Professional Documents
Culture Documents
2
MICROSCOPY
OBJECTIVES
At the end of this laboratory activity, the students will be able to:
Label the different parts of the microscope, and fill up the table below.
6.
A. Parts of the Microscope
7. Ocular Lens
7
(Eyepiece)
1Body Tube
2 Revolving Nosepiece
2. 8. Arm
8
3
3. Objectives
9. Stage
9
4. Stage Clips
4
10. Coarse
10 Adjustment Knob
5 Condenser
5.
11. Fine11
Adjustment Knob
6 Source
6. Light
12Base
12.
B. Functions of the Microscope
Parts of the Microscope Function
1. Body Tube Connects the eyepiece to the objective lenses.
2. Revolving Nosepiece The part of the microscope that holds two or more objective lenses that
can be rotated to easily change power
3. Objective lenses These are the major lenses used for specimen visualization
5. Condenser Lenses that are used to collect and focus light from the light source into
the specimen.
6. Light Source Shines up through the slide
7. Ocular Lense The top of the microscope where the viewer looks through
10. Coarse Adjustment Allows quick focusing by moving the objective lens or stage up and
Knob down
11. Fine Adjustment Knob Used to fine focus the image when viewing at the higher magnifications
12. Base Acts as microscope’s support
Answer: When examining your specimen under the microscope, you must always begin observing
them using the low power objective. Starting with this objective would provide you a wider field
of view, this would help you locate your specimen a lot easier. After locating the structure that is
needed to be observed and putting it at the center, you finally can switch to the higher objectives
for higher magnification.
5. You are planning to create a better compound light microscope—one that will enable
you to see objects smaller in diameter than 0.2 µm. You gather together the best lens
grinders in the world and put them to work in a lens-grinding laboratory having
unlimited resources. You instruct them to grind marvelous magnifying lenses and add
them to an existing compound light microscope. What’s wrong with this plan?
Answer: There are two things that are wrong in this plan. First, considering that the compound
microscopes use light to produce images and visible light has certain wavelengths (380nm -
750nm), it limits the microscope to produce images that are smaller than the half of the length
of the light waves. This means that it is impossible to make a compound light microscope that
enables someone to see objects smaller than 0.2 µm even using marvelous magnifying lenses.
Second, adding magnifying lenses to an existing compound light microscope is not also a good
plan. Given the fact that the compound light microscopes have limitations that were stated
previously and you have unlimited resources, you might as well create a different kind of
microscope that does not use visible light to produce images of objects smaller than 0.2 µm.
6. While using a compound light microscope, you have noticed that additional light is
required as you increase magnification. Describe three correct ways to increase the
amount of light entering the objective lens, and one incorrect way.
Answer: Increasing or reducing the light intensity (using the microscope rheostat control, moving
the condenser lens closer to or away from the sample (using the condenser adjustment knob),
and opening/closing the iris diaphragm are the three main ways to change the brightness. When
one has a condenser adjustment knob, it is incorrect to adjust the knob while using the
microscope to avoid the sample being in the wrong position.
Type Resolving Typical Uses of Microscope Sample Image Taken under the
Power Microscope (write the type of
organism)
Bright-field 0.2 μm ● used for viewing live and
stained specimen
Dendritic cells
Transmission 1 nm ● used to observe the
Electron detailed internal
Microscope structures of cells
Radiolarians
IV. Calibration of the Ocular Micrometer on a Microscope
Determine the calibration factors of each of the microscopes described below. Show solutions
(you can solve on a piece of paper, and then attach the photo of your solution to the google
classroom assignment) and write here the final answer.
3. At a total magnification of 100x, a student measures 16.4 ocular micrometer divisions per
millimeter. What is the distance, in micrometers, per ocular unit at 100x?
4. At a total magnification of 400x, a student measures 4.1 ocular micrometer divisions per
millimeter. What is the distance, in micrometers, per ocular unit at 400x?
6. Upon computing all the values for ocular divisions using all the objective lenses that came
with your microscope, is it necessary to always re-calibrate the lens every time you use
the microscope? When will it be necessary?
Answer: Every magnification has its own calibration settings, re-calibration is necessary when
the total magnification is changed. If images or measurements are taken using incorrect
calibration settings, the magnification indication and measurement values will also be incorrect,
requiring the whole process to be performed again.
Presented in the table below are the calibration factors presented in the video,
“Calibration of the Microscopic Ocular Micrometer” by Gilles Buldoc in 2017.
Determine the sizes of the following samples viewed under the microscope:
1. You are viewing the red blood cells under the 400x magnification. What is the
diameter/length of the red blood cells that are being pointed by the arrows?
1a
1b
Answer: a= 20 μm in diameter/length
b= 27.5 μm in diameter/length
Show how you computed this. You can do the computation on a paper, and you may attach a
photo to the google classroom of how you solved this.
Answer:
2. You are viewing the following cell in the scanner. What is the length of the sample?
Show how you computed this. You can do the computation on a paper, and you may attach a
photo to the google classroom of how you solved this.
3. You are viewing the following cell in the high-power objective. What is the length of the
sample?