Professional Documents
Culture Documents
Orthopaedic
Article Surger y
Journal of Orthopaedic Surgery
27(1) 1–8
ª The Author(s) 2019
Use of smartphone to improve Article reuse guidelines:
sagepub.com/journals-permissions
acetabular component positioning DOI: 10.1177/2309499019825578
journals.sagepub.com/home/osj
Abstract
Background: The position of the acetabular cup is important to the outcome of total hip athroplasty (THA). We
devised an instrument that uses the level indicator application of smartphone together with a mechanical alignment
guide to improve the precision of cup placement. This study aims to determine the percentage of acetabular cups
positioned in the Lewinnek safe zone comparing between the conventional technique (using a mechanical alignment
guide alone) and the smartphone technique (using a mechanical alignment guide combined with the devised
instrument and smartphone). Methods: A historical controlled trial was conducted among 82 patients who
underwent primary THAs through a posterolateral approach. In the conventional group, 41 cups were placed during
January 2013 and December 2014, whereas 41 cups in the smartphone group were placed during January 2015
and March 2016. Inclination and anteversion angles were measured in standardized pelvic radiographs. The cup
orientation was compared between groups. Results: The inclination angle in the smartphone group was significantly
lower than in the conventional group (40.9 (SD 3.8) vs. 46.3 (SD 6.7), p < 0.001), but the anteversion angle was higher
(19.6 (SD 4.4) vs. 16.5 (SD 6.1), p ¼ 0.010). The smartphone group had more cups positioned in the Lewinnek safe zone
(90.2% vs. 56.1%, p ¼ 0.001) and longer operative times (136 (SD 27) vs. 119 (SD 23) min, p ¼ 0.011). No significant
difference was found for blood loss (p=¼ 0.384) or dislocation rate (p ¼ 0.494). Conclusion: Using the computerized
function of smartphone could improve the precision of cup positioning. Most cups were placed within a narrow margin
inside the Lewinnek safe zone.
Keywords
acetabular cup positioning, pelvic inclinometer, smartphone, total hip arthroplasty
Date received: 18 May 2018; Received revised 27 September 2018; accepted: 30 December 2018
Introduction
The acetabular component positioning is highly important
to the function and outcome of total hip arthroplasty Department of Orthopaedic Surgery, Lampang Hospital and Medical
(THA). Achieving the optimal inclination and anteversion Educational Center, Lampang, Thailand
could potentially improve the longevity and range of
motion of a THA and decrease the dislocation rate.1–3 Corresponding author:
Anuwat Pongkunakorn, Department of Orthopaedic Surgery, Lampang
Lewinnek et al. advocated an inclination angle of 40 +10 Hospital and Medical Educational Center, 280 Paholyothin Road,
and an anteversion angle of 15 +10 as the safe zone for Mueang District, Lampang 52000, Thailand.
cup orientation.4 These goals have been the most widely Email: dranuwat@hotmail.com
Creative Commons Non Commercial CC BY-NC: This article is distributed under the terms of the Creative Commons
Attribution-NonCommercial 4.0 License (http://www.creativecommons.org/licenses/by-nc/4.0/) which permits non-commercial
use, reproduction and distribution of the work without further permission provided the original work is attributed as specified on the SAGE and Open
Access pages (https://us.sagepub.com/en-us/nam/open-access-at-sage).
2 Journal of Orthopaedic Surgery 27(1)
used targets for several decades although there is contro- group, 41 cups were prospectively placed during January
versy in the literature regarding the ideal orientation. The 2015 and March 2016 using a mechanical alignment guide
percentage of acceptably placed cups in the Lewinnek safe combined with our invented instrument and smartphone. In
zone varies from 25.7% to 70.5% reported in the literature.5 the historical-control group, 41 cups had been placed dur-
These outcomes resulted from the conventional method of ing January 2013 and December 2014 using a mechanical
cup placement that was guided by intraoperative anatomi- alignment guide alone. Exclusion criteria were patients
cal landmarks (free hand technique) or using mechanical with pelvic deformity or severe kyphoscoliosis.
alignment guides. Computer navigation can increase the Our pelvic inclinometer used a smartphone (Lenovo
percentage of placements in the safe zone up to 81%.6 This A390, China) and Smart Level application (version 1.0,
proven advantage must be weighed up against the argument androidboy1.blogspot.com). This program shows real-
of prolonged surgery times and higher costs.7 time tilting degrees of the smartphone along both the long-
A smartphone that combines features of a personal com- itudinal and the transverse axes. The phone was enclosed in
puter operating system with other features of a mobile a zip bag (SunZip, Thailand) and placed in a holder that
phone is an interesting choice to assist in surgery. The was made from a stretchable frame and adjustable adapter
intraoperative use of smartphone technology in THA for (Yunteng self picture monopod YT-188, China) with a
improving the accuracy of acetabular cup placement was stem made from an 11-mm diametered stainless tube
reported both in cadaveric and in clinical studies. Peters (AO/ASIF, Switzerland). The base was a stainless steel
et al. used the accelerometer and camera function of an frame with two screw holes at the lower end and a fixed
iPhone in a level indicator and protractor application to standard clamp (AO/ASIF, Switzerland) at the upper end
improve cup placement in 50 prospective cases.8 They (Figure 1). This frame could be connected firmly and dis-
found good results with all cups being placed within the connected quickly during the surgery (Figure 2).
Lewinnek safe zone. Kurosaka et al. confirmed this iPhone/ There were three steps in the smartphone group:
iPad technique in five cadavers and could achieve accep- Step 1. The pelvic inclinometer installation. After set-
table performance in determining the cup alignment ting the operating table in neutral position, a skin staple
regardless of the surgeon’s expertise.9 However, these two was put over the tip of the coccyx as a radiographic mar-
studies assumed that there are neither preoperative errors in ker. The patient was positioned in the lateral decubitus
pelvic positioning nor intraoperative pelvic motion before using two supports (posteriorly over the sacrum and ante-
cup placement. These conditions are unlikely to be met riorly over the anterior superior iliac spine (ASIS)). The
during THA. skin was prepared and partially draped. The inclinometer
We developed a pelvic inclinometer that uses the func- was securely fixed to the iliac tubercle by two self-drilling
tion of an accelerometer sensor of a downloaded bubble- screws (No.12 2.5 inch, hex washer headed, zinc plated,
level application in a smartphone. This application shows Taiwan). The operative field was then covered with a
real-time tilting degrees along two axes. When it is held in sterile clear-plastic sheet.
the phone holder and fixed to the pelvis, it can monitor Step 2. Standardization of the patient’s position. The
pelvic motion and present it in a digital number. For mea- pelvic position was verified using the cross-table fluoro-
suring the cup inclination, we used another smartphone scopy. A plumb was dangled on the image-reception screen
with a downloaded angle-meter application. This smart- and a long K-wire was taped on the screen to replace this
phone was placed on top of the cup positioner, and the plumb line (Figure 3(a)). The operating table was tilted so
alignment guide was pointed parallel to the pelvic inclin- that the inter-teardrop line was parallel to the K-wire and
ometer before final seating. We wanted to investigate the coccyx marker was in line with pubic symphysis. The
whether this instrument could improve the precision of cup C-arm was rotated horizontally until the positional relation-
placement in the Lewinnek safe zone. This study aims to ship between the superior margin of pubic symphysis and
determine the percentage of acetabular cups that were cor- the tip of coccyx on the fluoroscopic image was the same as
rectly positioned in the Lewinnek safe zone, comparing the preoperative supine anteroposterior radiograph (Figure
between the conventional method (using a mechanical 3(b)). This plane was marked using a cross-line laser level
alignment guide alone) and our technique (using a mechan- (a tool commonly used for installing floor tiles) to touch the
ical alignment guide combined with the invented instru- flat surface of the C-arm. It projected a laser line, perpen-
ment and smartphone). dicular to the axis of the C-arm, on the border of the smart-
phone (Figure 4(a)). The thumb-screw knob of the
smartphone holder was loosened and tightened again after
Materials and methods the axis of the smartphone was adjusted horizontally par-
A prospective non-randomized, historical-controlled study allel to the laser line (Figure 4(b)). The Smart Level appli-
was conducted among the adult patients who underwent cation was calibrated to the relative position of the phone to
cementless or hybrid THA from January 2013 to March 0 along both axes by touching the “calibrate” screen-icon
2016 through posterolateral approach by a single surgeon. (Figure 5). The inclinometer was then disconnected from
These patients were divided into two groups. In the study its base and the surgeon finished draping (Figure 2(b)). For
Pongkunakorn et al. 3
Figure 1. The pelvic inclinometer comprises (a) a smartphone (enclosed in a zip bag) and a plastic holder (stretchable frame and
adjustable adapter) and (b) a stainless steel tube stem and base.
Figure 2. (a) The upper part of the stainless steel base was firmly connected to the lower part that was fixed into the iliac tubercle by
two screws. (b) It could be disconnected quickly during the surgery.
eliminating the positive effect of fluoroscopy, the operating pointed 5 posterior to the border of the inclinometer smart-
table was tilted back to its initial position before starting phone by visual estimation before final cup seating.
the surgery. In the control group, the patient was fixed in lateral
Step 3. Control cup positioner and alignment guide. decubitus using two supports after setting the operating
Before the final cup placement, the inclinometer was table in neutral position as in the study group. No fluoro-
reconnected to its base again, and the operating table was scopy was used to calibrate the pelvic position. Cup pla-
tilted back to return the pelvis to its standardized lateral cement was performed using a mechanical alignment
decubitus position by obtain zero tilting degrees shown in guide alone at the operative inclination about 40 . The
both axes of the Smart Level application. The other smart- guide, set at 20 anteversion, was pointed 5 posterior to
phone (Lenovo A390, China) with the Clinometer applica- the shoulder position of the patient to achieve 25 of
tion (www.plaincode.com) was placed on top and parallel operative anteversion.
to the cup positioner. To achieve the 40 target angle of In both groups, the posterior capsule and short external
radiographic inclination, the positioner was moved verti- rotators were repaired. Antibiotic prophylaxis was intrave-
cally together with the smartphone until 36 of operative nous cefazolin before skin incision and at 6-h intervals for
inclination was shown in the application (Figure 6). The 24–48 h. Two doses of tranexamic acid (500 mg) were
positioner was then moved horizontally until the mechan- given intravenously, one before skin incision and the other
ical alignment guide, set at 20 of operative anteversion, before closure.
4 Journal of Orthopaedic Surgery 27(1)
Figure 3. (a) A long K-wire was taped on the image-reception screen parallel to the plumb line. (b) Using cross-table fluoroscopy, the
operating table was tilted until the inter-teardrop line was parallel to the K-wire, and the coccyx marker was in line with pubic
symphysis. The C-arm was then rotated horizontally until the relative position of coccyx and pubic symphysis on the fluoroscopic image
was similar to the pelvic radiograph.
Figure 4. (a) The pelvic tilt in sagittal plane of the patient was marked by placing a cross-line laser level on the flat surface of the C-arm.
A laser line was projected beside the border of the smartphone. (b) The smartphone holder was adjusted horizontally until the axis of
the smartphone was parallel to the laser line.
The sample size was calculated to detect a significant outcomes were radiographic inclination, radiographic ante-
difference in percentage of cup placement in Lewinnek’s version, and percentage of cup placement in the safe zone.
safe zone. We hypothesized that the smartphone technique Pubic symphysis-centered pelvic anteroposterior radio-
can achieve 81% in the safe zone, the same as with com- graphs were taken in the supine position at 6 weeks post-
puter navigation,6 whereas 52% of our previous THAs were operatively. The inclination angle was measured between
positioned in the safe zone. With a two-sided type-I error the inter-teardrop line and the long axis of the projected
level of 0.05 and an 80% statistical power of detection, the ellipse. The cup anteversion angle was measured and cal-
sample size was 41 hips in each group. culated using the Widmer method (Figure 7).10
Preoperative demographic data included patient age, All radiographic measurements were performed by two
gender, body mass index, diagnosis, and cup type. Primary orthopedic residents who were not involved with the
Pongkunakorn et al. 5
Age
Mean (SD) 51.1 (8.6) 54.5 (7.9) 0.069
Range 36–68 32–67
Gender
M:F 20:21 22:19 0.825
BMI
Mean (SD) 24.1 (4.2) 23.5 (3.8) 0.836
Range 17.4–33.3 17.5–31.1
Diagnosis, n (%)
ONFH 21 (51.2) 19 (46.3) 0.669
DDH 5 (12.2) 6 (14.6)
Inflammatory joint disease 4 (9.8) 5 (12.2)
Femoral neck fracture 1 (2.4) 4 (9.8)
Others 10 (24.4) 7 (17.1)
BMI: body mass index; SD: standard deviation; ONFH: osteonecrosis of
femoral head; DDH: developmental dysplasia of hip.
Figure 6. To achieve 40 radiographic inclination, the positioner
was moved vertically together with the other smartphone until Results
36 of operative inclination was shown in the Clinometer
application. There were 41 patients in each group enrolled in the study.
The patients’ baseline characteristics were not significantly
different between the two groups (Table 1). Plasmafit cups
surgery and repeated again 2 weeks later. The average (Aesculap, Germany) were used in 50 cases (61%) and
of four measurements was used for data analysis. The Trilogy cups (Zimmer, Warsaw, Indiana, USA) were used
intra-class correlation coefficients (ICCs) were calcu- in 32 cases (39%). The mean inclination angle in the smart-
lated for intra-observer and interobserver reliability. phone group was significantly lower than in the conven-
We used the two-way random-effects model and abso- tional group (40.9 (SD 3.8) vs. 46.3 (SD 6.7), p < 0.001).
lute agreement for ICC calculation. The exact probabil- The mean anteversion angle in the smartphone group was
ity test was used to compare categorical data, whereas significantly higher (19.6 (SD 4.4) vs. 16.5 (SD 6.1), p ¼
t-test was used to analyze continuous data, with a 0.010; Table 2). The smartphone group had more cup
p-value of 0.05 being regarded as significant. This placements in the Lewinnek safe zone than the conven-
study was approved by our institutional review board tional group (90.2% vs. 56.1%, p ¼ 0.001; Figure 8). The
and registered in the Thai Clinical Trials Registry. ICC for intra- and interobserver reliability of inclination
Informed consent was obtained from all patients who measurements was 0.97 (95% CI 0.95–0.98) and
participated in the study 0.93 (95% CI 0.91–0.95), respectively. The ICC for intra-
6 Journal of Orthopaedic Surgery 27(1)
Table 2. Comparison of radiographic measurements and Table 3. Comparison of secondary outcomes between the two
percentage in the Lewinnek safe zone between the two groups. groups.
positional relationship between the pubic symphysis and inclination and 19.6 anteversion in postoperative radio-
coccyx on the fluoroscopic image is similar to the preo- graphs of this study.
perative radiograph. This plane can be marked by adjusting The percentage of acetabular cups positioned in the
the lateral border of the smartphone to be parallel to the Lewinnek safe zone was significantly higher in the smart-
laser line, projecting perpendicularly from the C-arm as in phone group. The scatterplots showed less scattered dis-
our technique, or by marking the C-arm orientation on the tribution of cup orientation than in the conventional
floor with tape so that it can be seen clearly during the group. These improved inclination and anteversion angles
surgery as described by Nishikubo et al.16 They used this came from the use of both smartphones because the pos-
technique to correct the errors of pelvic tilt through repo- itive effect of fluoroscopy for pelvic standardization was
sitioning of the operating table using fluoroscopy before eliminated by the return of operating table to its initial
surgery and could place the acetabular component within position before starting the surgery. It could be assumed
an adequate zone (inclination 30 –50 and anteversion that the pelvic position at the time of skin incision was not
10 –30 ) in 84.3% of hips. different between groups. Two hips in the conventional
The second cause of cup malposition can be solved by group had posterior dislocation. One of them had antever-
our pelvic inclinometer. After the patient position is stan- sion of 1 and the other had inclination of 54 . We found
dardized in lateral decubitus, this position is memorized in no significant difference in dislocation and infection rates
the smartphone application by calibration to 0 . Intraopera- between the two groups, and this might be from the inad-
tive pelvic motion can be assessed by the level application, equate sample size. The calculated sample sizes with 80%
and the pelvis can be repositioned to 0 of tilting before cup power to compare the dislocation and infection outcomes
placement, by turning the operating table. This is the same at the same rates require 194 and 410 hips in each group,
principle as the gravity-assisted guidance system studied by respectively. Nevertheless, this study has a power of
Echeverri et al.17 and the pelvic tilt goniometer used by 95.1% (type I error 5%) to detect the outcome of safe
Asayama et al.18 Echeverri et al. used a bull’s-eye bubble zone outliers.
level fixed to a Shanz pin and placed in the iliac crest. It There are some limitations in this study. Firstly, it was
acts as a reference, identifying the initial yaw and roll not a concurrent study and so might be biased by the
positions throughout the operation. 17 However, the period effect especially from surgeon experience. How-
mechanical bubble level does not show a degree of tilting ever, those THAs in the conventional group were per-
and could not be calibrated to 0 at any position of the formed by the same surgeon who had passed the
device. Asayama et al. used a specially devised goniometer learning curve of such operations more than 10 years pre-
to measure pelvic motion in real time during surgery. The viously. Secondly, our technique had some complexity
device incorporated a digital compass with two goni- that needed a learning curve for fluoroscopic control and
ometers as well as a pendulum and target apparatus. It was instrument installation. The outcome might be better if the
fixed to a threaded Steinmann pin that inserted into the surgeon is familiar with them. Thirdly, we measured
ASIS.18 They suggested that the surgeon was better able radiographic anteversion angles using the Widmer
to accurately achieve the desired cup placement with the method,10 which might have variation in each measure-
real-time visual feedback about the pelvic orientation and ment. Currently, reformatted CT scan is considered to be
motion. Nevertheless, their pelvic tilt goniometer is more the gold standard for cup anteversion measurement,20 but
complicated, larger, and heavier than our invented inclin- it needs special software that is not available in our hos-
ometer. Fixation to the pelvis with only one threaded pin pital. However, measurements using the Widmer method
might be weaker than with two screws. Our device can be are the most similar to those using reformatted CT, with a
disconnected and does not interfere with the operative field mean difference of 0.9 .20 The ICC for reliability of
during surgical exposure. anteversion measurements in this study was also at a good
The intraoperative errors in the manual operation can be level. To the best of our knowledge, this is the first clinical
solved by our technique that uses a smartphone combined study that uses a smartphone as a pelvic tilt goniometer.
with an alignment guide. The smartphone is placed parallel Moreover, our method is a novel technique for cup orien-
to the cup positioner, and inclination measurement can be tation in the Lewinnek safe zone and capable of solving
done following the previously described method.8,9 Peters three major causes of cup malposition.
et al. advocated the operative inclination at 40 from their
surgical experience,8 while Kurosaka et al. recommended
38 following the Murray’s equation and cadaveric study Conclusion
with the Orthopilot navigation system.9,19 Both studies Using the computerized function of smartphone could
aimed the 15 target of radiographic anteversion. Our tech- improve the precision of cup positioning in THA. It can
nique aimed 36 operative inclination and 25 operative be used with different cup positioners of different compa-
anteversion following the Murray’s equation to achieve nies in the lateral decubitus position. Most acetabular cups
the targets of 40 radiographic inclination and 20 radio- were placed within a narrow margin inside the Lewinnek
graphic anteversion.9 It provided an average of 40.9 safe zone.
8 Journal of Orthopaedic Surgery 27(1)