Professional Documents
Culture Documents
The Performance of Alloy 625 in Long-Term Intermediate Temperature Applications
The Performance of Alloy 625 in Long-Term Intermediate Temperature Applications
&
Peter Tail
Methanex New Zealand Limited, New Plymouth, New Zealand
for feed stock superheaters which operated at Table 2. Mechanical properties of materials
between 480 and 560°C with the higher end of
New" Used
this range being reached only occasionally.
Temperatures were typically limited to 500°C and 20°C 500°C 20°(2 500°C
at the time that Alloy 625 was specified, no
UTS (MPa) 937 844 1 384 1 257
reference to hardening at this temperature was 0.2% Proof (MPa) 461 375 1 083 995
found and the general consensus from a variety % Elongation 46 40 6-7 21
of material suppliers and laboratories was that %ROA 60 37 7 31
Charpy impact (J/cm 2) 178 197 19 30
Alloy 625 would perform satisfactorily at these
Hardness H V (30) 230 444
temperatures.
At a routine plant outage, a small amount of " T h e test results for the new material were very similar to the
mechanical properties quoted on the test certificate for the used
maintenance welding was undertaken and it was material supplied at the time the vessel was manufactured.
found that the material weldability was poor and
cracking occurred. Subsequent on-site inspection
found that the material was significantly harder
than expected. holder used to accurately locate the specimen in
Samples of material which had been in service the testing machine and to which was strapped a
for approximately 5 0 0 0 0 h have been made calibrated 'K-type' thermocouple. The specimens
available and a programme of work was were removed from the furnace and as rapidly as
undertaken to characterise this material and possible placed in the testing machine. The actual
compare it with material supplied as new. time between removal from the furnace and
impact was determined by reference to cooling
curves that had previously been produced using a
MATERIALS standard test piece to which a 'K-type'
thermocouple had been spot welded close to the
The used Alloy 625 has been part of the shell of a V-notch.
feed stock superheater. It had been supplied with It can be seen by reference to Table 2 that the
original plant and had been in service for Charpy impact properties of the used material
approximately 5 0 0 0 0 h (6 years). The heat were poor. Consequently, it was decided to
treatment prior to delivery was 1 h at 1000°C measure the fracture toughness of the material.
followed by a quench. The new material tested Fracture toughness testing was conducted on
for comparison was supplied as being 'annealed samples of the used material. Crack tip opening
at 1050°C, rapid cooled'. The chemical composi- displacement (CTOD) measurements were de-
tions are listed in Table 1. termined using the technique described in BS
5762. 6 Using specimens of 1 3 m m nominal
Table 1. Material chemical composition thickness, the critical C T O D (maximum load)
Cr Mo Nb Fe C Si AI Ti Mn was determined at room temperature to be
0.022 mm.
Used 20. 8.4 3-78 4.2 <0-10 0.20 <0.05 0.08 0.13 C T O D testing is undertaken in order to
New 23 9-0 3-24 4.1 <0-10 0.22 <0.05 0-25 0.06
provide data to support engineering assessments
of pressure vessels and other similar equipment
under circumstances where linear elastic fracture
mechanics becomes invalid, i.e. when the
MECHANICAL PROPERTIES material in question exhibits some degree of
plasticity. In order to obtain a valid value of
The basic mechanical properties of the materials fracture toughness in most metallic materials, a
are listed in Table 2. certain minimum thickness of test piece must be
Charpy impact testing was conducted at both used. Below this minimum thickness, linear
room and plant operating temperatures of 500°C. elastic fracture mechanics (LEFM) do not apply
The elevated temperature testing was done by and valid measurements of fracture toughness,
soaking the specimens in a small fluidised-bed K~c cannot be obtained. The standard for plain
furnace held at approximately 520°C. The strain fracture toughness 7 testing requires that
specimens were held in a stainless-steel test piece the test piece thickness and crack length should
Alloy 625 in long-term temperature applications 43
both be greater than Table 3. Charpy impact properties of used material after
heat treatment (room temperature)
x (K'c) 2
2.5 --~y/ Heat treatment Impact energy Hardness
(J/cm 2) HV(30)
where try is the 0.2% proof stress. C T O D testing
600°C, 12 h 36 373
was chosen in the expectation that the used 650°C, 12 h 61 307
material would exhibit some ductility. The value 700°C, 12 h 57 302
of fracture toughness K at the m a x i m u m load was 650°C, 3h 54 317
650°C, 6h 56 303
calculated according to the procedures of either
BS5447 or BS5762 and found to be
2459 N / m m -3~2. The nominal thickness of the test
piece and the crack length were both 13 m m and Such a heat treatment would have to be similar
therefore the value of 2459 N / m m -3/2 just to an in-situ post-weld heat treatment, for
qualified as being a valid value of K,c for example, and was therefore limited by both
material of yield strength of 1083 MPa (see Table temperature and duration. Samples of the used
2). The above equation would require that the material were heat treated at 600, 650 and 700°C
test piece should be at least 12-9mm thick. In for 3,6 and 12h duration and the toughness
other words, the plasticity exhibited by the used assessed by Charpy impact testing and hardness.
material in a test piece of this size was sufficiently The results are presented in Table 3. C T O D
low that valid KI¢ values could be obtained. testing was undertaken on used material that had
Further evidence of the low ductility of the been heat treated for 12 h at 700°C in a similar
used material was provided during pre-cracking manner to that for the used as-received material
of one of the test pieces prior to C T O D testing. and in this case the critical C T O D (maximum
Fatiguing was conducted in accordance with the load) was found to be 0-155mm at room
British Standards 6,7 at a fatigue stress intensity temperature. In contrast with the very small flaw
(Kf) of 8 0 0 N / m m -3/2. Complete failure of a size considered critical in the used material (e.g.
particular test piece occurred during the fatiguing 2 mm long by 1 m m deep), the level of toughness
operation when the crack had grown to a length after heat treatment indicated that infinitely long
of 14.13mm (test piece thickness 12-85mm, cracks up to 2.5 m m deep could be tolerated.
width 25.70mm). U n d e r the higher strain rate Again, this was estimated using the procedures
conditions associated with the fatiguing opera- laid down in British Standards Institution
tion, the fracture toughness was considerably published document PD6493. 8
lower than when testing was conducted at usual
loading rates. Other than the fact that the crack
had been allowed to propagate slightly further MICROSCOPY
than was desirable, the conditions under which
the fatiguing operation was conducted were Optical microscopy
comfortably within the requirements laid out in
the standards. The optical microstructures of the new, used and
The C T O D results were used to indicate the used + heat treated (600°C, 12 h) are shown in
size of defect in the superheater that could lead Figs 1, 2 and 3, respectively. The microstructure
to catastrophic failure, using the techniques of the new material was as expected, having a
described in British Standards Institution docu- mainly equiaxed grain structure and the anneal-
ment PD6493. 8 The procedure indicated that ing twins commonly observed in austenitic
surface breaking defects of the order of 2 m m materials. A range of small discrete particles,
long by l m m deep were critical at room presumably carbides of the (Mo, Nb)C o r Cr23C6
temperature. types were present both on the grain boundaries
Following the discovery of the very low and within the grains. Occasional larger particles
ductility in the used material and therefore of having the characteristic pink colouration of
similar material that remained in service, an Ti(C, N) were also present.
attempt was made to recover some toughness by The microstructure of the used material was
a heat treatment that could be performed on the essentially the same except that in the particular
vessel in situ during a maintenance shut-down. etch used (15 parts HCI, 10 parts H N O 3 , 10 parts
44 Charles Thomas, Peter Tait
• "°, 2(~,tl
Fig. 1. Microstructure of new material. Fig. 4. Surface detail of polished and etched sample of used
as-received material as seen in the scanning electron
microscope.
• • •
• • •
• • •
Fig. 9. Bright field image of the used + heat treated (700°C,
12 h) material.
• • +,
made generation of a dark field image difficult. In
addition, the unidentified reflections that had
(a)
been present in the diffraction patterns of used
material were not seen in the heat-treated
material.
Figure 9 also illustrates needle like precipitates
020 22O that were not found in either the new or the
as-received used material. It is believed that
O
these are small 6 phase platelets, 6 being the
orthorhombic form of Ni3Nb. 12,13
O In the thin foils of new and used material,
other precipitates were not found to any
O o
0,o °O significant extent. Energy dispersive X-ray
analysis (EDX) analysis of those particles that
were found tended to indicate that they were rich
in particularly molybdenum and niobium. This
O would suggest that they were carbides of the M6C
000 20O type. 14
In the u s e d + h e a t treated materials, grain
boundary precipitates had been seen when
examined in the scanning electron microscope.
(b)
Fig. 8. (a) Diffraction pattern obtained from used as
received material [00]y zone axis. (b) Diagrammatic
representation of indexed pattern. Spots (340), (½10) and
(110) were used to create the composite dark field image in
Fig. 7(b).
Similarly, grain boundary precipitates were by use of crack tip opening displacement tests
observed in thin foils. However, grain boundary indicated that the vessels had negligible defect
precipitation was certainly not continuous in this tolerance and it was concluded without recourse
material and many grain boundaries in the thin to more detailed analyses, that salvation for these
foils appeared free of precipitates. Examples of a vessels lay in improving the material properties
precipitate is shown in Fig. 10. E D X analysis of rather than attempting to demonstrate their
these precipitates was inconclusive due to their integrity.
small size. However, in addition to the M6C Trial laboratory heat treatments were chosen
carbides seen in the new and used materials, in order that similar heat treatments could be
particles with a modest enrichment in molybde- conducted on-site during a plant maintenance
num and niobium relative to the general material shut down and were therefore constrained by
analysis were observed and it was suspected that temperature and duration. For example, there
these were blocky grain boundary 6 were other components in the vessel constructed
precipitates. 15 of carbon steel and, in addition, there was a
concern that at temperatures much above 700°C,
deformation of the vessel under its own weight
would occur. Fortunately, heat treatment at
DISCUSSION comparatively modest temperatures of 650-
700°C and for reasonable times of 6-12 h proved
After approximately 50000h of service at beneficial in regenerating some toughness. While
approximately 500°C, Alloy625 underwent sig- the increase in Charpy impact energy was not
nificant age hardening. The increase in strength massive (from 20 to 60J/cm2), it was most
was dramatic, the 0.2% proof stress more than welcome in increasing confidence as to the safety
doubling from 461 MPa to 1083 MPa (Table 1). of the vessels. The increase in CTOD value was
This degree of strengthening appears to be more dramatic from 0-02 mm to 0.15 mm and the
significantly in excess of other reported results improvement in defect tolerance as indicated by
for laboratory aged Alloy 625. 3"14"16 Indeed, only the procedures of PD6493 (Ref. 8) was highly
materials that have been deliberately alloyed to significant.
encourage an age hardening response by the Examination of the material microstructure
addition of titanium 4 or aluminium in Alloy 718 showed that the hardening mechanism was
(Ref. 17), for example, exhibited comparable caused by precipitation of 7". In the optical
tensile properties and hardnesses. The high examination, the precipitation was too fine to be
strength of the used material was also evident resolved but the presence of a precipitate was
when testing was conducted at the operating readily revealed by the heavy darkening of the
temperature of 500°C. matrix. Similarly, scanning electron microscopy
More importantly however, the increase in examination revealed a myriad of uniform and
strength was accompanied by a massive reduction extremely fine precipitates. It was however, the
in toughness and the consequences in terms of transmission electron microscopy that revealed
the integrity of the pressure vessels in question structures and diffraction patterns that confirmed
and indeed any similar plant, were of great the presence of 7".2'""~ However, at this time the
concern. Ductility of the used material as reason for the additional diffraction spots seen on
indicated by elongation and reduction of area the thin foils from the used material has not been
measurements, was significantly higher at the identified.
plant operating temperature. This trend was not Age hardening of Alloy625 is a recognised
strongly reinforced by the Charpy impact tests phenomenon ~,3 but not references to the
suggesting that elevated temperature toughness operation of Alloy 625 at 500°C have been found.
was strain rate dependant. The improvement in Typically, ageing experiments down to 600°C
ductility at operating temperature compared to have been conducted and at least two time-
room temperature was insufficient to allay fears temperature-transformation diagrams have been
concerning the integrity of pressure containing constructed for Alloy 625 (reproduced in Ref. 3).
plant in which Alloy 625 has been exposed to These differ largely in the predicted rate of
intermediate temperatures. precipitation of the M23C6 and M6C type carbides
Preliminary assessments of fitness for purpose but neither of them would indicate through
48 Charles Thomas, Peter Tait
extrapolation that y" should be expected at 500°C treatment. Other damaging precipitates were not
after 50 000 h. observed to any great extent.
Examples of Alloy 718 have been aged for up Once an age hardening material has been
to 50 000 h at temperatures as low as 538°C (Ref. over-aged, it is not possible to regenerate the
17) but the objective of that work was to assess maximum hardness without recourse to a
the possibility of softening by overageing of a solution heat treatment. In Alloy 625, this could
material that had previously been age hardened not occur at the low heat treatment temperatures
deliberately. The work showed that aging for up employed and it is therefore tempting to believe
to 50 000 h caused a slight increase in strength of that the problem will not re-occur since it is
age hardened Alloy718. Controlled aging of highly unlikely that 7" can or will re-precipitate
Alloy 625 has shown 3 that 0.2% proof strengths at the operating temperature. However, while it
of up to 800 MPa and hardness of approximately is undoubtedly true that the improvement in
350HV can be achieved but these are con- properties after heat treatment was most
siderably lower than observed in the material welcome relative to the very low toughness
aged for 5 0 0 0 0 h at 500°C. Information pre- observed in the used as-received material, the
viously available in the literature has indicated toughness remained low compared to new
that titanium and aluminium leading to the material. It is possible that this was due to the
precipitation of y' are necessary in order to formation of the orthorhombic 6 phase. Previous
generate such high strengths. work 3 has found that reduction in Charpy impact
The remedial heat treatments that were found energies from around 180 to J/cm 2 as seen here
to improve toughness were believed to have occurred as a result of intermediate temperature
worked due to an over-ageing effect. Following ageing and the ductility loss was attributed
heat treatment, transmission electron microscope primarily to the formation of the orthorhombic 6
examination of thin foils showed that the density phase.
of 7" precipitates fell dramatically and the At the ageing temperatures and short times
intensity of the superlattice streaks on the used in the heat treatments in this work, it is also
diffraction patterns due to y" was significantly possible that the formation of 6 Ni3Nb may not
reduced. The extra reflections that had been seen be complete and therefore, the reduction of
on diffraction patterns and were not identified toughness due to this mechanism should not be
disappeared after heat treatment and it is totally discounted.
therefore possible that whatever caused these The superheaters in question were sub-
reflections also contributed to the particularly sequently heat treated on-site by the use of
low toughness of the used material. heating pads and the complete vessels were
The heat treatments that were trialled and raised to 700°C. They were subsequently
found to improve toughness were quite short, the returned to service and have continued to
longest being 12 h duration. Previous work has operate without problem.
reported on the overageing of alloys hardened by
precipitation of 7" but generally, reduction in
hardness would be slower. For example, 2 a ACKNOWLEDGEMENTS
material aged at 650°C achieved a peak hardness
of 400 HV after about 500 h and the hardness fell The authors thank Margaret Borshevsky and
to only 350 HV after a total aging time of about Peter Gilberd for their assistance with the
1000h. 2 However, a t i m e - t e m p e r a t u r e para- electron microscopy, Neil Briggs and Terry Sage
meter such as Larson-Miller ( P = T(K+ who undertook much of the mechanical testing
log t)/1000; where T is absolute temperature, t and Methanex New Zealand Limited and
time (h), K is a constant and P the Larson-Miller Industrial Research Limited for permission to
parameter) which is frequently used to describe publish this paper.
stress rupture times for example suggests 12 h at
650°C is approximately equivalent to 150 000 h at
500°C (for K = 20). While the presence of the REFERENCES
orthorhombic 6 form of Ni3Nb was not confirmed
by diffraction pattern or other analytical tech- 1. Eiselstein, H. L. & Tillack, D. J., The invention and
niques, it was believed that this formed after heat definition of alloy 625. In Superalloys, 718, 612 and
Alloy 625 in long-term temperature applications 49
Various Derivatives, ed. Edward A. Loria. Minerals, 10. Paulonis, D. F., Oblak, J. M. & Duval, D. S.,
Metals and Materials Society, Warrendale, PA, 1991. Precipitation in nickel-base alloy 718. Trans. ASM, 62
2. Kirman, I. & Warrington, D. H., The precipitation on (1969) 611.
Ni3Nb phases in a N i - F e - C r - N b alloy. Met. Trans., 1 11. Mansfield, J., Convergent Beam Electron Diffraction of
(1970) 2667. Alloy Phases. Adam Hilger Ltd, Bristol, UK, 1984.
3. Kohler, M., Effect of the elevated temperature 12. Sjoberg, G. & Ingesten, N. G., Grain boundary
precipitation in alloy 625 on properties and microstruc- 6-phase morphologies, carbides and notch rupture
ture. In Superalloys 718, 612 and Various Derivatives, sensitivity of cast alloy 718. In Superalloys 718, 612 and
ed. Edward A. Loria. Minerals, Metals and Materials Various Derivatives, ed. Edward A. Loria. Minerals,
Society, Warrendale, PA, 1991. Metals and Materials Society, Warrendale, PA, 1991.
4. Cozar, R., Rouvy, M., Mayonobe, B. & Morizot, C., 13. Sundararaman, M., Mukhopadhyay, P. & Banerjee, S.,
Mechanical properties, corrosion resistance and micr- Precipitation of the 6-Ni3Nb phase in two nickel base
ostructure of both regular and titanium hardened 625 superailoys. Met. Trans. A, 19A (1988) 453.
alloys. In Superalloys 718, 612 and Various Derivatives, 14. Ferrer, L., Pieraggi, B. & Uginet, J. F., Microstructural
ed. Edward A. Loria. Minerals, Metals and Materials evolution during thermomechanical processing of alloy
Society, Warrendale, PA, 1991. 625. In Superalloys 718, 612 and Various Derivatives,
5. Thomas, C. W. & Tait, P., Failure prevention and life ed. Edward A. Loria. Minerals, Metals and Materials
prediction in New Zealand petrochemical industry. In Society, Warrendale, PA, 1991.
Proc. Syrup. Failures Repairs and Life Assessment in 15. Burke, M. G. & Miller, M. K., Precipitation in alloy
Pressure vessels and Pipework. 14-15 Nov. 1991, 718: A combined AEM and APFIM investigation.
IMMA, Melbourne, Australia. Superalloys 718, 612 and Various Derivatives, ed.
6. BSI, Methods for Crack Opening Displacement (COD) Edward A. Loria. Minerals, Metals and Materials
Testing. (BS 5762:1979). British Standards Institution, Society, Warrendale, PA, 1991.
London, UK. 16. Vernot-Loier, C. & Cortial, F., Influence of heat
7. BSI, Methods of Test for Plane Strain Fracture treatments on microstructure, mechanical properties
Toughness (Ktc) of Metallic Materials (BS 5447:1977). and corrosion behaviour of alloy 625 forged rod. In
British Standards Institution, London, UK. Superalloys 718, 612 and Various Derivatives, ed.
8. BSI, Guidance on Methods for Assessing the Accep- Edward A. Loria. Minerals, Metals and Materials
tability of Flaws in Fusion Welded Structures Society, Warrendale, PA, 1991.
(PD 6493:1980 and 1991). British Standards Institution, 17. Korth, G. E. & Trybus, C. L., Tensile properties and
London, UK. microstructure of alloy 718 thermally aged to 50,000
9. American Society of Metals, Metals Handbook (8th hours. In Superalloys 718, 612 and Various Derivatives,
edn, Vol. 8), American Society of Metals, Materials ed. Edward A. Loria., Minerals, Metals and Materials
Park, OH, p. 102. Society, Warrendale, PA, 1991.