Professional Documents
Culture Documents
The Lipmann Equation and The Characterization of Black Lipid Films.
The Lipmann Equation and The Characterization of Black Lipid Films.
4
Edited by E. D. Korn
Book available from: Plenum Publishing Corporation
227 West 17th Street, New York, New York 10011
R. FETTIPLACE, L. G. M. GORDON,
S. B. HLADKY, J. REQUENA, *
H. P. ZINGSHEIM,t and D. A. HAYDON
Physiological Laboratory
University of Cambridge
Cambridge, United Kingdom
The essential features of the horizontal cell are shown in Fig. 3. The
central vessel is again machined from PTFE and the hole in which the film
is formed is 0.05-0.15 cm in diameter. The depth of the hole should be
comparable to its diameter. Lipid solution is supplied to the hole through
the horizontal tube (about 0.03 cm bore) which is connected to either a
manometer or a micrometer syringe by a length of PTFE tubing. Poly-
ethylene tubing must not be used as it C~Ofb-par~f the hydrocarbon
solution, thus making control of the volume of the olution difficult.
After initial filling of the hole, lipid solution is withdra n until drainage
occurs. A great advantage of this technique, especially hen a syringe is
used, is that the area of the black film may readily be controlled and may
be changed quickly and reproducibly. The illumination and inspection of
the film are conveniently achieved by means of an incident-light metallurgical
microscope with a water-immersion objective. (The latter may be isolated
electrically from the rest of the microscope by means of a PTFE ring.)
Although awkwardly small working distances may be necessary, it is possible
Fig. 2. The vertical cell. a, The inner cell (hollow cylinder of PTFE); b, the outer cell
(borosilicate or perspex); c, collar to allow rotation of inner cell; d, curved supporting with this arrangement to view the black films under very high magnification.
rod, allowing change of tilt without affecting depth of film; e, electrodes; f, rotating
magnet; g, magnetic fleas. Lower plan view: h, light source; i, telescope.
light beam and the inspection telescope. For some purposes, it is necessary
to stir both interior and exterior aqueous phases, and this is accomplished
by means of the magnetic fleas shown in the figure. Electrodes in the two
compartments are also shown. These may be of blackened platinum for
AC measurements but otherwise should be silver-silver chloride or any
reversible electrode connected to the cell by salt bridges. It is nearly always PTFE
manometer
advantageous for the electrode area to be as large as possible to minimize I I
tube
't
-., -
lI~=~~-
-
--I I
t
ducing lipid into the hole from a pipette. In the latter technique, any excess
lipid is withdrawn until drainage occurs either spontaneously or under
the influence of passing air bubbles (Szabo et al., 1969).
The term on the right-hand side is often referred to as the disjoining pressure
Any thin one-component liquid film of finite tension is necessarily of the film and denoted II (see, e.g., Scheludko, 1967; de Feijter, 1973;
unstable with respect to the bulk liquid from which it is formed. The de Feijter and Vrij, 1972).
presence of other components may complicate the theoretical argument, It is easy to show that for a black lipid film P(he) is very small, and
but in general thin films collapse when slightly perturbed. The mechanism hence
of rupture or nonrupture following a thermal or other fluctuation has dal!
-- I ~O
been discussed by several authors in the soap film literature (de Vries, dh I!=I!.
1958a,b; Kitchener and Cooper, 1959; Vrij, 1966), but unfortunately there
is little beyond simple-minded calculations of the activation energy for The variatIOn of ail with film thickness arises from the interaction
hole formation that can be applied to the black lipid film case. However, through the London-van der Waals and adsorption forces of the two
the stability of a thin film relative to a thick one and the factors governing interfaces as they come together (for a formal analysis see Andrews, 1970).
the equilibrium thickness are rather better understood. The London-van der Waals forces operate in such a way as to reduce al!
For unit area of film, the free energy change (AI! - Aoo) of the system below its value at h = hoo of 2yoo. This effect is obviously destabilizing and
in which the thinning occurs is (Mackor and van der Waals, 1952; Andrews corresponds to a decrease in the free energy (AI! - AOO). The film stability
et al., 1970; Requena, 1974) originates from the lipid adsorbed or spread at the two interfaces and the
l! tendency for it to be squeezed out of the film when the space available for
AI! - Aoo = f I!~
Pdh + al! - 2yoo + L l1i(/-l/'
i
- /-liOO)
the normal deployment or thermal motion of its long chains becomes
restricted. The partial desorption of the lipid leads to an increase in al!
where hand hoo are, respectively, the thicknesses of the final and initial and hence also in (AI! - Aoo). When the rates of change of the two com-
films, P is the excess hydrostatic and capillary pressure tending to thin ponents of al! with h are equal and opposite,comlItlOn is satisfied and
the film, al! is the tension of the (two-sided) thin film, yOO is the tension of the film is at its equilibrium thickness, he. The above events are ~ematically
the interface between the bulk phases with which the film is in equilibrium,
11; is the total number of moles of "i" in the system (per unit film area),
and /-lil! and /-liOO are, respectively, the chemical potentials in the system
before and after the film has drained.
The summation on the extreme right-hand side of equation (I) is
zero for an infinite reservoir of bulk phase, and for practical purposes may
be neglected (see, e.g., Mackor and van der Waals, 1952). The term
in P depends on the choice of the reference thickness, hOO, and is therefore Id(If-AOO)
of arbitrary magnitude. The term (al! - 2yoo) is determined by the London- (.rq/cm2)
van der Waals forces tending to thin the film, and by the repulsive or sta- I
bilizing forces which arise from the presence of the polar lipid molecules.
The condition for film stability is
Fig. 4. Free energy change vs. thickness for glyceryl monooleate plus decane films in
saturated sodium chloride solutions. The dashed curves represent the separate London-
van del' Waals and steric interaction contributions. The Hamaker constant has been
taken to be 3.48 X 10-14 erg. Results of Andrews et al. (1970).
j1fustrated in Fig. I. An actual plot of (Ah - ACO) against h for a lipid film The areas of the films are measured with a graticule fitted into the
is shown in Fig. 4. For details of the construction of this curve, the original eyepiece of the microscope. The graticule consists of vertical and horizontal
publication (Andrews et of., 1970) should be consulted. It is sufficient here scales which are calibrated with a stage micrometer placed under the aqueous
to note that the onset of the stabilizing forces is very sharp and that, con- phase in the experimental cell. It is possible to estimate film diameters
sequently, the position of the free energy minimum is determined by the to about 10 fLm, so the film diameter must be I mm to achieve 1% accuracy
thickness at which these forces begin to operate. The depth of the free in its estimation. In practice, the determination of the film area is the
energy minimum (a" - 2yCO)h~h. is thus due almost entirely to the London- limiting factor in obtaining an accurate specific capacitance. Despite this,
van der Waals forces. it is possible to obtain accurately reproducible data on any given system
No mention has been made of the electrical double layer interaction and more elaborate procedures for the area measurement are not really
which is so important in the stabilization of many aqueous soap films. necessary.
While such electrical forces may not be wholly absent in the lipid film, it For AC measurements, the choice of electrodes is not critical so long
is possible to deduce from double layer theory and film conductivity data as their impedance is sufficiently low and they do not polarize. Thus,
that their contribution to the stabilization of the film is ordinarily negligible ideally, electrodes consisting of large areas (e.g., 2 cm2) of silver foil spot-
compared to the steric interactions of the chains (Haydon, 1968; Andrews welded on to silver wire might be used, but platinum black electrodes of
et aI., 1970). a similar size are usually as good provided they introduce minimal potential
differences across the films.
The system as seen by the AC bridge network includes the connecting
leads and electrodes, the PTFE pot, the aqueous phases both inside and
outside the pot, as well as the film itself. Hanai et of. (1964) obtained ap-
proximate values for the capacitances associated with the fixed parts of the
apparatus, and showed that, over the working range, these components
were frequency independent. They further demonstrated that the equivalent
circuit for the system could be reduced to that shown in Fig. 5. This circuit
The frequency range of interest for most black film systems is from may be analyzed in a variety of ways. For bridges, sucb~ of Wayne
50 to 107 Hz, although if only static dielectric constants or capacitances Kerr, which record the equivalent parallel capacitance C' and c;nuJtance
are required, a small frequency range in the region of 500 Hz may suffice. G of the circuit, the following treatment, taken from Hanai el of. 964),
In either instance, sine wave signal generators, bridges, and null point is convenient.
detectors covering the range in question are necessary. A large variety of The complex capacitance C* of the circuit (defined by i = jw * V)
suitable signal generators are available. Bridges such as the Wayne Kerr
Universal and Radio Frequency instruments have been found satisfactory
(Hanai el of., 1964). For the audiofrequency range, an oscilloscope with a
sensitivity of 100 fLYfcm is likely to be adequate as a detector, although
the use of a preamplifier between the bridge and the oscilloscope may be
beneficial. In the radiofrequency range, a communications receiver is an
appropriate detector.
With this apparatus, capacitances and conductances of black films of
approximate area 5 X 10 3 cm2 may usually be determined with an accuracy Cs
of better than 1%. Greater accuracy in the capacitance may be achieved
Fig. 5. The effective equivalent circuit for black film studies. Gilland Cm are, respectively,
without much difficulty but this would usually be superfluous owing to the the conductance and capacitance of the black film; G'Q and C'Q are the corresponding
uncertainties in the film area. properties of the aqueous phase and C. is the stray capacitance of the system.
C' = C" + Cs I-- (Cl - C,,)/(I + ah· 2)
2
103 c' (~F)
1 being the frequency. C"q, G"q, Cm, and Gm are the capacitances and
conductances for the aqueous solution and film, respectively, and Cs is the Fig. 6. Plots of the imaginary against .the real part of the complex capacitances of black
stray capacitance (see Fig. 5). films in NaCI solutions for egg yolk lecithin-decane films. (a) 103M NaCI. (b) 10-1
If Gtlw is always small and c., is independent of frequency, equation M NaCI. The arrows indicate the relaxation frequencies; the dashed curve in (a) is
the semicircle having its center on the C' axis. Res lilts of Hanai et al. (196<tJ.\
(5) gives a semicircular plot in the complex plane with C" (or G/w) tending
to zero as 1->- 0 and 1->- 00. At these extremes of the semicircle, it may be
shown that C' reduces to em + c" (I 0) and Cs (/= 00). Most of Cs reduced, G"q decreases and so also does j~. For very low electrolyte con-
may usually be trimmed out on the bridge before the film thins, and thus centrations, it may be impossible to obtain an accurate value of Cm directly,
the limiting value of C' as 1->- 0 gives the film capacitance Cm' since, at the low frequency required, the apparatus tends to be insufficiently
Examples of semicircular plots obtained for black films are shown in sensitive. Extrapolation of the semicircular plot from higher frequencies,
Fig. 6, and the plots of C' and G individually as a function of frequency however, often yields an acceptable result. Complications may also arise
are given in Fig. 7. Such results are typical of poorly conducting films. if materials are present which make the films highly conductive, so that
Provided Gm ~ G"q and C"q ~ Cm, equation (II) reduces to G/w is not negligible at all frequencies. Since G/w decreases with increasing
frequency, however, extrapolation from higher frequencies may again
circumvent the difficulty.
To conclude this section, it should be emphasized that, although the
Moreover, since Cm is often nearly independent of the electrolyte concen-
capacitance of the system of black lipid film and aqueous solution varies
trations used, the characteristic frequency, 10' of the system is determined
with frequency, that of the black film itself, Cm, is not necessarily frequency
solely by the specific conductivity of the aqueous phase. This effect is seen
dependent. In fact, several groups of authors have shown that for a variety
in the results of Figs. 6 and 7. Clearly, as the electrolyte concentration is
offilms Cm remains constant from effectively zero frequency to about 107 Hz
the applied potential V. The integral f: J dt is obtained by the graphical
estimation of the area under the curve of the current vs. time.
Although easy and relatively cheap to set up, the DC transient method
tends to be less accurate than the AC method. Thus unless signal averaging
techniques are used, the integration procedure, or the evaluation of the
relaxation time of the current transient, introduces uncertainties which are
usually substantially greater than I %. The DC transient method is also
clearly unsuitable for the examination of the possible frequency dependence
of the black film capacitance.
em = qjV= f~ JdtjV
J1
V
Fig. 8. Circuit for the measurement of the membrane capacitance by the DC transient
method.
normally be disregarded. For well-defined films, the specific capacitance is _---ionic qroup
a reproducible quantity, both within one experiment and between experi- o
ments. The specific capacitance is thus unlikely to be subject to any large 1+
_---,dipolor qroup
lipid
port 1
random error such as would be introduced by a considerable and variable of
0-qeqen-ion
number of lenses in a film. It is concluded that the area reduction produced membrane I
by the presence of lenses must fall within the experimental error. This
aqueous solution
o 1+
Fig. 10. Bilayer capacitance in solutions of uni-univalent electrolyte. The interfacial areas
per ion at the bilayer interfaces are, in A", (a) 208, (b) 2080, (c) 3360, (d) 6930, (e) 10,400,
(f) 13,900, (g) 20,800, (h) 00. The dashed line represents the geometrical capacitance
C /4nd to which the curves are asymptotic. Cm, d, and the membrane potential have been
t:ken as 2.07, 47.5 A, and 10 mY, respectively. The temperature was taken to be 20°C.
-. { 4nCm [( 4nCm )
Hydrocarbon
Capacitance (em)
thickness
n-Decanea 1.993 (fLF/cm2)
(A)
n-Dodecane a 2.020
n- Tetradecanea 2.037
Monoolein (C18:l)-decane 0.386 48.1
n-Hexadecanea 2.052
2.136 Monoolein (C18:l)-hexadecane 0.584 32.8
1-0ctadecene
1-0ctadecene equilibrated with water 2.137 Monopalmitolein (C16:l)-hexadecane 0.692 27.8
cis-9-Octadecene 2.139 Monomyristolein (C14:t)-hexadecane 0.795 24.2
Decane/l-octadecene, 1:1 v/v 2.056 Monolinolenin (C1s:3)-hexadecane 0.803 25.9
cis-9,12-0ctadecadiene 2.219 Dioleyl lecithin-decane 0.390 48.3
cis-9,1 2-0ctadecadiene equilibrated with water 2.204
Egg yolk lecithin-decane 0.387 48.6
cis-9, 12,15-0ctadecatriene 2.272
Egg yolk lecithin-hexadecane 0.603 32.2
I-Decene 2.107
I-Heptene 2.057
cis-3-Hexeneb 2.062
cis-5-Decene b 2.071 films is given in Table T. In Table II are shown estimates of the dielectric
constants and molecular volumes of the hydrocarbon residues of some
a Data from Landholt and Bornstein (1959) (T ~ 20°C). lipids.
b Data from Landholt and Bornstein (1959) (T ~ 25°C).
A selection of hydrocarbon thicknesses for some of the better-charac-
terized black films is shown in Table III.
3.2.4.1f'nvironment of Lipid Chains and the "Bulk Hydrocarbon" Assumption dencies would result in a higher capacitance than expected from the actual
thickness. Some evidence exists from partial molar volumes and spin lattice
Throughout the analysis of the capacitance data it has been assumed relaxation times of methylene protons that there is a degree of hydration
that the nonpolar core of the black film has the properties of bulk liquid of the methylene groups in the vicinity of the head group in long-chain
hydrocarbon. A paper by Montal and Mueller (1972) on the formation and surfactant micelles. This evidence has been used to support the contention
properties of solventless lipid bilayers reports specific capacitances of of water penetration, but it probably only reflects the exposure of these
2
0.9-1.0 [LF/cm • Owing to the nature of the lipids used in this work, the methylene groups to water at the micelle surface, since the area occupied
results are not directly comparable with those for the "solventless" lecithin- by each surfactant molecule in the micelle is larger than the cross-sectional
hexadecane membranes of Table III. It is nevertheless striking that the area of the headgroup.
capacitances reported for the two types of system are substantially different, The finding that specific capacitances of black films in saturated salt
and, in view of Montal and Mueller's speculation that water may penetrate solutions (where the water content of the film must be substantially lower)
appreciably into the hydrocarbon region of their solventless membranes, are, if anything, higher than those in 0.1 M salt would suggest that water
the matter merits some discussion. Repetition of the Montal and Mueller penetration is not a serious problem. The linear variation of inferred
experiments using the egg yolk lecithin of the systems in Table]" yielded thickness with chain length for the isosorbide esters and monoglycerides
specific capacitances in the range 0.66-0.81 [LF/cm2, the mean being 0.76 (Taylor and Haydon, 1966; Fettiplace, 1974) also seems inconsistent with
[LF/cm2 (Fettiplace, 1974). A basic difficulty of the technique is that owing water penetration. Although the situation might be somewhat worse for a
to the extremely low tensions of solventless membranes (the spontaneous pure lipid bilayer, particularly if the area per lipid molecule were larger,
expansion of even a lecithin-hexadecane membrane, Fettiplace et aI., the similarity of the value obtained for the water permeability of lecithin
] 971, is a manifestation of this) the slightest hydrostatic or capillary pressure black films to the value for pure bilayers (Fettiplace, 1974) indicates that
difference renders the membrane nonplanar and thus of uncertain area. the two systems are not qualitatively different.
Moreover, optical inspection has not so far proved of much value in checking
the question. It is therefore arguable that only in the ideal limiting case
is the membrane area equal to the hole area and that only the lower values 4. DETERMINATION AND CONTROL OF BLACK FILM
observed for the capacitances are significant. If this is so, the average COMPOSITION
specific capacitances given by Fettiplace and by Montal and Mueller are
likely to be overestimates. With this consideration in mind, it seems from 4.1. Introduction
Fettiplace's data that although there could still be a difference between A black film is stable as a consequence of the very strong adsorption
the solventless and hexadecane results, this is not necessarily greater than or spreading of the polar lipid or lipids at its interfaces. This adsorption
about 0.1 [LF/cm2• This discrepancy could be real, however, and it is still renders the composition of the black film quite different from the solution
of interest to consider the question of whether water penetrates sufficiently from which it is formed (it is relatively much richer in the lipid). Moreover,
into the hydrocarbon of either bilayers or black films to affect the assump- since the film is usually in equilibrium (or nearly so) with both this lipid
tion that fill may be taken as for bulk liquid hydrocarbon. solution in the Plateau border and the surrounding aqueous phase, any
The cross-sectional area of the lipid head group and chains in the pert~lrbation of the system is likely to change the composition of the film
crystalline state of lecithin is about 40 A2 as compared to an area per very rapidly. Techniques of analysis which involve mechanical sampling
molecule in the bilayer of 60-72 A2. This increase in area corresponds of the film, especially if they first entail fixing and embedding procedures,
to the exposure of five to seven methylene groups per chain at the water are therefore inherently suspect and to be avoided wherever possible.
interface. It has been suggested that water penetration in this region could Unfortunately, the alternatives are limited, particularly for complex
seriously affect the interpretation of specific capacitances, either by in- systems. Thus, radioautogra h has not been successfully used for a black
creasing the dielectric constant of the hydrocarbon interior adjacent to the film in situ. Absorption spectros y is insufficiently sensitive for most
interface or by decreasing the effective hydrocarbon thickness. Both ten- purposes, although it has been used detect molecules with unusually
high extinction coefficients, such as chlorophyll (Steinemann et at., 1971).
Reflection spectroscopy shows promise but has so far not been greatly
exploited (Cherry et aI., 1972). Fluorescence spectroscopy can have adequate
sensitivity but may require awkward calibration procedures and, like
absorption and reflection spectroscopy, is limited to certain rather unusual
types of molecule (see Section 8). For simple systems such as black films
consisting of one lipid and one solvent, however, surface chemical methods
may be successfully employed to estimate the lipid, and by using the hydro-
carbon thickness data derived from capacitance measurements, the solvent
also may be estimated. Although the application of this approach to more
complex systems is prohibitively tedious, it can be quite rigorous, at least
for the lipid, and hence may be used as a control for other techniques which
are more convenient but less well-founded.
A (A2 per
Solvent C (fLF/cm2) em d(A) ip
molecule)
bLTFEfilm
support
of system. The solvent content found by Pagano et al. is considerably
larger than that deduced by Fettiplace et al. This result almost certainly
aqueous
reflects the presence of the small lenses which are nearly always present in
phase
glyceryl monooleate-hexadecane black films and, indeed, in any other
black lipid film formed from highly water-insoluble solvents (see Sections
Fig. 13. Schematic representation of the apparatus used by Pagano et al. (1972) to sample
2 and 8). The present sampling technique is thus not necessarily suitable
black lipid films. Mercury droplets delivered from the micrometer syringe fall through for solvent estimation in the true black film if used directly. However, as
the film into the chloroform. The films may be observed under reflected light by means the results for the lipid are not in question, it could still be used indirectly,
of the lamp and telemicroscope. Pagano et al. thermostatted the aqueous solution. as in the surface thermodynamic-thickness approach.
Apart from the disadvantages just discussed, the sampling technique
film fragments coating the mercury droplets. Instead, the lipid solution is of Pagano et al. could, if used with discrimination, circumvent the almost
applied to the PTFE ring above the aqueous phase, and the ring is then insuperable complexity of applying the surface thermodynamic method to
placed in position by passing it obliquely through the air-water interface, multicomponent systems.
when the film blackens spontaneously. This procedure necessarily deposits
some of the radioactive film-forming solution on the surface of the aqueous
solution and this has to be rigorously removed by suction before the cup 5. CONTACT ANGLE AND FREE ENERGY OF FORMATION OF
of chloroform can be removed. After each droplet of mercury has fallen, THE BLACK LIPID FILM
the area of the black film contracts by an amount equal to the area of the
droplet, and time is allowed for the reblackening of the new border regions
before the next droplet is released. Sufficient droplets are passed through
Equation (I) of Section 2 gives the free energy change for a system in
the film to yield a convenient count in the chloroform. The surface area
which a thin film is formed from a layer of thick liquid, and the significance
of the mercury droplets is found from the drop weight and density of
of the various terms is briefly discussed there. For present purposes, the
mercury.
term of interest is that involving the tensions of the film (a) and bulk liquid
There are several hazards or points to be borne in mind in the above
(yOO)" as this represents the interaction between the two surfaces of the
technique, the more obvious of which are as follows: First, the passage of
thin film. Thus the Helmholtz free energy change I1A(he) associated with
the mercury droplet through the black film must stretch the interfaces to
this interaction for a film of unit area at its equilibrium thickness he is
some extent and so perturb the equilibrium mono layers. Second, the pres-
ence of the mercury surface may have some influence on the composition
of the coating film. Third, any lenses which happen to be trapped in the
film are likely to be included in the samples obtained. Fourth, chloroform A knowledge of L1A(he) can be valuable in connection with a wide
is relatively soluble in aqueous solutions and, given time, will partition range of problems. For a simple lipid film, it has been shown that L1A(he)
strongly into the black film and its Plateau border. I f this occurs, it could is dependent almost entirely on the London-van der Waals forces which
34 ) R. Fettiplace et al.
act ~he film and which are mainly responsible for the final stages of quantity and that the invocation of a hypothetical contact line requires
the drainage. The measurement of L1A(he) has, in these instances, yielded that a line tension term should appear in the force equation (de Feijter
accurate and detailed information concerning the London-van der Waals and Vrij, 1972), although for the lipid films under discussion, the line tension
forces in lipid-water systems (Haydon and Taylor, 1968; Andrews et al., is negligible.
1970; Requena, 1974). The incorporation into lipid films of, for example, The determination of fJ for aqueous soap films has been extensively
ions, polypeptides, or proteins may introduce interactions over and above discussed during the past few years, and several methods are available
those arising from London-van der Waals forces. Such effects have been (Mysels et aI., 1966; Princen, 1968; Kolarov et al., 1968; Prins, 1969;
examined for the lipid-soluble ion tetraphenylboride and for the polypeptide Clint et al., 1969; Princen and Frankel, 1971). In principle, these methods
gramicidin A (Requena, unpublished), and have provided useful indications may also be applied to lipid films, but in fact only an interference technique
as to the mechanism of action of these substances. In short, anything which comparable to that of Kolarov et al. (1968) has so far been described
modifies the attractive and repulsive forces across a lipid film will be re- (Haydon and Taylor, 1968; Kruglyakov et al., 1972; Requena, 1974).
flected in L1A(he). In this technique, interference fringes, arising from the illumination of the
The measurement of L1A(he) is most profitably achieved by determina- regions adjacent to the edge of the black film, are used to ascertain the
tion of the contact angle. This angle, fJ, is defined by the requirements of profile of the oil-water interface, and hence the angle of this interface with
the force balance in the equilibrium system, I.e., the film. Since even the first fringe occurs at a distance farther from the
edge of the film (about I [Lm) than the width of the transition region referred
to above, the angle obtained is the extrapolated value required for equa-
tion (26).
"Fig. 17. Schematic diagram of the optical path of the illuminating beam in the film-lens
system. The values for the geometrical parameters of a typical lens (see Fig. 16) are
drawn (not to scale) in order to give an idea of their magnitude. a and S are rays incident
Fig. 16. A lens of bulk lipid trapped in a thin film made from a solution of glyceryl mono- on the black film and lens, respectively, while b, c, T, and V are the corresponding reflected
oleate in /I-decane in aqueous 0.1 M NaCl at 20o~. rays. Other parameters are described in the text.
is so large and the contact angle so small that the incident light is effectively
normal to the lens surface. Referring to Fig. 18,
A plot of p against ,2
for the lens of Fig. 16 is shown in Fig. 19. The as-
sumption of spherical symmetry appears fully justified. Extrapolation of
H-)
]f the radius of the fringe is xp, it is possible to fit the height-radius data
to a parabola of the form
Th,e first derivative of equation (37) at the point Xo (the radius of the film)
gives the tangent of the contact angle. The radius of the film may be cal-
culated from equation (37) with Yp = O. (The black lipid films are suf-
ficiently thin for their thickness to be ignored in all the present equations.)
In practice, the center of the film is located and the diameter of several
dark rings measured. The heights of the points p as given by equation (36)
and their distance from the center of the film are then fitted to equation
(37) by a least-squares method.
]n conclusion, the contact angle obtained from the fringes of a lens
should not be identical to that from the Plateau border, because the curva-
tures' of the lens and Plateau border interfaces are different (in fact, of
opposite sign). Thus the hydrostatic pressures experienced by the film at
the boundaries of the two regions are also different. The film is not of
precisely the same thickness at the two points, and this is reflected in the
magnitude of the contact angle. However, for black lipid films, the effect
is theoretically very small and indeed has not been detected experimentally
since the two angles are equal to within the errors of the measurement.
The question is nevertheless important in principle since the lens can never
be in equilibr,ium with the Plateau border (Section 3), and a flux of lipid
Fig. 20. Photograph of the Plateau-Gibbs region. The film was formed from a solution
of glyceryl monooleate in n-decane in aqueous 0.1 M Nac;l at 20°C. and solvent must exist from the first to the second regions.
44 )
~stimation of Bulk Interfacial Tension, yOO too high, and from a knowledge of C (Section 3), the slope of the line
yields yoo.
Although the measurement of yOO is, at first sight, a routine problem
This approach has been verified for systems of known tension and has
to which there are a number of possible approaches, the use of materials
been successfully applied to phospholipid black films (Requena, 1974;
such as phospholipids presents difficulties. As pointed out in Section 4,
Requena and Haydon, to be published).
phospholipids do not usually form molecular solutions in the solvents used
for black film formation, and thus monolayers which arise from placing
dispersions of the lipid in contact with aqueous solutions tend to be ir- 6. OSMOTIC AND ISOTOPIC FLUXES: MEASUREMENT OF
reproducible. For this reason, unless the tension of the phospholipid WATER PERMEABILITY
dispersion which is actually in equilibrium with a black film is measured
in situ, estimations of dA(he) may be seriously in error (there is already
one completely erroneous value in the literature). While the direct in situ
Measurements of the water permeability of black lipid films have been
measurement of yOO is not an attractive proposition, there exists a very
described by a number of authors whose objectives have ranged from the
useful indirect method. This involves the application of the Lippmann
determination of the effects of lipid composition (Huang and Thompson,
equation to the black film system (Requena, 1974; Requena and Haydon,
1966; Finkelstein and Cass, 1967; Fettiplace, 1974) to the assessment of
to be published).
solute-solvent coupling in membranes containing the polyene antibiotic
An aqueous solution bounded by a black lipid film, in absence of
amphotericin B (Andreoli et al., 1971). Two techniques, osmotic flux
lipid-soluble ions or ionophores, is effectively an ideally polarizable elec-
measurement and isotope diffusion measurement, have been used. (The
trode. It is easy to show that the Lippmann equation in its usual form
hydrostatic pressures that black films will withstand produce negligible
(Aveyard and Haydon, 1973)
fluxes.) Apart from this clear subdivision, the basic features of the ex-
perimental methods described by the various authors are very similar.
It will therefore be convenient to devote the two sections below to the
is applicable to this system. (J is the film tension, E is the potential applied osmotic and isotopic approaches, respectively. Neither approach is entirely
across the film, and q is the polarization charge which arises from the ap- straightforward, both being complicated to some extent by the resistance
plication of the potential. If equation (38) is divided through by E and of the unstirred boundary layers of aqueous solution adjacent to the black
integrated, it becomes film. The isotope measurements in particular are subject to this difficulty
since the resistance of the unstirred layers in this approach may b~ larger
than that of the film itself. Much of the following discussion will be con-
cerned with the problem of the unstirred layers and of obtaining true mem-
where C", (= q/E) is the integral specific capacitance of the black film.
brane permeabilities from the water flux data. Although the movement of
For most black films, C", is effectively constant for small variations in E.
water specifically is discussed, the isotope flux technique is applicable
Moreover, if
generally to any labeled substance (see, e.g., Vreeman, 1966).
light source
\ stability. This problem may often be circumvented by minimizing the
-..,...- electrolyte concentration ratio, thus keeping the membrane potential down
while maintaining the required osmolarity difference. Even osmolarity
differences arising from impermeant nonelectrolytes such as urea or sucrose
may affect the film stability, either through the presence of trace surface-
/
qloss holder
\ ..
moqnetlc stirrer active impurities or through the significant differential effects that they
may have on the interfacial tensions or potentials of the two sides of the
Fig. 22. An apparatus for the study of osmotic flow across black lipid films (see text for
film.
description).
earlier, i1p f':::; 0 and, if the lipid membranes are known to be very im-
permeable to the solute, Js ~ Jv and a f':::; I. Thus equation (41) reduces to
where, e.g., for NaCl, 'J! = 2, and C1 and C2 are the concentrations of the
solutions on the two sides of the membrane; gl and g2 are the respective
rational osmotic coefficients and may, at least for NaCl, be obtained by
interpolation of the data in Robinson and Stokes (1959). The water per-
meability coefficient, Pf' is expressed in the form
lipid film. Provided this is done, and the stirring imposes reproducible : Pm I
the predominant influence on the boundary layer thickness. As calibrating Fig. 24. Circuits used in the determination of membrane conductance. (a) The simplest.
membranes, Everitt et al. (1969) found that the thinner varieties of cello- (b) The virtual ground circuit. (c) The equivalent circuit for the membrane in series
phane were suitable. with the aqueous solutions and electrodes.
of this circuit IS attractive, there is little else in its favor. In particular, after a change in either the applied potential Va or the membrane resistance
whenever the current is varying, the potential across the membrane is also Rm (Fig. 24c) to some new constant value, the current will relax to a new
changing unless Rk and Rs are each much less than Rm. Thus for an applied value according to standard equations of circuit theory,
potential of 100 m V, the largest measured potentials must be less than
1 m V (l % accuracy). If the conductance takes on a large range of values,
the smallest potentials which need to be measured may be much less than
10 [LV. Signals of such small size are difficult to handle and to measure.
A second disadvantage of this simple circuit is that if Rm and Rk are both
large (as in the single channel studies with gramicidin, where Rm ;;:; lOll Q
and Rk ;;:; 109 Q (Hladky and Haydon, 1972), then even for membranes where the dot indicates the time derivative.
with very small area (Cm 100 pF), the time constant for the response of
C-...J
While the gain is not actually a constant after a sudden change in
Vout to a change in Va or Rm is i = 1/RkCm > 10-1 s, which is too long. current, since the amplifier requires time to respond, so long as the gain
An alternative arrangement is available for determining the current, is large compared to I its actual value is not important. Thus in solving
which makes use of a high-gain amplifier with negative current feedback equations (54) and (55), it is permissible to use Vout = -A VI' The time
(Fig. 24b). As this circuit is the basis of all the measurements to be dis- constant for the single-exponential relaxation to the new steady value is
cussed, it is worth examining in some detail. In the steady state,
i = [Cm + (A + I)CkHI/Rm + (A + l)/Rk]-I (56)
provided that the gain has reached at least 100 within that time. For larger
currents (discussed in Section 7.3), the speed of response is limited by the
bandwidth and stability of the amplifier.
Thus the output voltage of the amplifier is proportional to the current while
the input voltage
is very small. It is for this reason that this arrangement is called a virtual The limitations on the detection of small conductance changes are the
ground circuit. For A = 105, the output may be as large as 10 V while various sources of noise, which may be conveniently grouped under four
the input voltage is still negligibly small. Thus so long as Rs is small (or headings: amplifier noise, physical vibration, electrical and magnetic
compensated; see below), this circuit acts as a voltage clamp for the mem- interference, and stray conductances in the membranes. These will be
brane. More complicated circuits which use separate pairs of electrodes discussed in turn.
to measure the potential and apply the current are an improvement only If the amplifier introduces noise equivalent to an added input of
insofar as they may be needed to avoid electrode polarization. amplitude vIn, then at a given angular frequency w
The time response to changes in the current across a membrane is
considerably improved by the use of a virtual ground ampli~er. Immediately
A(w) ,...., I
A(w) +I
_ n{ llRk + I/Rm + jw(Cm + C k) }
where j =V -J.
For low frequencies, the noise seen at the output is thus the same as
the noise voltage referred to the input, since Rk will be chosen somewhat
less than Rm. But at higher frequencies and assuming C", ~ C•.•
-13
2·10 amp
1=0"
Fig. 25. Fluctuations in the current across a glyceryl monooleate plus n-hexadecane mem-
Since Cm cannot be made much less than 100 pF and is commonly larger, brane containing a small (unknown) amount of gramicidin A. The applied potential was
and the minimum value of Rk is determined by the size of the signal to be 100 mY, the aqueous phases both contained M-IOO KCl, the membrane area was about
detected, the reduction of output noise to a tolerable level (about 1 m V 3 x 10 -4 em", and the temperature was 23°C. Similar experimental records at higher cur-
rent levels are, of course, less contaminated by noise (ef Hladky and Haydon, 1972).
apparent width on an oscilloscope) will require Ck '> I pF for Rk = 109
or 1010 Q. It should be emphasized that it is important to use membranes
of small area. quences of fluctuations at 10-11 Q-I with no contamination of the records
Mechanical vibration will cause small changes in membrane area and due to vibrations (see Fig. 25).
waves on the surface of the high-impedance aqueous phase connected to The three principal forms of electrical interference are all well known:
the virtual ground. Thus the membrane capacity and the capacity between movement of objects carrying static charge (e.g., variable capacitors such
the aqueous phase and ground will both vary. The magnitude of the resulting as the experimenter), electrical pickup of the supply frequency, and pickup
currents depends on the rates of the capacitance changes and, respectively, of magnetic fields in any closed conducting loops (ground loops). The
the applied potential or the surface (compensation) potential of the aqueous first two are simply eliminated by mounting the cell and the high-impedance
phase. These microphonic currents can completely mask those changes in lead to the current amplifier inside a grounded Faraday cage. Loops between
current which are due to conductance fluctuations. In order to minimize a signal lead and the accompanying ground are avoided by the use of coaxial
vibration it is necessary to mount the apparatus in such a way as to isolate cable. Closed loops of the ground leads can usually be avoided, or, if not,
it from the normal solid supports in the laboratory. A simple technique the area of the loop and thus the pickup can be reduced by twisting the
which is often adequate is to secure the apparatus to a heavy slab which is cables or sensibly arranging the various measuring devices.
then placed on an absorbent layer (foam or an inner tube) which, in turn, Of the various sources of noise, stray conductances in the membrane
is placed on the most massive support available. Using mountings of this are the most intractable in that their removal requires the use of highly
type, sequences of conductance fluctuations of the order of 10 10Q-I have purified materials. Nevertheless, with the use of glyceryl monooleate,
been recorded free from visible interference (Gordon and Haydon, 1972; dioleyl lecithin, egg phosphatidylserine, or bacterial phosphatidylethanol-
Eisenberg et al., 1973). An even more complete isolation has been ac- amine, membranes can be made which normally have specific conduc-
complished by suspending the slab from securely mounted girders using tances in MI I0 NaCI at pH 5.7 of less than 10-9 Q 1 cm-2 and for which
nylon rope. With this system it has been possible to record sustained se- the observed cond uctance is constant.
The choice of an amplifier to measure small conductance changes is
based on several criteria. In order to insure that all of the current goes
through the feedback loop and to minimize current noise, the input im-
pedance should be high (> 1012 Q) and the input bias current should be
small ( < I pA). Drift of the DC zero is only a minor problem, though it is
important to balance the amplifier so that zero input corresponds to zero
output since any offset potential will result in the same potential offset of
the amplifier input lead. As should be apparent from the discussion above,
a large bandwidth and low noise are both important. The Analog Devices
42 has been used extensively in the authors' laboratory for the smallest
currents, while the faster Philbrick 1011 and Analog 48K are useful for
higher currents. The amplifier is conveniently mounted in a metal box
with the inverting input connected both to the input lead and to the wiper
of a teflon rotary switch. The various contacts on the switch are then con-
nected to the amplifier output by a selection of values for Rk and CI.; in
parallel. For the high-resistance settings, Ck is chosen as a compromise
between noise and time response, while for low resistances Ck is chosen for
Fig. 26. The potassium current carried by trinactin as a function of time after steps in the
amplifier stability when controlling a capacitative input ("'-'2000 pF). Stable
membrane potential from 0 to 25, 50, 100, and 150 mY. The glyceryl monooleate plus
high-value resistances are available from Victoreen, while low-leakage n-hexadecane membranes were formed from a solution containing 105M trinactin; the
capacitors are available from many sources. aqueous phases contained 0.5 M KCI and, to reduce the series solution resistance, 0.5
M LiC!. The temperature was 23°C; the area of the membrane was about 10-2cm2.
The four traces were recorded in rapid succession using an Analog Devices 48K amplifier
with a 103 n feedback resistor. One large division equals 5 x 10 7 A vertically and 10 f.LS
horizontally.
As the conductances to be measured become progressively larger, the
various sources of noise other than the amplifier become less important
and the useful time response of the circuitry improves, primarily as a result sible, it is best to reduce Rs directly by the addition of a nonpermeating
of smaller values for Rk. There are, however, still limitations on the possible electrolyte.
time response which are imposed by the bandwidth of the available am- The resistance of the electrodes, which varies with time after a change
plifiers (of the order of 10 MHz), the stability of these amplifiers when in current and thus cannot be compensated, must be kept small by using
controlling the potential on a large capacitor, and the series resistance electrodes of very large surface area relative to the area of the membrane.
through which the membrane capacity must be charged. Amplifiers such as Silver-:silver chloride sheets with areas of about 4 cm2 have been satisfactory
the Philbrick-Teledyne 1011 and the Analog Devices 48K can establish for use with 10-2 cm2 membranes. When the resistance of the electrodes
a new value of the membrane potential to within 10% in less than 10!J.s cannot be made negligible, it is necessary to use a four-electrode voltage
for Cm c:::= 1000 pF and Rm = 106 Q (Fig. 26). clamp system (Hodgkin et aI., 1952; Moore and Cole, 1963; Bezanilla
The time required to charge a 2000-pF capacitor through a series et aI., 1970).
resistance of 1000 Q is of the order of 2 !J.s. In addition, for R. > 1O-2Rm, Whenever Rs is constant, it merely reduces the potential across the
the potential across the cell (Va - VI c:::= Va) is no longer the same as the membrane from Va to (Va - iRs)' The principle of increasing the applied
potential across the membrane even in the steady state. It is thus im- potential by iRs to compensate was well established in early voltage clamp
portant either to insure that Rs is less than the smaller of 1000 Q and experiments on squid giant axons (Hodgkin et al., 1952). For a two-electrode
10-2 Rm, or else to provide some form of compensation .. Whenever pos- system, the output of the current amplifier is inverted and a portion (equal
to Rsl Rk) is added to the applied potential. If a virtual ground circuit is where L1V is the potential across the diode, i.e., VI - Vout, lod is a constant
used to add the feedback signal and Va' the inversion is accomplished in for a given diode at a given temperature, and Vout -= -A VI' it follows
that VI ~ 0, and
the same process.
This circuit suffers from the inaccuracy of equation (62) and large tem-
Whenever the conductance varies over a large range, it is convenient to perature drifts. A more practical circuit (ef SGS-Fairchild, 1967) makes
record log current rather than the current itself. A voltage output signal use of the more accurate relation
proportional to log current is easily obtained by making use of the current-
voltage relation of a forward-biased pn junction or transistor. In principle,
the arrangement merely requires that a diode rather than a resistor be
placed in the feedback loop of a virtual ground circuit (Fig. 27a). Since, for the collector current of an NPN transistor with the base-emitter junction
forward-biased by a potential VBE.
for a forward-biased diode,
In the circuit shown in Fig. 27b, the input current is also the collector
current of the first transistor; therefore,
Because of the gain of the second amplifier, the output to the voltmeter
or recorder is therefore
FJ~ D D AD
special attention was paid to those problems which previously had in a :--;iF2--' i L
number of cases prevented quantitative intensity measurements. They '~------D---'-'- -- ------ ---~--O-~®)-
confined themselves to attempting a rigorous test of the technique, which Fig. 30. Schematic representation of the apparatus used by Zingsheim and Haydon (1973).
consisted of determining the adsorption of ANS to small planar black lipid L, Light source; M, mirrors; D, iris diaphragms; F heat filter; F2, excitation filter
"
(360 nm); AD, annular diaphragm for conical illumination in reflected light; M', half-
membranes. This system is more complicated than most of those just
silvered mirror; P,,2, movable prisms; E, eyepieces and auxiliary lenses for imaging;
described in that the dye contributes (with different quantum yields) to
F3,5,G, red suppression and cutoff filters; F" graded interference filter; C, camera; PM,
the fluorescence of both the aqueous phase and the membrane, thus making
photomultiplier; or, opaque illuminator; EI~" black film electronics; Elp, photometric
the requirements discussed above particularly stringent. Horizontal black electronics.
lipid membranes of approximately 2.5 mm diameter were formed in a
special cell on the stage of a fluorescence microscope (Fig. 29). An area about 0.4 mm In diameter was illuminated through an immersion dark-
ground condenser. Dark-ground illumination in incident light was also
to syri nqe or
manometer possible. Visible inspection of the membranes in visible light could be
microscope achieved either under dark-ground illumination or in reflected light. The
u
objective
fluorescence radiation was analyzed by passage through a graded strip
interference filter and detected by a photomultiplier. A variable diaphragm
in the image plane allowed selected portions of the membrane to be mea-
sured. Such a system (Fig. 30) is particularly attractive because of the pos-
sibilityof combining quantitative fluorescence, electrical, and microscopic
studies in a single experiment. A further valuable property which may
compensate for the higher cost is its versatility. The black film cell is easily
replaced by suitable chambers to study single cell membranes or by troughs
for the study of monolayers at the air-water or oil-water interfaces. The
system may of course be made more sophisticated, e.g., by the addition of a
Fig. 29. The black film cell used by Zingsheim and Haydon (1973) for fluorescence measure- reference light path and high-intensity light sources. If the graded interference
ments. The horizontal membrane is illuminated by an immersion dark-field condenser filter is replaced by a cutoff filter and if a high-pressure mercury arc or an ul-
through a thin window in the outer cell and is viewed by a microscope, the objective of traviolet laser is used for the illumination and photocounting techniques are
which is shown. employed, the sensitivity is easily sufficient to measure the fluorescence of as
from the interface between the equilibrium bulk lipid solution and the
aqueous phase was used, as discussed in Section 4. The adsorption of the
ANS was therefore determined by interfacial measurements and the ap-
plication of the Gibbs equation. For low membrane concentrations of the
dye «3 X 1013 moleculesjcm2), the fluorescence intensity was directly
proportional to the number of ANS molecules per unit membrane area
(Fig. 31). The sensitivity of the instrument was calibrated by measuring the
fluorescence of the ANS in the aqueous phase. For this purpose, the light
intensity distribution or, equivalently, the measured volume of the aqueous
phase had to be determined experimentally. Inner filter corrections were
applied and control experiments insured that artifacts such as discussed
earlier had been eliminated. Both the emission spectra and the quantum
yield agreed well with the respective data obtained from measurements on
sonicated aqueous dispersions of the lipid. This fact and the internal
consistency of the combined spectroscopic and thermodynamic measure-
ments satisfactorily confirm the validity of the thermodynamic approach
to the membrane composition through interfacial tension measurements
and the application of the Gibbs equation.