You are on page 1of 2

IN0477032105V1

POC CK-MB infinosis™


MB isoenzyme of creatine kinase
REF: IN047703 25 tests • The Infinosis™ CK-MB assay should be used with Infinosis™ 2020 FIA Step 3: Mixing
analyzer. Mix the specimen well with buffer for 5-10 seconds by tapping or inverting the
Intended use • The tests should be applied by professionally trained staff working in certified tube.
The Infinosis™ CK-MB is an immunofluorescence-based lateral flow assay for the laboratories at some remove from the patient and clinic at which the sample is Step 4: Loading
in vitro quantitative determination of the MB isoenzyme of creatine kinase (CK- taken by qualified medical personnel. Add 80 μL of sample mixture and load it onto the sample well of the test cassette.
MB) in human whole blood, serum, or plasma. • Infinosis™ CK-MB assay is single use only. Do not re-use it.
Step 5: Testing
• The Test Cassette and Analyzer should be used away from vibration and
Put the Test Cassette on the operation platform. 15 minutes later, insert the Test
Summary1-3 magnetic field. During normal usage, the Test Cassette may introduce minute
Cassette onto the Cassette Holder and click “Test”. The result will show in the
Creatine kinase (CK) is a dimeric enzyme which occurs in four different forms: a vibration, which should be regarded as normal.
display and print out when click “Print”.
mitochondrial isoenzyme and the cytosolic isoenzymes CK‐MM. • Use separate clean pipette tips and detector buffer vials for different specimens.
The determination of CK‐MB mass in serum is an important element in the The pipette tips and detector buffer vials should be used for one specimen only. Please refer to the Infinosis™ 2020 FIA analyzer Operation Manual for details.
diagnosis of myocardial ischemia, e.g. in acute myocardial infarction, the onset of • Do not smoke, eat, or drink in areas in which specimens or kit reagents are
cardiac symptoms and can remain detectable over a lengthy myocarditis, etc. CK‐ handled. Limitations - interference
MB is detectable in the blood about 3‐8 hours after the onset of cardiac symptoms • Blood specimens, used test cassettes, pipette tips and detector buffer vials are
• This test has been developed for testing human whole blood, serum, plasma
and can remain detectable over a lengthy period of time, depending on the course potentially infectious. Proper laboratory safety techniques, handing and specimen only.
of the condition. disposal methods should be followed in accordance with standard procedures
• The results of Infinosis™ CK-MB should be evaluated with all clinical and
CK‐MB may also appear in other clinical conditions, e.g. in rhabdomyolysis and relevant regulations observed by microbiological hazard materials. laboratory data available. If CK-MB test results do not agree with the clinical
andstroke. Within the scope of laboratory diagnostics, the determination of total • The results should be interpreted by the physician along with clinical findings evaluation, additional tests should be performed.
CK, troponin T and/or myoglobin can contribute to the differentiation of these and other laboratory test results.
• The false positive results may come from cross-reactions with some similar
clinical pictures. • The test will be applied on a routine basis and not in emergency situations. antibodies in blood, and similar epitopes from non-specific components in blood
The sensitivity of a CK‐MB determination is dependent upon the time at which a capturing fluorescent labeled antibodies.
sample was taken. Follow‐up assays are therefore meaningful. Storage and stability • The false negative results may from some unknown substance blocking epitope
• Store the Sample buffer at 4-30°C. The buffer is stable up to 24 months. adhering antibodies, unstable or degenerated CK-MB that cannot be identified
Test principle • Store Infinosis™ test cassette at 4-30°C, shelf life is up to 24 months. due to prolonged time and temperature and storage condition of sample and
Sandwich principle. Total duration of assay: 15 minutes • Test cassette should be used within 1 hour after opening the pack. reagent.
Sample is added to the sample well of the test, then the fluorescence-labeled • Other factors may interfere with Infinosis™ CK-MB and may cause erroneous
detector anti-CK-MB antibody binds to CK-MB antigen in blood specimen. As the Specimen collection and preparation results. These include technical or procedural errors, as well as additional
sample mixture migrates on the nitrocellulose matrix of test strip by capillary The test can be performed with either serum or plasma or whole blood. substances in blood specimens.
action, the complexes of detector antibody and CK-MB are captured to anti-CK- Whole Blood Collected by Venipuncture: For diagnostic purposes, the results should always be assessed in conjunction
MB antibody that has been immobilized on test strip. • Using standard phlebotomy procedure, collect a venipuncture whole blood with the patient’s medical history, clinical examination and other findings.
The more CK-MB antigen is in blood specimen, the more complexes are specimen using a blood collection tube with suitable anticoagulant (EDTA
accumulated on test strip. Signal intensity of fluorescence of detector antibody recommended) Limits and ranges
reflects amount of CK-MB captured and Infinosis™ FIA analyzer shows CK-MB • It is recommended that specimens should be tested immediately. Do not leave Measuring range
concentrations in blood specimen. The default results unit of Infinosis™ CK-MB the specimens at room temperature for prolonged periods. If the specimens are 1.0- 100 ng/mL (defined by the lower detection limit and the maximum of the
test is displayed as x ng/mL from Infinosis™ FIA analyzer. not tested immediately, they may be stored at 2-8°C. master curve). Values below the lower detection limit are reported as < 1.0 ng/mL.
• It’s not suitable to test the whole blood samples which have been stored at Values above the measuring range are reported as > 100 ng/mL.
Reagents 2-8°C for more than 2 days.
Materials provided Serum and Plasma: Lower limits of measurement
• Test cassette, 25 pcs, individually packaged • Separate the serum/plasma from blood as soon as possible to avoid hemolysis. Lower detection limit
• ID chip, 1 pcs • Test should be performed immediately after the specimens have been collected. Lower detection limit: 1.0 ng/mL

• Sample buffer, 25 vials • Do not leave the specimens at room temperature for prolonged periods. The detection limit represents the lowest analyte level that can be distinguished
• IFU, 1 copy Specimens may be stored at 2 - 8℃ for up to 3 days. For long-term storage, from zero. It is calculated as the value lying two standard deviations above that of
specimens should be kept below -20℃. the lowest standard (master calibrator, standard 1 + 2 SD, repeatability study, n =
Materials required (but not provided) 21).
• Infinosis™ 2020 FIA analyzer Quality control
• Transfer pipette set (100 μL size) Each Infinosis™ CK-MB test cassette contains internal control that satisfies Expected values
• Alcohol pads routing quality control requirements. This internal control is performed each time a Reference range
• Centrifuge (for plasma and serum only) patient sample is tested. This control indicates that the test cartridge was inserted <5.0 ng/mL
• Timer and read properly by Infinosis™ 2020 FIA analyzer. An invalid result from the Expected values may vary with age, sex, diet and geographical location. Each
internal control causes an error message on Infinosis™ 2020 FIA analyzer laboratory should determine its own expected values as dictated by good
Precautions and warnings indicating that the test should be repeated. laboratory practice.
• For in vitro diagnostic use only.
• Carefully follow the instructions and procedures described in this instructions Test procedure Specific performance data
before testing.
Refer to Infinosis™ 2020 FIA analyzer Operation Manual for the complete Representative performance data are given below. Results obtained in individual
• The test cassette should remain in its original sealed pouch until ready to use.
Do not use it if the pouch is damaged or the seal is broken. instructions on use of the test. The test should be operated in room temperature. laboratories may differ.
• Do not use reagents beyond the labeled expiry date. Step 1: Preparation
• Do not mix or use components from kits with different Lots. Check/insert ID Chip into the analyzer. Precision
• Don’t use Test Cassette if its Lot does not match with ID Chip that is inserted Intra-assay
Step 2: Sampling
onto the instrument. Determined by by using 10 replicates of specimen of 8.0 ng/mL CK-MB
Add 20μL of whole blood, serum or plasma to the buffer tube.
CV ≤ 15%

202105 V1 English 1/2 Infinosis™ CK-MB/IFU


IN0477032105V1

POC CK-MB infinosis™


MB isoenzyme of creatine kinase
Inter-assay
Determined by using 3 replicates for each of three lots using CK-MB specimen
levels at 8.0 ng/mL, CV ≤ 15%.

Linearity diasino
A serial concentration of CK-MB controls at 2.0 ng/mL, 5.0 ng/mL, 10 ng/mL, 20 DiaSino Laboratories Co., Ltd
ng/mL, 50 ng/mL, 80 ng/mL were each tested for three times the CV is No.68, Jingnansi Road, National Eco & Tech Development Area
r ≥ 0.9978 Zhengzhou, China. 450000
Technical Support: ts@diasino.com
Method comparison www.diasino.com
A comparison of the Infinosis™ CK-MB assay (y) with the Roche CARDIAC CK-
MB STAT assay (x) using 121 clinical samples gave the correlation: r=0.9520

References
1. Rozenman Y, Gotsman MS. The earliest diagnosis of acute myocardial
infarction. Annu Rev Med 1994;45:31-44.
2. Adams JE, Abendschein DR, Jaffe AS. Biochemical markers of myocardial
injury: Is MB creatine kinase the choice for the 1990s? Circulation
1993;88:750-763.
3. Apple FS. Diagnostic markers for detection of acute myocardial infarction and
reperfusion. Laboratory Medicine 1992;23(5):297-322.

In vitro diagnostic Refer to instruction


use for use

Expiry date Manufacturing date

Batch number Test per kit

Catalog number Do not re-use

Store between
Manufacturer
4-30℃

202105 V1 English 2/2 Infinosis™ CK-MB/IFU

You might also like