● Run CPU

TCD - 099 Double Input Signal Error


Point of Application ● DI120-150
Detection ○ Speed CPU ● DI180-420

[Meaning]When double input signal data from Run CPU and Speed CPU is maintained [Actions to take]
for 200ms in different status. This error detection function is to
prevent mis-operation caused by double input data defect. 1. Check connector insertion status between PCBs
1) MA : DOC_200 ↔ DOR-230 PCB
[Diagram] 2) VE : DOR-230 ↔ DPC-1XX(or DPC-300) PCB
3) INA, INB, and INC connector in DOR-230 PCB insertion
status
Input DOR-230 Run CPU Speed CPU Related
Signal Conn. No Input ADD Input ADD Circuit Diagram 2.Compare double input signal status.
IS_C1ST INA/15 0x501F 3x02227 I/O C[1] Compare and check the input signal status of Run and
IS_C1BR INA/16 0x501E Speed CPU shown in the left table.
IS_R5SC INC/1 0x5000
3.Check input buffer status.
IS_DL INA/14 0x5006
Operate, by force, the devices that consist the diagrams
IS_R2LD INC/2 0x5019 listed in the left table to check input signal variation.
IS_LDCX INB/1 0x5022 3x02228 I/O C[2] If no signal changes have been founded replace DOR-230
IS_DOLS INB/2 0x5026 PCB as the DPRAM elements between input buffer or Run
IS_CDS INB/4 0x5018 3x02219 CTL C[4] and Speed CPU are possibly been defected.