POWER ANALYZER

W E L C O M E T O T H E D E W E S O F T E X P E R I E N C E . O N E S O F T W A R E , O N E H A R D W A R E , O N E S O L U T I O N .
2 O N E H A R D W A R E .

HARDWARE HIGHLIGHTS
DEWESOFT HIGHEST ACCURACY HIGH SAMPLE RATE
7-YEAR WARRANTY Even though extremely small, Dewesoft Dewesoft data acquisition hardware features
Our warranty covers that the instruments power analyzer is highly accurate with 0.03 % high sampling rate amplifiers with a 1 MS/s
function as promised for a period of 7 years accuracy. sampling rate for the HS series and 15 MS/s for
from the day of the delivery. the XHS series.

FULLY ISOLATED
The worry free solution provides isolation on
the sensor side (channel to GND, as well as,
channel-to-channel) and even isolated sensor
excitation! Less noise, no ground loops, best
signal quality.
O N E H A R D W A R E . 3

SMALLEST POWER ANALYZER PLUG AND PLAY FLEXIBLE AND SCALABLE High Voltage
input
SIRIUS technology allows us to build the Any device, sensor or signal. Smart Dewesoft allows the user to customize the
ADC type 16 bit
world’s smallest power analyzer. sensors with TEDS are recognized instrument exactly to their needs. Select
automatically. the instrument and amplifiers and build Sampling rate Simultaneous 1 MS/s

FULLY SYNCHRONIZED your perfect measurement instrument. Ranges ±1600 V, ±800 V, ±400 V, ±200 V,
±100 V, ±50 V, ±20 V
Data from various sources are perfectly TOTAL SOLUTION Use the All-In-One Instruments with up to
Input impedance 10 MΩ in parallel 2pF 
aligned: Analog, Digital, Counter, Vehicle Combine your power measure­ments 64 input channels for high channel counts
Overvoltage In+ to In-: 4 kVpk-pk , Inx to GND: 2
buses, Video, ... with NVH, combustion analysis, Vehicle or synchronise any number of modular protection kVpk-pk , CAT II 1000V, CATIII 600V
Dynamics and other powerful Dewesoft SIRIUS units for distributed measure­ment Accuracy (DC –
VOLTAGE INPUTS tools. applications. 1kHz) 0.03% of reading & 0.02% of range

Measurement ranges up to
1600 V DC (CAT II 1000V). INTERFACES TEMPERATURE Low Voltage input
Intelligent Interfaces for simple and RANGE
reliable integration to other systems. Our instruments are the perfect solution ADC type 16 bit SAR with 100 kHz 5th order
analog AAF filter or bypass
DCOM, CAN, EtherCAT®, Ethernet, XCP. for summer and winter testing. The wide
Sampling rate Simultaneous 1 MS/s
operating temperature range allows using
±100V, ±50V, ±20V, ±10V, ±5V, ±2V,
them down to -30° C. Ranges ±1V, ± 500mV, ±200mV, ±100mV
and 50mV
Input impedance 10 (1) MΩ between IN+ or In- and
(100 V range) GND
2 to 30 V bipolar / 0 to 24 V
Sensor Excitation unipolar, sw programmable (16 bit
DAC), max 0.2 A / 2 W
Range < 10 V: 100V (200 V peak for
Overvoltage 10msec); Range ≥ 10 V: 300 V cont.;
protection 1000 V with banana plug
BNC, DSUB 9, Banana, Screw
Connector Connector
Accuracy (DC – 0.03% of reading & 0.02% of range
1kHz)
4 O N E H A R D W A R E .

HARDWARE OVERVIEW

SIRIUS® SBOX SIRIUS® R4/R4rt/R4-HUB


Synchronized, Integrated solution with 4 SIRIUS slices
highly reliable and powerful SBOX computer or USB hub
data logger and in one unit with real-time EtherCAT® slave
SIRIUS® MODULAR powerful data interface.
processing
Most flexible and
distributable single slices
computer. SIRIUS® XHS
with USB and EtherCAT®
interface. High-speed data acquisition system
(15 MS/sec) with the new Hybrid ADC
technology capable of high-bandwidth
transient recording and very high-
dynamic, alias-free data acquisition.

KRYPTON® 1
SERIES
Distribute your
ANALOG OUTPUTS measurements
SIRIUS slices can be configured with 8 analog down to a single
outputs and function as a multi-channel Our systems are completely customizable channel
function generator, can also do real-time signal to the customers’ needs. Amplifiers and
conditioning, analog replay of data in analysis, connectors can be chosen as well as HIGH ISOLATION
and perform manual or automated control High channel-channel and channel-ground
output with output voltage levels of up to number of signal inputs and the number isolation prevents ground loops and damage
+/- 10V. of measurement inputs. to the system from excessive voltage.
O N E H A R D W A R E . 5

DIGITAL/COUNTER/ENCODER INPUTS
Each counter channel is capable of 3x digital
inputs, 1x event counter, encoder, period,
pulse-width, duty-cycle, and precise frequency
SIRIUS® R8DB/R8rt and angle measurement using patented
SuperCounter® technology.
Integrated instrument with 8 SIRIUS slices,
powerful SBOX computer, optional 19” display
(R8D) and batteries (R8DB) and real-time SIRIUS® R1DB/R2DB SIRIUS® R3
EtherCAT® slave interface (R8rt).
Small-size instrument with embedded Up to 3 SIRIUS slices in a rack
computer, mounted lab unit with standard
12” display and batteries. easy-to-upgrade computer.

Our data acquisition systems are


versatile, modular, easy to use and can
work with any sensor with the highest
SENSOR POWER SUPPLY precision imaginable. Input channel
Amplifiers provide channel-independent, UNIVERSAL ANALOG INPUTS
programmable power supply for sensor configurations are flexible and the input A wide variety of universal and analog
excitation. channel count can vary from 1 to 1000’s amplifiers that accept voltage and full/half/
HYBRID ADC TECHNOLOGY of channels. Our measurement systems quarter bridge signals natively as well charge
Offers everything you ever wanted out of ISOLATED CAN BUS INTERFACE and IEPE accelerometers, thermocouple and
a high-end data acquisition system. High High-speed CAN 2.0b channels with 1 Mbit/sec are flexible and can grow with you at any RTD temperature sensors, current, resistance,
bandwidth and high dynamic mode available data throughput with additional support for time in your measurement process. and even LVDT sensors, with the use of DSI
software selectable per channel. CCP, OBDII, J1939, and CAN output. adapters.
6 O N E H A R D W A R E .

CLAMPS & TRANSDUCERS

-40°C TO +85°C HIGH ACCURACY


Dewesoft offers current transducers with wide
temperature range - ideal for both winter High-accuracy sensors for AC/DC current Highly precise zero flux current transducers or
(-40°C) and summer testing (+85°C).
UP TO 30 000 AMPS measurement and power analysis. From
fluxgate compensated clamps are a perfect fit
Dewesoft offers a very wide range of current for most demanding power measurements for
measurement ranges up to 30 000 (30 current clamps, high-precision zero E-mobility and inverter motors applications.
FLEXIBLE thousand) amps. flux current transducers, shunts and
Dewesoft instruments allow connecting any INTEGRATED SENSOR
type of current transducers. There are more AC/DC CURRENT MEASUREMENT Rogowski coils. POWER SUPPLY
than 1000 different current sensors available Dewesoft offers high-accuracy zero-flux Current clamps and zero-flux transducer can
on the market. If you want to connect your current transducers, Rogowski coils, current be powered straight from the DAQ instrument
own or other sensors we are happy to help clamps and shunts for AC and/or DC current like R2DB, R8 or with external SIRIUS slice
you. measurement. compatible chassis.
O N E H A R D W A R E . 7

CURRENT TRANSDUCERS

IT 60-S IT 200-S IT 400-S IN 500-S IT 700-S IN 1000-S IN 2000-S

Primary Current Range DC 60 A 200 A 400 A 500 A 700 A 1000 A 2000 A


RMS Sinus
Overload Ability Short Time (100 ms) 300 Apk 1000 Apk 2000 Apk 5000 Apk 3500 Apk 5000 Apk 10000 Apk
Max. burden resistor (100 % of Ip) 10 Ω 10 Ω 2.5 Ω 2.5 Ω 2.5 Ω 4Ω 3.5 Ω
di/dt (accurately followed) 25 A/μs 100 A/μs 100 A/μs 100 A/μs 100 A/μs 100 A/μs 100 A/μs
Temperature influence < 2.5 ppm/K < 2 ppm/K < 1 ppm/K < 0.2 ppm/K < 1 ppm/K < 0.3 ppm/K < 0.1 ppm/k
Output Ratio 100 mA at 60 A 200 mA at 200 A 200 mA at 400 A 666 mA at 500 A 400 mA at 200 A 666 mA at 1000 A 1 A at 2000 A
Bandwidth (0.5 % of Ip) DC ... 800 kHz DC ... 500 kHz DC ... 500 kHz DC ... 520 kHz DC ... 250 kHz DC ... 440 kHz DC ... 140 kHz
Linearity < 0.002 % < 0.001 % < 0.001 % < 0.003 % < 0.001 % < 0.003 % < 0.003 %
Offset < 0.025 % < 0.008 % < 0.004 % < 0.001 % < 0.005 % < 0.0012 % < 0.0012 %
Frequency Influence 0.04 %/kHz 0.06 %/kHz 0.06 %/kHz 0.01 %/kHz 0.12 %/kHz 0.1 %/kHz 0.1 %/kHz
Angular Accuracy < 0.025 ° + 0.06 °/kHz < 0.025 ° + 0.05 °/kHz < 0.025 ° + 0.09 °/kHz < 0.01 ° < 0.025 ° + 0.18 °/kHz < 0.01 ° + 0.05 °/kHz < 0.01 ° + 0.075 °/kHz
Rated isolation voltage RMS,
single isolation CAT III, 2000 V 2000 V 2000 V 1600 V
pollution deg. 2, 1000 V 1000 V 1000 V
1000 V 1000 V 1000 V 1000 V
IEC 61010-1 standards,
EN 50178 standards
Test voltage 50 / 60 Hz, 1 min 5.4 kV 5.4 kV 5.4 kV 4.2 kV 4.6 kV 4.2 kV 6 kV
Inner diameter 26 mm 26 mm 26 mm 38 mm 30 mm 38 mm 70 mm
Dimension 93 x 77 x 47 93 x 77 x 47 93 x 77 x 47 128 x 106 x 54 128 x 106 x 67 128 x 106 x 54 231 x 220 x 76
Dewesoft® Shunt 5Ω 5Ω 2Ω 1Ω 2Ω 1Ω 1Ω
PWR-MCTS2 needed Yes Yes Yes Yes Yes Yes Yes
8 O N E H A R D W A R E .

CURRENT CLAMPS AC/DC

DS-CLAMP-200DC DS-CLAMP-500DCS DS-CLAMP-500DC DS-CLAMP-1000DC DS-CLAMP-150DC DS-CLAMP-150DCS DS-CLAMP-1800DC

Current Range 200 A DC or AC RMS 500 A DC or AC RMS 500 A DC or AC RMS 1000 A DC or AC RMS 200 A DC or 150 A AC RMS 250 A DC or 150 A AC RMS 1800 A DC or AC RMS

Type Flux Gate sensor Flux Gate sensor Flux Gate sensor Flux Gate sensor Hall sensor Hall sensor Hall sensor
Sensitivity ± 10 mV/A ± 4 mV/A ± 4 mV/A ± 2 mV/A 20 mV/A 20 mV/A 1 mV/A
0 - 1000 A: ± 2.5 % of
reading ± 0.5 A
1000 - 1500 A: ± 3.5 % of
Accuracy 0.3 % of reading 0.3 % of reading 0.3 % of reading 0.3 % of reading 1 % + 2 mA 1 % + 2 mA reading
1500 - 1800 A: ± 5 % of
reading
Bandwidth DC to 500 kHz DC to 200 kHz DC to 100 kHz DC to 20 kHz DC to 100 kHz DC to 100 kHz DC to 20 kHz
TEDS Fully supported Fully supported Fully supported Fully supported Fully supported Fully supported Fully supported
Conductor Position max. ± 0.1 % of reading max. ± 0.1 % of reading max. ± 0.2 % of reading max. ± 0.2 % of reading ± 0.5 % ± 0.5 % ± 1.5 %
Sensitivity
± 10 mV ± 10 mV
Zero Offset (+25°C) max. ± 10 mA max. ± 250 mA max. ± 250 mA max. ± 50 mA -
max. + 1 mV/°C max. + 1 mV/°C
Dimensions [mm] 153 x 67 x 25 153 x 67 x 25 238 x 114 x 35 238 x 114 x 35 205 x 60 x 15 106 x 100 x 25 205 x 60 x 15
Max. Conductor Size 20 mm 20 mm 50 mm 50 mm 32 mm 25 mm 32 mm
Power Supply ± 11 V to ± 15 V ± 11 V to ± 15 V ± 11 V to ± 15 V ± 11 V to ± 15 V ± 10 V ± 10 V +9V
PWR-MCTS2 needed Yes Yes Yes Yes No No No
O N E H A R D W A R E . 9

CURRENT CLAMPS AC / ROGOWSKY COILS AC

DS-CLAMP-5AC DS-CLAMP-15AC DS-CLAMP-200AC DS-CLAMP-1000AC DS-FLEX-3000- DS- DS-


DS-FLEX-3000-17 DS-FLEX-3000-35 35-HS FLEX-3000-80 FLEX-30000-120

Type Iron-Core Iron-Core Iron-Core Iron-Core Type Rogowski coil Rogowski coil Rogowski coil Rogowski coil Rogowski coil
3 A, 30 A, 3 A, 30 A, 3 A, 30 A, 30 A, 300 A,
Current 3000 A
Range 5 A AC RMS 15 A AC RMS 200 A AC RMS 1000 A AC RMS 300 A, 3000 A 300 A, 3000 A 300 A, 3000 A 3000 A, 30000 A
Range AC RMS
AC RMS AC RMS AC RMS AC RMS

Sensitivity 100 mV/A 100 mV/A 10 mV/A 1 mV/A 3 A: 10 Hz to 3 A: 10 Hz to


3 A: 10 Hz to 10 kHz 10 kHz 5 kHz
Bandwidth 3 A: 10 Hz to 10 kHz Others: 10 Hz to 5 Hz - 1MHz Others: 10 Hz to Others: 10 Hz to
5 mA - 0.5 A 0.5 A - 10 A <1A 20 kHz 20 kHz 20 kHz
± 2 % of reading <1A ± 3.5 % of reading ± 2 % of reading
0.5 A - 1 A ± 2.5 % of reading 10 A - 100 A 10 A - 100 A Others: 10 Hz
Accuracy (+25 °C) 3 A: 10 Hz to 10 kHz ≤1,5 % ≤1,5 % ≤1,5 % ≤1,5 %
± 1 % of reading 1 A - 15 A ± 2.5 % of reading ± 0.5 % of reading to 20 kHz
1 A - 12 A ± 1 % of reading 100 A - 240 A 100 A - 1200 A
±0 .5 % of reading ± 1 % of reading ± 0.3 % of reading Others: 10 Hz 350 mm 800 mm 1200 mm
5 Hz - 1MHz 3 A: 10 Hz to 10 kHz
to 20 kHz (Ø 100 mm) (Ø 250 mm) (Ø 380 mm)
5 mA - 0.5 A 0.5 A - 10 A <1A
±2° <1A not specified not specified 30 A: 10 Hz to 5 kHz
0.5 A - 1 A ±5° 10 A - 100 A 10 A - 100 A Others: 10 Hz
Phase Error Others: 10 Hz to Not Supported Fully Supported not supported Not Supported
± 1° 1 A - 15 A ≤5° ≤1° to 20 kHz 20 kHz
1 A - 12 A ±3° 100 A - 240 A 100 A - 1200 A
±1° ≤ 2.5 ° ≤ 0.7 ° Accuracy ≤ 1.5 % ≤ 1.5 % ≤ 1.0 % ≤ 1.5 % ≤ 1.5 %
(+25 °C)
Bandwidth 2 Hz to 5 kHz 2 Hz to 10 kHz 2 Hz to 10 kHz 2 Hz to 10 kHz
Coil Length 170 mm 350 mm 350 mm 800 mm 1200 mm

Resolution 0.01 A 0.01 A 0.5 A 0.001 A TEDS Not supported Not supported Fully supported Not supported Not supported

TEDS Fully supported Fully supported Fully supported Fully supported

Overload Capability 1.5 x I nominal 1.3 x I nominal 1.3 x I nominal 1.3 x I nominal

Dimensions [mm] 102 x 34 x 24 135 x 51 x 30 135 x 51 x 30 216 x 111 x 45


DSIi-10A DSIi-20A DS-SHUNT-05

Type Current Transducer Current Transducer Shunt


Conductor Diameter 15 mm 20 mm 20 mm 52 mm
Current Range 10A AC/DC 20A AC/DC 5A
Other Current Transducers for AC and DC measurement from 300 mA up to 4000 A on request. Bandwidth 100 kHz 100 kHz -
Accuracy (+25 °C) 0.3 % 0.3 % 0.1 %
Resistance - - 50 mOhm
10 O N E H A R D W A R E .

SENSOR POWER SUPPLY DIRECT CONNECTION TO THE AMPLIFIER (LV)

Most of the current sensors can get CONNECTION VIA SIRIUSI-PWR-MCTS2


power supply directly out of the amplifier
(HS-LV or LV) like Rogowski Coils, AC/DC
Clamps, DSI-adapters. For current sensors
with high-power consumption like AC/ DS-CLAMP-200DC SIRIUSI-PWR-MCTS2
DC fluxgate clamps or the zero-flux
transducers there is an additional Power SIRIUS-4XHS-HV, 4XHS-LV
supply unit available, which can be fully
integrated into All-In-One Instrument
D9M-D9F-03M-CLAMP
(R8D, R2D) or as external SIRIUS box. the
most compact power supply solution on
SIRIUSi-PWR-MCTS2 / SIRIUSir-PWR-MCTS2
the market.
Power supply 9-36V DC
Max power 85 W CONNECTION VIA SIRIUSI-PWR-MCTS2
consumption
Physical dimensions 265 x 140 x 65 [mm]
Operating temperature -20 to 50°C
Storage temperature -40 to 85°C
SIRIUSI-PWR-MCTS 2
Humidity (@60°C) 95% RH non-condensing
4x Isolated Power supply
Output (1500V DC, 60sec) SIRIUSI-HS 4XHV, 4XLV
Output voltage +/-15V DC
Maximum output per 20 W
channel D9M-D9F-03M-CLAMP
Short circuit protection indefinite (automatic IT-400S
recovery) D9M-D9F-5M-MCTS
Over load protection 150 % of Iout max. typ
O N E H A R D W A R E . 11

XHS-PWR SIRIUS®

The SIRIUS XHS-PWR is a DAQ


device designed for direct in-vehicle
measurement of current, voltage, and
power. It features an integrated patented
DC-CT current transducer for very precise HIGH BANDWIDTH MODE
current measurements in the most This mode offers more than 5 MHz bandwidth
demanding applications such as very and 15 MS/s sampling rate, SIRIUS XHS can
acquire impulse, step, and square signals PERFECT SYNCHRONIZATION
high current peaks as well as leakage without any ringing or overshoot. Such an Even though you can select some channels to
current measurement. acquisition mode is perfect for transient be high bandwidth and some to be alias free,
recording and power analysis, and would filtering is made in the way that all signals are
usually be found in SAR ADCs. perfectly time aligned with zero phase shift.
12 O N E S O F T W A R E .

HIGHLIGHTS

THE DEWESOFT POWER


ANALYZER COMBINES
THE FUNCTIONALITY
OF MULTIPLE
INSTRUMENTS IN
JUST ONE COMPACT
INSTRUMENT.
Our Power Analyzer isn’t just the smallest
one in the world - it’s also the most
capable. Our flexible hardware combined
with Dewesoft X software creates a
whole new world of testing possibilities
for electrical measurement applications.
O N E S O F T W A R E . 13

HIGHLIGHTS

POWER ANALYZER RECORDER / DATA LOGGER FFT ANALYSIS POST PROCESSING


P, Q, S, PF, cos phi,… Possibility of more than Raw data storing at full sampling rate. 2D FFT, 3D FFT, Sideband Marker etc. Powerful analysis after measurement.
100 calculable values.

POWER QUALITY RAW DATA TRANSIENT RECORDING STATIC AND DYNAMIC TESTING
FFT, Harmonic FFT, Harmonics, Interharmonics, Raw Data is essential for a detailed analysis Triggering on analogue, math or power Our sophisticated Power calculation
Higher Frequencies, Flicker, Flicker emission of any electric machine. Transients and channels. algorithms ensure amazing results during
etc. oscillations can be captured continuously both static and dynamic recording conditions.
or with a trigger condition. Power values Analysis of both low-speed wind turbine
STANDARDS together with raw data allows the detection of power (<10 Hz), up to high speed electric
The instruments comply to a list of anomalies. vehicle motors (<3000 Hz) is possible. Detailed
international standards such as the IEC61000- analysis of period-based values is also
4-30 among others. included.
14 Y O U R S O L U T I O N .

POWER ANALYZER

Digital Signal

SOUND &
VIBRATION

High-Bandwidth Signal Conditioning

ELECTRIC MOTOR INVERTER BATTERY


SIRIUS® XHS 8xACC
Direct high voltage (2000V, CAT II 1000V isolated)

Direct current transducer connection (TEDS Supported)


SIRIUS® XHS 4xHV 4xLV

CURRENT SENSORS
We offer high-accuracy current sensors such as
FULLY ISOLATED 0.03 % ACCURACY High-sampling rate, high-bandwidth,
zero-flux current transducers, flux-gate current
Our worry-free solution provides sensor We offer high-accuracy amplifiers and sensors transducers, AC/DC current clamps, Rogowsky
isolation (channel-to-ground), as well as for voltage and current measurement with and high-accuracy hardware for coils and shunts with the power supply directly
from the system.
channel-to-channel isolation, and even accuracy as high as 0.03 %. power analysis on electric motors,
excitation isolation! Less noise, no ground
loops, and the best possible signal quality. inverters, transformers, switches and ADVANCED ONLINE AND OFFLINE
15 MS/s SAMPLING RATE any other electronic equipment. In MATH PROCESSING
2000 V DC / Dewesoft data acquisition hardware features conjunction sensors can be connected for
DewesoftX includes an easy-to-use
CAT II 1000 V/ CAT III 600 V high sampling rate amplifiers with 15 MS per mathematics engine. You can apply math
Direct input and acquisition of high voltage second sampling rate. Making it ideal for fast temperature, vibration, RPM and torque functions during the measurement, as well as
during post-processing.
signals. transient behavior. measurement, and more.
Y O U R S O L U T I O N . 15

EXTENDABLE MODULAR DESIGN

BATTERY INVERTER

DC POWER POWER MOTOR TORQUE / SPEED POWER LOAD

POWER CALCULATIONS
SCOPE AND FFT The system will calculate more than 100 power
EXTENDIBLE MODULAR DESIGN In addition to the power analysis other useful parameters such as P, Q, S, PF, cos phi and
Multiple power modules can be stacked tools and visualisations are available such as many others. All these calculations can be
to extend channel count - all perfectly a Scope, Vector Scope, Harmonic FFT, 2D FFT, done during the measurement or in post-
synchronized to each other. processing.
Our Power Analyzer isn’t just the and 3D FFT. For example the 3D FFT of a motor
run-up will yield valuable information about
DISTRIBUTABLE smallest one in the market - it’s also the behavior of the machine in a single plot. STATIC AND DYNAMIC TESTING
Several devices can be over long distances and the most capable. Flexible hardware Our sophisticated power calculation
still be perfectly synchronized for fault location
combined with DewesoftX creates a RAW DATA algorithms ensure amazing results during
analysis. Raw data storage is essential for detailed both static and dynamic recording conditions.
whole new world of testing possibilities analysis of all measured signals. Transients Analysis of both low-speed wind turbine
HIGH PRECISION POWER ANALYSIS for applications across a plethora of and Oscillations can be captured continuously power (<10 Hz), and high-speed electric
High-accuracy Dewesoft hardware combined or by means of a trigger. Power values in vehicle motors (>5 kHz) is possible. Detailed
with DewesoftX Power software guarantees industries. conjunction with raw data allow for immediate analysis with period-based values is also
reliable measurement results. anomaly detection. included.
16 Y O U R S O L U T I O N .

E-MOBILITY ADDITIONAL AUTOMOTIVE


data has the highest quality for the analysis.
The Dewesoft system is also able to measure TESTING POSSIBILITIES
various other types of signals such as GPS, The Dewesoft measurement devices can also
vibration, CAN, video, torque, acceleration etc. be used for a variety of other automotive
ELECTRIC VEHICLE MODULAR HARDWARE DESIGN
The modular hardware design makes it The data that is acquired is very versatile,
testing procedures see the list of applications.
More details can be found in the Automotive
TESTING possible to measure power (AC or DC) at
multiple measurement points perfectly
analysis can be done during the measurement
as well as in the powerful post processing
Testing brochure.
synchronised. This unique feature provides tool enabling analysis such as energy flow
ONLINE ANALYSIS OF EFFICIENCY – the flexibility to do a comprehensive analysis diagrams, influence factors on efficiency, • Autonomous driving
RECUPERATION – ENERGY BALANCE of different types of electric drive trains comparison to other vehicles, charging • Vehicle dynamics
The Dewesoft Power Analyzer combining the (Single motor, motor-generator, multiple analysis, comparison of different driver • Ride and handling tests
comprehensive DewesoftX software and the motor configuration from 2 to 4 motors). behaviour on different drive cycles. By virtue of • Brake testing
SIRIUS measurement device is the perfect All the auxiliary loads can be measured and the small form factor of the Dewesoft devices • Advanced driver assistance systems
solution for an all-in-one measurement analyzed simultaneously - including heating, they can be used in even the smallest electric • Pass-by noise
experience for electric vehicles. Measuring any air-conditioning, 24V loads, and 12V loads to vehicles. • Combustion analysis
type of motor, be it single phase or multiphase name just a few. • Torsional and rotational vibration
(up to 12 phases), inverter (DC/DC, AC/AC, ANALYZING DIFFERENT DRIVING • Order tracking
DC/AC) testing capabilities into the multiple THE HIGH SAMPLING RATE AND SITUATIONS • Road load data
hundred kHz region, as well as measuring BANDWIDTH There are various parameters that can • Performance testing
battery parameters. Please see the section: The high sampling rate (up to 15 MS/s) and influence the energy consumption of an • Component testing
Battery testing. high bandwidth (up to 5 MHz) ensures that the electric vehicle. These could be: • Modal analysis
• ambient influences, such as temperature • Structural testing
or weather, • Crash tests
• the quality of the road surfaces, or • Structural testing
different driving situations (uphill,
downhill, city, highway. overland or
combined drives), and On-road testing considering different
• the driver profile, as no two drivers have
the exact same driving behavior. driving conditions and driver profiles.
EL. MOTOR
The Dewesoft Power Analyzer is able to do
an energy analysis considering all of these
SIRIUSi XHS-PWR provides 1000 A
parameters during the test drives. calibrated range and up to 2000 A peak
current and over 500 kHz bandwidth.

BATTERY INVERTER

AC VOLTAGE,
CURRENT

DC VOLTAGE, CURRENT
Y O U R S O L U T I O N . 17

HYBRID & HYDROGEN CA-Scope (pressure over angle).


All CA specific calculations like the mean
isolated SIRIUSi charge inputs (with up to 24
Bit resolution) are in perfect sync with the
and de-pressurizing tests. The flow velocity
of the hydrogen to the fuel cell needs to be
effective pressure (IMEP, PMEP), heat release, Dewesoft Supercounters. This allows analysis monitored. Testing of pressure relief devices
TESTING start/ end of combustion (SOC, EOC), start/ of hybrid cars during real drive tests. that are temperature dependant for safety
end of injection (SOI, EOI), indicated power, reasons in the event of an emergency. These
COMBINED POWER maximum pressure (Pmax), derivate pressure ELECTRIC VEHICLE TESTING are just a few of the additional parameters that
AND COMBUSTION ANALYSIS (dp/da) are presented either as colour For more information on the measurements have to be measured.
Combining internal combustion engines and diagrams or as data tables. For more detailed of the electrical parameters please refer to the The flexibility and modular design of the
electric motors to propel vehicles can prove to analysis, statistical calculations per cylinder or electric vehicle testing section of this catalog. Dewesoft data acquisition devices makes
be quite the challenge when measurements over the complete engine can be performed. it possible to measure all the additional
of the two propulsion methods need to HYDROGEN TESTING parameters and more - GPS, torque, speed,
be compared. This is not the case with the KNOCKING DETECTION AND Fuel cell vehicles (FCV) use a fuel cell to vibration, CAN, video etc.
Dewesoft Combustion Analyzer and Power COMBUSTION NOISE store hydrogen that is then converted to
Analyzer. Dewesoft provides a dedicated knocking electrical power by a chemical reaction in
detection and combustion noise algorithm. the fuel cell that is used to drive the electric. High-accuracy combustion analyzer system
SYNCHRONISED MEASUREMENT The basis for all of these calculations are As a FCV has a tank that stores the hydrogen for engine research, development and
precise angle position data and cylinder there are a few additional parameters that
Due to the modular design of the Dewesoft
pressure measurement. Dewesoft provides must be measured. The pressure of the tank
optimization as well as testing of ignition
data acquisition devices these two very
different types of propulsion systems the ideal hardware for this: the galvanically needs to be monitored through pressurizing systems, exhaust systems, and valve
can be measured and analyzed perfectly control gear.
synchronised.
SIRIUS® XHS 4XHV 4XLV
COMBUSTION ANALYZER
The Dewesoft Combustion Analyzer enables
the user to display and compare measured
parameters using several different diagrams
such as, pV-diagrams (pressure of angle) or the

COMBUSTION MOTOR ELECTRIC MOTOR INVERTER BATTERY

Engine ECU
Crankshaft angle sensor
Injection detection
Cylinder pressure
Vibration sensors

CAN FD, LIN, XCP

SIRIUS® XHS-PWR
SIRIUS® HS 6xCHG 2xSTG+
18 Y O U R S O L U T I O N .

E-MOBILITY BATTERY TESTING Battery testing has a broad spectrum of testing


requirements, Dewesoft covers them all with
ease, whether only one or a combination, the
The batteries in battery electric vehicles Power Analyzer and Power Quality Analyzer
MOTOR/INVERTER All the data can be stored at the full-sampling
(BEVs) are exposed to conditions not
optimal for batteries. These include extreme
delivers the best results all of the time.

rate and all analysis can be done during the


TESTING BATTERY DEVELOPMENT
temperatures both hot and cold, humidity, as
measurement. Furthermore, the unique well as vibrations and shocks. These all have Cell characterisation, endurance and aging
post-processing functionality enables data an effect on the stability of power delivery and tests, shock and vibration, misuse tests such
Combined motor and inverter testing is the manipulation after the measurement (e.g. efficiency of the battery. as crash tests, short-circuit tests, overheating-,
Power Analyzers domain. It offers a high mathematics, power analysis settings, etc.). overloading-, overcharge- tests, forced
number of input channels for both voltage For instance if phase voltages were falsely EXTENSIVE TESTS discharge tests, impact/crush test, thermal
and current measurements, and provides connected this can be rectified, eliminating This makes it crucial to do extensive tests on misuse,
synchronized data acquisition on all channels. the need to repeat the measurement. batteries: starting from the cell-characteristics
leading up to the complete powertrain of the GENERAL BATTERY TESTING
The Dewesoft R8D Power Analyzer can TYPICAL CONFIGURATION BEV. Detailed analysis requires temperature Voltage and current, power and energy,
measure 8 x 3-phase systems simultaneously 12x Voltage and voltage measurement at multiple points temperature and humidity, pressure and
using a single measurement device. This 12x Current e.g. 50x cell voltage and 50x cell temperature vibration,
enables the measurement of an entire power Power Supply for Current transducer measurements.
system (e.g. electric vehicle, aircraft, ship etc.) 1x Torque The flexible and scalable solution from
completely synchronous. 1x Speed Dewesoft can be configured to encompass
BATTERY CHARGING ANALYSIS
AC/DC charging, charging energy, charge-/
over 1000 channels with many different
discharge efficiency, charging process and
The analyzer combines the functionalities of sensors - all synchronised for a detailed
time, harmonic analysis as well as inductive
motor and inverter testing and analysis as well analysis.
and conductive charging.
as the capability to measure other parameters
such as speed, torque, temperature, video,
GPS, and CAN. BATTERY TROUBLE SHOOTING
Such as voltage drops, voltage commutation
Earlier typical test bed applications required unbalance and inrush currents.
the use of multiple measurement instruments
- Power Analyzer, Scope, Data Logger, CAN IOLITE® R12 (TH) APPLICATIONS
reader etc. The Dewesoft Power Analyzer • Battery monitoring
facilitates the measurement and analysis of all PC • Transient recording
the data that would have been measured with • Charge and discharge analysis
such devices in a single measurement device.
TEMPERATURE MONITORING • Charging profiles
• Energy delivery
BATTERY DC POWER • Efficiency and losses
• Efficiency STORAGE NETWORK • State of charge
SYSTEM ADC • Cell voltages and temperatures
• Power & power quality analysis
• Analyzing 1-12 phase motors
• Raw data analysis
• Transient recording
• Data logging
• Scope
• Vectorscope
• Measurements such as speed, torque,
SIRIUS® XHS
temperature, etc...
Y O U R S O L U T I O N . 19

EV CHARGING ANALYSIS CHARGE AND DISCHARGE EFFICIENCY


During charging and discharging some
TESTBED / IN-VEHICLE USE
Measurements on BEVs under real-drive

CONDUCTIVE AND
energy is lost through heat. The quotient of
the amount of energy that is delivered by the ON-ROAD TESTING conditions require a powerful, mobile, and
extra compact measurement system due to
INDUCTIVE CHARGING battery and the amount of energy that was the constraints in space. The system also needs
Charging whether conductive (plug-in delivered to the battery, can be measured. Dewesoft data acquisition devices offer the to be able to power measurement sensors
charging) or inductive (wireless charging) can Additionally, the efficiency of the drive train modular design and flexibility that is suited for and other auxiliary systems such as screens
be analyzed with Dewesoft data acquisition from the battery over the inverter to the both testbench testing and in-vehicle testing. directly. Dewesoft Power Analyzers well-suited
devices, for both alternating current (AC) and electrical motor, auxiliary power consumption for this application.
direct current (DC). and finally the actual power that arrives at the SUPPLY FOR SENSORS
wheels, can be measured and analyzed. For real-drive tests no auxiliary power is TESTBENCH
With higher switching frequencies of the needed from the vehicle. The Dewesoft battery Test benches use several important interfaces
inverter (up to 150 kHz), high sampling rates packs - which are hot swappable - can power such as CAN, OPC-UA, DCOM, etc. to receive
of up to 15 MS/s ensure that even the fastest the Dewesoft system as well as the current and relay information. The Dewesoft NET
transients can be monitored and analyzed. transducers and other sensors. With SIRIUSi- option provides a remote control feature for
PWR-MCTS2 even zero flux transducers that Dewesoft data acquisition systems, enabling
need up to 20 W per unit can be powered. This you to control the entire test procedure from a
CHARGING PROFILE AND TIME ensures a true measurement without external single PC in the control room.
This types of tests include analyzing the influences.
charging station itself over the different
charging levels. Furthermore, they involve
analyzing the charging process of the battery
starting at the power delivery to the charging
station to bulk, as well as the absorption and
floating stages of charging.

TYPICAL CONFIGURATION
IOLITE® modules
for temperature monitoring
(8x TH, 8x RTD)

SIRIUS® XHS
for inverter monitoring
20 Y O U R S O L U T I O N .

GRID POWER ANALYZER POWER QUALITY ANALYSIS identification of numerous parameters and
The power quality can be analyzed and a faults. These include identifying big loads,
energy management plan for the operation inefficient equipment, standby consumption,
of both energy production plants and peak loads, harmonic interference, different
GRID ANALYSIS systems entering the power grid. However, the
design process and the equipment for such
energy consumers, can be compiled.
Energy management has the objective of
types of transients, estimation of line
parameters, detection of reverse power flows,
smart grids need comprehensive testing. voltage and frequency behaviour and many
optimizing the energy consumption for the
SMART GRID & ENERGY industrial as well as the residential sectors. more.
MANAGEMENT With synchronized and distributed
This includes many activities that will lead
In conventional power supply systems, measurement capability, the Dewesoft Power TYPICAL CONFIGURATION
to a stable decentralized power grid. These
the power is produced in big power Analyzer can measure both power generation SIRIUS R8
activities include reducing the overall energy
generation plants (thermal, nuclear, hydro...) and power consumption at multiple points in 4 x Voltage
consumption, move toward more efficient
and transported via lines to substations, the grid - a solid foundation for the design and 16 x Current
equipment and technologies and thereby
transformers to the customer. planning of smart grids.
also reducing the costs. The Dewesoft Power
Analyzer combined with the ability to measure
In recent years, the trend toward more
multiple points in the grid simplifies the
renewable energy power plants (wind, solar,
etc.) has pushed power supply systems to start
moving away from the centralized model. The
power grid operators are now facing some
inherent challenges.

The systems are designed and built for


centralized supply - not for the variable,
and intermittent power generation from
renewable energy sources, larger loads such as
heat pumps and EV charging stations, which
are causing transients and voltage quality
problems such as dips, swells, or sags.

There is a move toward so-called smart grids,


where producers and consumers of energy High Voltage Direct Current Cable
communicate and interact with each other
to avoid problems. They allow for a greater
increase in the number of renewable energy

POWER AC HVDC POWER


GRID GRID
HVDC AC
• Power and efficiency analysis
• Multiple power modules analyzed
simultaneously SIRIUS® XHS SIRIUS® XHS
• Synchronous distributed
data acquisition
PTP Ethernet PTP Ethernet
• Interaction between power generation
GPS GPS
& consumption
• Estimation of line parameters HVDC
transmission system analysis LOCATION 1 LOCATION 2
Y O U R S O L U T I O N . 21

MOTOR TESTING INVERTER TESTING Type of inverter


Industry inverter
Input
1 to 3~AC
Output
3~AC
MULTIPLE SENSOR MEASUREMENT
The IEC 60076-1 also requires the correction
of the temperature dependant power losses.
Motors have to fulfill higher and higher The Dewesoft Power Analyzer allows for Electric vehicle inverter DC 3 to 12~AC
With the Dewesoft Power Analyzer, the
requirements concerning energy efficiency. comprehensive and highly accurate analysis
Photovoltaic inverter DC 1 to 3~AC measurement of the temperatures for these
Since 2011 all asynchronous motors have to of all kinds of inverters. The combination
Wind power inverter 3~AC 3~AC losses (e.g. winding, oil, ambient, etc.) are
be at least level IE2 according to the IEC 60034. of modular, highly accurate hardware and
easy and completely synchronized to all
Before this standard was established losses powerful software can measure any in- and Electric two-wheeler DC 1 to 12~AC
other measured parameters. The corrected
were considered with 0.5 %. Now they have to output configuration up to a 7-phase AC
power losses can be calculated using the
be determined. system.
comprehensive
EFFICIENCY DETERMINATION • Efficiency analysis POWER TRANSFORMERS DewesoftX math toolbox. Furthermore, the
measurement of the power consumption of
There are two ways of determining the
efficiency of the motor, the direct and • Raw data analysis Power transformer testing as described in the
auxiliary loads (e.g. oil pumps, fan motors,
etc.) is also possible, as well as the sound level
indirect method. The direct method requires • Voltage rise time analysis (dU/dt) international standard IEC 60076, establishes according to the standard IEC 60551.
a measurement of the input power using • Transient recording multiple measurement parameters that must
a Power Analyzer, and a measurement of be done on the unit. With tools such as the
the output power by measurement of the • Filter analysis scope and vector scope, voltage ratios and
TYPICAL CONFIGURATION
8x Voltage, 8x Current
rotational speed and the torque that is applied phase displacements of various primary and
Fundamental frequencies from 0.5 Hz up to 3x Temperature
to the motor. secondary configurations (e.g star, delta,
3 kHz can be analyzed as well as switching Sound
and interconnected star) can be analyzed
frequencies into the multiple hundred kHz Additional Sensor calibration
The indirect method is based upon loss effortlessly.
segregation and then building a sum of all region. The analysis possibilities reach from
the losses. There are six measurement points efficiency determination to a detailed analysis POWER AND
of each switching pulse.
defined in the standard, these are summed POWER QUALITY ANALYSIS
together and subtracted from the power The transient recording functionality with the
input, yielding an output power for efficiency Raw data storing enables detailed analysis
ability to store all signals at the full sampling
determination. and depiction of each individual switching
rate (up to 15MS/s) ensures a detailed
pulse (e.g. transient behavior in the scope). The
analysis. Combining this with the trigger
POWER ANALYZER power quality library automatically calculates
functionality both failure and transient events
The Dewesoft Power Analyzer and the THD, harmonics, etc. with the click of a button.
(including long-term transformer testing)
capability of additional software sensor A transient recording captures voltage peaks
can be captured and analyzed with ease.
calibration guarantees the highest accuracy (e.g. with long cable lengths) or captures
The power quality library enables harmonic
measurement results. current peaks which can be any multiple of the
measurements of voltage and current into
nominal current.
the multiple hundred kHz region. The data
The modular hardware concept is able to can be represented as a percentage of the
measure multiphase (1 to 12-phase) motors High edge steepness of the inverter output (up
fundamental frequency as required for the no-
as well as the mechanical parameters such as to 10 kV/µs) can also create capacitive leakage
load current according to IEC 60076.
speed and torque. Furthermore, additional currents or high motor bearing currents (due
to the parasitic motor- and line capacities).
parameters such as vibration, sound level,
All these factors can harm the motor and
SYMMETRICAL COMPONENTS • Analysis according to IEC 60076
temperature, etc. can also be measured Additionally, the calculation of the zero-
effortlessly. often make the use of filters necessary. The
sequence impedance is required, this • Correction of power losses depending
Dewesoft Power Analyzer is not only capable
function is built into the DewesoftX power on temperature
POWER QUALITY of measuring all these parameters, the analysis
quality library. Power and efficiency analysis
Power Quality analysis (Fundamental of these parameters can be done during the
of transformers requires the highest
• Additional sensor calibration
actual measurement. Using the powerful
Power, Harmonics, THD, etc.) complements
integrated math library in DewesoftX, the
measurement accuracy for the phase angles. • Harmonics
the analysis capabilities. Dewesoft offers
a complete tool for motor analysis for voltage steepness (dU/dt) of every impulse can
This is especially important when analyzing • Symmetrical components
units with a low power factor, which is difficult
effortless analysis of motor efficiency, d/q be determined and statistically classified for
with conventional measurement equipment.
transformation, and efficiency mapping. example.
22 Y O U R S O L U T I O N .

GRID POWER ANALYZER STANDBY POWER by the pulsed current of the power supply
units. Furthermore, input filters often produce
reactive currents which can be a multiple
The IEC 62301, is the internationally of the active current. In older DAQ systems
FAULT/TRANSIENT frequency drives (VDF) and inverters. Power
quality parameters such as system unbalance,
recognised standard for the measurement of
the standby power consumed by consumer
these issues forced measurement ranges
to be set much higher than required by the
harmonics, THD, flicker, rapid voltage changes
RECORDING
electronic devices. It is estimated that pure sinusoidal signal, which decreased the
etc. can also be acquired and analyzed by the household appliances consume between 5 accuracy: The Dewesoft Dual-core technology
click of a button, all power quality parameters and 15 % of the entire power requirement of a (incorporates two 24-bit AD converters in
Power system conditions such as switching are measured according to international standard household, while on standby mode. parallel: One measuring the full input range
operations (capacitor bank connection or standards. Lowering the level of power consumed by and the other measuring 5 % of the range.
disconnection) or some unforeseen system standby power mode can be beneficial for the This ensures high accuracy for both high and
conditions can cause switching transients, DewesoftX TRIGGERING environment as well as financially. low ranges and makes it possible to have a
voltage interruptions, over-voltages, The possibility to acquire the data using the high range and best accuracy simultaneously.
harmonics, etc. These can affect the function of fast-on trigger, slow otherwise option data is MEASUREMENT REQUIREMENTS This technology is revolutionary for standby
power system operation, and lead to outages stored at a reduced rate (min, max, average There are several requirements set for power measurements and reaches never-seen
of delivery lines or even cause a complete and RMS), as soon as an event happens the measurement of the standby power. accuracies.
black-out of the entire power system. the data is stored at full speed enabling all Measurement devices must be able to
power parameters for analysis. In addition, measure very low currents (< 1mA) and TYPICAL CONFIGURATION
The effects of these unwanted conditions it is possible to utilize pre and post times very low power with specified accuracies 1x Voltage, 1x Current
more than often lead to devices seizing to which will store the data at full speed before depending on the wattage (<0.5 W with an Additional current transducer calibration
work or even getting completely destroyed, and after the event for as long as the user- accuracy of 0.01 W, >0.5 W with an accuracy of for 50 or 60Hz
this is especially true for sensitive electronic determined pre measurement. This reduces 2 %). Harmonic analysis up to the 49th order
equipment. The impact of these faults can be the amount of data that is stored which is (2.5 kHz @ 50 Hz fundamental) is required
very expensive, and time-consuming in the handy for long time measurements. and a Data logging capability is strongly
mitigation of these faults. recommended. The testing process requires
The possibility to use math channels to create the measurement of the power supply voltage,
combined trigger conditions offers the user as well as the THD, temperatures etc. all within
• High sampling rate up to 15 MS/s the freedom to determine the trigger events the specified limits. With the Dewesoft Power
• Storing raw data for individual events or combinations thereof. Analyzer all of the required parameters can be
Gain - 1
measured and analyzed.
• Triggering on different channels TYPICAL CONFIGURATION
ADC1
24 bit
ADC2

(analog, digital, math, power, DEWESOFT DUAL-CORE TECHNOLOGY

ISOLATION
SIRIUS R3 Fast

± 1000 V
Signal
200 kS CPU ADC1
input

power quality, etc.)


System
12 x Voltage The biggest challenge with measuring standby
ADC2
12 x Current power is measuring currents with a high
• Analysis of all line frequencies
24 bit

crest factor. The high crest factor is caused Gain - 2

(16,7 Hz, 50 Hz, 60 Hz, etc.)


DewesoftX SOFTWARE 1 MS/s 15 MS/s • DUAL-CORE-ADCR measurement for
Being able to store raw data for analysis and
the diverse triggering functions offered in the
low currents with high crest factor
DewesoftX software makes it easy to identify • Harmonics and THD
and analyze any kind of fault that may occur. • Data logging
The triggers in DewesoftX can be triggered on
any input channel (analog, digital, etc.), as well
as being triggered on mathematical channels
and standard power channels.
Analysis can be done on all line frequencies
(16,7 Hz, 50 Hz, 60 HZ, 400 Hz, 800 Hz ) as
well as variable frequencies for variable
*Comparison of a transient signal captured at 1 MS/s and 15 MS/s.
Y O U R S O L U T I O N . 23

GRID POWER QUALITY POWER QUALITY ANALYZER


Flicker emission is calculated according to the
IEC 61000-21 and is, therefore, able to evaluate
the flicker emission into the grid by wind
ANALYSIS turbines as well as other power generation
are in development.
units.
The different Power Quality parameters OVERVIEW OF THE POWER QUALITY
describe the deviation of the voltage from RAPID VOLTAGE CHANGES STANDARDS SUPPORTED BY
its ideal sinusoidal waveform at a certain Rapid Voltage Changes (RVC), describes DEWESOFT
frequency. The Dewesoft Power Analyzer is voltage changes that changes more than 3.3
rated to the IEC 61000-4-30 Class A acquisition % in a predefined time interval. This is added • IEC 61000-4-30, IEC 61000-4-7,
and measurement devices. When compared to as a supplement to the flicker standard. This is • IEC 61000-4-15, Requirements for Power
other Power Quality Analyzers Dewesoft offers a special calculation in DewesoftX, calculating Quality Analyzers, Calculation of Harmonics,
a more detailed power analysis considering the maximum voltage drop (Dmax), the Flicker etc.
its ability to store raw data, powerful post stationary deviation after the voltage drop • EN50160, EN 50163, IEE519, IEC 61000-2-4,
processing capability, analyzing behavior (dc) and the specific time where the voltage etc. Power Quality limits of public grid,
at faults, and the ability to calculate many dropped under 3.3 % of the nominal voltage as industries and railway applications
additional parameters in the extensive math specified in the IEC 61000-4-15. • IEC 61400-21, IEC 61400-12, FGW-TR3,
library. An overview of some of the analyses VDE-AR4105 etc. Power Quality Analysis of
that can be done at the click of a button is UNBALANCE - SYMMETRICAL Renewables
presented below. COMPONENTS • IEC 61000-3-3, IEC 61000-3-11 EMC of
Normally an electric power system operates voltage changes and Flicker
HARMONIC ANALYSIS in a balanced, symmetrical three-phase • IEC 61000-3-2, IEC 61000-3-12 EMC of
With the DewesoftX software harmonics sinusoidal steady-state mode. Unbalance in harmonic current
for voltage and current as well as active such a system means that the voltages and/or
and reactive power can be analyzed up to the currents are unsymmetrical. The Dewesoft
the 3000th order. All flicker calculations are Power Analyzer has the ability to measure over
implemented according to the IEC61000- 50 different parameters for a comprehensive
4-7 standard, and the number of sidebands analysis of an unbalanced system (positive,
and halfbands are user-definable. For higher negative and zero sequences ). This enables
frequency analysis harmonics can be grouped you to calculate multiple parameters for
into 200 Hz and 2 kHz bands up 150 kHz. The voltage, current, active, reactive, and apparent
calculation of THD (Total Harmonic Distortion) power. Unbalanced systems can cause the
for voltage and current up to the 3000th flow of current in the neutral line, heating of
order as well as interharmonics rounds up the conductors and equipment which decreases
harmonic analysis with Dewesoft. efficiency, as well as increasing harmonic
currents.
FFT WATERFALL ANALYSIS
In addition to the FFT analysis, the harmonic FREQUENCY DEVIATION
FFT analysis can also be depicted in either a Frequency deviation is caused by the
2D or 3D FFT Waterfall plot. The visualization connection and disconnection of generation • Harmonics and THD
is user-definable to be either linear or units and loads to the grid. These can have
logarithmic, 2D or 3D, and can also be severe consequences to the grid stability and up the 3000th order
sorted by harmonic order or frequency. This even lead to a blackout. The Dewesoft Power • Interharmonics & higher frequencies
visualisation is especially useful for the analysis Analyzer can be used to monitor the frequency
of variable-frequency drives (VFD).
• Flicker, Flicker emission, RVCs
on the grid, as well as for testing the frequency
behavior of power generation units when they • FFT, harmonic FFT, waterfall-FFT
FLICKER AND FLICKER EMISSION • Symmetrical components
The Dewesoft Power Analyzer measures flicker
according to the IEC 61000-4-15 standard.
24 Y O U R S O L U T I O N .

POWER QUALITY ANALYZER RENEWABLE ENERGY


applications.
MAINTENANCE AND SAFETY
These tests are done for user safety and
liability issues. Handling electrical equipment SOLUTION
ELECTRICAL safety standards.
FAULT DETECTION
could pose serious hazards such as electric
shock, fires, and even explosions. Furthermore, To maintain a stable and secure power grid
EQUIPMENT TESTING & Testing is the only way that faults in the
manufacturing can be identified before
these tests are done to ensure the quality of
the equipment and that they adhere to strict
operation with a large share of renewable
generation, certain standards must be adhered
the equipment is released into the field.
MAINTENANCE - BOTH Furthermore, it ensures that the electrical
equipment has the ability to be used for its
to. DewesoftX provides measurements
supported with reporting tools according to
FGW-TR3 and VDE-AR4105.
BIG AND SMALL intended function. Tests such as electrical
stability testing, fan, and pump testing, circuit
breaker and switch testing, filter analysis,
ACTIVE & REACTIVE POWER
ELECTRICAL TESTING castor testing, rod-drop testing, harmonics
The power analysis module provides a
The modular hardware design and the comprehensive set of power measurements.
analysis according to IEC 61000-3-2/-12,
powerful DewesoftX software offer a complete Combined with the statistics function the user
voltage changes according to IEC 61000-3-
solution for the testing of various types of can apply various statistic operations such as,
3/-11, and CE conformity of electrical devices
electrical equipment. The testing of modern minimum, maximum, averaged values, etc.
(Harmonics/Flicker) to name just a few.
electrical equipment in this day and age to any of the output power measurement
extends well beyond just fundamental testing. channels. This, for instance, allows you to
It encompasses so much more ranging capture maximum power, or check output
from safety, energy requirements, electrical power deviation, for a specific time interval.
performance, and operation to name just a
few. Monitoring In-Rush currents, voltage Power plants operate at different setpoints
transients, leakage, load tests, polarity for the injection of active and reactive power.
testing, currents, harmonics, and power Testing that the regulation controller behaves
quality analysis are just a few of the possible correctly under changing demand conditions
(i.e. if power reduction is within tolerance
as the frequency increases if reactive power
• Electrical stability testing injection increases as the voltage drops, etc.)
• Fan and pump testing is essential for ensuring a reliable network
operation with integrated renewables.
• Circuit breaker and switch testing DewesoftX enables you to perform
• Filter analysis measurement campaigns storing the data for
• Castor testing later in-depth analysis.

• Rod-drop testing POWER QUALITY


• Harmonics analysis Implemented parameters enable you to check
equipment compliance to power quality
according to IEC 61000-3-2/-12 standards such as IEC 61400-21.
• Voltage changes
according to IEC 61000-3-3/-11 Special functionalities are implemented to
calculate the flicker parameters (such as flicker
• CE conformity of electrical devices coefficient, flicker step factor, voltage change
(Harmonics/Flicker) factor, etc.). Switching operations are events
that can induce transient behavior, such as
oscillations that are not correctly dumped. The
power quality analyzer implements period
value calculations, for much of the calculations
Y O U R S O L U T I O N . 25

(P, Q, S, Urms, Irms, ...). Period values enable


you to minimize the averaging effect and are WIND POWER TESTING local and international standards and that all
the power quality emissions are within the • Power performance
ideal for the observation of transients. set limits. Other tests are also possible with • Transformer testing
TEST REQUIREMENTS the Dewesoft Power Analyzer these include
The Dewesoft power quality analyzer These tests include electrical measurements switching operations and low and high voltage
• Structural and sound level testing
also provides the classical power quality such as power performance, power quality, ride troughs. • Transformer, generator
parameters such as harmonics (up to 3000), and behavior at faults to name just a few.
and inverter analysis
interharmonics, and multiple harmonic Additionally, mechanical measurements TYPICAL CONFIGURATION
distortion measurements such as THD. need to be performed these include testing 3 x Voltage
structural integrity, power generation sound 3x Current
BEHAVIOUR AT FAULTS level, and many more. Additional:
Since our devices enable distributed and Wind direction, Windspeed,
synchronous data acquisition, they are ideal Power performance analysis according to the Temperature, pressure, etc.
for analyzing behavior at faults, such as IEC 61400-12, in order to determine the power
tripping of a protection relay, recloser, etc. Raw performance of the power plant requires
data from multiple points in the grid enables measurements of the voltage and currents
analysis of fault length, type, and specification but also includes measurements such as wind
of short-circuit current. speed, wind direction, and temperatures. The
averaged values from these measurements are
Particularly in distribution networks, sudden classified in BINS, from these BINS indicators
voltage dips can occur, which can lead to such as the performance factor (CP), and
unstable operation or even disconnection the annual energy production (AEP) can be
of certain generator types, the loss of determined.
generation causes the voltage to drop
even further and this can, in turn, cause a DewesoftX VOLTAGE,
cascading disconnection of other generators The flexible measurement screen visualization CURRENT
on the grid. The generators thus need to be setup in the DewesoftX software enables users
equipped with the so-called Low-Voltage to analyze graphs such as power factor over
Ride Through (LVRT) capability, which ensures wind speed, or tabling the different BINS, wind
that a generator can withstand certain supply speed over time, power performance factor,
WIND SPEED,
voltage dips. DewesoftX supports the most and any other required power parameters
WIND DIRECTION
common standards (IEC 61400-21, FGW TR3) round up the complete measurement
that describe the testing procedures for LVRT package.
RPM
capability.
COMMISSIONING THE POWER PLANT 100 m
Before commissioning wind power plants
STRAIN,
must fulfill the local requirements from local
VIBRATION
regulators and grid operators. The FGW-TR3
• Power analysis for AC and DC and the IEC 61400-12 are the most popular
• Raw data storing standards worldwide for testing the power
quality behavior and behavior at faults of
(Switching operations, faults) wind power plants. Parameters such as flicker
• Power quality analysis emission, flicker coefficient, harmonics,
interharmonics, and higher frequency
(Harmonics, Interharmonics, ect.) emissions are just a few of the test results
• Flicker, -emission, -coefficient, needed.
-step factor
These results are handed to the grid operators 170 m
• Symmetrical components, for confirmation that the wind power plant
period values fulfils all the necessary requirements set by
26 Y O U R S O L U T I O N .

POWER QUALITY ANALYZER


LIGHTING DEVICES The extensive math library is able to calculate
efficiency, energy consumption, and many
other parameters. For example, the current
The Dewesoft Power Analyzer is able to through a fluorescent lamp can be determined
measure both Efficiency and Power Quality as via the math library out of the secondary
well as do a full analysis of Lighting systems current and the cathode current.
using a single instrument. This is a new and
innovative lighting test experience. The
trend towards energy-saving lighting makes TYPICAL CONFIGURATION
fluorescent and LED lights more and more SIRIUS XHS
popular. In comparison to incandescent light 3x Voltage
bulbs, both fluorescent and LED technologies 3x Current
have higher efficiencies. 1x Low Voltage input for luminance
meter
DIFFERENT CHARACTERISTICS
These technologies have different
characteristics that have to be taken into
account when doing measurements on them.
For instance, fluorescent lighting makes use of
ballast units which make use of high switching
frequencies of up to 150 kHz.

LEDs are more energy-efficient than


incandescent bulbs, but they also have some
disadvantages. Using a light-emitting diode
that produces a non-linear load can impact
the power quality negatively, by introducing
noise into the grid. This puts unwanted strain
on the AC circuit.

DewesoftX
The high Sample Rate (15 MS/s) of the
Dewesoft Power Analyzer guarantees reliable
analysis of any kind of lighting system. The
power quality library automatically calculates
parameters such as Harmonics, THD, Flicker,
etc. • High switching frequencies
• High bandwidth Harmonics
• THD
• Flicker
• Energy
• Efficiency
Y O U R S O L U T I O N . 27

OVERVIEW
POWER ANALYSIS SOFTWARE FUNCTIONALITY
Functionality Dewesoft Power Analyzer Functionality Dewesoft Power Analyzer

Power Analysis for DC and AC  Power Analysis 


P, Q, S, PF, cos phi, D (Distortion), DH (Harmonic distortion), QH (reactive power of
Power Analysis harmonics) (for each phase and total)
Power Quality Analysis 
Fundamental Power P_H1, Q_H1, S_H1, cos phi_H1, phi_H1 (for each phase and total) Database Storing 
Voltage and Current RMS, RM, AVE (star and delta)
Energy Calculation Total, positive and negative (e.g. Recuperation)
Post Processing 
Efficiency 
Math Library 
DC, 1-phase, 2-phase, 3-phase delta, 3-phase star, 3-phase V, 3-phase Aron, 6-phase Data logging - Raw data storing 
Wiring Schematics (R2, R4, R8), 7-phase (R2, R4, R8, 12-phase (R4, R8)) (data Storing at Full Sampling rate of 15 MS/s per channel)

Scope 
Star-Delta Calculation (waveform and RMS values) (up to 8 graphs in one diagram, Zoom In- and Out)

Frequencies 16.7 Hz, 25 Hz, 50 Hz, 60 Hz, 400 Hz, 800 Hz, Variable from 0.5 Hz up to 3 kHz
Vector Scope 
Frequency Source Voltage, current, external (1-, 2-, 3-phase systems)

Period Values U, I, P, Q, S, symmetrical components for ½, 1, 2 or 4 periods and selectable Overlap up


to 99 %
FFT 
(up to ½ of Sampling Rate)
Number of Cycles for Power
Calculation 5 - 12
Harmonic FFT 
(up to ¼ of Sampling Rate)
Selectable - starting from 1ms , Multiple Averaging (e.g. 20 ms, 60 s, 600 s) possible
Power Averaging
Transient Recording 
(up to 15 MS/s)
Triggering Channels Analog, Digital, Counter, Math, Power, etc.

POWER QUALITY Triggering options


Simple edge (rising, falling), Window (two-levels: entering, leaving), Pulsewidth
(longer or shorter than duration), Window and Pulsewidth, Slope Trigger (rising or
falling slope with steepness)
Functionality Dewesoft Power Analyzer

Harmonics up to 150 kHz for voltage, current, active-, reactive power, phase angle
(according to IEC61000-4-7) and impedance

Variable Sidebands and Half Sidebands


(according to IEC61000-4-7)  Higher Frequencies up to 150 kHz (grouping in 200 Hz bands, 2 kHz bands optional available)
(according to IEC61000-4-7)
Harmonic Smoothing Filter (according to
IEC61000-4-7)  Flicker (according to IEC61000-4-15) selectable PST and PLT

Interharmonics (according to IEC61000-4-7)  Flicker Emission (according to IEC61400-21) 


Rapid Voltage Changes (according to selectable steady state and hysteresis
Total Harmonic Distortion (THD) IEC61000-4-15)
Voltage and current (Total, odd and even) - selectable up to 150 kHz
(according to IEC61000-4-7)
Symmetrical Components (according to Zero-, positive- & negative system for voltage and current (absolute or
IEC61000-4-30) relative to fundamental)
Total Interharmonic Distortion (TIHD) and
K-factor Voltage and current (Total, odd and even) - selectable up to 150 kHz Additional Symmetrical Components
(according to IEC61000-4-7) Active and reactive parts for zero-, positive- & negative system
(according to IEC61400-21)
DEWESOFT® WORLDWIDE: SLOVENIA, Austria, Belgium, Brazil, China, Denmark, France, Germany, Hong Kong,
India, Italy, Mexico, Russia, Singapore, Sweden, UK, USA and PARTNERS IN MORE THAN 50 COUNTRIES