Precision
Intra-assay
Determined by by using 10 replicates of specimen 500 pg/mL IL-6, CV ≤ 15%
Inter-assay
Determined by using 3 replicates for each of three lots using IL-6 specimen levels
at 500 pg/mL, CV ≤ 15%.
Linearity
A serial concentration of IL-6 controls at 10 pg/mL, 20 pg/mL, 50 pg/mL, 150 pg/
mL, and 500 pg/mL and 800 pg/mL were each tested for three times the CV is r ≥
0.9914
Method comparison
A comparison of the Infinosis™ IL-6 assay (y) with the Roche Elecsys IL-6 assay
(x) using 217 clinical samples gave the correlation: r=0.9650