Professional Documents
Culture Documents
Issue 1, 1996
Zoubair El Fallah
Table of Contents
1
optimisation (1). An intermediate gradient methods are avoided due to These gradient runs differ in their
approach to methods development the fact that they are very sensitive to slopes; that is, b1 and b2 while keeping
would take advantage of both the chro- the type of instrumentation used, and all other conditions the same. The first
matographer’s intuition and experience therefore, they are not easy to transfer gradient conditions (i.e., Ø0 and b1,
and the theory of chromatographic and to establish as routine rugged ana- have to be selected so that all peaks
retention. This is the essence of the lytical methods. Isocratic methods are are eluting primarily in the gradient por-
method development strategy presented preferred but they suffer from the fact tion; this supposes a relatively low Ø0,
here. It is based on the use of that for relatively complex samples, the and that enough resolution is obtained
Symmetry ® columns (2) with pH and sensitivity of the assay is relatively low, between all peak pairs or peaks of
solvent composition optimisation. especially for late eluting peaks. This interest, which means that b1 should be
Since the majority of pharmaceutical effectively limits peak capacity. around 0.5 to 1% organic change per
compounds are ionisable, selectivity is minute. b2, is typically either twice or
In reversed-phase HPLC, retention times
more likely to be affected by changes half b1, depending on the retention
obtained from linear solvent strength
in the pH of the mobile phase. range of the first run.
gradients are related to isocratic elution
Therefore, the possibility of working data and to parameters that depend on The combination of two sets of Eq.2;
over the entire pH range without the nature of the solute and solvent in that is, two different gradient runs,
adversely affecting the performance of use. This relationship between gradient leads to the following relationship
the system is an important consideration and isocratic retention is based on a which can be numerically solved for S:
in method development. This is espe- Linear Solvent Strength theory intro-
cially true when the solutes are of basic b2 exp(tr 1.S.b1 – 1) -
duced by Snyder (4,5). Using this
character. With conventional reversed- theory, it is possible to predict retention b1 exp (tr 2.S.b2 – 1) = 0 (Eq.3)
phase HPLC packings, the addition of times under isocratic or other gradient tr i is the reduced gradient retention time,
modifiers to block the silanols is neces- conditions. The theory is based on the obtained with the slope bi:
sary to achieve good peak shape, assumption that the variation of the ln k
especially when operating at neutral tr i = tgi – t 0 – td (Eq.4)
is linear with the volume fraction of the
pH. This has an impact on both the organic modifier: Once S is solved, ko can be obtained
sensitivity and the ruggedness of the from the following relationship:
method, which demonstrates that ln k = ln k 0 – S (Eq.1)
column quality and performance often where S is the slope of that variation k 0 = (1/ to.S.b1)(exp (tr 1.S.b1 +SØ0 )
dictate the strategy for method devel- and k 0 is the retention factor in pure - exp(SØ0 )) (Eq.5)
opment. With Symmetry columns, one water.
Therefore, for a given peak, the varia-
can use simpler mobile phases with The retention time of a given peak tion of k vs. fraction of organic is
low-to-moderate buffer concentrations, under gradient conditions is related obtained from retention data of two
and still achieve good peak shapes, to S and k0 through the following gradient runs by solving Eq.3 for S, and
even with strong bases. This makes it relationship: obtaining k0 from Eq.5.
possible to devise a simple strategy for
methods development that is primarily tg = t 0 + td + (1/S.b). This methodology of performing two
based on pH. Often, pH turns out to ln(1+ S.b.t 0.k 0.exp(-S. Ø0)) (Eq.2) runs is used by software packages like
be the principal parameter affecting DryLab™ (LC-Resources)
the separation. However, analysts tend t 0 is the retention time of an unretain- HIPAC™ (Phase Separations)
to examine its effect only when solvent ed solute, td is the delay time of the
HPLC system, b is the slope of the gra-
optimisation fails. We show that consid- Experimental Section
ering pH first, significantly speeds dient (% organic change/min) and Ø0
is the organic composition at the begin- The Waters HPLC system consisted
up the process of method development.
ning of the gradient. of the following: A 616 LC System,
a 717plus Autosampler, a 996
Theory In order to describe the retention of a
Photodiode Array Detector. Data acqui-
given compound in the mixture, one
Gradient elution chromatography is sition was done by the Millennium®
needs to know its corresponding para-
known to be a powerful tool for the 2010 Chromatography Manager.
meters, k 0 and S, in Eq.1. Since it is
analysis of complex mixtures. It is also not practical to perform isocratic runs For the separation of the mixture of
the method of choice when dealing to derive these parameters, due to the antidepressants and their metabolites,
with partially or totally unknown sam- fact that one needs to guess what com- a Symmetry C18 5µm (3.9 mm I.D. x
ples, which is the case when impurity position would lead to a reasonable 150 mm) column was used. Column
profiling is performed on a drug sub- retention window, the use of gradient temperature was maintained at 35 °C.
stance and/or product. Gradient chro- retention data, as shown in Eq.2, is a Eluent A was either 20 mM phosphate
matography makes it possible to better alternative. Two gradient runs buffer at pH 2.3 or pH 7.0 (MilliQ
achieve a higher peak capacity per unit using the selected solvent are needed water, Millipore), while solvent B was
time (3), and it has a better sensitivity in order to estimate k 0 and S, for each methanol. The flow rate was 1 mL/min.
than isocratic mode, as band concen- one of the peaks in the chromatograms. The effluent was monitored at 254 nm.
tration often occurs. However, in the
pharmaceutical laboratories,
2
Results and Discussion: Figure 1: Effect of PH on the separation of 2 antidepressants and their metabolites.
AU
reversed-phase columns. 0.10
0.05
Since Symmetry columns give good
peak shapes independent of pH, 0.00
3
10 minutes (i.e., retention time of Figure 3: Separation of two antidepressants and their metabilites at pH 7.0.
amitriptyline, the last eluting solute),
we reach the conclusion that 71% 71% Methanol / 29% 20 mM potassium phosphate buffer.
MeOH should be selected. However, 0.25
1 Nordoxepin
at this point we have to examine two
1 2 Doxepin
important things: 0.20
3 Nortriptyline
– do we have enough retention for
2 4 Amitriptyline
the first eluting peak? 0.15
3
AU
– do we have enough resolution 4
between all peak pairs? 0.10
Based on S and k0 values for nordox-
epin, we found that its retention factor, 0.05
k at 71% MeOH, is 0.64. Selectivity
values between adjacent peak pairs
0.00
were also calculated at 71% MeOH,
based on values reported in Table I, 0 2 4 6 8 10 12 14
and were found to be 2.4, 2.1 and Minutes
2.0, respectively. With today’s
columns, these selectivities are expect- The final chromatogram is obtained using the predicted mobile phase of 71% methanol
ed to give baseline separation. Figure 3 and 29% 20 mM phosphate buffer at pH 7.
shows the separation obtained at 71%
MeOH/29% 20 mM phosphate buffer
at pH 7.0. As expected, separation of Conclusion:
the four solutes was achieved with a run
Often it is necessary to include pH impurity profiles and assays for degra-
time of about 11 minutes. However, the
as a criterion in method development. dates. We have shown that following
retention factor of nordoxepin is much
Gradient runs at two different pH this strategy, the use of Symmetry
higher than predicted, and selectivity
values (acidic and neutral) in the early columns can significantly reduce the
values were 2.0, 1.3 and 2.1, respec-
stage of the optimisation are very time of methods development.
tively. There are several reasons that
useful in directing the analyst towards
could account for the difference
the right pH.
between predicted and observed
selectivities: A second gradient run is then per- References
formed at the selected pH, and its data
– the linear model is only an approxi- (1) A. Tchapla, P. Allard and S. Heron, LC-GC Intl. 6
can be successfully used to locate the (1993), 86.
mation, and quadratic behavior has
optimum composition, based on time
been observed (6). This is especially (2) U.D. Neue, D.J. Phillips, T.H. Walter, M. Capparella, B.
or/and resolution criteria. Once the pH Alden and R.P. Fisk, LC-GC 12 (1994), 468.
true with ionisable solutes, as they may
and retention range are selected,
elute under a mixed mode retention. (3) J.M. Davis, J.C. Giddings, Anal. Chem. 55 (1983),
mobile phase mixture optimisation can 418.
– errors in estimating delay volume will be carried out, when necessary. In
affect early eluting peaks more than (4) L.R. Snyder, In “ High Performance Liquid
order to carry out this strategy success- Chromatography, Advances and Perspectives”, Vol.1, C.
late eluters. fully, a column with good performance Horvath, ed., Academic Press, New York, (1980), pp 207-
at acidic as well as neutral pH is nec- 316.
– there are inherent proportioning
errors, typically 0.5%, both during gra- essary. This was a simple example (5) L.R. Snyder, J.W. Dolan and J.R. Grant, J. Chromatogr.
demonstrating the basic approach. 165 (1979), 3.
dient and final isocratic composition.
We have applied this strategy to signifi- (6) P.J. Schoenmakers, H.A.H. Billiet, R. Tijssen and L. De
Exact prediction is not the purpose of Galan, J. Chromatogr. 149 (1978), 519.
cantly more complex samples, such as
the strategy presented here. Data from
gradient runs are used only to guide us
to the retention window.
4
Ordering Information
Symmetry Analytical Steel Columns
Chemistry Particle Size Dimensions Part No. fi Automated gradient methods optimisa-
Symmetry C8 5 µm 2.1 mm x 150 mm WAT056955
tion assumes excellent integrity in HPLC
Symmetry C8 5 µm 3.0 mm x 150 mm WAT054230 system gradient generation. Efficient,
Symmetry C8 5 µm 3.9 mm x 150 mm WAT046970 precise and accurate gradient formation
Symmetry C8 5 µm 4.6 mm x 150 mm WAT045995 is a key criteria for HPLC solvent delivery
Symmetry C8 5 µm 4.6 mm x 250 mm WAT054270 systems, one in which Waters instruments
Symmetry C18 5 µm 2.1 mm x 150 mm WAT056975 excel. In Waters 616 LC System, unique
Symmetry C18 5 µm 3.0 mm x 150 mm WAT054200 gradient control software provides reliable,
Symmetry C18 5 µm 3.9 mm x 150 mm WAT046980 reproducible results.
Symmetry C18 5 µm 4.6 mm x 150 mm WAT045905 Waters patented Random Phase
Symmetry C18 5 µm 4.6 mm x 250 mm WAT054275 Synchronisation (RPS™) Software monitors
both eluent composition and flow rate,
Symmetry Steel Cartridge Columns (All cartridge columns require reusable endfittings.) adjusting automatically and instantaneously
Chemistry Particle Size Dimensions Part No. to current operating conditions to ensure
Symmetry C8 5 µm 3.9 mm x 50 mm WAT054240 optimal performance. Combined with a low
Symmetry C8 5 µm 3.9 mm x 150 mm WAT054235 system volume of <400 µL, changes in the
Symmetry C8 5 µm 4.6 mm x 150 mm WAT054255 eluent blend are quickly and accurately
Symmetry C8 5 µm 4.6 mm x 250 mm WAT054245 delivered to the HPLC column to facilitate
Symmetry C18 5 µm 3.9 mm x 50 mm WAT054220 fast separations for rapid methods
Symmetry C18 5 µm 3.9 mm x 150 mm WAT054205 development efforts.
Symmetry C18 5 µm 4.6 mm x 150 mm WAT054210
Symmetry C18 5 µm 4.6 mm x 250 mm WAT054215
Symmetry Validation Kit (three steel columns or steel cartridges from 3 different batches of packing materials)
Chemistry Particle Size Dimensions Part No. For more information on
Symmetry C8 column 5 µm 3.0mm x 150 mm WAT054434 Symmetry columns, please
Symmetry C8 column 5 µm 3.9 mm x 150 mm WAT046955 check box 4 on the Business
Symmetry C8 column 5 µm 4.6 mm x 150 mm WAT054435
Reply Card and return today.
Symmetry C8 column 5 µm 4.6 mm x 250 mm WAT054438
Symmetry C18 column 5 µm 3.0 mm x 150 mm WAT054446
Symmetry C18 column 5 µm 3.9 mm x 150 mm WAT047210
Symmetry C18 column 5 µm 4.6 mm x 150 mm WAT054448
Symmetry C18 column 5 µm 4.6 mm x 250 mm WAT054450 Symmetry Columns are also available
Symmetry C8 cartridge 5 µm 3.9 mm x 150 mm WAT054440 in narrow-bore. See article on page 6.
Symmetry C8 cartridge 5 µm 4.6 mm x 150 mm WAT054442
Symmetry C8 cartridge 5 µm 4.6 mm x 250 mm WAT054444
Symmetry C18 cartridge 5 µm 3.9 mm x 150 mm WAT054452
Symmetry C18 cartridge 5 µm 4.6 mm x 150 mm WAT054454
Symmetry C18 cartridge 5 µm 4.6 mm x 250 mm WAT054456
Part No.
End Connector Kit
(includes 2 sets of reusable endfittings, 2 C-Clips and Kalrez o-rings) WAT037525
* Sentry guard columns can be used with either the steel cartridge columns but require the appropriate Symmetry, Sentry, RPS, and Waters are trademarks of
guard holder. Waters Corp.
** For Waters 3.9 mm x 50 mm cartridge column, you must use the Universal Guard Holder. DryLab is a registered trademark of LC Resources, Inc.
HIPAC is a registered trademark of Phase Separation, Ltd.
5
Performance of Symmetry Narrow-Bore
Columns for Isocratic and Gradient Analyses
Dorothy J. Phillips
6
compromises to meet resolution and conditions are listed in Table 1. Flow Results and Discussion
sensitivity needs (Scheme 2). Transfer rates for all analyses were 0.6 mL/min
The four different diameter columns
of gradient methods from traditional for 3 mm I.D. columns and 0.29
gave good efficiencies for a neutral
analytical to narrow-bore columns mL/min for the 2.1 mm column. The
marker. The 4.6 mm and 3.9 mm I.D.
will be discussed in terms of system 3 mm Symmetry C18 columns were also
columns gave 15,000 and 13,500
demands and corrections for system used at 1 mL/min and compared to
plates for acenaphthene, respectively.
volume (Scheme 3). This report also the 3.9 mm and 4.6 mm columns. All
The 2.1 mm I.D. Symmetry C18 columns
discusses the use of the Waters four diameter Symmetry columns
had average efficiency for acenaph-
Sentry™ Guard column with the (2.1 mm, 3 mm, 3.9 mm, and 4.6 mm)
thene of 10,700 plates. The
narrow-bore columns. were compared at equal linear velocity
3 mm Symmetry C18 column efficiencies
(flow rates based on square of column
for acenaphthene averaged 11,420
Experimental diameter).
plates.
The HPLC system consisted of the 625 Alberta Peptide Mixture
Scheme 1: Speed, Back Pressure and
LC System, a 717plus Autosampler, The peptide mixture is from Alberta Solvent Consumption
and a 486 Tunable UV/Vis Peptide Institute at the University of
Typical back pressure and solvent
Absorbance Detector. The Millennium® Alberta in Edmonton, Alberta, Canada.
requirements for the four column geome-
2010 Chromatography Manager was The sample is called the Peptide
tries are given in Figures 1 and 2 and
used for system control and data acqui- Performance Standard for reversed-
Table 2. Table 2 contains the data for
sition. System suitability software was phase chromatography (Cat. RPS-
the sulfa drugs separated with a
used for data analysis. The convention- P0010). The five decapeptides are
methanol/pH 2.9 acetic acid mobile
al HPLC system band spread (extra- identified below. (3) S1 has three posi-
phase. The back pressures of these
column band spread from injector to tive (3+) charges and S2 to S5 have
columns, when operated at equal linear
detector cell) at 0.25 mL/min was two positive charges (2+). The sample
velocity with the 20% methanol mobile
90 µL. The microbore system had extra- is prepared by dissolving the lyophilised
phase, were 1960 psi and 2150 psi
column band spread of 40 µL with the mixture in 1 mL of 0.1% TFA or solvent
for 2.1 mm and 3 mm columns, respec-
microbore cell in the detector (2.7 µL A; this should give 100-200 injections
tively. The dramatic decrease in solvent
volume). The column is removed from at 0.1 AUFS at 210 nm.
consumption observed with the narrow-
the system to determine the extra-col- bore columns is immediately apparent
Ac-Arg-Gly-X-X-Gly-Leu-Gly-Leu-Gly-Lys-Amide
umn band spread; acetone is used as
the marker. The band spread is deter- S1 Arg-Gly-Ala-Gly- Gly-Leu-Gly-Leu-Gly-Lys-Amide
7
from Table 2. The 3 mm column however, the high back pressure limits 2000 psi for each. While the efficien-
required 56% less solvent than the the flow rate. The 2.1 mm column has cies show some decrease with column
4.6 mm column for analysis of the sulfa an even more limited flow rate range diameter, the differences are small
drugs. The 2.1 mm column requires due to high back pressures. The faster enough to be of no practical conse-
80% less solvent per analysis than the flow rates on the smaller diameter quence. The results obtained confirm
4.6 mm column. columns also result in a loss of efficien- that selectivity and resolution are inde-
Figure 1 shows the chromatography of cy since the columns are not being pendent of column diameter when all
the sulfa drugs on the different operated at optimum linear velocity columns are operated at equal linear
Symmetry columns at equal flow rate. (at minimum in van Deemter curve). velocity. Additional reductions in extra-
At 1 mL/min flow rate, the smaller the When the columns are operated at column bandspreading should make
column diameter, the higher the back equal linear velocity as shown in Figure the efficiencies for peak 5 even closer
pressure; the back pressure increases 2, the same chromatography can be in value.
from 1900 psi for the 4.6 mm column obtained on the 2.1 mm, 3.0 mm,
to 3150 for the 3 mm column. The nar- 3.9 mm and 4.6 mm I.D. columns. The
row-bore columns have the advantage back pressures for the four columns are
of shorter run times; the analysis is com- similar in this case at approximately
plete in 5 minutes on the 3 mm column
and requires 10 minutes on the 4.6 mm
column. The analysis time could be Figure 1: Performance of Symmetry C18 5µm Columns: Equal Flow Rates
even shorter on the 3 mm column;
Column: a) Symmetry C18 4.6 mm x 150 mm
b) Symmetry C18 3.9 mm x 150 mm
c) Symmetry C18 3.0 mm x 150 mm
Mobile Phase: water/methanol/glacial acetic acid 79:20:1
3 Flow Rate: 1mL/min
1 4 Detector: 254 nm
5
Sample: mixture of 6 sulfa drugs
H 2N SO2NH2 2 6
1: Sulfanilamide
H N a)
H 2N SO2N
N
0.0 2.0 4.0 6.0 8.0 10.0
2: Sulfadiazine
3
4 5 USP Plates USP Tailing Back Pressures
S
1
H2N SO2NH
Peak 4
N 2 6 a) 4.6 mm ID 7560 1.1 1900 psi
b) 3.9 mm ID 5710 1.1
3: Sulfathiazole c) 3.0 mm ID 3160 1.2 3150 psi
NH2 SO2NH
b)
N
CH3
4: Sulfamerazine 0.0 2.0 4.0 6.0 8.0
N
CH3 3 4 5
NH2 SO2NH
1 6
N 2
CH3
5: Sulfamethazine
S NHSO2 NHCOCH2CH2COOH c)
• H2O
N
0.0 2.0 4.0 6.0 8.0
6: Succinylsulfathiazole
Minutes
The smaller the column diameter, the higher the back pressure but the shorter the run time.
Efficiency decreases with column diameter.
8
Table 2: Performance of Symmetry® C18 5 µm Columns: Methanol as Organic Modifier
2.1 mm x 150 mm* 0.52 mL 0.29 mL/min 7.4 min. 2.1 mL 1960 psi (125 atm)
3.0 mm x 150 mm 1.06 mL 0.6 mL/min 7.4 min. 4.4 mL 2150 psi (137 atm)
1.0 mL/min 4.7 min. 4.7 mL 3150 psi (200 atm)
3.9 mm x 150 mm 1.79 mL 1.0 mL/min 7.4 min. 7.4 mL 2450 psi (156 atm)
4.6 mm x 150 mm 2.49 mL 1.0 mL/min 10.3 min. 10.3 mL 1900 psi (121 atm)
1.4 mL/min 7.6 min. 10.6 mL 2020 psi (129 atm)
3
4 5
1 Column: a) Symmetry C18 4.6 mm x150 mm
6 b) Symmetry C18 3.9 mm x150 mm
2
A c) Symmetry C18 3.0 mm x150 mm
d) Symmetry C18 2.1 mm x150 mm
Mobile Phase: water/methanol/glacial acetic acid 79:20:1
Flow Rates: a) 1.4 mL/min
b) 1.0 mL/min
c) 0.6 mL/min
0.00 10.00 d) 0.29 mL/min
3 Sample: mixture of 6 sulfa drugs, 10 to 39 µg/mL
4 5 Injection volumes: a) 14 µL
1 b) 10 µL
2 6
c) 6 µL
d) 3 µL
b)
0.00 10.00
3
4 5
1 6
2 USP Plates USP Tailing
Peak 4
a) 4.6 mm ID 5780 1.1
b) 3.9 mm ID 5710 1.1
c) c) 3.0 mm ID 4350 1.2
d) 2.1 mm ID 4140 1.2
0.00 10.00
3 5
4
1 6
2
d)
0.00 10.00
Minutes
At equal linear velocities, the run times are the same and the efficiencies are similar. The narrow-bore columns offer
reduced solvent consumption and require less sample to achieve equivalent detector response.
9
Sentry™ Guard Column Table 3: Effect of Sentry Guard Column on 3 mm I.D. column
The Sentry Guard columns (3.9 mm x
Compounds USP Plates USP Tailing Back Pressures
20 mm) were developed for use with
Factor
the 3.9 mm and 4.6 mm I.D. columns.
The fact that the guard column has the Acenaphthene
same or smaller I.D. than the analytical without guard column 10,650 1.2 1210 psi
column weighs heavily on its ability to with guard column 11,000 1.2 1300 psi
maintain or improve performance when
in line with the analytical column.(4) Butabarbital (peak 4 Fig. 3)
Therefore, we investigated the use of without guard column 7080 1.2 1610 psi
the Sentry Guard column with narrow- with guard column 7660 1.4 1760 psi
bore columns. As shown in Figures 3
and Table 3 the Symmetry C18 Sentry
Guard column did enhance the perfor-
mance of the 3 mm I.D. column with Figure 3: Effect of Sentry Guard Column on 3 mm I.D. Column
the same chemistry. The retention times
are 25% longer with the guard column Columns: a. Symmetry C18 3mm x 150mm
b. column a with Symmetry C18
as expected; the 20 mm addition to the
Sentry Guard Column (3.9mm x 20mm)
total column length led to a 23% Mobile Phase: 0.1M potassium phosphate, pH7/
increase in total column volume acetonitrile/water 20:30:50 v/v/v
(change 1.06 mL to 1.3 mL). The plate Flow Rate: 0.6 mL.min
Detection: 214 nm
counts for both acenaphthene in the
Injection Volume: 6µL of sample with each barbiturate
plate count test and for the barbiturates at 25 µg/mL
increased 3% to 8%. Therefore, the
Sentry Guard column can be used with a. Symmetry C18 3 mm column
the 3 mm I.D. columns. 3
1
2
0.20 4
H 5
O 7
O N O
C2H 3 6
HN
C2H3 CH2=CHCH2 0.10 8
NH
N O (CH3)2CHCH2
O
H O
4
0.20 5
H CH3 7
O N O O N O 6
C2H3 CH3CH2 NH 0.10 8
NH
O O
0.00
4: Butabarbital 8: Secobarbital
10
Scheme 2: Sensitivity, Load and Resolution Figure 4: Sensitivity as a Function of Column Diameter
Different column geometries are needed
to meet the user's separation demands 4
in terms of resolution, speed, load, sen-
3.5
sitivity and solvent consumption. (1) The
maximum concentration (Cmax) of sam- 3
2.1 mm x 150 mm
ple in the detector at the peak apex is 2.5
inversely proportional to the square of 2
the column diameter and the column
1.5
length (Cmax ¬ 1/(d2 x L)). Smaller 3.9 mm x 150 mm
diameter columns provide an increase 1
AU d. 2.1 mm I.D.
0.0000
5.00 10.00
Minutes
At constant injection volume, the peak height increases as column diameter decreases.
The 2.1 mm column has the highest sensitivity but shows a loss in resolution.
11
Larger diameter columns have more
capacity, allowing separation to be
maintained at larger sample loads.
However, the larger diameter columns (CH3)2NCH2CH3O
CH2COOH
require more sample than the narrow- C=C • HO – C – COOH
bore columns for a given sensitivity or CH2COOH
C2H5
detector response. For the tamoxifen
assay, the 2.1 mm I.D. column requires
only 20 µL of a given concentration to Tamoxifen
12
Figure 6: Effect of Column Diameter on Gradient Analyses at Equal Linear Velocity
– 0.15
a. 4.6mm I.D.
– 0.25
– 0.35
Columns: Symmetry C18 5µm columns; 4.6 mm, 3.9 mm,
10.00 15.00 20.00 25.00 3 mm and 2.1 mm I.D. x 150mm
b. 3.9mm I.D. Flow Rates: a. 1.4 mL/min,
b. 1 mL/min,
0.10
c. 0.59 mL/min,
0.05 d. 0.29 mL/min
Mobile Phase: A. 0.1% TFA aqueous
0.00 B. acetonitrile with 0.1% TFA
10.00 15.00 20.00 25.00 Gradient: 10% to 40% B in 30 minutes, step to 60%
B for 5 min, 100% A for 40 min
c. 3mm I.D. Sample: Alberta Peptide mix
0.10
a. 7µL
b. 5µL
0.05 c. 3µL
d. 2µL
0.00
Detector: 214nm
10.00 15.00 20.00 25.00 Temperature: 35 ˚C
d. 2.1mm I.D.
0.10
0.05
0.00
10.00 15.00 20.00 25.00
Minutes
The system delay volume causes the retention times to become longer as the column diameter decreases; at slower flow rates
it takes longer for Solvent B to reach the head of the column.
Figure 7: Effect of System Delay Volume on Gradient Analysis with Different Diameter Columns
0.10 3
1
2
0.05
0.00
After modification of the HPLC System, shorter retention times were obtained with the 2.1 mm column compared to before the
modification. They were also more similar to what was obtained with a 4.6 mm column.
13
Figure 8: Gradient Analyses with Different Diameter Columns using the Waters 616 LC Delivery System
a) 4.6mm I.D.
0.10 Columns: Symmetry C18 5 µm columns;
4.6 mm, 3.9 mm.
3.0 mm and 2.1mm I.D. x 150 mm
AU
0.05 c. 3µL
d. 2µL
Detector: 214nm
0.00
Temperature: 35 ˚C
10.00
Minutes
0.05
0.00
10.00
Minutes
0.05
0.00
10.00
Minutes
The low system volume of the Model 616 system gives similar chromatography regardless of the column diameter. The removal of the
static mixers further reduces the system delay volume.
References
1. U. Neue, "Selecting the Best Column Dimension for a Chromatography", LC Magazine 4(3) (1986), pp. 250-
Separation", Waters Column VI(1) 1993, pp. 1-5. 254.
2. "Small Bore Liquid Chromatography Columns: Their 4. M. Capparella, W. Foster, M. Larrousse, D. Phillips, A.
Properties and Uses", R. P. W. Scott, editor, John Wiley & Pomfret and Y. Tuvim, "Characteristics and Applications of a
Sons, New York, 1984. New High Performance Liquid Chromatography Guard
Column", J. Chromatogr. A 691 (1995), pp. 141-150
3. C. T. Mant and R. S. Hodges, "Requirement for Peptide
Standards in Reversed-Phase High Performance Liquid
14
Ordering Information
Symmetry Analytical Steel Columns
Symmetry Validation Kit (3 steel columns from 3 different batches of packing materials)
Rapid Po
lyolefin
Additive
Separat
Objective: ion Using
The object Nova-Pa
demon ive of
strate the this application k ®C
HPLC for utility of note
reversed-pha is to 18
Analysis protected large scale purific
synthetic a-tions se
AccQ • of™ Hy
oligonucleot of DMT 1.5 AU
e Serum
oligonucleot
thod for used in
a
ranging wide variety
ides are
success-fully
Alb
10
Objective:
Amino of applic
Bondap
demon ive of
strate the this applicMost ationrecentalysis Waters
polymerase
chain reactio ncing
6
Eluent
A:
Radial-P
ak® HC18
ak™ cartridgHA (8 x 100mm
rbanc
Eluent thanol
1 B: Methan (95:5, , pH
synthetic bonea-tions ter, P=0) 140
rd (i.e. 9 Sample al/Milli v/v)
rifi of DMT
oligonucleot 7 : -Q® water
modifi or Phosph
Antisense cation of Ph Details: P=S) oligon ides. ed (i.e. phospphosphole back- 1 = AMQ 12 protecte
orothioa
ted 21 (95:5,
2 = Asp 7 = NH3 Injectio mer (DMT v/v)
the produ ucleotides to horo-thioated 3 = Ser 8 = Arg 13 = Val n: 12 d)
4 mLs, contain
Therape tic oligon cells. As 10 = Ala 15 = Lys units (21 ing 624
targete inhibit
6 = His 11 = Pro 2 mgs) of O.D.26
used in ucleotides such, “antise d protein 16 = IIe 3
hio
13 in 70% oligonu 0nm
17
Objective: from use of to multigram s-fully gations”
applic related 18 injected
into
e Respo
polymerase Bondap
DNA seque and cost s to se HPLC
scenc
ive of 6
provides tic
demon Most ak® HC18
strate the this application recently,
4 8
reaction of ncingeffective techni
chainisolatio Eluent Radial-P
a
10
A: ak™ cartridgHA (8 x 100mm
Fluore
icatio ns
logy investi 14 30% B, n by-pro
to multig s-fully related ducts using
Note s
from use of applic gations” Column injected
utilization as hybrid DNA.a-tions ram amoun require : Bondap
into Bondap
m
ts of purifie milligram 20 ak
and the as primers for ization Revers ak® HC18
e, 300n
ted failures
and protectin require l oligonu al/Milli
the ioated full ter or g groups
in
d for antisen an 8 x cleotid : Phosph -Q® water
bone modifi ter, P=0) or rd (i.e.ucleot length se therape 100mm Radial- es can Injectio orothio (95:5,
applications ide products
phosp utic investig
be
Pak cartrid rapidly
protecte
n: purified d)
ated 21
mer (DMT v/v)
P=S) oligon ed (i.e. phosp hole back- . for these ations. ge. This 12 mLs, from contain
same undesi
the produ ucleotides to horo-thioated units packin
(21 mgs) ing 624
red synthes
g isof availab O.D.26 is
selectively ,
cells. As ction of targete
in 70% oligonule 0nm reactio
A and in a variety
cleotide n by-prod
d protein inhibit
0
30% B,
techno
such, “antise DMT blocked
injected of column ucts using Bonda
logy investi nse and s within 0 failures into dimens pak
to multig related 5 ion for gram
gations” 10
scale
ram amoun require 15
DNA. milligram 20
Reversed-pha ts of purifie Milligr 25
rapid and se HPLC d synthetic
Minutes 30
am quantit Moffat,
isolation cost effective HC18 35
provid Anti-Sen A.S. 1988
HA reverseies of DMT protect 40
of technique es a purifica Genetic se Technolo (Nov/Dec).
phosphoroth either phosp for the tions as d-phas
e materia ed synthetic Klauser. Engineering gy In Quest “Researchers
Bio/Tech A. 1990 News. for Novel Pursue
hodies require
d for antisenl in an 8 x
oligonu nology. . “Antisen Vol 8. Pages Drugs and
oligonucleot ioated full ter or cleotid se
se therape 100mm Radial- es can be rapidly
Vol 8. 1 and Agriprod
length Pages Start-Ups Surveyed 7. ucts”,
applications ide products utic investig Pak cartrid purified
303 - 304.
”,
. for these ations. ge. This from undesi
same packin red
g is availab synthesis reactio
le in a n by-prod
variety
of column ucts using Bonda
Wa t e r s
dimens pak
ion for
gram scale
Moffat,
Anti-Sen A.S. 1988
Genetic se Technolo (Nov/Dec).
Klauser. Engineering gy In Quest “Researchers
Bio/Tech A. 1990 News. for Novel Pursue
nology. . “Antisen Vol 8. Pages Drugs and
Vol 8. se 1 and Agriprod
Pages Start-Ups Surveyed 7. ucts”,
303 - 304.
”,
APPLICATIO
NS
NOTES
15
A New Solid Phase Extraction Method for
Determination of Triazine Herbicides and
Metabolites in Aqueous Samples
Michael S. Young
Introduction
Atrazine and related triazine herbicides Compared with traditional C18 silica or but the more polar metabolites, such as
are used in great quantities throughout styrene divinylbenzene polymeric phas- desisopropylatrazine, are not suitable
the world for preemergence weed con- es, the Porapak RDX material utilised in for extraction using C18 silica.
trol. In the central United States, for this study shows enhanced retention Sep-Pak® cartridges containing Porapak
example, millions of kilograms of tri- for organic compounds with polar RDX resin were first developed for the
azines are applied each year. In areas functionalities. Among the more polar high efficiency extraction of nitramine
of heavy usage, surface water supplies pesticides, atrazine and simazine are explosives from water (4). In this
are often affected by runoff of these compounds now routinely measured in research, the Porapak RDX cartridge
substances and their transformation drinking water according to EPA has been shown to be a superior SPE
products (1). Therefore, a number of methods 507 and 505 (2). These device for determination of triazine
these compounds are included in the compounds are included in the class herbicides and metabolites in aqueous
list of drinking water contaminants to be of triazine herbicides. Often it is impor- samples. This method has been devel-
monitored under the Safe Drinking tant not only to measure a particular oped for use with natural waters
Water Act. There is also significant analyte such as atrazine, but also the (surface and groundwater) as well as
interest on the part of other agencies, biological transformation products of finished waters (treated municipal
such as the USGS, regarding the fate that substance. For example, a number drinking water supplies). Samples (up
and transport of the triazine herbicides of metabolites are of interest to to 1.0 L) are adjusted to pH 7 with
in the natural environment. In Europe, researchers studying the effects of wide- phosphate buffer and are eluted
where groundwater is utilised for a high spread atrazine use in the Mississippi through the Sep-Pak cartridge at 10
proportion of drinking water supplies, watershed (1,3). Figure 1 shows mL/min. After extraction, the cartridge
the EC has established more stringent atrazine and its principle transformation is eluted using 1:1 methanol/acetoni-
limits than has the US EPA. Currently, products of interest to environmental trile and the resulting extract is analysed
the US limit is 3 g/L for atrazine; the researchers. Atrazine and its sister tri- by HPLC/UV using a Symmetry C18,
European limit is 0.1 g/L for atrazine azine compounds, (simazine, 5µm column.
or any individual regulated pesticide, propazine) are easily extracted from
and 0.5 g/mL for the sum of all pesti- water using C18 silica SPE procedures,
cides. Because groundwater levels in
agricultural areas were consistently
above this limit, Germany banned the Figure 1:
use of Atrazine in 1991, and has rec-
ommended banning the use of this her-
bicide throughout the European (C2H5)NH N NH2(C3H7) NH2 N NH2(C3H7)
HYDROXYATRAZINE DESISOPROPYLATRTAZINE
16
Experimental Section water was 500-1000 mL while the components and were analysed by the
sample size for non-filtered surface new Porapak RDX Sep-Pak method. The
Chemical Standards. Standard triazine
water was 250 mL. results of this experiment indicated that
herbicides and metabolites were
Recovery and detection limit samples the detection limit for these compounds
obtained from AccuStandard Inc.
were first adjusted to pH 7 by addition is lower than 100 ng/mL.
(New Haven CT). The analytes included
in this study were the herbicides of 2 mL of 10-3 M phosphate buffer. Best Solvent for Sample Elution. Initial
atrazine, propazine, cyanazine, and The Porapak RDX Sep-Pak cartridges experiments were performed using pure
simazine, and the metabolites desethy- were then preconditioned with 15 mL acetonitrile to elute the analytes from
latrazine, desisopropylatrazine and of methanol followed by 30 mL of the Sep-Pak cartridges. However,
hydroxyatrazine. MilliQ water. The buffered samples inconsistent recovery of hydroxyatrazine
were then processed through the Sep- led to an investigation of various sol-
Sources of Water. Detection limit studies
Pak cartridges at approximately 10 mL vents and mixtures to yield consistent
were conducted on high purity reagent
per minute. Following sample enrich- and high recovery of all compounds.
water obtained from a MilliQ ® system
ment, the cartridges were air dried for It was observed that the addition of a
(Millipore). Natural pond water was
10 minutes before elution with 5 mL of few drops of phosphoric acid to the
obtained from a small manmade lake in
1:1 acetonitrile/methanol. The sample acetonitrile gave high recovery of the
central Massachusetts. Finished drinking
extracts were then diluted 1:1 with hydroxyatrazine, but this procedure
water was obtained at Waters
MilliQ water before analysis by HPLC. resulted in an extract which required
Corporation, Milford, MA.
pH adjustment before HPLC analysis.
HPLC Conditions. All analyses were An alternative elution solvent, 1:1
Results and Discussion
performed on a Waters HPLC system acetonitrile/methanol gave virtually
comprised of a 600 Multi-solvent Effect of pH. Two separate experiments the same recovery as the acidified elu-
Delivery System, a 486 Tunable were performed which showed little or ent and required no further adjustments
Absorbance Detector, a 712 WISP ™ no relation between recovery and pH of the extract for chromatographic
Autosampler, and a 616 LC system. in the range from 2 to 8.5. Preliminary analysis.
The Millenium® 2010 Chromatography experiments have shown that the
Porapak RDX cartridge is highly effective Recovery from Pond Water. Four
Manager was used for system control,
for extraction of phenolics at pH 2. replicate 250 mL samples were pre-
data acquisition, and data analysis.
Therefore, the Porapak RDX cartridge pared from the Lake Boon water
The analytical column was a Waters
may be used for the simultaneous source. This was a typical surface
Symmetry C18, 5µm 3.9 mm x 150 mm.
extraction of triazines and metabolites water with considerable particulate
The aqueous portion of the mobile
along with acidic contaminants such as matter, biota, and a noticeable green-
phase (Solvent A) was a pH 4.7 buffer
phenol. ish brown hue of dissolved and sus-
prepared from ammonium acetate and
pended natural organic matter. The
acetic acid, each at 1 mM concentra- Detection Limit. According to a proce- results of this experiment are presented
tion in MilliQ water, and adjusted to dure recommended by the EPA (5), the in Table 1. These results confirm that
5% methanol. The organic portion of detection limit for an analyte may be triazine herbicides and their principle
the mobile phase (Solvent B) was 5% estimated from the magnitude of the metabolites, including hydroxyatrazine,
methanol in acetonitrile. All solvents standard deviation in the results are well recovered from a difficult
were obtained from J.T. Baker obtained from seven or more replicate sample matrix.
(Philipsburg N.J.) and were HPLC spiked samples. In this experiment,
grade; all reagents were also from seven identical 1 L samples were
Baker and were ACS grade. spiked at a level of 0.400 g/L of all
The injection volume was 50 L and
the detector was set to 223 nm for all
analytes with the exception of hydroxy- Table 1 Recovery of Triazine Herbicides and Metabolites from a Pond Water Source
atrazine, which was monitored at (spike level 2 g/mL, 5 replicates)
240 nm.
Compound % Recovery % RSD
Solid Phase Extraction with Porapak RDX
Cartridges. The conditions chosen for desisopropylatrazine 106 2.8
the extraction of triazine herbicides and
hydroxyatrazine 110 11
metabolites were based on the proce-
desethylatrazine 100 4.3
dure developed for analysis of nitroaro-
simazine 106 4.4
matic and nitramine explosives (4). All
cartridge conditioning and sample cyanazine 105 2.5
enrichment was performed at 10 mL per atrazine 107 5.4
minute. A Waters SPE manifold was propazine 103 1.1
used for processing up to 12 cartridges
at a time. The sample size for finished
17
Figure 2 shows a typical HPLC chro- Figure 2: HPLC Analysis of Spiked Pond water Processed on Porapak RDX Sep-Pak Cartridge
matogram obtained from the spiked
pond water experiments. Column: Symmetry C18, 5µm 3.9 x 150 mm
Recovery from a Municipal Drinking Mobile Phase: A) 5% methanol in acetonitrile
Gradient: 100% A for 0.5 min step to 20% B,
Water. Five replicate 1.0 L samples hold for 5 min, then linear gradient to 75% B, for 25 min
were prepared at a spike level of 0.4 0.040 Flow Rate: 1.2 mL/min
g/mL using Milford, MA tap water. Injection Volume: 50 µL
0.035 Detection: 223 nm (240 nm for hydroxyatrazine)
Before spiking, each sample was treat-
Sample: 250 mL of pond water spiked
ed with 30mg of sodium sulphite to
11.858 deethylatrazine
The high RSD value for recovery of 0.020
9.575 hydroxyatrazine
16.392 simazine
17.158 cyanazine
desisopropylatrazine resulted from
23.142 propazine
19.908 atrazine
0.015
one data point out of range. If that
data point is eliminated, the result for 0.010
recovery of that analyte is 101% with
18.867
20.400
an RSD of 6.5%. 0.005
18
Ordering Information
Porapak RDX
19
Measurement of Trace Formaldehyde in
Workplace Air using Waters XPoSure
Aldehyde Sampler Cartridges
Donna M. Martin and Pamela C. Iraneta
A Short Term Exposure Limit (STEL) difficulty is due to the short sampling • The method needs to quantify
method for the analysis of formaldehyde time of the method and the background formaldehyde in the range of 0.02
in workplace air was developed using levels of the derivatisers. In the STEL to 2 ppm under low and high
Waters XPoSure™ Aldehyde Sampler method, air samples are taken for 15 humidity conditions.
Cartridges. The method involves collect- minutes with portable sampling pumps • The method needs to meet the
ing air onto a cartridge that contains a ideally placed in the breathing zone. National Institute for Occupational
solid phase derivatising agent specific The Waters team discussed STEL mea- Safety and Health accuracy
for aldehydes and ketones, eluting the surements for formaldehyde in work- criterion.
sample and analysing it by HPLC. The place air with several experts in industri-
high flow rate capability and guaran- A cutaway view of the XPoSure
al hygiene and occupational safety. Aldehyde Sampler cartridge is shown
teed low backgrounds allow for the The following is a list of performance
quantitation of formaldehyde down to in Figure 1. It has male and female luer
criteria for the ideal sampling device: connectors for easy attachment to
the 0.02 ppm range. Results from
method validation studies demonstrate • The solid phase derivatisation pumps and is bi-directional with respect
that XPoSure meets the National Institute reaction must be fast. to sample collection and desorption.
for Occupational Safety and Health • The device needs to be compatible The solid phase is 2,4-dinitrophenylhy-
(NIOSH) accuracy criterion. with existing portable sampling drazine (DNPH) coated silica which,
pumps used in workplace under acidic conditions, reacts rapidly
monitoring. with aldehydes and ketones to form its
Introduction
Formaldehyde is receiving increased • The device needs to accommodate
attention as a toxic substance and an a large sample throughput to pre
occupational hazard. It is widely used concentrate formaldehyde on the
as a textile treatment agent, as an inter- cartridge.
mediate in the manufacture of urea-
formaldehyde, and in making phenol- Figure 1: Cutaway View of an XPoSure Aldehyde Sampler Cartridge.
formaldehyde resins which are often
used in manufacturing fiberboard, parti-
cleboard and plywood.
The Occupational Safety and Health
Administration (OSHA) reduced the 8
Female luer connector
hour permissible exposure limit (PEL)
for formaldehyde from 1 ppm to 0.75
ppm and the American Conference of Polyethylene frit 100 µm pore size
Governmental Hygienists (ACGIH) has
adopted a new threshold limit value
(TLV) of 0.3 ppm.(1,2) This trend Acidified DNPH coated silica
500 -1000 µm particle size
toward lower exposure limits is expect-
ed to continue. In particular, Short Term
Exposure Limit (STEL) measurements for Polyethlene frit 100µm pore size
formaldehyde in the sub ppm range Aluminum compression ring
have been difficult to quantify with
certainty using older solid phase
derivatisation technologies such as 2- Male luer connector
(hydroxymethyl) piperidine (3). The
20
corresponding DNPH-hydrazone deriv-
ative (shown below). The DNPH-hydra- NO2 NO2
zone can then be eluted from the car-
R1 H+ R1
tridge with acetonitrile and analysed C=O H2NNH NO2 C = NNH NO2 + H2O
by HPLC. R R
Since 1991, Waters has marketed a Carbonyl Compound 2,4-Dinitrophenylhydrazine DNPH Derivative
DNPH-coated product that measures (Aldehyde or Ketone) (DNPH)
trace aldehydes and ketones in car
exhaust for the automotive industry.
XPoSure is similar in design, however,
a larger silica particle size and larger Figure 2: Chromatographic Separation used to Analyse Samples
frit pore size are used to obtain low
back pressure and high sample
Column: Symmetry™C18 5 µm 3.9 x 150 mm
throughput compatible with portable Mobile Phase: 45:55 acetonitrile/water
sampling pumps. Flow Rate: 1.3 mL/min
Detection: 360nm
Injection Volume: 20 µL
Experimental Calibration Standard Level 4
21
Figure 3: Schematics for Permeation Tube Systems used to Generate Atmospheres of Known Formaldehyde Concentrations
0.2 L/min
N2
Permeation Tube
1.3 L/min
0.2 L/min
N2
Permeation Tube
1.3 L/min
H2 O
22
Figure 4: Precision and Bias Data for 15 Minute STEL Measurements of Laboratory Generated Controlled Atmospheres Containing
Formaldehyde
Ordering Information
XPoSure
Reference
1. U.S. Department of Labor, Occupational Safety and Cincinnati, OH: NIOSH 1984. Method 2541-
Health Administration (OSHA): Code of Federal Regulations Formaldehyde.
Title 29 CRF 1919.1048 Formaldehyde, Government
Printing Office, Washington, D.C., July 1, 1993. 4. P. C. Iraneta, presentation at AICHE 1995, Kansas City,
pp. 410-442 May 1995
2. American Conference of Industrial Hygienists (ACGIH): 5. E.R. Kennedy, T.J. Fischbach, R. Song, P.M. Eller, and S.
1991 – 1992 Threshold Limit Values and Biological A. Shulman, “Guidelines for Air Sampling and Analytical
Exposure Indices. Cincinnati, OH: ACGIH, 1991. Method Development and Evaluation”. Cincinnati, OH:
NIOSH July 13, 1994. p.38.
3. National Institute for Occupational Safety and Health
Administration (OSHA): NIOSH Manual of Analytical
Methods, 3rd ed. (DHHS/NIOSH Pub. No. 84-100)
23
Literature Corner:
Clinical:
Andrew Clifford and his colleagues at UCal/Davis have developed an innovative
“Protocol for Collection and HPLC application for Waters Sep-Pak ® DNPH-Silica Cartridges. They collect normal human
Analysis of Volatile Carbonyl breath over a 15-20 minute period, draw a 25-litre aliquot of this sample through a
Compounds in Breath” DNPH-Silica Cartridge, elute the hydrazone derivatives of trapped aldehydes and
ketones with 3 mL of acetonitrile, and identify the compounds by HPLC analysis.
Y. Lin, S.R. Dueker, A.D. Jones, S.E. Ebeler, Recoveries were typically >90%, and CVs for the entire sampling/analysis protocol on
and A.J. Clifford, Clin. Chem. 41(7) duplicate samples were <11%. This new protocol is noninvasive and will enable many
(1995), 1028. in vivo studies on the links between levels of particular carbonyl compounds in human
breath and various metabolic or pathological states, for example, acetone (diabetes
mellitus), acetaldehyde (alcohol consumption.)
Patrick McDonald
Industrial/Environmental:
Peter Jackson and collaborators have demonstrated a variety of ways in which HPLC
“Studies on System Performance and systems can be configured with combinations of hardware, suppressor, and column
Sensitivity in Ion Chromatography” components for anion analysis. Dual-piston, reciprocating pumps fitted with low-pressure
pulse dampeners minimised baseline noise seen with conductivity detection, whether
P.E. Jackson, J.P. Romano, and B.J. or not post-column suppression was used. Column temperature control was critical to
Wildman, J. Chromatogr. A 706(1+2) successful sub-ppb detection limits in IC without suppression. An optimum system for
(1995), 3. selectivity as well as detection limits was the Waters IC-Pak™ Anion HR column (4.6 x
75 mm) with a Waters IC-Pak Cation Guard Column (4.6 x 50 mm) used as a
suppressor. Though solid-phase suppressors have a finite lifetime, before regeneration is
required, high-capacity, well-packed suppressor columns like the IC-Pak C column
demonstrate an optimum combination of sensitivity, linearity, and analytical efficiency.
Patrick McDonald
fl fl fl fl fl fl fl fl fl fl fl fl fl fl fl fl fl fl fl fl fl fl fl fl fl fl fl fl fl fl fl fl fl fl fl fl
On-line and off-line solid-phase (SPE) extraction methods were developed for
“Degradation of Organophosphorous determination of organophosphorous insecticides and metabolites in natural waters.
Pesticides and their Transformation The methods were used to investigate the degradation rates of these materials in the
Products in Estuarine Waters” environment. Following SPE on C18 silica, separation and detection was performed
using a number of techniques including HPLC coupled with photodiode array detection
S. LaCorte, S. Lartiges, P. Garrigues, and (PDA) and thermospray mass spectrometry (TSMS). The authors observed that the on-line
D. Barcelo, Environmental Science & HPLC based methods employing sensitive detectors such as the Waters 996 Photodiode
Technology 29, no.2 (1995), pp. 431-438 Array Detector were the most sensitive of the techniques investigated, allowing detection
limits of 0.1 g/L using only 150 mL of sample. They also obtained these detection limits
for analytes which are too thermally labile or polar for traditional GC based analysis.
Based on their analyses, the authors have concluded that many of the insecticides are
more persistent in the environment than previously thought.
Michael S. Young
Biochemistry:
Purification of milligram amounts of RNA is necessary for biochemical and biophysical
“Synthesis and Purification of Milligram studies, such as x-ray crystallography and NMR. The authors describe a one-column
Quantities of Short RNA Transcripts” method that uses a 90 min step gradient to separate runoff RNA transcript molecules
that differ by one nucleotide (nt) out of 69 nt. The isolated material is pure enough to
R. Kim, E. L. Holbrook, J. Jancarik and S.
be used for the biophysical studies. A double stranded DNA template made by
Kim, BioTechniques 18(6) (1995), 992 polymerase chain reaction (PCR) was used to synthesise the RNA chains by in vitro
transcription. The RNA was precipitated, solubilised and then fractionated on a
DEAE-5PW ion-exchange column (Waters Protein-Pak™ DEAE-5PW) on the Waters
650 Advanced Protein Purification system. The 70 nt was separated from the 69 nt.
All unincorporated nucleotides elute early in the low salt buffer and any aggregates
elute in the high salt buffer. The authors also state that the method has been used in
their laboratory to transcribe and purify several different RNA molecules ranging in
size from 32-to 69-nt long.
Dorothy J. Phillips
24
Pharmaceutical:
Renin inhibitors are potentially better for controlling hypertension than the currently used
“Determination of the Renin Inhibitor
angiotensin converting enzyme (ACE) inhibitors. Renin inhibitors have the potential to
Ro 42-5892 in Human Plasma by produce fewer clinical side effects; therefore, the search for them has become an active
Automated Precolumn Derivitisation, research area. Ro 42-5892 has been identified as a very potent renin inhibitor in man
Reversed-Phase High Performance and required a sensitive assay method for the evaluation of its pharmacokinetics. The
Liquid Chromatographic Separation authors report a very sensitive and selective HPLC assay for analysis of Ro 42-5892
and Electrochemical Detection after in human plasma. The Ro 42-5892 was extracted from plasma with dichloromethane,
Post-column Irradiation” derivatised with 2,4-dinitrofiuorobenzene, chromatographed and then detected by elec-
J. Leube and G. Fisher, J. Chromatogr. B trochemical oxidation after UV irradiation. Identifying the stationary phase that gave the
665 (1995), 373
desired selectivity was a challenge. Fifteen different silica-based phases (3 to 10 µm
particle size, different chemistries) and three other phases were tested in 24 different
column formats. Only Waters Nova-Pak® C18 4 µm 3.9 mm x 150 mm column gave
the required selectivity between interfering peaks and the analyte. The mobile phase
contained pH 7 buffered acetic acid with disodium-EDTA and acetonitrile (100:85).
The combination of the excellent selectivity on the Waters Nova-Pak C18 column and
the electrochemical detection achieved a limit of quantitation of 0.3 ng/mL.
Quantitative recovery of Ro 42-5892 of 98% and 107% were noted at 0.5 and 3
ng/mL concentrations, respectively. The assay was used successfully for over 250
samples from a pharmacokinetic study in hypertensive patients.
Dorothy J. Phillips
fl fl fl fl fl fl fl fl fl fl fl fl fl fl fl fl fl fl fl fl fl fl fl fl fl fl fl fl fl fl fl fl fl fl fl fl
Researchers at McGill University and Lady Davis Institute of Medical Research have
“Simultaneous microassay of alfentanil,
developed a rapid HPLC assay that simultaneously quantitates alfentanil, fentanil
fentanil and sufentanil by high-perfor- and sufentanil in plasma and urine samples of newborn babies. A Nova-Pak CN 5µm
mance liquid chromatography” cartridge, 8 x 100 mm was used along with a guard column of the same packing.
R. Bansal and J.V. Aranda, J. Liq. Chrom. Detection limits as well as linearity range were estimated. Due to the speed and
18 (1995), 339. sensitivity of the assay (less than 10 minutes), it can be easily used for therapeutic
drug monitoring of these drugs.
Zoubair El Fallah
Food:
There is concern that organophosphorous pesticides, such as methyl parathion, may
“Rapid Determination of Methyl contaminate dairy animals through ingestion of contaminated grasses or by percuta-
Parathion and Methyl Paraoxon in Milk neous absorption. A rapid solid phase extraction (SPE) procedure has been developed,
by Gas Chromatography with Solid- utilising Waters Sep-Pak C18 cartridges, for the isolation and concentration of methyl
Phase Extraction and Flame parathion and methyl paraoxon residues from milk prior to GC analysis. When com-
Photometric Detection” pared with the traditional liquid-liquid extraction (LLE) procedure, the SPE procedure
gave similar detection limits, fewer interferences, and used far less solvent. A typical
R. Baynes and J. Bowen, Journal of
sample could be prepared by SPE in 75 minutes compared with over 3 hours for
AOAC International, 78 no 3 (1995),
the LLE procedure.
pp. 812-815 Michael S. Young
General Chemistry:
Because of its non destructive nature and ease of calibration, researchers at Rice
“ Determination of C60 and C70 University used HPLC with photodiode array detection to quantitate fullerenes in
Fullerenes in geologic materials geologic materials. A Waters Nova-Pak C18, 4 µm 3.9 x 300 mm column was used.
by high performance liquid The mobile phase was 40/60, toluene/methanol and quantitation was performed
chromatography” using a Waters 996 Photodiode Array Detector at 330 nm. Limits of detection of 5
and 20 ng were estimated for C60 and C70, respectively. Due to the high organics
D. Heymann, L.P. Felipe Chibante and R.E. content in real samples, a preparative step was necessary. It was performed on a Prep
Smalley, J. Chromatogr. A 689 (1995), Nova-Pak C18 6 µm 19 mm x 300 mm column. Collected fractions of C60 and C70
157. were then reinjected on the analytical column. Recoveries on the order of 90% were
obtained. Due to possible photochemical decomposition of early eluting organics,
which could contaminate the detector cell, the authors suggest running the preparative
step with the PDA turned off.
Zoubair El Fallah
25
Conversations with Chemists
surfactant structure on the separation of
enantiomeric mixtures by CE. Over
eighty proprietary chiral surfactants
were synthesised by me, and their
propensity to separate enantiomers via
a MEKC format were tested by Jeff
Mazzeo and Mike Swartz. It has been
exciting to work on the separation of
Jeff Mazzeo Michael Swartz Ed Grover enantiomeric mixtures by CE. Not only
is the study of chirality very relevant to
industry and the consumer marketplace,
What role did you play in the it also provides an intriguing academic
development of EnantioSelect? challenge.
In the last issue of the Waters Column, Jeff Mazzeo: My initial role on the What is the potential impact of this
chiral CE team was to evaluate the technology?
we published an article on the
ability of novel chiral surfactants pre- Ed Grover: If we are ultimately able to
"Separation of Enantiomers pared by Ed Grover to separate enan- develop a complementary set of chiral
by Micellar Electrokinetic Capillary tiomers via MEKC. Since many of the surfactants (five to six surfactants) which
Chromatography with Novel Chiral chiral compounds which we are inter- can separate most of the chiral com-
Surfactants.” The response was over- ested in separating are charged, I pounds in existence today, we believe
developed equations to calculate the that we would greatly simplify the ana-
whelming, so we decided to follow-up
capacity factors and alpha values of lytical chemist's approach to separating
the article by interviewing the charged compounds, and also exam- enantiomeric mixtures.
Waters' scientists who developed ined the effect of migration window on
EnantioSelect™. EnantioSelect reagents resolution for both neutral and charged Michael Swartz: Although the true
guarantee exact enantiomer migration compounds. During initial evaluation, I impact is yet to be seen, I think that we
determined that optimum partitioning have brought a new way of thinking
order reversal for improved quantita-
was just as important as enantioselec- about the assay of chiral compounds to
tion and peak identification. tivity in determining resolution. the scientific community. While compar-
Specifically, partitioning of anionic isons to other additives will be drawn,
analytes was low due to charge repul- the advantages of our approach--simul-
sion. To overcome this problem, work taneous achiral and chiral compound
would have to be done at a pH below analysis, peak reversals for identifica-
the pKa of the acidic group in the ana- tion and quantitation, and simplified
lyte structure, which led Ed to synthe- method development--should not be
size anionic surfactants which main- underestimated.
tained full ionisation at acidic pH’s.
Jeff Mazzeo: The unique ability to
Michael Swartz: I first investigated the exactly reverse enantiomer migration
separation of chiral compounds using order will improve quantitation of enan-
CE in 1989 using cyclodextrins as tiomeric separation and permit confir-
If you would like to receive the last buffer additives. These early efforts mation of chiral separation. This benefit
issue of the Waters Column, please were frustrated by a lack of selectivity is clearly seen in comparison to hetero-
check box 11 on the Business Reply and applicability--every new com- geneous chemistry such as chiral LC
Card and return today. pound required an intense application phases, as well as natural product CE
effort that was too often unsuccessful. additives such as derivatised cyclodex-
Upon the realisation that synthesised trins. I believe that once we generate a
chiral surfactants solved many of the family of chiral CE additives which will
frustrating problems associated with complement each other in terms of their
previously investigated additives, I've ability to separate chiral compounds,
focused my energies working on creat- and will share the benefits described
ing method validation protocols for above, enantiophoresis technology will
EnantioSelect. revolutionise the way chiral separations
are performed.
Ed Grover: During the research phase,
I was responsible for the systematic
design and synthesis of chiral surfac-
tants to understand the influence of
26
What’s new
Welcome to Waters on the World Wide Web
Visit us on the internet at http://www.waters.com/
27
How to reach us
Sales Offices Mexico Waters S.A. DE C.V.
Austria and European Export (Central Europe, Tel.: 525 359 0121
CIS, India and India Subcontinent) Fax: 525 359 0394
Waters Ges.m.b.H. The Netherlands Waters Chromatography B.V.
Tel.: 1 8771807 Tel.: 7650 87200
Fax: 1 8771808 Fax: 7650 87280
Australia Waters Australia Pty. Limited Norway Waters AS
Tel.: 2 933 1777 Tel.: 2 219 5360
Fax: 2 898 1455 Fax: 2 219 5362
Belgium and Luxembourg Waters S.A. - N.V. Peoples Republic of China - Beijing
Tel.: 02 7261000 Waters China Ltd.
Fax: 02 7261100 Tel.: 10 592 8918
Brazil Waters Comercial LTDA Fax: 10 506 0344
Tel.: 011 55 11 543 7788 Poland Waters Sp.zo.o
Fax: 011 55 11 530 6413 Tel.: 48 2 669 1225
Canada Waters Limited Fax: 48 2 663 7033
Tel.: 800 252 4752 Puerto Rico Waters
Fax: 905 678 9237 Tel.: 809 747 8445, 790 2225
CIS Waters Representation Office, Moscow Fax: 809 747 8448
Tel.: 7 095 336 7000 Singapore Waters Asia Ltd. (Asia Headquarters)
Fax: 7 095 336 7000 Tel.: 225 4855
Czech Republic Fax: 225 4655
Waters Ges.m.b.H. Spain Waters Cromatografia S.A.
Tel.: 42 2 35 02 27 Tel.: 3 325 9616
Fax: 42 2 35 23 75 Fax: 3 325 9896
Denmark Waters A/S Sweden Waters Sverige AB
Tel.: 46 598080 Tel.: 08 6282440
Fax: 46 598585 Fax: 08 6282455
Finland Waters Oy Switzerland Waters AG
Tel.: 0 506 4140 Tel.: 62 889 2030
Fax: 0 5064 1411 Fax: 62 889 2059
France Waters S.A. Taiwan Waters Asia Ltd.
Tel.: 1 30487200 Tel.: 2 706 8766
Fax: 1 30487201 Fax: 2 706 9704
Germany Waters GmbH UK and Ireland Waters Ltd.
Tel.: 06196 40 06 00 Tel.: 1 923 816700
Fax: 06196 48 23 88 Fax: 1 923 219012
Hong Kong Waters China Ltd.
Tel.: 2964 1800 All other countries
Fax: 2549 8956 Waters Corporation
Hungary Waters kft 34 Maple Street
Tel.: 1 1620686 Milford, MA 01757 USA
Fax: 1 1853236 Tel.: 508 478 2000
India Waters Pvt. Ltd. Toll-free: 1 800 252 4752
Tel.: 80 3341944 Fax: 508 872 1990
Fax: 80 3340527
Italy Waters S.p.A.
Tel.: 02 2500565
Fax: 02 2501827
Japan Nihon Waters Ltd.
Tel.: 3 3471 7191
Fax: 3 3471 7116
Malaysia Waters Asia Ltd.
Tel.: 3 704 8600
Fax: 3 704 8599
28
Paraquat/Diquat
Objective:
The current EPA method for analysis of
aqueous samples for paraquat and diquat 360.00
is method 549.1. This method specifies a
340.00 Mobile phase: 70% methanol/30% water with
photodiode array detector (PDA) because the
320.00 PIC B-7, pH = 3.0
HPLC procedure chosen in the method does 308 nm
300.00
AU
not fully resolve the peaks. Paraquat is moni- Flow Rate: 1.0 mL/min
280.00
tored at 257 nm, while diquat is monitored at Column: Symmetry C18, 5 µm 3.9 mm x 50 mm
260.00 257 nm
308 nm. The EPA method used a C18 column Temperature: 30 °C
240.00
and a mobile phase containing diethylamine
220.00 Detection: 257 nm and 308 nm
(DEA), to maintain good peak shape. The
objective was to provide an improved chro- Data System: Millennium 2010
matographic separation without the need of Paraquat
Chromatography Manager, Ver. 2.01
0.040
an amine modifier.
Details: 0.030
Diquat
®
A Symmetry C18 5µm column was used for
AU
0.020
this analysis because it does not require any
0.010
amine modifier to prevent tailing. Also, its
high level of retention for the analytes allowed 0.000
the use of a 50 mm column. The short col- 6.00 7.00 8.00 9.00 10.00 11.00
umn, along with the elevated temperature uti- Minutes
lized, allowed the analysis to be accom-
plished in under 15 minutes.
System:
A Waters 600 Multi-Solvent Delivery System with a column heater was
used with a 717 Plus Autosampler and a 996 Photodiode Array
Detector. System Control, data acquisition and reporting were
accomplished with the Millenium 2010 Chromatography Manager,
Version 2.01.
Waters Corporation 34 Maple Street, Milford, Massachusetts 01757-3696 U.S.A. 508 478-2000
Millennium 2010 Chromatography Manager Waters 996 Photodiode Array Detector
Advanced computer technology simplifies accessing, Complete Spectral Data-Complete Confidence • You can centralize operations through single-keyboard
interpreting, and reporting your results. in Peak Purity and Identity. system programming with either the new Waters 600E
controller or with an integral, on line PC.
The productivity of today’s chromatography is centered With Waters 996 Photodiode Array Detector, you can be
confident that if there is an impurity in your sample, you will • For comprehensive programming of all pump,
around the ability to find, generate, summarise, and report
find it; that your compound is really what you report it to be autosampler, and detector parameters directly from the
results in a way that meets your needs and complies with
and not something else. Every spectral detail of your sample controller keypad, the PowerLine™ software embedded
any regulatory requirements. The Millenium relational data-
is rapidly, easily, and accurately detected. Every peak is in the 600 controller allows you to store up to 15
base lets you create your own view filters to customise the
analysed with mathematical precision so you can be sure to completed instrument methods for easy method retrieval.
way you sort your information, such as “lotCode”.
asses the purity of peaks and confirm their identity. And you • By linking your 600E to the Millenium 2010
• Using the database, you can track a method history will never have to trade off resolution against sensitivity. Chromatography Manager, you can take advantage of
giving you audit trail capability that ensures total security
• The 996 PDA Detector lets you look at your results in a complete system control and information management.
of your raw data and results. In addition, you can
specify “Access Type” which allows you to “lock” your variety of ways - even while a separation is under way.
methods for even greater security. • Absorbency measurement over the entire 190-800 nm
• The database keeps track of your instruments, triggering wavelength range.
an alarm when scheduled maintenance is required. The • The MaxPlot feature of the Millenium PDA Software
Millenium relational database lets you view the injection automatically detects UV/Vis absorbing compounds
information on any calibration point at any particular anywhere within the 190-800 nm range. It measures
point in time on a processing sequence. This gives you their absorbence maxima, and plots a peak for each
a detailed account of the processing method for that compound at its absorbence maximum, creating a
injection, ensuring an accurate audit trail for regulatory composite chromatogram of all of the compounds it has
compliance. detected.
• The Millenium Chromatography Manager can provide
acquisition and control for as many as four chromato-
graphic systems, including three HP5890 GC. Waters™ 600E Multisolvent Delivery System
• The Millenium Report Publisher has the capability to give A no-compromise approach to single pump
you exactly what you want. gradient systems
Objective:
As more and more drinking water production
facilities turn to ozonation as a disinfectant, Column: Symmetry C18 5µm 3.9 mm x 150 mm
there has been a concomitant need for repro-
ducible and sensitive analytical methods for Mobile Phase: A: 50% acetonitrile in water
determination of oxonation related disinfection B: Acetonitrile
byproducts (DPBs) in finished waters. Among
Gradient: 100% A for 4 min,
the most important of these DPBs are aldehydes
Then linear gradient to 100% B in 25 min.
and ketones. EPA method 554 is a recom-
mended procedure for determination of aldehy- Flow Rate: 1.2 mL/min
des and ketones as dinitrophenylhydrazine Detection: UV @ 360 nm (0.12 AUFS)
derivatives by HPLC following solid phase
extraction. Injection Volume: 20 µL
Details:
1: Formaldehyde-DNPH 7: Pentanal-DNPH
This chromatographic procedure has been 2: Acetaldehyde-DNPH 8: Hexanal-DNPH
developed to provide a superior separation of
3: Propanal-DNPH 9:Heptanal-DNPH
the 12 EPA method 554 analytes in under 30
minutes. 4: Crotonaldehyde-DNPH 10: Octanal-DNPH
System:
Waters Corporation 34 Maple Street, Milford, Massachusetts 01757-3696 U.S.A. 508 478-2000
Millennium 2010 Chromatography Manager The Waters 486 Tunable UV/VIS Absorbance
Detector
Advanced computer technology simplifies accessing,
interpreting, and reporting your results. The performance you expect with the flexibility you need.
The productivity of today’s chromatography is centered The Waters 486 Tunable UV/VIS Absorbance Detector is
around the ability to find, generate, summarise, and report the highest performing UV/VIS detector with the flexibility to
results in a way that meets your needs and complies with accommodate virtually all your HPLC applications. Our
any regulatory requirements. The Millenium relational data- exclusive Taper-Cell® flow cell design eliminates refractive
base lets you create your own view filters to customise the index effects, providing exceptional baseline stability.
way you sort your information, such as “lotCode”.
• The 486 detector is tunable from 190 to 600 nm
• Using the database, you can track a method history wavelength range with reproducibility and precision.
giving you audit trail capability that ensures total security
• The unique modular cell assemblies for microbore,
of your raw data and results. In addition, you can
analytical, preparative, high-pressure mass
specify “Access Type” which allows you to “lock” your
spectrometry, and nonmetallic flow paths give your
methods for even greater security.
application flexibility.
• The database keeps track of your instruments, triggering
an alarm when scheduled maintenance is required. The
Millenium relational database lets you view the injection
information on any calibration point at any particular
point in time on a processing sequence. This gives you
a detailed account of the processing method for that
injection, ensuring an accurate audit trail for regulatory
compliance.
• The Millenium Chromatography Manager can provide
acquisition and control for as many as four chromato-
graphic systems, including three HP5890 GC.
• The Millenium Report Publisher has the capability to give
you exactly what you want. For more information
For more information on these
products check boxes 14-18
on the Business Reply Card.
Determination of EPA Method 604 Phenols by HPLC
Objective:
Phenols are important industrial wastewater 9
contaminants which must be monitored for com- Column: Symmetry C8 5µm 3.9 mm x 150 mm
pliance with discharge regulations. Phenols, Mobile Phase: A: 1% acetic acid in water
particularly nitrophenols, often display inconsis- B: 1% acetic acid in acetonitrile
tent response when analysed by GC, even
5 Gradient: (T,%B)(0,30),(20,100)
when time consuming derivatisation steps are 4
employed. Flow Rate: 1.2 mL/min
Detection: UV @ 280 nm (0.08 AUFS)
Details:
2 Injection Volume: 40 µL
EPA Method 604 phenols are separated in
Concentration: 1.0 µg/mL each component
20 minutes using a Symmetry ® C8 5 µm
column. A Model 486 Variable Wavelength UV
Detector was used for this separation; for 6 1: Phenol 7: 4-Chloro-3-Methylphenol
3
greater sensitivity, a Model 460 pulsed 78 2: 4-Nitrophenol 8: 2,4-Dichlorophenol
electrochemical detector may by used. 1
10 3: 2-Chlorophenol 9: 4,6-Dinitro-2-Methylphenol
System:
A Waters™ 600 S Controller, 600 pump, 612
WISP Autosampler and a 486 detector was
used for this analyses. System control, data
acquisition and reporting were accomplished
using the Millennium 2010 Chromatography
Manager.
Waters Corporation 34 Maple Street, Milford, Massachusetts 01757-3696 U.S.A. 508 478-2000
Millennium 2010 Chromatography Manager The Waters 486 Tunable UV/VIS Absorbance
Detector
Advanced computer technology simplifies accessing,
interpreting, and reporting your results. The performance you expect with the flexibility you need.
The productivity of today’s chromatography is centered The Waters 486 Tunable UV/VIS Absorbance Detector is
around the ability to find, generate, summarise, and report the highest performing UV/VIS detector with the flexibility to
results in a way that meets your needs and complies with accommodate virtually all your HPLC applications. Our
any regulatory requirements. The Millenium relational data- exclusive Taper-Cell® flow cell design eliminates refractive
base lets you create your own view filters to customise the index effects, providing exceptional baseline stability.
way you sort your information, such as “lotCode”.
• The 486 detector is tunable from 190 to 600 nm
• Using the database, you can track a method history wavelength range with reproducibility and precision.
giving you audit trail capability that ensures total security
• The unique modular cell assemblies for microbore,
of your raw data and results. In addition, you can
analytical, preparative, high-pressure mass
specify “Access Type” which allows you to “lock” your
spectrometry, and nonmetallic flow paths give your
methods for even greater security.
application flexibility.
• The database keeps track of your instruments, triggering
an alarm when scheduled maintenance is required. The
Millenium relational database lets you view the injection
information on any calibration point at any particular
point in time on a processing sequence. This gives you
a detailed account of the processing method for that
injection, ensuring an accurate audit trail for regulatory
compliance.
• The Millenium Chromatography Manager can provide
acquisition and control for as many as four chromato-
graphic systems, including three HP5890 GC.
• The Millenium Report Publisher has the capability to give
you exactly what you want. For more information
For more information on these
products check boxes 14-18
on the Business Reply Card.