Versaflux System: Operating Instructions
Versaflux System: Operating Instructions
Operating Instructions
Original instructions
Page intentionally left blank
Table of Contents
Table of Contents
1 Introduction .......................................................................................................... 5
1.1 Important user information ............................................................................................................. 6
1.2 Associated documentation .............................................................................................................. 9
4 Installation ............................................................................................................ 94
4.1 Site requirements .................................................................................................................................. 95
4.2 Transport .................................................................................................................................................. 97
4.3 Unpack VERSAflux ................................................................................................................................ 99
4.4 Setup ........................................................................................................................................................... 101
4.4.1 Assembly ............................................................................................................................................ 102
4.4.2 Process component connections ............................................................................................ 104
4.4.3 Power supply ................................................................................................................................... 106
1 Introduction
In this chapter
This chapter contains the following sections:
All users must read the entire Operating Instructions before installing, operating or
maintaining VERSAflux.
Always keep the Operating Instructions at hand when operating VERSAflux.
Do not operate VERSAflux in any other way than described in the user documentation.
If you do, you may be exposed to hazards that can lead to personal injury and you may
cause damage to the equipment.
WARNING
Do not operate VERSAflux in any other way than described in the
Operating Instructions.
Prerequisites
In order to operate VERSAflux safely, and according to the intended purpose, the following
prerequisites must be met:
• You should be acquainted with the use of bioprocessing equipment and with the
handling of biological materials.
• The system must be installed according to the instructions in Chapter 4 Installation,
on page 94.
Safety notices
This user documentation contains WARNINGS, CAUTIONS and NOTICES concerning the
safe use of the product. See definitions below.
Warnings
WARNING
WARNING indicates a hazardous situation which, if not avoided,
could result in death or serious injury. It is important not to proceed
until all stated conditions are met and clearly understood.
Cautions
CAUTION
CAUTION indicates a hazardous situation which, if not avoided,
could result in minor or moderate injury. It is important not to pro-
ceed until all stated conditions are met and clearly understood.
Notices
NOTICE
NOTICE indicates instructions that must be followed to avoid
damage to the product or other equipment.
Typographical conventions
Software items are identified in the text by bold italic text. A colon separates menu levels,
thus File:Open refers to the Open command in the File menu.
Hardware items are identified in the text by bold text (for example, Power).
System-specific documentation
In addition to the Operating Instructions, the documentation package supplied with
VERSAflux also includes detailed specifications and traceability documents.
The most important documents in the documentation package with regard to technical
aspects of VERSAflux are:
2 Safety instructions
Important
WARNING
Before installing, operating or maintaining VERSAflux, all users
must read and understand the entire contents of this chapter
to become aware of the hazards involved.
Failure to do this may cause human injury or death, or damage to
the equipment.
In this chapter
This chapter contains the following sections:
2.2 Labels 20
Introduction
The safety precautions in this section are grouped in the following categories:
• General precautions, on page 11
• Flammable liquids and explosive environment, on page 12
• Personal protection, on page 13
• Installing and moving, on page 14
• Power supply, on page 16
• System operation, on page 16
• Maintenance, on page 18
Note: Some of the safety precautions in this chapter may concern components or
situations described in other VERSAflux product documents.
General precautions
WARNING
Perform a risk assessment for any risks due to the process or pro-
cess environment. Evaluate the effects the use of VERSAflux and
the operational processes may have on the classification of the
hazardous area. The process might cause the area to increase or
the zone classification to change. Implement the risk reduction
measures needed, including use of personal protection equipment.
WARNING
Make sure that all installation, maintenance, operation and inspec-
tion is carried out by qualified personnel who are adequately
trained, understand and adhere to local regulations and the oper-
ating instructions, and have a thorough knowledge of VERSAflux
and the entire process.
WARNING
VERSAflux is designed for fixed installation. No main power supply
cable plug is provided. Connect the system directly to the electrical
cabinet or use other type of fixed installation.
WARNING
Do not operate VERSAflux in any other way than described in the
Operating Instructions.
WARNING
Protective earth. VERSAflux must always be connected to
protective earth when energized.
WARNING
Do not use VERSAflux if it is not working properly, or if it has suffered
any damage, for example:
• damage to the power cord or its plug
• damage caused by dropping the equipment
• damage caused by splashing liquid onto it
Personal protection
WARNING
When using hazardous chemical and biological agents, take all
suitable protective measures, such as wearing protective glasses
and gloves resistant to the substances used. Follow local and/or
national regulations for safe operation and maintenance of
VERSAflux.
WARNING
Personal Protective Equipment (PPE). Whenever packing, unpack-
ing, transporting or moving VERSAflux, wear:
• Protective footwear, preferably with steel lining.
• Working gloves, protecting against sharp edges.
• Protective glasses.
WARNING
Only personnel authorized by GE Healthcare may open the cabinet
doors. There is high voltage inside the cabinet that can cause
human injury or death.
CAUTION
Do not touch the system while pumping fluid through the system
that has a temperature above the normal working temperature.
Do not touch the system until you are sure that this can be done
without risk and when all components in the system have reached
the normal working temperature range.
CAUTION
Avoid spillage and overflow. Make sure that the waste tubing is
inserted in an appropriate waste container and secured in place.
CAUTION
Use ear protection whenever working close to the system in oper-
ation.
WARNING
VERSAflux must be installed and prepared by GE Healthcare per-
sonnel or third party authorized by GE Healthcare.
WARNING
Move transport crates. Make sure that the forklift has capacity to
safely lift the crate weight. Make sure that the crate is properly
balanced so that it will not accidentally tip when moved.
WARNING
Heavy object. Because of the significant weight of VERSAflux, great
care must be taken not to cause squeezing or crushing injuries
during movement. At least two, but preferably three or more, per-
sons are recommended when moving the unit.
WARNING
Access to power switch and power cord. The power switch must
always be easy to access. The power cord must always be easy to
disconnect.
CAUTION
To prevent bacterial growth, VERSAflux may be partly filled with
denatured Ethanol (18% C2H5OH (Ethanol), 2% C3H7OH (Isopropanol)
and 80% H2O (water)) or 0.1 M NaOH (Sodium Hydroxide) at delivery.
The mixture is hazardous to humans if consumed. Flush out the
denatured Ethanol or Sodium Hydroxide before assembling, testing
or integrating VERSAflux into the intended process context.
CAUTION
The wheels of VERSAflux should be locked during normal use. The
wheels should be unlocked only when moving the unit.
CAUTION
Make sure that all tubing, hoses and cables are placed so that the
risk for tripping accidents is minimized.
CAUTION
VERSAflux is designed for indoor use only.
CAUTION
Do not use VERSAflux in a dusty atmosphere or close to spraying
water.
CAUTION
Make sure that correct air pressure is always maintained. Too high
or too low air pressure may be hazardous and may cause erroneous
results and leakage.
Power supply
WARNING
VERSAflux must always be connected to a grounded power outlet.
WARNING
National Codes and standards (NEC, VDE, BSI, IEC, UL etc.) and local
codes outline provisions for safely installing electrical equipment.
Installation must comply with specifications regarding wire types,
conductor sizes, plug, branch circuit protection and disconnect
devices. Failure to do so may result in personal injury and/or
equipment damage.
WARNING
All electrical installations must be performed by authorized
personnel only.
System operation
WARNING
Safe distance. Always maintain a safe distance from
VERSAflux during drainage or other activities that may involve
splashing.
WARNING
Before operation, all process connections and the piping system
must be tested for leakage at maximum pressure for continued
protection against injury risks due to fluid jets, burst pipes or poten-
tially explosive atmosphere.
WARNING
Never exceed the operating limits stated in this document and on
the system label. Operation of VERSAflux outside these limits may
damage equipment and bodily harm or death may occur.
WARNING
Emergency stop. Pressing the EMERGENCY STOP will not shut off
mains power to the cabinet.
WARNING
Power failure. During a power failure, or if the EMERGENCY STOP
button is pressed, VERSAflux may remain pressurized. Opening a
line or vessel at this point could result in the release of potentially
hazardous process or cleaning fluid, and cause bodily harm.
When recovering from a power failure or emergency shutdown,
make sure all lines and vessels are depressurized before opening.
WARNING
Alarm signals. All alarm signals must be set within the limits
specified in the system documentation. Pressure and temperature
control must be activated while the system is in use to prevent the
piping system to leak or break.
NOTICE
Only use chemicals that have been proven not to be harmful to
the wetted parts of the unit.
NOTICE
Make sure that large particles are prevented from entering the
VERSAflux system during operation, otherwise the filters may be-
come clogged and the feed pump may malfunction.
NOTICE
Do not start the rotary lobe pump before it is primed.
Maintenance
WARNING
The electric cabinet doors may only be opened when VERSAflux is
taken out of operation and subject to LOCK OUT / TAG OUT.
WARNING
Only personnel authorized by GE Healthcare may perform service,
installation, and maintenance of components inside the VERSAflux
cabinet.
WARNING
LOCK OUT / TAG OUT! Before any maintenance or decommission-
ing work is performed on VERSAflux, make sure that:
• it is empty and depressurized.
• it is disconnected from process feed, electrical power and
pneumatic supply.
• it is prevented from accidentally becoming re-energized during
maintenance.
• it is clearly tagged as taken out of operation.
• all process wetted areas are clean and decontaminated.
WARNING
For continued protection against injury risks due to fluid jets, burst
pipes or potentially explosive atmosphere, the piping system must
be tested for leakage at maximum pressure:
• After assembly or maintenance
• Before operation or CIP
CAUTION
Do not climb on any parts of VERSAflux except where clearly al-
lowed. Follow local regulations and make sure that equipment is
properly secured when inspecting VERSAflux at high level.
NOTICE
Although system filter type reconfiguration can be performed by
a properly skilled user, GE Healthcare recommends that a GE
Healthcare service engineer is assigned for this work.
Contact your GE Healthcare representative for more information
on changing filter configurations if required.
2.2 Labels
Introduction
This section describes the various labels on VERSAflux and their meaning.
Safety labels
The table below describes the various safety labels that may be found on VERSAflux.
Symbol/text Description
Introduction
This section describes how to perform an emergency shutdown of VERSAflux, and the
result in the event of power failure.
Precautions
WARNING
Emergency stop. Pressing the EMERGENCY STOP will not shut off
mains power to the cabinet.
WARNING
Power failure. During a power failure, or if the EMERGENCY STOP
button is pressed, VERSAflux may remain pressurized. Opening a
line or vessel at this point could result in the release of potentially
hazardous process or cleaning fluid, and cause bodily harm.
When recovering from a power failure or emergency shutdown,
make sure all lines and vessels are depressurized before opening.
B A
VERSAflux 400
Part Function
Emergency shutdown
Step Action Result
• Buzzer is activated.
• No data is lost.
2 If required, also switch off the • The entire system, including the HMI
mains power supply using the and PLC, becomes powerless.
MAIN SWITCH, power (B), dis-
connect the power cord or • Data and run status may be lost.
switch off the fixed power sup-
ply circuit breaker.
Power failure
The system power is lost if the MAIN SWITCH on the cabinet is turned off or the mains
cable is disconnected or the power supply is lost.
Pump stops if the electrical power to the system is lost. All valves will immediately revert
to default positions. Any data that has not been saved at that time may be lost.
Step Action
1 Make sure that the condition that caused the power failure or emergency
stop is corrected.
NOTICE
If the system does not restart as expected, contact your local GE
Healthcare representative for assistance.
Introduction
This section contains information about the decommissioning of VERSAflux.
Decontamination
VERSAflux shall be decontaminated before decommissioning and all local regulations
shall be followed with regard to scrapping of the equipment.
Recycling of hazardous
substances
VERSAflux contains hazardous substances.
Disposal of electrical
components
Waste electrical and electronic equipment must not be disposed as unsorted municipal
waste and must be collected separately. Please contact an authorized representative
of the manufacturer for information concerning the decommissioning of equipment.
Introduction
The following product pollution control information is provided according to SJ/T11364-
2006 Marking for Control of Pollution caused by Electronic Information Products.
根据SJ/T11364-2006《电子信息产品污染控制标识要求》特提供如下有关污染 控制
方面的信息
Label Meaning
Label Meaning
该标志表明本产品含有超过SJ/T11363-2006《电子信息产品中有毒
有害物质的限 量要求》中限量的有毒有害物质。标志中的数字为本
产品的环保使用期,表明本 产品在正常使用的条件下,有毒有害物
质不会发生外泄或突变,用户使用本产品 不会对环境造成严重污染
或对其人身、财产造成严重损害的期限。单位为年。
为保证所申明的环保使用期限,应按产品手册中所规定的环境条件
和方法进行正 常使用,并严格遵守产品维修手册中规定的期维修和
保养要求。
产品中的消耗件和某些零部件可能有其单独的环保使用期限标志,
并且其环保使 用期限有可能比整个产品本身的环保使用期限短。应
到期按产品维修程序更换那 些消耗件和零部件,以保证所申明的整
个产品的环保使用期限。
本产品在使用寿命结束时不可作为普通生活垃圾处理,应被单独收
集妥善处理
Value Meaning
• 此表所列数据为发布时所能获得的最佳信息
VERSAflux X 0 0 0 0 0
3 System description
In this chapter
This chapter contains the following sections:
3.1 Overview 32
3.1 Overview
General
These Operating Instructions cover the VERSAflux 120 and 400 systems.
VERSAflux is a series of cross flow filtration (CFF) systems that utilize WinCC™ Advanced
(TIA Portal) V11 software for semi automation, with data logging capabilities over the
entire cross flow process.
VERSAflux systems are intended for pilot to production scale biological separations.
The systems are configured to operate hollow fiber cartridges suited for microfiltration
applications such as cell clarification/harvesting, or hollow fibers for ultrafiltration appli-
cations, such as protein concentration and diafiltration in downstream unit operations.
Material compliance
All plastic and polymer materials that come in contact with buffers and samples are
compliant with FDA CFR 21 part 177, or USP Class VI.
Used materials are traceable back to their production batches.
You can find information about the design and materials used in your system in the
System-specific documentation, on page 9.
In this section
This section contains the following subsections:
8 1 2
7
3
6 4
Part Function
3 Indicator lamps
4 Electrical connectors
15
14
9
13
12
10
11
Part Function
12 Electric cabinet
13 System label
15 Connectors on top
7 8 9 10 11
1 2 3 4 5 6 12
Connector Function
3 PT-113
4 TT-161
5 FIT-142 Signal
6 FIT-142 Power
8 PT-111
9 PT-112
10 FIT-141 Signal
Connector Function
11 FIT-141 Power
12 USB connector
Top connectors
1 2 3
Connector Function
6 2
3
5
Part Function
3 Indicator lamps
11 12
10
Part Function
10 System label
11 Connectors on top
13
8 9 10 11 12
1 2 3 4 5 6 7
Connector Function
4 PT-113
5 TT-161
6 FIT-142 Signal
7 FIT-142 Power
9 PT-111
10 PT-112
Connector Function
11 FIT-141 Signal
12 FIT-141 Power
13 USB connector
Top connectors
1 2 3
Connector Function
In this section
This section contains the following subsections:
3.3.2 Automation 45
Skid
The rigid stainless steel structure supports and protects all process components and
the electric cabinet. The structure is designed for handling in a production environment
and is easy to move and keep clean.
The structure occupies a small box-shaped space that makes it easy to fit into any loca-
tion in the production facility.
Electric cabinet
The electric cabinet contains all electrical and pneumatic equipment. When the door is
closed, the entire cabinet is ingress protected according to IP54.
3.3.2 Automation
Control system
The VERSAflux System is automated by means of the WinCC Advanced (TIA Portal) V11C
control platform. Once the required control mode is selected and parameters entered,
a user can safely operate the system.
HMI
The HMI is built in to the cabinet and has high IP protection from the environment.
The HMI automatically starts when the system power is turned on, no specific starting
procedure is required for the HMI.
For HMI specifications, refer to the product documentation package.
User interface
The display and input devices are ergonomically designed for use in a clean production
environment.
Communications
All communication with controlled components mounted outside the cabinet uses the
HART™ industry standard communication protocol and hardware.
Description
A standard VERSAflux system is provided with the following process connections:
• Feed inlet
• Filter inlet, outlet and drain connections
• Retentate outlet
• Permeate outlet
• Pump outlet located before filters and meters.
Illustration
VERSAflux 120 VERSAflux 400
P1
P5
P2
B1
P4
P3
Description
The main purposes of the feed line is to transfer product (or feed) from the recirculation
tank, to the filter, FILTER and to raise the pressure of the feed so that correct trans
membrane pressure (TMP), filter pressure and feed flow is achieved. This is done by the
feed pump, P-201.
To protect the system against over pressure, one pressure sensor, PT-111 is installed
after the pump. If the system pressure rises over pre-set value, an alarm will be triggered
in HMI and the pump will stop, to make sure that the system is not subject to over pres-
sure. The feed line for VERSAflux 120 has a temperature sensor, TT-161 (optional), installed
before the pressure sensor.
The feed line can be drained through the HV-301 valve.
The internal diameter of the feed line is selected to keep the flow velocity of the product
within a favorable range with regard to the flow range for each system.
ME1
B1
F
P-201
PT-111
HV-301
TT-161
P3
ME1
B1
P-201
PT-111 HV-301
P3
Description
Particles / molecules too large to pass through the filter pores are retained by the filter
as retentate.
Retentate is led out from the filter (FILTER) through the retentate connection and then
guided to the retentate outlet.
A sensor for pressure, PT-112 is located on the retentate outlet line. For VERSAflux 400
System, a sensor for temperature, TT-161 (optional), is located on the retentate line.
TMP is controlled by the pressure control valve PCV-341 (optional) regulated by control
software, or is controlled by the manual diaphragm valve PCV-331 in alternative option.
The entire retentate line can be shut off by closing the retentate valve HV-302 (optional)
combined with PCV-341 (optional), or by closing PCV-331 in alternative option.
The retentate line is drained through the drain valve HV-306.
All retentate drain valves on VERSAflux 120 and 400 use uniformed tag, HV-303.
HV-303 P5 P1
F PT-112 FIT-141
HV-303 P5 PCV-331 P1
HV-303 P5 TT-161 P1
HV-303 P5 TT-161 P1
Description
Particles / molecules small enough to pass through the filter pores are collected as
permeate.
Permeate is led out from the filter through the permeate connection (FILTER) and guided
to the permeate outlet through valve HV-304.
Sensors for pressure, PT-113 (optional), and flow, FIT-142 (optional) are located on the
permeate outlet line. The temperature sensor, TT-161, for VERSAflux 120 is located on
the Feed line.
About options
The VERSAflux suite of filtration systems comes in a basic setup, as described in previous
sections.
Additionally, a selection of options are available to extend the functionality of the instru-
ment.
In this section
This section contains the following subsections:
Description
Dummy filters allow the system to operate with liquid but without filtering capability.
These can be either purpose made dummies or decommissioned filters.
Usage can be for tests and set-up where running the system is required but actual filtra-
tion is not.
Contact your local GE Healthcare representative for more information regarding dummy
filters.
Description
To connect the feed tank to a corresponding VERSAflux 120 or 400 system, a set of hoses
is required. Hose kits are available as an option from GE Healthcare. The hoses are made
of reinforced PT-cured silicone and are designed to meet the tank and system specifica-
tions.
Overview
When the VERSAflux systems are operated with hollow fiber cartridges, required compo-
nents of the piping system must be installed.
The retentate line must be in the upper position.
In this section
This section contains the following subsections:
3.5.3 HF cartridges 63
4 3 4 3 4
75 55 35
An optional spool piece kit is needed to mount size 55 and 35 hollow fibers to the system,
see feed line extension pipes in the next table.
Description of HF components
Part Function
1 Retentate line
4 3 4 3 4
85 65 45
An optional spool piece kit is needed to mount sizes 65 and 45 hollow fibers to the system,
see feed line extension pipes in the Description of HF components, on page 61.
Description of HF components
Part Function
1 Retentate line
3.5.3 HF cartridges
Illustration: HF cartridge
Description of HF cartridges
Hollow fiber (HF) cartridges contain numerous very thin hollow fibers glued together at
the ends.
Each fiber is open at the ends and the feed flows inside the fibers and out the other end.
The size of the pores determines the filtering characteristics of the cartridge, that is, the
cut-off size for particles that either will be retained or filtered out with the permeate.
The permeate flows outside the fibers and is collected through one of the side connectors.
HF cartridge draining
Pressure sensors
Three pressure sensors are provided. The pressure sensor, PT-111, provides pressure
data for the feed pump outlet and filter feed inlet pressure, PT-112, measures retentate
pressure and PT-113 (optional) measures permeate pressure.
The pressure drop (delta P) is calculated by the control software by subtracting the re-
tentate pressure from the feed pressure.
The transmembrane pressure (TMP) is calculated by the control software by subtracting
the permeate pressure from the average of the feed and retentate pressure.
The pump outlet pressure sensors also provide overpressure protection for the feed
pump and system.
Flow sensors
Two flow sensors are provided. A magnetic flow meter for the retentate, and a magnetic
flow meter for the permeate flow.
FIT-141 (optional) measures the retentate flow and FIT-142 (optional) measures the
permeate flow.
The feed flow is calculated by adding the permeate flow to the retentate flow.
Temperature
On VERSAflux 120 system, temperature is measured in the feed line by sensor TT-161
(optional).
On VERSAflux 400 system, temperature is measured in the retentate line by sensor
TT-161 (optional) .
In this section
This section contains the following subsections:
Introduction
This section describes the external connections between VERSAflux 120 and 400 and
the optional tank.
Connection diagram
P1
P5
P2
B1 ME1
P4
P3
Legend
Blue Outlets
Orange Inlets
Battery limits
"Battery limits" represent all interface points between the VERSAflux and the customer
process plant.
System battery limits require hose connections not provided by GE Healthcare.
Normally, the user provides hoses for these connections. Refer to Section9.1 Specifications,
on page 143 for details of hose and connection types and sizes.
P2
HV-304
P4
FIT-142
P F
PT-113
HV-305.3 HV-305.4
HV-305.1
FILTER
HV-306
HV-304.4
HV-304.3
HV-304.1
FILTER
HV-305
PT-111
P
HV-301
PT-112 TT-161
P
P3
P5
HV-303
FIT-141
F
P-201
PCV-341
ME1
PCV-331
HV-302
B1
P1
P2
HV-304
P4
FIT-142
P F
PT-113
HV-307.3 HV-307.4
HV-307.1
FILTER
HV-308
HV-306.3 HV-306.4
HV-306.1
FILTER
HV-307
HV-305.3 HV-305.4
HV-305.1
FILTER
HV-306
HV-304.4
HV-304.1
HV-304.3
FILTER
HV-305
PT-111
P
HV-301
PT-112
P
P3
P5
FIT-141
T F
HV-303
TT-161
P-201
HV-302 PCV-341
PCV-331
ME1
B1
P1
Component description
Tag Description
B1 Feed
P1 Retentate
P2 Permeate
P3 Feed drain
P4 Cartridge drain
P5 Retentate drain
Introduction
VERSAflux is operated and monitored using the HMI (Human machine interface) that is
displayed on the screen. The screens and the structure of the HMI are described below.
All set range and data in this chapter are for VERSAflux 400 system.
Cabinet indicators
1 2 3
Welcome screen
Step Action
2 Wait until the system is ready, as shown by the System Starting countdown
and appearance of the Logon button.
Step Action
4 Enter the correct information for User and Password, then press OK.
Result:
5 Confirm that USB memory stick is installed on the cabinet, then click YES.
Result: The system moves on to the Home screen.
Home screen
The Home screen provides an overview and is the main control screen for system
operation.
All instruments with corresponding settings can be found in this area, as shown in the
following table (Range data for VERSAflux 400 system):
Note: If an optional instrument is not selected, it will disappear from the PFD area
together with its corresponding settings and related processes.
• TMP value
• TMP control
• Acc. VOL1
• Acc. VOL2
• Delta P control
The Live Trend area displays instance values for the following:
The Instance Alarm & Acknowledge area indicates instance alarm, warnings and system
events. The alarm background will become yellow when acknowledged.
The Acknowledge & Reset area comprises the buttons that are used to acknowledge
and reset information, as shown in the following table.
Button Function
Alarm Ack. Alarms acknowledged by operator, after which the alarm dis-
appears from the window
The Control mode area displays the different control modes that can be selected.
Control group 1:
Control group 2:
The Start and Stop Buttons area allows the user to start and stop the pump.
Name Function
The Foot Display area shows up on all screens. It displays the different screen selection
possibilities for the system.
Setting1 screen
The Setting1 screen includes the Alarm and Warning settings range and enable/disable
functions for the installed in-line instruments.
Click the Setting1 button in the Foot display area to go to the Setting1 screen.
2 Alarm parameter set- User settings for each Need Admin account to
tings instrument modify; data will be re-
Alarm/Warning tained during power
off.
The detailed information for instrument Alarm/Warning settings are shown in the following
table:
WIT-176 1 HA 0.0~1200.0 1 kg
WIT-176 LA 0.0~1200.0 1 kg
WIT-176 HW 0.0~1200.0 1 kg
WIT-176 LW 0.0~1200.0 1 kg
1 WIT-176 is the load cell signal from a tank. It is valid only when the system is connected to a specific tank and the Enable
Tank button is enabled on the Setting2 page.
Setting2 screen
The Setting2 screen includes the Alarm and Warning settings range, the enable/disable
functions and the PID settings for the different control modes together with the Misc
function settings (such as unit switch from bar to PSI, tank load cell enable/disable
function, Buzzer enable/disable function, Compress air alarm enable/disable function,
and the Abnormal flow alarm enable/disable function).
Click the Setting2 button in the Foot display area to go to the Setting2 screen.
• TMP Alarm
• Enable Tank
• Buzzer enable/disable
Note: Once the Enable Tank button has been enabled, a Tank will appear on the
PFD on the Home page. Otherwise the Tank is hidden.
Trend screen
All the installed instruments´ historical data trend can be found on the Trend page.
Click the Trend button in the Foot display area to go to the Trend screen.
In the trend diagram area, the Y axis to the left shows the measurement for flowrate
and temperature (FIT-141, FIT-142, TT-161), and the Y axis to the right shows the mea-
surement for pressure (PT-111, PT-112, PT-113, TMP, Delta P).
The user can drag the marker over the trend curve to read the process data from the
data table below the trend curve area.
AlarmLog screen
The AlarmLog screen records all the warnings and alarms that ocur.
Click the AlarmLog button in the Foot display area to go to the AlarmLog screen.
All warnings will display with a white background. When an alarm occurs, the background
will turn yellow. It will remain yellow when the user acknowledges the alarm and will
turn blue when the fault is corrected and the system is reset.
Step Action
8 Enter the automatic Logoff time for the new user in minutes (default).
Functional Buttons
The Language, Log Off and Exit buttons in the Foot display area are all functional buttons.
Language
By clicking the Language button in the Foot display area the user can switch between
English and Chinese language display.
Log Off
By clicking the Log Off button in the Foot display area, the current user will be logged
off from the system and the user information in the Head display area will disappear,
although the system will keep the current working condition.
When the operation of the system software is resumed, the user will be asked to log on
to the system.
Exit
Only the administrator can exit the current filtration software interface and go back to
the Window CE interface by clicking the Exit button.
Data recording
The system operation data will be recorded on the USB memory stick in *.csv file format.
All the instrument readings will be recorded as Process Data[X].csv, where X stands for
a number, starting from 0, 1, 2, etc.
All the warnings and alarms will be recorded as Alarm Log[X].csv, where X stands for a
number, starting from 0, 1, 2, etc.
The *.csv files can be opened and edited by the Microsoft® Excel® software on any
other computer.
4 Installation
Precautions
WARNING
Before attempting to perform any of the procedures described in
this chapter, you must read and understand all contents of the
corresponding section(s) in the Section 2.1 Safety precautions, on
page 11 as listed below:
In this chapter
This chapter contains the following sections:
4.2 Transport 97
Ambient environment
Avoid:
• Direct sunlight.
• Strong magnetic or electric fields.
• Vibrations.
• Corrosive gas.
• Dust.
Electrical power
Refer to Section4.4.3 Power supply, on page106 and Section9.1 Specifications, on page143
for power, voltage and phase requirements.
Compressed air
• Dry and particle free air for system supply.
• Refer to Section 9.1 Specifications, on page 143 regarding capacity requirements.
• Refer to Connect compressed air supply, on page 102 regarding connections to the
instrument.
Media supply
Refer to Section 9.1 Specifications, on page 143 regarding requirements for media supply
and delivery.
Supply must be arranged so that piping dimensions, piping lengths, valves and height
differences do not obstruct processing.
Computer
VERSAflux 120 and 400 has an industrial built-in HMI installed.
4.2 Transport
WARNING
Move transport crates. Make sure that the forklift has capacity to
safely lift the crate weight. Make sure that the crate is properly
balanced so that it will not accidentally tip when moved.
WARNING
Personal Protective Equipment (PPE). Whenever packing, unpack-
ing, transporting or moving VERSAflux, wear:
• Protective footwear, preferably with steel lining.
• Working gloves, protecting against sharp edges.
• Protective glasses.
Use a pallet jack or fork-lift with a minimum capacity to match the empty weight of the
system plus the transport crate. Refer to Section 9.1 Specifications, on page 143 regarding
system weight.
Note: Make sure that intended openings and apertures are large enough to allow
passage of the crate when lifted from the floor for transport.
Unpacked
WARNING
Heavy object. Because of the significant weight of VERSAflux, great
care must be taken not to cause squeezing or crushing injuries
during movement. At least two, but preferably three or more, per-
sons are recommended when moving the unit.
• VERSAflux can be rolled by hand on hard and level surface with wheel brakes released.
• If the floor quality does no allow rolling the VERSAflux on its own wheels, it can be
moved with a pallet jack or fork-lift.
• For minimum door aperture size, refer to Section 9.1 Specifications, on page 143.
Tools required
• Suitable lifting device, forklift or similar. Refer to Section 9.1 Specifications, on page 143
for dimensions and weight to be handled.
• 13 mm wrench (or ratchet with 13 mm socket).
• Electrical screwdriver no. 2 bit.
• Knife.
• Lever.
Visual inspection
Check:
• that all equipment is enclosed in the crate according to the packing list.
• the equipment for any apparent damage and document carefully if found.
If any equipment is missing or damages are found, contact your GE Healthcare repre-
sentative immediately.
Precautions
CAUTION
To prevent bacterial growth, VERSAflux may be partly filled with
denatured Ethanol (18% C2H5OH (Ethanol), 2% C3H7OH (Isopropanol)
and 80% H2O (water)) or 0.1 M NaOH (Sodium Hydroxide) at delivery.
The mixture is hazardous to humans if consumed. Flush out the
denatured Ethanol or Sodium Hydroxide before assembling, testing
or integrating VERSAflux into the intended process context.
Step Action
1 Remove the front, side and back panels by loosening the bolts.
3 Use a forklift or similar lifting equipment to move the system out from the
crate.
4.4 Setup
In this section
This section contains the following subsections:
Precautions
CAUTION
To prevent bacterial growth, VERSAflux may be partly filled with
denatured Ethanol (18% C2H5OH (Ethanol), 2% C3H7OH (Isopropanol)
and 80% H2O (water)) or 0.1 M NaOH (Sodium Hydroxide) at delivery.
The mixture is hazardous to humans if consumed. Flush out the
denatured Ethanol or Sodium Hydroxide before assembling, testing
or integrating VERSAflux into the intended process context.
4.4.1 Assembly
Locate your air inlet as in the previous figure, and connect to the compressed air supply.
For air supply requirements, refer to Compressed air, on page 95.
Control system
VERSAflux 120 and 400 are delivered with the WinCC Advanced (TIA Portal) V11 system
control software.
WinCC Advanced (TIA Portal) V11 is pre-installed and configured on the built-in HMI with
no additional specifications required.
For more information see Section 3.3.2 Automation, on page 45.
Filter setup
If the system is configured for the intended filter type, there is no need for further config-
uration upon arrival at the process site or inbetween processing runs.
If the system needs to be reconfigured for different size of HF cartridges, refer to Sec-
tion 7.4 Disassembly and assembly, on page 134.
Introduction
This section provides guidelines for the connection of process components to VERSAflux.
Use hoses delivered with the tank to connect the tank and the system to each other.
Use customer supplied hoses to connect the combined tank and system assembly battery
limits to the plant interface points.
Refer to Section 9.1 Specifications, on page 143 for information regarding connection type
and size.
• The chemical properties of the tubing used shall meet or exceed the properties
specified in Section 9.3 Chemical resistance, on page 148.
Power requirements
The power supply requirements are specified in Section 9.1 Specifications, on page 143.
Installation
Connect the system power cord to either:
• A fixed power supply by means of a permanent connection which can be detached
only by the use of a tool.
• An EN 60309-2 compliant connector.
Circuit breaker
A circuit breaker must be present in the fixed power supply.
Breaking the power supply to the VERSAflux system by using this breaker must be
equivalent to disconnecting the power cord for a non-fixed connected instrument.
WARNING
Only personnel authorized by GE Healthcare may perform service,
installation, and maintenance of components inside the VERSAflux
cabinet.
Neutral Black N
Live Black L
10 A 1 mm2
16 A 1.5 mm2
25 A 2.5 mm2
32 A 4 mm2
More information
Voltage, power, fuse requirements and the tripping current for the fixed power supply
ground fault protector can be found in the General Specification provided in the product
documentation package.
5 Run preparations
In this chapter
This chapter contains the following sections:
Precautions
WARNING
Before attempting to perform any of the procedures described in
this chapter, you must read and understand all contents of the
corresponding section(s) in the Section 2.1 Safety precautions, on
page 11 as listed below:
NOTICE
Only use chemicals that have been proven not to be harmful to
the wetted parts of the unit.
Refer to Section 9.3 Chemical resistance, on page 148 for more information.
NOTICE
Make sure that large particles are prevented from entering the
VERSAflux system during operation, otherwise the filters may be-
come clogged and the feed pump may malfunction.
Recommendations
The table below describes the situations where the GE Healthcare range of ULTA™ NFF
capsules can be used.
Part Function
F Permeate out
Part Function
Starting VERSAflux
Step Action
1 Make sure that the air supply to the system is turned on.
3 Make sure that the USB stick has been plugged in.
4 Make sure that the cabinet doors are closed and locked.
6 Wait until the HMI has started up and the Logon button is displayed.
7 Click the Logon button, then enter the correct user name (User) and
password (Password).
8 Click Yes when the USB Memory Installed window pops up.
Step Action
1 Make sure that the system is configured for using HF cartridges and car-
tridges of the intended size, refer to Section 7.4 Disassembly and assembly,
on page 134 if required.
3 Install the cartridge into the system, refer to illustrations in Section 3.5
Hollow fiber (HF) configuration, on page 60.
Introduction
Before using your VERSAflux system and cartridge to process product, you should prime
the rotary lobe pump as follows.
Priming is done for VERSAflux 120 and 400.
NOTICE
Do not start the rotary lobe pump before it is primed.
Prepare connection
Step Action
1 Close the retentate valve, permeate valve, retentate drain valve, permeate
drain valve and cartridge drain valve.
4 Connect the feed inlet with a hose to a container full of pure water.
5 Open the feed supply to let feed flow into the pump slowly.
Run priming
Step Action
1 Run the pump while watching the flow through the membrane cartridge.
6 Operation
In this chapter
This chapter contains the following sections:
Precautions
WARNING
Before attempting to perform any of the procedures described in
this chapter, you must read and understand all contents of the
corresponding section(s) in the Section 2.1 Safety precautions, on
page 11 as listed below:
NOTICE
Only use chemicals that have been proven not to be harmful to
the wetted parts of the unit.
Refer to Section 9.3 Chemical resistance, on page 148 for more information.
NOTICE
Make sure that the actions listed below are completed before
VERSAflux is started.
• Check that the pump is filled with correct lubricant oil before start, refer to the pump
supplier documentation found in the document package for details.
• Check that all inlets and outlets to the system are connected, closed or in the state
that they are supposed to be before the system is started.
- Inlets that are not in use shall be closed.
- Outlets that are not in use shall be fitted with tubing and directed to waste.
• Check the condition of all connections and gaskets.
• Check that no chemicals that may be harmful to the system will be used.
Start VERSAflux
Step Action
1 Make sure that the cabinet doors are closed and locked.
2 Make sure that the Emergency Stop button has been reset.
5 Wait until the HMI has started up and the Logon button is displayed.
6 Click the Logon button, then enter the correct user name (User) and password
(Password).
7 Click Yes when the USB Memory Installed window pops up.
8 Check that the standby state of the HMI has no error message.
Prime VERSAflux
Perform the priming as described in Section 5.4 Priming the rotary lobe pump, on page 116.
In this chapter
This chapter contains the following sections:
6.2.1 Introduction
Process
vp
sp
I
P
D
te Time
This illustration shows the behaviour of the P, I and D components of a typical PID process
control regulator.
• Process is the process to be controlled.
• Time is elapsed time.
• sp is the "set point" where the process should operate.
• te is a point in time where the process is disturbed and the regulator will act.
• vp is the process value.
• P (orange) is the proportional component of the regulator.
• I (cyan) is the integral component of the regulator.
• D (purple) is the derivate component of the regulator.
P Proportional Guides the process to Reduces the effect of an error but does not
its setpoint with an completely eliminate it. A simple P-regulator
input that varies results in a stable stationary error between
proportinally to the actual and requested flow or pressure.
error.
• Too large P-component will make the
system unstable.
• Too small P-component will make the
regulator slow and prevent it from
eliminating the error entirely.
I Integrating Samples the error over The I parameter eliminates the stationary
time and reinforces error, but results in a slight instability leading
the regulator output to oscillations in the actual flow or pressure.
as long as the error is The I parameter can have values between 0
not eliminated. and infinity.
D Derivating Delivers an inital "kick" The D parameter can reduce the oscillations
to the regulator with introduced by a PI-regulator. D can have
an amplitude that is values between 0 and infinity.
proportional to the
• Too large D-component may cause the
error change rate.
system to overshoot initally and make the
system sensitive to noise.
• Too small D-component may cause for
the system to respond slowly to errors and
possiblly oscillate.
If Then...
Adjustment rate
No drastic changes should be applied.
An appropriate measure is to change the I parameter by a factor of 0.5 or 2, respectively,
while initially keeping the P parameter constant.
Faster regulation required
When optimizing the parameters to obtain a faster regulation, a recommended start
may be to:
• Decrease I parameter by dividing by 2.
• Increase P parameter by multiplying by 1.2.
More stable regulation required
When the regulation is too fast, for example noisy curves, it may be necessary to optimize
the parameters in the opposite way, i.e., increase I parameter and decrease P parameter.
Troubleshooting
I Large The system will oscillate after the error has been elimi-
nated.
1 Click the green OFF button on the Home screen in WinCC Advanced (TIA
Portal) V11.
7 Maintenance
In this chapter
This chapter contains the following sections:
Precautions
WARNING
Before attempting to perform any of the procedures described in
this chapter, you must read and understand all contents of the
corresponding section(s) in the Section 2.1 Safety precautions, on
page 11 as listed below:
Component Action
Protective earth Make sure that the protective earth wiring is not disconnected
or damaged.
Monthly or as needed
Component Action
Connection and Check for leakage. Replace seals if needed. Do a leakage test
seals at maximum operating pressure.
Emergency stop Check that the Emergency stop button functions correctly.
Annually or as needed
This section covers maintenance actions required annually or as needed.
Component Action
Complete stainless Inspect stainless steel systems for rust. If needed, passivate
steel system the steel by recirculating a solution of 5% phosphoric acid in
the system overnight at room temperature and then rinse with
purified water until the pH of the outlet liquid is neutral.
Pump Replace all wear and tear parts. Replace drive element lubricant.
Refer to the System Documentation for details.
7.2 Cleaning
General
A suitable frequency of routine cleaning is determined by the nature of the starting
material and the type of process. However, routine cleaning shall be performed at
intervals aimed at prevention rather than cleaning VERSAflux (and connected equipment)
from microbial growth or contamination.
Cleaning-in-place (CIP)
It is possible to run CIP manually on the VERSAflux system.
After completed CIP, rinse the VERSAflux interior and all cleaned components thoroughly
with water of desired quality, for example Water For Injection (WFI), to remove all traces
of the cleaning agent.
Filters
• GE Healthcare Hollow fiber cartridges cleaning procedures are provided in the Hollow
fiber cartridges for membrane separations Operating Handbook, GE Healthcare article
no 18-1165-30.
• Filter units provided by other manufacturers shall be cleaned according to the in-
structions from the respective manufacturer.
External surfaces
The VERSAflux is designed to be operated in a clean environment and the external sur-
faces should not normally accumulate any substantial amount of dust or dirt.
Regularly, wipe the outside of the VERSAflux with a clean cloth. Use a mild cleaning agent
such as water, followed by 20% Ethanol. Regular wiping and care of the equipment will
help keep the surfaces uncorroded.
It is not recommended to spray or splash liquids towards external surfaces.
Do not spray or splash liquids towards filters or the fan on the electric cabinet.
7.3 Storage
Step Action
Step Action
To prevent microbial growth, the storage solution shall be replaced regularly if VERSAflux
is stored for long periods of time.
Filters
• GE Healthcare Hollow fiber cartridges storage procedures are provided in the Hollow
fiber cartridges for membrane separations Operating Handbook, GE Healthcare article
no 18-1165-30.
• Filter units provided by other manufacturers shall be stored according to the
instructions from the respective manufacturer.
User-allowed actions
This section covers all disassembly and assembly procedures that the end user is allowed
to perform without support from GE Healthcare.
WARNING
Only personnel authorized by GE Healthcare may open the cabinet
doors. There is high voltage inside the cabinet that can cause
human injury or death.
WARNING
The electric cabinet doors may only be opened when VERSAflux is
taken out of operation and subject to LOCK OUT / TAG OUT.
WARNING
LOCK OUT / TAG OUT! Before any maintenance or decommission-
ing work is performed on VERSAflux, make sure that:
• it is empty and depressurized.
• it is disconnected from process feed, electrical power and
pneumatic supply.
• it is prevented from accidentally becoming re-energized during
maintenance.
• it is clearly tagged as taken out of operation.
• all process wetted areas are clean and decontaminated.
WARNING
For continued protection against injury risks due to fluid jets, burst
pipes or potentially explosive atmosphere, the piping system must
be tested for leakage at maximum pressure:
• After assembly or maintenance
• Before operation or CIP
NOTICE
While system filter type reconfiguration can be performed by a
properly skilled user Technician, GE Healthcare recommends that
a GE Healthcare service engineer is assigned for this work.
Contact your GE Healthcare representative for more information
on changing filter configurations if required.
HF size 75 Install it directly by connecting to the permeate hose and the car-
tridge drain hose.
Switch to HF 1 Connect the HF55 supporting tube with the HF55 filter on the
size 55 top.
2 Install the filter on the system with a tri-clamp tighten.
3 Connect the filter to the permeate hose (same as for HF75) and
the HF55 permeate port.
4 Install the unique HF55 drain hose on the cartridge drain mani-
fold and connect to the HF55 drain port.
Switch to HF 1 Connect the HF35 supporting tube with the HF35 filter on the
size 35 top.
2 Install the filter on the system with a tri-clamp tighten.
3 Connect the filter to the permeate hose (same as for HF75) and
the HF35 permeate port.
4 Install the unique HF35 drain hose on the cartridge drain mani-
fold and connect to the HF35 drain port.
HF size 85 Install it directly by connecting to the permeate hose and the car-
tridge drain hose.
Switch to HF 1 Connect the HF65 supporting tube with the HF65 filter on the
size 65 top.
2 Install the filter on the system with a tri-clamp tighten.
3 Connect the filter to the permeate hose (same as for HF85) and
the HF65 permeate port.
4 Install the unique HF65 drain hose on the cartridge drain mani-
fold and connect to the HF65 drain port.
Switch to HF 1 Connect the HF45 supporting tube with the HF65 filter on the
size 45 top.
2 Install the filter on the system with a tri-clamp tighten.
3 Connect the filter to the permeate hose (same as for HF85) and
the HF45 permeate port.
4 Install the unique HF45 drain hose on the cartridge drain mani-
fold and connect to the HF45 drain port.
Other components
Other than described in this chapter, the VERSAflux is not designed to be disassembled
or assembled by the user.
If the need for further disassembly should arise, always contact your GE Healthcare
representative for advice before attempting any actions not described by this document.
8 Troubleshooting
Precautions
WARNING
Before attempting to perform any of the procedures described in
this chapter, you must read and understand all contents of the
corresponding section(s) in the Section 2.1 Safety precautions, on
page 11 as listed below:
System
Power Power failure during Check circuit breaker, both in system and for
a run. external supply, as applicable.
Compressed Compressed air fail- Insufficient air pressure caused by, for in-
air ure during a run. The stance, malfunctioning air supply equipment
alarm “No air supply or a leaking air hose.
to the system” is dis- a) Rectify the air supply problem.
played and the sys- b) Restart the run by pressing the green ON
tem is stopped. button on the Home screen.
Pressure curve
Erratic flow, noisy base- Gas bubbles passing Check that there is sufficient
line signal, irregular through or trapped in supply of liquid.
pressure trace. the pump.
Check all connections for leaks.
9 Reference information
In this chapter
This chapter contains the following sections:
9.1 Specifications
Dimensions
Note: All dimensions are presented in cm where nothing else is noted.
VERSAflux 120
205
144 80
VERSAflux 400
235
197 95
Technical specifications
Overvoltage category II
Pollution degree 2
Operating limits
Property Value
Environmental limits
Parameter Value
Ingress protection IP 54
Battery limits
Process
The indicated dimensions represent the TriClamp flange size and the minimum inner
diameter for used hoses required to match battery limit connections.
Pneumatic
Electrical
Power consumption
120 3A 5A
400 3A 18 A
Introduction
The table below gives allowed exposure concentrations and times for various chemicals
that may be used in GE Healthcare BioProcess instruments in general. For information
regarding hollow fiber chemical resistance, please refer to Hollow fiber cartridges for
membrane separations Operating Handbook, GE Healthcare article no 18-1165-30.
Some of the chemicals listed may not be applicable for your instrument.
WARNING
Flammable liquids. VERSAflux is not approved to handle flammable
liquids.
WARNING
Some of the chemicals used with VERSAflux may be flammable
under certain conditions. Make sure to use chemicals only under
conditions where they are not flammable. Refer to local and/or
national classifications of flammable liquids.
Remaining aspects
Regarding
• Training
• Service
• Method optimization
• Ordering information
• Other issues not covered by this manual
Please contact your local GE Healthcare representative for advice.
Contact information is found on the back cover of this manual.
Index
A Method optimization, 150
More information, 150
Accessories, 150
Ambient environment, 95 N
C NFF, reasons for , 111
NFF, recommendations, 111
Checks, final before start, 119 NFF, sample handling, 112
Chemical resistance, 148 Notes and tips, 8
Cleaning
Cleaning agents, 130 O
External surfaces, 131
Operating limits, 144
General, 130
Ordering information, 150
Compressed air, 95
D P
Personal protection, 13
Decontamination, 27
PID, faster regulation, 125
Disassembly, other compo-
PID, more stable regula-
nents, 137
tion, 125
Disposal, general instruc-
PID, parameters descrip-
tions, 27
tion, 123
Disposal of electrical compo-
PID, trouble shooting, 126
nents, 27
PID, tuning and adjust-
E ments, 125
Power failure, 26
Electrical power, 95 Prerequisites, 7
Emergency Purpose of this document, 5
Procedures, 23
Restarting the system, 26 R
Shutdown, 25
Recycling of hazardous sub-
Stop, 25
stances, 27
H Reference information, 142
Restart after emergency, 26
Hollow fiber cartridges, installa- Rinsing, after CIP, 130
tion, 115
S
I
Safety
Installing and moving, precau- notices, 7
tions, 14 Safety labels, 22
L Safety precautions, 11
Service , 150
Labels Space and floor load, 95
Safety labels, 22 Spare parts, 150
Starting, VERSAflux, 114
M Storage
Maintenance schedule, 129 Long term, 132
Media supply, 96
imagination at work
29-0900-27 AA 01/2014