Professional Documents
Culture Documents
User Manual
BK-200mini
Revision statement
This document applies to the latest and higher version of the software listed. If
the subsequent software version changes the information in this file, a new
version will be released.
Initial version: V1.0, 2019.2
Software version: M2.1.1.1
The purpose of creating this document is to improve the content and usability of
the user manual.
Intellectual property
The intellectual property of this user manual and its corresponding products
belongs to Shandong Biobase Biotech Co., Ltd. (hereinafter referred to as
―Company‖)
No individual or organization may reproduce (including photographing,
reprinting, transcription, etc.), copy, modify or translate any part of this manual
without the prior written consent of the company.
If the contents of this user manual are changed, the user will not be notified.
Statement
Shandong Biobase Biotech Co., Ltd (hereinafter referred to as ―Company‖) has
the final interpretation right of this user manual.
The company is responsible for the safety, reliability and performance of the
product when all of the following requirements are met:
Assembly operations, expansion, re-adjustment, improvement, repair and
replacement of parts are performed by professionals recognized by the
company.
All repairs involving replacement parts and supporting accessories and
consumables are original (original) or approved by the company.
The relevant electrical equipment complies with national standards and the
requirements of this user manual.
Product operation is carried out in accordance with this user manual.
User
The readers of this user manual are the following laboratory specialists.
Personnel who perform daily operations of the system.
Personnel performing system maintenance and troubleshooting.
Person who learns system operation.
Please read the contents of this user manual carefully before using the product
and use the product correctly. Please keep this user manual in a safe place so that
you can check it out at any time. If the precautions stated in this user manual are
not followed during use, no warranty will be given.
When the instrument is in the packaging status, take care to prevent rain and sun
exposure during transportation to prevent severe impact, heavy pressure and
dumping.
If the instrument has been unpacked and needs to be moved, repack the
instrument before shipping.
Storage
The packaged instrument should be stored at -10 ℃ ~ 40 ℃, relative humidity
is not more than 85%, no corrosive gas and well ventilated environment.
Contact information
Preface .............................................................................................................................................. I
Security information ......................................................................................................................... I
1 System Overview .......................................................................................................................... 2
1.1 Installation requirements and steps ..................................................................................... 3
1.1.1 Installation requirements .......................................................................................... 3
1.1.2 Instrument installation .............................................................................................. 6
1.2 Instrument structure ............................................................................................................ 7
1.2.1 Instrument appearance ............................................................................................. 8
1.2.2 Instrument cover internal structure .......................................................................... 9
1.2.3 Instrument movement structure .............................................................................. 10
1.2.4 Overall structure and function ................................................................................ 11
1.2.5 Reagent & Sample Processing System................................................................... 12
1.2.6 Reaction system ..................................................................................................... 18
1.2.7Washing system ...................................................................................................... 20
1.2.8 Optical Detection System ....................................................................................... 21
1.2.9 Mixing system ........................................................................................................ 22
1.2.10 Operation System ................................................................................................. 23
1.2.11 Data output ........................................................................................................... 23
1.2.12 Accessories&consumable .................................................................................... 23
1.2.13 Data manage software .......................................................................................... 23
1.3 Optional parts .................................................................................................................... 24
1.3.1 Printer ..................................................................................................................... 24
1.3.2 Water purifier ......................................................................................................... 24
1.3.3 Bar code reader installation .................................................................................... 24
1.3.4 IP address configuration ......................................................................................... 25
1.3.5 Power setting .......................................................................................................... 26
1.3.6 Power on ................................................................................................................ 29
1.4 Software introduction&operation...................................................................................... 29
1.4.1 Software interface .................................................................................................. 29
1.4.2 Mouse Usage .......................................................................................................... 33
1.5 System Parameter .............................................................................................................. 33
2 Basic Operation .......................................................................................................................... 36
2.1 Operate chart ....................................................................................................................... 2
2.2 Check before starting .......................................................................................................... 2
2.2.1 Water Checking........................................................................................................ 2
2.2.2 Power Checking ....................................................................................................... 2
2.2.3 Printer Paper checking ............................................................................................. 3
2.2.4 Waster Connection ................................................................................................... 3
2.2.5 Probe Checking ........................................................................................................ 3
2.3 Turn On Analyzer ............................................................................................................... 3
2.3.1 Turn On the power ................................................................................................... 3
2.3.2 Software ................................................................................................................... 4
2.4 Status Confirm .................................................................................................................... 6
2.5 Reagent Preparation ............................................................................................................ 6
2.5.1 Reagent preparation ................................................................................................. 7
2.6 Calibration........................................................................................................................... 9
2.6.1 Calibration Testing ................................................................................................. 10
2.6.2 Calibrator preparation ............................................................................................ 11
2.6.3 Calibration Start ..................................................................................................... 11
2.7 QC ..................................................................................................................................... 12
2.7.1 QC Testing ............................................................................................................. 12
2.7.2 QC Preparation ....................................................................................................... 12
2.7.3 QC Testing Start..................................................................................................... 13
2.8 Sample Testing .................................................................................................................. 13
2.8.1 Normal sample testing............................................................................................ 13
2.8.2 Sample Preparation ................................................................................................ 16
2.8.3 Start Testing ........................................................................................................... 17
2.9 Emergency Sample Testing(STAT) .................................................................................. 17
2.9.1 Emergency Sample Entry ....................................................................................... 17
2.9.2 Emergency Sample Preparation ............................................................................. 19
2.9.3 Testing Start ........................................................................................................... 20
2.10 Testing condition and control.......................................................................................... 20
2.11 Daily Maintenance .......................................................................................................... 22
2.12 Turn Off .......................................................................................................................... 22
2.13 Operation after turn off analyzer ..................................................................................... 23
3 System Setting .............................................................................................................................. 1
3.1System Setting...................................................................................................................... 2
3.1.1 Brief introduction ..................................................................................................... 2
3.1.2Sample information and testing information setting ............................................... 2
3.1.3 LIS settings ...................................................................................................................... 2
3.1.4 User Settings .................................................................................................................... 2
3.2 Item Setting ......................................................................................................................... 2
3.2.2 Parameter Setting ..................................................................................................... 3
3.2.3Judging parameter settings ........................................................................................ 6
3.2.4 Correction factor setting ........................................................................................... 7
3.3 Calibration Setting .............................................................................................................. 7
3.3.1 Introduction .............................................................................................................. 7
3.3.2 Edit calibrators, set calibrator concentration, set calibration mode .......................... 7
3.3.3 Delete calibration ................................................................................................ 9
3.4 QC setting ........................................................................................................................ 9
3.4.1 Introduction .......................................................................................................... 9
3.4.3 Item Selection ..................................................................................................... 10
3.4.4 Set the concentration parameters of QC products ............................................... 11
3.4.5 Set QC Rules ............................................................................................................ 11
3.4.6 Delete the QC ......................................................................................................... 12
4 Calculation Method.................................................................................................................... 13
4.1 Introduction ....................................................................................................................... 14
4.2 Analysis method ................................................................................................................ 14
4.2.1 Brief introduction ................................................................................................... 14
4.3 End point method .............................................................................................................. 15
4.3.1 Brief introduction ................................................................................................... 15
4.3.2 One point end point method ................................................................................... 16
4.3.3 Two point end point method .................................................................................. 16
4.4 Fixed time method ............................................................................................................ 18
4.4.1 Brief introduction ................................................................................................... 18
4.4.2 Calculation ............................................................................................................. 18
4.5 Rate method ...................................................................................................................... 20
4.5.1 Brief introduction ................................................................................................... 20
4.5.2 Calculation ............................................................................................................. 20
4.6 Calibration method ............................................................................................................ 21
4.6.1 Brief introduction ................................................................................................... 21
4.6.2 linear method.......................................................................................................... 21
4.6.3 Nonlinear method ................................................................................................... 24
4.7 Prozone check ................................................................................................................... 30
4.7.1 Brief introduction ................................................................................................... 30
4.7.2 Antigen addition method ........................................................................................ 30
4.7.3 Reaction rate ratio method ..................................................................................... 31
5 Reagent application.................................................................................................................... 32
5.1 Overview ........................................................................................................................... 33
5.1.1 Brief introduction ................................................................................................... 33
5.1.2 Reagent Information Interface Overview ............................................................... 33
5.2 Reagent margin alarm limit setting ................................................................................... 34
5.2.1 Brief introduction ................................................................................................... 34
5.2.2 Reagent margin alarm limit .................................................................................... 34
5.3 Reagent residue detection ................................................................................................. 34
5.3.1 Brief introduction ................................................................................................... 34
5.3.2 Detection reagent balance ...................................................................................... 34
5.3.3 Manually refresh the reagent balance..................................................................... 35
5.3.4 Cancel margin detection ......................................................................................... 36
5.3.5 Set the reagent balance to automatically refresh .................................................... 36
5.4 Print reagent information .................................................................................................. 36
5.4.1 Brief introduction ................................................................................................... 36
5.4.2 Print reagent information ....................................................................................... 36
5.5 Loading reagent................................................................................................................. 37
5.5.1 Brief introduction ................................................................................................... 37
5.5.2 Reagent registration ............................................................................................... 37
5.5.3 Loading reagent...................................................................................................... 38
5.6 Replacement reagent ......................................................................................................... 38
5.6.1 Brief introduction ................................................................................................... 38
5.6.2 Replacement reagent .............................................................................................. 39
5.7 Unloading reagent ............................................................................................................. 39
5.7.1 Brief introduction ................................................................................................... 39
5.7.2 Unloading biochemical reagents ............................................................................ 39
6 Calibration test ........................................................................................................................... 40
6.1 Overview ........................................................................................................................... 41
6.2 Calibration parameter setting ............................................................................................ 41
6.3 Calibration item directory ................................................................................................. 42
6.4 Calibration test .................................................................................................................. 44
6.5 Calibration result confirmation ......................................................................................... 44
6.5.1 View response curve .............................................................................................. 44
6.5.2 Query calibration result .......................................................................................... 46
7 QC test......................................................................................................................................... 47
7.1 Overview ............................................................................................................................. 2
7.2 QC rules setting ................................................................................................................... 2
7.2.1 Brief introduction ..................................................................................................... 2
7.2.2 QC rules setting ........................................................................................................ 2
7.3 QC information login .......................................................................................................... 5
7.4 Quality control item directory ............................................................................................. 6
7.5 Quality control test .............................................................................................................. 7
7.6 QC results confirmation ...................................................................................................... 8
7.6.1 View response curve ................................................................................................ 8
7.6.2 Query QC Result ...................................................................................................... 9
7.6.3 Analysis of out of control ....................................................................................... 11
8 Sample Test................................................................................................................................... 2
8.1 Overview ............................................................................................................................. 2
8.2 Sample Test Method ........................................................................................................... 2
8.2.1 Introduction .............................................................................................................. 2
8.2.2 Sample entry ............................................................................................................ 2
8.2.3 Additional Sample Test ............................................................................................ 3
8.2.4 Modify/ Add Testing Items ...................................................................................... 4
8.2.6 Sample Processing ................................................................................................... 6
Load the sample into the reagent&sample tray. ................................................................ 6
8.3 Cancel Sample Test ............................................................................................................. 7
8.4 Sample Test ......................................................................................................................... 7
9 Data Processing .......................................................................................................................... 12
9.1 Data Output ......................................................................................................................... 2
9.1.1 Introduction .............................................................................................................. 2
9.1.2 Derived Data to LIS Host ......................................................................................... 2
9.1.3 Data Backup ............................................................................................................. 3
9.3 Sample Report Printing ....................................................................................................... 4
9.3.1 Introduction .............................................................................................................. 4
9.3.2 Sample report .......................................................................................................... 4
9.6 QC Report Print .................................................................................................................. 7
9.6.1 Introduction .............................................................................................................. 7
9.6.2 QC Result ................................................................................................................. 7
10 Item Parameters ......................................................................................................................... 8
11 System Function ......................................................................................................................... 4
11.1 User and password settings ............................................................................................... 2
11.1.1 Introduction ............................................................................................................ 2
11.1.2 Adding Users.......................................................................................................... 2
11.1.3 Deleting Users ........................................................................................................ 2
11.1.4 Changing the password .......................................................................................... 3
12 How to Use LIS........................................................................................................................... 4
12.1 Overview ........................................................................................................................... 2
12.3 Sample test connected to LIS ............................................................................................ 2
12.3.1 Introduction ............................................................................................................ 2
12.3.2 Sample Testing Connected to LIS .......................................................................... 3
13 Maintenance ............................................................................................................................... 4
13.1 Overview ........................................................................................................................... 5
13.2 Regular Maintenance ........................................................................................................ 7
13.2.1 Introduction ............................................................................................................ 7
13.3 Daily Maintenance ............................................................................................................ 8
13.3.1 Check reagent&sample probe / stirrer / cleaning needle / cleaning tank ............... 8
13.3.2 Check pure water connection ............................................................................... 10
13.3.3 Check waste connection ....................................................................................... 10
13.4 Weekly Maintenance....................................................................................................... 12
13.4.1 Cleaning reagent&sample probe / cleaning needle / stirrer outer wall ................. 12
13.5 Monthly maintenance ................................................................................................... 14
13.5.1 Cleaning tank ....................................................................................................... 14
13.6 Maintenance every three months............................................................................. 15
13.7 Maintenance every six months ................................................................................ 17
13.8 Unscheduled maintenance ....................................................................................... 20
Steps ................................................................................................................................ 21
14.1 Troubleshooting Methods ....................................................................................... 34
14.1.1 Introduction .......................................................................................................... 34
14.1.2 Observing instrument failure prompt ................................................................... 34
Appendix D: Cross Contamination Reference Sheet ................................................................. 50
Security information
This chapter introduces the safety symbols used in the user manual and their
meanings. It summarizes the safety hazards and precautions when using the
instrument, as well as the labels and specific meanings attached to the
instrument, and lists the components included in the instrument. Whether the
content of toxic or hazardous substances or elements meets relevant
standards.
Safety symbol
Various safety symbols are used in this user manual andchemical analyzer to
remind you of the things you need to be aware of during operation. As shown
in the following table:
Symbol Sign language Description
Used for reagent&sample probes and waste drains.
Biological
indicates a risk of biological infection, and if not
infection risk
followed, there may be a risk of biological infection.
Used for power switch, 220V/110V power supply
Prevent electric
position. indicating the risk of electric shock, if
shock
contact, may cause personal injury.
Used for halogen lamp position. indicates a burn
Prevent burns hazard and may be burnt if contacted or not
followed.
Used to indicate a static-sensitive device or to
Electrostatic
indicate a device or connector that has not been
sensitive device
tested for antistatic.
Used for the position of moving parts such as sample
arm, mixing arm, cleaning mechanism, etc..
Prevent moving
indicating potential danger, the operator must be
parts
trained, if not in accordance with the instructions,
may cause personal injury.
For use of this instrument safely, please read the following safety precautions
carefully. Any operation that violates the following safety precautions may
result in personal injury or damage to the instrument.
Warning:
In all cases marked with this warning sign, the user manual is required to
clarify the nature of the potential hazard and any countermeasures that must
be taken. If you do not follow the instructions in this user manual, the
protective measures provided by this instrument may be invalid.
Prevent burns:
Do not touch the light source after the system is turned on.
When replacing the halogen lamp, the lamp must be replaced after the
power is turned off, otherwise the high temperature halogen lamp and
the light source box may cause burns.
Waste treatment:
Some substances in reagents, controls, calibrators, cleaning fluids, and
waste liquids are subject to pollution regulations and emission standards.
Please comply with local emission standards and consult the relevant
reagent manufacturer or distributor.
When handling waste, be sure to wear gloves, wear overalls to prevent
infection, and wear protective glasses if necessary.
Prevent fires and explosions:
Alcohol is flammable and must be used with great care.
Processing instrument:
Some substances in discarded instruments are subject to pollution regulations.
Dispose of used instruments in accordance with local waste disposal
standards.
Operational precautions
Instrument use
Warning:
This instrument is used to quantitatively analyze the clinical chemistry of
serum, plasma, urine, cerebrospinal fluid and other samples.
When considering clinical results based on the results of the analysis, please
consider clinical symptoms or other test results.
Use environment
Be careful:
The electromagnetic environment should be evaluated before operating
the equipment.
Please install the instrument correctly according to the installation
environment specified in the user manual. Do not place the device in a
location where it is difficult to operate the disconnect device. Installation
and use of this instrument outside of the specified conditions may result
in unreliable results and may result in damage to the instrument.
If you need to change the system status, please contact the company's
customer service center or distributor in your area.
System installation
Warning:
This product is a permanently connected device that uses switches and circuit
breakers as disconnect devices. In order not to affect the normal operation of
the instrument, before installation, ensure that the building in the installation
location is equipped with a switch or circuit breaker that meets the
requirements of GB4793.1-2007 as the disconnect device. The switch or
circuit breaker in the building should clearly indicate that the switch or
circuit breaker is the disconnect device of the instrument.
Be careful:
Do not place equipment that emits abnormal noise near the instrument.
Please turn off the equipment that emits electromagnetic waves, such as
cell phones, radio transceivers, etc., in the room where the instrument is
located, and do not use other CRT monitors near the instrument. Noise
and electromagnetic interference may cause instrument malfunction.
Do not use other medical instruments near this instrument.
Electromagnetic waves emitted by this instrument may cause
malfunction of other medical instruments in the vicinity.
Instrument use
Be careful:
Please follow the instructions in the user manual to use the instrument.
Improper use may result in inaccurate measurements and may even
result in damage to the instrument or personal injury.
Before using the instrument for the first time, please set the calibration
and then carry out QC to confirm that the instrument is working
properly.
When using the instrument, QC procedures must be performed,
otherwise the reliability of the results cannot be guaranteed.
Do not open the sample/reagent cover during the analysis
The network port of the analyzer is set to be connected to the network
port of the computer. Do not connect to cables other than any other
device. Please use the dedicated cable provided by the company to
connect the analyzer to the computer.
The computer is a platform for operating the instrument-specific
operating software. Installing any software or hardware other than the
company's designated content on this computer may prevent the
instrument from functioning properly. Do not run other software while
the instrument is in operation.
Computer viruses can destroy software and data. Please do not use your
computer for other purposes or to connect to the Internet.
Do not touch your computer's monitor, mouse, or keyboard with wet or
chemically-friendly hands.
Do not turn the power switch back on within 10 seconds of turning off
the analyzer's total power, otherwise the instrument may enter protection.
If the instrument enters the protection status, please turn it off and then
turn it on again.
Instrument maintenance:
Be careful:
Follow the instructions in the user manual to maintain the instrument.
Improper maintenance may result in incorrect analysis and may even
result in damage to the instrument or personal injury.
The instrument may be placed for a long time and may have dust on the
surface. When cleaning, use a clean soft cloth soaked in water, wring it
out gently, and if necessary, dip a small amount of soap. Do not use
organic solvents such as alcohol. After cleaning, dry the surface with a
dry cloth.
Before cleaning, please turn off all power to the instrument and unplug
the power cord. During the cleaning process, take necessary measures to
prevent water droplets from entering the instrument, otherwise the
instrument may be damaged or personal injury.
Check the main components, such as replacement of halogen lamps,
reagent&sample probes, stirrer, and injection components, and
calibration analysis must be performed.
If the instrument needs to be repaired due to malfunction, please contact
company customer service center. During maintenance, the instrument
may need to be taken out of service or transported. Please be careful to
avoid biological hazards, electric shock hazards, and moving parts
hazards due to maintenance.
When replacing the light source lamp, it is necessary to wait for more
than 20 minutes after the power is turned off to perform the lamp
replacement operation. Otherwise, the high temperature light source
lamp and the light source box may cause burns.
Parameter settings
Be careful:
The instrument needs to set parameters such as sample size, reagent amount,
and measurement wavelength. When setting these parameters, please follow
the instructions in the user manual and refer to the instructions provided with
the reagents.
Sample
Be careful:
Use a complete serum sample and a urine sample that does not contain
suspended solids. If the serum sample contains fibrin, or if the urine
sample contains insoluble impurities such as suspended solids, it may
block the reagent&sample probe and affect the analysis results.
Drugs, anticoagulants, preservatives, etc. present in the sample may
interfere with certain analytical results
Hemolysis, jaundice, chylomicron, etc. in the sample may affect the
analysis results. It is recommended to do a sample blank test.
Please use the correct sample storage measures. Improper sample storage
measures may alter the composition of the sample and result in incorrect
analysis results.
To prevent the sample from evaporating, do not leave the sample open
for a long time. If the sample evaporates, it may result in incorrect
analysis results.
Some samples may not be analyzed based on test parameters and
reagents used. For these samples, please consult the reagent
manufacturer or distributor and the distributor of Shandong Biobase
Biotech Co., Ltd
The sampling quantity is required for the analysis of this instrument.
When sampling, determine the appropriate sample size according to the
instructions in this user manual.
Before analyzing, please confirm that the sample is placed in the correct
sample position, otherwise you may not get the correct result.
Be careful:
When using this instrument for analysis, you need appropriate reagents,
calibrators, and controls.
Please select the matching reagent according to this instrument. If you
are not sure if the reagent is available, please consult your company or
company's distributor.
Reagents, calibrators, use and storage of QC products, etc., please follow
the instructions of the relevant reagent manufacturer or distributor.
If the reagents, calibrators, and controls are not properly stored, even
within the validity period, the correct test results and the best instrument
performance may not be obtained.
After checking the reagents, please perform calibration analysis. Without
calibration analysis, the correct analysis results may not be obtained.
During analysis, cross-contamination of reagents may affect the results
of the analysis. Consult reagent manufacturers or companies for reagent
cross-contamination information.
Data backup
Note:
The instrument has the function of automatically storing data on the
computer's hard disk, but the computer hard disk data is deleted or the hard
disk is damaged due to other reasons, which may result in data being
unrecoverable. Please periodically back up the analysis data and
measurement parameters to other mobile storage devices.
Note:
Please refer to the instructions for use of the computer and printer.
1 System Overview
System Overview-2
System Overview
UPS
Outlet
The connecting line in the figure indicates the corresponding pipe or wire.
In order to facilitate the operation, maintenance and repair of the instrument, the
Auto chemical analyzer must meet the following conditions during installation:
The distance between the left and right sides of the instrument and the wall
should be no less than 500mm.
The distance between the rear panel and the wall of the instrument should be
no less than 500mm.
The distance between the front of the instrument and other instruments
should be no less than 1000mm.
Guarantee the space for the waste liquid device and the pure water supply
device during installation.
Minimum distance 500mm
Computer
425
Analyzer
Minimum distance 1000mm
Front
mm
System Overview-4
System Overview
Figure 1-3
Warning:
Protective grounding must be good to prevent electric shock and instrument
failure
Warning:
The accuracy of normal operation and test data cannot be guaranteed for the
instrument in environments other than those described above. If the
System Overview-5
System Overview
temperature and humidity do not meet the above requirements, please use air
conditioning equipment and humidification equipment.
The instrument generates heat during operation and is discharged through the
rear of the instrument. The working environment should be well ventilated
and ventilation equipment should be used if necessary. However, airflow
should be avoided to blow directly to the instrument, otherwise it may affect
the accuracy of the instrument test.
Warning:
The water quality must meet the water supply requirements. Otherwise,
insufficient water purity may affect the test results.
Ensure that the water supply hole and pipeline installation of the instrument
should be kept unobstructed. In addition, the inlet of the L-frame of the
instrument should be higher than the pure water bucket, and the height difference
from the liquid level of the feed water should be less than 50cm.
Ensure that the drainage hole and pipeline installation of the instrument should be
kept unobstructed, and the outlet of the instrument L frame should be higher than
the waste liquid barrel mouth (or special discharge port for waste liquid), and the
length of the waste pipe should not exceed 2m.
Warning:
In order to ensure the normal operation of the instrument after installation,
the installation and initial setting of the instrument , pls contact us.
System Overview-6
System Overview
After the instrument arrives, please check the packaging of the instrument
carefully to see if there is physical damage. If there is any damage, please contact
BIOBASE agent . After confirming that there is no external damage, follow the
steps below to unpack:
Erect the arrow on the box upright.
Open the accessory box and check whether the object is complete according
to the packing list. If there is any missing, please contact BIOBASE
company or local agent
Check the appearance of the instrument carefully. If there is any damage,
please contact BIOBASE company directly.
2 4
1
5
10
3
6 7 8
11 13 12
14
1
15
1 16
System Overview-8
System Overview
2 Observation window Here you can observe the working conditions of the
internal sample loading system, detection unit, etc.
4 Right side panel The power switch is located on the right side panel
5 Power Board Turn the power and power cord connections on or off
14 Aviation joint Connect pure water and waste liquid float level switch
to realize pure water and waste liquid alarm
15 Outlet Connect the snake skin tube and drain the waste liquid
5 2
2 4
System Overview-9
System Overview
Figure 1-8
4 Mixing system Stir the reagent and sample mixture in the cuvette
5 Reaction system The reaction cup is fixed and the inside is kept at a
fixed temperature to provide reaction conditions
1 2 3 4 5 6 7 8 9
10 11 12 13 14 15 16
Figure 1-9
System Overview-10
System Overview
Analyzer
Figure 1-10
System Overview-11
System Overview
Figure 1-11
1-Sample position.2-Reagent circle
The reagent&sample tray is divided into sample position and reagent circle, and
the sample position is 37,The standard cup containing the sample, calibrator, and
control can be placed in the set position,the reagent&sample tray is rotated and
sent to the sample arm for the reagent and sample position.
2
3
Figure 1-12
System Overview-12
System Overview
1
3
2 4
Figure 1-13
1-Reagent&sample tray positioning plate.2-motor. 3-Reagent&sample disk axis,
4-Reagent&sample tray fixed base.
2 3
Figure 1-14
1-Reagent&sample tray, 2-Reagent&sample tray circle-, 3-Reagent&sample tray upper
ring bracket, 4-Reagent&sample tray handle
System Overview-13
System Overview
click “Y” to execute the detection. "Return to zero" means returning to the
initial position.
Reagent incubator
Semiconductor
refrigeration chip
Heat sink
Cooling fan
Figure 1-15
The semiconductor refrigerating sheet has two faces, hot and cold. When
installing, pay attention to distinguishing different faces.
The main component of the refrigeration control system is the reagent&sample
tray. The sample reagent tray is kept at a low temperature for 24 hours through a
cooling tray, a cooling fan, and a heat sink.
System Overview-14
System Overview
2
7
3
8
4
9
10
11
12
5
Figure 1-17
1-Reagent&sample probe cover.2-Ball spline shaft.3-Sample arm main frame.4-Left and
right stepper motor.5-Adapter plate.6-Upper and lower stepper
motor.7-Reagent&sample probe-.8-Left and right timing belt.9-Up and down timing
belt.10-Bearing compression sleeve.11-Pipe clamp.12-Move up and down
Sample arm has the function of adding sample and adding reagent.
The sample loading function of the sample arm is to take a set amount of sample
from the sample cup and then fill it into the reaction cup. The sample is sampled
at a dose of 2μl to 70μl with an accuracy of 0.1μl. At the same time of
sampling, the reagent&sample probe will open the liquid level detection function
System Overview-15
System Overview
to detect a sample size of 50μl, and the minimum test sample size is 50μl or
more.
The reagent loading function of the sample arm is to take a set amount of reagent
from the reagent bottle and add it to the reaction cup. At the same time, the
reagent&sample probe opens the liquid level sensing function. The remaining
amount of the reagent is calculated by the distance the reagent&sample probe is
dropped, and the remaining amount of the reagent is displayed in the "Reagent
Information" window. As shown in figure 1-18.
Figure 1-18
During reagent sampling, to prevent the reagent from being diluted, at each time
aspirate the reagent. Reagent&sample probes are drawn according to the "set
amount + balance", only the set amount of reagent when adding reagent to the
reaction cup. The reagent setting amount ranges from 20 μl to 350 μl and can
be set in units of 1μl.
The liquid level detection function of the reagent&sample probe can detect the
remaining amount of the sample or reagent. Auto collision protection is activated
when a similar collision occurs during the descent. The reagent&sample probe
returns to the highest position in the vertical direction and no longer drops.
Anti-collision protection can only be resumed by power-off, and then reset after
power-on reset. After restarting, it is necessary to re-adjust the mechanical
position of the reagent&sample probe, and thoroughly check the instrument
without any problem before starting the test again.
When performing the sample test, the reagent&sample probe is sequentially
moved in the order of "sample cup (or reagent bottle), reaction cup,
reagent&sample probe cleaning tank".
Warning:
System Overview-16
System Overview
When the system is running, do not place any hand or other parts of the body
on the path of the sample arm or place any obstacles on the path of the
sample arm. Failure to do so may result in personal injury or system damage.
Up fixed screw
Spring
Adapter sleeve
Anti-collision base
System Overview-17
System Overview
may cause aspiration error. If the sample size is less than the dead volume,
transfer the sample to a small sample tube before testing. The minimum sample
size of the sample tube is the sample size required for the test.
The dead volume of each sample tube is shown in the following table.
Sample container Specification Dead volume
Sample cup Φ12×37mm, 2ml 50ul
Original blood collection Φ12×68.5 mm Higher than the
tube/Plastic test tube Φ12.7×75 mm unavailable sample
Φ13×75 mm 8mm
System Overview-18
System Overview
5
1
3
4
1-Reaction cuvette position, 2-Identification film, 3-Gland, 4-Cuvette joint, 5-Cuvette
Figure 1-23 Top cover assembly
System Overview-19
System Overview
1 5
2
3 6
1-Photoelectric switch, 2-Big Gear, 3-Fixed Base, 4-Axis, 5-Small Gear, 6-Step Motor
Figure 1-25 Base part of reaction tray
As figure 1-22, 1-23, 1-24, 1-25 display, reaction tray included up cover, tray,
step motor, drive motor, axis fix part, fiber fix board., etc.
After the power on, the reaction tray rotates counterclockwise, and the probe is
rotated until the reagent&sample probe can swing to the position of the reaction
cup to the 1st position.
The cuvette can be recycled for holding the reaction solution. After each test, the
system Automatically 4-steps cleans and dries the cuvette for the next test.
1.2.7Washing system
1 6
7
8
2 9
3
4 10
11
5
12
1-Washing Probe, 2-Linear guide slider mount, 3-Photoelectric switch, 4-Step motor,
5-Base of washing arm, 6-Compression screw for washing probe,
7-Fixed axis of washing probe, 8-Driven wheel, 9-Driven belt,
10-Up and down wheel, 11-Motor fixed board, 12-Fixed column.
Figure 1-26 Structure of Washing System
System Overview-20
System Overview
When the power is turned on, the cleaning arm first rises to the zero position,
then vertically descends to the reaction cup, cleans the reaction cup, and after
rising, rises and stops above the reaction cup.
The reaction cup automatic cleaning system uses a cleaning agent and deionized
water to perform a 4-step automatic cleaning of the reaction cup to ensure that the
reaction cup is free from cross-contamination and drying during the test. The
process of 4-step cleaning is as follows:
First, second and third step cleaning: wash the cuvette with deionized water and
blot the cuvette
Fourth step cleaning: wiping the reaction cup
After the washing is completed, the washing waste liquid is discharged through
two stages: high concentration waste liquid and low concentration waste liquid.
The system supports the high-concentration cleaning waste liquid level detection
function. When the waste liquid of the high-concentration waste liquid tank is
detected exceeds a certain amount, an alarm is issued, prompting to empty the
waste liquid tank.
For system checking, click the ―Instrument check‖, and select the ―Cleaning
Arm‖ –―Lift Y‖ function to check it the cleaning needle in right positions
Cuvettes
Optical signal
Light source detection
Figure 1-27 Optical Detection System
The optical system consists of a light source, a colorimetric system, and a
spectroscopic component to provide monochromatic light of sufficient intensity
and a stable and reliable colorimetric optical path structure.
System Overview-21
System Overview
1
6
2
8
3
4 9
5
1-Probe Cover, 2-Stirrer, 3-Up&Down Base.4-Up&Down Pulley, 5-Main Support,
6-Ball Spline Shaft, 7-Left&Right Pulley, 8-Left&Right Motor, 9-Up&Down Motor
Figure 1-28 Structure of mixing arm
System Overview-22
System Overview
After the power is turned on, the stirrer rises vertically, swings left and right,
stops above the cleaning position, and then vertically descends to the cleaning
tank for cleaning. After completion, it rises and stops above the cleaning tank.
The initialization process is the same as the motion status after power-on.
During the sample test, the mixer arm is vertically lowered to the cleaning
position for cleaning, and then swinged to the corresponding reaction cup of the
reaction tray to stir the solution.
1.2.12 Accessories&consumable
Accessories and consumables refer to the components necessary for the
instrument to perform the sample test. Always check to ensure that the quantity is
sufficient and replenish and replace if necessary. To ensure personal safety and to
ensure system performance, please use accessories and consumables
manufactured or recommended by the company. If necessary, please contact our
customer service department or distributor in your area.
For details on accessories and consumables, refer to "13.1.2 Accessories
Information" (page 2).
1.3.1 Printer
Used to print test results and other data. The instrument supports any of inkjet
printers and laser printers. The printer is not a standard configuration. If you need
to choose the option, please contact our customer service department. If you need
to purchase it separately, please make sure to purchase the printer that meets the
requirements.
Before using the operating software for printing, check as follows:
1) Check whether the printer driver is installed.
2) Check whether the data cable connection between the printer and the
analyzer is correct.
3) Verify the printer put into the appropriate printing paper. Switch on the
printer, and start printing.
System Overview-24
System Overview
BIOBASE series auto chemistry analyzer and scanner are all used the USB
interface.
System Overview-25
System Overview
System Overview-26
System Overview
System Overview-27
System Overview
System Overview-28
System Overview
1.3.6 Power on
After installation, install the pure water float switch and washing solution float
switch, so that the floats to keep floating status. Connected to the following
power:
Total power supply and analysis unit power on the right panel of the analyzer,
computer and monitor power supply, the printer power supply.
About 20 minutes after turning on the power supply (Wait for the temperature
and light source stability), instrument into standby mode.
△! Note:
When the instrument is installed for the first time, please perform the water pipe
exhaust 20 times before use to ensure that the air bubbles in the pipe are drained.
Stand by
Workspace
Menu bar
System Overview-29
System Overview
According to the function selected by the user, the interface window of the
corresponding function will appear. For example, if the item entry button is
clicked in the menu bar, the ‗sample entry‘ software interface shown in Figure
1-41 is displayed.
System Overview-31
System Overview
Data Processing Export data Export test data and other information
BK-
Data Maintenance Maintain print report data and other’s
200mini
Auto Test Results Correction Query the test results and correct the results
Chemistry
Analyzer Check the information of various parts, and
Instrument check
troubleshoot
Adjustment Adjust the information of mechanical position
Wash && Background Clean the cup and read the background information
Sample Status Display the status, location, and type of the sample
System Overview-32
System Overview
Wavelength 340~800nm
Accuracy ±1nm
Basic
Reaction Temp 37℃±0.1℃
System Overview-33
System Overview
Performance Standard
Sample Position 37
Reagent Position 28
Cuvette Qty 48
System Overview-34
System Overview
Performance Standard
Net Weight 36Kg
Installatio
n&Storag Storage: -10℃~40℃, <85%RH,
e Storage&Using Working:15℃~30℃,35%RH~80%RH,
Altitude: less than 2000m
System Overview-35
2 Basic Operation
This chapter describes the basic operation methods and daily operation
procedures of the instrument, including the following steps:
Check before starting
Turn on analyzer
Confirm instrument status
Loading reagent
Calibration application and testing
QC application and testing
Regular sample application and testing
Emergency sample application and testing
Test status and test control
Maintenance
Shutdown
Operation after shutdown
Basic operation method
1. Check the high-concentration waste liquid tank and check whether the waste
liquid in the barrel is empty. If not empty, empty the waste container.
2. Check the connection of low-concentration waste liquid to ensure that the
waste liquid pipe is not bent, and the drain discharge port of the sewer is not
higher than the waste liquid outlet of the instrument.
2.3.2 Software
1. Click the conductor of software, when your first time running this software, it
will require software configuration.
Click conductor again, and the login the software, original user name and
password is 1000
Note:
If the normal privileged user password is forgotten, you can log in to the system
as an advanced privileged user, delete the username, and then reset it. or contact
Biobase after sale engineer.
3.After the correct login and power-on detection are normal, the main interface of
the operation software is displayed. At this point, the boot process ends. If the
environment is found to be inconsistent during the boot detection process, a
message will pop up. Please take corresponding measures according to the
information displayed on the interface.
After the boot, the instrument acts as follows:
1. The stirring arm is lifted and the cleaning arm is lifted.
2. Stir and swing back and forth to the cleaning position.
3. The reaction disk is reset to zero.
4. Repeat steps 1 and 2 once.
5. The sample arm is lifted and swings left and right.
6. The reagent&sample disc rotates clockwise after returning to zero.
7. Repeat steps 5 and 6 once, after which the instrument enters the standby status.
Note:
After the instrument is powered on and enters the standby status, each
position is not in the zero position. You need to enter the software and click
the initialization command to perform an initialization operation before you
can correspond!
To ensure accurate test results, please display “Standby” in the system status
area and turn it on for approximately twenty minutes before starting the test
operation to ensure stable light source and temperature control.
For the first time, please first exhaust the water pipe 20 times to ensure that
the air bubbles in the pipe are drained.
Warning
Note:
Note:
1. Confirm the system status and perform reagent loading operations according to
different status.
Standby: go directly to the next step.
Test: Wait for all items to be tested, and then load the reagents when the
instrument enters standby mode.
2. Select [Reagent] - [Reagent Information].
3. Select the location where the reagent needs to be loaded.
4. Enter the loaded reagent information in the [Reagent Information Edit] dialog
box, including:
Bar code
Project name
Reagent type
Bottle size
5. Click [Add] to save the input reagent information.
6. Click on the blank location of the other reagents list to load reagents for other
items on the instrument.
7. Open the reagent tray cover.
8. Control the reagent loading list, put each reagent into the corresponding
position on the reagent&sample tray, and then open the reagent bottle cap.
9. Cover the reagent tray cover.
10. Refresh the reagent remaining amount information.
a)Close system reagent loading
Click the REAGENT in software interface, and then display reagent information
menu.as Figure2-8. Scan the reagent barcode with the scanner as reagent tray no.
and click ‗Add‘.
2.6 Calibration
The calibration test is used to calculate the calibration parameters to participate in
the calculation of the sample results. In general, a calibration test is recommended
when any of the following conditions occur:
Create a new project.
When the reagents, calibrators and controls are still within the validity period, the
QC test generates an alarm.
Replace the reagent lot number or bottle number.
The project exceeded the calibration validity period.
Modified calibration rules including: calibration method, number of repetitions,
calibrator concentration, and calibrators used.
The light source lamp, syringe, reagent&sample probe, etc. were replaced.
If the following parameters are modified, they must be calibrated:
Main wavelength
Secondary wavelength
Blank time
Reaction time
Amount of reagent
Sample size, diluted sample size, and dilution amount corresponding to the
standard amount
Analytical method
Sample type
Reaction direction
Sample blank and result unit
For the method of setting the calibration, refer to "3.3 Calibration Setting" (Scale
Setting - page 6)
Be careful:
After the power is turned on, the light source lamp needs to be stable for 20
minutes, and the test can be started when the instrument enters the standby
status.
After applying the calibration test and properly placing the required calibrator,
you can start the calibration test.
1. Click [Set Current Calibration] on the calibration parameter page.
Basic operation method-11
Basic operation method
2.7 QC
QC results are an important tool for monitoring whether the instrument is
performing well. In order to judge whether the instrument test performance is
stable, it is recommended to conduct QC tests every day. Manual QC testing is
possible. The QC product is allowed to add new items under any status. When the
QC item is not tested, the QC information can be modified in the test status, only
the additional items are allowed, and the QC information is not allowed to be
modified.
2.7.1 QC Testing
It is allowed to apply for QC test according to the QC product. You can choose
the QC product, the QC product position and the sample cup type. You must
select at least one item, otherwise you will not be allowed to save the application.
1. Select [Program Input] - [ Calibration&QC Input] - [QC Input]
2.7.2 QC Preparation
Improper use of the calibrator may result in infection. Do not touch the
calibrator directly with your hands. Always wear gloves, work clothes to
prevent infection, and wear protective glasses if necessary. If the calibrator
comes into contact with the skin, please follow the user's working standards
immediately and consult a doctor.
Be careful:
After the power is turned on, the light source lamp needs to be stable for 20
minutes, and the test can be started when the instrument enters the standby
status.
After applying for the QC test and correctly placing the required QC, you can
start the QC test.
After clicking [Start] on the page toolbar, the following prompt box will pop up.
After confirming that the number of QC items is correct, click [Start] to start the
test.
10. The number of times the sample needs to be repeated in [Repeat Times].
The input range is from 1 to 90, and the default is one.
11. If an item in the sample has an application request that is different from the
other items, enter the following information:
[Sample size]: Select the sample size required for the project test. This sample
size type is consistent with the sample size type defined when setting up the
project. If the incremental sample size and the reduced sample size are defined,
the increment or decrement can be selected here. Otherwise only the standard
amount is allowed to be selected.
[Number of repetitions]: Enter the number of repeated tests for the item.
[Sample blank]: Set whether to perform sample blank test separately for the
project.
Click [Save].
Be careful:
Note:
Before loading the sample, make sure there are no bubbles in the sample tube
to avoid inaccurate test results.
1. Click [Data Processing] - [Report Print].
2. After entering the interface, the sample corresponding to the sample is placed
in the reagent&sample tray according to the sample number and the patient
information.
After the power is turned on, the light source lamp needs to be stable for 20
minutes, and the test can be started when the instrument enters the standby
status.
After the sample is completed, the temperature is stable. After all other test
conditions are set, click the start button in the toolbar. The pop-up dialog
box is shown in Figure 2-17. Click ‗Start‘ to enter the sample test process.
a) b)
Figure 2-17 Sample Testing Start
1. Click [Item Entry] on the toolbar, enter the [Sample Entry] form, and select
[Add] on the right side.
2. In the [Test Items] column on the left, select the items to be tested, click to
select the ―Selected Test Items" column, and then fill in the patient's letter ‗name‘,
‗gender‘, ‗age‘ "Wait, edit the information of the "submission department",
"sample type‘, "send the doctor", etc. After the login is completed, click [Save] to
complete the operation. Enter the sample number in [No.].
The number can consist of numbers or letters and numbers, is not case sensitive,
and cannot exceed 10 digits in length.
3. Enter the sample position, the emergency sample cannot be set with the sample
number that is duplicated with the sample being tested.
4. In the following order, fill in the patient's letter "name", "gender", "age", etc.,
and edit the information of "sent inspection department", "send inspection doctor"
and so on.
5. Select the type of sample in the [Sample Type] drop-down list.
Options include: serum, plasma, urine, cerebrospinal fluid and others.
6. Select the sample size to be drawn in the [Sample Size] drop-down list in the
sample attribute area. The options include standard, incremental and decrement.
7. Set whether a sample blank test is required.
8. Select the type of sample cup to use in the [Sample Cup] drop-down list.
Options include micro cups and standard cups.
9. Choose whether to [dilute].
10. Select the [Emergency] check box.
11. The number of times the sample needs to be repeated in [Repeat Times].
The input range is from 1 to 90, and the default is one.
12. If an item in the sample has an application request that is different from the
other items, enter the following information:
[Sample size]: Select the sample size required for the project test. This sample
size type is consistent with the sample size type defined when setting up the
project. If the incremental sample size and the reduced sample size are defined,
the increment or decrement can be selected here. otherwise only the standard
amount is allowed to be selected.
[Number of repetitions]: Enter the number of repeated tests for the item.
[Sample blank]: Set whether to perform sample blank test separately for the
project.
Click [Save].
1. Click [Item Entry] on the toolbar, enter the [Sample Entry] form, and select
[Batch Entry] at the bottom right of the interface.
2. Enter the number of the first emergency sample in [Sample No.].
3. Select the bit number of the first emergency sample in [Sample No.].
4. Enter the number of emergency samples for batch testing in [Quantity].
5. Select the type of sample in the [Sample Type] drop-down list.
6. Select the project to be tested in the project list.
7. Select the combined project to be tested in the list of combined projects.
8. Set the following parameters as needed:
Sample size
Sample blank
Sample cup
Repeat times
9. Select the [Emergency] check box.
10. If an item in the sample has an application request that is different from the
other items, enter the following information:
[sample size]
【repeat times】
[sample blank]
11. Click [OK].
Be careful:
Note:
Before loading the sample, make sure there are no bubbles in the sample tube
to avoid inaccurate test results.
After the power is turned on, the light source lamp needs to be stable for 20
minutes, and the test can be started when the instrument enters the standby
status.
After the sample is completed, the temperature is stable. After all other test
conditions are set, click the start button in the toolbar. The pop-up dialog
box is shown in Figure 2-18. Click ‗Start‘ to enter the sample test process.
a) b)
Figure 2-18 Sample Testing
During the sample test, click on the menu bar to monitor the status of the
instrument sample tray, reagent tray, and reaction tray in real time, as shown in
Figure 2-19 b).
Sample Condition
b) Status Monitor
Figure 2-19 Sample Testing
Sample tray monitoring: Click “Sample tray status” under the “Monitor” form
to display the status of the sample tray on the left side of the form, as shown in
Figure 2-20, where the sample tray uses different colors to distinguish the status
of the sample. After you have applied for a sample test and placed the required
samples correctly, you can start the test. To view the sample results, refer to "8.12
Sample Results Viewing and Processing" (page 8-40).
Reagent Disk Monitoring: Click“Reagent Disk Status”under the“Monitoring”
form to display the reagent disk detection status on the left side of the form, as
shown in Figure 2-21:
Reaction tray monitoring: The reaction tray displays the status of the reaction in
real time during the sample detection process. Click on “Reaction Disk Status”
under the “Monitor” form, as shown in Figure 2-22, where different colors
indicate that the cuvette is in different working status.
2. Select [Exit] on the left side of the main interface, and click Shutdown on the
Windows operating system interface to close.
3. Turn off the power in the following order:
Printer power supply
Operation unit display power supply
Analysis main power supply
Computer monitor power supply with data management software (optional) or
LIS (optional)
Water supply module (optional) power supply
After the main power of the analysis department is turned off, the refrigeration
system continues to work. If the instrument will be out of service for a long time
or not used for more than 7 days, turn off the mains power
This chapter describes the basic setup methods for the instrument, including:
System settings
Project parameter setting
Calibration settings
QC setting
3.1System Setting
3.1.1 Brief introduction
This section mainly introduce some of the setup options for system settings.
2
coefficient, other project parameters can only be viewed, and modification and deletion are not
allowed. [Project parameters] interface as shown below:
The following sections detail the settings for user-defined projects and the settings for various
project parameters.
3
Item Number
The item number is a unique number for the item and is not allowed to be repeated. The numbers
must all consist of numbers.
Chinese Name
Refers to the full name of the project name. You can enter any character and you can enter up to
36 characters. The input is not case sensitive. The full name of the project can be empty or
repeated.
Minimum absorbance, maximum absorbance, normal high value, normal low value, lowest test
value, highest test value
The specific information is filled in according to the requirements of the reagent manual.
Details
Detailed information including gender, sample type, upper and lower limits, normal high and low
values are determined by the reagent instructions.
Sample type
The sample type refers to the type of sample to which the project applies, including: serum,
plasma, urine, cerebrospinal fluid, and others. The [Sample Type] drop-down box contains the
options for the sample types supported by the project. The default sample type is displayed by
default. The system supports setting parameters for multiple sample types for the same project,
including basic parameters and result judgment parameters. The sample type of the closed project
is imported through the parameter table, and the sample type of the open project can be
customized by the user. When setting project parameters for various sample types of open projects,
you must first set parameters for the serum samples and then set the parameters for the other
sample types. The parameters of the serum sample type are used by default when calibrating.
Analysis Method
The analysis method refers to the analysis method used to calculate the test results when the
project is tested. Options include: one-point endpoint method, two-point endpoint method,
fixed-time method, and rate method.
Table 3-1 Analysis Method
Quantitative analysis of the substance based on the absorption spectrum
End point
characteristics of the reaction product at the equilibrium of the reaction
method
and its magnitude of light absorption.
It means that the reaction rate is proportional to the primary concentration
of the substrate during a certain reaction time. As the substrate is
continuously consumed, the overall reaction rate is continuously reduced,
Fix Time
and the increase or decrease in absorbance is becoming smaller and
smaller. This type of reaction takes a long time to reach equilibrium and it
takes a period of delay to enter the stable reaction period.
It is used to continuously measure the multi-point data of the concentration
of a certain reaction product or substrate in the enzymatic reaction with
Rate time, determine the initial velocity of the enzyme reaction, and indirectly
calculate the enzyme activity concentration. Mainly used for the
determination of enzyme activity.
Main Wavelength
The dominant wavelength should be selected based on the characteristics of the light absorption of
the particular reaction product and used to detect the light absorption intensity of the reactants.
Options include: 340 nm, 405 nm, 450 nm, 510 nm, 546 nm, 578 nm, 630 nm, and 700 nm.
4
Sub Wavelength
The secondary wavelength is used to correct the absorbance value measured at the dominant
wavelength and to reduce the effects of noise such as flickering, drifting, and scratching of the
cuvette. The secondary wavelength cannot be the same as the dominant wavelength. Options
include: vacant, 340 nm, 405 nm, 450 nm, 510 nm, 546 nm, 578 nm, 630 nm, and 700 nm.
Decimal number
The number of decimal places is the number of digits retained after the decimal point in the result
value. 0 to 3 decimal places available. Options include: 0, 0.1, 0.01, and 0.001.
Sample size, standard amount, sample size released, volume of sputum release, increment,
reduction
The sample size, the standard amount, refers to the amount of sample that needs to be added in a
standard test. The range is from 2 μl to 70 μl, in increments of 1 μl, and the default is 3 μl.
You can enter up to one decimal place.
The diluted sample size refers to the original amount of the sample involved in the dilution.
The amount of diluent refers to the amount of diluent involved in the dilution. The range is from 0
μl to 200μl, and the default is 0μl. You can enter up to one decimal place.
Note:
If the diluted sample size and dilution amount are set, make sure that the sum of the two is within
100μl ~ 280μl. otherwise it cannot be saved.
The diluted sample size input method for standard, incremental, and decrement tests is the same.
Decrease refers to the amount of sample that needs to be added during the decrement test. The
range is from 0 μl to 70 μl, in increments of 0.1 μl, and the default is 0 μl. You can enter up
to one decimal place.
Increment is the amount of sample that needs to be added in an incremental test. The range is from
0 μl to 70 μl, in increments of 0.1 μl, and the default is 0 μl. You can enter up to one
decimal place.
Note:
If the diluted sample size and dilution amount are set, the diluted sample will be used for normal,
incremental or decrement testing. if the diluted sample size and dilution amount are not set, the
normal, incremental or decremented sample size will be used accordingly. Test.
Sample blank
The sample blank is similar to the normal sample test except that the reagent is changed to the
same amount of normal saline and then tested in the same procedure as the normal sample. The
sample blank test was used to rule out the effects of non-colorimetric reactions such as sample
interference (hemolysis, jaundice, and lipemia) on absorbance. Sample blanks are only valid for
the single-agent endpoint method. Select the check box before [Sample Blank] to indicate that the
project must be sample blank test before the test, and the project is automatically selected in the
‗Sample Blank‘ column of the [Project Options] and [Retest] screens, and cannot be modified.
R1 amount, R2 amount
Input range of R1: 20μl ~ 350μl for conventional reagents, 240μl by default, in increments of 1μl.
The input range of R2: 0μl ~ 255μl, the default is 60μl, in 1μl increments.
R1 and R2 can be input at the same time, regardless of the order.
5
3.2.3Judging parameter settings
This section describes how to set the judgment parameters of the project.
Linear limit
The linear limit is only valid for the rate method. For rate method items, the absorbance during the
reaction time should be linear with the time (response curve). If the substrate is depleted, or the
photometer fluctuates or the agitation is uneven, it will lead to erroneous test results. Therefore,
the system calculates the linearity of the measurement time, and compares it with the linear limit
to detect whether the linear range of absorbance participating in the calculation of the reactivity is
linear.
If the linear range of reaction data does not satisfy the linear judgment, the system will add the
mark "UNLINE" to the result report.
The linear limit ranges from 0 to 1 and is accurate to two decimal places. The default is empty,
indicating that the linear limit is not determined.
The substrate depletion limit is only valid for the kinetic method and the fixed time method and is
calculated by the following equation:
among them,
L1: sample test, the main wavelength absorbance of the first metering point after the sample is
added
Lb: reagent blank or 0 concentration calibrator test, adding the main wavelength absorbance of the
first metering point after sample agitation
When L1-Lb ≤ 0, or when no reagent blank test or 0 concentration calibrator test is performed, no
correction will be made. The calibration test does not correct for substrate depletion.
For the positive effect, when the main wavelength absorbance of the photometric point is greater
than the corrected substrate depletion limit, the substrate is depleted at this point. for the negative
effect, the main wavelength absorbance of the photometric point is less than the corrected
substrate consumption. When the time limit is reached, the substrate is exhausted at this point.
Front belt inspection
There are two methods for anterior examination: rate checking and antigen addition.
Rate check method:
If the rate check method is selected, six auxiliary judgment values need to be set, namely: Q1, Q2,
PC, and ABS. Units should be consistent with reaction time and blank time.
The four-value input range is:
Single reagent item: 5≤q1<q2≤33, 5 is the first metering point after the sample is stirred.
Double reagent item: 17≤q1<q2≤33, 17 is the first metering point after adding R2 to stir.
PC: Any value between -99999999~99999999, the precision is 4 digits after the decimal point.
ABS: Any integer between -99999999 and 99999999.
Antigen addition method:
If the antigen addition method is selected, it is necessary to set three judgment values of Q1, Q2
and PC, and the other three judgment values are not allowed to be input.
6
The input ranges of the three judgment values are:
Q1: 39 ≥ Q1 ≥ the end of the reaction time.
Q2: 69 ≥ Q2 ≥ 41. PC: Any value between -99999999 and 99999999, the precision is 4 digits after
the decimal point.
7
corresponding to the same item is the same on each instrument. The project calibration test is only
allowed if the calibrator position and concentration are set.
1. Select [Project Parameters] - [Scale Parameters].
2. Select the calibrator you want to modify in the item list, and click [Scale Setting] - [Modify].
3. The following information of the calibrator can be modified in the middle of the interface:
Project name
Calibration mode
Specimen type
4. The calibration mode includes:
Options include:
A little linear
Two-point linear
Multi-point linearity
Logistic-Log 4P
Logistic-Log 5P
Factor
Polyline (but origin)
Spline
5. If the factor method is selected, enter the K factor in [Factor].
This field is only active when a single-point linear calibration rule is selected. After inputting the
K factor, the calibration result is calculated as Y=K*X. Y represents the calibration result, K
represents the factor, and X represents the degree of reactivity. As long as the K factor is entered,
it can be used to calculate sample results without performing a calibration test.
6. Write the number of repetitions of the calibration test in [Repeat Times], which is generally 1
by default.
7. After selecting the calibration mode, determine the number of calibrations.
A project can select up to 10 calibrators (including water), and the number of calibrators should
correspond to the selected calibration rules, as shown in the following table:
Figure3-2 relationship between calibration rules and the number of calibration products
Calibration Calibration
Example
mode Quantity
Factor — CK-MB
8、After determining the number of calibration, set the corresponding absorbance, concentration
and cup size
9、Click [save] to save the Settings of calibration.
8
3.3.3 Delete calibration
In addition to the system's own WATER, delete calibration information is allowed.
After the calibration product is deleted, All Settings are deleted. Calibration products
can no longer participate in the calibration application, but the historical results of
calibration products can still be searched according to the project. Only if the
calibration product is not tested, it can be deleted.
1、Select [project parameters] - [calibration parameters]
2、Select the calibration item you want to delete in the calibration item list area
3、Click "scale delete"
4、Click "save" to complete the delete operation
3.4 QC setting
3.4.1 Introduction
QC Settings include the following steps:
Add/edit QC products
Select applicable items
Set the concentration parameters of QC products
Set QC rules
3.4.2 Add/edit QC products
When setting, the name of QC product must be selected. The name and batch number
combination of QC products cannot repeat. If the batch number is not set for the QC
product, it is not allowed to define the QC product with the same name. Only when
the system is in the standby status, only allow to add or edit QC.
9
2、Click [add].
10
4、Select the item applicable to QC products in the middle list, and click the left and right arrow
buttons to import it into [QC items].
5、Confirm the [QC digit] of the QC item (i.e. the position of the reagent&sample tray).
6、Click [save] to save the Settings of QC products.
7、To set up more QC products, click [add] and repeat steps 2 to 6.
8、If you want to cancel an item, then re-select it and click [cancel].
11
Figure 3-8 QC rules
12
4 Calculation Method
This chapter briefly introduces the measuring principle of the instrument, including:
Analysis method
Verification type and parameter calculation method
13
Calculation method
4.1 Introduction
The instrument is a fully Auto, discrete, random clinical chemical analyzer, the whole process by
the computer control. With the help of various calculation methods and measurement principles,
can quickly complete the test.
The data analysis and calculation process of the system is shown in the following figure:
AD
Absorbance
Reactivity
Calibration parameters
QC judgment
First, the system tested the intensity of the light through photoelectric conversion, linear
amplification and AD conversion, and then calculated the absorbance of the reaction solution
and the change rate of absorbance according to the intensity of light, i.e., the reactivity, and
then calculated the calibration parameters according to the reactivity. Finally, carry out QC
testing, calculate the QC results, judge whether the system is stable, and then calculate the
sample test results according to the calibration parameters.
Attention:
analysis. It is not sensitive to small changes in reaction conditions (such as enzyme amount, pH,
temperature, etc.), as long as the change does not affect the equilibrium of the reaction within a
certain period of time.
B
R1 S R2
C
Time
Figure 4-2 One endpoint reaction curve
(a) Metering point:[L]-[M]-[0]-[0](1<M<L≤42)
(b) Absorbance calculation:
Take the metering point L&M absorbance average, calculated is follow:
AL AL 1 ... AM 1 AM
AX
(L M )
(c) Concentration calculation
CX {K ( AX A1 )} IFA IFB
Cx is a sample concentration to be test, A1 is the first point absorbance value, K is K
factor, B is the absorbance of reagent blank, IFA and IFB is a constant of the
instrument, are represented by the slope and intercept
(d) Analysis Item:
Like TP, ALB etc.
Absorbance
R2
Cup blank
B1
S
R1
B1
Time
That is, in a fixed time interval TL~TM, the amount of change in substrate concentration is
proportional to the initial concentration of the substrate. This is the generality of a reaction. The
increase (or decrease) of the absorbance during the period and the concentration of the analyte The
proportional time method is also called initial rate method, first-order dynamic method, two-point
dynamic method, etc. According to the input form of the metering point, the fixed time method
can be divided into single interval fixed time method and double interval fixed method. Time
method. The double interval fixed time method can deduct the sample blank in real time, that is,
the absorbance change between two points in a certain period of time is used as the sample blank
deduction.
4.4.2 Calculation
Reaction curve shows as figure 4-4:
Absorbance
B1
Time
B is cup blank, R1~R2 is add position of reagent, Ax is the average metering point
between L and M change in absorbance per minute, Cx is sample concentration to be
test, C1 is the concentrations of calibration solution 1 (reagent blank), K is K factor, B
is the calibration solution 1 (reagent blank) absorbance, IFA and IFB is a constant of
the instrument, are represented by the slope and intercept.
(d) Analysis Item:
BUN, Picric acid method CRE, etc
4.5.2 Calculation
Reaction curve show as 4-5:
Absorbance
R1 S
R2
Cup blank
B1
AL AM
Reaction boundary level
B1
Time
Figure 4-5 Rate method reaction curve
Ax
B STD(1)
C1 Cx Concentration
K: Input value
C1: The concentration of the calibration solution 1 is the input value.
(d) Calculation of Concentration
A2 STD(2)
Ax Sample
B STD(1)
C1 Cx C2 Concentration
A10
A9
A8
A7
A6
A5
A4
A3
Ax
A2
B
C1 C2 Cx C3 C4 C5 C6 C7 C8 C9 C10 Concentration
X Cr
S1ABS ( B) A
Y
Y
K
X
n
X Cri Cr Ai A
i 1
n
Y Cri Cr
2
i 1
n
A Ai / n
i 1
n
Cr Cri / n
i 1
A1, A2 is the two measured values of calibration solution (1), n is the number of
calibration solution N×2, Cri is the concentration calibration solution (i) .
(c) Calculation of concentration
C X K AX B C1 IFA IFB
Logit-log4P
Suitable for the working curve that absorbance showed convergence with the
concentration increase, Logit-log4P (nonlinear method) calibration curve shown in
figure 4-9:
Absorbance
AN STD(N)
Ax Sample
A3 STD(3)
A2 STD(2)
B
STD(1)
C1 C2 C3 Cx CN Concentration
K
AX B
1 aC b
1 K ( AX B)
C b
a AX B
A Ai ,
N 2 2
ij
i 1 j 1
SD
2N 4
(N=4~6, j=1 or2 )
(Aij-Ai‘) is d-value of absorbance between Ai‘ from fitting equation and Aij from
testing or the d-value between Aij and A12. Every calibration fluid test twice, and the
maximum value of Aij is 12.
(e) Applicable methods
One endpoint method, fixed time method, two endpoint method, rate method.
Logit-log5P
Same characteristic with Logit-log4P, and Logit-log5p has one more calculate
parameter, so the result is more accuracy. Calibration is shown curve as figure 4-10:
Absorbance
AN
A3
Ax
A2
C1 C2 Cx C3 CN Concentration
AX B
a b lnC c C ln 0
K AX B
Get C from Newton Approximation
C X (C C1) IFA IFB
K
AX B
1 exp a b l n C c C
A Ai ,
N 2 2
ij
i 1 j 1
SD
2N 5
(N=5~6, j=1 or 2)
(Aij-Ai‘) is d-value of absorbance between Ai‘ from fitting equation and Aij from
testing or the d-value between Aij and A12. Every calibration fluid test twice, and the
maximum value of Aij is 12.
(e) Applicable methods
One endpoint method, fixed time method, two endpoint method, rate method.
Polyline method
Test from fluid 1 to fluid 5 or fluid6, and get the curve line, and link those points
with straight line. As figure 4-11 shown:
Absorbance
STD(6)
STD(2)
STD(1)
C1 C2 CA C6 Concentration
Spline method
In this line, every value of calibration linked to a complete curve, and the error is
also fitting in the curve, so the curve fitting is better than poly line. The calibration
curve is shown as figure 4-12:
Absorbance
AN
Ax
A3
A2
C1 C2 C3 Cx CN Concentration
f (C X C ( I )) a ( I ) b ( I ) (C X C ( I )) d ( I ) (C X C ( I )) 2 d ( I ) (C X C ( I )) 3 AX
A Ai ,
N 2 2
ij
i 1 j 1
SD
2N 4
(N=5~6, j=1 or 2)
(Aij-Ai‘) is d-value of absorbance between Ai‘ from fitting equation and Aij from
testing or the d-value between Aij and A12. Every calibration fluid test twice, and the
maximum value of Aij is 12.
(e) Applicable methods
One endpoint method, fixed time method, two endpoint method, rate method.
Reactivity R
(antigen excess equivalent (antibody excess
zone) zone zone)
Antigen
concentration C
Figure 4.5 Dose response curve of antigen-antibody reaction
In the reaction of the antigen-antibody, the ratio of the insoluble antigen-antibody complex
produced to the antigen-antibody is closely related, and the amount of the insoluble
antigen-antibody complex formed at the appropriate ratio is the largest, and the light transmitted at
the time is the least, which corresponds to the maximum absorbance. When the ratio is less than
this ratio, the amount of the insoluble antigen-antibody complex is reduced, the transmitted light is
increased, and the absorbance is decreased. In this case, the two samples (antigens) having a large
difference in concentration, the amount of the insoluble antigen-antibody complex formed. Can be
equal, if the front test is not carried out, the same measurement results will be obtained. Therefore,
for antigen-antibody reaction, it is necessary to carry out the front test.
The prozone limit refers to the maximum or minimum PC value allowed when there is no
excess antigen.
Prozone check parameters include:
Prozone limit PCM, and q1、q2
Prozone check distinction abs. low limit
There are two methods for prozone check: antigen re-addition method and reaction rate ratio
method. The two methods will be described in detail below.
The above q 1, q2 are metering points, and satisfy 69≥q2≥41,39≥q1≥end of reaction time.
If one of PCM, q1, and q2 has no input, no antigen re-addition judgment is made.
Sample PC value: PC=Aq2-k×Aq1.
k is the volume correction factor.
Single reagent project: k=(VR1+VS)/(VR1+2VS).
Dual reagent project: k=(VR1+VS+VR2)/(VR1+2VS+VR2).
For the ascending reaction, if the PC <PCM. drops the reaction, if PC> PCM, the front check
abnormal "PRO" mark will be given and the alarm will be given.
Sample PC value: . If PC>PCM, give the front check error "PRO" mark
and alarm.
Metering point input requirements:
Single reagent project: 5≤q1 <q2≤33, 5 is the first metering point after mixing the sample.
Dual reagent project: 17≤q1 <q2≤33, 17 is the first metering point after adding R2 to stir..
If one of PCM, q1, and q2 has no input, the reaction rate ratio is not judged.
The system will no longer perform the prezone check if the following two conditions occur:
(Sample end point absorbance - reaction start point absorbance)<【ABS】
Sample reactivity is not within the calibrator response range (valid for sample and QC
tests on non-linearly calibrated items).
This chapter describes the functions and operating methods associated with
reagent applications.
32
Reagent application
5.1 Overview
5.1.1 Brief introduction
This chapter describes the advanced application functions of the reagent module.
Please select the following operations according to actual needs.
Reagent allowance alarm limit setting
Reagent residue detection
Print reagent information
Loading reagents
Replace reagents
Unloading reagents
Reagent application-33
Reagent application
Reagent application-35
Reagent application
Click [One button refresh] to refresh the remaining amount and measurable number of
reagents in all positions to the maximum, and update the status of the entire reagent tray.
Click [Refresh] to refresh the remaining amount and measurable number of reagents in the
currently selected position to the maximum, and update the corresponding reagent status.
Reagent application-36
Reagent application
Warning:
Biological infection:
Be sure to wear gloves and overalls to prevent infection and wear
protective glasses if necessary.
Do not touch the reagent directly, as it may cause skin damage or
inflammation.
Reagent application-37
Reagent application
3. Click [Add]. The corresponding location in the reagent information list box will display
the corresponding reagent information.
4. If the reagent input is wrong, select [Delete] and delete the item to rescan the code.
Open reagent channel reagent information record
1. Select [Reagents] - [Reagent Information].
2. Select the reagent tray number, in the [Reagent Information Edit] list box, select the
reagent, bottle size and reagent type information in sequence, as shown below.
Warning:
Reagent application-38
Reagent application
Biological infection:
Be sure to wear gloves and overalls to prevent infection and wear
protective glasses if necessary.
Do not touch the reagent directly, as it may cause skin damage or
inflammation.
Reagent application-39
Calibration test
6 Calibration test
This chapter describes the various functions and methods of operation related to
calibration, including:
Calibration parameter settings
Calibration entry
Calibration test
Calibration result confirmation
6.1 Overview
Calibration refers to the determination of the reactivity of a known concentration of the
calibrator. Based on the mathematical relationship between the concentration and the
reactivity (ie, the calibration method), the coefficient (ie, the calibration parameter) in the
relationship is calculated to determine the concentration and the reactivity. Specific
mathematical expressions. Ordinary samples are used to determine the sample
concentration based on the known expression of the coefficient and the measured
reactivity.
[Save] button on the left to pop up the save dialog box. [Confirm] button to delete the
calibration item.
When the ― ‖ button is displayed in the start test dialog box, it means that the test
condition is not met. Putting the mouse on the button will display the corresponding
prompt. According to the information prompt, complete the pre-test check and restart the
calibration test.
3. During the test, the status bar shows the test process, as shown below.
1. In the menu bar on the left side of the main interface, select [Data Processing],
click [Response Curve] - [Scale] to enter the calibration response curve
interface.
2. Select the item in the results list for which you want to view the response
curve.
7 QC test
This chapter introduces various functions and methods of operation related to QC,
including:
Overview
QC rule setting
QC information login
QC items
QC test
QC results confirmation
QC adjustment
QC tes-47
QC tes
7.1 Overview
QC test refers to the sample provided by the authoritative department or reagent developer,
and provides the concentration range of various substances to be tested in the sample. The
results obtained by testing the sample on the instrument are compared with the given range.
This determines whether the instrument status is normal and the test result is reliable.
After each calibration test, reagent replacement, maintenance, and troubleshooting
operations, QC testing is recommended to ensure stable instrument performance.
determined by the mean value ( X ) and standard deviation ( SD ) of the known specimen
(usually the control solution) by the controlled analysis method. X 2SD is the warning
QC tes-2
QC tes
No
1 2S In control
Yes No
No No No No
1 3S 2 2S R 4S 4 1S 10 -
x
Yes Yes Yes Yes Yes
Out of control
QC tes-3
QC tes
QC tes-4
QC tes
QC tes-5
QC tes
QC tes-6
QC tes
When the ― ‖ button is displayed in the start test dialog box, it means that the test
condition is not met. Putting the mouse on the button will display the corresponding
QC tes-7
QC tes
prompt. According to the information prompt, complete the pre-test check and restart
the quality control test.
3. During the test, the status bar displays the test process as shown below.
Xi
i 1
Target( X ): N
N
Xi average
i 1
2
Standard deviation(SD): N 1
SD
100%
Coefficient of variation(CV%): average
Deviation: Xi-(average)
Deviation
100%
%Error: average
Among them:
N is the number of measurements, Xi is the test result.
QC tes-8
QC tes
QC tes-9
QC tes
QC tes-10
QC tes
QC tes-11
Sample test
8 Sample Test
This chapter describes the various functions and operation methods related to sample testing,
including:
Sample test
Modify/append sample and project tests
Sample retest
Load/unload sample
Cancel sample request
View sample tray location
Sample results viewing and processing
Sample test-2
Sample test
8.1 Overview
The system applies sample testing for the reagent&sample tray and supports the following test
methods, including: single sample application, batch application, retest application, additional
application, and emergency application. The applied test project may be a single biochemical
project, a computing project, or a project portfolio composed of commonly used projects. Before
starting the sample analysis, you can set the project test properties as needed. during the test, input
patient information and view the sample test status. The system also provides the function to
delete samples and their application information and test results.
These features and operation methods are detailed in the following sections.
Precaution:
Sample test-2
Sample test
The system can enter a single sample or batch sample, as well as a combination of items.
Single sample entry: Click [Add], and select the items to be tested in the upper left column, click
the button, and then select to the "Selected test items" list column. The sample bit number
generally starts from 1 and is accumulated in turn. The serial number cannot be repeated, and the
patient's information ―name‖, ―gender‖, ―age‖, etc. are filled in order, and the information such as
―sent inspection department‖, ―sample type‖, ―to be sent to the doctor‖ is edited. After the login is
completed, click [ Save] to complete the operation.
Batch sample entry: If there are multiple samples, you need to test the same project. You can use
batch input, put the samples on the sample tray continuously, click [Batch Entry], fill in the
quantity to be tested in the ―Bulk Entry‖ information input box. Start number, as shown below.
The sample bit number input starts from 1 to 37, and the sample bit number changes from 37 to 1
and then increments from 1. In the upper left column, select the items to be tested, click the button,
select to the "Selected test items" list column, and click [Batch Save] to complete the operation.
Click on the top left corner of the interface, the start test dialog box will pop up, then click
[Start] to start the test.
Sample test-3
Sample test
If the system is in the "standby" status, click the button in the upper left corner of the
main interface to start the test.
Sample test-4
Sample test
Sample test-5
Sample test
Retest].
Only the current day can be re-tested.
status, click the button in the upper left corner of the main
interface, and then perform the next step.
tinue to the next step.
3. Confirm that the sample tray and reagent&sample probe have stopped moving.
4. Open the sample tray cover.
5. Insert the sample cup into the sample tray until the bottom of the sample cup is in full contact
with the annular groove of the lower ring of the sample tray.
6. If you need to continue loading more samples, load the sample cup according to step 5 until the
loading is completed.
7. Install the sample tray cover.
Loading samples from the reagent&sample tray
Chemical Infection Risk:
When handling, be sure to wear gloves and overalls to prevent infection and
wear protective glasses if necessary.
1. Confirm that the sample tray and reagent&sample probe are in a stopped status.
2. Confirm system status:
f the system is in the test status, click the button in the upper left corner of the main interface,
and then perform the next step.
Sample test-6
Sample test
1. In the upper left corner of the main interface, the " " button pops up to start the test dialog
box.
Sample test-7
Sample test
When the ― ‖ button is displayed in the start test dialog box, it indicates that the test condition
is not met. Putting the mouse on the button will display the corresponding prompt. According to
the information prompt, complete the pre-test check and restart the sample test.
1. During the test, the status bar displays the test process as shown below.
Sample test-8
Sample test
Sample test-9
Sample test
Sample test-10
Sample test
Sample test-11
9 Data Processing
This chapter describes the backup of various test result data, print template settings, Auto
printing and manual printing methods, and describes the style of printing reports.
The report examples provided in this chapter are for demonstration purposes only. Please
refer to the actual printed report.
Data processing
Precaution:
Do not turn off the analysis main power or exit the operating software
during data export.
1. Select [Data Processing] - [Data Export].
2. Check the data before exporting the data.
3. Select its export mode, network port or serial port in the [Export Mode] drop-down menu.
4. Network port export:
Select the network port export mode, you need to set the corresponding server address, port
number, encoding format and other data.
Confirm whether it is transmitted in real time.
Then click [Connect] to check whether the status is connected successfully.
Serial port export:
Select port export mode, you need to set the corresponding port and baud rate.
After clicking [Open], a prompt box for confirming whether to transmit in real time will appear. If
you need real-time transmission, click [OK], otherwise click [Cancel].
Data processing -3
Data processing
The relevant data is listed in the sample list. Click [Export] to back up the data. The backup file is
named by default with the date and time of the backup. The format is .csv.
9.2.1 Introduction
Various test results and data can be printed out via the printer and the specified template. Can set
the printer type and default.
Identify the printer and the order in which the items are printed.
9.2.2 Changing the print paper type
Open the biochemical instrument software installation folder, find the configuration file
, open the configuration file config.ini to find the field [print page], set the print
paper type, select the paper type as A4_zh, A5_zh, B6_zh, if the paper type is set to A5_zh, field
Data processing -4
Data processing
Data processing -5
Data processing
Method two:
1. Select [User Management] in the menu bar on the left side of the main interface to open
the "Test Results Query" interface, as shown below:
Data processing -6
Data processing
9.6.2 QC Result
After the QC test is finished, you can select the print result and print it through the [Data
Processing] interface.
1. Select [Data Processing] - [QC Result Query].
2 Query the QC result that needs to be printed.
3 Select the QC result.
4 Click [Print QC Chart] or [Print QC Report].
Data processing -7
Item parameters
10 Item Parameters
This chapter introduces the various advanced application functions of the project. Its main
purpose is to integrate some of the results tested by other instruments into one inspection
report. include:
Manually enter project settings and applications
Calculation project settings and applications
Combination project setup and application
Item parameter -8
Item parameter
full set of liver function and a full set of kidney functions, just click on the name of the project
portfolio. Complete the registration function of multiple items to facilitate quick entry during
sample registration.
Item parameter-3
11 System Function
This chapter describes how to perform various instrument commands and advanced system
settings to make better use of the instrument.
Advanced system setup options include:
System function
Description:
The user privilege level is divided into three levels: normal, advanced, and super advanced.
Ordinary permissions are used by hospital operators, advanced privileges are used by hospital
administrators, and super-privilege is used by customer service engineers.
The advanced permission group can set different permissions for each user of the normal
permission group, allowing them to perform operations corresponding to the assigned rights.
If the normal privileged user password is forgotten, you can log in to the system as an advanced
privileged user, delete the username, and then reset it. or contact the company's customer service
department or the distributor in your area. If the administrator password is forgotten, please
contact the company's customer service department or distributor in your area.
System function -3
12 How to Use LIS
12.1 Overview
This chapter describes how to use LIS.
The Laboratory Information System (LIS) is an external host connected to the analyzer through a
fixed interface for downloading sample application information to the analysis module and
accepting sample test results transmitted from the analysis module.
Before using the LIS to download sample application information and transmit sample results, you
need to set the communication parameters of the LIS and the result transmission method.
Before using the LIS feature, make sure you have selected the LIS. If not, please contact our
customer service department or distributor in your area.
selected, after obtaining and downloading the application information, it is necessary to manually
set the location for the sample.
This chapter describes how to maintain the instrument, including common maintenance
instructions and regular maintenance. A detailed description of the purpose of each maintenance
project, timing of use, required supplies, instrument status, precautions, and operating procedures
are provided.
maintenance
13.1 Overview
13.1.1 Introduction
To ensure reliable performance, good working condition and longevity of the system, please
operate and regularly maintain the system in strict accordance with the requirements of this user
manual. Even if you are only an operator, it is important to understand the maintenance and
troubleshooting knowledge in this chapter. In-depth study will enable the instrument to achieve
optimal operation and optimum performance during use.
The system provides biochemical maintenance instructions and maintenance lists. Through the
maintenance command, you can perform various maintenance operations on the instrument. You
can use the maintenance list to master the periodic maintenance items, record the time and content
of each maintenance, and record the abnormalities or other important events that occur during the
maintenance process for later review. For unresolved problems encountered during use and
maintenance and repair issues not covered in this chapter, please contact the company's customer
service department or the distributor in your area.
Warning:
Do not perform maintenance work that is not explicitly stated in this chapter.
Failure to do so may result in system damage and personal injury.
Do not touch parts that are clearly documented and can be operated and
maintained by the user.
Unauthorized repairs to the system can result in system damage and personal
injury, and the terms promised in the repair contract are no longer valid.
After the maintenance work is completed, please confirm that the system is
working properly.
Do not spill liquids such as water or reagents on the mechanical or electrical
parts of the system.
If you do not use the instrument for a period of time (more than one week) or
need to move the instrument, please contact the company's customer service
department or the distributor in your area to perform on-machine maintenance
to ensure that the instrument will still perform well after the next power-on.
maintenance-5
maintenance
For 1
No Item For once
year 1 6
day timely weekly 3 month yearly
month month
1 Sample tray ●
2 Reagent&sample probe ○
Reagent&sample probe
3 cleaning tank ○
Stirrer cleaning tank
4 6 24
Cuvette (8/group) ○ ●
(note a) group group
Reaction tank and
5 reaction tank heating ○
belt
6 halogen lamp (light
6 1 2 ●
source lamp)
maintenance-7
maintenance
Period
For 1
No Item For once
year 1 6
day timely weekly 3 month yearly
month month
8 Stirrer ○
Reagent&sample probe
9 ●
pump
Reagent disk
11 ○
refrigeration unit
12 Cooling fan ○
Warning:
stirrer rotates normally. Check whether the water in the cleaning needle and the cleaning tank is
normal.
Maintenance opportunity
It is recommended to perform this maintenance operation before starting the test every day.
Instrument status
When performing this maintenance operation, make sure the instrument is in standby.
Precautions
Warning:
Operation Steps
1. Open the top cover of the analyzer.
2. Check the reagent&sample probe/stirrer/clean needle for any dirt on the outer wall. If there is
dirt, wipe the outer wall of the needle with a cotton swab dipped in alkaline cleaning solution, and
wipe the surface of the needle with a cotton swab with pure water, as shown in Figure 13-1:
maintenance-9
maintenance
Figure 13-2 Normal and abnormal water flow direction of the needle
5. Observe whether the cleaning water out of the cleaning tank is normal, whether the water
volume is suitable, and whether the water flow in the cleaning tank can be washed up to about 5
mm above the needle tip. If the effluent is normal, proceed to the next step, otherwise contact
customer service engineer.
6. Click the ―Initialize‖ icon to observe whether the operation of each module is normal. If it is
normal, you can test it later. If the modules are not working properly, please contact customer
service engineers.
Precaution:
Chemical Infection Risk:
During maintenance work, be sure to wear gloves and work clothes to prevent
infection and wear protective glasses if necessary.
When disposing of waste liquid, dispose of high-concentration waste liquid in
accordance with local regulations.
Operation Procedures:
1. Check whether the waste liquid discharge system is normal, keep the waste liquid pipeline
unbending, discharge smoothly, and treat the high and low concentration waste liquid properly
(the waste liquid treatment method refers to local regulations).
2. Confirm that the waste liquid guiding tube is unblocked and not bent. Otherwise, the waste
liquid may overflow from the analysis module panel due to poor drainage, which may cause
damage to the analysis module.
3. If the liquid leakage still occurs after the above operation, please contact our customer service
department or the distributor in your area.
4. Emptying the waste liquid in the high-concentration waste liquid tank.
13.3.4 Reaction Cup Detection
After long-term use of the reaction cup, substances such as protein or debris may remain on the
inner surface and cannot be cleaned, which may affect the light transmittance of the reaction cup.
In addition, if the inner or outer wall of the reaction cup is contaminated or the reaction cup is
scratched or cracked. Both affect the transmittance or uniformity of the cuvette, which in turn
affects the accuracy and stability of the absorbance test results. Therefore, it is necessary to
confirm the use status of the cuvette.
Purpose
Check if the cuvette is contaminated and the transmittance is reduced to avoid affecting the test
results.
Maintenance time
It is recommended to perform this maintenance daily, after changing the cuvette or after
cleaning the cuvette.
Instrument status
Before performing this maintenance, make sure the system is powered on for more than 20
minutes and is in standby. Also confirm that the cuvette has been placed in all cups. If not, please
add a reaction cup.
Precaution:
Precaution:
In order to ensure the performance of the light source lamp, please handle or
replace it as soon as possible after the cuvette is judged to be a dirty cup.
After the replacement is completed, the cuvette detection function should be
performed again. Since the residual material in the reaction cup will affect the
reaction result of the reaction cup, it is recommended to perform the reaction
cup test after completing the ―cleaning background‖.
Operation Procedure:
Make sure light source has opened more than 20min, open —— check
showing status of cuvette. Check all cup positions, record the cup position information marked in
red, and perform [Cleaning Background] or [Replacement Cuvette] Maintenance. For details,
maintenance-11
maintenance
please refer to “13.3.4 Reaction Cup Detection” and “13.6.1 Replacing the Reaction Cup”. The
screen displays all the reaction cups and identifies the dirty cups by color:
White: free cuvette
Pink: add R1 already
Blue: add R1 and sample already
Purple: add R1, R2 and sample already
Green: finished inspection
Red: dirty
Warnings:
Be careful to avoid scratching your hand with the tip of the needle. Prevent
bending or scratching of the reagent&sample probe and reagent&sample
probe. If the above situation occurs, the reagent&sample probe or the
reagent&sample probe should be replaced immediately, otherwise the test
result will not be guaranteed.
maintenance-12
maintenance
Warning:
Do not hold the probe at the top of the probe cover by hand. The wrong
operation is shown below.
maintenance-13
maintenance
Steps
1. Open the top cover of the analyzer.
2. Manually remove the reagent&sample probe and stirrer cross arm so that the
reagent&sample probe and the stirrer leave the cleaning tank.
warning
Do not hold the needle at the top of the needle cover by hand. The wrong
operation is shown below.
maintenance-14
maintenance
Figure 13-7
4. Then, about 100 mL of pure water is injected into each of the washing tanks for rinsing.
5. The system performs the cleaning operation of the outer wall of the needle.
6. Observe whether the water in the cleaning pool is normal, that is, whether the cleaning pool can
reach about 5mm above the needle tip. if it is not normal, please contact the customer service
engineer.
not to affect the accuracy of the measurement results and the detection speed.Or software in
the case that the dirty cup status of the reaction tray reaches 1/3 or more, the new reaction cup
needs to be replaced.
Purpose
Replace the cuvette to ensure the accuracy of the results.
Maintenance opportunity
This maintenance is recommended every three months.
Maintenance supplies
Reaction cups (8/group)*6 groups.
Instrument status
Make sure the analyzer is powered off during this maintenance.
Precautions:
Steps
1. Turn off the main power switch of the instrument and remove the reaction disk cover.
2. Unscrew the fastening screw, as shown in Figure 9-32
maintenance-15
maintenance
Figure 13-8
3. Remove the six sets of reaction cups, as shown in Figure 13-9:
Figure13-9
1. Install six new cuvettes on the reaction tray in reverse order.
2.
Warning:
If the used cuvette is exposed to the air for a long time, the
contaminants may condense on the wall of the cup. Therefore, the
reaction tray cover should be covered in time. In addition, if an
emergency stop occurs during the test, the cuvette that has not been
cleaned should be cleaned or rinsed with pure water to prevent the
reaction solution from remaining on the cuvette for a long time.
Never use organic solvents (benzenes, alcohols, etc.) to scrub or soak the
cuvette.
13.6.2 Cleaning of the reaction tank
The reaction tank should be cleaned in time for long-term use to prevent dust and other factors
from affecting the test results. In addition, especially in the case of water intake, it should be
wiped clean in time to prevent contamination of the reaction cup and light path, affecting the test
results.
Purpose
Clean the reaction tank to prevent dust and other effects from affecting the test results
Maintenance opportunity
This maintenance is recommended every three months.
Maintenance supplies
Gauze, pure water.
Instrument status
Make sure the analyzer is powered off during this maintenance.
maintenance-16
maintenance
Precautions:
Biological infection risk:
During maintenance work, be sure to wear gloves, work clothes to
prevent infection, and wear protective glasses if necessary.
Steps
1. Turn off the main power switch of the instrument and remove the reaction disk cover.
2. Unscrew the fastening screw, as shown in Figure 13-8 above.
3. Remove the six sets of reaction cups and place them in pure water or clean place, as shown
in Figure 13-9 above.
4. Wipe the reaction cell with a clean, wet gauze (be careful not to wipe the metering
window), as shown in Figure 13-10:
After the reaction tank is cleaned, install the reaction cup and cover the reaction tray cover.
Click on the red alarm button ,The alarm message window is displayed as shown in Figure
13-11. The halogen lamp should be replaced.
maintenance-17
maintenance
Figure 13-11
Purpose
Replace the light source to ensure the accuracy of the test results
Maintenance opportunity
This maintenance is recommended every six months.
Maintenance supplies
Halogen lamp, gauze, pure water, alcohol, cross flower screwdriver.
Instrument status
Make sure the analyzer is powered off during this maintenance.
Steps
1. Prepare a new halogen lamp, as shown in Figure 13-12:
Figure13-12
Warning
Do not touch the surface of the halogen lamp, otherwise it will affect the
amount of light. If the surface is found to have smudges such as
fingerprints, wipe it with a gauze dampened with alcohol.
2. Turn off the main power switch of the instrument. After about 30 minutes (to completely cool
the lamp chamber), proceed to the next step to avoid burns.
3. Use a cross-blade screwdriver to unscrew a screw on the panel 1.
maintenance-18
maintenance
Panel 1
Terminal
block
Fastening screw
Wire 1
Wire 2
Tie
maintenance-19
maintenance
Warning:
Warning:
Do not spill liquid on the analyzer to prevent liquid from immersing and
causing damage to the instrument.
maintenance-20
maintenance
Steps
1. Confirm that the instrument is in a non-test status and open the upper cover of the
analysis module.
2. Gently wipe the sub-module table top and the disc cover with a gauze dipped in alcohol.
3. Use a special cleaning agent to clean the keyboard and other parts.
4. Cover the upper cover of the analysis module.
13.8.2 Cleaning reagent&sample tray
When the sample or reagent is accidentally spilled into the pan, or when visually inspecting the
inner wall of the bin for dust and condensed water during the cooling process, it should be cleaned
in time to reduce the risk of cross-contamination.
Purpose
Clean the reagent&sample tray assembly to keep the working environment and the table clean and
tidy, reducing the risk of cross-contamination.
Maintenance opportunity
Perform this maintenance when a sample or reagent accidentally spills into the chamber and there
is dust on the inside of the visual chamber and condensation occurs during cooling.
Maintenance supplies
Clean gauze, pure water, alcohol, cotton swab
Instrument status
When performing this maintenance operation, make sure the instrument is in a standstill or
standby status.
Precautions
Warning
Steps
1. Confirm that the instrument is in the stop or standby status.
2. Open the reagent&sample tray cover, unscrew the reagent&sample tray handle
counterclockwise, and take the reagent&sample tray and place it in a safe and reliable
position.
Reagent&sample tray handle
maintenance-21
maintenance
3. Wipe the inner wall of the tray with gauze with a small amount of deionized water or alcohol. If
necessary, use a gauze with a small amount of neutral detergent to wipe.
4. Use gauze, a small amount of deionized water or alcohol to wipe the reagent&sample tray. For
dirt on the sample and reagent positions, use a cotton swab dipped in a small amount of alcohol.
5. Wipe the condensate from the reagent&sample tray to the condensate drain.
Condensate drain
maintenance-22
maintenance
maintenance-23
maintenance
Figure13-22
5. Unplug the tubing at the end of the reagent&sample, as shown in Figure 13-23.
maintenance-24
maintenance
Sample cup
The front end of
dispensing probe
maintenance-25
maintenance
反应杯
Cuvettes
Figure13-28
Note: If the probe tip of the reagent&sample probe is not at the center of the reaction cup, please
contact customer service engineer.
19. Sample position vertical position check: Click sample position “Y”, the reagent&sample
probe is left to stop above the sample position, click the vertical position “Y”, the
reagent&sample probe will drop until the tip of the sample just touches the bottom of the sample
cup. , then click the vertical position "Y", lift the reagent&sample probe up to the top of the
sample position, then click the zero position "Y", and the reagent&sample probe swings back to
the initial position.
20, reagent outer ring vertical position check: click the reagent outer ring "Y", the reagent&sample
probe is left to stop above the reagent outer ring, click the vertical position "Y", the
reagent&sample probe will always drop until the probe tip just touches the reagent Click the
vertical position "Y" at the bottom of the cup, lift the reagent&sample probe over the outer
circumference of the reagent, and then click the zero position "Y", and the reagent&sample probe
swings back to the initial position.
21, the vertical position of the reaction cup: click the reaction cup "Y", the reagent&sample probe
is placed right to the top of the reaction cup to stop, click the vertical position "Y", the
reagent&sample probe will drop, so that the Teflon layer is coated Partially all enter the cuvette,
click the vertical position "Y", the reagent&sample probe is lifted above the cuvette, then click the
zero position "Y", and the reagent&sample probe swings back to the initial position.
13.8.4 Replacing the reagent&sample probe
maintenance-26
maintenance
Warning
maintenance-27
maintenance
If not, check if the spring is stuck, or if the screw is pressed too tightly to correct the problem.
13. After the maintenance operation is completed, turn on the power of the analyzer and observe
whether the LED green light of the number D2 on the circuit board on the cross arm of the probe
is lit.
If it is lit, it indicates that the liquid level detection system is normal.
If it is not normal, please contact the company's customer service department or the distributor
in your area.
14. Install the cross arm cover until you hear the click of the card, confirm that the installation is
in place.
15. Turn on the main power switch of the instrument and initialize the instrument.
16. Click “Maintenance” in the menu bar to open the “Instrument Detection” form and find
the sample arm in the test item list, as shown in Figure 13-29.
样本杯
针的前端
maintenance-28
maintenance
Position, as shown in Figure 13-31, then click the zero position "Y", the reagent&sample probe
swings back to the initial position.
Figure 13-31
Note: If the probe tip of the reagent&sample probe is not at the center of the reaction cup, please
contact customer service engineer.
20, sample position vertical position check: click the sample position "Y", the reagent&sample
probe is placed to the left of the sample position to stop, click the vertical position "Y", the
reagent&sample probe will continue to drop until the tip of the probe just touches the bottom of
the sample cup. , then click the vertical position "Y", lift the reagent&sample probe up to the top
of the sample position, then click the zero position "Y", and the reagent&sample probe swings
back to the initial position.
21, reagent outer ring vertical position check: click the reagent outer ring "Y", the reagent&sample
probe is left to stop above the reagent outer ring, click the vertical position "Y", the
reagent&sample probe will always drop until the probe tip just touches the reagent Click the
vertical position "Y" at the bottom of the cup, lift the reagent&sample probe over the outer
circumference of the reagent, and then click the zero position "Y", and the reagent&sample probe
swings back to the initial position.
22, the vertical position of the reaction cup: click the reaction cup "Y", the reagent&sample probe
is placed right to the top of the reaction cup to stop, click the vertical position "Y", the
reagent&sample probe will drop, so that the Teflon layer is coated Partially all enter the cuvette,
click the vertical position "Y", the reagent&sample probe is lifted above the cuvette, then click the
zero position "Y", and the reagent&sample probe swings back to the initial position.
13.8.5 Replacing the stirrer
If the stirrer is damaged or scrapped, it cannot be repaired. or if it is bent and scrapped, it should
be replaced in time to avoid affecting the test.
Purpose
Replace the stirrer.
Maintenance opportunity
Perform the maintenance when the stirrer is damaged and scrapped and cannot be repaired.
Maintenance supplies
Alkaline cleaning solution, pure water, clean gauze, new stirrer, Phillips screwdriver
Instrument status
When performing this maintenance operation, make sure the instrument is in an idle or faulty
status.
Precautions
Warning
maintenance-29
maintenance
maintenance-30
maintenance
Mixing arm
M2 screw Stirrer
Figure 13-34 Installing the stirrer
Note: The stirrer pin must be securely secured to prevent instrument malfunction.
8. Lift the stirring arm to the top and rotate the stirring arm in the direction of the reaction cup by
hand. The distance between the agitating stirrer and the cuvette is approximately 1.3 cm.
9. Turn on the instrument power switch and initialize the instrument. /0, in the menu bar
"Maintenance", open the "Instrument Detection" form, find the mixing arm, as shown below, click
the reaction cup "Y", check the level of the reaction cup of the stirrer, confirm whether the
position of the stirrer is Located in the center of the reaction cup, if you click the zero position
“Y” in the center, the stirrer can swing back to the top of the cleaning tank.
Stirrer
Cuvettes
Figure 13-36
Note: If you are not in the center of the cuvette, please contact the customer service engineer.
11. Vertical position of the stirrer: In the “Instrument Detection” form, click the reaction cup
“Y”, the stirrer is placed to the left of the reaction cup, then click the vertical position “Y”, the
stirrer is lowered, and then click Vertical position "Y", the stirrer is lifted, and finally click the
zero position "Y", the stirrer can swing back to the top of the cleaning tank.
13.8.6 Replacing the Reaction Cup
When the cuvette is contaminated with stains such as serum or debris, scratches or cracks, it
will affect the accuracy of the test absorbance. Therefore, it is necessary to test the reaction cup. If
the cup is found to be abnormal, it should be replaced in time.
Purpose
Make sure the cuvette is normal, free from contamination, scratches or cracks.
maintenance-31
maintenance
Maintenance opportunity
Replacement of the cuvette is an occasional maintenance. Replace it in time when:
After performing the reaction cup test, the cup was found to be abnormal.
After performing the cuvette flushing operation, the cuvette is still unusable.
I found that the light passing surface of the reaction cup was scratched or broken.
Maintenance supplies
Fiber-free gloves, dry dust-free fiber cloth or gauze, spare reaction cup
Instrument status
Make sure the analyzer is powered off during this maintenance.
Precautions:
Biological infection risk:
During maintenance work, be sure to wear gloves, work clothes to
prevent infection, and wear protective glasses if necessary.
Steps
1. Turn off the main power switch of the instrument and remove the reaction disk cover.
2. Unscrew the fastening screw, as shown in Figure 13-37.
Figure 13-37
3. Use the index finger and thumb to remove the reaction cup of the corresponding cup along the
reaction disk, and find the corresponding cup according to the cup number silk screen.
Bit
4. Put the supplied reaction cup or cleaned reaction cup into the reaction tray and press it to the
bottom of the cup.
stop.
5. Put the corresponding joint into the corresponding position of the reaction tray and tighten the
fixing screws.
Warning
When installing the reaction cup, be careful not to scratch the reaction
cup. Do not touch the middle and lower parts of the light-passing surface
of the reaction cup. Otherwise, the light-passing surface will be
contaminated, which will result in inaccurate absorbance data.
When installing the cuvette, ensure that the light passing surface of the
reaction cup is placed along the radial direction of the reaction disk.
Always use gloves without fibers and powders during operation to
ensure that the clear side of the cuvette is not contaminated.
maintenance-32
maintenance
This chapter describes how to view the fault log and how to determine the source and solution of
the fault in the event of a failure. Read this chapter carefully and master the fault identification
method to help you better use the instrument.
maintenance-33
Alarm and fault handling
14.1.1 Introduction
When the instrument fails, it will be manifested in various ways. The following sections describe
troubleshooting methods to guide you through troubleshooting and troubleshooting when you find
an instrument failure.
In general, troubleshooting requires the following steps:
.
instrument.
.
implement solutions and implement effective solutions.
The reagent&sample tip is dirty. Wipe the probe with a cotton swab
Reagent&sample tip The pipe or filler of the dipped in an alkaline cleaning solution.
with water droplets sampling and filling mechanism Perform maintenance checks.
has leakage or full bubble
Check the interface area and vent the
The cleaning mechanism line water pipe.
Water drops on the leaks or is completely bubbled. Perform maintenance on the cleaning
cleaning needle The nozzle and pipeline are mechanism. If you need to replace the
blocked. hose, please contact customer service
engineer.
1. Carry out maintenance of the cleaning
No water flowing out 1. The nozzle and pipeline are mechanism. If you want to replace the
of the cleaning nozzle blocked. hose, please contact customer service
engineer.
1. Carry out maintenance of the cleaning
1. The nozzle of the cleaning
Water overflow in the mechanism. If you want to replace the
mechanism and the pipeline are
cuvette hose, please contact customer service
blocked.
engineer.
1. The interface part is not 1. Confirm the leak and reinstall it.
Syringe pump leakage properly installed. 2. Replace the syringe pump.
2. Water leakage in the pump.
Confirm that the air enters and reinstall.
Perform exhaust in the system
1. The interface part is not
maintenance. If there are tiny bubbles
There are bubbles in properly installed.
that cannot be removed, you can gently
the syringe pump 2. The filling device is not fully
tap the syringe pump while the reagent
exhausted.
or washing water is flowing, and use
vibration to eliminate it.
1. Poor contact of the level plate
Check if the level plate interface line is
interface.
in good contact.
Abnormal liquid level 2. There is a problem with the
Check if the grounding is connected.
detection instrument grounding.
Check for large electromagnetic
3. There is a large
interference around.
electromagnetic interference.
Alarm Alarm
description Handling suggestions
number source
Insufficient water
outside the instrument
Net water Add enough clean water to the outside of
2305 results in insufficient
bucket the instrument.
water in the clean
water tank.
Net water The water bucket is
2306 Suspend the water purifier.
bucket too full.
Send data timeout to
3335 system Please link with the LIS is normal.
LIS serial port.
The water gap is not
Reaction up to standard and the
3584 Clean the cuvette.
tray reaction cup may have
dirt.
Abnormal The bulb energy is Check the lamp and retest. If it still alarms,
4097
lamp abnormal. please contact the maintenance staff.
Data
collection
Did not receive the
4098 communic It is recommended to check the line.
maximum cup number.
ation is
abnormal
Data
collection
The OD module does
4099 communic It is recommended to check the line.
not adjust the gain.
ation is
abnormal
Instrument Start test signal Check the sample and reagent balance and
4353
failure timeout. re-run after reset.
Instrument The reaction disk stop
4354 It is recommended to re-run after reset.
failure signal 1 times out.
Instrument The cleaning arm
4356 It is recommended to re-run after reset.
failure action timed out.
Instrument
4358 Timer 1 timed out. It is recommended to re-run after reset.
failure
Instrument The reaction disk stop
4359 It is recommended to re-run after reset.
failure signal 2 times out.
Instrument The sample arm action Check the sample and reagent balance and
4360
failure timed out. re-run after reset.
Instrument The stirring arm action
4362 It is recommended to re-run after reset.
failure 2 times out.
Reagent bit = {0}, bar
Reagent Please re-scan after adding project
200015 code = {1}: unknown
tray parameters
new reagent!
Reagent bit = {0}, bar
Reagent Please confirm the reagent barcode and
200016 code = {1}: Reagent
tray rescan
bar code is invalid!
Alarm Alarm
description Handling suggestions
number source
An error occurred Please check the corresponding reagent
data
300000 while querying the parameter settings to ensure that the project
processing
reference value. reference values are all set.
An error occurred
data Please check the corresponding reagent
300001 while calculating the
processing parameter settings.
test request result.
An error occurred Please check if the database service is
data
300002 while generating the correct or contact customer service
processing
test request. engineer.
An error occurred
data while calculating the Please check the corresponding reagent
300003
processing project results for the parameter settings.
sample.
An error occurred Please check if the database service is
data
300004 while updating the correct or contact customer service
processing
sample status. engineer.
The test result is out of
the valid range. Time
data = {0}, sample number Please check the status of the corresponding
300007
processing = {1}, item code = sample on the sample tray.
{2}, repeat number =
{3}.
Please check the previous fitting parameters
An error occurred
data and status of this item and ensure that the
300008 while auto-scaling the
processing number of standards is sufficient for the
reagent {0}: {1}.
current fit.
data The reagent bottle is Please check the reagent bottles at these
300009
processing empty, position: {0}. locations.
data Electrolyte calibration Please apply for electrolyte calibration
300018
processing failed again.
The requested barcode
data Please confirm that the LIS return
300020 {0} does not contain
processing information is correct.
any test items
Failed to send data to
data
300023 LIS via database. Please check if the LIS link is working.
processing
Sample number = {0}
The version of the
upper computer and
Communi the lower computer do
400022 cation not match, the upper Contact customer service engineer
board computer version {0},
the lower computer
version {1}
data Insufficient sample
400070 Add sample
processing size for {0} location
Alarm Alarm
description Handling suggestions
number source
data No sample at {0}
400071 Add sample
processing location
data Insufficient reagent
400072 Add reagent
processing volume
data
400073 No reagent Add reagent
processing
Sample Sample driver module
400074 driver communication is Please check the sample driver module
module abnormal
Reagent Reagent R1 driver
400075 R1 drive module Please check the reagent R1 drive module
module communication error
Reaction Reaction disk module
400076 disk communication is Please check the reaction panel module
module abnormal
AD module
AD
400077 communication is Please check the AD module
module
abnormal
Exhaust is not
400078 Circuit Please check the circuit
completed
Network connection Please turn off the PC software first, then
error, network card is start the instrument until the network
60001 adapter
disabled or not connection is normal, then re-run the PC
connected software.
Communi Please close the PC software first, then
Communication is
60002 cation restart the instrument, and then re-run the
abnormal!
board PC software.
Communi
The data return length Please check the communication line or
60003 cation
is wrong! communication board.
board
Please check the communication line and
60004 adapter Not online!
re-run the PC software.
Please turn off the PC software, then start
The port number is
60005 adapter the instrument until the network connection
off!
is normal, and then re-run the PC software.
Please turn off the PC software, then start
The network
60006 adapter the instrument until the network connection
connection is broken!
is normal, and then re-run the PC software.
Communi The software does not
Please contact customer service engineer to
60010 cation match the instrument
re-configure the software.
board number!
When the probe is hit,
the instrument moves
60011 firing pin abnormally and the Please contact customer service engineer.
sample loading has
stopped.
Alarm Alarm
description Handling suggestions
number source
Reagent information Please check the parameters of the reagent
Reagent
100000 was not found. and confirm that the reagent has been laid
tray
Reagent code = {0}. out on the reagent tray.
Test An error occurred Please check if the communication line is
100001 applicatio while downloading the normal and make sure the instrument is in
n test request. standby or working normally.
An error occurred
Test Please check if the communication line is
while the instrument
100002 applicatio normal and make sure the instrument is in
was starting to test the
n standby.
sample.
An error occurred
Please check the database parameters and
data while loading and
100010 status and contact customer service
processing refreshing the sample
engineers.
tray status.
Please check the database parameters and
data
100011 Auto review failed. status and contact customer service
processing
engineers.
data Invalid fit type code. Please check the previous fitting parameters
100012
processing Item = {0}. and status of this item.
The reaction tray
temperature is too low
Reaction
100013 to ensure that the test
tray
results are correct and
the test has stopped.
The reaction tray
temperature is too high
Reaction
100014 to ensure that the test
tray
results are correct and
the test has stopped.
The reaction tray
Reaction temperature is low and
100015
tray the test results may be
affected.
The reaction tray
Reaction temperature is too high
100016
tray and the test results
may be affected.
The reaction tray
Reaction temperature sensor is
100017
tray not connected or is
damaged.
data Invalid instrument Please try to reinstall the program or contact
100022
processing type. a customer service engineer.
Loading The loading arm is Please reboot or contact customer service
100023
arm initialized incorrectly. engineer.
Alarm Alarm
description Handling suggestions
number source
Reaction Reaction disk moving Please reboot or contact customer service
100024
tray part initialization error engineer.
Mixing Stirring arm Please reboot or contact customer service
100025
arm initialization error engineer.
Unable to get alarm
Please check if the database service is
basic information from
100026 other correct or contact customer service
database <=
engineer.
AlarmID={0}
data Verification failed Please try to reinstall the program or contact
100031
processing after initialization a customer service engineer.
Instrument It is recommended to restart the lower
100032 other
initialization failed. computer.
The cup blank value is
Reaction below the lower limit Please clean the cuvette and replace the
200000
tray and the reaction cup cuvette if necessary.
number = {0}.
The cup blank value is
Reaction above the upper limit Please clean the cuvette and replace the
200001
tray and the reaction cup cuvette if necessary.
number = {0}.
Stirring Abnormal mixing
400080 drive module Please check the mixing module
module communication
43
Appendix B Liquid Road Connection Diagram
44
Appendix C Product Support Reagents
See the table below (Xin Beisi Company's commonly used biochemical reagent parameter table).
Analy
Sampl Sub-E Norma
Sample R1 R2 Primary Wa sis Deci-mal Max Pri-Star Pri-End Sub-Sta
Item eType Unit Min Abs nd l High
Volume Volume Volume vele-ngth Meth Place Abs t Point Point rt Point
s Point Value
od
One
ALB 5 300 0 578 Point serum 1 g/L 0 3.3 30 30 0 0 55
End
Rate
ALP 6 240 60 405 metho serum 0 U/L 0 3.3 22 30 0 0 135
d
Rate
ALT 22 240 60 340 metho serum 1 U/L 0 3.3 23 33 0 0 41
d
Rate
AMY 7 250 90 405 metho serum 0 U/L 0 3.3 23 33 0 0 104
d
Two
ApoA1 5 225 75 340 Point serum 2 g/L 0 3.3 35 35 12 13 1.9
End
Two
ApoB 5 225 75 340 Point serum 2 g/L 0 3.3 35 35 12 13 1.5
End
Two
ASO 5 240 60 578 Point serum 0 IU/mL 0 3.3 28 29 19 20 166
End
45
Analy
Sampl Sub-E Norma
Sample R1 R2 Primary Wa sis Deci-mal Max Pri-Star Pri-End Sub-Sta
Item eType Unit Min Abs nd l High
Volume Volume Volume vele-ngth Meth Place Abs t Point Point rt Point
s Point Value
od
metho
d
Fixed
time
BMG 5 225 75 578 serum 1 mg/L 0 3.3 21 28 0 0 1.8
metho
d
Rate
CHE 5 250 50 405 metho serum 0 U/L 0 3.3 22 30 0 0 12600
d
One
CHO 4 300 0 510 Point serum 2 mmol/L 0 3.3 15 15 0 0 5.2
End
Rate
CK 15 240 60 340 metho serum 0 U/L 0 3.3 22 30 0 0 190
d
Rate
CK-MB 15 240 60 340 metho serum 1 U/L 0 3.3 22 30 0 0 25
d
Fixed
time
CREA 15 240 60 510 serum 1 umol/L 0 3.3 22 30 0 0 115
metho
d
Two
CRP 20 225 75 340 serum 2 mg/dL 0 3.3 35 35 12 13 0.8
Point
46
Analy
Sampl Sub-E Norma
Sample R1 R2 Primary Wa sis Deci-mal Max Pri-Star Pri-End Sub-Sta
Item eType Unit Min Abs nd l High
Volume Volume Volume vele-ngth Meth Place Abs t Point Point rt Point
s Point Value
od
End
Two
DBIL 9 240 60 450 Point serum 2 umol/L 0 3.3 35 35 12 13 6.8
End
One
GLU 6 300 0 510 Point serum 2 mmol/L 0 3.3 34 35 0 0 6.4
End
Two
origina
HbA1c 10 225 75 630 Point 2 % 0 3.3 33 33 20 21 5.8
l blood
End
Rate
HCY 24 240 60 340 metho serum 1 umol/L 0 3.3 21 29 0 0 15
d
Two
HDL-C 5 225 75 546 Point serum 2 mmol/L 0 3.3 35 35 12 13 2.25
End
Rate
LDH 6 240 60 340 metho serum 0 U/L 0 3.3 22 30 0 0 225
d
Two
LDL-C 5 225 75 546 Point serum 2 mmol/L 0 3.3 35 35 12 13 3.35
End
Fixed
TBA 5 225 75 405 serum 1 umol/L 0 3.3 22 28 0 0 20
time
47
Analy
Sampl Sub-E Norma
Sample R1 R2 Primary Wa sis Deci-mal Max Pri-Star Pri-End Sub-Sta
Item eType Unit Min Abs nd l High
Volume Volume Volume vele-ngth Meth Place Abs t Point Point rt Point
s Point Value
od
metho
d
Two
TBIL 24 9 225 75 Point serum 0 umol/L
End
0 3.3 35 35 12 13 20.5
One
TG 5 300 0 510 Point serum 2 mmol/L 0 3.3 35 35 0 0 1.7
End
Two
UA 5 240 60 546 Point serum 0 umol/L 0 3.3 35 35 12 13 480
End
Fixed
time
UREA 5 225 75 340 serum 1 mmol/L 0 3.3 22 30 0 0 8.3
metho
d
One
TP 6 300 0 546 Point serum 1 g/L 0 3.3 34 35 0 0 88
End
Rate
α-HBDH 5 240 60 340 metho serum 0 U/L 0 3.3 22 30 0 0 182
d
Rate
GGT/γ-GT 6 225 75 405 metho serum 0 U/L 0 3.3 22 30 0 0 47
d
48
49
Appendix D: Cross Contamination Reference Sheet
Proj
Method
ect Method
principl Method principle
nam principle
e
e
Oxidase Enzymatic cycling
TG → TBA
methods assay
Oxidase Enzymatic cycling
TC → TBA
methods assay
CH Substrate
→ TG Oxidase method
E hydrolysis
LD Direct method of GLU(O
→ Oxidase method
L-C determination X)
HD Direct method of GLU(O
→ Oxidase method
L-C determination X)
The above cross-contamination only is taken for example when BIOBASE reagent is
tested on the BIOBASE series of Auto Chemical Analyzer.
The reagent formula‘s changing in cross-contamination, so the above test is only for
reference, if not, please refer to the actual test situation.
50
Jinan Biobase Biotech Co., Ltd
Biobase Biodustry (Shandong) Co., Ltd
Add: NO. 51 South Gongye Road, Jinan, China
Tel: +86-531-81219801/03 Fax: +86-531-81219804
E-mail: export@biobase.com
Web: www.biobase.com/www.biobase.cc
51