You are on page 1of 2

IN0477012104V1

POC Troponin I infinosis™


Cardiac Troponin I
REF: IN047701 25 tests • The tests should be applied by professionally trained staff working in certified Mix the specimen well with buffer for 5-10 seconds by tapping or inverting the
laboratories at some remove from the patient and clinic at which the sample is tube.
Intended use taken by qualified medical personnel. Step 4: Loading
The Infinosis™ Troponin I (cTnI) is an immunofluorescence-based lateral flow • Infinosis™ cTnI assay is single use only. Do not re-use it. Add 80 μL of sample mixture and load it onto the sample well of the test cassette.
assay for the in vitro quantitative determination of cardiac troponin I (cTnI) in • The Test Cassette and Analyzer should be used away from vibration and
Step 5: Testing
human whole blood, serum or plasma. magnetic field. During normal usage, the Test Cassette may introduce minute
Put the Test Cassette on the operation platform. 15 minutes later, insert the Test
vibration, which should be regarded as normal.
Cassette onto the Cassette Holder and click “Test”. The result will show in the
Summary1-4 • Use separate clean pipette tips and detector buffer vials for different specimens.
display and print out when click “Print”.
Cardiac troponin I, often denoted as cTnI, is presented in cardiac muscle tissue by The pipette tips and detector buffer vials should be used for one specimen only.
a single isoform with a molecular weight of 23.9 kDa. It consists of 209 amino acid • Do not smoke, eat, or drink in areas in which specimens or kit reagents are Please refer to the Infinosis™ 2020 FIA analyzer Operation Manual for details.
residues. The theoretical pI of cTnI is 9.05. cTnI differs from other troponins due to handled.
its N-terminal extension of 26 amino acids. This extension contains two serines, • Blood specimens, used test cassettes, pipette tips and detector buffer vials are Limitations - interference
residues 23 and 24, which are phosphorylated by protein kinase A in response to potentially infectious. Proper laboratory safety techniques, handing and
• This test has been developed for testing human whole blood, serum, plasma
beta-adrenergic stimulation and important in increasing the inotropic response. disposal methods should be followed in accordance with standard procedures specimen only.
cTnI has been shown to be phosphorylated by protein kinase A, protein kinase C, and relevant regulations observed by microbiological hazard materials.
• The results of Infinosis™ cTnI should be evaluated with all clinical and
protein kinase G, and p21-activated kinase 3.A significant part of cTnI released • The results should be interpreted by the physician along with clinical findings laboratory data available. If cTnI test results do not agree with the clinical
into the patient's blood stream is phosphorylated. For more than 15 years cTnI and other laboratory test results. evaluation, additional tests should be performed.
has been known as a reliable marker of cardiac muscle tissue injury. It is • The test will be applied on a routine basis and not in emergency situations.
• The false positive results may come from cross-reactions with some similar
considered to be more sensitive and significantly more specific in diagnosis of the antibodies in blood, and similar epitopes from non-specific components in blood
myocardial infarction than the "golden marker" of last decades – CK-MB, as well Storage and stability capturing fluorescent labeled antibodies.
as total creatine kinase, myoglobin and lactate dehydrogenase isoenzymes. • Store the Sample buffer at 4-30°C. The buffer is stable up to 24 months. • The false negative results may from some unknown substance blocking epitope
• Store Infinosis™ test cassette at 4-30°C, shelf life is up to 24 months. adhering antibodies, unstable or degenerated cTnI that cannot be identified due
Test principle • Test cassette should be used within 1 hour after opening the pack. to prolonged time and temperature and storage condition of sample and
Sandwich principle. Total duration of assay: 15 minutes reagent.
Sample is added to the sample well of the test, then the fluorescence-labeled Specimen collection and preparation • Other factors may interfere with Infinosis™ cTnI and may cause erroneous
detector anti-cTnI antibody binds to cTnI antigen in blood specimen. As the The test can be performed with either serum or plasma or whole blood. results. These include technical or procedural errors, as well as additional
sample mixture migrates on the nitrocellulose matrix of test strip by capillary Whole Blood Collected by Venipuncture: substances in blood specimens.
action, the complexes of detector antibody and cTnI are captured to anti-cTnI • Using standard phlebotomy procedure, collect a venipuncture whole blood For diagnostic purposes, the results should always be assessed in conjunction
antibody that has been immobilized on test strip. specimen using a blood collection tube with suitable anticoagulant (EDTA with the patient’s medical history, clinical examination and other findings.
The more cTnI antigen is in blood specimen, the more complexes are recommended)
accumulated on test strip. Signal intensity of fluorescence of detector antibody • It is recommended that specimens should be tested immediately. Do not leave Limits and ranges
reflects amount of cTnI captured and Infinosis™ FIA analyzer shows cTnI the specimens at room temperature for prolonged periods. If the specimens are Measuring range
concentrations in blood specimen. The default results unit of Infinosis™ cTnI test not tested immediately, they may be stored at 2-8°C. 0.1 - 40 ng/mL (defined by the lower detection limit and the maximum of the
is displayed as x ng/mL from Infinosis™ FIA analyzer. • It’s not suitable to test the whole blood samples which have been stored at master curve). Values below the lower detection limit are reported as < 0.1 ng/mL.
2-8°C for more than 2 days. Values above the measuring range are reported as > 40 ng/mL.
Reagents Serum and Plasma:
Materials provided • Separate the serum/plasma from blood as soon as possible to avoid hemolysis. Lower limits of measurement
• Test cassette, 25 pcs, individually packaged • Test should be performed immediately after the specimens have been collected. Lower detection limit
• ID chip, 1 pcs • Do not leave the specimens at room temperature for prolonged periods. Lower detection limit: 0.1 ng/mL

• Sample buffer, 25 vials Specimens may be stored at 2 - 8℃ for up to 3 days. For long-term storage, The detection limit represents the lowest analyte level that can be distinguished
• IFU, 1 copy specimens should be kept below -20℃. from zero. It is calculated as the value lying two standard deviations above that of
the lowest standard (master calibrator, standard 1 + 2 SD, repeatability study, n =
Materials required (but not provided) 21).
Quality control
• Infinosis™ 2020 FIA analyzer Each Infinosis™ cTnI test cassette contains internal control that satisfies routing
• Transfer pipette set (100 μL size) quality control requirements. This internal control is performed each time a patient Expected values
• Alcohol pads sample is tested. This control indicates that the test cartridge was inserted and Reference range
• Centrifuge (for plasma and serum only) read properly by Infinosis™ 2020 FIA analyzer. An invalid result from the internal < 0.3 ng/mL
• Timer control causes an error message on Infinosis™ 2020 FIA analyzer indicating that Expected values may vary with age, sex, diet and geographical location. Each
the test should be repeated. laboratory should determine its own expected values as dictated by good
Precautions and warnings laboratory practice.
• For in vitro diagnostic use only. Test procedure
• Carefully follow the instructions and procedures described in this instructions Specific performance data
before testing. Refer to Infinosis™ 2020 FIA analyzer Operation Manual for the complete
The test cassette should remain in its original sealed pouch until ready to use. instructions on use of the test. The test should be operated in room temperature. Representative performance data are given below. Results obtained in individual

Do not use it if the pouch is damaged or the seal is broken. Step 1: Preparation laboratories may differ.
• Do not use reagents beyond the labeled expiry date. Check/insert ID Chip into the analyzer.
• Do not mix or use components from kits with different Lots. Step 2: Sampling
Precision
• Don’t use Test Cassette if its Lot does not match with ID Chip that is inserted Intra-assay
Add 20μL of whole blood, serum or plasma to the buffer tube.
onto the instrument. Determined by by using 10 replicates from same batch to test with 0.3 ng/mL cTnI
Step 3: Mixing control. CV ≤ 10%.
• The Infinosis™ cTnI assay should be used with Infinosis™ 2020 FIA analyzer.

202105 V1 English 1/2 Infinosis™ Troponin I/IFU


IN0477012104V1

POC Troponin I infinosis™


Cardiac Troponin I
Inter-assay

Determined by using 3 replicates from random 3 continuous batches to test with


0.3 ng/mL cTnI control CV ≤ 15%.
diasino
Linearity DiaSino Laboratories Co., Ltd
A serial concentration of cTnI controls at 0.1 ng/mL, 0.5 ng/mL and 1.5 ng/mL, 5.0 No.68, Jingnansi Road, National Eco & Tech Development Area
ng/mL, 15 ng/mL, 25 ng/mL were tested, the Correlation Coefficient is r ≥ 0.9942. Zhengzhou, China. 450000
Technical Support: ts@diasino.com
Method comparison www.diasino.com
A comparison of the Infinosis™ Troponin I assay (y) with the Roche Troponin I
STAT assay (x) using 119 clinical samples gave the correlation: r=0.9720

References
1. Kozlowski, LP (21 October 2016). "IPC - Isoelectric Point Calculator". Biology
Direct. 11 (1): 55. doi:10.1186/s13062-016-0159-9. PMC 5075173. PMID
27769290.
2. Solaro RJ, Moir AJ, Perry SV (1976). "Phosphorylation of troponin I and the
inotropic effect of adrenaline in the perfused rabbit heart". Nature. 262 (5569):
615–616. doi:10.1038/262615a0. PMID 958429.
3. Layland J, Solaro RJ, Shah AM (2005). "Regulation of cardiac contractile
function by troponin I phosphorylation". Cardiovascular Research. 66 (1): 12–
21. doi:10.1016/j.cardiores.2004.12.022. PMID 15769444.
4. Labugger R, Organ L, Collier C, Atar D, Van Eyk JE (2000). "Extensive troponin
I and T modification detected in serum from patients with acute myocardial
infarction". Circulation. 102 (11): 1221–1226. doi:10.1161/01.cir.102.11.1221.
PMID 10982534.

In vitro diagnostic Refer to instruction


use for use

Expiry date Manufacturing date

Batch number Test per kit

Catalog number Do not re-use

Store between
Manufacturer
4-30℃

202105 V1 English 2/2 Infinosis™ Troponin I/IFU

You might also like