SERVICE MANUAL

3 - LOCATION & CONNECTOR PINS DESCRIPTION

3-4 HV CAPACITOR BOARD

Inhibition
Inhibition reference
J1 connector J2 connector
(14 pins) (4 pins)
J3 J4
J5

Connector Pin Signal


J1 1 ; 2 ; 13 ; 14 Power supply input (+24V)
3 Inhibition
4 1 pulse voltage
5 2 pulses voltage
6 Reference output following the pulses number
7 High voltage input
8 - 12 Power supply input (ground)
J2 1 Power supply input (+24V)
2 Power supply input (ground)
3 Reference high voltage (+)
4 Reference high voltage
5 Inhibition
J3 1 High voltage power supply
J4 1 High voltage ground
J5 1 High voltage power supply
J6 1 High voltage ground

Rev. 05/2005 P - 31
SERVICE MANUAL

3 - LOCATION & CONNECTOR PINS DESCRIPTION

3-5 HV POWER SUPPLY BOARD

S7

S5
S4
S3
S6
S2
S1

Connector Pin Signal


S1 1 Power supply input (+24V)
S2 1 Power supply input (ground )
S3 1 Ref+
S4 1 Ref
S5 1 Inhibit
S6 1 High voltage out
S7 1 High voltage (ground)

P - 32 Rev. 05/2005
CHAPTER 4

INSTALL
INSTALLATION
ALL ATION
SERVICE MANUAL

4 - INSTALLATION
a) Place the chinrest and screw it on the table with the two long six-point
socket screws.
b) Insert a metal shaft in the hole at the side of cross-slide table.
Position two pieces of gears at both ends of the metal shaft and fix it with
six-point socket screws, placing a groove between each gear and groove.

Cross slide table


Groove
Metal shaft Groove
Screw Gear Gear Screw

c) Place the slit lamp on the table.


Position the gears correctly on the rack of the table to avoid malfunction of
the cross-slide table.
Make certain that bottom sides (viewed from operator) of the
cross-slide table are parallel to the side-lines of white plate. In the
opposite case the table can not move.

S T OP
S T OP

Good position Bad position

d) Place the slit generator axis on the slit


lamp axis and screw it.

DANGER
CAUTI O N : FEDERAL USA L AW RESTRI CTS THI S DEVI CE TO SAL E BY O R O N THE
LASER RADI ATI O N
O RDER O F A PHYSI CI AN O R AN O PTO M ETRI ST L I CENCED BY THE L AW O F THE Avoi d eye or ski n exposure t o
STATE I N W HI CH HE PRACTI SES TO USE O R O RDER THE USE O F THI S DEVI CE di rect or scat t ered radi at i on
DANG ER : EXPLO SI O N HAZARD. DO NO T USE I N PRESENCE O F
FLAM M ABLE ANESTETI CS O R O XI G EN. RI CH ATM O SPHERE Nd : YAG 1064 nm 30mJ M ax( 3 p u l ses 4
CAUTI O N : ELECTRI C SHO CK HAZARD. DO NO T O PEN. REFER Di ode 637 nm 1 mW
SERVI CI NG TO Q UALI FI ED SERVI CE PERSO NNEL.
CLASS I V LASER PRO DUCT

P - 34 Rev. 05/2005
SERVICE MANUAL

4 - INSTALLATION

e) Place the both gear covers sliding it under the rail guides of the cross slide
table.

Gear cover Gear cover

STOP

Rail guide Rail guide

f) Connect the following cables and remote :


- Sub-D 25 cable - Power supply cable
- Slit Lamp cable - Joystick interlink cable or footswitch cable
- High voltage cable - Remote
- Ground cable

Slit lamp power High voltage


supply connector connector

Joystick or footswitch Remote Ground Main


connector connector connector switch

Joystick Slit Interlock H.V. Viridis


GND
Footswitch Lamp Voltage Interlock
0
I

GROUND

+ Useonly with a 250V fuse


Employer uniquement avec
un fusible de 250 V
+
SN :
51 02 02

Sub-D 25 pins Hole threaded Hole threaded


connector to screw knurled to screw knurled

g) Close the rear panel with the large screws passing the cables in the holes
on the rear panel.
Large screws to keep in place rear panel

REF : OPTIMIS II THIS PRODUCT


COMPLIES WITH 21 CFR 10 40
SN :
QUANTEL MEDICAL CAUTION :
21, rue Newton, Z.I. du Brézet,
0

U.S. F EDERAL LAW


I

63039 Clermont-Ferrand cédex 2


FRANCE RESTRICTS THIS
Tel : +33 (0) 473 745 745 - Fax : + 33 (0) 473 745 7 00 OR ON ORDER
OF A PHYSICIAN

! IP 20

+ Use only witha 250 V fuse


Employer uniquementavec +
MADE IN FRANCE un fusible de 250 V
AP LASER C LASS4/IV
G N d:YAG Las er : 1064nm,
0459
30mJ Max , (3 puls es x 4ns )
LASER C LASS2/II
100 - 230V~ D iode Las er : 637nm, <1mW Max
VISIBLE AND/OR INVISIBLE LASER RADIATION
50/60 Hz 4 AT AVOID E YE OR SK IN EXPOS URE TO
150 VA,1 Φ 250 V DIRECT ORSCATTERED RADIATION
CLASS IV LASER PRODUCT

Hole to pass the cables Hole to pass the cables

Rev. 05/2005 P - 35
SERVICE MANUAL

4 - INSTALLATION
i) Switch On the master switch.
Check the panic switch is not activated.
Turn the key to ON, the following screens will be displayed :

Running Selftest
Please Wait

Running Selftest
-----

Running Selftest
- - - - - - - - - - - - OK

0.5mJ 1 Pulse
Count 0

To check OPTIMIS II after installation, see chapter 8-1


“CHECKING AFTER INSTALLATION”.

P - 36 Rev. 05/2005
CHAPTER 5

SERVICE MODE
SERVICE MANUAL

5 - SERVICE MODE

5-1 ACCESSING SERVICE MODE


Accessing the SERVICE mode is only possible after opening the laser
console.

To access at the service mode :

- Insert jumper connector on the CPU board,


- Turn On the key : the laser is in service mode.

Sub-D connector to
connect a PC computer
(linkage RS232)

Place here the


jumper connector

5-2 SCREEN IN SERVICE MODE


When the OPTIMIS II is in service mode, the energy value is measured by
the cell of the energymeter board and displayed in real time on the
OPTIMIS II screen.

You can check if the measured energy is in the safety windows (+20%, -
20% and -50%).

Screen in service mode, when you pulse :

Number of pulses
Energy selectionned by Energy value read by
the energy knob the energy meter board,
!displayed in real time)

1.6mJ 1 E1.6mJ
Count 0

Counter value

P - 38 Rev. 05/2005
SERVICE MANUAL

5 - SERVICE MODE

5-3 DEFECTS AND SOLUTIONS LIST


Message Explanation and solutions
Remove Slit Lamp The slit generator is in front of the treatment beam.

To pulse, the generator must be on a side.


When you remove generator, a Bip is emitted and you
can pulse.

Energy > 3.0mJ When you turn ON the key, the selected energy must be
inferior to 5.0 mJ if you want the laser start (safety).
- Adjust energy at a lower value.

Error 10 Shutter error.


- Check the shutter position.
- Check the shutter control.
- Check / change the shutter.

Error 11 A too LOW power is measured by the energymeter:


below (-50 %) the expected one
Service :
- Select the service mode.
- Check the displayed energy and real energy with a
joulemeter.
- Adjust the potentiometers of the energymeter board
(calibration).
- Change and adjust the energymeter board.
- Adjust the cavity of the laser head.

Error 12 A too HIGH power is measured : over (+20 %) the expected


one
Service :
- Select the service mode.
- Check the displayed energy and real energy with a
joulemeter.
- Adjust the potentiometers of the energymeter board.
- Change and adjust the energymeter board.
- Adjust the cavity of the laser head.

Error 13 High voltage error.


Service :
- Measure the high voltage values and re-adjust it with the
3 potentiometers. Call the Quantel Medical Technical
Department to know the different values of voltage for 1,
2 and 3 pulses.
- Change the high voltage capacitor board.
- Change the high voltage power supply.

Error 14 Transmission decoder error.


- Call for service.

No error message but : Key is activated and screen is turned OFF.

The remote circuit is open or is not properly closed (door or other safety device).
- Close the remote circuit (door or other safety device : Yag 532nm delivery
system in 532nm position).
Check the remote control connection.

Rev. 05/2005 P - 39
SERVICE MANUAL

5 - SERVICE MODE

5-4 QUIT THE SERVICE MODE

PROCEDURE :
1 - Remove the jumper connector of the CPU board.

2 - Select 0.5mJ with the energy knob.

3 - Turn the key to ON.

4 - The laser will pass automatically in normal mode.

CAUTION : if you forget to select 0.5mJ with the energy knob, the laser will stay in
service mode : - the measured energy will display on the screen,
- the energy safeties will not activated.

P - 40 Rev. 05/2005