Professional Documents
Culture Documents
Statement
Beijing Aeonmed Co., Ltd. (Aeonmed for short) holds the copyrights to this manual, which is
non-public, published, and reserves the rights to keep it as a secure document. Refer to this
manual when operating, maintaining and repairing products only. Anyone other than Aeonmed
may not make it known to others.
Proprietary materials protected by the copyright law are included in this manual. Any section of it
cannot be reproduced, copied, or translated into other languages without any prior written
approval from Aeonmed who reserves the copyright.
Everything written in this manual is considered to be correct. Aeonmed is not legally responsible
for any mistakes printed within and any damages caused by incorrect installation and operation.
Aeonmed does not supply privileges endowed by the patent law to any other parties. Aeonmed is
not legally responsible for the results caused by patent law breaking and any rights of the third
party violating.
Refer to this manual before the product is used. The manual includes operating procedures which
must be performed with cautiously, operations that may result in non-normal working conditions
and the dangers which may damage equipment or cause bodily harm. Aeonmed is not
responsible for the security, reliability and function of the equipments in case that the dangers,
damages and non-normal phenomenon mentioned in this manual happen. Free repairs for these
malfunctions will not be provided by Aeonmed.
Aeonmed has the rights to replace any content in this manual without notice.
Manufacturer Responsibility:
Aeonmed is responsible for the security; reliability and function of the equipments when to
following conditions are adhered to:
i
Aeon8800A Anesthesia Machine User Manual Statement
CAUTION: Malfunctioning equipment may become invalid and cause bodily injury if a
set of effective and approving repairing proposals cannot be submitted by
the institution which is responsible for using this equipment.
The paid theoretical framework diagram will be supplied according to customer requirements by
Aeonmed, plus calibrating method and other information to help the customer, under the
assistance of qualified technicians, repair the equipment parts where can be done by customer
himself based on the stipulation by Aeonmed.
Warranty:
For a period of one year from the date of original delivery, the components and assemblies of this
product is warranted to be free from defects manufacturing techniques and materials, provided
that the same is properly operated under the conditions of normal use and regular maintenance.
The warranty period for other parts is three months. Expendable parts are not included. Our
obligation under the above warranties is limited to repairing free of charge.
Free Obligations:
Our obligation under the above warranties does not include the freight and other fees;
Aeonmed is not responsible for any direct, indirect or final product broken and delay
which result from improper use, alteration by using the assemblies unratified and
maintenance by anyone other than Aeonmed;
Improper use;
ii
Aeon8800A Anesthesia Machine User Manual Statement
Aeonmed is not responsible for the safety, reliability and operating condition of this product in
case that:
This product is not operated correctly in accordance with the manual instruction. The
power supply used or operating environment does not follow the requirements in this
manual.
Return
Follow the steps below in case that the product needs to be returned to Aeonmed:
Contact our customer service by informing them the number and type of the product. The number
is marked on the surface of the product. Return is unacceptable if the number cannot be identified.
Enclose a statement of the number, type and the reason of return as well.
2. Transportation charges
Transportation and insurance charges must be prepaid by the user for transporting the product to
Aeonmed for repairing. (Customers charges are added with regard to the products sold to
non-Chinese mainland users).
iii
Aeon8800A Anesthesia Machine User Manual Statement
Use Notice
To use this product correctly and effectively, the user must read this user manual carefully before
using this product.
When the user is using this product, this user manual must be understood totally and observe
strictly.
This product is only applied to the use this user manual described.
Only trained and professional service persons are allowed to repair and examine this product.
If there is any case happened in the process of use, please call us and intense service will be
provided.
If there is any change in product specification, another informs will not be given.
iv
Aeon8800A Anesthesia Machine User Manual Statement
Tel: +86-10-83681616
Fax: +86-10-63718989
Website: http://www.aeonmed.com
E-mail: service@aeonmed.com
v
Aeon8800A Anesthesia Machine User Manual CONTENTS
CONTENTS
1. Product Description .............................................................................................................................. 1-1
vi
Aeon8800A Anesthesia Machine User Manual CONTENTS
5.6.1.4. Synchronized Intermittent Mandatory Ventilation with Volume control (SIMV-VC) .......................... 5-15
5.6.1.6. Pressure control ventilation with Volume Guaranteed (PCV-VG) --Optional ...................................... 5-18
5.6.1.7. Synchronous Intermittent Mandatory Ventilation with Pressure Control (SIMV-PC) --Optional ........ 5-20
vii
Aeon8800A Anesthesia Machine User Manual CONTENTS
viii
Aeon8800A Anesthesia Machine User Manual CONTENTS
ix
Aeon8800A Anesthesia Machine User Manual CONTENTS
x
Aeon8800A Anesthesia Machine User Manual CONTENTS
xi
Aeon8800A Anesthesia Machine User Manual Product Description
1. Product Description
Aeon8800A Anesthesia Machine is a continuous flow anesthesia system which offers manual or
automatic ventilation, easily adjustable fresh gas delivery, anesthetic agent delivery, ventilation
monitoring, convenient ergonomics, and state-of-the-art safety systems.
Aeon8800A Anesthesia Machine is intended for use by licensed clinicians, for patients requiring
anesthesia within a hospital. This product is suitable for infants, children and adults.
1-1
Aeon8800A Anesthesia Machine User Manual Product Description
1.2. Symbols
The following table provides descriptions of symbols that are used on the device and/or within this
manual.
Power switch on ON
Dangerous Voltage
Earth (ground)
1-2
Aeon8800A Anesthesia Machine User Manual Product Description
Equipotential
Date of manufacture
Manufacturer
Catalogue number
Serial number
Lock
Unlock
Insp. flow
Exp. flow
1-3
Aeon8800A Anesthesia Machine User Manual Product Description
O2 flush O2+
Fuse
MANUAL/AUTO
Recycle
1-4
Aeon8800A Anesthesia Machine User Manual Product Description
Alarm icon
Type B equipment
Input, entrance
1-5
Aeon8800A Anesthesia Machine User Manual Product Description
1.3. Abbreviations
Abbreviation Definition
A
AGSS Anesthetic gas scavenging system
APL Adjustable pressure limiting
B
BDU Basic data unit
C
C Compliance
CGO Common gas outlet
cmH2O Centimeters of Water
CO2 Carbon oxide
D
Des. Desflurane
E
EEPROM Electrically erasable programmable read only memory
Enf. Enflurane
ETCO2 End-expiratory CO2 concentration
F
FiO2 Fraction of Inspired Oxygen
Flow-t Flow-time wave
Flow Flow
Freq Frequency
FreqMIN Minimum Frequency in PS mode
G
GUI Graphic user interface
H
Hal. Halothane
I
I:E Ratio of Aspiratory time to Expiratory Time
INSCO2 Inspiratory CO2 concentration
Iso. Isoflurane
IPPV Intermittent positive pressure ventilation
L
L/min Liter per minute
L Liter
M
MAC Minimum Alveolar Concentration
Manual Manual ventilation
1-6
Aeon8800A Anesthesia Machine User Manual Product Description
1-7
Aeon8800A Anesthesia Machine User Manual Product Description
Please read and adhere to all warnings, CAUTIONs and NOTICEs listed here and in the
appropriate areas throughout this manual.
WARNING: statement gives important information that, if ignored, could lead directly to
personal injury.
CAUTION: statement gives important information that, if ignored, could lead directly to
equipment, the user should carefully read this manual before operating:
WARNING: Do not use the machine if any test failures. Contact Aeonmed’s authorized
service representative for maintaining the machine.
1-8
Aeon8800A Anesthesia Machine User Manual Product Description
WARNING: Aeon8800A is not suitable for use in a magnetic resonance imaging (MRI)
environment.
WARNING: To avoid the risk of electric shock, this equipment must only be connected
to a supply main with protective earth.
WARNING: Breathing tube, mask, SUCTION filter and so on, adapter of mainstream
CO2 module, water trap of sidestream CO2 module etc. are disposable
accessories. Before using check the pipeline to ensure the disposal
subassembly is out of using .Check the tube which will be used before use
and avoid reusing.
WARNING: When the suction flow is insufficient, the suction port filter should be
checked or replaced.
WARNING: Do not use anti-static and/or electric breathing tubes or masks; otherwise,
WARNING: Do not incline the machine at the angle of more than 10 degree.
WARNING: Remove all equipments on the top cover before moving the machine. Use
WARNING: Ensure that the machine does not incline when it is upgrading,
1-9
Aeon8800A Anesthesia Machine User Manual Product Description
WARNING: Remove all the equipments attached at the flank of the machine before
WARNING: Apply the castor brakes when the anesthesia machine is in use.
WARNING: Do not put the item over 25 kg on the top board of Aeon8800A anesthesia
machine.
WARNING: Do not put the item over 12 kg on the work table of Aeon8800A anesthesia
machine.
WARNING: The whole width of machine should be less than 0.8m when hanging some
stuff.
WARNING: The medical gas pipeline system malfunction could cause one or more
applicable to the anesthetic machine that only uses cylinders for gas
supply.
WARNING: The parts of machine in contact with gas to be inhaled by patients don’t
to reproduction.
1-10
Aeon8800A Anesthesia Machine User Manual Product Description
enflurane and isoflurane, which are compatible to other gas are allowed to
use in anesthetic system.
WARNING: Possible fire hazard. Fuses (i.e., additional sockets) must only be replaced
WARNING: Possible electric shock hazard. The machine may only be opened by
WARNING: Electric shock and fire hazard. Do not clean the machine while it is on and
WARNING: Disconnect the power plug from the mains supply before removing the rear
WARNING: Malfunction of the central gas supply system may cause more than one or
WARNING: Use a cleaning and sterilizing schedule that confirms to your institution’s
A damaged oxygen sensor may lead to leakage of its highly caustic contents
that contain potassium chloride. Wear safety gloves and safety eye glasses
per manufacturer’s recommendations.
Do not inhale fumes that may result from any sterilization process.
1-11
Aeon8800A Anesthesia Machine User Manual Product Description
WARNING: Use extreme care while handling the absorbent as it is a caustic irritant.
WARNING: Use care in lifting and manipulating vaporizers during the mounting
process as their weight may be greater than expected, based on their size
and shape.
WARNING: All gas volume, flow and leakage specifications have been tested under
WARNING: After the device replacement is finished, all calibrations must be done
WARNING: If the anesthetic machine is equipped with Anesthetic Gas Delivery System,
WARNING: The medical device connected to the anesthesia system must conform to
CAUTION: For your safety and that of your patients, strictly follow this user manual.
CAUTION: Perform the tests specified on the Pre-use Preparation and, in case of a fault,
do not operate the machine until the fault has been corrected.
CAUTION: Before starting the machine, users must be familiar with the information
1-12
Aeon8800A Anesthesia Machine User Manual Product Description
CAUTION: If the machine does not function as described, it must be examined and
CAUTION: Handle the machine with care to prevent damage or functional faults.
CAUTION: Ensure that the gas supply of the machine always complies with the
technical specifications.
CAUTION: Before clinical use, the machine must be correctly calibrated and the
CAUTION: After servicing, all calibrations must be performed again and only the
WARNING: The USB interface of Aeon8800A anesthesia machine isn’t suitable for
CAUTION: The anesthesia ventilator used together with anesthesia system conforms
to ISO 80601-2-13:2011.
CAUTION: The anesthesia system is used together with the following monitoring
• Conform to the O2, CO2, and anesthetic gas monitor of ISO 80601-2-55:2011.
1-13
Aeon8800A Anesthesia Machine User Manual Product Description
1-14
Aeon8800A Anesthesia Machine User Manual Components
2. Components
2.1. Configuration
Touch screen
E-Flowmeter
Auxiliary flowmeter
Top light
Heater function
Configuration
Battery
Vaporizer(VP300/VP500)
External lamp
YOKE
2-1
Aeon8800A Anesthesia Machine User Manual Components
1
16
15
2 14
13
3
12
4 11
10
5
9
6 8
2-2
Aeon8800A Anesthesia Machine User Manual Components
Counterclockwise rotation of
flowmeter rotary knob increases
the flow; clockwise rotation of
flowmeter rotary knob decreases
the flow.
When flow meter is connected
with gas supply, rotate O2 or Air
knob, the digital screen displays
Electrical
4 data; there is O2 or Air flowing
flowmeter
out.
When O2 flow≥0.25L/min, turn
N2O knob, there is N2O flowing
out;
When O2≥0.5L/min, increase
2-3
Aeon8800A Anesthesia Machine User Manual Components
2-4
Aeon8800A Anesthesia Machine User Manual Components
WARNING: When
15 Spotlight adjusting the external lamp,
don’t touch the lamp surface
for 1s.
2-5
Aeon8800A Anesthesia Machine User Manual Components
2-6
Aeon8800A Anesthesia Machine User Manual Components
Auxiliary O2 output
3 WARNING: The requirement flow
port
of the device connected to the
auxiliary drive gas outlet interface
cannot exceed 90L/min.
2-7
Aeon8800A Anesthesia Machine User Manual Components
2-8
Aeon8800A Anesthesia Machine User Manual Components
WARNING: The
connection of electrical
equipment to the auxiliary
mains outlet socket may
Auxiliary mains reduce the system safety
7.
outlet socket level; therefore the
electrical equipment
connected to the auxiliary
mains outlet socket must
conform to the IEC
60601-1.
2-9
Aeon8800A Anesthesia Machine User Manual Components
CAUTION: Breathing system used with Aeon8800A complies with ISO 80601-2-13:2011:
2011.
Breathing system is mainly used to store fresh gas including anesthetic gas, oxygen, and absorb
waste gas. It directly connects to airway to support patient’s respiration.
2-10
Aeon8800A Anesthesia Machine User Manual Components
6 CO2
Absorber
It is used to absorb CO2 patient having exhaled.
Canisters
2-11
Aeon8800A Anesthesia Machine User Manual Components
NOTE:
Inspiratory and expiratory valve guarantee the single direction of gas flow in the breathing system.
When the anesthesia machine is working normally:
In the inspiratory phase, the inspiratory valve is acting and the expiratory valve is at rest; In the
expiratory phase, expiratory valve is acting and the inspiratory valve is at rest.
2. Applied Part of Aeon8800A anesthesia machine is the part between mask and Y-piece.
2-12
Aeon8800A Anesthesia Machine User Manual Components
2.5.1.APL Valve
WARNING: Keep all inlets/cables away from the APL valve, and do not put lines/
cables underneath the APL valve in order to adjust the APL valve
smoothly.
CAUTION: The APL valve is automatically excluded from the breathing system
APL valve has two functions; one is that it limits the maximum pressure during manual ventilation;
the other is that the airway pressure can be exhausted quickly by lifting the APL valve.
The APL valve has a labeled knob for selecting between 0cmH2O to 70cmH2O and for indicating
approximate pressure settings. Click sound will be heard when adjusting the APL valve. In manual
mode, the APL valve knob can be rotated to change the pressure threshold at which gas will flow
through the pressure threshold at which gas will flow through the valve and into AGSS. Clockwise
rotation of the APL valve knob increases the pressure threshold, and counterclockwise rotation of
the APL valve knob decreases the pressure threshold. Lifting the top of the APL valve knob will
temporarily relieve pressure. The leakage from the anesthetic breathing in all operational modes
when tested at pressure of 3.0kPa (30cmH2O) is 40ml.
2-13
Aeon8800A Anesthesia Machine User Manual Components
Fig.2-6 Vaporizer
CAUTION: The vaporizer used with Aeon8800A Anesthesia Machine complies with ISO
80601-2-13:2011.
CAUTION: For more operating and maintaining information of the vaporizer, refer to
User Manual.
2-14
Aeon8800A Anesthesia Machine User Manual Components
WARNING: Do not block the pressure relieve hole on top of the gas tank when using
AGSS.
CAUTION: The anesthetic gas scavenging system used with Aeon8800A Anesthetic
Machine complies with ISO 80601-2-13:2011.
CAUTION: If the AGSS used together with anesthesia system is produced by other
manufacturer, make sure that the AGSS conforms to ISO 80601-2-13:2011.
CAUTION: For more operating and maintaining information of AGSS, refer to AP1000
User Manual.
2-15
Aeon8800A Anesthesia Machine User Manual Components
An optional suction device is available to remove secretions, blood or other semi-liquid fluid from
the patient’s respiratory tract.
Before performing suction, adjust the negative pressure on the Suction device to an appropriate
value. When using the Suction device, connect the gas inlet to the Aeon8800A auxiliary gas
supply port. If the Suction device is connected with other gas supplies, the pressure of gas supply
should be less than 0.4MPa.
CAUTION: The negative pressure suction device used together with anesthesia system
conforms to EN ISO 10079-3.
CAUTION: For more information about operation and maintenance of the negative
pressure suction device, please refer to the Instruction for Use provided
with the negative pressure suction device.
2-16
Aeon8800A Anesthesia Machine User Manual Components
As shown in Figure 2-19 the mechanical device used to adjust flow control. The flow setting is
displayed to the left of the adjustment knob and also below the corresponding flow tube on the
display. The Total Flow Tube will display the sum of O2, N2O, and air flow. The flow controller is
equipped with a control valve for the ratio of O2 and N2O, so as to ensure that the O2 concentration
is no less than 21% when the user is using O2 and N2O at the same time. With no O2 supply
pressure, N2O is cut off. The effective range of the flow controller is 0~15L/min and the accuracy
is ±10%. When the value is out of the range, the flowmeter readings are for reference only.
2-17
Aeon8800A Anesthesia Machine User Manual Components
2.10. O2 flush
Figure 2-10
As shown in Figure2-10, by pressing the button, high flow of 35L/min--50L/min can be provided to
fresh gas outlet. When ―O2+‖ is pressed, between 35 L/min and 50 L/min of O2 flow will be
delivered to the Breathing System. Flush is available at all times when the machine is connected
to a Central Gas Supply System or Back-up Cylinder.
2-18
Aeon8800A Anesthesia Machine User Manual User interface
3. User interface
The panel of the anesthetic machine is composed of user interface, function keys and a knob. As
shown in figure 3-1.
3
2
No Name Description
1 Touch screen Touch screen.
The user interface can display real-time pressure and flow
waveforms, respiratory loop, monitoring data, Electronic Flowmeter
settings and anesthesia System and provides both sound and visual
alarms to assist the user.
2 Knob The user can set the parameter values by rotating the knob and
confirm set values by pressing the knob. Rotating the knob clockwise
increases values and rotating the knob counter-clockwise decreases
values.
3 MANUAL/ In Standby mode, first press MANUAL/AUTO knob, the system switch
AUTO to Manual mode, then secondly switch to IPPV mode.
In automatic mode, first press MANUAL/AUTO knob, the system
switch to Manual mode, then secondly switch to the original mode.
When entering or exiting the Manual mode, the user information area
displays according to the original model.
3-1
Aeon8800A Anesthesia Machine User Manual User interface
Gas: O2.
CAUTION: The anesthesia ventilator which is used with the anesthetic system complies
with ISO 80601-2-13:2011.
1 2 3
RS-232/VGA For connection of a VGA display which will display the same
2
interface information as the GUI screen.
3-2
Aeon8800A Anesthesia Machine User Manual User interface
The user interface is divided into eleven areas, see Figure 3-3.
3-3
Aeon8800A Anesthesia Machine User Manual User interface
3-4
Aeon8800A Anesthesia Machine User Manual User interface
3-5
Aeon8800A Anesthesia Machine User Manual User interface
3-6
Aeon8800A Anesthesia Machine User Manual User interface
Name description
Ventilation mode display: ventilation mode area shows current ventilation mode. The mode setup
menu will be open by clicking the ventilation mode area via touch screen.
3.2.1.2 There are five states for the mode buttons as shown below:
Focusing state - when rotating the knob to the purpose mode, the
White focus state - this state has been selected and then rotate knob
3-7
Aeon8800A Anesthesia Machine User Manual User interface
Parameter setting area is located at the bottom of the screen, with green/blue background and
round white buttons, which are applied for real-time respiratory parameter adjustment, including
six setting buttons.
3-8
Aeon8800A Anesthesia Machine User Manual User interface
The ventilation parameters of all models are arranged from left to right, as shown in
the chart below:
SIMV-PC
(Optional) ΔP Trigger Blank Blank Blank Back
(Optional)
ALARM
Manual Blank Blank Blank Blank Blank
on/off
3-9
Aeon8800A Anesthesia Machine User Manual User interface
2.The second status displays when the button has been selected by
the rotation of the knob;
3. The third status displays when the button has been pressed.
3-10
Aeon8800A Anesthesia Machine User Manual User interface
Alarm indication: when an alarm appears, an alarm bell icon shows in the warning area. For a
senior-level alarm, the alarm bell is red with white lettering. An intermediate-level alarm is yellow
with black lettering. A low-level alarm is blue-green with black lettering. At most, four alarms can
display on the screen at one time.
By pressing MUTE button, an ― ‖ shows on the alarm icon. Below the alarm bell a countdown
timer appears. The ― ‖ and countdown second display the same color as the alarm bell.
If there is an alarm, press alarm bell and the silence will keep for 120 seconds; press the key
again, and the silence is cancelled. If the alarm has a change but the silence less than 120s, the
silence will be cancelled.
The flickering frequency of higher priority alarm among all alarms indicates all alarms, alarm bell
and countdown seconds (if any). The alarm information text keeps own color respectively (if any).
3-11
Aeon8800A Anesthesia Machine User Manual User interface
Current date indicates at date display area, with format ―YYYY/MM/DD, MM/DD/YYYY or
DD/MM/YYYY‖.
Current time indicates at the time display area, with 12 hour or 24 hour format.
NOTE: Time and date format and value can be set in system setup menu.
Anesthesia machine should be able to automatically detect a USB device.
When the machine detects USB access, this icon shows. When the device is
unplugged or fails to detect USB, this icon does not show.
Power status includes AC power and backup battery state, which will be always
shown. When AC power is normal, the white plug icon shows consistently without
flashing. When net power is disconnected or fails, the white plug icon disappears.
3-12
Aeon8800A Anesthesia Machine User Manual User interface
Waveform display area is located in the center of the screen, including waveform and breathing
loop.
(CO2-t)
Breathing Loops
3-13
Aeon8800A Anesthesia Machine User Manual User interface
The patient message area is to the right of the alarm zone, and
to the left of the information area. In this area, there are two
icons: patient type and trigger type.
When patient triggers, the trigger logo will show in the patient
information area and disappear after 250ms. The background color
will restore the original color.
3-14
Aeon8800A Anesthesia Machine User Manual User interface
3-15
Aeon8800A Anesthesia Machine User Manual User interface
The electronic flow meter displays real-time N2O, air and O2 flow
values. Every flowmeter has 2 virtual flow tubes that display flow.
Each flow tube has the same icon, scale, structure and color (grey).
The units and numerical readings are displayed in white. The fill color
of virtual flow tubes is the same with gas icon color.
3-16
Aeon8800A Anesthesia Machine User Manual User interface
Message prompt area is in the upper left corner of the parameter settings area. Message prompt
area is used to display critical information or to provide real-time help to the user.
3-17
Aeon8800A Anesthesia Machine User Manual User interface
The shortcut key area includes shortcuts to alarm settings, system settings, calibration, trend,
breathing loop, self test, configuration and main interface.
Alarm settings: when user presses alarm setting key, alarm menu
appears. If this button is pressed in open state, alarm setting menu
closes, and returns to main interface.
Trend: by pressing trend key, the trend menu appears. If this button is
pressed in open state, trend menu disappears and returns to main
interface.
Loops: by pressing Loops key, the loops menu appears. If this button is
pressed in open state, the functional menu disappears and returns to
main interface.
3-18
Aeon8800A Anesthesia Machine User Manual User interface
3-19
Aeon8800A Anesthesia Machine User Manual User interface
3-20
Aeon8800A Anesthesia Machine User Manual Pre-Operative Preparation
4. Pre-operative Preparation
First step
Connect one end of the sampling gas tube to
the breathing system integration bottom;
connect the other end of sampling gas tube to
outlet module.
Fig-4-1
Second step:
Connect CGO connection cable of the
breathing system with CGO interface in the left
of anesthetic machine.
Fig.4-2
Third step:
Connect one end of heating cable with the
interface below the breathing system; the other
end connected with interface marked ―Heater‖
in the outlet module.
Fig.4-3
4-1
Aeon8800A Anesthesia Machine User Manual Pre-operative Preparation
Fourth step:
Connect one end of drive gas tube with the
interface below the breathing system.
Fig.4-4
Fifth step:
Connect the other end of drive gas tube with
interface marked ―Ventilator‖ in the outlet
module.
Fig.4-5
4-2
Aeon8800A Anesthesia Machine User Manual Pre-Operative Preparation
Step 1:
Hold breathing bag to bag arm directly.
Fig.4-6
Second step:
Connect manual bag to manual bag arm with an
upper force.
Fig.4-7
4-3
Aeon8800A Anesthesia Machine User Manual Pre-operative Preparation
First step:
Test plug is between inspiratory port and
expiratory port of the breathing system ―T‖ is
marked above the plug.
Fig.4-8
Second step:
Connect Y-piece to the Y-Piece seal.
Fig.4-9
4-4
Aeon8800A Anesthesia Machine User Manual Pre-Operative Preparation
4.1.4.Connection of O2 sensor
First step:
Pull out O2 cell plug from inspiratory valve cover
and then put O2 cell plug into the hole on the
bottom of breathing system integration.
Fig.4-10
Fig.4-11
Second step:
Connect the transparent joint which connects
O2 sensor by cable to the top port of O2 sensor.
Fig.4-12
Fig.4-13
4-5
Aeon8800A Anesthesia Machine User Manual Pre-operative Preparation
Third step:
Connect O2 sensor with inspiratory valve cover.
Fig.4-14
Fourth step:
Connect the other end of O2 sensor connecting
cable to the port marked ―Oxygen‖ of the outlet
module.
Fig.4-15
4-6
Aeon8800A Anesthesia Machine User Manual Pre-Operative Preparation
CAUTION: The anesthetic gas/CO2 monitoring module has mainstream and sidestream.
The two types can’t be used at the same time.
CAUTION: Please refer to the IRMA CO2/AX+ user guide, which is delivered together
with the IRMA CO2/AX+ sensor and include all the information about
operations and technical specifications of these two products.
Step 1:
Snap the probe on top of the airway adapter. It
will click into place when properly seated.
Fig.4-16
Fig.4-17
4-7
Aeon8800A Anesthesia Machine User Manual Pre-operative Preparation
Fig.4-18
Step 3:
Fig.4-20
Step 4:
Connect the other end of the adapter to the face
mask.
Fig.4-21
4-8
Aeon8800A Anesthesia Machine User Manual Pre-Operative Preparation
Step 1:
Fig.4-23
Step 3:
Connect the other end of the elbow to the
breathing bag, or to patients through tube.
Fig.4-24
Step4:
Connect the other end of the elbow to the
sampling tube, and make a connection between
the sampling tube and the CO2 or anesthesia
gas water trap.
Fig.4-25
Step 5:
Connect the other end of the sampling tube to
the gas inlet of Aeon8800A.
Fig4-26
4-9
Aeon8800A Anesthesia Machine User Manual Pre-operative Preparation
CAUTION: Make sure AP1000 installs at a place where operator can get a
clear view of it.
Fig.4-27
Step 2:
Hold AGSS with both hands ensuring the fixity
faces the wide slot. Slide AGSS horizontally to
rail till its limiting position.
Fig.4-28
4-10
Aeon8800A Anesthesia Machine User Manual Pre-Operative Preparation
Step 3:
Connect the two ends of screw thread to
EVAC port beneath breathing system and
AGSS respectively.
Fig.4-29
Step 4:
Connect exhaust pipeline to exhaust port of
AGSS, screw it clockwise tightly.
Fig.4-30
4-11
Aeon8800A Anesthesia Machine User Manual Pre-operative Preparation
equipment, the user should carefully read this manual before operating:
WARNING: Do not use the machine if any test failures. Contact Aeonmed’s authorized
4-12
Aeon8800A Anesthesia Machine User Manual Pre-Operative Preparation
WARNING: Make sure that the breathing system is in good connection and with no
damage.
3. Make sure the breathing system is correctly connected and has no damage; the absorber
canister is fully filled with soda lime.
4. Make sure the vaporizer is well locked and is filled with adequate anesthetic agent.
5. Make sure the pipeline gas supply system is correctly connected and is at proper pressure.
8. Make sure the castors are not loose and the forefront two castors are locked.
9. Make sure the anesthesia machine is connected to power socket and power indicator is light.
There is no power supply if power indicator does not light.
1. Turn on power switch; user interface is light, STANDBY interface displays after self-test is
finished.
2. After the anesthesia machine operates under STANDBY mode for five minutes, pull out
power cord
3. Make sure power failure alarm occurs (alarm bell sounds or user interface displays POWER
FAILUREURE message).
4-13
Aeon8800A Anesthesia Machine User Manual Pre-operative Preparation
CAUTION: Operator should make sure gas supplies are correctly connected during
WARNING: The malfunction of the center gas supply system may cause one or even all
1. Cut off pipeline gas supply. Perform the following steps if pipeline gauge does not read zero:
2. Connect O2 supply.
6. Make sure O2 cylinder gauge reading restores to zero. Low gas source alarm occurs when O2
pressure is decreasing.
4.2.3.Vaporizer Test
CAUTION: For information of vaporizer’s function test, refer to VP300 User Manual.
4-14
Aeon8800A Anesthesia Machine User Manual Pre-Operative Preparation
4.2.4.Alarm Test
4. Make sure:
6. Make sure:
Disassemble O2 sensor, make sure that O2 takes up 21% of room air according to
the test result.
4-15
Aeon8800A Anesthesia Machine User Manual Pre-operative Preparation
Put O2 sensor in pure oxygen; make sure that O2 takes up 100% according to the
test result.
Set low minute volume alarm limit to 10.0L/min, make sure low minute volume alarm
is eliminated.
Make sure continuous PAW alarm occurs when continuous pressure is at 10cmH2O
for 15 seconds.
4-16
Aeon8800A Anesthesia Machine User Manual Operating guide
5. Operating guide
5.1. Opening
WARNING: Make sure the assembly of patient breathing tube and controlling setting is
Fig.5-1
5-1
Aeon8800A Anesthesia Machine User Manual Operating guide
Fig.5-2
After the system self-test:
1) Select and confirm ―Continue‖, the
Automatic Test begins.
Fig.5-3
2) Select and confirms ―Skip‖, the
―Manual Test‖ interface appears.
Fig.5-4
5-2
Aeon8800A Anesthesia Machine User Manual Operating guide
Fig.5-5
Select ―Continue‖, the system
enter into the Automatic Test
interface.
―Automatic Test…‖appears in the
user interface.
Fig.5-6
5-3
Aeon8800A Anesthesia Machine User Manual Operating guide
2. Select ―Continue‖ and press the knob ,the system begins to re-test, The interface displays
shown as above picture.
Fig.5-7
2. If the bellows is not full, the
message ―Push ―O2+‖ flush to
completely fill the Bellows‖
appears on the user interface.
1) Select ―Continue‖ to perform
Manual Test again.
2) Select ―Skip‖ to proceed to
Manual Test.
Fig.5-8
5-4
Aeon8800A Anesthesia Machine User Manual Operating guide
Fig.5-9
2. If the leakage is at the range of
100—400ml/min, the interface
shows as the picture on the right.
Fig.5-10
3. If the leakage is more than
400ml/min, the interface shows as
the picture on the right.
Fig.5-11
5-5
Aeon8800A Anesthesia Machine User Manual Operating guide
Fig.5-12
5. Automatic Test failed
5-6
Aeon8800A Anesthesia Machine User Manual Operating guide
5-7
Aeon8800A Anesthesia Machine User Manual Operating guide
Fig.5-17
2. If the manual test fails for three
times, the interface displays as the
picture on the right.
Fig.5-18
3. If the Y-piece port and manual
bag arm connector are not occluded,
the interface after test is shown as
the picture on the right.
5-8
Aeon8800A Anesthesia Machine User Manual Operating guide
Fig.5-20
5.Manual test success
If the manual test succeeds, the
interface displays as shown on the
right.
The system will enter into Standby
mode automatically after 5s count
time.
Fig.5-21
5-9
Aeon8800A Anesthesia Machine User Manual Operating guide
The patient type of the Aeon8800A anesthesia system has 2 type options: Child and Adult. The
user can select the needed.
CAUTION: The changing of the patient type is only allowed in STANDBY mode
Fig.5-22
Press the knob to select the patient
type button in the bottom, left corner
of the screen. This button will show
yellow when selected.
Fig.5-23
5-10
Aeon8800A Anesthesia Machine User Manual Operating guide
Fig.5-24
Fig.5-25
5-11
Aeon8800A Anesthesia Machine User Manual Operating guide
Standard:
Optional:
In Manual mode, ventilation can be carried out by manual operation; there are ventilation
waveforms and monitoring values at present.
5-12
Aeon8800A Anesthesia Machine User Manual Operating guide
All the breathings are provided by the anesthesia machine, the patient cannot switch them by self.
The ventilation is managed by the predetermined ventilation volume, breathing frequency and
ratio of inspiratory time to expiratory time to keep the tidal volume stable, thereby assure the
stable MV.
5-13
Aeon8800A Anesthesia Machine User Manual Operating guide
Ventilation is managed according to a set goal pressure. Ventilation capacity (and velocity) is
changed accordingly, and related to respiratory system compliance and airway resistance. Tidal
volume changes with pulmonary compliance and airway resistance. Flow is slow-moving wave.
Pulmonary alveoli are full at the beginning of inspiratory which is helpful for the gas exchange in
the lung.
5-14
Aeon8800A Anesthesia Machine User Manual Operating guide
SIMV is a ventilation mode which combines spontaneous breathing with control ventilation. Within
the trigger window, the patient can trigger the command positive pressure ventilation
synchronizing with the spontaneous breathing. Between the two command ventilation cycle, the
patient is allowed to breath spontaneously. The command breathing follows the presets volume.
By setting frequency and tidal volume to assure the minimum MV, cooperating with the patient’s
spontaneous breathing to reduce the rivalry between patient and machine and negative effect of
blood dynamics, and prevents the potential syndrome, such as air pressure injury and so on.
By changing the presets frequency to change the breathing support level, in other words, from
entire support to partial support, the machine can be used for long-term patient to retreat from
machine.
5-15
Aeon8800A Anesthesia Machine User Manual Operating guide
5-16
Aeon8800A Anesthesia Machine User Manual Operating guide
Fig.5-29 Waveform of PS
5-17
Aeon8800A Anesthesia Machine User Manual Operating guide
In PCV-VG, breaths are controlled by the anesthesia machine (mandatory) and will be delivered
at a set rate and set volume (VT). Inspiratory pressure will be regulated to achieve the operator set
volume (Tidal Volume). The pressure range that the anesthesia machine will use is between the
PEEP + 2 cmH2O level on the low end and 5 cmH2O below Pmax on the high end. The inspiratory
pressure change between breaths is a maximum of +/- 3 cmH2O. PCV-VG begins by first
delivering a volume breath at the set tidal volume. The patient’s compliance is determined from
this volume breath and the inspiratory pressure level is then established for the next PCV-VG
breath. Inspiratory pressure from the last 3 breaths will be averaged to determine the pressure
needed to maintain the set tidal volume. Breath to breath inspiratory pressure change will be in
increments of 3 cmH2O up to PMAX - 5 cmH2O. When PMAX is reached the pressure is maintained
at the PMAX value until the end of inhalation. The inspiratory and expiratory valves will adjust to
maintain the PMAX value. When PMAX is reached, the next breath cycle will be normal PCV-VG
control and should not maintain PMAX.
5-18
Aeon8800A Anesthesia Machine User Manual Operating guide
5-19
Aeon8800A Anesthesia Machine User Manual Operating guide
In SIMV - PC breaths will be controlled by the anesthesia machine (mandatory) or triggered by the
patient (spontaneous). When controlled by the anesthesia machine, breaths will be pressure
limited and time cycled, resulting in an operator set pressure (PTARGET) being delivered for an
operator set period (TINSP). The anesthesia machine delivers the mandatory breath in synchrony
with the patient’s inspiratory effort. If no inspiratory effort is detected, the anesthesia machine will
deliver a mandatory breath at the scheduled time. Mandatory breaths will be PCV-type breaths
with decelerating flow waveform. Between mandatory breaths the patient will be able to breathe
spontaneously. These breaths can be pressure supported. When PMAX is reached the pressure is
maintained at the PMAX value until the end of inhalation. The inspiratory and expiratory valves
should adjust to maintain the PMAX value.
5-20
Aeon8800A Anesthesia Machine User Manual Operating guide
5-21
Aeon8800A Anesthesia Machine User Manual Operating guide
The user can select the ventilation mode according to the patient’s actual situation.
CAUTION: When ventilation mode is switched, the contact time of preselected state and
confirmation state need to be more than 1s.
Fig.5-32
If you wish to choose PS mode, click
the PS mode button. The button will be
at pre-selection state.
NOTE:
When the ventilation mode button is at
pre-selection state, the corresponding
ventilation parameters that can be set
appear on the bottom of the screen.
Fig.5-33
Click the PS button again to confirm
your selection.
Fig.5-34
5-22
Aeon8800A Anesthesia Machine User Manual Operating guide
Fig.5-35
Turn the knob to adjust the wanted
parameter value.
Fig.5-36
Click on ΔP button again the setting.
Fig.5-37
5-23
Aeon8800A Anesthesia Machine User Manual Operating guide
CAUTION: The menu window covers the second and third waveform.
CAUTION: When selected, a short cut key will display the selected menu. When Normal
is selected the all menus will be removed.
5-24
Aeon8800A Anesthesia Machine User Manual Operating guide
5.7.2.Alarm setting
The alarm settings can be opened by touching the Alarm shortcut key. The alarm setting menu
includes four sub-options: Vent, Gas, Agent, and Log. The default alarm sub-option on start-up is
Vent Alarm limits.
Vent:
Gas:
Fig.5-40
Agent:
Fig.5-41
5-25
Aeon8800A Anesthesia Machine User Manual Operating guide
Log:
Fig.5-43
5-26
Aeon8800A Anesthesia Machine User Manual Operating guide
5.7.3.System setting
System setting menu includes three sub-menus: Display, Info. and Setup. Display submenu is
default menu.
EFM RT Sens Mon: view the real-time monitoring value of the oxygen flow meter sensor, the N2O
flow meter sensor, the air flow meter sensor, and the pressure sensor.
RT Vent Mon: view the real-time monitoring value of the inspiratory volume, the expiratory volume,
the system resistant, the system compliance, and the tube compliance.
5-27
Aeon8800A Anesthesia Machine User Manual Operating guide
Fig.5-47
5-28
Aeon8800A Anesthesia Machine User Manual Operating guide
5.7.4.Calibration
CAUTION: Remove the oxygen sensor from the breathing system and expose the
sensor to room air for at least three minutes before proceeding.
Note: A plug is provided on a chain attached to the side of the breathing system for
plugging the O2 sensor port on the inspiratory valve during this operation.
CAUTION: The alarm “O2 Cal Due” will be displayed when the oxygen sensor has not
been calibrated for over 72 hours.
Fig.5-49
5-29
Aeon8800A Anesthesia Machine User Manual Operating guide
Fig.5-51
5-30
Aeon8800A Anesthesia Machine User Manual Operating guide
Fig.5-52
If gas module is activated, select ―Start‖
to calibrate.
Fig.5-53
If the calibration successes, the
interface will display ―Calibration
Successful‖, the running bar’s color will
be green, and the status indicator will be
right.
Fig.5-54
If the calibration fails, the interface will
display ―Calibration Failed‖, the running
bar’s color will be red, and the status
indicator will be red.
Fig.5-55
5-31
Aeon8800A Anesthesia Machine User Manual Operating guide
Fig.5-56
Select "Start", then calibrate touch screen
according to the instruction.
5-32
Aeon8800A Anesthesia Machine User Manual Operating guide
5.7.5.Trend
When the trend shortcut is selected, the trend data can be viewed. The machine will record the
last 8 hours of trend data at intervals of 30 seconds.
The trend time will appear on the left side of the menu, and the trend data will appear on the right
side of the menu, including Ppeak, Pplat, Pmean, PEEP, Freq, Vt, MV, FiO2, etCO2, FiCO2, Agent1,
Agent2, N2O, MAC, FG-O2, FG-AIR and FG-N2O.
Fig.5-58
The next page includes data from 7
parameters, as shown in Figure.
Fig.5-59
5-33
Aeon8800A Anesthesia Machine User Manual Operating guide
5.7.6.Check
In standby mode, when the ―Check‖ button is selected, the system test will be performed including
Leak test and Compliance Test.
Fig.5-60
Fig.5-61
If the Leak Test is success, the
interface is shown as the picture on
the right.
Select ―Finish‖ to perform
Compliance Leak
Fig.5-62
5-34
Aeon8800A Anesthesia Machine User Manual Operating guide
Fig.5-63
5-35
Aeon8800A Anesthesia Machine User Manual Operating guide
Fig.5-65
Fig.5-66
5-36
Aeon8800A Anesthesia Machine User Manual Operating guide
Fig.5-67
Fig.5-68
5-37
Aeon8800A Anesthesia Machine User Manual Operating guide
5.7.7.Configuration
Fig.5-69
Enter password ―2020‖ and click on
― ‖, the interface shown as the picture
on the right.
Fig.5-70
Select one ventilation mode, the
interface shown as the picture on the
right appears.
There are two sections in the menu, one
for adult and one for child. The user can
configure the default parameters for
each ventilation mode.
Settings entered will be used as the
default settings after unit power up.
Fig.5-71
"Load Factory Settings" button when
pressed , will change the default
settings to the factory defaults
5-38
Aeon8800A Anesthesia Machine User Manual Operating guide
Fig.5-72
Enter password ―2020‖ and click on
― ‖, the interface shown as the picture
on the right appears
Limits.
Fig.5-74
5-39
Aeon8800A Anesthesia Machine User Manual Operating guide
5.8. Shutdown
Turn OFF the gas supplies, until the gas in the system has gone out completely;
Turn the power switch to OFF to shut down the anesthesia system.
5-40
Aeon8800A Anesthesia Machine User Manual Alarm and troubleshooting
CAUTION: If alarm occurs, the patient’s safety must be protected firstly, then do
troubleshooting or necessary remedy.
CAUTION: When several alarms occur at the same time, alarm message only displays
the front two in a sequence of “High” to “Low”.
CAUTION: When alarm silencing, the alarm bell has dashed “X” on itself. After
120s, the alarm bell renews to the old state; If the alarm is not dealt in time,
alarm continues.
CAUTION: If alarm occurs when the anesthetic machine works normally, visible and
WARNING: Do not set alarm limit exceeding the extremes, otherwise, alarm system will
failure.
WARNING: A potential hazard can exist if different alarm presets are used for the same
or similar equipment in any single area, such as ICU or cardio tic operating
room.
WARNING: The operator should examine if the current alarm setting is suitable for every
patient.
WARNING: When power supply is interrupted, system will still renew the last alarm
settings when opening again.
6-1
Aeon8800A Anesthesia Machine User Manual Alarm and troubleshooting
The alarm states have 3 priorities: high, medium, low. High alarm needs to be dealt in time.
The operator can judge if the alarm system is functional by visible and audible alarm message.
The distance that the operator can hear the alarm sound or distinguish the priority of alarm is 4
meters at least. And the distance to see the alarm message is not less than 1 meter at the front of
the machine.
The top area of the user interface displays alarm message, see figure 6-1.
Table 6-1
6-2
Aeon8800A Anesthesia Machine User Manual Alarm and troubleshooting
The alarm method includes technical alarm and functional alarm; technical alarm includes startup
test list and alarm occurring during normal operation.
6-3
Aeon8800A Anesthesia Machine User Manual Alarm and troubleshooting
Table 6-2
Low gas supply High Drive gas pressure is less Check pipeline gas supply and
pressure than 2.9psi±15% replace pipeline.
Monitor O2
concentration≥10%,
cancel the alarm
immediately.
6-4
Aeon8800A Anesthesia Machine User Manual Alarm and troubleshooting
EFM High GUI receives and sends Switch to the manual mode
communication data from EFM unit failure
Call service Rep
lasts for 2s
failure
PSU High PSU send command to Switch to manual mode and call
communication GUI every 2secs, the service Rep.
failure alarm appears.
Gas module Low Pull out adapter probe in Call service Rep.
probe failure gas module for several
seconds.
CAUTION: If the gas
module communication
failures or interior error
in gas module, the alarm
message will not occur.
6-5
Aeon8800A Anesthesia Machine User Manual Alarm and troubleshooting
6-6
Aeon8800A Anesthesia Machine User Manual Alarm and troubleshooting
6-7
Aeon8800A Anesthesia Machine User Manual Alarm and troubleshooting
Table 6-3
6-8
Aeon8800A Anesthesia Machine User Manual Alarm and troubleshooting
6-9
Aeon8800A Anesthesia Machine User Manual Alarm and troubleshooting
No O2 sensor
calibration for 72 There is no calibration for O2 sensor for Recalibrate the O2
Low
72 hours. sensor.
hours
Bellows cannot Bellows cannot be full filled only in Push the ―O2+‖ button or
be full filled. High IPPV and SIMV-VC mode when adjust the flowmeter
ventilating. knob to increase flow.
6-10
Aeon8800A Anesthesia Machine User Manual Alarm and troubleshooting
6-11
Aeon8800A Anesthesia Machine User Manual Alarm and troubleshooting
6.2.4.Self-test alarm
Table 6-4
O2 supply sensor failure O2supply sensor failure Replace the O2supply pressure
board.
Air supply sensor failure Air supply sensor failure Replace the Air supply pressure
board.
O2supplyfailure Disconnection of O2 supply Check the connection of
O2supply.
Air supply failure Disconnection of O2 supply Check the connection of Air
supply.
Please close flowmeter The flowmeter is open. Close the flowmeter.
O2flowmeter failure O2flowmeter is invalid. Replace the E-flowmeter
assembly.
Air flowmeter failure Air flowmeter is invalid. Replace the E-flowmeter
assembly.
N2Oflowmeter failure N2Oflowmeter is invalid. Replace the E-flowmeter
assembly.
EFM pressure sensor EFM pressure sensor is Replace the E-flowmeter
failure invalid. assembly.
PSU communication PSU send command to GUI Call service Rep.
failure every 2secs, the alarm
appears.
Breathing system heating Breathing system heating Call service Rep.
failure function is invalid.
Breathing system Breathing system temperature Call service Rep.
temperature sensor failure sensor is invalid.
EFPROM chip failure EFPROM cannot be read or Call service Rep.
written.
PAW sensor failure The data of the pressure Call service Rep.
sensor is incorrect.
Ambient pressure sensor The data of the ambient Call service Rep.
failure pressure sensor is incorrect.
PEEP valve failure Incorrect PEEP valve data Call service Rep.
6-12
Aeon8800A Anesthesia Machine User Manual Alarm and troubleshooting
6-13
Aeon8800A Anesthesia Machine User Manual Alarm and troubleshooting
Table 6-5
2. Incorrect installation of
absorber canister.
4. Incorrect installation of O2
sensor or O2 sensor plug, or the
damage of O-ring of O2 sensor.
6-14
Aeon8800A Anesthesia Machine User Manual Cleaning and Disinfection
WARNING: Turn off Aeon8800A power switch before cleaning and disinfecting.
WARNING: Use a cleaning and sterilizing schedule that conforms to your institution’s
The O2 sensor may leak and burn (by Chlorine Potassium Oxide) if
damaged.
WARNING: Do not inhale any agents generated during cleaning and disinfecting.
CAUTION: User cleaning agent sparingly. Excess fluid could enter the machine,
causing damage.
CAUTION: Do not autoclave any parts of the machine unless specifically identified as
autoclave able in the User Manual. Clean the machine only as specified in
the User Manual.
Never use abrasive agents (i.e. steel wool or silver pool ish) to clean
components.
7-1
Aeon8800A Anesthesia Machine User Manual Cleaning and Disinfection
Do not immerse the bellows for more than 15 minutes as this may cause
expansion or accelerated aging.
All cleaning solutions used must have a pH between 7.0 and 10.5.
CAUTION: Never use the oxygen sensor or its connector in any type of liquid.
CAUTION: The valve disc is fragile and must, therefore, be handled with care while
removing the valve cage from the valve assembly.
CAUTION: If moisture remains in the bellows after cleaning, it may become tacky.
CAUTION: Prior to use after cleaning or disinfecting, power up the machine and follow
the on-screen prompts to perform the Leak Test and the Compliance Test.
CAUTION: The PAW gauge cannot withstand immersion or the heat and pressure of
autoclaving.
CAUTION: Peracetic acid and formaldehyde fumigation will damage external surfaces.
7-2
Aeon8800A Anesthesia Machine User Manual Cleaning and Disinfection
7.1. Guideline
Operator should have a better understanding of WARNINGS and CAUTIONS before cleaning
and disinfecting. Different methods will be used for different parts. Parts need to be completely
dried after cleaning and disinfecting.
After cleaning and disinfecting, the peruse calibration must be done before using after the
anesthesia machine is reinstalled, refer to ―Preoperative preparation‖ in Aeon8800A User Manual.
The recommended disinfection time of the machine is 100~300, according to the use condition
and sterilization time of anesthesia machine specifically.
ISO17664 conformity
Cleaning Information
Cleaning is supposed to do before using the breathing system for the first time after installation.
If there is potential infective object in the breathing system, such as blood or secretion, please
clean it with disposable cloth and authorized sterilant instantly.
Only the components marked 134℃ are heat-resistant and pressure-resistant parts that are
capable of withstanding autoclave sterilization, all parts expect O2 sensor and PAW gauge can be
washed with moderate sterilant in the automatic washing machine.
O2sensor and PAW gauge can be only cleaned on surface, O2sensor, cannot be autoclave
sterilized and cleaned with automatic washing machine.
7-3
Aeon8800A Anesthesia Machine User Manual Cleaning and Disinfection
Foundational Action
maintenance
every day clean anesthetic machine’s surface
every week Ventilate for the system, open flow meter to make floating-point move in
order to flow meter pipe block or floating-point conglutination.
every month Leak test of bellow assembly; 100% O2 concentration calibration.
Cleanness and Check if assembly is damaged, replace or repair if necessary.
installation
According to actual Open drain valve to replace soda lime.
situation Pipeline replacement: User replaces breathing pipeline in time according
to actual situation to ensure normal running (no gas leak in breathing
pipeline‖) after breathing pipeline is used for a certain time.
7-4
Aeon8800A Anesthesia Machine User Manual Cleaning and Disinfection
Fig.7-1,7-2
Take the heater cable from the outlet module.
Fig.7-3,7-4
Take the heater cable from the heater port in
the breathing system.
Fig.7-5,7-6
7-5
Aeon8800A Anesthesia Machine User Manual Cleaning and Disinfection
Fig.7-7,7-8
Take the screwed pipe from the breathing
system and store safely.
Fig.7-9,7-10
Take the integrated gas sampling pipe from the
outlet module.
Fig.7-11,7-12
One hand holds the bracket, the other hand
holds the other side of the breathing system,
and lift the breathing system vertically, take the
breathing system from the anesthesia machine.
Fig.7-13
7-6
Aeon8800A Anesthesia Machine User Manual Cleaning and Disinfection
Fig.7-14
Fig.7-15
Take PAW gauge from the breathing system
and store it safely.
Fig.7-16,7-17
7-7
Aeon8800A Anesthesia Machine User Manual Cleaning and Disinfection
Fig.7-18,7-19
Turn the bellows counterclockwise and take it
from the breathing system.
Fig.7-20
Take the bellows and store safely.
Fig.7-21
7-8
Aeon8800A Anesthesia Machine User Manual Cleaning and Disinfection
Fig.7-22
7-9
Aeon8800A Anesthesia Machine User Manual Cleaning and Disinfection
B: Washing Machine (Wet pasteurization at 70℃, 158℉, for 30 minutes after detergent cleaning)
D: Autoclaving (Including steam r hot air at 134℃, 273℉). Use your manufacturer’s or your
facility’s recommendations.
Table 7-1
7-10
Aeon8800A Anesthesia Machine User Manual Cleaning and Disinfection
1. Breathing Circuit
After using on every patient, clean breathing circuit with 70%~90% alcohol, wash with clean
water and dry, or place it in autoclave staircase for sterilization.
After using on every patient, clean silicon screw tube and breathing bag with clean water,
completely dry and put them in autoclave staircase for sterilization.
CAUTION: Do not use ultra-violet wave to sterilize silicon tube or breathing bag in case
of aging.
Either vaporing or immersion disinfection can be used in practice; in case of immersion all
sterilized parts must be dried with the high pressure air or oxygen before reuse.
4. Insp./Exp. Valve
Dismount the cover of the inspiratory and expiratory valves by rotating it counter-clockwise, then
takes away the valve patch, clean valve seat and circle together. And clean cover of the
inspiratory and expiratory valves and valve patch with the gauze soaked with water soluble
sterilizing agent, after all parts cleaned and dried recover it in original integration. Then one must
check the leakage and the movement of the inspiration and expiration valves in accordance with
the required regulation and checking procedure. Please handle all parts with care preventing any
damage.
WARNING: Take care and do not damage or crash valve disc during cleaning.
7-11
Aeon8800A Anesthesia Machine User Manual Cleaning and Disinfection
WARNING: Never separate the diaphragm and the valve seat in a pressure-relief valve.
CAUTION: Do not immerse them more than 15 minutes to prevent inflation or aging.
CAUTION: Dry by hanging while fully spread. If moisture is left in the bellows, they may
become tacky.
1) Disassembling.
2) To prevent component damage, clean them lightly. Put the recommended non-enzyme mild
agent used for latex and plastic in hot water.
4) Check the components if they are broken or damp, then perform the assembling and function
test.
CAUTION: For more cleaning and disinfecting information of AGSS, refer to AGSS
User Manual.
7-12
Aeon8800A Anesthesia Machine User Manual Regular maintenance
8. Regular maintenance
WARNING: In the process of removing product, strike and quiver must be avoided.
WARNING: Exhausted and old instruments (such as battery and LCD screen) harmful
8-1
Aeon8800A Anesthesia Machine User Manual Regular maintenance
This chapter provides message needed by Aeon8800A anesthetic machine in term maintenance.
Replace or adjust some components before doing some calibration or adjustment. If user uses
improper components such as improper specification causes malfunction, our company will not
take responsibility. Do not use malfunction device. Replacement and maintenance should be
finished by authorized service engineer or qualified, trained persons with repairmen experience.
After maintenance, the device should be tested and ensure its normal function and comply with
the manufacture’s regulation. Use components manufactured or sold by our company if
replacement is needed, then test and ensure device comply with the manufacture’s specification.
If service support is needed, please call local service engineer. In any case, the repairmen cost
will include the current price and reasonable personnel cost, but except repairmen items within
Aeonmed company’s warranty.
CAUTION: Person without repair experience about such device must not repair this
device.
8-2
Aeon8800A Anesthesia Machine User Manual Regular maintenance
Table 8-1
corresponding parts.
WARNING: The ineffective sodium lime has an obvious change in color when it’s
ineffective.
8.2.1.Appearance checklist
1. Assure the appearance of Aeon8800A no damage, the anesthesia machine can run
normally.
2. Assure absorbent canister install correctly and fill enough soda lime. Assure enough
anesthetic in vaporizer.
3. Assure the exhaust gas absorber pipe connected with APL valve no damage and dry
inside.
4. Assure O2, N2O, AIR, VAC and EVAC connection pipe no damage.
5. Assure O2, N2O, AIR, VAC and EVAC flexible pipe no damage (if they are be used).
6. Assure AC power cable no damage.
8-3
Aeon8800A Anesthesia Machine User Manual Regular maintenance
CAUTION: The life-span of all components above is the result in normal circumstance
8-4
Aeon8800A Anesthesia Machine User Manual Regular maintenance
WARNING: Exhausted and old O2 sensor must be treated according to the present
8.4.1.Replacement of O2 sensor
3. Connect new O2 sensor to cable connection port and install it on inspiratory valve of
breathing system.
manufacture.
O2 sensor can be used to detect partialO2 concentration of the anesthetic machine. O2sensor
belongs to consumptive products, so attentions should be paid to its useful-life.and use as the
characteristics and technical requirements provided by manufacture. Main technical requirements
of O2 sensor in Aeon8800A as follows:
8-5
Aeon8800A Anesthesia Machine User Manual Regular maintenance
Performance parameters:
Interface form and definition: FCC-68 4 core telephone pin (RJ11-4), see the following figure.
Resolution: 1mBar O2
Long-term output excursion: <5%(in 100% O2 circumstance and use more than one year)
8-6
Aeon8800A Anesthesia Machine User Manual Regular maintenance
ALARM: Disconnect AC power before fuse replacement, otherwise, it may cause injury
even death.
WARNING: When fuse replacement, fuse of the same type and size must be used,
CAUTION: Fuse belongs to damageable assembly, proper force must be used when
replacement.
CAUTION: Use screwdriver to swirl fuse and not use force largely and suddenly,
Replacement steps:
8-7
Aeon8800A Anesthesia Machine User Manual Regular maintenance
CAUTION: Only service engineer authorized by our company can replace battery. If the
anesthetic machine is not used in a long time, please call service engineer
to cut power supply connection.
CAUTION: The exhausted and old battery must be treated according to present local
regulations.
Battery configuration
Charging: when the AC power is connected, the system will charge the battery automatically,
the recommended charging time is no less than 8 hours.
Discharging: when the device uses battery to supply power, normally the work time is 120
minutes. When the battery’s electric capacity is insufficient, the device will display ―Low
power‖ alarm until power off. When the ―Low power‖ alarm occurs, the user should connect
AC power immediately and avoid the system closes (considering safety, manual startup
should be done to renew operation after the device closes automatically.)
Do not disassembly the battery causally and short battery connection line to avoid danger.
Battery’s Preservation
If the battery is not used for a long time, it should be preserved after the battery is charged
fully.
When the preservation time exceeds 3 months, charge the battery at least every 3 months.
Improper maintenance will cause battery broken, the battery should be replaced immediately
to avoid fluid leak erode device. Call manufacture when replace battery.
Battery’ removal
Use screw driver to remove the screws of the back cover board, disconnect the battery
module with device to remove battery.
8-8
Aeon8800A Anesthesia Machine User Manual Regular maintenance
Since installation, consumptive components names and code in every time dimension as the table
in below:
8-9
Aeon8800A Anesthesia Machine User Manual Technical Data
9. Technical Data
CAUTION: All technical specifications should be updated based on the real state of
CAUTION: All display values are measured under dry environment and proper
pressure.
WARNING: Cylinder valve should be closed during pipeline gas supply, so that gas
Pneumatic system plays the role of supplying O2, N2O, AIR and anesthetic agent to breathing
system, ensuring their pressures are within normal range respectively, so that normal operation of
anesthesia machine and patient’s safety can be guaranteed.
9-1
Aeon8800A Anesthesia Machine User Manual Technical Data
9-2
Aeon8800A Anesthesia Machine User Manual Technical Data
9-3
Aeon8800A Anesthesia Machine User Manual Technical Data
Gas Supply
Table 9-1
Flow Speed
Table 9-2
9-4
Aeon8800A Anesthesia Machine User Manual Technical Data
Aeon8800A Anesthesia Machine can be divided into three parts: UI, drive control and circuit. UI
consists of computer system, auxiliary control circuit, power and etc, which mainly takes charge of
display, alarm, keyboard input and interface expansion. Drive control has BDU as its core.
Exterior sensor, signal processing, amplifier, power and battery are part of drive control which
mainly used for data acquisition and ventilation mode control. UI and drive control should be
connected by cables.
Top light
Back light
Heater
9-5
Aeon8800A Anesthesia Machine User Manual Technical Data
CAUTION: Do not put heavy stuff on the top panel or in the drawers.
Table 9-3
Dimensions: 1403mm(H)×695mm(W)×855mm(D)
Weight: 110kg±5kg(not include vaporizer and
System
backup cylinder)
Top Panel Load: 25kg
Castor 125mm, each front castor with brake
Drawer 142mm(H)×376.5mm(W)×438mm(D)
Display 15’ TFT LCD
Cylinder Gauge ( AIR, O2, Range: 0MPa -1MPa. Stepping: 0.05MPa
N2O) Accuracy: ±2.5% of full range
Range: -10 cmH2O - 100cmH2O. Stepping: 200Pa
PAW Gauge
Accuracy: ±2.5% of full range
Device Lifetime 8 years
9-6
Aeon8800A Anesthesia Machine User Manual Technical Data
CAUTION: The anesthesia machine should be stored in draught room where there is
Table 9-4
Storage: 70 kPa~106kPa
Transport 70 kPa~106kPa
Power AC100-240V,50Hz/60Hz
9-7
Aeon8800A Anesthesia Machine User Manual Technical Data
Class I device
Common device/IP21
Traveling device
Continuous running
NOTE:
The applied Part of Aeon8800A anesthesia machine is the part between mask and Y-piece.
9-8
Aeon8800A Anesthesia Machine User Manual Technical Data
NOTE:
1. When the anesthesia system stops transferring, the pipeline supplying pressure should
be 280KPa to 600KPa.
2. When main power fails, the system switches to battery power automatically, and all the
3. When main power and internal battery both fails, the gas output is not affected.
9.6.1.Electrical
the connections between at least one medical device and one or more
non-electrical devices. Even if no functional connections between the single
devices, when using one auxiliary power outlet, they form a medical electrical
system. User must be aware of the risk of increased leakage current and
temperature and other increased risk may exist, when the device is connected
to the auxiliary socket-outlet.
WARNING: When earth wire has failure, if the device connects with auxiliary power supply,
WARNING: If the exterior protective grounded wire has questions during wiring, interior
9-9
Aeon8800A Anesthesia Machine User Manual Technical Data
Table 9-5
Supply Voltage 100-240VAC, 50/60Hz
Maximum Input Current 8A
AC power fuse F10AH 250VAC, φ5X20(F)
Auxiliary AC power plug 4
Auxiliary AC power plug fuse T2AH 250VAC, φ5X20(F)
Maximum output valve of auxiliary 1.5A(single plug); 6A(in total)
AC power plug
Grounded resistance Small than 0.2Ω
USB interface
Table 9-6
Number of USB Interfaces Standard: 2
9.6.2.Pneumatic
Table 9-7
9-10
Aeon8800A Anesthesia Machine User Manual Technical Data
Table 9-8
9-11
Aeon8800A Anesthesia Machine User Manual Technical Data
9.7.1.Compliance correction
In all of software version, the anesthesia machine is able to carry out a compliance correction
which compensates for volume "lost" in hoses and in the breathing system.
The anesthesia machine determines the system compliance (hoses and breathing system) and
required later for compliance correction during the self test or standby leak test. The hose of
standard equipped compliance has a fixed value of 11mL/kPa.
Having the system compliance (Csys) and ΔP, Anesthesia machine can calculate which volume
(ΔV) has been "consumed" by the system, or, in other words, which volume has not been
delivered to patient.
ΔV = Csys x ΔP
The anesthesia machine increases the volume to be delivered by the value ΔV.
VT corr = VT + ΔV
The anesthesia machine repeats this procedure during each breathing phase so that the V T
delivered to the patient gradually approaches the set VT. This procedure is completed as soon as
the plateau value stops changing (±0.5 cmH2O).The displayed flow and the resulting values (for
example, MV) are updated accordingly.
9-12
Aeon8800A Anesthesia Machine User Manual Technical Data
Table 9-9
9-13
Aeon8800A Anesthesia Machine User Manual Technical Data
9.8.2.Ventilation parameter
Table 9-10
9-14
Aeon8800A Anesthesia Machine User Manual Technical Data
9.8.3.Gas dynamics
Table 9-11
Label 1: Within rated outlet pressure range and under double maximum rated inlet pressure condition.
Table 9-12
According to ISO8835-2, test under peak flow 60L/min, fresh gas 10L/min.
9-15
Aeon8800A Anesthesia Machine User Manual Technical Data
9.8.4.Monitor performance
Table 9-13
Item Range Stepping Accuracy
Vt 0mL - 3000mL 1mL 20-100mL, error is ±20mL;
100-3000mL, error is ±20%
Freq. 0bpm - 110bpm 1bpm 1bpm
PEAK -20cmH2O - 99cmH2O 1cmH2O ±5%, at least1cmH2O
MEAN -20cmH2O - 99cmH2O 1cmH2O ±5%, at least1cmH2O
PLAT -20cmH2O - 99cmH2O 1cmH2O ±5%, at least1cmH2O
FiO2 18% - 100% 1% Error is ±3 % ( concentration
value)
C 0mL -100mL/cmH2O 1mL/cmH2 Error is 20% or 5mL/ cmH2O,
O choose the maximum one
PEEP -20cmH2O - 99cmH2O 1cmH2O 2cmH2O
PAW -20cmH2O -99cmH2O 1cmH2O Error is±3cmH2O or ±10%,
choose the maximum one
Battery 100%, 75%, 50%, 25%, 0%.
voltage When battery is used to supply power, this sign displays available remaining electric
state voltage.
display When the vehicle is connected with AC power, this sign means charging.
Pressure monitor range: 0cmH2O - 80cmH2O.According to different airway pressure,
gain the wave form displays is different.
0cmH2O ~ 10cmH2O, pressure axis gain: 5cmH2O
Paw-t
0cmH2O ~30cmH2O, pressure axis gain: 10cmH2O
wave form
0cmH2O ~80cmH2O, pressure axis gain: 20cmH2O
Time axis is a fixed range.(the axises of Flow-t, Paw-t is same.):
When gas module is opened, 0s ~15s;Gas module is closed, 0s ~20s.
Display range of flow rate: -90L - 90L/min, gain: 45L/min. On time axis, the positive
Flow-t
axes stands for inspiratory direction; under the time axis, the negative axes stands
Wave form
for expiratory direction. Flow rate is 0L/min, which means no gas flow rate in airway.
CO2 display range is 0mmHg - 76mmHg, gain is 38mmHg. Time axis is a fixed
CO2 time
range.
wave form
When gas module is opened, 0s ~15s;Gas module is closed, 0s ~20s.
Horizontal abscissa displays Paw: fixed range: -20cmH2O~ 80cmH2O, gain is
P/V Loop 20cmH2O.
Vertical abscissa displays tidal Volume: fixed range is 0ml~ 1500ml, gain is 250ml.
Horizontal abscissa displays tidal Volume: -1400mL ~ 0mL, gain is700mL.
Vertical abscissa stands for Flow: fixed range: -90L - 90L/min, gain is 45L/min; The
F/V Loop
positive axes stands for aspiratory flow rate, the negative axes stands for expiratory
flow.
9-16
Aeon8800A Anesthesia Machine User Manual Technical Data
Table 9-14
9-17
Aeon8800A Anesthesia Machine User Manual Technical Data
The gas monitoring module is used to connect other medical devices, provide monitor data,
monitor and display CO2, N2O and 5 anaesthetic gases (halothane, isoflurane, enflurane,
sevoflurane, desflurane).
The gas monitoring module is connected to the patient’s breathing pipeline to monitor the
inspiratory and expiratory gas and can be applied to Adult/Child and Infant.
Minimum alveolar concentration (MAC) is a standard for comparing the potency of inhalation
anesthetics. The MAC value represents the alveolar concentration of an anesthetic (at one
atmosphere) that, in 50 percent of a tested population, prevents gross muscular movement in
response to a painful, standardized stimulus.
If a mechanism to determine MAC values is implemented in the host device, the algorithms used
for this calculation must be adequately documented.
The MAC value may be calculated and displayed by using end-tidal (Et) gas concentrations
according to the following formula:
Note: Altitude, patient age and other individual factors are not considered in the
formula above.
9-18
Aeon8800A Anesthesia Machine User Manual Technical Data
The partial pressure and the volume percentage of CO2, N2O, O2 and anesthetic agents depend on the
amount of water vapor in the measured gas. The O2 measurement will be calibrated to show 20.8 vol%
at actual ambient temperature and humidity level, instead of showing actual partial pressure. 20.8 vol%
O2 corresponds to the actual O2 concentration in room air with 0.7vol% H2O concentration (at 1013 hPa
this equals for example 25°C and 23% RH). The measurement of CO2, N2O, and anesthetic agents (e.g.
all gases measured by the IR-bench) will always show the actual partial pressure at the current
humidity level.
In the alveoli of the patient, the breathing gas is saturated with water vapor at body temperature
(BTPS).
When the breathing gas is sampled, and passing the sampling line, the gas temperature will get close
to the ambient temperature before reaching the ISA sidestream gas analyzer. As the Nomoline
removed all condensed water, no water will reach the ISA gas analyzer. The relative humidity of the
sampled gas will be about 95%.
If CO2 values at BTPS are required, the following equation can be used:
Where:
EtCO2 = EtCO2 value sent from ISA [vol %]
Pamb = Ambient pressure sent from ISA [kPa]
3.8 = Typical partial pressure of water vapor condensed between patient circuit and ISA [kPa]
EtCO2(BTPS) = EtCO2 gas concentration at BTPS [vol%]
O2 is assumed to be room air calibrated at a humidity level of 0.7 vol% H2O.
Spectral broadening
The presence of oxygen and nitrous oxide can cause some interference in the CO2 measurement.
This is known as spectral broadening.
Nitrous oxide, N2O:
N2O compensation is not included in the system. Therefore, for the N2O broadening effect on CO2
measurement accuracy, please refer to the below table.
Oxygen, O2:
9-19
Aeon8800A Anesthesia Machine User Manual Technical Data
Gas concentration is reported in units of volume percent. The concentration is defined as:
The total pressure of the gas mixture is measured by a cuvette pressure sensor in the ISA gas
analyzer. For conversion to other units, the actual atmospheric pressure sent from the ISA
sidestream analyzer may be used, e.g.
CO2 in mmHg = (CO2 concentration) x (atm. pressure value in kPa from ISA) x (750 / 100).
Example: 5.0 vol% CO2 @ 101.3 kPa 0.05 x 101.3 x 750 / 100 = 38 mmHg
9-20
Aeon8800A Anesthesia Machine User Manual Technical Data
Table 9-15
Working principle of gas The gas module is a mainstream gas monitoring module, connecting to
module patient to monitor gas concentration of breath.
The gas module can be used for adults, children and infants.
9-21
Aeon8800A Anesthesia Machine User Manual Technical Data
Table 9-16
Basic Configuration
38×37×34mm(1.49″×1.45″× 1.34″)
Storage and
transportation
IRMA CO2: -40-75℃, -40-167℉
temperature
IRMA AX+: -20-75℃, -4-167℉
9-22
Aeon8800A Anesthesia Machine User Manual Technical Data
transport).
IRMA AX+:
4.5-5.5VDC, Max1.4W
Disposable infant:
NOTE 1: After being in a condensing atmosphere, this module shall be stored for more
The humidity range 50-100% is valid for the temperature range -40℃ - 40℃.
9-23
Aeon8800A Anesthesia Machine User Manual Technical Data
Table 9-17
Data Output
Fi and ET The newest breathing value (Fi and ET) is continually displayed after one
breath.
Table 9-18
Gas Analyzer
Probe 2-9 channel NDIR type gas analyzer measuring at 4-10 μm. Pressure,
temperature and full spectral interference correction.
9-24
Aeon8800A Anesthesia Machine User Manual Technical Data
Table 9-19
Table 9-20
Gas Accuracy
NOTE 1: The accuracy specification is valid for the operating temperature and humidity
NOTE 2: The accuracy specification for IRMA AX+ is not valid if more than two agents
9-25
Aeon8800A Anesthesia Machine User Manual Technical Data
Table 9-21
Metered dose Not for use with metered dose inhaler propellants
inhaler propellants3)
conditions” above.
9-26
Aeon8800A Anesthesia Machine User Manual Technical Data
NOTE 2: Interference at indicated gas level. For example, 50vol% helium typically
9-27
Aeon8800A Anesthesia Machine User Manual Technical Data
ISA samples gas from the respiratory circuit through the Nomoline Family sampling line at a rate
of 50sml/min, making measurements of CO2 possible for adult, pediatric and infant patients.
The Nomoline Family sampling lines incorporate a unique water separation (N2O Moisture)
section, which removes condensed water, The NOMO section is also fitted with a bacteria filter
that protects the gas analyzer from water intrusion and cross contamination.
As long as no sampling line is connected, the ISA gas analyzer remains in a low-power sleep
mode. Once the sampling line is connected, the ISA gas analyzer switches to measuring mode
and starts delivering gas data.
The Nomoline Family sampling lines are available in a wide variety of versions for both intubated
and spontaneously breathing patients and in both disposable and re-sposable configurations –
intubated patients can for instance be monitored using the disposable Nomoline Airway adapter
Set or a combination of the multiple patient use Nomoline Adapter and a disposable Nomoline
Extension / T-adapter. Spontaneously breathing patients could similarly be monitored using a
disposable Nomoline Nasal CO2 Nasal CO2 Cannula with Luer Connector.
The Nomoline Adapter may be used with other third party sampling line and canulas. Please
however note that the Nomoline Family or sampling line are designed for optimal performance
and measurement fidelity when used with the ISA gas analyzers. For instance, when connecting
to a respiratory circuit, the Masimo T-adapter provides a central gas sampling point thereby
minimizing the risk of sampling line occlusion
Note: Using sample tubes or cannulas with larger inner diameter than 1mm will increase ISA’s
total system response time.
For the Nomoline Family sampling line replacement, please call the service engineer authorized by our
company.
9-28
Aeon8800A Anesthesia Machine User Manual Technical Data
To set up the medical backboard device for gas analysis, follow these steps:
3. Connect a Nomoline Family sampling line to the ISA analyzer input connector.
5. Connect the gas sample exhaust port to a scavenging system or return the gas to the
patient circuit to prevent pollution of the operation room when N2O and/or anesthetic
agents are being used. Due to the risk of patient cross-infection, always use a bacteria
filter on the exhaust port side if sampled gas is intended to be re-breathed.
7. A green LED indicates that the ISA analyzer is ready for use.
Note: Returning the ISA’s exhaust gas to the patient circuit is not allowed in the USA.
1. Before connecting the Nomoline Family sampling line to the breathing circuit, do the
following:
3. Check that the gas inlet connector is lit with a steady green light.
4. Breathe briefly into the sampling line and check that the medical backboard device
displays a valid CO2 waveform and valid values.
5. Occlude the sampling line with a fingertip and wait for 10 seconds.
6. Check that an occlusion alarm is displayed and that the gas inlet connector shows a
flashing red light.
7. Perform a tightness check of the patient circuit with the sampling line attached.
9-29
Aeon8800A Anesthesia Machine User Manual Technical Data
9.10.4.3. Manitenance
For the ISA analyzer maintenance, please call the service engineer authorized by our company.
9.10.4.4. Zeroing
The gas analyzer needs from time to time to establish a zero reference level for the gas
measurements and the flow.
ISA performs zeroing by switching the gas sampling from the respiratory circuit to ambient air. The
automatic zeroing is performed 1 to 3 times per day, and takes less than 3 seconds for ISA CO2
gas analyzers and less than 10 seconds for ISA Multigas analyzers.
During zeroing, if ISA’s exhaust gas returned to the patient circuit, the returned gas level will be
different from the gas level at the sampling site.
9.10.4.5. Alarms
See User Manual for the medical backboard device for a complete description of the alarm
handling related to the ISA sidestream gas analyzer.
Indication Status
9-30
Aeon8800A Anesthesia Machine User Manual Technical Data
9.10.4.6. Cleaning
The ISA sidestream gas analyzers and the Nomoline Adapter may be cleaned using a cloth
moistened (not wet) with max 70% ethanol or isopropyl alcohol.
To prevent cleaning liquids and dust from entering the ISA gas analyzer throught its sampling gas
inlet connector, keep the Nomoline Family sampling line fitted while cleaning the analyzer.
Warning: Never sterilize or immerse the ISA sidestream gas analyzer in liquid.
Warning: The ISA sidestream gas analyzer is intended for use by authorized healthcare
professionals only.
Warning: Carefully route the sampling line to reduce the risk of patient entanglement or
strangulation.
Warning: Do not lift the ISA gas analyzer by the sampling line as it could disconnect
from the ISA, causing the ISA gas analyzer to fall on the patient.
Warning: Dispose Nomoline Family sampling lines in accordance with local regulations
for biohazardous waste.
Warning: Use only airway T-adapters with the sampling point in the center of the
adapter.
Warning: Do only sample lines intended for anesthetic agents if N2O and/or anesthetic
agents are being used.
Warning: Do not user T-adapter with infants, as this adds 7ml dead space to the patient
ciruit.
Warning: Do not use the ISA gas analyzer with metered-dose inhalers or nebulized
medications as this may clog the bacteria filter.
Warning: Since a successful zeroing requires the presence of ambient air(21% O2 and
0% CO2), ensure that the ISA is placed in a well ventilated place. Avoid
breathing near the ISA sidestream gas analyzer before or during the zeroing
procedure.
Warning: Never sterilize or immerse the ISA sidestream gas analyzer in liquid.
9-31
Aeon8800A Anesthesia Machine User Manual Technical Data
Warning: The ISA sidestream gas analyzer is intended only as an adjunct in patient
assessment. It must be used in conjunction with other assessments of
clinical signs and symptoms.
Warning: Replace the sampling line if the sampling line input connector starts flashing
red, or the medical backboard device displays a “Check sampling line”
message.
Warning: The ISA sidestream gas analyzer is not designed are not designed for MRI
environments.
Warning: During MRI scanning, ISA must be placed outside the MRI suite.
Warning: Do not apply negative pressure to remove condensed water from the
Nomoline Family sampling line.
Warning: Too strong positive or negative pressure in the patient circuit might affect the
sample flow.
Warning: Strong scavenging suction pressure might affect the sample flow.
Warning: Due to the risk of patient cross-infection, always use a bacteria filter on the
exhaust port side if sampled gas is intended to be re-breathed.
Warning: Do not place the ISA gas analyzer in any position that might cause it to fall on
the patient.
9-32
Aeon8800A Anesthesia Machine User Manual Technical Data
Warning: Do not re-use disposable single-patient use Nomoline Family sampling lines
due to the risk of cross contamination.
Warning: Do not operate the ISA sidestream gas analyzer if the enclosure is damaged.
Warning: Do not use the Nomoline Airway Adapter Set Infant with adult/pediatric
patients.
Caution: The ISA analyzers should be securely mounted in order to avoid the risk of
damage to the ISA.
Caution: Do not operate the ISA sidestream gas analyzer outside the specified
operating environment.
(US Only) Caution: Federal law restricts this device to sale by or on the order of a physician. For
professional use. See instructions for use full prescribing information, including indications
contraindications, warnings, precautions and adverse events.
Table 9-22
Working principle of gas The gas module is a sidestream gas monitoring module, connecting to
module patient to monitor gas concentration of breath by electrical machine
inhaling trace gas.
The gas module can be used for adults, children and infants.
9-33
Aeon8800A Anesthesia Machine User Manual Technical Data
Table 9-23
Basic Configuration
Weight 70g
(95%RH at 30℃)
ISA AX+:
9-34
Aeon8800A Anesthesia Machine User Manual Technical Data
NOTE 1: After condensation, the unit shall be stored for more than 24 hours in an
environment with relative moisture content below 95%RH (non-condensing).
Table 9-24
Data Output
ISA AX+: CO2, N2O, primary and secondary agents(HAL, ENF, ISO,
SEV, DES)
9-35
Aeon8800A Anesthesia Machine User Manual Technical Data
Table 9-25
Gas Analyzer
Sensor head 2-9 channel NDIR type gas analyzer measuring at 4μm -10μm.
ISA OR+/AX+
CO2 ≤300ms
HAL ≤500ms
9-36
Aeon8800A Anesthesia Machine User Manual Technical Data
Table 9-26
The following accuracy specifications are valid for dry, single gases at 22±5℃ and 1013±40 hPa.
15%-25% Unspecified
8%-25% Unspecified
10%-25%
22%-25%
Table 9-27
Gas Accuracy
NOTE 1: The accuracy specification is not valid if more than two agents are present in
the gas mixture. If more than two agents are present, an alarm will be set.
9-37
Aeon8800A Anesthesia Machine User Manual Technical Data
Table 9-28
Interference gas and vapor effects
Metered dose Not for use with metered dose inhaler propellants
4)
inhaler propellants
conditions” above.
9-38
Aeon8800A Anesthesia Machine User Manual Technical Data
NOTE 3: Interference at indicated gas level. For example, 50vol% helium typically
decreases the CO2 readings by 6%. This means that if measuring on a mixture
containing 5.0vol% CO2 and 50vol% helium, the measured CO2 concentration
will typically be (1-0.06)*5.0vol%=4.7vol% CO2.
9-39
Aeon8800A Anesthesia Machine User Manual Technical Data
Mainstream gas
2.4m Unshielded 1 Set DC Power
monitoring module cable
Sidestream gas
0.5m Unshielded 1 Set DC Power
monitoring module cable
Aeon8800A needs special precautions regarding EMC and put into service according to the EMC
information provided in the user manual; Aeon8800A conforms to this IEC 60601-1-2:2014
standard for both immunity and emissions. Nevertheless, special precautions need to be
observed:
The use of accessories, and cables other than those specified by AEONMED, with the
exception of accessories and cables sold by AEONMED of Aeon8800A as replacement parts
for internal components, may result in increased EMISSIONS or decreased IMMUNITY of the
Aeon8800A.
WARNING: Use of this equipment Aeon8800A adjacent to or stacked with other equipment
should be avoided because it could result in improper operation.‖
9-40
Aeon8800A Anesthesia Machine User Manual Technical Data
Table 1 - Emission
CISPR 11
RF emissions Professional healthcare facility environment
Group 1, Class A
IEC 61000-3-2
Harmonic distortion Professional healthcare facility environment
Class A
NOTE The EMISSIONS characteristics of this equipment make it suitable for use in industrial
areas and hospitals (CISPR 11 class A). If it is used in a residential environment (for which
CISPR 11 class B is normally required) this equipment might not offer adequate protection
to radio-frequency communication services. The user might need to take mitigation
measures, such as relocating or re-orienting the equipment.
9-41
Aeon8800A Anesthesia Machine User Manual Technical Data
9-42
Aeon8800A Anesthesia Machine User Manual Technical Data
Electrical fast ±2 kV
IEC 61000-4-4
transients/burst
100kHz repetition frequency
Surges
IEC 61000-4-5 ±0.5 kV, ±1 kV
Line-to-line
Surges
IEC 61000-4-5 ±0.5 kV, ±1 kV, ±2 kV
Line-to-ground
3V, 0.15MHz-80MHz
Conducted
disturbances induced IEC 61000-4-6 6V in ISM bands between 0.15MHz and 80MHz
by RF fields
80%AM at 1kHz
0% UT; 1 cycle
Voltage dips IEC 61000-4-11
and
Single phase: at 0º
Electrical fast ±1 kV
IEC 61000-4-4
transients/burst
100kHz repetition frequency
3V, 0.15MHz-80MHz
Conducted
disturbances induced IEC 61000-4-6 6V in ISM bands between 0.15MHz and 80MHz
by RF fields
80%AM at 1kHz
9-43
Aeon8800A Anesthesia Machine User Manual Technical Data
9-44
Aeon8800A Anesthesia Machine User Manual Technical Data
8. 230000075 Bag(3L) 3L
9-45
Aeon8800A Anesthesia Machine User Manual Technical Data
9-46
This manual No.: 130014952
Directive 93/42/EEC
Tel: +8610-83681616
Fax: +8610-63718989
E-Mail: service@aeonmed.com
Website: http://www.aeonmed.com
Edition 00.01
March 2019