You are on page 1of 2

IN0177022105V1

POC T3 infinosis™
Triiodothyronine
REF: IN017702 25 tests • Don’t use Test Cassette if its Lot does not match with ID Chip that is inserted Step 5: Testing
onto the instrument. Put the Test Cassette on the operation platform. 15 minutes later, insert the Test
Intended use • The Infinosis™ T3 assay should be used with Infinosis™ 2020 FIA analyzer. Cassette onto the Cassette Holder and click “Test”. The result will show in the
The Infinosis™ T3 is an immunofluorescence-based lateral flow assay for the in • The tests should be applied by professionally trained staff working in certified display and print out when click “Print”.
vitro quantitative determination of triiodothyronine (total T3) in human serum or laboratories at some remove from the patient and clinic at which the sample is
plasma. taken by qualified medical personnel. Please refer to the Infinosis™ 2020 FIA analyzer Operation Manual for details.
• Infinosis™ T3 assay is single use only. Do not re-use it.
Summary1-7 • The Test Cassette and Analyzer should be used away from vibration and Limitations - interference
Triiodothyronine (T3) is the hormone principally responsible for the development magnetic field. During normal usage, the Test Cassette may introduce minute • The assay is unaffected by icterus (bilirubin < 600 μmol/L or < 35 mg/dL),
of the effects of the thyroid hormones on the various target organs. T3 (3,5,3’- vibration, which should be regarded as normal. hemolysis (Hb < 0.559 mmol/L or < 0.9 g/dL), lipemia (Intralipid < 1200 mg/dL),
triiodothyronine) is mainly formed extrathyroidally, particularly in the liver, by • Use separate clean pipette tips and detector buffer vials for different specimens. and biotin < 94 nmol/L or < 23 ng/mL.
enzymatic 5’-deiodination of T4. Accordingly, the T3 concentration in serum is The pipette tips and detector buffer vials should be used for one specimen only. • Criterion: Recovery within ± 10 % of initial value.
more a reflection of the functional state of the peripheral tissue than the secretory • Do not smoke, eat, or drink in areas in which specimens or kit reagents are • Heterophilic antibodies and rheumatoid factors in samples may interfere with
performance of the thyroid gland.
 handled. test results. Heterophilic antibodies in human serum can react with reagent
A reduction in the conversion of T4 to T3 results in a decrease in
 • Blood specimens, used test cassettes, pipette tips and detector buffer vials are immunoglobulins, interfering with in vitro immunoassays. Patients routinely
the T3 concentration. It occurs under the influence of medicaments such as potentially infectious. Proper laboratory safety techniques, handing and exposed to animals or animal serum products can be prone to this interference
propanolol, glucocorticoids or amiodarone and in severe non-thyroidal illness disposal methods should be followed in accordance with standard procedures and anomalous values may be observed. Additional information may be
(NTI), and is referred to as “low T3 syndrome”. As with T4, over 99 % of T3 is and relevant regulations observed by microbiological hazard materials. required for diagnosis. This kind of samples is not suitable to be tested by this
bound to transport proteins. However, the affinity of T3 to them is around 10-fold • The results should be interpreted by the physician along with clinical findings assay.
lower. and other laboratory test results. • Performance of this test has not been established with neonatal samples.
The determination of T3 is utilized in the diagnosis of T3-hyperthyroidism, the • The test will be applied on a routine basis and not in emergency situations. • Serum T3 concentration is dependent upon a multiplicity of factors:
detection of early stages of hyperthyroidism and for indicating a diagnosis of hypothalamus gland function and its regulation, TBG concentration, and the
thyrotoxicosis factitia. Storage and stability binding of T3 to TBG. Thus, total T3 concentration alone is not sufficient to
• Store the sample buffer at 4-30°C. The buffer is stable up to 24 months. assess clinical status.
Test principle • Store Infinosis™ test cassette at 4-30°C, shelf life is up to 24 months. • A decrease in total T3 values is found with protein-wasting diseases, certain
• Test cassette should be used within 1 hour after opening the pack. liver diseases and administration of testosterone, diphenylhy- dantoin or
Competitive principle. Total duration of assay: 25 minutes
salicylates. A table of interfering drugs and conditions, which affect total T3
Sample is added to the sample well of the test, the fluorescence-labeled detector
Specimen collection and preparation values, has been compiled by the Journal of the American Association of
T3 antibodies bind to T3 antigens in blood specimen and form immune
Clinical Chemists.
complexes. As the complexes migrate on the nitrocellulose matrix by capillary • The test can be performed with either serum or plasma.
action, it can’t be captured by T3 antigens that have been immobilized on test • Collect serum samples in accordance with correct medical practices.
strip, otherwise the excess unbound fluorescence-labeled detector T3 antibodies • Using standard phlebotomy procedure, collect a venipuncture whole blood
Limits and ranges
are captured. Thus the more T3 in blood, the less unbound fluorescence-labeled specimen using a blood collection tube. If collecting plasma use a blood Measuring range
antibodies accumulated on test strip. Signal intensity of detector T3 antibodies collection tube containing suitable anticoagulant (EDTA recommended). 0.45-6.0 ng/mL (defined by the lower detection limit and the maximum of the
reflect the amount of antigens and are processed in the Infinosis™ FIA system to master curve). Values below the detection limit are reported as < 0.45 ng/mL.
• Separate the serum/plasma from blood as soon as possible to avoid hemolysis.
determine the T3 concentration in blood. Values above the measuring range are reported as > 6.0 ng/mL
• Test should be performed immediately after the specimens have been collected.
Do not leave the specimens at room temperature for prolonged periods.
Reagents Specimens may be stored at 2-8°C for up to 3 days. For long-term storage, Lower limits of measurement
specimens should be kept below -20°C. Lower detection limit
Materials provided
Lower detection limit: 0.4 ng/mL

• Test cassette, 25 pcs, individually packaged The detection limit represents the lowest analyte level that can be distinguished
• ID chip, 1 pcs Quality control
from zero. It is calculated as the value lying two standard deviations above that of
• Sample buffer A, 1 vial, 2 mL Each Infinosis™ T3 test cassette contains internal control that satisfies routing
the lowest standard (master calibrator, standard 1 + 2 SD, repeatability study, n =
• Sample buffer B, 1 vial, 1 mL quality control requirements. This internal control is performed each time a patient
21).
• Centrifuge tube, 25 sample is tested. This control indicates that the test cartridge was inserted and
• IFU, 1 copy read properly by Infinosis™ 2020 FIA analyzer. An invalid result from the internal
Expected values
control causes an error message on Infinosis™ 2020 FIA analyzer indicating that
Materials required (but not provided) the test should be repeated. 0.8-2.11 ng/mL
These values correspond to the 2.5th and 97.5th percentiles of results obtained
• TSH control (Infinosis™ control is recommended)
from a total of 352 healthy test subjects examined.
• Infinosis™ 2020 FIA analyzer Test procedure
We have not studied the reference intervals in children, adolescents and pregnant
• Transfer pipette set (100 μL size) Refer to Infinosis™ 2020 FIA analyzer Operation Manual for the complete
women.
• Centrifuge (for plasma and serum only) instructions on use of the test. The test should be operated in room temperature.
Each laboratory should investigate the transferability of the expected values to its
• Timer Step 1: Preparation own patient population and if necessary determine its own reference ranges.
Check/insert ID Chip into the analyzer.
Precautions and warnings
Step 2: Sampling Specific performance data
• For in vitro diagnostic use only. Add 80 μL of Sample buffer A, 30 μL of Sample buffer B, and 20 μL of plasma/ Representative performance data are given below. Results obtained in individual
• Carefully follow the instructions and procedures described in this instructions serum to the Centrifuge tube.
before testing. laboratories may differ.
• The test cassette should remain in its original sealed pouch until ready to use. Step 3: Mixing
Do not use it if the pouch is damaged or the seal is broken. Mix the specimen well with buffer A and B for 10 minutes at room temperature Precision
• Do not use reagents beyond the labeled expiry date. (18-25℃) Intra-assay
• Do not mix or use components from kits with different Lots. Step 4: Loading Determined by using 10 tests in the same batch to test with T3 control, CV ≤ 15%
Add 80 μL of sample mixture and load it onto the sample well of the test cassette. Inter-assay

202105 V1 English 1/2 Infinosis™ T3/IFU


IN0177022105V1

POC T3 infinosis™
Triiodothyronine
Determined by using 3 tests in 3 random and continuous batches to test with T3
control, CV ≤ 20%

Method comparison
A comparison of the Infinosis™ T3 assay (y) with the Roche Elecsys T3 (x) using
diasino
clinical samples gave the following correlation: DiaSino Laboratories Co., Ltd
No.68, Jingnansi Road, National Eco & Tech Development Area
Number of samples measured: 312 Zhengzhou, China. 450000
Linear regression Technical Support: ts@diasino.com
www.diasino.com
y = 1.0309X - 0.2539
r = 0.9855
Analytical specificity
For the antibody derivative used, the following cross-reactivities were found: D-T3
100 %; L-T4 < 0.18 %; D-T4 < 0.18 %; L-rT3 < 0.05 %; L-T2 < 0.9 %.

References
1. Wheeler MH, Lazarus JH. Diseases of the Thyroid. London, Glasgow,
Weinheim, New York, Tokyo, Melbourne, Madras: Chapman and Hall Medical,
1994:107-115.
2. Pfannenstiel P, Saller B. Schilddrüsenkrankheiten Diagnose und Therapie.
Berliner Medizinische Verlagsanstalt GmbH, 1995;2:30-32,60-62.
3. Fisher DA. Physiological variations in thyroid hormones; physiological and
pathophysiological considerations. Clinical Chemistry 1996;42:135-139.
4. Tietz NW. Clinical Guide To Laboratory Tests. 3rd ed. Philadelphia, Pa: WB
Saunders Co, 1995:612.
5. Surks MI, Chopra IJ, Mariash CN, Nicoloff JT, Solomon DH. American Thyroid
Association guidelines for use of laboratory tests in thyroid disorders. JAMA
1990;63:1529-1532.
6. Becker DV, Bigos ST, Gaitan E, Morris JC, Rallison ML, Spencer CA, et al.
Optimal use of blood tests for assessment of thyroid

function (letter). JAMA 1993;269:273.
7. Klee GG. Clinical usage recommendations and analytic performance goals for
total and free triiodothyronine measurements. Clinical Chemistry
1996;42:155-159.

In vitro diagnostic Refer to instruction


use for use

Expiry date Manufacturing date

Batch number Test per kit

Catalog number Do not re-use

Store at 4-30℃ Manufacturer

202105 V1 English 2/2 Infinosis™ T3/IFU

You might also like