You are on page 1of 8

Original Article

Proc IMechE Part H:


J Engineering in Medicine
2021, Vol. 235(5) 515–522
Total knee replacement wear during Ó IMechE 2021
Article reuse guidelines:
simulated gait with mechanical and sagepub.com/journals-permissions
DOI: 10.1177/0954411921991269

anatomic alignments journals.sagepub.com/home/pih

Chase Maag , Ioan Cracaoanu, Jason Langhorn and Mark Heldreth

Abstract
Total knee replacements (TKR) have historically been implanted perpendicular to the mechanical axis of the knee joint,
with a commensurate external rotation of the femur in flexion relative to the posterior condylar axis (PCA). Although
this mechanical alignment (MA) method has typically offered good long-term survivorship of implants, it may result in
alignment of the implant that departs significantly from the native Joint Line (JL) in extension and flexion for a consider-
able portion of the patient population. There is a growing interest with surgeons to implant TKR components more
closely aligned to the natural JL (Anatomic Alignment-AA) of the patient’s knee joint to reduce the need for soft tissue
releases during surgery, potentially improving knee function and patient satisfaction. Using a previously-validated finite
element model of the lower extremity, implant- and alignment-specific loading conditions were developed and applied in
a wear experiment via a six-degree-of-freedom joint simulator. MA was defined as 0° Joint Line (JL), 0° varus hip-knee-
ankle (HKA) angle, and 3° external femoral rotation. AA was defined as 5° varus JL, 3° varus HKA, and 0° femoral rota-
tion. The experiment returned wear rates of 3.76 6 0.51 mg/million cycles (Mcyc) and 2.59 6 2.11 mg/Mcyc for
ATTUNEÒ cruciate-retaining (CR) fixed bearing (FB) in MA and AA, respectively. For ATTUNE posterior-stabilized (PS)
FB in AA, the wear rate was 0.97 6 1.11 mg/Mcyc. For ATTUNE CR rotating platform (RP), the wear rates were
0.23 6 0.19 mg/Mcyc, 0.48 6 1.02 mg/Mcyc in MA and AA respectively. Using a two-way ANOVA, it was determined that
there was no significantly difference in the wear rates between AA and MA (p = 0.144) nor the wear rate of ATTUNE PS
FB in AA significantly different from either ATTUNE CR FB or ATTUNE CR RP.

Keywords
Total knee replacement, wear testing, experimental simulator, kinematics, load control

Date received: 13 July 2020; accepted: 6 January 2021

Introduction degrees-of-freedom (DOF) possible in the knee joint;


flexion, anterior-posterior (A-P) displacement, interior-
Total knee replacement (TKR) is a surgical process in exterior (I-E) rotation, and vertical load leaving the
which a patient’s damaged knee joint is replaced with a other two axes uncontrolled (Varus-Valgus (V-V) and
prosthetic implant. Approximately 700,000 TKR sur- Medial-Lateral (M-L)).3,4 By only controlling four axes
geries are performed annually in the United States; this these protocols are only valid for testing implants in a
number is projected to grow due to the increased preva- simulation of mechanical alignment (MA), with com-
lence of degenerative joint diseases such as arthritis, ponents aligned perpendicular to the neutral mechani-
and the success of prosthetic knee implants in improv-
cal axis of the human knee in extension (0° Hip-Knee-
ing a patient’s function and quality of life.1,2 Predictive
Ankle (HKA) alignment, 0° Joint Line (JL)) and
wear testing is a necessary pre-clinical screening test to
femoral posterior condyles parallel to the insert during
ensure the efficacy of these total joint replacement
flexion (Figure 1). With improvements in materials,
devices. Joint replacement wear testing is typically per-
formed in vitro on experimental simulators with inputs
derived to reproduce normal human gait and published DePuy Synthes Joint Reconstruction, Warsaw, IN, USA
by international standards organizations such as ISO
Corresponding author:
and ASTM.3–5 Chase Maag, DePuy Synthes Joint Reconstruction, 700 Orthopaedic Dr,
Inputs for the wear testing of TKR, defined in ISO Warsaw, IN 46582-3994, USA.
14243-1:2009 and 14243-3:2014, include four of the six Email: cmaag@its.jnj.com
516 Proc IMechE Part H: J Engineering in Medicine 235(5)

Figure 1. Mechanical alignment diagram (top) and anatomic


alignment diagram (bottom) (Depuy-Synthes surgical technique
DSUS/JRC/0617/2179).

process, and design, implant wear has become increas-


ingly robust to the standard ISO wear protocol. This
has led to the need for more clinically relevant wear Figure 2. Attune constructs - CR fixed bearing (top left), PS
loading profiles to be developed in order to exercise fixed bearing (TOP RIGHT), CR rotating platform (bottom
devices outside of these standardized gait profiles.6 An center).
effective and efficient way to produce these loading
profiles has not yet been documented.
In TKR, MA has been historically considered the Telemetric implants such as those developed by the
most ideal alignment for implant survivorship due to its Charité Universitätsmedizin, Berlin have also provided
low reported revision rates in well aligned implants.7 In a point of comparison for these models.13 While tele-
contrast to MA which aligns the implants perpendicu- metric implants have generated insight into the loading
lar to the neutral mechanical axis of the knee, an ana- of a single implant, they have the limitation of repre-
tomic alignment (AA) surgical technique aims to more senting a single alignment scenario in vivo. As such an
closely recreate the patient’s natural JL.8 Implementing advanced lower-limb FE model described by
the AA surgical technique in a varus knee typically Fitzpatrick et al.12 was used to produce implant and
results in a more varus tibial component and a more alignment specific loading conditions. A cruciate-
valgus femoral component to maintain the HKA align- retaining (CR) TKR design in both fixed bearing (FB)
ment within 0° to 3° overall varus. In addition, the and rotating platform (RP) were evaluated for wear
rotation of the femoral component in flexion is based using MA and AA alignments (Figure 2). Additionally,
upon restoring the posterior condyles parallel to the a posterior stabilized (PS) FB TKR design was evalu-
PCA. This rotation in flexion maintains harmony with ated for wear in AA and compared to the previously
the more natural JL. It is hypothesized that the AA sur- stated wear experiments. The purpose of this study was
gical technique may also reduce the amount of soft tis- to compare the loading conditions and the resulting
sue releases required to effectively balance the knee, wear of MA to that of AA in a CR FB and a CR RP
and studies have shown that the reduction in soft tissue design and to compare the AA loading and wear of the
release leads to reduced pain and improved function.9,10 CR FB design to a PS FB design. A representation of
As such, the AA surgical technique has recently been the experimental flow can be seen in Figure 3.
gaining popularity with TKR surgeons. AA technique
should not be confused with kinematics alignment
(KA). Kinematic alignment is a different technique that Methods
has a different focus and will not be discussed in this Defining AA parameters
paper.
Finite element (FE) models and their application to Anatomic alignment surgical cut parameters were
experimental simulators are important in understand- defined as 3° varus HKA angle, 5° varus JL, and 0°
ing and advancement of joint replacement devices. femoral rotation (Figure 1). The implant manufactur-
Advanced FE models such as the Kansas knee simula- er’s ATTUNEÒ Knee System Femur First measured
tor model and Denver lower-limb model have provided resection AA surgical technique, as well as the
insight into the loading and kinematics of TKRs.11,12 ATTUNE Knee System Tibia First balanced approach
Maag et al. 517

Figure 3. Computational/experimental flow chart.

patient specific alignment surgical technique, expressly


limit this technique to a maximum of 3° varus HKA
angle, no more than 3° varus JL and femoral rotation
is recommended to be held at 0°. A 5° varus JL was
chosen for this study to simulate inadvertent surgical
variation using conventional surgical instruments.14–17
These parameters are believed to produce a worst-case
implant orientation using this technique.

Computational simulation
Gait loading data (K1L-110108_1_86p) was retrieved
from the Charité Universitätsmedizin Berlin database
(Orthoload), collated and trimmed to a single cycle for
further analysis at DePuy Synthes (Warsaw, IN). It is
noted that raw data collected by Orthoload does not
describe the knee flexion/extension during activity, and
as such the flexion was measured at DePuy Synthes via
the video record of the activity undertaken by the
patient.13 Implant loads and kinematics were then pro-
cessed for Abaqus/Explicit (SIMULIA, Providence,
RI) using a custom MATLAB (MathWorks, Natick,
MA) script. This custom MATLAB script moved the
data into a useable form for the Abaqus/Explicit solver
to read. Using the previously described lower limb
model of Fitzpatrick et al.12 (Figure 4) physiological
hip, ankle, and muscle loading conditions were calcu-
lated for the gait cycle described above (Figure 5).
These physiological hip, ankle, and muscle loading con-
ditions were then applied through the same lower limb
model to an ATTUNE CR TKR in MA (0° HKA, 0°
JL, 3° femoral rotation), as well as AA (3° HKA, 5°
JL, 0° femoral rotation) and to an ATTUNE PS TKR
in the same AA alignment. All geometry was meshed
using rigid triangular elements with fundamental ele- Figure 4. Finite element model representation of the lower
ment length of 1 to 1.2 mm, which has been demon- limb.12
strated to be sufficiently fine for mesh convergence.18,19
Contact was modeled using a calibrated pressure over-
closure model detailed in Fitzpatrick et al.12 and a fric- kinematics were extracted from the simulation model.
tional coefficient of 0.04. Components were modeled as Measured kinetics and kinematics were then processed
rigid and as such have no material properties attributed through a custom MATLAB script to develop the
to them. After completion of the FE simulations the 6- implant and alignment specific joint level loading con-
DOF loads measured at the tibial interface and ditions. These loading conditions were then applied
518 Proc IMechE Part H: J Engineering in Medicine 235(5)

Figure 5. Input waveforms (load and kinematics) used on AMTI VIVO 6-DOF joint simulator.

Figure 6. AMTI vivo simulator (Top), experimental test setup-fixed bearing CR (bottom left) rotating platform CR (bottom right).

through an AMTI VIVO 6-DOF joint simulator ATTUNE CR and PS implants (Figure 6). For each of
(AMTI, Watertown, MA) for experimental wear the load cases determined for measurement (ATTUNE
testing. CR FB and RP MA, ATTUNE CR FB and RP AA,
and ATTUNE PS FB AA) four implant sets (femur,
tibial base, and polyethylene insert) were used; three
Experimental testing wear sample sets and one loaded soak sample. The
Gravimetric wear testing was performed by means of physical wear simulation was run in five axis force con-
an AMTI VIVO 6-DOF joint simulator (AMTI, trol, with flexion being the only displacement-
Watertown, MA) using production finished size 5 controlled axis, at a clinically relevant gait frequency of
Maag et al. 519

Table 1. Results.

Implant design/alignment Wear rates (mg/Mcyc)

ATTUNE CR FB MA 3.76 6 0.51


ATTUNE CR FB AA 2.59 6 2.11
ATTUNE PS FB AA 0.97 6 1.11
ATTUNE CR RP MA 0.23 6 0.19
ATTUNE CR RP AA 0.48 6 1.02

0.8 Hz. Test lubricant was bovine calf serum with a


protein concentration of approximately 17.25 g/L sup-
plemented with sodium azide (0.2%) for bacteria
growth retardation and EDTA (20 mM) to prevent cal-
cium precipitation. The temperature of the serum was
maintained at 37°C 6 2°C throughout the duration of
the test to simulate in vivo test conditions. Gravimetric
wear of the tibial inserts was quantified following meth-
ods detailed in ISO 14243-2:2009 with measurement
points being at test commencement and every 0.5 mil-
lion cycles (Mcyc) thereafter.20 Wear rates were calcu-
lated via linear regression of the gravimetric loss.

Figure 7. Average wear rate (top) and average wear (bottom)


Statistical analysis of mechanical alignment CR FB, anatomic alignment CR FB,
Statistical analysis is undertaken using a two-way mechanical alignment CR RP, anatomic alignment CR RP and
ANOVA with interaction term and a confidence interval anatomic alignment PS FB.
of 0.95. The factors of implant design and alignment
were used in the analysis. If the interaction term is components was found to be scratched during the one
deemed to be insignificant it will be removed, and analy- Mcyc measurement interval. It is believed that this was
sis will proceed using a two-way ANOVA. Additionally, caused by a fluid containment failure that inadvertently
in-order to understand how the implant designs compare happened during one of the testing intervals. This led
a Tukey pairwise comparison will be done. to increased wear of this sample set and as such wear
rates were calculated using the two remaining samples
rather than all three.
Results
Computational simulation
Statistical analysis
Using the lower limb model previously described, inde-
Initially the interaction of implant design and alignment
pendent implant and alignment loading profiles were
were checked for significance, the p value was greater
created for the AMTI VIVO joint simulator. These
than 0.05 and found to not be significant. As such, the
independent loading profiles can be found in Figure 5.
interaction term of the two-way ANOVA was removed
The lower limb model was successfully used to translate
from the final analysis. The analysis determined that
telemetric implant data to the specific implant and
there is no significant difference between MA and AA
alignments being used in the wear experiment.
(p = 0.144) while there was a significant difference
between implant designs (p = 0.003). Investigating the
Experimental testing implant design differences via a Tukey pairwise com-
parison, CR RP had significantly lower wear rates than
The AMTI VIVO 6-DOF joint simulator was observed
CR FB while both CR FB and CR RP did not have sig-
to effectively reproduce the loading profiles generated
nificant differences with PS FB. Wear scars were photo-
by the lower limb model procedure described above.
graphed at the end of testing and can be found in
Following 4.5 million cycles (Mcyc) of wear testing the
Table 2.
linear wear rates were calculated to be 3.76 6 0.51 mg/
Mcyc, 2.59 6 2.11 mg/Mcyc, 0.23 6 0.19 mg/Mcyc,
0.48 6 1.02 mg/Mcyc, and 0.97 6 1.11 mg/Mcyc for
Discussion
ATTUNE CR FB MA, ATTUNE CR FB AA,
ATTUNE CR RP MA, ATTUNE CR RP AA, and Testing of orthopaedic knee implants in alignments
ATTUNE PS FB AA; respectively (Table 1). Average other than MA have not been extensively reported in
wear for all sample groups can be seen in Figure 7. It is literature. It is believed that this is a consequence of the
noted that one of the ATTUNE CR FB MA femoral inherent difficulty running in vitro testing in alignments
520 Proc IMechE Part H: J Engineering in Medicine 235(5)

Table 2. Sample group wear scars.

Sample group Wear scar


Medial Lateral

Mechanical
alignment CR FB

Figure 8 Comparison of compartmental loading based on


percent medial loading.

Anatomic
other than MA. Utilizing a computational model to
alignment CR FB calculate loading at the interface of the tibial implant
enabled the 6-DOF simulator to induce the change in
alignment from MA to AA without changing the physi-
cal fixturing. This study has shown that a combination
of computational and experimental techniques can be
readily used to better understand the wear and loading
of different alignments in the knee joint. Coupling com-
putational and experimental in this fashion may aid in
reducing design phase iterations, and better understand
how variations in implant orientation may affect per-
Anatomic formance in vivo.
alignment PS FB Wears scars presented in Table 2 give insight into
how the different implants and alignments effect wear
on the polymer inserts. One of the first comparisons
that can be drawn from the wear scars are changes
between mechanical and anatomic alignment of the CR
FB implants. It can be seen from that wear scar outline
that the anatomic alignment wear scar has much more
lateral movement when compared to the mechanical
alignment wear scar. This is an expected outcome when
changing the alignment methods (Figure 8). Anatomic
Mechanical alignment targets the patients pre-operative state and as
alignment CR RP such this implant alignment produces kinematics more
similar to the natural knee. It is noted that the natural
knee tends to have an increased medial pivot move-
ment, where the medial condyle experiences less A-P
movement than the lateral condyle.8,21,22 When com-
paring wear scars between MA and AA alignment for
the CR RP inserts, a similar pattern is found. Anatomic
alignment PS inserts do not seem to exhibit the same
wear pattern, and it is surmised that this may be due to
PCL removal in the finite element simulations and/or
Anatomic post-cam interactions in the PS design.
alignment CR RP
It is noted that this technique would make it possible
to produce loading profiles for additional alignment sce-
narios or additional activities and investigate other
potential consequences of implant alignment such as
implant fixation. This technique may also offer insight
on the effect of implant alignment on ligamentous struc-
tures in the knee joint. While in this study the ligamen-
tous structure insertion points were held constant, it may
be possible to perturb these locations to see the effects
on tibial loads with difference alignments and implants.
Maag et al. 521

These aspects will be investigated in future studies. Ioan Cracaoanu: Investigation, Formal Analysis, Data
Another application of this technique that is not dis- Curation
cussed is investigation into mal-alignment. Implants may Jason Langhorn: Writing – Review & Editing,
not always be perfectly aligned when initially implanted Supervision, Resources
or may migrate after implantation. As such there may be Mark Heldreth: Conceptualization, Project administra-
an interest in studying the effects of mal-alignment on tion, Writing – Review & Editing
the implant. Mal-alignment is suspect for early fixation All Authors have read and approved the final sub-
failure, increased wear, and other early revision factors. mitted manuscript.
Mal-alignment was not being directly investigated in this
study but using this technique it is possible to implant in Declaration of conflicting interests
any alignment position to see the outcome in the joint.
The author(s) declared the following potential conflicts
Ligamentous structure information could be extracted
of interest with respect to the research, authorship, and/
from the model, along with tibial interface loading and
or publication of this article: Four of the authors (CM,
its effects on the cement mantle. Investigation and better
IC, JL, MH) are employees of DePuy Synthes.
understanding of the joint effects of mal-alignment may
be studied in the future.
This technique, as all do, has its limitations; the pri- Funding
mary example being it is limited to only being able to The author(s) received no financial support for the
produce loading profiles from already recorded in vivo research, authorship, and/or publication of this article.
measured tibial plateau loads. Though this limitation is
present, if external boundary conditions for an activity
ORCID iD
were known for an activity one could apply those to the
model to find the loads at the tibial interface. There is Chase Maag https://orcid.org/0000-0003-1246-987X
also an assumption that the external boundary condi-
tions of the Denver lower-limb model do not change References
when switching implants and alignments. This assump-
1. Williams SN, Wolford ML and Bercovitz A. Hospitaliza-
tion is justified by the fact that if the same patient per- tion for total knee replacement among inpatients aged 45
forms that activity (with a different implant/different and over: United States, 2000–2010. NCHS data brief, no
alignment) then the external boundary conditions allow 210. 2015. Hyattsville, MD: National Center for Health
that activity to be achieved. One could argue that as the Statistics.
alignment of the implants change, the body (external 2. American Academy of Orthopaedic Surgeons. ‘‘98 percent
boundary conditions) will realign to reduce the inherit of total knee replacement patients return to life, work fol-
force balances in the joint. This possibility is acknowl- lowing surgery.’’ ScienceDaily, www.sciencedaily.com/
edged, however by not rebalancing for the new align- releases/2013/03/130321082857.htm (2013, accessed 10
ment these conditions would be more of a ‘‘worse’’ case March 2020).
3. ISO 14243:2009. Implants for surgery -wear of total
condition. In the lower limb model the knee insert is
knee-joint prosteheses. Part 1: loading and displacement
considered a rigid body to reduce computation time
parameters for wear-testing machines with load control
necessary to execute. In the situation where kinematics and corresponding environmental conditions for test.
and kinetics are of interest rather than wear metrics this 4. ISO 14243:2014. Implants for surgery -wear of total
is an appropriate simplification. knee-joint prosteheses. Part 3: loading and displacement
From the results detailed in this publication it can parameters for wear-testing machines with displacement
be concluded that AA, as defined in this study, does control and corresponding environmental conditions for
not adversely affect the wear rates of ATTUNE CR tests.
FB or RP when compared to MA and measured using 5. ASTM F3141-17a, Standard guide for total knee replacement
the AMTI VIVO 6-DOF simulator. Additionally, it loading profiles. West Conshohocken, PA: ASTM Interna-
was found that ATTUNE CR RP had a significantly tional. https://www.astm.org/Standards/F3141.htm (2017).
6. Reinders J, Sonntag R, Vot L, et al. Wear testing of
lower wear rate than that of ATTUNE CR FB; this
moderate activities of daily living using in vivo measured
may be a result of decoupling the internal-external
knee joint loading. Plos One 2015; 10(3): e0123155.
rotational torque in the RP implants. It can also be 7. Bonner TJ, Eardley WGP, Patterson P, et al. The effect
concluded that the wear rate of ATTUNE PS FB in of post-operative mechanical axis alignment on the sur-
AA not significantly different from either ATTUNE vival of primary total knee replacements after a follow-
CR FB or ATTUNE CR RP. up of 15 years. J Bone Joint Surg 2011; 93-B: 1217–1222.
8. Cherian J, Kapadia B, Banerjee S, et al. Mechanical,
anatomical, and kinematic axis in TKA: concepts and
Author contributions
practical applications. Curr Rev Musculoskelet Med 2014;
Chase Maag: Formal Analysis, Project Administration, 7: 89–95.
Software, Methodology, Investigation, Data Curation, 9. Dossett HG, Estrada NA, Swartz GJ, et al. A rando-
Visualization mized controlled trial of kinematically and mechanically
522 Proc IMechE Part H: J Engineering in Medicine 235(5)

aligned total knee replacements. Bone Joint J 2014; 96-B: 15. Bankes MJ, Back DL, Cannon SR, et al. The effect of
907–913. component malalignment on the clinical and radiological
10. Howell SM, Papadopoulos S, Kuznik KT, et al. Accu- outcome of the kinemax total knee replacement. Knee
rate alignment and high function after kinematically 2003; 10: 55–60.
aligned TKA performed with generic instruments. 16. Depuy-Synthes surgical technique DSUS/JRC/0617/2179
Knee Surg Sports Traumatol Arthrosc 2013b; 21: 2271– 17. Depuy-Synthes surgical technique 097081-181130
2280. 18. Godest AC, Beaugonin M, Haug E, et al. Simulation of
11. Baldwin MA, Clary CW, Fitzpatrick CK, et al. Dynamic a knee joint replacement during a gait cycle using explicit
finite element knee simulation for evaluation of knee finite element analysis. J Biomech 2002; 35: 267–275.
replacement mechanics. J Biomech 2012; 45: 474–483. 19. Halloran JP, Petrella AJ and Rullkoetter PJ. Explicit
12. Fitzpatrick CK, Baldwin MA, Clary CW, et al. Evaluat- finite element modeling of total knee replacement
ing knee replacement mechanics during ADL with PID- mechanics. J Biomech 2005a; 38: 323–331.
controlled dynamic finite element analysis. Comput Meth- 20. ISO 14243:2009. Implants for surgery -wear of total knee-
ods Biomech Biomed Eng 2012; 17: 360–369. joint prosteheses. Part 2: methods of measurement.
13. Bergmann G (Ed.) Charite-Universitaetsmedizin Berlin 21. Lee YS, Howell SM, Won YY, et al. Kinematic align-
(2008) ‘‘Orthoload’’. http://Orthoload.com (2008, ment is a possible alternative to mechanical alignment in
accessed 1 July 2011). total knee arthroplasty. Knee Surg Sports Traumatol
14. Barrett W, Hoeffel D, Dalury D, et al. In-vivo alignment Arthrosc 2017; 25(11): 3467–3479.
comparing patient specific instrumentation with both 22. Roussot MA, Vles GF and Oussedik S. Clinical out-
conventional and Computer Assisted Surgery (CAS) comes of kinematic alignment versus mechanical align-
instrumentation in total knee arthroplasty. J Arthroplasty ment in total knee arthroplasty: a systematic review.
2014; 29: 343–347. EFFORT Open Rev 2020; 5: 486–497.

You might also like