Professional Documents
Culture Documents
Quality Control
b) Inter-assay precision
Good laboratory practice requires that quality control specimens
(controls) be run with each calibration curve to verify assay Expected Values Serum Sample 1 2 3
performance. To ensure proper performance, control material should
be assayed repeatedly to establish mean values and acceptable # Replicates 5 5 5
It is recommended that each laboratory establish its own normal range
ranges. based on the patient population, geography, dietary and environmental Mean cTnI (ng/ml) 0.714 4.86 18.8
factors, and to reflect current practice and criteria for AMI-diagnosis. S.D. 0.074 0.504 2.07
Calculation of Results However, based on published literature, the diagnostic cut-off for AMI
patients is determined to be 1.5ng/ml12. CV (%) 10.4 10.4 11
2. Sensitivity Analyte Concentration 6. Clinical Correlation
The minimal detectable concentration of cTn-I by this assay is Heparin 14 IU/mL A statistical study using 41 clinical patient serum samples, ranging in
estimated to be 0.10 ng/ml. Warfarin 10 ug/mL cTn-I concentration from 0 ng/ml to 87 ng/ml as analyzed using
Chemux Troponin-I ELISA Test, demonstrated equivalent correlation
EDTA 18 mg/mL with a commercially availabel kit as show below:
3. Interference and Cross-Reactivity Red Blood Cells < 100 per mL
The following were tested for cross-reactivity at concentrations up to Hemolysate < 0.05% Correlation Coefficient 0.9296
the levels indicated below. No cross-reactivity was observed for any of Total proteins 30mg/mL Slope 1.1592
the components.
Intercept 2.7361
4. Hook Effect Mean (Chemux) 25.3
ANALYTE TEST CONCENTRATION It has been demonstrated at Troponin I levels up to 20,000 ng/ml, this Mean (Reference) 24.3
Rabbit skeletal muscle troponin C 2,500 ng/ml EIA Kit will produce a concentration measurement above 100 ng/mL,
which is the upper limit of its linear range. However in view of the
Human cardiac troponin T 2,500 ng/ml limitation of optical measurements in our EIA system, absence of the
Human skeletal muscle troponin T 2,500 ng/ml Hook effect cannot be clearly demonstrated beyond the O.D. reading of
Human skeletal muscle troponin I 2,500 ng/ml 3.000. It is recommended that appropriate sample dilutions be made
so that accurate troponin I concentrations can be determined through
Hemoglobin 200mg/ml the precise reading within the linear range of this EIA system. For any
Biotin 200 ng/ml sample that either: produces an O.D. reading above 3.000, has a
measured concentration above 100ng/mL or is clinically suspected to
Bilirubin 1mng/ml contain Troponin I level in excess of 100ng/mL, we recommend diluting
patient samples 1:10 before further analysis.
In vitro testing of the following commonly-used drugs revealed no 5. Linearity and Parallelism study
interference within the normal therapeutic range:
A study was conducted to demonstrate linearity of the assay. Two
positive patient samples were serially diluted. Ng/ml values were
Analyte (10 µg/ml Final Concentration) calculated for individual OD readings of the diluted samples. The
linearity of R² values and slope values are listed in the following table:
Acetaminophen Digitonin
Acetylsalicyclic acid Digoxin The linear regression graph of above two positive samples:
Limitations of the Procedure
Adenine Dopamine
Albumin (bovine) Erythromycin Sample Neat 1:2 1:4 1:8 1:16 1:32 slope R²
Alloprinol Gentistic Acid 1.Reliable and reproducible results will be obtained when the assay
1 17.0 10.5 4.72 2.21 0.99 0.54 1.034 0.99
procedure is carried out with a complete understanding of the package
Ambroxol Isoproterenol
Ampicillin Isosobide dinitrate insert instructions and with adherence to good laboratory practice.
2 10.9 5.16 2.17 1.02 0.65 0.40 1.048 0.99
Ascorbic Acid Nifedipine 2.Troponin-I levels can be increased in any conditions resulting in
Atenolol Nystatin myocardial cell damage. Diagnostic results obtained from this assay
Atropine Oxazepam should be used in conjunction with other diagnostic procedures and
Caffiene Oxytetracycline information available to the physician such as addition clinical testing,
Captopril Propanolol ECG, symptoms, and clinical observations.
Chloramphenicol Theophiline 3.Serum samples demonstration gross hemolysis, lipemia, or turbidity
Cinnarizine L-Thyroxine should not be used with this test.
4.The wash procedure is critical. Insufficient washing will result in poor
Cyclophosphamide Urea
precision and falsely elevated absorbance readings.
Cyclosporine Uric Acid
5.Patient samples may contain human anti-mouse antibodies (HAMA)
Veraparmil
which are capable of giving falsely elevated or depressed results with
assays that utilize mouse monoclonal antibodies. Reliable results in
specimens with HAMA levels above 50μg/mL cannot be guaranteed in
In the absence or presence of 20 ng/ml ternary complex troponin I in these specimens, and all test results should be used in conjunction
normal human serum, the above substances do not show significant with additional clinical observations.
interference to the expected test results. 6.Test results that are inconsistent with the clinical picture and patient
history should be interpreted with caution.
The following materials commonly found in serum specimens exhibited 7.In view of the limitation of optical measurements in our EIA system,
no interference with test results at levels below the specified absence of the Hook effect cannot be clearly demonstrated beyond the
concentrations: O.D. reading of 3.000. It is recommended that appropriate sample
O.D. reading of 3.000. It is recommended that appropriate sample 14.Cardiac Markers Panel, in: Catalog of NyTest, Inc., pg. 4.
dilutions be made so that accurate troponin I concentrations can be
determined through the precise reading within the linear range of this 15.Christensen, R.H. et. al.: Standardization of cardiac troponin I
EIA system. For any sample that either: produces an O.D. reading assays: Round robin of ten candidates reference materials. Clin.
above 3.000, has a measured concentration above 100ng/mL or is Chem., 47:431, 2001.
clinically suspected to contain Troponin I level in excess of 100ng/mL,
we recommend diluting patient samples 1:10 before further analysis. 16.Bodor, G.S.: Cradiac troponin I: a highly biochemical marker for
myocardial infarction. J. Clin. Immunoassay, 17: 40-4, 1994.
References 17.Adams, J., Bodor, G., Davila-Romain, V., et. al.: Cadiac troponin-I:
A marker for cardiac injury. Circulation, 88:101-6, 1993.
1.Perry SV: The regulation of contractile activity in Muscle. Biochem
Soc Trans 1979; 7:593 18.Joint European Society of Cardiology/American Cllege of
Cradiology: J.Am.Coll.Cardio., 36(3), "Myocardial Infarction Redefined",
2.William JM, Grand RJA: Comparison of amino acid of troponin I from 2000.
different striated muscles. Nature 1978; 271:31
19.BLODD TESTS FOR RAPID DETECTION OF HEART ATTACK ©
3.Hartner KT, Petle D:Fast and slow isoforms of troponin I and troponin 2000 American Heart Association, Inc. Guideline, September, 2000.
C. Distribution in normal rabbit muscles and effects of chronic
stimulation. Eur J Biochem 1990; 188:261 20.Ellestad, M,H., The diagnostic power of four chemical markers on
admission to the chest pain center. In Differential Diagnosis and
4.Wu AHB, ed: Cardiac Markers. Totowa, NJ, Human Press, 1998, pp Management of Patients with Chest Pain: A Multiple Biochemical
300 Marker Approach. A Symposium Sponsored by Baylor College of
Medicine, held during the 68th Scientific Session of the American Heart
5.Hanfer S et al: Cardiac troponins in serum in chronic renal failure. Association, November 11, 1995, Anaheim, California.
Clin Chem 1994; 40:1790
11/2014
6.Bhayana V et al: Disordence between results of serum troponin T and
I. Clin Chem 1995; 41:312
8.Wu AHB, Feng YJ, Moore R, Apple FS, McPherson PH, Buechler KF,
Bodor G: Characterization of cardiac troponin subunit release into
serum after acute myocardial infarction and comparisons of assays for
triponin T and I. Clin Chem 1998; 44:1198-2008