Basic Operation
This vapor-actuated gage features a sealed capillary completes a circuit when it touches the adjustable limit
tube and sensing bulb. When subjected to heat, the contacts. Contacts have self-cleaning motion to ensure
liquid in the sensing bulb expands to vapor creating electrical continuity. A toggle switch is provided on
pressure against a bourdon tube mechanism. The SPLC models to override the low contact for equipment
bourdon tube translates this vapor pressure into a start-up.
mechanical gage reading. Models 45TE and 45TEF have internal snap-acting
For models SPLC and SPLFC, the gage pointer acts SPDT switches.
as a pressure indicator and as one switch pole, which
Dimensions
Temperature Ranges
Accuracy (SPL and 45 Series Models) Minimum Sensing Bulb
Available (dual scale dials)
Insertion into Process
Fahrenheit Celsius First 1/3 Middle 1/3 Upper 1/3
15° to 250° 9° to 121° ±8°F/±4°C ±2°F/±1°C ±2°F/±1°C 5 in. (127 mm)
130° to 350° 54° to 177° ±8°F/±4°C ±2°F/±1°C ±3°F/±1.5°C 2-1/2 in. (64 mm)
260° to 450° 127° to 232° ±8°F/±4°C ±2°F/±1°C ±3°F/±1.5°C 2-1/2 in. (64 mm)