Professional Documents
Culture Documents
Abstract
Total knee replacements (TKR) have historically been implanted perpendicular to the mechanical axis of the knee joint,
with a commensurate external rotation of the femur in flexion relative to the posterior condylar axis (PCA). Although
this mechanical alignment (MA) method has typically offered good long-term survivorship of implants, it may result in
alignment of the implant that departs significantly from the native Joint Line (JL) in extension and flexion for a consider-
able portion of the patient population. There is a growing interest with surgeons to implant TKR components more
closely aligned to the natural JL (Anatomic Alignment-AA) of the patient’s knee joint to reduce the need for soft tissue
releases during surgery, potentially improving knee function and patient satisfaction. Using a previously-validated finite
element model of the lower extremity, implant- and alignment-specific loading conditions were developed and applied in
a wear experiment via a six-degree-of-freedom joint simulator. MA was defined as 0° Joint Line (JL), 0° varus hip-knee-
ankle (HKA) angle, and 3° external femoral rotation. AA was defined as 5° varus JL, 3° varus HKA, and 0° femoral rota-
tion. The experiment returned wear rates of 3.76 6 0.51 mg/million cycles (Mcyc) and 2.59 6 2.11 mg/Mcyc for
ATTUNEÒ cruciate-retaining (CR) fixed bearing (FB) in MA and AA, respectively. For ATTUNE posterior-stabilized (PS)
FB in AA, the wear rate was 0.97 6 1.11 mg/Mcyc. For ATTUNE CR rotating platform (RP), the wear rates were
0.23 6 0.19 mg/Mcyc, 0.48 6 1.02 mg/Mcyc in MA and AA respectively. Using a two-way ANOVA, it was determined that
there was no significantly difference in the wear rates between AA and MA (p = 0.144) nor the wear rate of ATTUNE PS
FB in AA significantly different from either ATTUNE CR FB or ATTUNE CR RP.
Keywords
Total knee replacement, wear testing, experimental simulator, kinematics, load control
Computational simulation
Gait loading data (K1L-110108_1_86p) was retrieved
from the Charité Universitätsmedizin Berlin database
(Orthoload), collated and trimmed to a single cycle for
further analysis at DePuy Synthes (Warsaw, IN). It is
noted that raw data collected by Orthoload does not
describe the knee flexion/extension during activity, and
as such the flexion was measured at DePuy Synthes via
the video record of the activity undertaken by the
patient.13 Implant loads and kinematics were then pro-
cessed for Abaqus/Explicit (SIMULIA, Providence,
RI) using a custom MATLAB (MathWorks, Natick,
MA) script. This custom MATLAB script moved the
data into a useable form for the Abaqus/Explicit solver
to read. Using the previously described lower limb
model of Fitzpatrick et al.12 (Figure 4) physiological
hip, ankle, and muscle loading conditions were calcu-
lated for the gait cycle described above (Figure 5).
These physiological hip, ankle, and muscle loading con-
ditions were then applied through the same lower limb
model to an ATTUNE CR TKR in MA (0° HKA, 0°
JL, 3° femoral rotation), as well as AA (3° HKA, 5°
JL, 0° femoral rotation) and to an ATTUNE PS TKR
in the same AA alignment. All geometry was meshed
using rigid triangular elements with fundamental ele- Figure 4. Finite element model representation of the lower
ment length of 1 to 1.2 mm, which has been demon- limb.12
strated to be sufficiently fine for mesh convergence.18,19
Contact was modeled using a calibrated pressure over-
closure model detailed in Fitzpatrick et al.12 and a fric- kinematics were extracted from the simulation model.
tional coefficient of 0.04. Components were modeled as Measured kinetics and kinematics were then processed
rigid and as such have no material properties attributed through a custom MATLAB script to develop the
to them. After completion of the FE simulations the 6- implant and alignment specific joint level loading con-
DOF loads measured at the tibial interface and ditions. These loading conditions were then applied
518 Proc IMechE Part H: J Engineering in Medicine 235(5)
Figure 5. Input waveforms (load and kinematics) used on AMTI VIVO 6-DOF joint simulator.
Figure 6. AMTI vivo simulator (Top), experimental test setup-fixed bearing CR (bottom left) rotating platform CR (bottom right).
through an AMTI VIVO 6-DOF joint simulator ATTUNE CR and PS implants (Figure 6). For each of
(AMTI, Watertown, MA) for experimental wear the load cases determined for measurement (ATTUNE
testing. CR FB and RP MA, ATTUNE CR FB and RP AA,
and ATTUNE PS FB AA) four implant sets (femur,
tibial base, and polyethylene insert) were used; three
Experimental testing wear sample sets and one loaded soak sample. The
Gravimetric wear testing was performed by means of physical wear simulation was run in five axis force con-
an AMTI VIVO 6-DOF joint simulator (AMTI, trol, with flexion being the only displacement-
Watertown, MA) using production finished size 5 controlled axis, at a clinically relevant gait frequency of
Maag et al. 519
Table 1. Results.
Mechanical
alignment CR FB
Anatomic
other than MA. Utilizing a computational model to
alignment CR FB calculate loading at the interface of the tibial implant
enabled the 6-DOF simulator to induce the change in
alignment from MA to AA without changing the physi-
cal fixturing. This study has shown that a combination
of computational and experimental techniques can be
readily used to better understand the wear and loading
of different alignments in the knee joint. Coupling com-
putational and experimental in this fashion may aid in
reducing design phase iterations, and better understand
how variations in implant orientation may affect per-
Anatomic formance in vivo.
alignment PS FB Wears scars presented in Table 2 give insight into
how the different implants and alignments effect wear
on the polymer inserts. One of the first comparisons
that can be drawn from the wear scars are changes
between mechanical and anatomic alignment of the CR
FB implants. It can be seen from that wear scar outline
that the anatomic alignment wear scar has much more
lateral movement when compared to the mechanical
alignment wear scar. This is an expected outcome when
changing the alignment methods (Figure 8). Anatomic
Mechanical alignment targets the patients pre-operative state and as
alignment CR RP such this implant alignment produces kinematics more
similar to the natural knee. It is noted that the natural
knee tends to have an increased medial pivot move-
ment, where the medial condyle experiences less A-P
movement than the lateral condyle.8,21,22 When com-
paring wear scars between MA and AA alignment for
the CR RP inserts, a similar pattern is found. Anatomic
alignment PS inserts do not seem to exhibit the same
wear pattern, and it is surmised that this may be due to
PCL removal in the finite element simulations and/or
Anatomic post-cam interactions in the PS design.
alignment CR RP
It is noted that this technique would make it possible
to produce loading profiles for additional alignment sce-
narios or additional activities and investigate other
potential consequences of implant alignment such as
implant fixation. This technique may also offer insight
on the effect of implant alignment on ligamentous struc-
tures in the knee joint. While in this study the ligamen-
tous structure insertion points were held constant, it may
be possible to perturb these locations to see the effects
on tibial loads with difference alignments and implants.
Maag et al. 521
These aspects will be investigated in future studies. Ioan Cracaoanu: Investigation, Formal Analysis, Data
Another application of this technique that is not dis- Curation
cussed is investigation into mal-alignment. Implants may Jason Langhorn: Writing – Review & Editing,
not always be perfectly aligned when initially implanted Supervision, Resources
or may migrate after implantation. As such there may be Mark Heldreth: Conceptualization, Project administra-
an interest in studying the effects of mal-alignment on tion, Writing – Review & Editing
the implant. Mal-alignment is suspect for early fixation All Authors have read and approved the final sub-
failure, increased wear, and other early revision factors. mitted manuscript.
Mal-alignment was not being directly investigated in this
study but using this technique it is possible to implant in Declaration of conflicting interests
any alignment position to see the outcome in the joint.
The author(s) declared the following potential conflicts
Ligamentous structure information could be extracted
of interest with respect to the research, authorship, and/
from the model, along with tibial interface loading and
or publication of this article: Four of the authors (CM,
its effects on the cement mantle. Investigation and better
IC, JL, MH) are employees of DePuy Synthes.
understanding of the joint effects of mal-alignment may
be studied in the future.
This technique, as all do, has its limitations; the pri- Funding
mary example being it is limited to only being able to The author(s) received no financial support for the
produce loading profiles from already recorded in vivo research, authorship, and/or publication of this article.
measured tibial plateau loads. Though this limitation is
present, if external boundary conditions for an activity
ORCID iD
were known for an activity one could apply those to the
model to find the loads at the tibial interface. There is Chase Maag https://orcid.org/0000-0003-1246-987X
also an assumption that the external boundary condi-
tions of the Denver lower-limb model do not change References
when switching implants and alignments. This assump-
1. Williams SN, Wolford ML and Bercovitz A. Hospitaliza-
tion is justified by the fact that if the same patient per- tion for total knee replacement among inpatients aged 45
forms that activity (with a different implant/different and over: United States, 2000–2010. NCHS data brief, no
alignment) then the external boundary conditions allow 210. 2015. Hyattsville, MD: National Center for Health
that activity to be achieved. One could argue that as the Statistics.
alignment of the implants change, the body (external 2. American Academy of Orthopaedic Surgeons. ‘‘98 percent
boundary conditions) will realign to reduce the inherit of total knee replacement patients return to life, work fol-
force balances in the joint. This possibility is acknowl- lowing surgery.’’ ScienceDaily, www.sciencedaily.com/
edged, however by not rebalancing for the new align- releases/2013/03/130321082857.htm (2013, accessed 10
ment these conditions would be more of a ‘‘worse’’ case March 2020).
3. ISO 14243:2009. Implants for surgery -wear of total
condition. In the lower limb model the knee insert is
knee-joint prosteheses. Part 1: loading and displacement
considered a rigid body to reduce computation time
parameters for wear-testing machines with load control
necessary to execute. In the situation where kinematics and corresponding environmental conditions for test.
and kinetics are of interest rather than wear metrics this 4. ISO 14243:2014. Implants for surgery -wear of total
is an appropriate simplification. knee-joint prosteheses. Part 3: loading and displacement
From the results detailed in this publication it can parameters for wear-testing machines with displacement
be concluded that AA, as defined in this study, does control and corresponding environmental conditions for
not adversely affect the wear rates of ATTUNE CR tests.
FB or RP when compared to MA and measured using 5. ASTM F3141-17a, Standard guide for total knee replacement
the AMTI VIVO 6-DOF simulator. Additionally, it loading profiles. West Conshohocken, PA: ASTM Interna-
was found that ATTUNE CR RP had a significantly tional. https://www.astm.org/Standards/F3141.htm (2017).
6. Reinders J, Sonntag R, Vot L, et al. Wear testing of
lower wear rate than that of ATTUNE CR FB; this
moderate activities of daily living using in vivo measured
may be a result of decoupling the internal-external
knee joint loading. Plos One 2015; 10(3): e0123155.
rotational torque in the RP implants. It can also be 7. Bonner TJ, Eardley WGP, Patterson P, et al. The effect
concluded that the wear rate of ATTUNE PS FB in of post-operative mechanical axis alignment on the sur-
AA not significantly different from either ATTUNE vival of primary total knee replacements after a follow-
CR FB or ATTUNE CR RP. up of 15 years. J Bone Joint Surg 2011; 93-B: 1217–1222.
8. Cherian J, Kapadia B, Banerjee S, et al. Mechanical,
anatomical, and kinematic axis in TKA: concepts and
Author contributions
practical applications. Curr Rev Musculoskelet Med 2014;
Chase Maag: Formal Analysis, Project Administration, 7: 89–95.
Software, Methodology, Investigation, Data Curation, 9. Dossett HG, Estrada NA, Swartz GJ, et al. A rando-
Visualization mized controlled trial of kinematically and mechanically
522 Proc IMechE Part H: J Engineering in Medicine 235(5)
aligned total knee replacements. Bone Joint J 2014; 96-B: 15. Bankes MJ, Back DL, Cannon SR, et al. The effect of
907–913. component malalignment on the clinical and radiological
10. Howell SM, Papadopoulos S, Kuznik KT, et al. Accu- outcome of the kinemax total knee replacement. Knee
rate alignment and high function after kinematically 2003; 10: 55–60.
aligned TKA performed with generic instruments. 16. Depuy-Synthes surgical technique DSUS/JRC/0617/2179
Knee Surg Sports Traumatol Arthrosc 2013b; 21: 2271– 17. Depuy-Synthes surgical technique 097081-181130
2280. 18. Godest AC, Beaugonin M, Haug E, et al. Simulation of
11. Baldwin MA, Clary CW, Fitzpatrick CK, et al. Dynamic a knee joint replacement during a gait cycle using explicit
finite element knee simulation for evaluation of knee finite element analysis. J Biomech 2002; 35: 267–275.
replacement mechanics. J Biomech 2012; 45: 474–483. 19. Halloran JP, Petrella AJ and Rullkoetter PJ. Explicit
12. Fitzpatrick CK, Baldwin MA, Clary CW, et al. Evaluat- finite element modeling of total knee replacement
ing knee replacement mechanics during ADL with PID- mechanics. J Biomech 2005a; 38: 323–331.
controlled dynamic finite element analysis. Comput Meth- 20. ISO 14243:2009. Implants for surgery -wear of total knee-
ods Biomech Biomed Eng 2012; 17: 360–369. joint prosteheses. Part 2: methods of measurement.
13. Bergmann G (Ed.) Charite-Universitaetsmedizin Berlin 21. Lee YS, Howell SM, Won YY, et al. Kinematic align-
(2008) ‘‘Orthoload’’. http://Orthoload.com (2008, ment is a possible alternative to mechanical alignment in
accessed 1 July 2011). total knee arthroplasty. Knee Surg Sports Traumatol
14. Barrett W, Hoeffel D, Dalury D, et al. In-vivo alignment Arthrosc 2017; 25(11): 3467–3479.
comparing patient specific instrumentation with both 22. Roussot MA, Vles GF and Oussedik S. Clinical out-
conventional and Computer Assisted Surgery (CAS) comes of kinematic alignment versus mechanical align-
instrumentation in total knee arthroplasty. J Arthroplasty ment in total knee arthroplasty: a systematic review.
2014; 29: 343–347. EFFORT Open Rev 2020; 5: 486–497.