassembling the Display from Main Control Unit ................................................................ 212

VIII
Aeon8800A Service Manual Table of Contents

6.5.6. Removal of Components in the Display ................................................................................. 213


6.5.7. Spotlight ............................................................................................................................... 215
6.5.8. Top light ............................................................................................................................... 216
6.5.9. Vaporizer Manifold ............................................................................................................... 217

7. How to Get Help................................................................................................................................... 218

IX
Aeon8800A Service Manual Product Description

1. Product Description

1.1. General Information


The Aeon8800A Anesthesia Machine is a continuous flow anesthesia system which offers manual
or automatic ventilation, easily adjustable fresh gas delivery, anesthetic agent delivery, ventilation
monitoring, convenient ergonomics, and state-of-the-art safety systems.
In terms of theory of operation, the Aeon8800A Anesthesia Machine, driven by pneumatics and
controlled by electricity, is a device used to administer to a patient, continuously or intermittently, a
general inhalation anesthetic and to maintain a patient’s ventilation. The integrated breathing
system delivers the mixture of O2, N2O, AIR and anesthetic agents for patient breathing control,
while electrical control system functions as monitoring patient parameters.
The Aeon8800A Anesthesia Machine is intended for use by licensed clinicians, for patients
requiring anesthesia within a hospital, and can be used in both adults and pediatric populations
weighing above 5kg.

1.2. Anesthesia machine components

Aeon8800A Anesthesia Machine consists of mainframe, anesthetic ventilator, breathing system,


flowmeter and battery.
The optional parts consist of CO2 module,O2 concentration monitor module, anesthetic agent
module, auxiliary inspiration O2 device, air drive, anesthetic gas scavenging system(AGSS) and
vaporizer,Suction device,VP300(Enflurane, Isoflurane, Sevoflurane. interface type: seleca-tec,
using double metal sheet to make temperature compensation, Pour-fill infusion).

1
Aeon8800A Service Manual Product Description

1.2.1. Front View

1
16
15
2 14
13
3
12
4 11
10
5
9

6 8

Fig. 2-1 Aeon8800A Front View

No. Name No. Name

1 Display screen 9 Push-and-pull Handle

2 Knob 10 Auxiliary Work Table

3 Auxiliary O2 flowmeter 11 Cylinder Gauge

4 Electrical flowmeter 12 Power Switch

5 Breathing System 13 Gas Supply Gauge

6 O2 flush 14 Vaporizer Manifold

7 Caster (with brake) 15 Spotlight

8 Drawers Assembly 16 Top light

2
Aeon8800A Service Manual Product Description

1.3. Side View

Fig. 2-2 Aeon8800A Side View

No. Name No. Name


1 SUCTION (Optional) 6 Absorber canister assembly
2 Auxiliary O2 port 7 Breathing system heater cable
socket
3 Auxiliary O2 output port 8 Drive gas interface
4 O2 sensor socket 9 AGSS (Optional)
5 Common gas outlet(C•G•O) 10 Caster

3
Aeon8800A Service Manual Product Description

1.4. Back View

Fig. 2-3 Aeon8800A Back View

NO. Name NO. Name


1 Gas inlet module 5 Caster
2 Power cable 6 Potential equalization connector
3 YOKE 7 Auxiliary mains outlet socket
4 Battery 8 Fuse

4
Aeon8800A Service Manual Product Description

1.4.1. Explosive Veiw

No. Name No. Name

1 Gas Input Assy 3 Electric Box Assy

2 AC Power Assy

5
Aeon8800A Service Manual Product Description

1.4.2. Breathing system components

NO. Name NO. Name


1 Bellows Assembly 8 Inspiratory Port
2 APL(adjust pressure limitation) 9 Test Plug
Valve
3 Breathing Bag Arm 10 Expiratory Port
4 Inspiratory Valve 11 Hook
5 Water drain valve 12 O2 Cell Plug
6 Expiratory Valve 13 Canister
7 Airway Pressure Gauge 14 Canister Release Handle

6
Aeon8800A Service Manual Product Description

1.5. Symbols
The following table provides descriptions of symbols that are used on the device and/or within this
manual.

Symbol Description Symbol

CAUTION ! Attention, consult accompanying documents./


Refer to user manual.

Turn on power switch ON

Turn off power switch OFF

"ON" for a part of equipment

"OFF" for a part of equipment

Alternating current (AC)

Direct current (DC)

Dangerous voltage

Earth (ground)

Protective earth (ground)

7
Aeon8800A Service Manual Product Description

Equipotentiality

Date of manufacture

Manufacturer

Catalogue number

Serial number

Rotation in two directions

Rotate to adjust(increase counter-clockwise)

Lock

Unlock

Insp. Flow

Exp. Flow

8
Aeon8800A Service Manual Product Description

O2 flush O2+

Breathing bag/manual ventilation

Fuse

Read the reading of the top floating-point

Draining valve operation guide

Outer power supply has been connected

Recycle

This mark means that pneumatic and electric device rubbish is


not allowed to be dealt as unclassified city rubbish and should be
collected separately.

Battery charging condition

Alarm mute icon

9
Aeon8800A Service Manual Product Description

Alarm icon

Interference many occur in the vicinity of equipment marked with


this symbol

Authorized representative in the European community

This device complies with the requirements of Medical Device


Directive, 93/42/EEC.

10
Aeon8800A Service Manual Product Description

1.6. Abbreviations
Abbreviation Definition
A
AGSS Anesthetic gas scavenging system
APL Adjustable pressure limiting
B
BDU Basic data unit
C
C Compliance
C·G·O Common gas outlet
cmH2O Centimeters of Water
CO2 Carbon oxide
D
Des. Desflurane
E
EEPROM Electrically erasable programmable read only memory

Enf. Enflurane
ETCO2 End-expiratory CO2 concentration
F
FiO2 Fraction of Inspired Oxygen
Flow-t Flow-time wave
Flow Flow
Freq Frequency
FreqMIN Minimum Frequency in PS mode
G
GUI Graphic user interface
H
Hal. Halothane
I
I:E Inspiratory to Expiratory Ratio
INSCO2 Inspiratory CO2 concentration
Iso. Isoflurane
IPPV Intermittent positive pressure ventilation
L

11
Aeon8800A Service Manual Product Description

L/min Liter per minute


L Liter

M
MAC Minimum Alveolar Concentration
Manual Manual ventilation
PMEAN Mean pressure
mL Mill Liter
MRI Magnetic resonance imaging
MV Minute volume
N
N2O Nitro oxide
O
O2 Oxygen
P
Paw Airway pressure
Paw-t Pressure-time wave
Pb Plumbum
PLAT Plat pressure
PCV Pressure Control Ventilation
PEEP Positive End Expiratory Pressure
Pmax Maximum pressure
PMEAN mean pressure
PPEAK Peak pressure
PS Pressure Support Ventilation
PTARGET Target Pressure
S
SIMV Synchronized Intermittent Mandatory Ventilation
Sev. Sevoflurane
T
TINSP Inspiratory Time
TP Inspiratory Pause Time
Trigger Flow trigger
TSLOPE Inspiratory Slope Time

12
Aeon8800A Service Manual Product Description

V
VT Yidal volume
W
WDT Watch dog timer
ΔP Differential Pressure

13
Aeon8800A Service Manual Operation Theory

2. Operation Theory

2.1. Pneumatic system

Pneumatic system plays the role of supplying O2, N2O, AIR and anesthetic agent to breathing
system, ensuring their pressures are within normal range respectively, so that normal operation of
anesthesia machine and patient’s safety can be guaranteed.

14
Aeon8800A Service Manual Operation Theory

2.1.1. Pneumatic Diagram

Aeon8800A Pneumatic Diagram

15
Aeon8800A Service Manual Operation Theory

1 Inlet filter 21. Safety valve


2 Pressure gauge 22 PEEP valve
3 Check valve 23 Ambient pressure
4 N2O Pressure regulator 24 Flow sensor
5 AIR Pressure regulator 25 Airway pressure
6 O2 Pressure regulator 26 Sensor board
7 Pressure sensor 27 AGSS
8 Pressure regulator 28 Exhaust valve
9 Pressure regulator 29 Bellows check valve
10 Auxiliary breathing O2 30 Bellows
11 Auxiliary supply O2 31 Reversing valve
12 O2, N2O ratio valve 32 APL valve
13 E-flowmeter 33 Breathing bag
14 O2 flush valve 34 Flow sampling
15 Suction 35 Absorber container
16 Check valve 36 O2 sensor
17 Vaporizer mount 37 Insp. Check valve
18 Proportional solenoid valve 38 Exp. Check valve
19 Solenoid valve 39 Breathing system
20 Throttle

16
Aeon8800A Service Manual Operation Theory

2.1.2. Gas tube connection diagram

17
Aeon8800A Service Manual Operation Theory

Tube connection diagram


Technical requirement:
1. Connect the tube connectors marked a same number with a correspond tube.
2. Use a proper force when connecting gas tube to connectors.
3. To ensure the tube depth of the inserted tube.
4. Before connecting 15# and 16# tube, insert the spring (130004100) into the tube.

No. Part Number Part Name Label


1 230000668 PU tube 8/5.5(green) 1
2 230000668 PU tube 8/5.5(green) 2
3 230000582 PU tube 4/2.5(green) 3
4 230000674 PU tube 6/4(yellow) 4
5 230000581 PU tube 4/2.5(yellow) 5
6 230000673 PU tube 6/4(blue) 6
7 230000580 PU tube 4/2.5(blue) 7
8 230000668 PU tube 8/5.5(green) 8

9 230000668 PU tube 8/5.5(green) 9

10 230000668 PU tube 8/5.5(green) 10

11 230000668 PU tube 8/5.5(green) 11

12 230000673 PU tube 6/4 (blue) 12

13 230000580 PU tube 4/2.5(blue) 13

14 230000674 PU tube 6/4(yellow) 14

15 230000581 PU tube 4/2.5(yellow) 15

16 230000668 PU tube 8/5.5(green) 16

17 230000582 PU tube 4/2.5(green) 17

18 230000582 PU tube 4/2.5(green) 18

19 230000582 PU tube 4/2.5(green) 19

18
Aeon8800A Service Manual Operation Theory

20 230000034 Silica gel tubeØ15×2(inside 190mm spring 20


130004100)
21 230000034 Silica gel tubeØ15×2(inside 290mm spring 21
130004100)
22 230000616 Sampling silica tubeØ3×1.75 22

23 230000616 Sampling silica tubeØ3×1.75 23

24 230000616 Sampling silica tubeØ3×1.75 24

25 230000616 Sampling silica tubeØ3×1.75 25

26 230000616 Sampling silica tubeØ3×1.75 26

27 230000582 PU tube 4/2.5(green) 27

28 230000668 PU tube 8/5.5(green) 28

30 230000356 PU tube 8/5.5(transparent) 30

31 230000356 PU tube 8/5.5(transparent) 31

32 230000039 Silica gel tube Ø 1.5×1.5

33 230000712 Double direction, different diameter connector 32

34 230000616 Sampling silica gel tube Ø3X1.75 33

35 230000033 Sampling silica gel tube5/2 34

36 230000616 Sampling silica gel tube Ø3X1.75 35

37 230000616 Sampling silica gel tube Ø3X1.75 36

38 230000033 Sampling silica gel tube5/2 37

39 230000033 Sampling silica gel tube5/2 38

40 230000616 Sampling silica gel tube Ø3X1.75 39

NOTE:Between tube 38 and tube 39, there is a gas resistance which is shaded by arrow.

19
Aeon8800A Service Manual Operation Theory

2.1.3. Pipeline gas inlet module assembly

Pipeline gas supply system consists of gas inlet module and pressure relief valve. Gas inlet
module has three inlet ports for O2, Air and N2O. The pressure regulator can decrease the inlet
gas pressure to meet the need of patients.

20
Aeon8800A Service Manual Operation Theory

No. Part Number Part Name


1 130007976 Gas Input Assembly Fixing Plate
2 122003752 O2 Input Assembly
3 122003753 AIR Input Assembly
4 122003754 N2O Input Assembly
5 122003757 Driver Pressure-Relief Valve Assembly
6 122003756 N2O Pressure-Relief Valve Assembly
7 122003756 AIR Pressure-Relief Valve Assembly
8 122003755 O2 Pressure-Relief Valve Assembly

21
Aeon8800A Service Manual Operation Theory

2.1.4. Cylinder gas inlet assembly (Optional)

Cylinder gas inlet assembly consists of YOKE assembly and pressure regulators. The pressure
regulator can decrease the high pressure gas from cylinder, and then the release gas goes into
pressure relief valve.

Fig. YOKE Assembly Explosive diagram

No. Part Number Part Name


1 240000542 N2O pressure regulator Assembly
2 240000543 Air pressure regulator Assembly
3 240000541 O2 pressure regulator Assembly
4 130011892 YOKE bracket

NOTE: N2O pressure regulator Assembly includes the two nuts and the subpart marked with (1).
Air and O2 pressure regulator Assembly are the same as N2O pressure regulator Assembly.

22
Aeon8800A Service Manual Operation Theory

2.1.5. Electronic Flowmeter

Electronic flowmeter, the mechanical device used to adjust flow control. The flow setting is
displayed to the left of the adjustment knob and also below the corresponding flow tube on the
display. The Total Flow Tube will display the sum of O2, N2O, and air flow. The flow controller is
equipped with a control valve for the ratio of O2 and N2O, so as to ensure that theO2 concentration
is no less than 21% when the user is using O2 and N2O at the same time. With no O2 supply
pressure, N2O is cut off. The effective range of the flow controller is 0~15L/min and the accuracy
is ±10%. When the value is out of the range, the flowmeter readings are for reference only.

E-flowmeter Explosive Diagram1

23
Aeon8800A Service Manual Operation Theory

E-flowmeter Explosive Diagram 2

Flow Sensor Assembly Explosive Diagram

24
Aeon8800A Service Manual Operation Theory

NO. Part Number Part Name


1 122005055 Flowmeter interface assembly
2 122006845 Total flow tube assembly
3 122006841 Flowmeter display board
4 122004100 Oxygen Ratio Control Valve
5 130008323 Oxygen Ratio Control Valve Bracket
6 122006844 Flow Regulator Assembly
7 122006846 Flow Sensor Assembly
8 130011970 Fixed Board
9 210003471 Flow Sensor
10 130012246 Flow Sensor Manifold

NOTE: When replacing Part1 and Part 2, please refer to Diagram1; when replacing Part4, Part6
and Part7, please refer to Diagram 2.
NOTE: Don’t leave filtration and sealing gasket behind when replacing flow sensor.

25
Aeon8800A Service Manual Operation Theory

2.1.6. Vaporizer Manifold

As shown in figure, vaporizer manifold can be used to support anesthetic vaporizer. When
vaporizer manifold is installed with anesthetic vaporizer, the anesthetic gas in it can be input to
anesthesia system through vaporizer manifold. When anesthetic vaporizer is unloaded, vaporizer
manifold can be used as gas flow channel.
Part Number: 122005534

Fig. Back View

1. Gas inlet port, through which gas goes into the vaporizer.
2. Gas outlet port, through which the fresh gas flow to the breathing system.

26
Aeon8800A Service Manual Operation Theory

2.1.7. O2 Flush Valve

As shown follow figure, by pressing the button, high flow of 35L/min~50L/min can be provided to
fresh gas outlet. When connected to Central Gas Supply System or Back-up Cylinder, even if the
machine is power off, press the “O2+” button, the high flow O2 could be sent to breathing system
or breathing bag.

O2 flush valve

1 Tube connector connected to the oxygen pressure regulator, when O2flush button is pressed,
O2 flows into the connector (1) from oxygen pressure regulator
2 Tube connector connected to C.G.O; when O2 flush button is pressed, O2 flows from
connector (1) to connector (2), and finally to the patient.

27
Aeon8800A Service Manual Operation Theory

2.1.8. Auxiliary O2 output device

Fig. Auxiliary O2 output device


1. Auxiliary O2 output port

Auxiliary O2 flow-meter
Auxiliary O2 flow-meter is fixed on the machine with 2screws.there is rotary knob, with which user
can adjust the flow of O2.
2. Auxiliary drive gas output port
When the range of inlet gas pressure is 280—600kpa,the flow of drive gas output is more than
90L/min.

WARNING: The requirement flow of the device connected to the auxiliary drive gas outlet
interface cannot exceed 90L/min.

28
Aeon8800A Service Manual Operation Theory

2.1.9. Breathing System

The breathing system is integrated into a compact aluminum block, directly connecting with
patient’s respiratory tract to help complete breathing process. The main function is to store
anesthesia gas, oxygen and air, discharge waste gas, absorbed carbon dioxide. Also, this block is
heated to body temperature to prevent condensing of humidified gases within the breathing
system, thus improving airway climatization for the patient's re-breathing of respiratory gases.

Fig. Breathing system assembly

NO. Part Name NO. Part Name


1 Bellows Assembly 8 Inspiratory Port
2 APL Valve 9 Test Plug
3 Breathing Bag Arm 10 Expiratory Port
4 Inspiratory Valve 11 Hook
5 Water drain valve 12 O2 Cell Plug
6 Expiratory Valve 13 Canister
7 Airway Pressure Gauge 14 Canister Release Handle

29
Aeon8800A Service Manual Operation Theory

2.1.9.1. Partial structure of breathing system

NO. Part Number Part Name


1 130010341 AC200 Assembly
2 122005511 Airway Pressure Gauge
3 122005512 Inspiratory Valve Assembly
4 122001761 APL Valve
5 122005513 Exspiratory Valve Assembly
6 301000098 Cross recessed pan head screw M3×12

30
Aeon8800A Service Manual Operation Theory

2.1.9.2. Detail Structure of Inspiratory/Expiratory Valve

122005513
Expiratory
valve

122005512
Inspiratory
valve

Fig Inspiratory/Expiratory Valve

NO. Part Number Part Name


1 130006265 Valve nut
2 130007763 Expiratory valve cover
3 130002188 Valve support arm
4 130004102 Valve film
5 130004147 Valve gasket
6 130010350 Valve base
7 304000108 O-type latex gasket Ø42X1.8
8 304000102 O-type latex gasket Ø22X1.8
9 130007762 Inspiratory valve cover

31
Aeon8800A Service Manual Operation Theory

2.1.9.3. Bellows assembly

NO. Part Number Part Name

1 130004086 Bellows dome

2 122005548 Bellows assembly

3 130004091 U-type gasket

4 130010369 Bellows fixing base

32
Aeon8800A Service Manual Operation Theory

2.1.9.4. Interior structure of bellows

NO. Part Number Part Name


1 130010369 Bellows fixing base
2 130010341 AC200 integration
3 130010379 AC200 below cover
4 122005518 Valve assembly
5 122001737 Switch assembly
6 122005519 Bellows check valve assembly
7 130010378 O-type sealing gasket Ø62.5X1.8

33
Aeon8800A Service Manual Operation Theory

8 304000096 O-type silica gel sealing gasket Ø9X1.8


9 304000068 O-type latex gasket 28×1.8
10 304000101 O-type silica gel sealing gasket Ø20×1.8
11 301000333 Hexagon lobular socket head cap screw M4×10

2.1.9.5. Absorber Canister

NO. Part Number Part Name

1 130004092 Absorber hose


2 130006429 Absorber mount assembly
3 130006258 Elastic gasket (1)

4 130003231 Canister soda lime

5 130004098 Absorber screen, metal

6 130006417 Elastic gasket (2)

34
Aeon8800A Service Manual Operation Theory

2.2. Electrical system

Aeon8800A machine is divided into two parts, the upper part and down part. The upper parts of
machine provide the operation, alarm and system management functions. The down parts of
machine realize the functions of signal collection, signal process and critical controlling (mode
algorithm) to improve the system integration and stabilization grade, and realize the function of
power management and flowmeter electrical displayed through using MCU.
First, the gas flow and pressure signal convert to electric signal via the flow and pressure sensor.
After the magnification and regulation circuit, the electronic signal converts to digital signal
through the AD conversion circuit. The controller manages the electrical element to realize the
ventilation function. Aeon8800A anesthesia machine is a multi-processor system, the down part
of machine can complete the controlling function while the upper part of machine can realize the
operator-machine intercommunication function; the controller on the communication board can
deal with the keyboard input data and manage the power supply, also can monitor the working
state of host through the communicating data.

35
Aeon8800A Service Manual Operation Theory

2.2.1. Electrical components overview

Outlet with Fuse Battery 24V

Line Power
Power Cord AC Fuse
filter Supply
Power Board

LVDS
12V VGA Board
Touch control
Board Power 5V PC104 5V

Single

Power
COM1 COM1 with me

COM2 with Main


COM2
controller Board

COM3 with
UI PC104 COM3 Information
systems
LCD
COM4 COM4 with CO2
Communication Flowmeter
Board Connect Board Board
USB USB Reservation
Communication
with Power Board
Internet
Internet
Reservation

Alarm driver
CCFL POWER 12V
CCFL

Single
and Controller

Power
Information
systems
interface
CO2
Insulation Insulation
Interface
interface Board
Speaker
interface Main controller Board O2 O2 Sensor
Sensor

WoodWard PEEP Manual


Control Control Control

WoodWard PEEP Manual


Valve Valve Valve

Fig. Electrical components diagram

36
Aeon8800A Service Manual Operation Theory

2.2.2. Cable connection diagram

Input AC Touching
panel
LCD

J3
Connect W33
J1 W7 W28
Gas Supply W17 Key Board
Check Board W27
W8 PC104

J4 J3
Touch

J5 JJ1
W5 W32 Control
Switcher Board
Encode
LJ3
Fuse 250V/10A Breathing W34 W6 W25
J6 W35
System W29
W23
AC Indicate Top Light W26 W30

J2

J3
Socket- LED
Isolate Board Speaker
Fuse 250V/2A J4 W22 Top Light
outlet W16 J11 J1 J7 Switch
LED Switcher LED
J13J12 P6 W21 EFM Display Board
Socket- J6 J8
Fuse 250V/2A J1 P4 P1 P4
outlet W19
W11

J3
ConnectA W31

J9
Socket- J2 J5 Flow Sensor
Fuse 250V/2A
outlet
W10 W12 LJ4
Socket- W20 W24 O2

J12
Fuse 250V/2A

J4
Sensor
outlet

J2
J4

Power Supply Main Control Board


W18

J13
Filter Ambient

J5
W9 Board
Pressure
P2

J1
Board

J1
J3 J2 J1
W4
W15 W14
LJ2 W13
LJ1 W2 W3
W1 Switcher Battery Solenoid Woodward
PEEP
Power Group Valve Valve

Cable connection diagram

37
Aeon8800A Service Manual Operation Theory

Aeon8800A Anesthesia Machine Cables

Part
NO. Part Name Connector Ⅰ Connector Ⅱ
Number
Switching Power Supply input Switching Power
W1 122006831 LJ1
cable Supply
Switching Power
Switching Power Supply output PSB
W2 122006832 Supply output
cable (PSB) J2
connector

W3 122005710 Battery assembly SLA-5AH24V-3 LJ2 Battery

W4 122005711 3-Power supply cable of PSB J1 on PSB LJ2

J12 on B side of
W5 122006825 Host switch cable Host switch SW1
Interface Board
Interface Board –Breathing J13 on B side of
W6 122006835 LJ3
system heating cable Interface Board
LED
W7 210003473 LED-PC104 cable PC104
LCD G1211L04
J4 on
W8 210003474 Screen backlight cable Communication Screen
Board
J3 on Interface
W9 122006821 PSB-Interface Board cable P2 on PSB
Board
PSB- Interface Board J2 Interface
W10 122006822 J4 on PSB
communication cable Board
Main control
J9 on A side of
W11 122006823 board(MCB)-Interface board J4 on MCB
Interface Board
power supply cable
MCB-Interface board
W12 122006824 J5 on MCB J3 on Interface
communication cable
Solenoid Valve assembly J8 on MCB
W13 122001791 Solenoid Valve
(AMISCO24V) PE to valve
J1 on MCB
W14 240000129 Proportional valve 1350-5083 Inspiratory valve

W15 122002026 IP valve assembly J2 on MCB Expiratory valve

38
Aeon8800A Service Manual Operation Theory

Aeon8800A Anesthesia Machine Cables

J11 on B side of
W16 122006826 LED indicator assembly LED signal light
Interface Board
J14 on B side of
W17 122006828 GSPB cable GSPB J1
Interface Board

W18 122006827 APB cable MCB J13 APB J1

EFM board power supply and


W19 122006830 Interface board J1 EMF board J1
communication board
5-core shield cable
W20 210003052 MCB J12 LJ4
LE.BA105.ABE
Interface –light power supply Interface board
W21 122005731 Light switch
cable J7、1W/LED
J8 on Interface
W22 122006829 Top light switch assembly Boat switch
board B side
J6 on Interface
W23 122005733 Back light assembly of top light top light board
board B side
5-core shield cable LEMO 5-core O2 sensor 、
W24 210003053
LE.BA106.AAW socket Shield layer
Communication board- interface J4 on Interface Communication
W25 122006820
board cable board B side board J6
Communication
W26 122006819 Communication board- ISB cable ISB J2
board JJ1
Communication
W27 PC104
board

27-1 COM1 COM1


122006818 OM1-COM4 cable
27-2 COM2 COM2

27-3 COM3 COM3

27-4 COM4 COM4

Touching panel-touching control Touching control


W28 Touching panel
board cable board
PC104-Touching control board Touching control
W29 122006842 PC104
cable board

39
Aeon8800A Service Manual Operation Theory

Aeon8800A Anesthesia Machine Cables

Speaker
W30 122006833 speaker assembly ISB J3
(8ohm/0.5W)
Flow sensors
SFM7050 O2、
W31 122006847 SFM7050 flow sensor cable EFM board
SFM7050 AIR、
SFM7050 N2O
Communication
W32 122002061 Encoder assembly Encoder
board J3
Breathing system
W34 122001624 Host- breathing system cable LJ3
heating LJ6
Communication board -PC104 Communication
W35 122002194 PC104 J5
power supply cable board J5

NOTE:
GSPB Gas Supply Pressure Board

MCB Main Control Board

EFM Electronic Flowmeter Board

APB Ambient Pressure Board

ISB Isolation Board

TLB Top Light Board

40
Aeon8800A Service Manual Operation Theory

2.2.3. PSB (PSB)

Connectors Quick Reference

2 1

J1

P2

J4

J2

3 4

Top View of PSB

41
Aeon8800A Service Manual Operation Theory

Bottom View of PSB


J1: Battery Connector
Connector type: 3.96mm pitch 1×3 pin header

Pin Name Function


1 BAT Battery Supply Input+
2 GND Power Ground
3 BATV+ Battery Test Input

J2: Power output Connector


Connector type: 3.96mm pitch 1×4 pin header

Pin Name Function


1 +24V 24V DC Input
2 +24V 24V DC Input
3 GND Power Ground
4 GND Power Ground

42
Aeon8800A Service Manual Operation Theory

J4: Communication Connector


Connector type: D-SUB, DCH-9P

Pin Name Function


1 /PWON Power On Enable, Low active
2 RS232 TXD RS232A Transmit Data
3 P3.6V_Enable Power 3.6V Enable, High active
4 GND Ground
5 TSN Temperature Data Sample
6 RS232 RXD RS232A Receive Data
7 GND Ground
8 P12V_Enable Power 12V Enable, High active
9 PRESSURE_AD2 No Function

P2: Power Supply Connector


Connector type: 4.20mm pitch 2×7 pin header

Pin Name Function


1 P3.6V Power 3.6V
2 GND Power Ground
3 P12V Power 12V
4 GND Power Ground
5 A12V Analog Power 12V
6 GND Power Ground
7 +5V Analog Power 5V
8 GND Power Ground
9 RETURN Heating Sample Signal
10 POUT Heating Power Output
11 +24V 24V DC Input
12 PWBUS Power Bus
13 GND Power Ground
14 GND Power Ground

43
Aeon8800A Service Manual Operation Theory

LED Indicating Definition

NO. Designator Status Function


1 DS1 GREEN AC Power Input is active
2 DS2 GREEN Power Bus Output is active
3 DS6 RED Heating Power Output is active
4 DS7 GREEN PSB is OK
5 DS8 GREEN Battery charging is progress

T19 T7 T16

T26

T3
T8

T1 T5

T10
T4

T2
T6

T18

T21
T22
T14

T24
T12

T25 T11 T13 T9 T15

44
Aeon8800A Service Manual Operation Theory

Test Point Definition


Pin Name Function
T1 30V 30V for charge
T2 24V 24V DC Input
T3 PWBUS Power Bus
T4 A12V Analog Power 12V
T5 P12V Power 12V
T6 +5V Analog Power 5V
T7 P3.6V Power 3.6V
T8 BAT Battery Supply Input+
T9 VDD Analog Power 5V
T10 POWER 24V
T11 5VAUX 5V Auxiliary
T12 POUT Heating Power Output
T13 BAT_M Battery Monitor Signal
T14 DQ Temperature Data Sample
T15 DQ1 Ambient Temperature Data Sample
T16 GND Power Ground
T17 GND Power Ground
T18 PWEN Power On Enable, High active
T19 ___ voltage of charge
T20 GND Power Ground
T21 GND Power Ground
T22 GND Power Ground
T24 GND Power Ground
T25 PWM Heating Control Signal
T26 BAT Battery Supply Input+

45
Aeon8800A Service Manual Operation Theory

2.2.4. Main control board (MCB)

Connectors Quick Reference

1
J5 J4

J1

2、3、4
J12

J2

J3
J13

J7

Top View of MCB

46
Aeon8800A Service Manual Operation Theory

Bottom View of MCB

J1: Woodward valve Connector


Connector type: 2.54mm pitch 1×4 pin header

Pin Name Function


1 P12V Power 12V
2 NC NC
3 NC NC
4 Inhale Inspirarory valve

J2: IP valve Connector


Connector type: 3.96mm pitch 1×2 pin header

Pin Name Function


1 P3.6V Power 3.6V
2 Exhale Expiratory valve

47
Aeon8800A Service Manual Operation Theory

J3: Solenoid valve Connector


Connector type: 2.54mm pitch 1×2 pin header

Pin Name Function


1 PWBUS Power Bus
2 Manual Manual/AUTO Valve

J4: Power Supply Connector


Connector type: 4.20mm pitch 2×5 pin header

Pin Name Function


1 P3.6V Power 3.6V
2 GND Power Ground
3 GND Power Ground
4 PWBUS Power Bus
5 P12V Power 12V
6 GND Power Ground
7 GND Power Ground
8 GND Power Ground
9 A12V Analog Power 12V
10 +5V Analog Power 5V

J5: Communication Connector


Connector type: D-SUB,DCH-15P

Pin Description Function


1 GND Power
2 BDU_UI_RS232_RXD BDU Receive Data
3 EFM_RX EFM Receive Data
4 GND Power Ground
5 P3.6V_Enable Power 3.6V Enable, High active
6 Flower meter Flower meter Signal
7 NC NC
8 S_IN(No Function) NC
9 BDU_UI_RS232_TXD BDU Transmit Data

48
Aeon8800A Service Manual Operation Theory

10 GND Power Ground


11 EFM_RX EFM Transmit Data
12 P12V_Enable Power 12V Enable, High active
13 GND Power Ground
14 GND Power Ground
15 PRESSURE1 Pressure of Gas Supply

J12: O2 Sensor Connector


Connector type: 2.54mm pitch 1×4 pin header

Pin Description Function


1 O2_N O2 Sensor signal Input-
2 O2_N O2 Sensor signal Input-
3 O2_P O2 Sensor signal Input+
4 GND Power Ground

J13: Ambient Pressure Signal Connector


Connector type: 2.54mm pitch 1×3 pin header

Pin Description Function


1 GND Power Ground
2 Pair Ambient Pressure Signal
3 +5V Analog Power 5V

J7: Turbine Connector (no function)


Connector type: 2.54mm pitch 1×3 pin header

Pin Description
1 +5V
2 GND
3 Turbine signal

49
Aeon8800A Service Manual Operation Theory

LED Indicating Definition


Designator Status Function
1 D8 GREEN Analog Power 3.3V is active
2、3、4 D10、D11、D12 GREEN(Flash) MCB is OK

T4 T3 T10 T6 T9

T2

T8
T1

T5

T12

T11

T7

Test Point Definitions


Designator Name Function
T1 2.5V Reference Voltage of DAC
T2 AR10V Analog Power 10V
T3 Ex _En Power 3.6V Enable, High active
T4 In _En Power 12V Enable, High active
T5 DA _In DA Voltage of Inspire Valve

50
Aeon8800A Service Manual Operation Theory

T6 A12V Analog Power 12V


T7 DA _Ex DA Voltage of Expire Valve
T8 Ex _AD AD Voltage of Expire Valve
T9 I _In DA Voltage of Inspire Valve
T10 E _In DA Voltage of Expire Valve
T11 M _Valve Manual/AUTO Valve Check
T12 M_EN Manual/AUTO Valve Enable, Low active
T13 NC NC
T14 NC NC
T15 Flower mete _AD Flower mete _AD Voltage
T16 CCP Not Use
T17 GND Power Ground
T18 3.3V Analog Power 3.3V
T19 1.8V Analog Power 1.8V
T20 In _AD AD Voltage of Inspire Valve
T21 CLK Clock Signal
T22 CPU_RST CPU RESET Signal
T23 TX_B BDU Transmit Data
T24 RX_B BDU Receive Data
T25 NC NC
T26 NC NC
T27 NC NC
T28 NC NC
T29 5V Analog Power 5V
T30 5.5V Analog Power 5.5V
T31 GND Power Ground
T32 REF Reference Voltage 5V
T33 INSP_AD Inspiration flow voltage
T34 S_INSP Inspiration Sensor Signal
T35 2.048V Reference Voltage 2.048V
T36 NC NC
T37 GND Power Ground
T38 PAIR Ambient Pressure Signal
T39 GND Power Ground

51
Aeon8800A Service Manual Operation Theory

T41 GND Power Ground


T42 PAW Pressure inside airway
T43 S_IN Not Use
T44 SUP_AD Not Use
T45 S_PAW Airway Sensor Signal
T46 O2 O2 Sensor Signal
T47 NC NC
T48 O2 Sensor O2 Sensor Amplified Signal
T49 AAA Not Use
T50 AAA Not Use
T51 24V 24V
T52 GND Power Ground
T53 NC NC
T54 P3.6V Power 3.6V
T55 GND Power Ground
T56 P12V Power 12V
T57 GND Power Ground
T58 SUP1_AD AD voltage of Gas Supply
T59 GND Power Ground
T60 GND Power Ground
T61 GND Power Ground
T62 GND Power Ground
T63 PRESSURE1 Signal of Gas Supply
T64 GND Power Ground

52
Aeon8800A Service Manual Operation Theory

2.2.5. Interface Board

Connectors Quick Reference

J11

J13

J12

J1

P6

J14

J10

J7

J6

J8
J4

Top view of Interface Board

53
Aeon8800A Service Manual Operation Theory

J3

J2

J5

J3

Bottom View of Interface Board

J11: AC Power LED Indicators Connector


Connector type: 2.55mm pitch 1×2 pin header

Pin Description
1 GND
2 LED voltage

54
Aeon8800A Service Manual Operation Theory

J13: Heating Connector


Connector type: 3.96mm pitch 1×3 pin header

Pin Description
1 RETURN
2 DQ
3 OUT

J12: Switcher Connector


Connector type: 2.54mm pitch 1×3 pin header

Pin Description
1 GND
2 NC
3 ON

J1: EFM Connector


Connector type: 2.54mm pitch 1×6 pin header

Pin Description
1 +12V
2 GND
3 +5V
4 GND
5 EFM_RX
6 EFM_TX

55
Aeon8800A Service Manual Operation Theory

P6: Flow Sensor Connector


NOTE: Available for Aeon8600A and Aeon8300A, invalid for Aeon8800A
Connector type: 2.0mm pitch 1×4 pin header

Pin Description
1 Flow meter signal
2 +10V
3 GND
4 NC

J14: Gas supply Signal Connector


Connector type: 2.54mm pitch 1×5 pin header

Pin Description
1 PRESSURE1
2 +12V
3 GND
4 +12V
5 S_IN

J10: Flow Meter Backlight Connector


Connector type: 2.54mm pitch 1×2 pin header

Pin Description
1 +24V
2 GND

J7: LED Connector


Connector type: 3.96mm pitch 1×2 pin header

Pin Description
1 +24V
2 ON/OFF

56
Aeon8800A Service Manual Operation Theory

J6: Top Light Connector


Connector type: 2.54mm pitch 1×2 pin header

Pin Description
1 +24V
2 Top light

J8: Top Light Switch Connector


Connector type: 2.54mm pitch 1×4 pin header

Pin Description
1 Top light
2 OFF
3 Low
4 High

J4: Power and communication Connector


Connector type: D-SUB, DCH-25P

Pin Description Pin Description


1 GND 2 BDU_UI_RS232_RXD
3 RS232 TXD 4 GND
5 NC 6 NC
7 NC 8 NC
9 NC 10 +24V
11 +24V 12 GND
13 GND 14 BDU_UI_RS232_TXD
15 GND 16 RS232 TXD
17 NC 18 NC
19 NC 20 NC
21 NC 22 +24V
23 +24V 24 GND
25 GND

57
Aeon8800A Service Manual Operation Theory

J3: Power Supply Connector


Connector type: 4.20mm pitch 2×7 pin header

Pin Description
1 +3.6V
2 GND
3 P12V
4 GND
5 A12V
6 GND
7 +5V
8 GND
9 RETURN
10 POUT
11 +24V
12 PWBUS
13 GND
14 GND

J2: Communication Connector


Connector type: D-SUB, DCH-9P

Pin Description
1 /PWON
2 RS232 TXD
3 P3.6V_Enable
4 GND
5 TSN(temperature data sample)
6 RS232 RXD
7 GND
8 P12V_Enable
9 PRESSURE_AD2(No Function)
J5: Communication Connector
Connector type: D-SUB, DCH-15P

58
Aeon8800A Service Manual Operation Theory

Pin Description
1 GND
2 BDU_UI_RS232_RXD
3 EFM_RX
4 GND
5 P3.6V_Enable
6 Flow meter signal
7 NC
8 S_IN(No Function)
9 BDU_UI_RS232_TXD
10 GND
11 EFM_RX
12 P12V_Enable
13 GND
14 GND
15 PRESSURE1

J9: Power Supply Connector


Connector type: 4.20mm pitch 2×5 pin header

Pin Description
1 +3.6V
2 GND
3 GND
4 PWBUS
5 P12V
6 GND
7 GND
8 GND
9 A12V
10 +5V

59
Aeon8800A Service Manual Operation Theory

2.2.6. Communication Board

Connectors Quick Reference

J3 J4 1

J1

J6

COM1
COM2
COM3

J5 JJ1 COM4 J7

Top View of Communication Board

Bottom View of Communication Board

60
Aeon8800A Service Manual Operation Theory

J4: Power and communication Connector


Connector type: D-SUB, DCH-25P

Pin Description Pin Description


1 GND 2 BDU_UI_RS232_RXD
3 RS232 TXD 4 GND
5 NC 6 NC
7 NC 8 NC
9 NC 10 +24V
11 +24V 12 GND
13 GND 14 BDU_UI_RS232_TXD
15 GND 16 RS232 TXD
17 NC 18 NC
19 NC 20 NC
21 NC 22 +24V
23 +24V 24 GND
25 GND

J5: PC104 Power Supply Connector


Connector type: 3.96mm pitch 1×4 pin header

Pin Description
1 +5V
2 GND
3 GND
4 +12V

61
Aeon8800A Service Manual Operation Theory

JJ1: Isolation Board Connector


Connector type: D-SUB, DCH-15S

Pin Description
1 RX-
2 TX-
3 USB+5V
4 PC_D-1
5 OUTA
6 GND
7 COM4RXD
8 COM3RXD
9 RX+
10 TX+
11 PC_D+1
12 OUTB
13 +5V
14 COM4TXD
15 COM3TXD

COM1—COM2: Serial Port Connectors


Connector type: 2.00mm pitch 2×5 box header

Pin Description Pin Description


1 DCD# 2 RXD
3 TXD 4 DTR#
5 GND 6 DSR#
7 RST# 8 CTS#
9 R1# 10 N/C

62
Aeon8800A Service Manual Operation Theory

J7: CO2 Connector (Optional)


Connector type: 3.96mm pitch 1×5pin header

Pin Description
1 TXD
2 RXD
3 GND
4 +5V
5 NC

J1: Keyboard Connector


Connector type: 2.54mm pitch 2×5 box header

Pin Description Pin Description


1 GND 2 K1
3 K2 4 K3
5 /PWON 6 GND
7 K5 8 K6
9 K7 10 N/C

J4: Inverter Connector


Connector type: 1.25mm pitch 1×6pin header

Pin Description
1 +12V
2 +12V
3 GND
4 GND
5 ON/OFF
6 DIM

63
Aeon8800A Service Manual Operation Theory

J3: Encode Connector


Connector type: 2.54mm pitch 1×6pin header

Pin Description
1 GND
2 TSP
3 GND
4 TSB
5 TSA
6 ECV

LED Indicating Definition


NO. Designator Status Function
1 D8 GREEN(Flash) Communication Board is OK

64
Aeon8800A Service Manual Operation Theory

T1 T4 T3

T2

Test Point Definitions


Designator Name Function
T1 +5V Analog Power 5V
T2 P12V Power 12V
T3 +3.3V Analog Power 3.3V
T4 GND Power Ground
T5 I2C1_SCL I2C1 Clock Signal
T6 I2C1_SDA I2C1 Data Signal
T7 OUTB Speaker -
T8 OUTA Speaker +

65
Aeon8800A Service Manual Operation Theory

2.2.7. Ambient Pressure Board (APB)

Connectors Quick Reference

J1

Top View of APB

Bottom View of APB


J1: Ambient Pressure Signal Connector
Connector type: 2.54mm pitch 1×3 pin header

Pin Description
1 GND
2 Ambient Pressure Signal
3 +5V

66
Aeon8800A Service Manual Operation Theory

2.2.8. Gas Supply Pressure Signal Board

Connectors Quick Reference

J1

Top View of Gas Supply Pressure Signal Board

Bottom View of Gas Supply Pressure Signal Board

J1: Gas Supply Pressure Signal Connector


Connector type: 2.54mm pitch 1×4 pin header

Pin Description
1 Gas Supply Pressure Signal
2 A12V
3 NC
4 NC

67
Aeon8800A Service Manual Operation Theory

2.2.9. Breathing System Heating Board

Connectors Quick Reference

Top View of Breathing System Heating Board

Bottom View of Breathing System Heating Board

Connector type: LEMO4, PKG. NO.4SL.LG

Pin Description
1 RETURN
2 TSN
3 POUT
4 PE

68
Aeon8800A Service Manual Operation Theory

2.2.10. Monitor Interface Board Components Overview

1 J3 Connector Single Board Computer CN27 CRT/VGA/LCD connection

2 J6 Connector Single Board Computer CN15 USB1/2 connection

3 J2 Connector Single Board Computer CN26 LAN connection

4 J4 Connector VGA interface

5 J5,J7 USB interface


Connector

6 J1 Connector NET interface

69
Aeon8800A Service Manual Operation Theory

2.2.11. Touch Screen Control Board

1 Touch Screen connection

2 Industrial personal computer USB connection

70
Aeon8800A Service Manual Operation Theory

2.3. Electric Box

Fig. Electric box of Aeon8800A

NO. Part Number Part Name


1 122006807 Pneumatic box
2 210003567 Switching power supply
3 122006836 PSB
4 122006843 Interface board
5 122006871 Gas module assembly
6 122007196 Battery assembly
7 122005725 IP valve assembly
8 122006838 Main control board

71
Aeon8800A Service Manual Operation Theory

2.3.1. Gas module assembly

NO. Part Number Part Name


1 130003915 Gas Exhaust port
2 122000834 Safety valve
3 304000120 O-shape rubber sealing gasket 25 X 1.8
4 130003913 Position pointed ferrule
5 304000099 O-shape rubber sealing gasket16×1.8
6 130003914 Inhalation gas port (Drive gas)
7 240000097 Solenoid Valve
8 220000426 Right angle gas connector
9 130003912 Gas integration module
10 304000096 O-shape rubber sealing gasket 9X1.8
11 210003135 Proportional Valve
12 122006848 Ambient Pressure Board

72
Aeon8800A Service Manual Operation Theory

2.3.2. IP valve

IP valve fixed to the machine with 4 screws on it, shown as Fig. The main function is controlling
PEEP valve, peak pressure and plateau pressure. The PEEP valve regulates the pressure at
which the exhaust valve opens, therefore the exhaust valve opens only when the pressure
exceeds the set PEEP pressure.

Fig. IP Valve

73
Aeon8800A Service Manual Operation Theory

2.3.3. Battery Assembly

Figure Battery Assembly

NO. Part Number Part Name


1 130007963 Battery Box
2 210002830 Battery
3 130007965 Battery Shield

74
Aeon8800A Service Manual Operation Theory

2.4. AC power supply assembly

No. Part No. Part Name


1 210002134 Power cable
2 210000181 Mains socket
3 210001737 Fuse 250V/10A
4 210000208 Fuse 250V/2A
5 130011885 Power supply bracket
6 122001805 Speaker assembly
7 122006840 Isolation board
8 130007943 Socket fixing plate
9 210001746 European-type Terminal Assy
10 210001787 Power supply waveform filter
11 130011887 Power supply cover

75
Aeon8800A Service Manual Operation Theory

2.5. Screen assembly

NO. Name NO. Name


1 Knob board 7 Monitor interface board
2 Knob LED & Mute Key Board 8 Industrial control board
3 Monitor cable 9 Alarm LED board
4 Display connector 10 Monitor Control Board
5 USB connector 11 LCD Backlight inverter board
6 NET connector

76
Aeon8800A Service Manual Checkout

3. Checkout

3.1. Inspect the system

WARNING:Make sure that the breathing system is in good connection and with no damage.

 Make sure the anesthesia machine has no damage.


 Make sure all parts are correctly connected.
 Make sure the breathing system is correctly connected and has no damage; the
absorber canister is fully filled with soda lime.
 Make sure the vaporizer is well locked and is filled with adequate anesthetic agent.
 Make sure the pipeline gas supply system is correctly connected and is at proper
pressure.
 Make sure backup equipments are available and functioning.
 Make sure anesthetic agent and emergent medicine are available.
 Make sure the castors are not loose and the forefront two castors are locked.
 Make sure the anesthesia machine is connected to power socket and power
indicator is light. There is no power supply if power indicator does not light.

3.2. Electrical safety tests

CAUTION: Perform the electrical safety inspection as the last step after completing a repair or

after routine maintenance. Perform this inspection with all covers, panels, and
screws installed.
1. Unplug the power cables from the convenience receptacles at the rear of Aeon8800A.
2. Plug Aeon8800A into a safety analyzer.
3. Connect the case ground lead of the analyzer to the U-blade ground of one of the
convenience receptacles. Perform the following tests with the case grounded:
 Normal polarity
 Normal polarity with open neutral
4. Perform the following tests with the case ungrounded:
 Normal polarity
 Normal polarity with open neutral
 Reverse polarity
5. Verify that the maximum leakage current does not exceed 300μA (0.3mA).

77
Aeon8800A Service Manual Checkout

3.3. System test

3.3.1. Set-up test

WARNING: Make sure the assembly of patient breathing tube and controlling setting is correct

before ventilating.

1. Plug the power cord to an electrical outlet.


2. Connect the pipeline gas resource or cylinder gas resource (open the cylinder valve).
3. Close O2, N2O, Air flowmeter and the auxiliary O2 flowmeter.
4. Turn the switch to “ON”.

The user interface is lighted,


entering into the system self-test
interface.

Fig.3-1

3.3.1.1. System Self–test

Enter into the system self-test


interface after opening.

Fig.3-2

78
Aeon8800A Service Manual Checkout

After the system self-test:


1) Select and confirm “Continue”, the
Automatic Test begins.

Fig.3-3
2) Select and confirms “Skip”, the
“Manual Test” interface appears.

Fig.3-4

79
Aeon8800A Service Manual Checkout

3.3.1.2. Automatic Leak Test

After System Self Test, the


automatic Test begins.
Follow instructions on the user
interface.
Select “Continue” to perform
Automatic Test.
2) Select “Skip” to proceed to
Manual Test.

Fig.3-5
Select “Continue”, the system
enter into the Automatic Test
interface.
“Automatic Test…”appears in the
user interface.

Fig.3-6

80
Aeon8800A Service Manual Checkout

These situations may appear during the automatic test,


If these situations appear, please:
1. Do according to the prompt message on the user interface.
2. Select “Continue” and press the knob ,the system begins to re-test, The interface displays
shown as above picture.

1. If the E-flowmeter isn’t closed, the


message “Please close the flowmeter”
appears on the user interface.

Select “Continue” to perform


Automatic Test again.
2) Select “Skip” to proceed to Manual
Test.

Fig.3-7
2. If the bellows is not full, the
message “Push “O2+” flush to
completely fill the Bellows” appears on
the user interface.
1) Select “Continue” to perform
Manual Test again.
2) Select “Skip” to proceed to Manual
Test.

Fig.3-8

81
Aeon8800A Service Manual Checkout

Automatic Test results


1. Automatic Test failed

If the pressure sensor (Ambient


pressure sensor and Airway
pressure sensor) is disabled, the
automatic test will fail and the
prompt message will appear on the
user interface.

In the situation, only the manual


Fig.3-9
ventilation is available.

Select “Manual” to enter into


Standby screen.
2. Automatic Test failed
If the safety valve in the gas
module is disabled the automatic
test will fail, and the interface as
the picture on the right appears.
Select “Continue” to perform
Manual Test, the automatic
ventilation is disabled.

Fig.3-10

82
Aeon8800A Service Manual Checkout

3. Automatic Test failed


If the pressure rises during
automatic test, the test will fail and
the interface as the picture on the
right appears.

1) Select “Retry” to do the


Automatic Test again.
2) Select “Skip” to proceed to
Manual Test, the automatic
Fig.3-11
ventilation is disabled.
4. If the leakage is more than
400ml/min, the interface shows as
the picture on the right.

1) Select “Retry” to perform the


Automatic Test again.
2) Select “Skip” to proceed to the
Manual Test, the automatic
ventilation is disabled.

Fig.3-12
5. If the leakage is at the range of
200—400ml/min, the interface
shows as the picture on the right.

1) Select “Retry” to perform the


Automatic Test again.
2) Select “Skip” to proceed to the
Manual Test, the automatic
ventilation is disabled.

Fig.3-13

83
Aeon8800A Service Manual Checkout

6. If the safety valve is valid and


the leakage is less than
200ml/min, the automatic test is
passed and then the system
proceeds to the Manual Test.

Fig.3-14

84
Aeon8800A Service Manual Checkout

3.3.1.3. Manual Test

1. If the automatic test is not done


before performing the manual test,
the user interface of manual test will
display as the picture on the right.

Follow the instructions on the user


interface.

Fig.3-15
Select “Skip”, the system will enter
into normal interface.

2. If system enters into the Manual


Test when partial automatic test is
done, the user interface will display
as the picture on the right.

Press “O2+” button according to the


instructions on the user interface
during 60s.

Fig.3-16

Select “Skip”, the system will enter


into normal interface.

85
Aeon8800A Service Manual Checkout

Manual test results

1. If the PAW gauge value is bellow


20 cmH2O, the interface will display
as the picture on the right.

During the 60s count time proceed


the procedure according to the
instructions on the user interface.

Select “Skip”, the system will enter


into Standby interface.
Fig.3-17
2. If the manual test fails for three
times, the interface displays as the
picture on the right.

Select “Skip”, the system will enter


into Standby interface.

Fig.3-18
3. If the Y-piece port and manual
bag arm connector are not occluded,
the interface after test is shown as
the picture on the right.

During the 60s count time proceed


the procedures according to the
instructions on the user interface. Or
the system will enter into Standby
interface automatically after 60s.
Fig.3-19

Select “Skip”, the system will enter


into Standby interface.

86
Aeon8800A Service Manual Checkout

4. There is no operation of pushing


“O2+” to fulfill the bellows during the
first 60s count time after manual test
interface appears, the interface on
the right follows.

Select “Skip”, the system will enter


into Standby interface.

Fig.3-20
5. If no operation is done during the
first two 60s count down after the
Manual test interface appears, the
interface displays shown as the
picture on the right.

Select “Skip”, the system will enter


into Standby interface.

Fig.3-21
6.Manual test success
If the manual test succeeds, the
interface displays as shown on the
right.
The system will enter into Standby
mode automatically after 5s count
time.

Fig.3-22

87
Aeon8800A Service Manual Checkout

3.3.2. Check

In standby mode, when the “Check” button is selected, the system test will be performed including
Leak test and Compliance Test.

CAUTION: Check can only be performed in standby mode.

3.3.2.1. Leak Test

After the “Check” button is selected,


the interface is as shown in the
figure on the right.
Follow instructions on screen to
perform test

CAUTION: Leak Test can only be

performed in standby mode.


Fig.3-23

Select “Start” to perform Leak Test


after completing the procedures on
the screen.

Fig.3-24

88
Aeon8800A Service Manual Checkout

If the Leak Test is success, the


interface is shown as the picture on
the right.
Select “Finish” to perform
Compliance Leak

Fig.3-25

If the Leak Test fails, the interface is


shown as the picture on the right.
Select “Repeat” to perform the Leak
Test again.

Select “Return” to enter into normal


interface.

Fig.3-26

89
Aeon8800A Service Manual Checkout

3.3.2.2. Compliance Test

Select “Compliance Test” sub-menu, if


the interface shown as the picture on the
right, select the “Leak Test” sub-menu to
perform leak test.
Select “Return” to enter into normal
interface.

NOTE: Perform the Leak Test prior to the


Compliance Test when starting the Fig.3-27
system test.
If the Leak Test is performed first, the
interface as the picture shown on the right
appears after selecting Compliance Test
sub-menu.

Follow instructions on screen to perform


test.

Fig.3-28
Select “Start” to perform Compliance Test
after completing the procedures on the
screen.

Fig.3-29

90
Aeon8800A Service Manual Checkout

If the Compliance Test is success, the


interface is shown as the picture on the
right.
Select “Return” to finish the test.

Fig.3-30

If the Compliance Test fails, the interface


is shown as the picture on the right.
Select “Repeat” to perform the Leak Test
again.
Select “Return” to enter into normal
interface.

Fig.3-31

91
Aeon8800A Service Manual Checkout

3.4. Gas Delivery System Tests

3.4.1. O2 Flush Verification

1. Attach the digital flow meter to the common fresh gas outlet.
2. Verify that the O2 flush flow is between 35~50L/min when pressing the O2 flush valve.

3.4.2. O2:N2O Ratio System

1. Set the O2 and N2O flow control valves to minimum.


2. Rotate the N2O flow control valve until the top of the O2 flow meter’s floater rise to 1L/min.
3. Verify that the N2O flow is not higher than 3.7L/min.
4. Lower the N2O flow to minimum.

3.5. Pneumatic Leak Tests

3.5.1. N2O Cylinder Leak Test

1. Remove the N2O line pressure hose form the line pressure inlet on Aeon8800A.
2. Mount a full N2O cylinder to the rear panel yoke.
3. Open the N2O cylinder until its pressure gauge indicates cylinder pressure.
4. Close the N2O cylinder.
5. The N2O cylinder pressure gauge should not drop more than 10% of its initial pressure over 1
minute.

3.5.2. O2 Cylinder Leak Test

1. Remove the O2 line pressure hose form the line pressure inlet on Aeon8800A.
2. Mount a full O2 cylinder to the rear panel yoke.
3. Open the O2 cylinder until its pressure gauge indicates cylinder pressure.
4. Close the O2 cylinder.
5. The O2 cylinder pressure gauge should not drop more than 10% of its initial pressure over 1
minute.

92
Aeon8800A Service Manual Checkout

3.5.3. AIR Cylinder Leak Test

1. Remove the AIR line pressure hose form the line pressure inlet on Aeon8800A.
2. Mount a full AIR cylinder to the rear panel yoke.
3. Open the AIR cylinder until its pressure gauge indicates cylinder pressure.
4. Close the AIR cylinder.
5. The AIR cylinder pressure gauge should not drop more than 10% of its initial pressure over 1
minute.

3.5.4. Line Supply Check Valve Test

1. Connect and open a full gas cylinder on each yoke.


2. Verify the cylinder gauges operate.
3. Remove each line pressure hose one at a time from the wall supply inlet.
4. Verify no gas is escaping from any of the three central supply high pressure hoses connected
to Aeon8800A using a digital flow meter.
5. Verify there is no external damage to the gas supply inlet coupling on Aeon8800A.
6. Reattach each central supply gas hose to the wall supply inlets.

3.5.5. N2O Line Pressure Leak Test

1. Remove the N2O cylinder from Aeon8800A.


2. Connect the N2O line pressure hose to the line pressure inlet on Aeon8800A.
3. Pinch the N2O line pressure hose to stop AIR line flow.
4. Remove the N2O line pressure hose from the N2O line source.
5. The N2O line pressure gauge should not fall more than 2psi in 20 seconds.
6. Reconnect the N2O line pressure and remove the pinch in the hose.

3.5.6. O2 Line Pressure Leak Test

1. Remove the O2 cylinder from Aeon8800A.


2. Connect the O2 line pressure hose to the line pressure inlet on Aeon8800A.
3. Pinch the O2 line pressure hose to stop AIR line flow.
4. Remove the O2 line pressure hose from the O2 line source.
5. The O2 line pressure gauge should not fall more than 2psi in 20 seconds.
6. Reconnect the O2 line pressure and remove the pinch in the hose.

93
Aeon8800A Service Manual Checkout

3.5.7. AIR Line Pressure Leak Test

1. Remove the AIR cylinder from Aeon8800A.


2. Connect the AIR line pressure hose to the line pressure inlet on Aeon8800A.
3. Pinch the AIR line pressure hose to stop AIR line flow.
4. Remove the AIR line pressure hose from the AIR line source.
5. The AIR line pressure gauge should not fall more than 2psi in 20 seconds.
6. Reconnect the AIR line pressure and remove the pinch in the hose.

3.5.8. Cylinder Supply Check Valve Test

1. Close and remove all gas cylinders from Aeon8800A.


2. Verify the cylinder gauges return to zero.
3. Using the digital flow meter, verify that no gas is escaping from any of the three yoke
connections on Aeon8800A.
4. Verify there is no external damage to the gas cylinder or yoke pins.
5. Reconnect the cylinders as necessary.

3.6. Breathing System Checks

3.6.1. Internal Gas Connections Test

1. Close and remove all gas cylinders from Aeon8800A.


2. Connect only the AIR line pressure hose to Aeon8800A from the wall supply. Leave all other
line pressure hoses disconnected.
3. With Aeon8800A powered OFF, rotate the AIR flow control knob to ensure a continuous flow
increase throughout its full range.
4. Disconnect the AIR line pressure hose form Aeon8800A, and connect the O2 line pressure
hose from the wall supply.
5. Fully rotate the N2O flow control knob and verify that there is not flow.
6. Fully rotate the AIR flow control knob and verify that there is no flow.

94
Aeon8800A Service Manual Checkout

3.6.2. Drive Gas Pressure Loss Alarm, N2O Cut off Valve Test

1. Set the O2 flow to 2L/min using the flow control valve.


2. Set the N2O flow to 2L/min using the flow control valve.
3. Set the AIR flow to 2L/min using the flow control valve.
4. Interrupt the O2 supply to Aeon8800A.
5. Verify that the flow of N2O and O2 stops within 2 minutes and that the flow of AIR (if available)
continues to flow at 2L/min.
6. Verify the following alarms are activated:
 O2 Supply Failure appears on the screen
 An alarm tone sounds

3.7. Vaporizers

3.7.1. Vaporizer Interlock Test

1. Attach two vaporizers to the vaporizer mounting manifold and lock them in place.
2. Rotate either of the vaporizer dial to 3%.
3. Verify that the other vaporizer dial cannot be rotated to a setting.
4. Set both vaporizer dials to 0.
5. Rotate the other vaporizer dial to 3%.
6. Verify that the first vaporizer dial cannot be rotated.
7. Rotate both vaporizer dials to T and remove both vaporizers.
8. Verify that the locking spring is intact.
9. Reconnect both vaporizers to the vaporizer mounting manifold.

3.7.2. Vaporizer Leak Test

CAUTION Verify the system has passed the leak test without the vaporizers prior to this test.

1. Verify that vaporizers are mounted correctly.


2. Ensure that the gas pressure for O2, N2O and AIR are at 50±10psi.
3. Power ON Aeon8800A.
4. Attach a reusable-silicone –rubber breathing circuit to the Breathing System.
5. Tightly connect the Y-fitting on the breathing circuit to the test port.
6. Attach a breathing bag to the bag arm.

95
Aeon8800A Service Manual Checkout

7. Set the APL valve to the fully closed position (70cmH2O).


8. Rotate the O2 flow control valve until 50cmH2O pressure is observed on the airway pressure
gauge.
9. Verify that the necessary flow to stabilize the pressure is less than 300mL/min.

3.8. Performance Verification

3.8.1. STANDBY Mode Ventilation Test

NOTE For testing purposes always use a reusable breathing circuit.

1. Ensure that the gas pressure for O2, N2O, and AIR are at 50±10psi.
2. Power ON Aeon8800A.
3. Perform the start up tests per the on-screen instructions. Ensure successful completion.
4. Attach a breathing circuit and test lung to the Y-fitting of the breathing circuit.
5. Set the APL valve to 15cmH2O.
6. Set the AIR flow to 5L/min using the flow control valve.
7. Squeeze the breathing bag once every 10 seconds to inflate and deflate the test lung to
approximately 20cmH2O of pressure.
8. Verify the inflation and deflation of test lung.

3.8.2. Manual Mode Ventilation Test

1. Set the ventilation mode to MANUAL.


2. Set the APL valve to approximately 25cmH2O. Push the O2 flush button to fill the breathing
bag.
3. Set the AIR flow to 1L/min using the flow control valve.
4. Squeeze the breathing bag once every 3 seconds.
5. Verify the inflation and deflation of the test lung.
6. Verify that an airway pressure waveform and all numeric values appear on screen during bag
compressions.
7. Stop squeezing the breathing bag and set the APL valve to the open position.

96
Aeon8800A Service Manual Checkout

3.8.3. APNEA Alarm Test

1. While in the manual ventilation mode, stop ventilating the test lung.
2. Verify that the following APNEA alarm signals activates at approximately 60 seconds from the
last bag compression.
 APNEA appears on the screen.
 An alarm tone sounds.

3.8.4. Alarm MUTE Test

1. While the APNEA alarm is sounding, press the MUTE key.


2. Verify the audio portion of the alarm stops and resumes after 2 minutes.

3.8.5. IPPV Ventilation Mode Test

3.8.5.1. IPPV Adult Ventilation Mode Test

CAUTION For testing purposes always use a reusable breathing circuit.

1. Attach a breathing circuit and breathing bag.


2. Attach an adult test lung to the Y-fitting of the breathing circuit.
3. Attach a vent tester between the expiratory port and the expiratory hose.
4. Set the O2 flow to 2L/min and set the N2O and AIR flow rates to minimum flow.
5. Set the ventilation controls to:

Ventilator Controls Ventilator Settings


Patient Type Adult
Ventilation Mode IPPV
Tidal Volume-VT 600
Breath Rate-freq 8
I:E Ratio-I:E 1:2
Plateau -Tp 10
PEEP OFF
Pmax 40

97
Aeon8800A Service Manual Checkout

6. Select IPPV again to begin ventilation.


7. Verify that the pressure waveform、Tidal Volume、Mean or Plateau Pressure、 Resp. rate and
minute volume vales appear on the screen.
8. Verify the Tidal Volume display is within 15% of the set value within approximately 1 minute
from the start of ventilation.
9. Verify the Tidal Volume display is within 15% of the delivered volume measured with the vent
tester within approximately 1 minute from the start of ventilation.
10. Verify the measured O2 concentration is at least 97% after 5 minutes.
11. Set the AIR flow to 3L/min and set the N2O and O2 flow rates to minimum flow.
12. Verify the measured O2 concentration is 21% ±3% vol.% after 5 minutes.

98
Aeon8800A Service Manual Checkout

3.8.5.2. IPPV Child Ventilation Mode Test

CAUTION: Limit the volume in the test lung to provide sufficient airway pressure to satisfy the

Low Peak Pressure alarm. Or reduce the Peak Pressure alarm limit to a lower
value to prevent the alarm when using an adult test lung.
1. Attach a breathing circuit and breathing bag.
2. Attach an adult test lung to the Y-fitting of the breathing circuit.
3. Attach a vent tester between the expiratory port and expiratory hose.
4. Set the O2 flow to 2L/min and set the N2O and AIR flow rates to minimum flow.
5. Set the ventilation controls to:

Ventilator Controls Ventilator Settings


Patient Type Child
Ventilation Mode IPPV
Tidal Volume-VT 120
Breath Rate-freq 20
I:E Ratio-I:E 1:2
Plateau-Tp 10
PEEP OFF
Pmax 40

6. Select IPPV to begin ventilation.


7. Verify that the pressure waveform, Tidal Volume, Mean Pressure, Resp. rate and minute
volume values appear on the screen.
8. Verify the Tidal Volume display is within 25mL or±15% of the delivered volume (whichever is
greater) measured with the vent tester within approximately 1 minute from the start of
ventilation.
9. Verify the delivered volume as measured by a vent tester at the expiratory port, is within 17%
of the Tidal Volume set value within approximately 1 minute from the start of ventilation.

99
Aeon8800A Service Manual Checkout

3.8.6. Airway Disconnect Alarm Test

1. While the ventilator is running, disconnect the expiratory limb from the expiratory port on the
Breathing System.
2. Verify the following airway pressure disconnect alarm signals activate:
 APNEA message appears on the screen.
 Low Airway Pressure message appears on the screen.
 An alarm tone sounds.

3.8.7. PCV Adult Ventilation Mode Test

CAUTION For testing purposes always use a reusable breathing circuit.

1. Attach a breathing circuit and breathing bag.


2. Attach an adult test lung to the Y-fitting of the breathing circuit.
3. Attach a vent tester between the expiratory port and the expiratory hose.
4. Set the O2 flow to 3L/min and set the N2O and AIR flow rates to minimum flow.
5. Set the ventilator controls to:

Ventilator Controls Ventilator Setting


Patient Type Adult
Ventilation Mode PCV
Target Pressure -PTARGET 20
Breath Rate -freq 8
I:E Ratio-I:E 1:2
PEEP OFF
Inspiratory Slope-Tslope 0.5

6. Select PCV to begin ventilation.


7. Verify the Peak Pressure reading of the display is within ±4cmH2O of the Peak Pressure
measured with the vent tester.
8. Verify that the pressure waveform Tidal Volume, Resp. Rate and minute volume values
appear on the screen.
9. Verify that the Peak Value reaches 20±4cmH2O within five breaths from the start of
ventilation.

100
Aeon8800A Service Manual Checkout

3.8.8. Pressure Support (PS) Ventilation Mode Test

CAUTION For testing purposes always use a reusable breathing circuit.

1. Attach a breathing circuit and breathing bag.


2. Attach an adult test lung to the Y-fitting of the breathing circuit.
3. Attach a vent tester between the expiratory port and the expiratory hose.
4. Set the O2 flow to 1L/min and set the N2O and AIR flow rates to minimum flow.
5. Set the ventilator controls to:

Ventilator Controls Ventilator Setting


Patient Type Adult
Ventilation Mode PS
Differential Pressure -ΔP 5
PEEP OFF
Flow Trigger -Trigger 3
Inspiratory Slope-Tslope 0.5
Minimum frequency -freqMIN 4

6. Set PS to begin ventilation.


7. Begin triggering breath by slightly squeezing the test lung and releasing. Maintain a
continuous breath rate.
8. Verify that a pressure waveform and all ventilation parameters appear on the screen.
9. Verify that the Peak Pressure reading on the display is ±4 the value of ΔP+PEEP.
10. Stop triggering breaths.
11. Verify that after 30 the ventilator delivers a breath and displays the message APNEA
BACKUP.
12. Verify the system ventilates with a frequency of 2bpm.

101
Aeon8800A Service Manual Checkout

3.9. Alarms and Failsafe Functions

3.9.1. Set Up

CAUTION For testing purposes always use a reusable breathing circuit.

1. Ensure that the gas pressure for O2, N2O and AIR are at 50±10psi.
2. Power ON Aeon8800A.
3. Perform the Startup Tests per the on-screen instructions. Ensure successful completion.
4. Attach a breathing circuit and breathing bag.
5. Attach an adult test lung to the Y-fitting of the breathing circuit.
6. Set the O2 flow to 2L/min and set the N2O and AIR flow rates to minimum flow.
7. Set the ventilator controls to:

Ventilator Controls Ventilator Setting


Patient Type Adult
Ventilation Mode IPPV
Tidal Volume -VT 600
Breath Rate -freq 8
I:E Ratio -I:E 1:2
Plateau -Tp 10
PEEP OFF
Pmax 40

8. Select IPPV to begin ventilation.

3.9.2. Low FiO2 Alarm Test

1. Set the flow FiO2 Alarm limit to 50% O2.


2. Set the AIR flow control valve to 5L/min.
3. Set the FiO2 flow controller to minimum flow.
4. Verify the following Low FiO2 alarm signals activate, within three ventilation cycles:
 Low FiO2 message appears on the screen.
 An alarm tone sounds.
5. Set the Low FiO2 alarm limit to 18%.
6. Verify the alarm signals cease.

102
Aeon8800A Service Manual Checkout

3.9.3. High FiO2 Alarm Test

1. Set the high FiO2 Alarm limit to 49% O2.


2. Set the FiO2 flow control valve to 5L/min.
3. Set the AIR flow controller to minimum.
4. Verify the following High FiO2 alarm signals activate:
 High FiO2 message appears on the screen.
 An alarm tone sounds.
5. Set the high FiO2 alarm limit to the max setting.
6. Verify the alarm signals cease.

103
Aeon8800A Service Manual Checkout

3.9.3.1. Peak Pressure Alarms Test

1. Set the PAW Low alarm to the lowest setting.


2. Set the PAW High alarm limit set point about 5 to 8 digits below the Peak Pressure displayed
on the upper left of the screen.
3. Verify the following (high) peak pressure alarms activate:
 High Airway Pressure message appears on the screen.
 An alarm tone sounds.
 Inspiration ends and expiration begins as the pressure meets the high alarm limit.
4. Set the PAW High alarm limit set point to 65cmH2O.
5. Verify the alarms signals cease.
6. Set the PAW Low alarm limit set point to 50cmH2O.
7. Verify the following (low) peak pressure alarms activate:
 Low Airway Pressure message appears on the screen.
 An alarm tone sounds.
8. Set the PAW Low alarm limit to 12cmH2O.
9. Verify the alarm signals cease.

3.9.3.2. Minute Volume Alarm Test

1. Set the MV Low alarm limit set point to the highest value.
2. Verify the following alarms activate:
 Low MV message appears on the screen.
 An alarm tone sounds.
3. Set the MV Low alarm limit to minimum setting.
4. Verify the alarm signals cease.
5. Set the MV High alarm set point to the lowest value.
6. Verify the following alarms activate:
 High MV message appears on the screen.
 An alarm tone sounds.
7. Set the MV High alarm limit set point to the highest value.
8. Verify that the alarm signals cease.

104
Aeon8800A Service Manual Checkout

3.10. Miscellaneous Tests

3.10.1. Test the Line Voltage Alarm

1. Before starting this test, verify that the battery is fully charged. The battery icon in the upper
right corner of the screen is solid when the battery is fully charged.
2. Interrupt AC line voltage.
3. Verify that the following alarms activate:
 An alarm tone sounds.
 AC Power Failure message appears on the screen.
4. Verify that a fully charged battery operates the ventilator for a minimum of 45 minutes.
5. Plug Aeon8800A into AC line voltage.
6. Verify that the alarm signals cease.
7. Verify the presence of the battery charging icon in the upper right corner of the screen (status
bar moving from right to left).

3.10.2. Wheel Brakes Test

1. Verify that each from wheel brake operates.

3.10.3. Work Light Test

1. Turn on the work light located on the bottom side of the top panel.
2. Verify that it lights in both on positions.

3.10.4. Auxiliary Flow Meter

1. Verify an AIR flow of 15L/min can be obtained by connecting the auxiliary AIR hose to the
pressure source and opening the flow meter.
2. Verify an O2 flow of 15L/min can be obtained by connecting the auxiliary O2 hose to the
pressure source an opening the flow meter.

105
Aeon8800A Service Manual Checkout

106
Aeon8800A Service Manual Configuration menu

4. Configuration menu
This section describes the Service level function of the anesthesia machine.
The Configuration menu has three parts which are password protected, including “Vent Defaults”,
“Alarm Defaults” and “Service Mode”.

CAUTION: Configuration button is effective only in standby mode.

Click on “ ”menu on the right column of the screen, the correspond sub-menu appears
on the screen

Fig.4-1

4.1. Vent Defaults


Vent Defaults submenu is for super-user and service engineer.

When the “Configuration” button is


selected, the Vent Defaults interface
appears.

Fig.4-2

107
Aeon8800A Service Manual Configuration menu

Enter the password “2020”and click on


“ ”, the interface shown as the picture
on the right appears.
If enter wrong password, click on “ ”
to clear the entered number.
There are four ventilation modes in the
“Vent Defaults” tab, including IPPV,
PCV, SIMV and PS.
Fig.4-3
Select one ventilation mode, the
interface shown as the picture on the
right appears.
There are two sections of ventilation
parameters in the menu, one for adult
and the other for child.
The user can set the default parameters
for each ventilation mode.
Settings entered will be used as the Fig.4-4
default settings after unit power up.
"Load Factory Settings" button when
pressed , will change the default
settings to the factory defaults

108
Aeon8800A Service Manual Configuration menu

4.2. Alarm Defaults


Vent Defaults submenu is for super-user and service engineer.

Select the Alarm Defaults sub-menu,


the interface shown as the picture on
the right appears.

Fig.4-5
Enter the password “2020” and click on
“ ”, the interface shown as the picture
on the right appears
If enter wrong password, click on “ ”
to clear the entered number.

There are four alarm limits which can be


set as needed in the “Alarm Defaults”
tab, including Adult Vent Alarm Limits, Fig.4-6
Child Vent Alarm Limits, Gas Alarm
Limits and Agent Alarm Limits.

Settings entered will be used as the


default settings after unit power
up."Load Factory Settings" button when
pressed, will change the default settings
to the factory defaults.

109
Aeon8800A Service Manual Configuration menu

4.3. Service Mode


Service Mode submenu is only for the service engineer.
Use the service-level password “2010” to access the Service menu, which including “Calibration”,
“SW Update”, “Optional Functions”, “Optional Language” and “Setup”.

Select the Service Mode sub-menu, the


interface shown as the picture on the
right appears.

Fig.4-7
Enter the right password and click on
“ ”, the interface shown as the picture
on the right appears.

If enter wrong password, click on “ ”


to clear the entered number.

Fig.4-8

110
Aeon8800A Service Manual Configuration menu

4.3.1. Calibration

The calibration submenu including Flow Valve, Flow Sensor, Pressure Sensor, PEEP Valve
and Cal Data. Select “Calibration” submenu, the screen displays as bellow picture.

Fig.4-9

111
Aeon8800A Service Manual Configuration menu

4.3.1.1. Flow Valve

Remove Bellows, Set flowmeters to 0


Restore the cover

Remove I and E valve disks,


Disconnect hose from
reinstall valve covers
Insp. port

Seal Exp. ports

Fig.4-10 Flow valve calibration Diagram

Select “Calibration” -- “Flow valve” to


enter the flow valve calibration screen.

Select “start” to begin flow valve


calibration, and select “ ” to return to
Calibration menu.

Fig.4-11

112
Aeon8800A Service Manual Configuration menu

Do the procedure according to the


screen displayed message and “Flow
valve calibration Diagram”.

Then select “start” the screen shown as


the picture on the right appears.

Fig.4-12

Select “start” to begin flow valve


calibration, the system calibrating
screen shows as the picture on the right.

Fig.4-13

If calibration success, the screen shown


as the picture on the right appears
Select “ ” back to Calibration menu.

Fig.4-14

113
Aeon8800A Service Manual Configuration menu

If calibration failure, the screen shown


as the picture on the right appears

Select “Repeat” to do the flow valve


calibration again.

Select “ ” back to Calibration menu.

Fig.4-15

114
Aeon8800A Service Manual Configuration menu

4.3.1.2. Flow Sensor

Remove Bellows,
Restore the cover

Close flow meter

Remove I and E valve


disks, reinstall valve covers
Flow Monitor
(TSI)

Seal Exp. ports

Flow Regulator
Fig.4-16 Inspiratory flow sensor calibration diagram

1. Connect one tube of flow regulator to Auxiliary O2 port.


2. Connect the other tube of flow regulator to one end of drive gas tube, of which the other end is
connected to breathing system bottom.
3. Connect flow monitor (TSI) to Insp. Port

NOTE: Before connecting TSI to anesthesia machine, please set “Gas” of TSI to “Oxygen”, and
set “Flow Units” of TSI to “Volumetric L/min”. The TSI window have been set is shown as
following picture.

115
Aeon8800A Service Manual Configuration menu

Fig.4-17 TSI Window

Select “Calibration” -- “Flow Sensor”


to enter the flow sensor calibration
screen.

Select “start” to choose the flow sensor


to be calibrated, and select “ ” to
return to Calibration menu.

Fig.4-18
Select “start”, the screen shown as the
picture on the right appears.

There are two bars on the screen, one


for inspiration sensor and the other for
expiration sensor.

Select one sensor to calibrate.


Fig.4-19
Select “ ” to return to Calibration
menu.

116
Aeon8800A Service Manual Configuration menu

Do the procedure according to the


screen displayed message and
“Inspiratory sensor calibration diagram”.

Then select “Start” the system enters


into the calibration interface.

Fig.4-20
Select “Start”, the screen shown as the
picture on the right appears.

Calibrate 0, 3, 12 and 30 in turns.

If the calibration success, the calibrated


character is grey, other while the
calibrated character is black.
Fig.4-21
Select “Start”, the system begins the
total calibration procedure. Select “ ”
to return to Calibration menu.
Regulate the throttle valve to make the
flow to 0L/min and then press “0” button.
The system begins 0 L/min calibration.

If the 0 L/min calibration success, the


interface shown as the picture on the
right appears.

Regulate the throttle valve to make the Fig.4-22


flow to other values and then press the
corresponding button to do other
calibrations.

117
Aeon8800A Service Manual Configuration menu

If the 0 L/min calibration failure, the


interface shown as the picture on the
right appears.

Press “0” button to do 0 L/min


calibration again.

Select “Start”, the system begins the


total calibration procedure. Fig.4-23

Select “ ” to return to Calibration


menu.
Select “Start”, the screen shown as the
picture on the right appears.

Select “ ” to return to Calibration


menu.

Fig.4-24
If all the calibrations of 0 , 3 , 12 ,
30L/min success, the total calibration is
successful and the interface shown as
the picture on the right will appear.

Select “ ” to return to Calibration


menu.
Fig.4-25

118
Aeon8800A Service Manual Configuration menu

If one of 0, 3, 12, 30calibration failure,


the total calibration will fail and the
interface shown as the picture on the
right will appear.

Select “Repeat” the system goes back


to the flow sensor initialization
calibration interface.
Fig.4-26
Select “ ” to return to Calibration
menu.

119
Aeon8800A Service Manual Configuration menu

Expiratory flow sensor calibration


Close flow meter

Remove Bellows,
And then restore
the cover

Seal Ins. ports

Remove I and E valve Connect flow monitor to


disks, reinstall valve Exp. Port
covers

Tube connector
opens to AIR

Fig.4-27 The flow monitor connector position

1. Connect one tube of flow regulator to Auxiliary O2 port.


2. Connect the other tube of flow regulator to flow monitor (TSI).
3. Connect flow monitor (TSI) to Exp. Port
4. Connect a tube longer than 1m to the drive gas port at the bottom of breathing system, and
make the other connector of tube open to AIR

NOTE: Before connecting TSI to anesthesia machine, please set “Gas” of TSI to “Oxygen”, and
set “Flow Units” of TSI to “Volumetric L/min”. The TSI window have been set is shown as
following picture.

120
Aeon8800A Service Manual Configuration menu

Fig. 4-28 TSI Window

Select “Calibration” -- “Flow Sensor” to


enter the flow sensor calibration screen.

Select “Start” to choose the flow sensor to


be calibrated, and select “ ” to return to
Calibration menu.

Fig.4-29

Select “Start”, the screen shown as the


picture on the right appears.

There are two bars on the screen, one for


inspiration sensor and the other for
expiration sensor

Select expiration sensor to calibrate.


Fig.4-30
Select “ ” to return to Calibration menu.

121
Aeon8800A Service Manual Configuration menu

Do the procedure according to the screen


displayed message and “Expiratory sensor
calibration diagram”.

Then select “Start” the system enters into


the calibration interface.

Fig.4-31
Select “Start”, the screen shown as the
picture on the right appears.

Calibrate 0, 3, 12 and 30 in turns.

If the calibration success, the calibrated


character is grey, other while the calibrated
character is black.
Fig.4-32
Select “Start”, the system begins the total
calibration procedure. Select “ ” to return
to Calibration menu.

Regulate the throttle valve to make the flow


to 0L/min and then press “0” button.
The system begins 0 L/min calibration.

If the 0 L/min calibration success, the


interface shown as the picture on the right
appears.

Regulate the throttle valve to make the flow Fig.4-33


to other values and then press the
corresponding button to do other
calibrations.

122
Aeon8800A Service Manual Configuration menu

If the 0 L/min calibration failure, the


interface shown as the picture on the right
appears.

Press “0” button to do 0 L/min calibration


again.

Select “Start”, the system begins the total


calibration procedure. Fig.4-34

Select “ ” to return to Calibration menu.

Select “Start”, the system begins the total


calibration procedure and the screen is
shown on the right.

Select “ ” to return to Calibration menu.

Fig.4-35

If all the calibrations of 0 , 3 , 12 , 30


success, the total calibration is successful
and the interface shown as the picture on
the right will appear.

Select “ ” to return to Calibration menu.

Fig.4-36

123
Aeon8800A Service Manual Configuration menu

If one of 0,3,12,30calibration failure,


the total calibration will fail and the
interface shown as the picture on the right
will appear.

Select “Repeat” the system goes back to


the flow sensor initialization calibration
interface.
Fig.4-37
Select “ ” to return to Calibration menu.

124
Aeon8800A Service Manual Configuration menu

4.3.1.3. Pressure Sensor

Set APL valve to 0


Close flow meter

Disconnect any hoses


from the inspiratory and
expiratory ports

Fig.4-38

Select “Calibration” -- “Pressure


Sensor” to enter the pressure sensor
calibration screen.

Select “start” to begin flow valve


calibration, and select “ ” to return to
Calibration menu.

Fig.4-39

125
Aeon8800A Service Manual Configuration menu

Do the procedure according to the


screen displayed message and
“Pressure sensor calibration Diagram”.

Then select “start”, the screen shown


as the picture on the right appears.

Fig.4-40

If calibration success, the screen shown


as the picture on the right appears
Select “ ” back to Calibration menu.

Fig.4-41

If calibration failure, the screen shown


as the picture on the right appears

Select “Repeat” to do the flow valve


calibration again.

Select “ ” back to Calibration menu.

Fig.4-42

126
Aeon8800A Service Manual Configuration menu

4.3.1.4. PEEP Valve

Set Fresh Gas


Connect the breathing Flow to 1.0L/min
circuit and connect a
1L bag to Y-piece
Push “O2+” to full fill
the Bellows

Fig.4-43

Select “Calibration” -- “PEEP valve” to


enter the PEEP valve calibration screen.

Select “start” to begin flow valve


calibration, and select “ ” to return to
Calibration menu.

Fig.4-44

127
Aeon8800A Service Manual Configuration menu

Do the procedure according to the


screen displayed message and “PEEP
valve calibration Diagram”.

Then select “start”, the screen shown


as the picture on the right appears.

Fig.4-45

If calibration success, the screen shown


as the picture on the right appears
Select “ ” back to Calibration menu.

Fig.4-46

If calibration failure, the screen shown


as the picture on the right appears

Select “Repeat” to do the flow valve


calibration again

Select “ ” back to Calibration menu.

Fig.4-47

128
Aeon8800A Service Manual Configuration menu

4.3.1.5. Cal Data

Select “Calibration” -- “Cal Data” to


enter the Cal Data screen.

The screen has five items; each item


has some corresponding monitoring
values which are just for reference.

Select“ ” to return to Calibration


menu. Fig.4-48

4.3.2. SW Update

Select “Service Mode” -- “SW Update”


to open the interface on the right.

Do according to the promote message


on the screen.

The “Start” button is invalid.


Select“ ” to return to Service Mode
menu. Fig.4-49

129
Aeon8800A Service Manual Configuration menu

Insert the USB in which has the newest


software of anesthesia machine.

After connecting USB to machine the


display shows as the right picture.

Select “Start” to begin SW Update


procedure,
Select “ ”to close the sub-menu Fig.4-50

After selecting “Start” the software


update interface covers the full screen.
The server can update all software.

When doing the update do not break the


update process.

Select “Update All Components” to


repeat the update if the update process
is broken abnormally.
Fig.4-51
The interface displays “Success!” or
“Failure!” in the running bar area after
updating.

The success color is green while the


failure is red.

Fig.4-52

130
Aeon8800A Service Manual Configuration menu

The message “Re-cycle to implement


the update software” appears after
updating software. The server should
re-cycle the machine to perform the GUI
in the Update Software.

Fig.4-53

131
Aeon8800A Service Manual Configuration menu

4.3.3. Optional Functions

Select “Service Mode” -- “Optional


Function” to open the interface on the
right.

“ ” indicates that machine has this

function.

“ ” indicates that machine has no this

function. Fig.4-54

Choose the needed function and press


“Activate function” to achieve the
needed functions.

Select“ ” to return to Service Mode


menu.
Press “Activate function” to open the
interface on the right.

Remember the ID NO.:


NEGD-I68Q-JTNL-GTID-RV produced
by the machine according to its current
configuration.

Send the ID NO. and required function Fig.4-55


to manufacturer to ask for the register
number.

132
Aeon8800A Service Manual Configuration menu

Enter the register number from Function


Configuration Software into the white
blank on the machine interface shown
as right picture.

Press” ”, if the register number

is correct, the interface as the right


picture appears.
Fig.4-56

The message “Application new


configure after reboot” appears on the
screen.

If the register number is wrong, the


interface as the right picture appears.

The message “The password is


incorrect” appears on the screen.

Fig.4-57

133
Aeon8800A Service Manual Configuration menu

4.3.4. Optional Language

Optional Language sub-menu has


several languages.

Before each language there is a□, Click


on “□”to select the language or not.

Shows the language has been

selected.

Fig.4-58

Select “Save” to choose the needed


language.

Select “ ” back to Service Mode


menu.

Fig.4-59

134
Aeon8800A Service Manual Configuration menu

4.3.5. Setup

Select “Setup” to enter the setup


screen.

There are three items in the setup


screen, including Demo ,Flow Meters
and Gas Standard

Demo only has an “OFF” menu.


Fig.4-60

Flow Meters have three items, US, CE


and CN.

The gray spot following the item shows


the current type of flow meter.

Fig.4-61

Gas Standard has two items, ATP and


BTPS.

The gray spot following the item shows


the current type of gas standard.

Fig.4-62

135
Aeon8800A Service Manual Configuration menu

136
Aeon8800A Service Manual Troubleshooting

5. Troubleshooting

5.1. General Troubleshooting


The following table shows symptoms and corrective actions for problems when troubleshooting Aeon8800A. The information given indicates failures
during startup and runtime.

137
Aeon8800A Service Manual Troubleshooting

Trouble Occurrence Phase Cause Analysis Solutions


1. The power supply switch is turned to ON, 1.Turn the power supply switch to OFF and power on again
immediately after be turned to OFF, which after waiting for 1~2seconds.
makes the time too short for the BDU control
board power to reboot.
BDU 2. The data wire is disconnected or 2. Check whether the data wire (signal wire at backside of
Communication Startup test damaged. display) is disconnected or is damaged.
FAIL 3. The MCB is defective. 3. Check whether D10, D11 and D12 on MCB are flashing and
D8 is lighting, if not, there is something wrong with MCB.
Replace the MCB and perform each item of calibration and
startup test again.
4. MCB crashes or resets. 4. Turn off the power supply and power on again.

138
Aeon8800A Service Manual Troubleshooting

Trouble Occurrence Phase Cause Analysis Solutions


1. The power supply switch was turned ON, 1. Turn the power supply switch to OFF, wait 1~2 seconds,
and OFF quickly. then power on again.

2.GUI、BDU and Keyboard are incompatible. 2. Update software version. Enter SERVICE manual to check
out the real software version. If the version is 00.00, turn the
power supply switch to OFF, and power on again.

3. The data wire is disconnected or 3. Check whether the data wire (signal wire at backside of
damaged. display) is disconnected or is damaged.

4. MCB is defective. 4. Check whether D10, D11 and D12 on MCB are flashing and
Software
Startup test D8 is lighting, if not, there is something wrong with MCB.
Version FAIL
Replace the MCB and perform each item of calibration and
startup test again.
5. The keyboard board is defective. 5. Check whether keyboard indicator is flashing when
keyboard software is not suited to machine, There is
something wrong with keyboard if indicator is not flashing.
Replace the user interface assembly.
6. PC104 board is defective. 6. If all the software communication fail while keyboard and
MCB works well; Replace PC104 board with the compatible
software version.

139
Aeon8800A Service Manual Troubleshooting

Trouble Occurrence Phase Cause Analysis Solutions


1. Data stored in EEPROM on MCB is lost 1. Startup, and then perform each item of the calibration and
BDU EEPROM
Startup test (checksum error). startup test again.
Data FAIL
2. MCB is defective. 2. Replace MCB.
1. The wire that connects to the vent/manual 1. Check whether the connection wire of the vent/manual valve
valve is disconnected or is damaged. is disconnected or is damaged.
2. The vent/manual solenoid valve is 2. Measure the resistance of solenoid valve, if the resistance is
damaged. not within the range of 200~400Ω, the solenoid valve is fail.
Replace the vent/manual solenoid valve.

Vent/Manual 3. MCB, interface board power supply cable 3. Check whether the connection of cable is correct with a
Startup test
Valve Failure failure multimeter. Or measure the voltage of MCB yellow cable to
earth. If the voltage is not beyond 22-25V, the cable is
defective.
4. MCB is defective 4. Check the connection cable and replace MCB. Then
perform each item of the calibration and startup tests again.

Keyboard 1. The data wire is disconnected or is


1. Check whether keyboard COM1 is correctly connected.
Communication Startup test damaged.

FAIL 2. The keyboard board is defective. 2. Replace keyboard.

140
Aeon8800A Service Manual Troubleshooting

Trouble Occurrence Phase Cause Analysis Solutions

1. The connection wire of the MCB is 1. Check whether the connection wire of the MCB is
disconnected or is damaged. disconnected or damaged.
Pressure
Startup test
Sensor FAIL
2. The connection of the MCB is defective. 2. Check the power supply. Replace the MCB.
3. The tubing kinked or occluded. 3. Check the tubing for kinks and remove any occlusions.
1. The O2 concentration is below 16%. 1. Increase O2 concentration.

2. The O2 sensor is out of range. 2. Replace the O2 sensor.


3. The main unit O2 sensor cable is 3. Check whether the main unit O2 sensor cable id
disconnected or damaged. disconnected or damaged.
O2Sensor FAIL Startup test
4. The MCB is defective. 4. Replace the MCB, and then perform each item of the
calibration and startup test again.
5. Weeping occurs on O2 sensor. 5. Replace the O2 sensor.
6. The O2 sensor is not compatible with the 6. Replace the MCB. And then perform each item of the
MCB. calibration and startup test again.

141
Aeon8800A Service Manual Troubleshooting

Trouble Occurrence Phase Cause Analysis Solutions


1. The connection wire of the inspiration 1. Check whether the PEEP valve is disconnected or
valve is disconnected or damaged. damaged.
2. The PEEP valve is defective. 2. Replace the PEEP valve.

3. The MCB is defective. 3. Replace the MCB and then perform each item of the
PEEP Valve
Startup test calibration and startup test again.
FAIL
4. The PEEP valve calibration fails. 4. Clean the check valves inside the Breathing System, and
then perform the flow sensor calibration and flow valve
calibration.
5. The PEEP valve is blocked. 5. Clean or replace the PEEP valve.

1. The connection wire of the inspiration 1. Check whether the inspiration valve s disconnected or
valve is disconnected or damaged. damaged.
Inspiration 2. The inspiration valve is defective. 2. Replace the inspiration valve.
Startup test
Valve FAIL
3. The MCB is defective. 3. Replace the MCB and then perform each item of the
calibration and startup test again.

142
Aeon8800A Service Manual Troubleshooting

Trouble Occurrence Phase Cause Analysis Solutions


1. The flow sensor does not reach the 1. Reinstall the absorber canister.
minimum value.
2. The AD value of flow sensor is incorrect. 2. Check whether the flow sensor is correctly connected.
Flow Sensor
Startup test
FAIL 3 The flow meter leaks. 3. Replace the flow meter.

4. MCB is defective 4. Replace MCB and then perform each item of the calibration
and startup test again.
1. The connection wire to the speaker is 1. Check whether the connection wire of the speaker is
disconnected or damaged. disconnected or damaged.
2. The speaker assembly is not connected. 2. The communication board is not pretty welded, weld the
Alarm
Startup test communication board again.
Speaker FAIL
4. The loop of the speaker is turned off. 4. Replace the speaker.
5. The communication board is defective. 5. Replace the communication board, and confirm the software
version is compatible with the new one.

143
Aeon8800A Service Manual Troubleshooting

Trouble Occurrence Phase Cause Analysis Solutions


1. The pneumatic tube is disconnected. 1. Check whether the pneumatic tube is connected or
damaged.
2. Gas cannot flow fluently in the Breathing 2. Adjust or replace the Breathing System.
During manual System.
leak test, the 3. The bellows moves downward. There are 3. Reduce the pressure on some ports, such as CGO port and
bellows moves Startup test leaks happened on some ports in the 1.75 pressure regulator.
when it Breathing System. The pressure of 1.75bar
shouldn’t. pressure regulator is so high that leaks
occur.
4. The bellows moves upward, the safety 4. Replace the safety valve.
valve leaks.
In leakage 1. The pressure gauge is defective. 1. Replace the pressure gauge.
detection, the
APL valve
detection 2. The Breathing System leaks in manual 2. Replace the APL valve or replace the Breathing System.

pressure mode.
Startup test
displayed on
the pressure
gauge does not
stay within
25~35cmH2O.

144
Aeon8800A Service Manual Troubleshooting

Trouble Occurrence Phase Cause Analysis Solutions


1. The proportional valve on the airway 1. Calibrate the inspiration valve again.
In leakage module fails.
detection, 2. The safety valve in the airway module 2. Replace the safety valve or the airway module.
Startup test
safety valve fails.
detection fails. 3. The local altitude is above 4000 meters. 3. Replace the application software and pressure sensor board
to adapt to the local altitude.
1. The tubing connection is not tight. 1. Check the connection on the tubing.
2. The absorber canister is not correctly 2. Check the installation of the absorber canister.
In leakage
connected.
detection, the
3. The pressure gauge or oxygen sensor on 3. Check the installation of the pressure gauge or oxygen
leakage Startup test
the circuit is not installed tightly. sensor on circuit.
volume is
4. There is a leakage in the fresh gas supply. 4. Check the fresh gas supply for leakage.
400mL~1L.
5. The bellows moves upward or downward. 5. Reduce the output pressure on circuit, or replace the safety
valve.
1. The tubing connection is not tight. 1. Check the connection on the tubing.
In leakage
2. The absorber canister is not correctly 2. Check the installation of the absorber canister.
detection, the
connected.
leakage Startup test
3. The pressure gauge or oxygen sensor on 3. Check the installation of the pressure gauge or oxygen
volume is
the circuit is not installed tightly. sensor on circuit.
larger than 1L.
4. There is a leakage in the fresh gas supply. 4. Check the fresh gas supply for leakage.

145
Aeon8800A Service Manual Troubleshooting

Trouble Occurrence Phase Cause Analysis Solutions


5. The gate valve on vaporizer mounting 5. Replace the gate valve.
manifold is defective.
1. The Y-fitting is not sealed or the Breathing 1. Check the connection and leakage of respiration tubes.
System has leakage.
2. The flow meter is set incorrectly. 2. Calibrate the flow meter again.
In compliance
3. The respiration tubes are too long. 3. The respiration tubes cannot be too long, replace the long
test,
Startup test tubes.
compliance test
4. The calibration of pressure sensor is 4. Calibrate the pressure sensor again.
fails.
inaccurate
5. The airway module is defective. 5. Replace the airway module.
6. The gas module is defective. 6. Replace the gas module.
1. Gas supply is not connected. 1. Check whether gas supply is connected or is switched on.
Observe the gas supply pressure gauge to find whether the
pressure is higher than 0.28Kpa.
2. The gas supply sensor connection cable is 2. Replace the gas supply cable.
Startup test
O2 supply fails. damaged.
/Runtime
3. The MCB is defective. 3. Connect gas supply and make the pressure is higher than
28Kpa. Then test the voltage of T63 and T58, if the valve is
lower than 0.83V, MCB is defective. Replace the MCB, and
perform each item of calibration and startup.

146
Aeon8800A Service Manual Troubleshooting

Trouble Occurrence Phase Cause Analysis Solutions


1. AC outlets are not connected.
AC Outlets are Startup test 1. Check whether AC outlets are connected correctly.
non functional /Runtime
2. A fuse is tripped. 2. Replace the fuse.
1. Battery capacity is low. 1. Connect the battery to AC power. Make sure the battery is
fully discharged and charged every 3 months.
Low Battery Runtime 2. The battery is not functional. 2. Replace the battery.
3. The connection cables or power board are 3. Check the connection cables or replace the power board.
defective.
1. Battery capacity is low. 1. Connect the battery to AC power. Make sure the battery is
fully discharged and charged every 3 months.
2. The connection cable of battery is not 2. Check whether the battery connection cable is not
The battery
connected or damaged. connected or damaged.
icon always
Runtime 3. The battery is not functional. 3. Replace the battery.
shows charging
4. The power board is defective or the 4. Replace the power board.
status.
surrounding environment is too high.
5. The cables on the power board are 5. Reconnect the cables correctly.
defective.

147
Aeon8800A Service Manual Troubleshooting

Trouble Occurrence Phase Cause Analysis Solutions


1. Mains outlet is not connected correctly. 1. Check whether mains outlet is connected correctly.
2. The connection cables of the Breathing 2. Replace the connection cables of the Breathing System.
System are defective.
The heating of 3. The heating control board is defective. 3. Replace the heating control board or the output cable.
the Breathing Runtime
System fails. 4. The circuit inside the breathing system is 4. Replace the breathing system.
defective.
5. The light on the heating control board 5. Replace the heating control board.
cannot be lighted or is always light.
The O2 reading
1. The O2 sensor calibration is incorrect, or
is out of Runtime 1. Calibrate the O2 sensor again or replace the O2 sensor.
the O2 sensor is defective.
tolerance.
The pressure
1. The pressure sensor calibration is
reading is out Runtime 1. Calibrate the pressure sensor again.
incorrect.
of tolerance.
1. PEEP valve calibration is incorrect. 1. Calibrate PEEP valve again.
PEEP reading
2. Pressure sensor calibration is incorrect. 2. Calibrate the pressure sensor again.
is out of Runtime
3. PEEP value is above the normal level. 3. Calibrate the PEEP valve again.
tolerance.
4. PEEP valve is not tightly connected. 4. Tighten the screw at the bottom of PEEP valve.

148
Aeon8800A Service Manual Troubleshooting

Trouble Occurrence Phase Cause Analysis Solutions


1. The compliance test isn't accurate after 1. Perform the compliance test again.
start-up.
2. The supply gas pressure is too low, so the 2. Advance the supply gas pressure.
VT is inaccurate.
3. The inspiratory VT is too large or the 3. Examine the check valve for one direction character, clean
expiratory VT is too small; The disk in check the check valve disk.
becomes sticky, cleaning is needed.
4. The bellows does not move steadily. 3. Heating the bellows or replace the bellows.
5. The inspiratory VT is too small or the 5. Replace or reconnect the pneumatic hose, settle the
The Vt reading
expiratory VT is too large; When the airway pipeline to avoid it blocked; replacing O-ring of the connector to
is out of Runtime
pressure is 0 and 30cmH2O, the value of avoid the leakage.
tolerance.
Insp. /Exp. sensors in engineer mode AD
monitoring is far away from 0.27V.
6. The Expiratory wave on screen is smaller 6. If the water in the circuit, use MANUAL Mode, block the
or disappeared; it means there is some water outlets of circuit, disconnect the mainframe port of pneumatic
in the circuit sensor part or in the mainframe hose. Press the O2flush valve to blow the water out; If the
pipeline. water is in the pipeline, use injector assembly to fix this.
7. The VT is inaccurate. 7. Clean the block in the Insp. /Exp. Airway.
8. The calibration of flow sensor is 8. Recalibrate the flow sensor again.
inaccurate.

149
Aeon8800A Service Manual Troubleshooting

The Vt reading 9. Recalibrate the flow sensor again. 9. The calibration of pressure sensor is inaccurate.

is out of Runtime
tolerance. 10. Recalibrate the inspiratory valve again. 10. The calibration of inspiratory valve is inaccurate.

O2flush is
Runtime 11. Replace MCB. 11. MCB is defective.(No flow rate wave)
defective.

150
Aeon8800A Service Manual Troubleshooting

Trouble Occurrence Phase Cause Analysis Solutions

The power switch 1. The power switch or the connection cable is 1. Replace the power switch or the connection
on the main unit is All defective. cable.
non-functional. 2. The communication board is defective. 2. Replace the communication board.

All or some of the 1. The key board is defective. 1. Replace the key board.
keys are All 2. The communication board or COM 1 connection
2. Check and replace the communication board.
non-functional. cable is defective.
1. The encoder is defective. 1. Replace the encoder.
The encoder is
All 2. The communication board or COM 1 connection
non-responsive. 2. Check and replace the communication board.
cable is defective.
1. The work light circuit board or the cable is 1. Replace the work light circuit board or check
All or some of the defective. the wire connection.
work lights are All 2. The flow meter drive backlight is overloaded, which
2. Decrease the flow meter backlight overload.
non-functional. leads to trigger protection.
3. The work light switch is defective. 3. Replace the work light switch.
1. The power supply to the light is defective. 1. Replace the power supply to the light.
The flow meter 2. The foil of light is defective. 2. Replace the foil of light.
light is All 3. The inverter is defective. 3. Replace the inverter.
non-functional. 4. Replace the power board or replace the
4. The flow meter light is still light after power off.
communication board.

151
Aeon8800A Service Manual Troubleshooting

Trouble Occurrence Phase Cause Analysis Solutions


1. The maximum value of flow meter is far too low. 1. Adjust the throttle.
2. The flow meter leaks. 2. Replace the flow meter.
3. Adjust the proportional gear, adjust pin valve or
3. The ratio of O2 and N2O is inaccurate.
The flow meter is replace the flow meter.
All
defective. 4. The color of the flow meter back light is reversed. 4. Rotate the back light board.
5. The label of the rotation knob is damaged. 5. Replace the knob label.
6. Replace the O2 gas inlet, or adjust the O2flow to
6. There is basic flow in the flow meter.
0.
The Aux. AC outlet
All 1. The fuse is damaged. 1. Replace the fuse.
is non-functional.
1. The GUI program has crashed. 1. Reboot the system.
2. The signal cable is disconnected. 2. Check the signal cable.
3.PC104, back light power, communication board or 3. Check and replace PC104, back light power,
inverter is defective. communication board or inverter.
The display screen All
4. The display screen is damaged. 4. Replace the display screen.
is a solid color.
5. The missing or incorrect of CMOS setting data 5. Recover CMOS default setting by short-circuit
leads to main board inactivated. jumper JBAT1.
6. The connection cable of display screen is 6. Replace the connection cable of the display
defective. screen.

Calibration fails. All 1. The flow valve calibration does not pass. 1. Power off and power on, and then calibrate again,

152
Aeon8800A Service Manual Troubleshooting

Trouble Occurrence Phase Cause Analysis Solutions


or replace the gas module.

2. Prolong the silicone tube connecting to Insp.


2. AD difference is too large.
/Exp. ports.
3. The flow valve cannot be calibrated. 3. Replace MCB.
Ventilation Mode
All 1. Ventilation test under IPPV adult mode fails. 1. Clean the check valve.
test fails.

153
Aeon8800A Service Manual Troubleshooting

Trouble Occurrence Phase Cause Analysis Solutions


1. Noise arises as increasing AIR flow under IPPV 1. Adjust the spring on the AIR port of the gas
mode. inlet.
2. Noise arises as starting up the machine. 2. Replace the PEEP valve or flow valve.
There is noise
3. Noise of high pressure gas supply leads to low 3. Replace the reducing valve of high-pressure
coming out of the All
pressure alarm. O2.
machine.
4. Noise arises at the N2O gas inlet. 4. Replace the N2O cutoff-valve.
5. Noise arises as the bellows moving downside
5. Replace the PEEP valve.
under PCM mode.
Other setting
parameters are All 1. PTARGET is inaccurate in PCV mode. 1. Adjust the upper limit of pressure alarm.
incorrect.

154
Aeon8800A Service Manual Troubleshooting

Trouble Occurrence Phase Cause Analysis Solutions


1. Rotate the knob, if the screen works normally,
1. Panel is defective
the panel is defective; Replace the panel.
Panel is defective, 2. Rotate the knob, and there is no action on the
including MUTE screen; Check whether PC104-Communication
and All 2. Communication Board is defective. cable is connected well with a multimeter, if the
MANUAL/AUTO cable works well, the communication board is
key defective. Replace communication board.
3. Cable connected to PC104 and communication
3. Replace communication cable
Board is defective

1. Remove the speaker assembly from Isolation


Speaker is Board, measure the resistance between two
invalid.(when connectors of speaker assembly, if the resistance
1. Speaker is defective
alarm appears, the value is not within the range of 6-10Ω,the speaker
alarm icon assembly is defective. Replace speaker
All
displayed on the assembly,
screen while alarm
2. Do the above operation, if the resistance value
sound can’t be
2. Communication Board- Isolation Board cable is is within the range of 6-10Ω,power off machine
heard)
disconnected or damaged. and connect speaker assembly to Isolation Board,
Test the connectivity of T7 and T8 on

155
Aeon8800A Service Manual Troubleshooting

communication board with multimeter, if there is


no connectivity between T7 and T8, the
Communication Board- Isolation Board cable is
defective.
Replace Communication Board- Isolation Board
cable

3. Do the above operation if there is connectivity


3. Communication Board is defective between T7 and T8, Communication Board is
defective and replace it.

156
Aeon8800A Service Manual Troubleshooting

Trouble Occurrence Phase Cause Analysis Solutions


1. Rotate the knob, if the screen works normally,
1. Panel is defective
the panel is defective; Replace the panel.
Panel is defective, 2. Rotate the knob, and there is no action on the
including MUTE screen; Check whether PC104-Communication
and All 2. Communication Board is defective. cable is connected well with a multimeter, if the
MANUAL/AUTO cable works well, the communication board is
key defective. Replace communication board.
3. Cable connected to PC104 and communication
3. Replace communication cable
Board is defective
1. Check whether the breathing system heating
cable is disconnected or damaged with a
multimeter.
2. Remove breathing system cable from machine,
1. The breathing system heating cable is connect AC power, and power on, check whether
Breathing system
disconnected from heater cable socket. DS7 indicator of PSB is flashing or not. If not, the
heating function is All
2. PSB is defective. PSB is defective.
invalid
3. Breathing system heating assembly is defective. 3. The breathing system heating cable and PSB
work normally, connect heating cable to machine,
if DS7 indicator of PSB is still flashing or no longer
alight, breathing system heating assembly is
defective.

157
Aeon8800A Service Manual Troubleshooting

Trouble Occurrence Phase Cause Analysis Solutions


1. Close E-flowmeter, if there is gas flow at CGO
oulet, E-flowmeter is leaking. Replace
E-flowmeter.

1.E-flowmeter leaks 2. If E-flowmeter has no leakage and the


displayed digit has no change when rotating flow
E-flowmeter is
All 2. E- flowmeter Board is defective. regulator clockwise, E- flowmeter board is
defective
defective. Replace E- flowmeter board.
3. Flow sensor is defective
3. If E-flowmeter has no leakage and the
displayed digit increases when rotating flow
regulator clockwise, flow sensor is defective.
Replace flow sensor.

158
Aeon8800A Service Manual Troubleshooting

Trouble Occurrence Phase Cause Analysis Solutions


1. Measure the resistance of MANUAL/AUTO
valve at its connector. If the resistance is not
within the range of 200-400Ω, the connector of
valve is disconnected or MANUAL/AUTO valve is
defective. Reconnect valve cable or replace
MANUAL/AUTO valve.
1. MANUAL/AUTO valve cable is loose.

MANUAL/AUTO 2. Power on, and measure the voltage between


Startup test 2. MCB-Interface Board cable is defective.
valve fails T51 and T52 of PSB.

3.MCB is defective,
If the voltage is not within the range of 22-25V,
MCB-Interface Board cable is defective, replace
MCB-Interface Board cable

If the voltage is within the range of 22-25V, MCB


is defective. Replace MCB.

159
Aeon8800A Service Manual Troubleshooting

Trouble Occurrence Phase Cause Analysis Solutions


1. Measure the resistance of expiratory valve at its connector. If the
resistance is not within the range of 2-5Ω, the connector of valve is
disconnected or expiratory valve is defective. Reconnect valve cable
or replace expiratory valve.

2. Test the connectivity between connector and cover of expiratory

1. PSB is defective. valve with a multimeter. If there is a good connectivity, the expiratory
valve is defective, replace expiratory valve.

2. MCB is defective.
3. Remove MCB-Interface Board signal cable from the machine.

Expiratory valve Power on, and measure the voltage between T7 and T16 of PSB.
Startup test 3. Expiratory valve connector is
fails
loose. If the voltage is not within the range of 3-4.5V, PSB is defective,
replace PSB.
4. Expiratory valve is defective.
If the voltage is within the range of 3-4.5V, measure the voltage
between T54 and T55 of MCB, If the voltage is not within the range of
3-4.5V, MCB-Interface Board power supply cable is defective.
Replace MCB-Interface Board power supply cable

4. If all the above malfunctions are not existed, MCB is defective,


replace MCB.

160
Aeon8800A Service Manual Troubleshooting

Trouble Occurrence Phase Cause Analysis Solutions


1. Measure the resistance of Inspiratory valve at its connector. If the
resistance is not within the range of 70-80Ω, the connector of valve is
disconnected. Reconnect valve cable.

2. Remove MCB-Interface Board signal cable from the machine.


1. PSB is defective.
Power on, and measure the voltage between T5 and T17 of PSB.

2. MCB is defective. If the voltage is not within the range of 11-13V, PSB is defective,
Inspiratory valve replace PSB.
Startup test
fails 3. Inspiratory valve connector is
loose. If the voltage is within the range of 11-13V, measure the voltage
between T56 and Ground of MCB, If the voltage is not within the
range of 11-13V, MCB-Interface Board power supply cable is
defective. Replace MCB-Interface Board power supply cable

3. If all the above malfunctions are not existed, MCB is defective,


replace MCB.

161
Aeon8800A Service Manual Troubleshooting

Trouble Occurrence Phase Cause Analysis Solutions


1. Remove gas tube from pressure sensor, Measure the voltage
between T42 and AGND of MCB, if the voltage is not 0.5V, regulate
VR1 until T42 voltage gets to 0.5V. If the voltage can’t get to 0.5V or
has no changes when regulating VR1, MCB is defective and replace
MCB.

1. Ambient pressure board is


2. Skip startup test and enter into Manual Test interface, push “O2+”
defective.
Pressure sensor button to make the airway pressure up to 30cmH2O, measure the
Startup test
fails voltage between T42 and AGND of MCB with a multimeter.
2. MCB is defective.
If the voltage is not within the range of 1.2-1.4V, MCB is defective
and replace it.

If the voltage is within the range of 1.2-1.4V, power on the machine


again, and pressure sensor still fails after startup test, ambient
pressure board is defective. Replace ambient pressure board.

162
Aeon8800A Service Manual Troubleshooting

Trouble Occurrence Phase Cause Analysis Solutions


1. Measure the voltage of pin of J4 to ground;

1 to Ground: 11-13V;
2 to Ground: 11-13V
3 and 4 have connectivity to ground.
1. LED is defective. 5 to Ground: 4-5V;
6 to Ground: 2-5V;
2. Backlight is defective.
One of above condition can’t be satisfied, communication board is
defective, replace the communication board.
Black Screen All 3. PC104 is defective

2. All the above conditions can be satisfied, disconnect LED cable


4. Communication board is
from PC104.
defective
If the screen is black, backlight is defective, replace the backlight.
If the screen is white, the backlight is normal, LED or PC104 is
defective, replace the invalid part.
MCB.

163
Aeon8800A Service Manual Troubleshooting

5.2. General Troubleshooting Guide


Problem 1 Something wrong with Startup
Phenomenon:
The display, top light and spotlight can’t light up while the E-flowmeter is normal.
Action:
The phenomenon above shows that there is something wrong with 24V power supply while other
power supplies are normal.
1. Connect the AC power supply, rotate the switch to “ON”.
2. Open the lower panel of machine.
3. Check the indicating light Z2 (PWBUS STATE) on the PSB. The position of Z2 is shown as
the picture bellow.

164
Aeon8800A Service Manual Troubleshooting

If Z2 (DS2) indicator doesn’t light up, the PSB is with malfunction


If Z2 (DS2) indicator lights up, there is something wrong with the cable connected to interface
board, the cable is shown as the follow picture.

165
Aeon8800A Service Manual Troubleshooting

Problem 2 Something wrong with Startup when using the battery supply only

Phenomenon
Disconnect AC power supply to use the battery only, and rotate the switch to “ON” to find that the
machine can’t startup.

Action:
1. Rotate main switch to “OFF”, Disconnect the AC power.
2. Remove the lower rear panel of the machine.
3. Disconnect the cable connected to the battery.
4. Test the voltage between the middle pin and upper pin; then the voltage between the lower
and the middle pin; Shown as the following picture.

If both the two voltage are over 24V, that’s to say PSB is with a malfunction.

166
Aeon8800A Service Manual Troubleshooting

If one or two voltage is less than 24V, that’s to say cable connected to the battery is with a
malfunction.

Problem 3 Automatic Leak Test failure--- Startup Self-test

Phenomenon:
The automatic leak test fails during the startup self-test.

Action:
1. Observe the airway pressure gauge value during Automatic Leak Test.
The normal airway pressure values changing during Automatic Leak Test:
First, the airway pressure ascends to 80cmH2O, and then quickly descends to 50cmH2O, and
then descends to 30cmH2O slowly and stays at 30cmH2O for a while.
2. If the airway pressure ascends slowly or even doesn’t ascend, the flow valve needs
calibration.
3. If the airway pressure descends from 30cmH2O without staying, that to say the machine has
a leakage.

Problem 4 Flow valve calibration failure

Phenomenon:
The flow valve calibration in the Service Mode fails.

Action:
1. Make sure the gas supply pressure is normal.
2. Make sure flowmeter is fully closed; the drive gas tube, CGO, absorber canister O2 sensor
and airway pressure gauge are all well connected or installed.
3. Make sure the bellows and Insp. /Exp. valve disks are removed and the bellows cover and
Insp./Exp. Valve covers are installed.
4. Seal the Expiratory port.

167
Aeon8800A Service Manual Troubleshooting

If the flow valve calibration failure is displayed quickly after performing the calibration, at the same
time, the threshold voltage and the maximum flow are very small, recalibrate the flow sensor.

If the flow valve calibration failure is after a while, and the maximum flow rate is bellow 90L/min,
replace the pressure regulator.

Problem 5 Unsteady tidal volume (Vt)

Phenomenon:
It needs 5 periods to get to the set value after setting Vt value, and the Vt always changes in the
ventilating process.

Action:
Enter into Service Mode and recalibrate the flow valve.

Problem 6 High gas supply pressure

Phenomenon:
High O2 supply pressure over 0.5Mpa in hospital may lead to the unsteady tidal volume.

Action;
Recalibrate flow valve if the O2 supply pressure is over 0.5Mpa.

168
Aeon8800A Service Manual Troubleshooting

Problem 7 detect the machine leakage

Phenomenon:
The machine has one or some leakage, the leakage place should be found.

Action:
1. First do the manual leak test.
 Set to “STANDBY” mode.
 Seal Insp. /Exp. Ports and bag arm port.
 Rotate the flowmeter knob to make the airway pressure up to 30cmH2O.
 Close the flowmeter, observe the airway pressure value to find if the value descends or
not.
2. Second do the automatic leak test.
Set to “STANDBY” mode, open the “Check” sub-menu and to perform the automatic leak test.
 If leakage is found both in manual and automatic leak test progress:
Check absorber canister, O2 sensor interface, airway pressure gauge and Insp. /Exp.
Ports valve covers.
 If leakage is only found in manual leak test process
Check the APL valve and connector of the bag arm.
 If leakage is only found in automatic leak test process
Check the bellows, bellows cover, U-shape sealing ring and drive gas tube.

169
Aeon8800A Service Manual Troubleshooting

Problem 8 Bellows move downwards abnormally

Phenomenon:
In the progress of automatic leak test during startup, the bellows falls rather than stays at the top
of bellows cover after the bellows is full filled with gas when pressing “O2+”.
The bellows falls abnormally by accident during normal ventilation, at the same time the leak test
result is normal.

Action:
The phenomenon above indicates that there is something wrong with the bellows check valve.

Replace the bellows check valve:


1. Remove the bellows cover, bellows and disconnect breathing system heating line, drive gas
tube and corrugated hose.
2. Remove four M4 screws of breathing system shown as the follow picture.

NOTE: The four screws are in the four deep holes. Don’t remove the breathing system from
machine when disassembly breathing system. First, loose the screws and then remove them
when disassembly.
3. Open the bellows fixing plate shown as the following pictures.

170
Aeon8800A Service Manual Troubleshooting

4. Replace the bellows check valve.


NOTE: Insert the screwdriver into the check valve and then rotate to remove the check valve,
don’t pull it up directly, shown as the following pictures.

NOTE: Daub a little silica gel on the O-ring surface and push into the slot softly when installing
bellows check valve.
Install the bellows fixing plate
1. Put the expiratory valve into the bellows fixing board.
2. Make sure there are 5 O-rings and every surface of O-rings is glazed without any damage.
3. Put all the screws into the breathing system holes and the tighten them.

171
Aeon8800A Service Manual Troubleshooting

Expiratory valve

4. Connect drive gas tube, breathing system heating cable and corrugated hose.
5. Install the bellows and bellows cover.
6. Perform the leak test.

Problem 9 Absorber canister ascend-descend assembly got stuck

Phenomenon:
Absorber canister ascend-descend assembly got stuck and can’t descend.

Action:
Absorber canister lock assembly needs to be replaced.
1. Open the lock assembly and remove absorber canister.
2. Disconnect corrugated hose and chain, and then remove ascend-descend assembly, shown
as the following pictures.

3. Loose and remove the two M4 screws shown as the following picture and replace the lock
assembly.

172
Aeon8800A Service Manual Troubleshooting

4. Install ascend-descend assembly.


5. Connect corrugate hose and chain.
6. Install the absorber canister and perform the leak test.

Problem 10 Leakage from loose breathing system lock assembly

Phenomenon:
The loose breathing system lock assembly leads to the leakage from absorber canister.

Action:
Check the two screws on the suspender and two screws on the lock assembly to make sure
whether the screws are loose or not, if so, tighten the screws and check the leakage; if not
regulate the lock assembly again.
1. Remove the absorber canister, corrugated hose and ascend-descend assembly
2. Remove the two M4 screws on the lock assembly and then remove the lock assembly.
3. Loosen the screws on the suspender, rotate to regulate the suspender, make the length of the
suspender 284mm~284.5mm, and then tighten the screws.

173
Aeon8800A Service Manual Troubleshooting

4. Reinstall the lock assembly, ascend-descend assembly, absorber canister and corrugated
hose.
5. Perform leakage test.

174
Aeon8800A Service Manual Maintenance Information

6. Maintenance Information

6.1. Maintenance introduction


This chapter of the Service Manual provides the necessary technical information to perform
repairs to the system. The most important prerequisites for effective troubleshooting are a
thorough understanding of the system's functions, as well as understanding its principles of
operation.

Warning: Whenever using anesthetic gases, nitrous oxide, oxygen, or any hospital gas

always follow the appropriate agent evacuation/collection procedures. Use the


hospital gas evacuation system.

Caution: This device uses high pressure compressed gas. When attaching or

disconnecting backup gas cylinders, the user should always turn the cylinder
valves slowly. Use Aeon8800A flow meters to bleed down the pressure,
watching the cylinder gauge indicate the depleting cylinder pressure, before
disconnecting the cylinder from the yoke. Always open and close cylinder
valves fully.

Caution: This device operates using compressed gas at high pressures from the hospital

central supply. When connecting gas supply lines attach the hose connection to
the machine before connecting the quick disconnect fitting to the hospital source.
Disconnect the supply hose from the hospital source connection prior to
disconnecting it from Aeon8800A gas connection fittings.

175
Aeon8800A Service Manual Maintenance Information

6.2. Maintenance Guidelines


Identify the problem—Due to the wide variety of potential symptoms, certain problems may be
more subtle than others. Following the guidelines of the tests will help determine the problem, if
one exists.
Avoid shorting component leads together—during repair procedures, it can be tempting to make a
quick series of measurements. Always turn the power off before connecting and disconnecting the
test leads and probes. The accidental shorting of leads can easily stress the components and
cause a second failure (aside from the safety risk).
Use the proper equipment—the equipment listed in "Special Tools Required" is suggested to fulfill
a wide range of troubleshooting requirements. It is imperative to use the designated equipment in
order to ensure proper results of any and all test procedures.
Clean up the repair area—after any repair, clean off the repair area.

6.3. Repair backup parts list

Replacing Field Factory


No. Part No. Part Name Qty. Remark
Period Repair Repair

Top light assembly


Top light backlight
1 122003763 1 Yes
Assy
Top Light Switch
2 122003764 1 Yes
Assy
Top Light
3 130007957 1 Yes
Penetrable Board

Screen assembly

Touch Screen
5 122012063 1 Yes
Assy
LCD Controlled
8 210004112 1 As needed Yes
NLB150XG01L-01 component
Communication Controlled
9 122006839 1 Yes
board component
10 122005425 Screen Controller 1 Yes

E-flowmeter Assy.

176
Aeon8800A Service Manual Maintenance Information

Replacing Field Factory


No. Part No. Part Name Qty. Remark
Period Repair Repair
Important
11 122006793 E-flowmeter Assy. 1 Yes
component

Auxiliary flowmeter assembly

Auxiliary
12 130007947 Flowmeter 1 Yes
Penetrable Board
Auxiliary
13 122003737 1 Yes
Flowmeter Assy

Gas inlet module

N2O Input
14 122003754 1 Yes
Integrated Assy
N2O Input
15 130007975 1 Yes
Connector
AIR Input
16 122003753 1 Yes
Integrated Assy
AIR Input
17 130007974 1 Yes
Connector
O2 Input Integrated
18 122003752 1 Yes
Assy
19 130007973 O2 Input Connector 1 Yes
Gas Supply
20 122005527 1 Yes
Pressure Board
Drive Gas
21 122003757 Pressure-Relief 1 Yes
Valve Assy
Single-Output
22 122003756 Pressure-Relief 1 Yes
Valve Assy
Double-Output
23 122003755 Pressure-Relief 2 Yes
Valve Assy
24 130003930 Filter 3 3 years

177
Aeon8800A Service Manual Maintenance Information

Replacing Field Factory


No. Part No. Part Name Qty. Remark
Period Repair Repair

Pressure Gauge assembly

High Pressure
25 122003071 N2O Gauge 1 Yes
Assembly
High Pressure AIR
26 122003070 1 Yes
Gauge Assembly
High Pressure O2
27 122003069 1 Yes
Gauge Assembly
Pressure Gauge
28 130008025 6 Yes
Bracket
Low Pressure N2O
29 122005495 1 Yes
Gauge Assy
Low Pressure AIR
30 122005496 1 Yes
Gauge Assy
Low Pressure O2
31 122005497 1 Yes
Gauge Assy

Knob assembly

32 122011958 Encoder Assy 1 Yes

33 130007865 Knob 1 Yes

Switch Assembly

34 210001743 Switch 1 Yes


Switch End
35 210001744 1 Yes
Module, 1NO

Outlet module

Outlet Module
36 122003760 1 Yes
Assy
Interface board-
breathing system
37 122006835 1 Yes
heating cable of
outputting

Power supply assembly

178
Aeon8800A Service Manual Maintenance Information

Replacing Field Factory


No. Part No. Part Name Qty. Remark
Period Repair Repair
AC Power Assy
38 122006797 1 Yes
(E)
AC Power Assy
39 122006798 1 Yes
(C)
AC Power Assy
40 1 Yes
122006799 (U)
Controlled
41 122006840 Isolation Board 1 Yes
component
42 122006833 Speaker Assy 1 Yes

43 210003583 Fuse 250V/10A 2 Yes

44 210003583 Fuse 250V/2A 8 Yes

O2 Sensor assembly

45 210001975 O2 Sensor 1 1 Year Yes


Five-Core Shield
Cable
46 210003053 1 Yes
LE.BA106.ABE-16
898
O2 Sensor
47 130008105 1 Year Yes
Connector
O-shape silica gel
304000101
sealing

Cylinder gas inlet assembly

N2O pressure
240000542 regulator 1
Assembly
Air pressure
240000543 regulator 1
Assembly
O2 pressure
240000541 regulator 1
Assembly

Pneumatic Box Assembly

179
Aeon8800A Service Manual Maintenance Information

Replacing Field Factory


No. Part No. Part Name Qty. Remark
Period Repair Repair

Pneumatic Box
49 122006806 1 Yes
Assy
50 122005725 IP Valve Assy 1 Yes

51 122006871 Gas Module Assy 1 Yes

52 122000834 Safety Valve 1 Yes


Soneloid Valve Controlled
53 240000097 1 Yes
AMISCO 24V component
Ambient pressure Controlled
54 122006848 1 Yes
board component
Proportional valve Controlled
55 210003135 1 Yes
1350-5083 component
Main Control Controlled
56 122006838 1 Yes
board component
Controlled
57 122006836 PSB 1 Yes
component
Switching power Controlled
58 210003567 1 Yes
supply component
Battery Assy Controlled
59 122007196 1 3 Years Yes
SLA-4AH24V-3 component

Breathing System

AC200 Breathing Important


60 122004128 1 Yes
System component
61 122005509 Bag arm Assy 1 Yes

62 122001761 APL Valve 1 Yes


Pressure Gauge
63 122005511 1 Yes
Assy
Inspiratory Valve
64 122005512 1 Yes
Assy
Inspiratory Valve
65 130007762 1 Yes
Cover
66 130004102 Valve Disk 1 Yes

67 122005513 Expiratory Valve 1 Yes

180
Aeon8800A Service Manual Maintenance Information

Replacing Field Factory


No. Part No. Part Name Qty. Remark
Period Repair Repair

Assy

Expiratory Valve
68 130007763 1 Yes
Cover
69 122001737 Switch Assy 1 Yes
Expiratory Valve
70 122005518 1 Yes
Assy
Bellow Check
71 122005519 1 Yes
Valve Assy
72 130004091 U-Type Gasket 1 3 Years Yes

73 130004086 Bellows Cover 1 Yes

74 122005548 Bellows Assembly 1 3 Years Yes


O2 Concentration
75 122005521 1 Yes
Plug Assy
Ascend-descend
76 122003782 1 Yes
Assembly
77 122003068 Lock Assembly 1 Yes
Absorber Canister
78 122003216 1 Yes
Assy
79 130004092 Corrugated Tube 1 1 Year Yes

80 130006719 Elastic Gasket 2 1 1 Year Yes

81 130004147 Valve Seal Gasket 1 1 Year Yes


Used in O2
O-shape silica gel concentrati
82 304000100 1 1 Year Yes
sealing18×1.8 on plug
Assy
Used in
O-shape silica gel
83 304000096 1 1 Year Yes pressure
sealing9×1.8
gauge Assy
Used in
O-shape silica gel
inspiratory
84 304000102 sealing22×1.8 1 Yes
valve Assy,
expiratory

181
Aeon8800A Service Manual Maintenance Information

Replacing Field Factory


No. Part No. Part Name Qty. Remark
Period Repair Repair

valve Assy

Used in
inspiratory
O- shape silica gel
85 304000108 1 Yes valve Assy、
sealing t 42X1.8
expiratory
valve Assy

Other components

Used in
O-shape silica gel
86 304000044 1 Year Yes vaporizer
sealing14×2.65
mount Assy
87 210003051 LED Signal Light 1 Yes

88 122003740 C·G·O Assy 1 Yes

89 122003739 O2 Flush 1 Yes


Auxiliary O2
90 122003738 1 Yes
Supply Assy
91 130011896 Panel(C)

92 130011897 Panel(E)
Controlled
93 122006843 Interface board
component
94 122003762 Spotlight Assy 1 Yes
Auxiliary Work
95 122003761 1 Yes
Table Assy
O2 Sensor Output
96 122003741 1 Yes
Assy
Five core shielded
210003052 cableLM.BA105.A 1
BE
Integrated Silica
97 130010364 1 3 Years Yes
Tube
Host-breathing
98 122001624 1 Yes
system Cable
99 122005534 Vaporizer 1 Yes

182
Aeon8800A Service Manual Maintenance Information

Replacing Field Factory


No. Part No. Part Name Qty. Remark
Period Repair Repair

Mount(two kettles)

Cylinder Bracket
100 130003864 3 Yes
Gasket
Silica gel screwed
101 230000318 1 3 Years Yes
conduitφ17
102 220000280 YOKE Gasket 3 1 Year Yes

183
Aeon8800A Service Manual Maintenance Information

6.4. Special Tools Required

No. Part No. Name of Tools Quantity


1 --- Slotted Screwdriver (large,medium, 1 for each
small)
2 --- Crisscross Screwdriver(large, 1 for each
medium, small)
3 --- Inner Hexagon Spanner 1
4 --- Open spanner 1
5 --- Meter Screw driver 1
6 --- TSI 4043E 1
7 --- Multimeter 1
8 130004396 R15 hook spanner 1
9 240000460 TSI sensor of quality and flow 1
10 240000461 SS20 handheld pressure gauge 1
11 230000193 Silica gel corrugated pipe Φ22 1
(1000mm,light Green, standard)
12 230000189 Silica gel corrugated pipe Φ22 2
(600mm,light Green, standard)
13 230000188 Test lung 1L(Sponge lung) 1
14 230000075 Breathing bag(3L,black,4-piece) 1
15 122003766 O2 concentration ,monitoring kits 1
16 122002003 Throttle Assembly 1
17 122002002 Prolong tube block component 2
18 130006181 Test connector of expiratory valve 1
19 130005546 Connector of service and repair kit-2 1
20 130005547 Connector of service and repair kit-3 1
21 130005700 Connector of service and repair kit-4 1
22 122002946 Pressure sampling frock 1
23 130006182 Frock adapter of pressure sampling 1
frock
24 130006183 Pressure sampling connector 1
25 220000018 Straight-in flash-plug pipe tie-in 1
26 304000014 O- ring 10.6×1.8 1

184
Aeon8800A Service Manual Maintenance Information

27 301000347 Hexagon pattern countersunk nut 1


M4×6
28 220000080 German-style external angle hook 1
29 301000095 Plug Pan Head Screw M2×6 1
30 304000102 O- ring 22×1.8 1
31 122003112 Connect assembly of manual gas 1
seal test of integrated module
32 122003113 Two-connector assembly 1
33 304000097 O- ring 12×1.8 1
34 122003114 Four- connector assembly 1
35 122003115 Connector assembly-1 1
36 122003116 Connector assembly of expiratory 1
valve test
37 304000100 O- ring 18×1.8 1
38 304000125 O- ring 30×1.8 1
39 230000011 PU Tube 8/6(blue) 1 meter
40 230000273 High vacuum silica gel 1

185
Aeon8800A Service Manual Maintenance Information

6.5. Maintenance of machine

WARNING: During installation, make sure the polarity and the plug-in connections are correct

to avoid the risk of damage due to reversed polarity.

WARNING: Make sure not to kink or squeeze any hoses when reassembling units to avoid the

risk of malfunction due to kinked or squeezed hoses.

6.5.1. Bleed gas pressure from the machine

Before disconnecting pneumatic fittings, bleed all gas pressure from the machine.
1. Close all cylinder valves and disconnect all pipeline supplies from the source.
2. Set the main switch to ON.
3. Ensure that all cylinder and pipeline pressures read zero.
4. Establish a flow for the affected gas to bleed down the pressure.
5. Set the system to Standby mode.

6.5.2. Disassembly the rear panels

WARNING: Before removing rear panels.

1. Bleed all gas pressure from the machine.


2. Disconnect AC power cord and all electrical cables.
3. Disconnect all pipeline or cylinder gas source.

Disassembly the upper rear panel


1. Rotate and loosen the fixing screw anticlockwise with screwdriver, and then remove the six
screws from the rear panel.
2. Remove the upper rear panel from the machine with much care.
3. Remove the screws on the gas inlet module panel and power supply panel shown as the
following picture and open the two panels could repair the upper parts of anesthesia machine.

186
Aeon8800A Service Manual Maintenance Information

Figure 6-1
Disassembly the lower rear panel
1. Remove the two fixing screws near battery box with screwdriver.
2. Then remove the other ten screws around the panel.
3. Remove the lower rear panel from the machine with care.

187
Aeon8800A Service Manual Maintenance Information

Figure 6-2 Lower rear panel

Disassembly the lower YOKE panel


The YOKE panel has four screws on the each corner as shown in the bellow figure.

Figure 6-3
1. Rotate the screwdriver anticlockwise and then remove the fixing screws on the panel.
2. Open the YOKE panel.

188
Aeon8800A Service Manual Maintenance Information

6.5.3. Anesthesia Machine Upper Parts Repairing

WARNING: During installation, make sure the polarity and the plug-in connections are correct

to avoid the risk of damage due to reversed polarity.

WARNING: Make sure not to kink or squeeze any hoses when reassembling units to avoid the

risk of malfunction due to kinked or squeezed hoses.

6.5.3.1. Power supply panel repairing

1. Remove the screws on the power supply panel. (Right side of above picture)
2. Open the power supply panel Shown as the following picture. (Left side)
3. Loose four screws on the box and remove the cover. Shown as the following picture.(Right
side)
4. Repair or replace any parts in the box as need.

Note: It is only necessary to remove the AC Input Panel when replacing the flow sensor

assembly. As necessary, proceed to next step.

Figure 6-4 power supply box

1 ISB(Isolation Board) 3 AC power socket


2 Fuse 4 Power Switch

189
Aeon8800A Service Manual Maintenance Information

6.5.3.2. Gas inlet integration module panel repairing

Gas inlet integration module includes O2 inlet integration module, AIR inlet integration module and
N2O inlet integration module.
Disassembly of O2 inlet integration module is an example of the replacement of gas inlet
integration module.

Figure 6-5 O2 Inlet Integration Module

Figure 6-6 O2 Inlet Integration Module

O2 Inlet Integration Module Replacement


1. Disconnect all the gas tubes from the module.
2. Disconnect GSPB cable (W17) from J14 on Interface Board and GSPB.
3. Remove two screws securing GSPB to O2 Inlet Integration Module
4. Remove GSPB from the module.
5. Loosen and remove the two screws marked 1, 2 on the picture above.
6. Remove O2Inlet Integration Module from the machine bracket.
7. Replace GSBP and O2 Inlet Integration Module as needed with a new one.
8. Reassembly GSBP and O2Inlet Integration Module in the reverse order of disassembly.

190
Aeon8800A Service Manual Maintenance Information

NOTE: The detail tubes connection and cables connection refer to chapter 2.1.2 and chapter
2.2.2.

6.5.3.3. Replacement of filter in gas inlet integration module

Warning: The filter replacing period is 3 years in normal use.

Warning: After replacing the filter, check and verify the sealing ability of gas inlet integration

module.

Warning: Daub proper glue to the screw when installing the filter to avoid the connector loosing

in long term of use.


Disassembly:

1. Remove two screws securing the gas


connector to the gas inlet integration module.

Figure 6-7
2. Remove the gas connector from the module
and put on a safe place.

3. Pull out the O-ring sealing gasket (1) and Figure 6-8
filter (2). Replace with a new filter. 1 O-ring sealing gasket
2 Filter
3 Fixing screw

191
Aeon8800A Service Manual Maintenance Information

Reassembly:

1. Put O-ring sealing gasket and filter into the


module.

Caution: Be careful when installing the filter

to avoid O-ring and filter dropping out.

Figure 6-9
2. Daub proper glue to the screw surface.

Figure 6-10
3. Align the gas connector to other parts of the
module and secure the gas connector with two
fixing screws.

4. Check the sealing ability of the gas inlet


integration module. Refer to chapter
3.5.4~3.5.7 Figure 6-11

192
Aeon8800A Service Manual Maintenance Information

6.5.3.4. Drive Gas Pressure Regulator Replacement

Figure 6-12 Drive Gas Pressure Regulator Position

Figure 6-13 Pressure regulators


1. Disconnect all the gas tubes connected to the regulator.
2. Loose two screws securing the regulator bracket to panel.
3. Remove Regulator Assembly.
4. Remove nut securing bracket to Regulator.
5. Record the revision and serial number of the new regulator.
6. Reassembly the new regulator to bracket and secure with two nuts.
7. Secure Regulator Assembly with (2) screws to the panel.
8. Connect a gas tube to the upper side gas connector of regulator.
9. Attach a pressure meter to the other side of the regulator.
10. Verify regulator is 1.75 +/- 0.5 bar (adjust regulator as necessary).
11. Remove Pressure Meter and reinstall tube to the bottom gas connector of regulator.

193
Aeon8800A Service Manual Maintenance Information

6.5.3.5. Replacement of E-flow meter assembly

W19

W31

Figure 6-14 E-flow meter Assembly

1. O2/N2O ratio-controlled valve


2. Gas flow sensors

194
Aeon8800A Service Manual Maintenance Information

Figure 6-15
a: O2 flow sensor
b: Air flow sensor
c: N2O flow sensor

Disassembly:
 E-flowmeter assembly
1. Disconnect gas tubes connected to gas regulators.
2. Disconnect gas tube connected to vaporizer.
3. Disconnect W19 connected to Interface Board for gas supply and communication.
4. Loose and remove four screws secured E-flowmeter assembly to machine on the four corners.
5. Remove E-flowmeter assembly from anesthesia machine.
 Gas flow sensors
1. Disconnect all the gas tubes connected to gas flow sensors.
2. Disconnect W31 connected to EFM.
3. Loose and remove four screws secured gas flow sensors to E-flowmeter assembly.
4. Remove gas flow sensors from E-flowmeter assembly.
 O2/N2O ratio-controlled valve
1. Disconnect gas tubes connected to O2/N2O ratio-controlled valve.

195
Aeon8800A Service Manual Maintenance Information

2. Loose and remove two screws on the manifold that fixing O2/N2O ratio-controlled valve to
E-flowmeter assembly.
3. Remove O2/N2O ratio-controlled valve from E-flowmeter assembly.
.
Assembly:
1. Replace the trouble parts as needed.
2. Record the serial number of the new parts.
3. Assembly the E-flowmeter assembly in reverse order of disassembly procedure
4. Secure E-flowmeter assembly to the machine.
5. Connect all the gas tubes and cable to E-flowmeter assembly.

196
Aeon8800A Service Manual Maintenance Information

6.5.4. Maintenance of electrical box

6.5.4.1. Replacement of MCB

W13
W18

W15

W20

W14

W11 W12

Figure 6-16 MCB


Disassembly and Reassembly:
1. Disconnect W13 from J3 connected to Solenoid valve
2. Disconnect W15 from J2 connected to IP valve
3. Disconnect W14 from J1 connected to Woodward valve
4. Disconnect W11 from J4 connected to Power Supply
5. Disconnect W12 from J5 connected to Interface board A side.
6. Disconnect W20 from J12 connected to LJ4
7. Disconnect W18 from J13 connected to Ambient Pressure Board.
8. Remove all the gas tube connected sensors.
9. Remove4 screws fixing the MCB to machine
10. Replace new MCB and record the new version and serial number of new MCB.
11. Reassembly the new MCB in reverse order of disassembly procedure.

197
Aeon8800A Service Manual Maintenance Information

6.5.4.2. Replacement of PSB (PSB)

Interface Board

PSB

Switching power supply

Figure 6-17 PSB position

198
Aeon8800A Service Manual Maintenance Information

W2

W10

W4
W9

Figure 6-18 PSB

Disassembly and Assembly:


1. Disconnect W10 from J4 connected to Interface Board A side
2. Disconnect W9 from P2 connected to Interface Board A side.
3. Disconnect W2 from J2 connected to Switching Power Supply
4. Disconnect W4 from J1 connected to Battery
5. Remove four screws fixing PSB to machine and remove PSB.
6. Replace the new PSB and record the new version and serial number.
7. Reassembly the new PSB in reverse order of disassembly procedure.

199
Aeon8800A Service Manual Maintenance Information

W1

Figure 6-19

Switching power supply

W2

Figure 6-20
Disassembly and Assembly:
1. Remove 2 screws fixing Switching Power Supply to machine frame.
2. Disconnect W2 connected to PSB
2. Disconnect W1 from connected to Interface Board A side.
3. Disconnect W2 from J2 connected to LJ1
6. Replace the new Switching Power Supply
7. Reassembly the new one in reverse order of disassembly procedure.

200
Aeon8800A Service Manual Maintenance Information

6.5.4.3. Replacement of Gas Module

WARNING: During installation, make sure the polarity and the plug-in connections are correct

to avoid the risk of damage due to reversed polarity

WARNING: Make sure not to kink or squeeze any hoses when reassembling units to avoid the

risk of malfunction due to kinked or squeezed hoses.

WARNING: In case of the risk from the wrong connection of tube, pay more attention when

connecting the tubes.

Figure 6-21 Gas Module

201
Aeon8800A Service Manual Maintenance Information

Inspiratory W18

Valve

Solenoid Valve

Figure 6-22

Figure 6-23 Gas Module


Disassembly:
1. Remove the tubes on the right side of Gas Module
2. Remove two tubes which has bigger diameter from right side of Gas Module.
3. Remove four tubes from front side of Gas Module.
4. Disconnect one end of W15 from J6 on SBD, disconnected the other end from Ambient
Pressure Sensor Board.
5. Disconnect W12(Insp. - Valve), W13(Solenoid Valve) cable from JJ1、JJ2 interface on MCB
6. Remove two screws securing Gas Module to the machine.
7. Remove the Ambient Pressure Sensor Board secured to the Gas Module with two screws

202
Aeon8800A Service Manual Maintenance Information

8. Record the new version and serial number of new Gas Module
Assembly:
1. Fix the Ambient Pressure Sensor Board to the Gas Module with two screws
2. Fix Gas Module with two screws to the machine
3. Connect W12 of Insp.-Valve、W13of Solenoid Valve cable to JJ1, JJ2 interface on MCB
4. Connect one end of W15 to J6 on SBD, connect another end of W15 to Ambient Pressure
Sensor Board
5. Connect two tubes to connectors on the left side of Gas Module
6. Connect two tubes which has bigger diameter to connectors on the right side of Gas Module.
7. Connect four tubes to connectors on the front side of Gas Module.

6.5.4.4. Replacement of proportion valve

CAUTION: The screws fixing proportion valve have plastic gasket, please pay more attention

when replacing the proportion valve to avoid the gasket missing.

CAUTION: Line up the proportion valve and holes in the module when reinstalling the

proportion valve to avoid the reverse installation.

Figure 6-24 Proportion Valve


Replacement:
1. Remove the gas module from machine. The detail disassembly procedure refers to the chapter
6.5.4.3
2. Remove the two fixing hexagon screws with hexagon screwdriver, see the figures as bellow.

203
Aeon8800A Service Manual Maintenance Information

Figure 6-25
3. Replace and the new proportion valve
4. Record the version and serial number of the new proportion valve.
5. Put the hexagon screws into the proportion valve holes.
6. Put the plastic gasket on the screws.
7. Line up the proportion valve and the module.
8. Rotate the screws counter clockwise with screwdriver and fix proportion valve.
9. Check the gas module sealing ability, refer to chapter 4.3.1.

204
Aeon8800A Service Manual Maintenance Information

6.5.4.5. Replacement of solenoid valve

CAUTION: Do not forge the sealing gasket when reassembling the solenoid valve to insure the

sealing ability.

Figure 6-26 Solenoid valve

Figure 6-27 Sealing gasket

Replacement:

1. Remove the gas module from machine. The


detail disassembly procedure refers to the

205
Aeon8800A Service Manual Maintenance Information

chapter 6.5.4.3.

2. Rotate the fixing screws counter clockwise


with a screwdriver, and then remove the
solenoid valve from the module. See the
picture shown on the right.

Figure 6-28
3. Remove the fixing screws of the solenoid valve and the remove the valve from the module.
4. Replace the solenoid valve.
5. Install solenoid valve use the reverse procedure of disassembly
6. Check the gas module sealing ability, refer to chapter 4.3.1.

206
Aeon8800A Service Manual Maintenance Information

6.5.4.6. Replacement of safety valve

CAUTION: Check the sealing gasket when replacing the safety valve. If the sealing gasket is

damaged on the surface or out of use, please replace it.

Figure 6-29 Safety valves


Disassembly
1. Remove the gas module from machine. The detail disassembly procedure refers to the chapter
6.5.4.3

2. Rotate the screws counterclockwise with the


screwdriver and remove the two fixing screw shown as
the picture on the right.

3. Remove the gas exhausted port

4. Pull out the old safety valve and replace it.


Figure 6-30

5. Install the new safety valve in the reverse order of


disassembly.

6. Check the gas module sealing ability, refer to chapter.4.3.1.

207
Aeon8800A Service Manual Maintenance Information

6.5.4.7. Replacement of Ambient Pressure Board (APB)

WARNING: Disconnect W15 cable before replacing APB.

CAUTION: Replace APB directly when the ambient pressure sensor doesn’t work normally.

CAUTION: The screws fixing APB have plastic gasket, please pay more attention when

replacing APB to avoid the gasket missing.

Figure 6-31 APB

Disassembly and Installation

1. Remove the gas module from machine.


The detail disassembly procedure refers to
the chapter 6.5.4.3

2. Rotate the fixing screws counter clockwise


with a screwdriver, and remove APB. The
APB position seen the picture shown on the Figure 6-32
right.
3. Replace APB, and record the .version and serial number of the new APB.

4 Install APB using the reverse procedure of disassembly.

208
Aeon8800A Service Manual Maintenance Information

6.5.4.8. Replacement of PEEP Valve

Figure 6-33 IP Valve


Disassembly:
1. Remove two tubes connected to left side of valve.
2. Remove one tube connected to front side of valve.
3. Remove one tube connected to right side of valve.
4. Disconnect W15 cable connected to J2 on MCB
5. Remove two screws (1) securing IP valve bracket to gas box.
6. Remove four screws (2) securing IP valve to bracket.
7. Replace the new IP valve and record its revision and serial number.
Assembly:
1. Secure bracket to IP valve with four screws.
2. Secure IP valve assembly to gas box with two screws.
3. Connect W15 of IP valve to J2 on MCB
4. Connect tube to rear connector on left side of the valve.
5. Connect tube to front connector on left side of the valve.
6. Connect tube to connector on front of the valve.
7. Connect tube to connector on right side of the valve.

209
Aeon8800A Service Manual Maintenance Information

6.5.4.9. Replacement of battery

Warning: Disconnect AC power supply before replacing the battery.

Warning: Rotate main switch to “OFF” to close the anesthesia machine when replacing the

battery.

Caution: The disposal of used batteries must be conducted in accordance with local

regulation requirements.

Caution: The battery has the same specifications with Aeon8800A Anesthesia machine is

recommended

Step 1 Step 2

Step 3 Step 4

210
Aeon8800A Service Manual Maintenance Information

Step 5

Figure 6-34、35、36、37、38

Replacement:
1. Disconnect the AC power supply;
2. Set the main switch to “OFF”
3. Remove four screws fixing the battery assembly to machine frame.
4. Pull the battery assembly out.
5. Disconnect battery cable connected to LJ2.
6. Remove the two screws on the securing plate and then remove the securing plate.
7. Pull out the old battery and disposal of it in accordance with local regulation requirements.
8. Replace the old battery with a new one.
9. Reinstall the new battery in the reverse order of disassembly.

211
Aeon8800A Service Manual Maintenance Information

6.5.5. Disassembling the Display from Main Control Unit

1 Screw

2 Spring
lock

Figure 6-39 Display

To separate the display from the Main Control Unit

a. Loose the screws (No.1)

b. Pullout the spring lock(No.2)

c. Slide out the display along the direction of the bracket.

212
Aeon8800A Service Manual Maintenance Information

6.5.6. Removal of Components in the Display

Four screws
Two screws

Figure 6-40 Figure6-41

a. Remove the four screws in Figure 6-40 and remove the mounting plate;
b. Remove the two screws in Figure 6-41;

Figure 6-42 Figure 6-43

c. Pry open the back-shell of display with a flat blade screwdriver at the areas circled in Figure
6-42.Be careful not to damage the display housing.

d. Open the back-shell slowly and remove speaker by removing the two screws in the
Figure 6-43, then remove the back-shell.

Screws

PCB mask

Figure 6-44

213
Aeon8800A Service Manual Maintenance Information

e. Remove the eight screws, and then remove the PCB mask, Figure6-44;

f. Figure6-45, shows the components in the display that can be easily removed.

Figure 6-45

NO. Name NO. Name

1 Alarm light panel 7 Display interface panel

2 Display Control Panel 8 Touch screen control

3 W36 cable 9 Aperture and mute Panel

PC/104 single board computer suite,


4 10 W31 cable
industrial grade CF card

5 Speaker Component 11 Front shell component

6 1G memory card

Test: When replacing any of the above components, you need to check if the screen is working
normally.

214
Aeon8800A Service Manual Maintenance Information

6.5.7. Spotlight

Figure 6-46
Replacement:
1. Disconnect the AC power supply;
2. Set the main switch to “OFF”;
3. Remove the upper rear panel;
4. Remove the two screws (circled).
5. Disconnect the cables as marked in the above figure.
6. Replace the spotlight with a new one.
7. Reinstall the new spotlight with the reverse procedure of disassembly.

215
Aeon8800A Service Manual How to Get Help

6.5.8. Top light

Figure 6-47、48、49
Disassembly:
1. Disconnect the AC power supply;
2. Set the main switch to “OFF”;
3. Remove the screws (circled) around the coping and then remove the coping.
4. Remove the four screws (circled) which fix the top light board on the top plate.
5. Replace the top light backlight board.
6. Record the version and serial number of the new top light board.
7. Reassembly and reinstall the top light board with the reverse procedure of disassembly.

216
Aeon8800A Service Manual Maintenance Information

6.5.9. Vaporizer Manifold

Figure 6-50、51
1. Remove the upper rear panel of machine.
2. Open the gas inlet integration module panel.
3. Pull out the two pipelines as the direction the arrows marked.
4. Remove the four screws (circled) with the socket wrench.
5. Pull out the vaporizer mount.
6. Replace the vaporizer mount.
7. Reassembly the vaporizer manifold with the reverse procedure of disassembly.

217
Aeon8800A Service Manual How to Get Help

7. How to Get Help

Aeonmed maintains a network composed of professional maintenance personnel who have


received factory trainings. Perform all the operation inspections to check appropriate control
settings before requesting repair. If operation problem still exists, please contact Aeonmed
distributor or service department for help.

Manufacture: Beijing Aeonmed Co., Ltd.


Address: 11B2, Feng tai Science Park, Beijing, China
P.C.: 100070
Tel: 010–83681616
Fax: 010–63718989
Service Hotline: 800-810-8333
Service apartment service@aeonmed.com
e-mail address:
Website: http://www.aeonmed.com

218