You are on page 1of 11

Designation: D 2837 – 98a An American National Standard

Standard Test Method for


Obtaining Hydrostatic Design Basis for Thermoplastic Pipe
Materials1
This standard is issued under the fixed designation D 2837; the number immediately following the designation indicates the year of
original adoption or, in the case of revision, the year of last revision. A number in parentheses indicates the year of last reapproval. A
superscript epsilon (e) indicates an editorial change since the last revision or reapproval.

1. Scope priate safety and health practices and determine the applica-
1.1 This test method describes a procedure for obtaining a bility of regulatory limitations prior to use.
hydrostatic design basis for thermoplastic pipe materials, by 2. Referenced Documents
evaluating stress rupture test data derived from testing pipe
made from the subject material. The method is applicable to all 2.1 ASTM Standards:
known types of thermoplastic pipe and for any practical D 1598 Test Method for Time-to-Failure of Plastic Pipe
temperature and medium. Under Constant Internal Pressure2
1.2 Unless the data approximate a straight line, when E 29 Practice for Using Significant Digits in Test Data to
calculated using log-log coordinates, it is not possible to assign Determine Conformance with Specifications3
a hydrostatic design basis to the material. Data that exhibit high 3. Terminology
scatter or a “knee” give low (that is, conservative) extrapolated
values when evaluated. In addition, the lower confidence level 3.1 Definitions:
limits are not met and the data are classed as unsuitable. 3.1.1 pressure—the force per unit area exerted by the
1.3 The values stated in inch-pound units are to be regarded medium in the pipe.
as the standard. 3.1.2 hoop stress—the tensile stress in the wall of the pipe
in the circumferential orientation due to internal hydrostatic
NOTE 1—Over 1200 sets of data, obtained with thermoplastic pipe and pressure.
piping assemblies tested with water, natural gas, and compressed air, have 3.1.3 The following equations shall be used for the relation
been analyzed. None of the compounds in the lists of Recommended
Hydrostatic Strengths and Design Stresses for Thermoplastic Pipe and
between stress and pressure:
Fittings Compounds in PPI Technical Report TR4, issued at intervals for S 5 P~D — t!/2t for outside diameter controlled pipe (1)
over 12 years by the Plastics Pipe Institute, exhibit knee-type plots, that is,
deviate from a straight line in such a manner that a marked drop occurs in or
stress at some time when plotted on equiscalar log-log coordinates. Data
S 5 P~d 1 t!/2t for inside diameter controlled pipe (2)
have been obtained for test periods over 120 000 h. It might be noted that
some thermoplastic compounds that are not suitable or recommended for where:
piping components do exhibit knee-type plots at 23°C (73°F); in these
S = stress,
cases, very low results are obtained when the data are analyzed by this test
method. Futher information on piping compounds may be found in the
P = pressure,
references at the end of this test method. D = average outside diameter,
d = average inside diameter, and
1.4 The experimental procedure to obtain individual data t = minimum wall thickness.
points shall be as described in Test Method D 1598, which 3.1.4 failure— bursting, cracking, splitting, or weeping
forms a part of this test method. When any part of this test (seepage of liquid) of the pipe during test.
method is not in agreement with Test Method D 1598, the 3.1.5 long-term hydrostatic strength (LTHS)— the esti-
provisions of this test method shall control. mated tensile stress in the wall of the pipe in the circumferen-
1.5 General references are included at the end of this test tial orientation that when applied continuously will cause
method. failure of the pipe at 100 000 h. This is the intercept of the
1.6 This standard does not purport to address all of the stress regression line with the 100 000-h coordinate.
safety concerns, if any, associated with its use. It is the 3.1.6 hydrostatic design basis (HDB)—one of a series of
responsibility of the user of this standard to establish appro- established stress values for a compound. It is obtained by
categorizing the LTHS in accordance with Table 1.
3.1.7 service (design) factor—a number less than 1.00
1
This test method is under the jurisdiction of ASTM Committee F-17 on Plastics (which takes into consideration all the variables and degree of
Piping Systems and is the direct responsibility of Subcommittee F17.40 on Test
Methods.
2
Current edition approved December 10, 1998. Published March 1999. Originally Annual Book of ASTM Standards, Vol 08.04.
3
published as D 2837 – 69. Last previous edition D 2837 – 98. Annual Book of ASTM Standards, Vol 14.02.

Copyright © ASTM, 100 Barr Harbor Drive, West Conshohocken, PA 19428-2959, United States.

1
D 2837
TABLE 1 Hydrostatic Design Basis Categories calculated values, based on experimental data, are available for
NOTE 1—The LTHS is determined to the nearest 10 psi. Rounding temperatures both above and below the temperature of interest.
procedures in Practice E 29 should be followed. 4.4 Hydrostatic design stresses are obtained by multiplying
Range of Calculated LTHS Values Hydrostatic Design Basis the hydrostatic design basis values by a service (design) factor.
psi (MPa) psi (MPa)
4.5 Pressure ratings for pipe may be calculated from the
hydrostatic design stress (HDS) value for the specific material
190 to < 240 ( 1.31 to < 1.65) 200 ( 1.38)
240 to < 300 ( 1.65 to < 2.07) 250 ( 1.72) used to make the pipe, and its dimensions using the equations
300 to < 380 ( 2.07 to < 2.62) 315 ( 2.17) in 3.3.
380 to < 480 ( 2.62 to < 3.31) 400 ( 2.76)
480 to < 600 ( 3.31 to < 4.14) 500 ( 3.45) 5. Procedure
600 to < 760 ( 4.14 to < 5.24) 630 ( 4.34)
760 to < 960 ( 5.24 to < 6.62) 800 ( 5.52) 5.1 General—Generated data in accordance with Test
960 to <1200 ( 6.62 to < 8.27) 1000 ( 6.89) Method D 1598.
1200 to <1530 ( 8.27 to <10.55) 1250 ( 8.62)
1530 to <1920 (10.55 to <13.24) 1600 (11.03)
5.2 Stress Rupture—Obtain the data required for 4.2.1 and
1920 to <2400 (13.24 to <16.55) 2000 (13.79) 4.2.2 as follows:
2400 to <3020 (16.55 to <20.82) 2500 (17.24) 5.2.1 Obtain a minimum of 18 failure stress-time points for
3020 to <3830 (20.82 to <26.41) 3150 (21.72)
3830 to <4800 (26.41 to <33.09) 4000 (27.58)
each environment. Distribute these data points as follows:
4800 to <6040 (33.09 to <41.62) 5000 (34.47) Hours Failure Points
6040 to <6810 (41.62 to <46.92) 6300 (43.41) <1000 At least 6
6810 to <7920 (46.92 to <54.62) 7100 (48.92) 10 to 1000 At least 3
1000 to 6000 At least 3
After 6000 At least 3
After 10 000 At least 1
safety involved in a thermoplastic pressure piping installation)
NOTE 3—When the long-term stress regression line of a compound is
which is multiplied by the HDB to give the HDS.
known, this method may be used, using fewer points and shorter times, to
3.1.8 hydrostatic design stress (HDS)—the estimated maxi- confirm material characteristics, or to evaluate minor process or formu-
mum tensile stress in the wall of the pipe in the circumferential lation changes. See also PPI TR3, Policies and Procedures for Developing
orientation due to internal hydrostatic pressure that can be Recommended Hydrostatic Design Stresses for Thermoplastic Pipe Ma-
applied continuously with a high degree of certainty that failure terials.
of the pipe will not occur. 5.2.2 Analyze the test results by using, for each specimen,
3.1.9 pressure rating (PR)—the estimated maximum pres- the logarithm of the stress in psi and the logarithm of the
sure that the medium in the pipe can exert continuously with a time-to-failure in hours as described in Appendix X1 (Note 4).
high degree of certainty that failure of the pipe will not occur. Calculate the strength at 100 000 h. Include as failures at the
3.1.9.1 The PR and HDS are related by the equations given conclusion of the test those specimens which have not failed
in 3.3. after being under test for more than 10 000 h if they increase
the value of the extrapolated strength. Accomplish this by first
4. Significance and Use
obtaining the linear log-log regression equation for only the
4.1 The procedure for estimating long-term hydrostatic specimens that failed, by the method of least squares as
strength is essentially an extrapolation with respect to time of described in Appendix X1. Then use the stress in psi for each
a stress-time regression line based on data obtained in accor- specimen that has been under test for more than 10 000 h, and
dance with Test Method D 1598. Stress-failure time plots are that has not failed, with this regression equation to calculate the
obtained for the selected temperature and environment: the time in hours. If this time is less than the hours the specimen
extrapolation is made in such a manner that the long-term has been under test, then use the point. Determine the final line
hydrostatic strength is estimated for these conditions. for extrapolation by the method of least squares using the
NOTE 2—Test temperatures should preferably be selected from the failure points along with those non-failure points selected by
following: 40°C; 50°C; 60°C; 80°C; 100°C. It is strongly recommended the method described above. Unless it can be demonstrated that
that data also be generated at 23°C for comparative purposes. they are part of the same regression line, do not use failure
4.2 The hydrostatic design basis is determined by consider- points for stresses that have failure times less than 10 h.
ing the following items and evaluating them in accordance with Include failure points excluded from the calculation by this
5.4. operation in the report, and identify them as being in this
4.2.1 Long-term hydrostatic strength at 100 000 h, category. Refer also to Appendix X8.
4.2.2 Long-term hydrostatic strength at 50 years, and NOTE 4—It should be noted that contrary to the custom in mathematics,
4.2.3 Stress that will give 5 % expansion at 100 000 h. it has been the practice of those testing plastics pipe to plot the
4.2.4 The intent is to make allowance for the basic stress- independent variable (stress) on the vertical (y) axis and the dependent
strain characteristics of the material, as they relate to time. variable (time-to-failure) on the horizontal (x) axis. The procedure in
4.3 Results obtained at one temperature cannot, with any Appendix X1 treats stress as an independent variable.
certainty, be used to estimate values for other temperatures. 5.2.3 Determine the suitability of the data for use in
Therefore, it is essential that hydrostatic design bases be determining the long-term hydrostatic strength and hydrostatic
determined for each specific kind and type of plastic compound design basis of plastic pipe as follows:
and each temperature. Estimates of long-term strengths of 5.2.3.1 Extrapolate the data by the method given in Appen-
materials can be made for a specific temperature provided that dix X1, to 100 000 h and 50 years, and record the extrapolated

2
D 2837
stress values (4.2.1 and 4.2.2), and 5.3.4 Calculate the stress corresponding to a circumferential
5.2.3.2 Calculate, by the method given in Appendix X2, the expansion of 5.00 % in accordance with 5.3.3 and Appendix
lower confidence value of stress at 100 000 h. X4. The stress is the antilog of r in the equation c 5 a9 1
5.2.3.3 If the lower confidence value at 100 000 h differs b9 r in Appendix X4. Use the values for a9 and b9 as calculated
from the extrapolated LTHS value by more than 15 % of the in Appendix X4 and 0.6990 for c. This stress may be obtained
latter, or M in Appendix X2 is zero or negative, or b in the by calculation or read from the circumferential expansion-
equation h = a + bf in Appendix X1 is positive, consider the stress plot obtained in 5.3.3. In cases of disagreement, use the
data unsuitable. calculation procedure.
5.3 Circumferential Expansion—Obtain the data required 5.4 Hydrostatic Design Basis—The procedure for determin-
for 4.2.3 as follows: ing the HDB shall be as follows (see also Appendix X7):
5.3.1 Initially test at least three specimens at a stress of 5.4.1 Calculate the hydrostatic strength at 100 000 h
50 % of the long-term hydrostatic strength determined in (LTHS) in accordance with 5.2.
5.2.3.1 until the circumferential expansion exceeds 5 % or for 5.4.2 Calculate the hydrostatic strength at 50 years in
2000 h, whichever occurs first. Measure the expansion of the accordance with 5.2.3.1.
circumference in the center of that section of the pipe specimen 5.4.3 Estimate the long-term hydrostatic strength using
that is under test to the nearest 0.02 mm (0.001 in.) periodically expansion test data and in accordance with 5.3.
(Note 5) during the test, unless the expansion at some other
NOTE 6—For all the presently used stress rated thermoplastic pipe
point is greater, in which case measure the section with the materials in North America the 5 % expansion strengths are not the
maximum expansion. Calculate the changes in circumference limiting factor. Therefore, this measurement is not required for such
for each specimen as a percentage of the initial outside materials.
circumference. Calculate the expansion at 100 000 h for each
5.4.4 Determine the hydrostatic design basis (HDB) by
specimen by the method given in Appendix X3 or by the
categorizing, in accordance with Table 1, the applicable hydro-
plotting technique described in 5.3.3. If the calculated expan-
static strength value as specified below:
sion for one or more of the specimens tested exceeds 5 %, then
5.4.4.1 Use the LTHS value (5.4.1) if it is less than 125 % of
use the hydrostatic stress as determined from circumferential
the 50-year value (5.4.2), and less than the expansion strength
expansion measurements as the stress value to be categorized
value (5.4.3).
to establish the hydrostatic design basis.
5.4.4.2 Use the 50-year value if it is less than 80 % of the
NOTE 5—It is suggested that these measurements be made once every LTHS value, and less than the expansion strength value.
24 h during the first 5 days, once every 3 days during the next 6 days, and 5.4.4.3 Use the expansion strength value if it is less than the
once a week thereafter. The periods shall be selected on the basis of past LTHS and 50-year values.
experience with the type of pipe so that they will be reasonably distributed
to obtain a good plot.
5.5 Hydrostatic Design Stress—Obtain the hydrostatic de-
sign stress by multiplying the hydrostatic design basis by a
5.3.2 The stresses and distribution of specimens used to service (design) factor selected for the application on the basis
determine hydrostatic stress from circumferential expansion of two general groups of conditions. The first group considers
measurements shall be as follows: the manufacturing and testing variables, specifically normal
Approximate Percent of Long-Term Minimum Number of variations in the material, manufacture, dimensions, good
Hydrostatic Strength (See 5.2) Specimens
20 3
handling techniques, and in the evaluation procedures in this
30 3 test method and in Test Method D 1598 (Note 7). The second
40 3 group considers the application or use, specifically installation,
50 3
60 3
environment, temperature, hazard involved, life expectancy
desired, and the degree of reliability selected (Note 8). Select
Subject the specimens to test until the circumferential the service factor so that the hydrostatic design stress obtained
expansion exceeds 5 % or for 2000 h, whichever occurs first. provides a service life for an indefinite period beyond the
5.3.3 The results may be calculated by the methods given in actual test period.
Appendix X3 and Appendix X4 or plotted by the following
procedures. Plot the percent changes in circumference against NOTE 7—Experience to date, based on data submitted to PPI, indicates
that variation due to this group of conditions are usually within 610 %,
time in hours on log-log graph paper. Draw a straight line by
for any specific compound.
the method of least squares, with time as the independent
NOTE 8—It is not the intent of this standard to give service (design)
variable as described in Appendix X3. Calculate the expansion factors. The service (design) factor should be selected by the design
of the circumference in percent at 100 000 h for each specimen engineer after evaluating fully the service conditions and the engineering
by the equation from Appendix X3: properties of the specific plastics under consideration. Alternatively, it
c 5 a8 1 5.00 b8 (3) may be specified by the authority having jurisdiction.
It is recommended that numbers selected from ANSI Standard Z17.1-
Do not use extrapolations of curves for specimens that 1973 for Preferred Numbers, in the R10 series (25 % increments) be used,
expand more than 5 % in less than 1000 h. Plot the correspond- namely, 0.80, 0.63, 0.50, 0.40, 0.32, 0.25, 0.20, 0.16, 0.12, or 0.10. If
ing expansion-stress points from the 100 000 h intercept on smaller steps seem necessary it is recommended that the R20 series (12 %
log-log graph paper and draw a line representative of these increments) be used, namely, 0.90, 0.80, 0.71, 0.63, 0.56, 0.50, 0.45, 0.40,
0.36, 0.32, 0.28, 0.25, 0.22, 0.20, 0.18, 0.16, 0.14, 0.12, 0.112, or 0.10.
points by the method of least squares with stress as the
independent variable as described in Appendix X4. 5.6 Supplemental Validation of Long-Term Hydrostatic

3
D 2837
Strength for Polyethylene Materials—Apply one of the two above. The reference stress shall be established by three
following procedures to PE material to validate the 23°C specimens all failing in the ductile mode at the same tempera-
(73°F) LTHS calculated by step 5.2. Use Procedure I when it is ture.
practical to develop sufficient slit failure mode elevated tem- 5.7 Determination and Validation of the Hydrostatic Design
perature data for analysis by rate process equations. Procedure Basis (HDB) for Elevated Temperatures for Polyethylene
II may be elected if the composition will not fail in the slit Piping Materials
mode within 6000 h at temperatures of 80°C (176°F) or higher. 5.7.1 Standard Method— Determination and Validation of
Conduct all validation tests with water inside the pipe speci- Elevated Temperature HDB—Develop data in accordance with
mens. 5.2 for the temperature at which an HDB is desired.
5.6.1 Procedure I: 5.7.1.1 If a brittle/slit failure occurs before 10 000 h, this
5.6.1.1 Select an elevated temperature appropriate for the Standard Method is not applicable and the Alternate Method in
polyethylene material. The maximum temperature chosen 5.7.2 shall be used. Analyze the data to determine the linear
should not be greater than 95°C (203°F). regression equation. Extrapolate this equation to 100 000 h to
5.6.1.2 Select a stress at this temperature at which all determine the LTHS. If the 97.5 % LCL at 100 000 h is less
failures occur in the slit mode (a crack through the pipe wall than 90 % of this LTHS, consider the data unsuitable for use by
with no visible evidence of material deformation). This set of this method. If all conditions are satisified, use Table 1 to
temperature and stress is called Condition I. Test at least six determine the HDB category at this temperature.
pipe specimens at this Condition I until failure. 5.7.1.2 When the HDB category has been determined, use
5.6.1.3 At the same temperature, select another stress about Tables 2-5 to define the time and stress requirements needed to
75 to 150 psi lower than for Condition I. Test at least six validate this HDB. Test ate least six specimens at the stress
specimens at this Condition II until failure. level determined by these tables. These specimens must have a
5.6.1.4 Select a temperature 10°C (18°F) to 20°C (36°F) minimum log average time exceeding the value shown in the
lower than the one in Condition I and use the same stress as table to validate the elevated temperature HDB. For example,
Condition I. This is Condition III. Initiate testing for six to validate an HDB of 1000 psi at 140°F, this required time is
specimens at this Condition III. Ideally, the selected tempera- 3800 h at 193°F (90°C)/600 psi or 11 300 h at 176°F
ture for Condition III should result in specimens that are on test (80°C)/675 psi.
for at least 1000 to 5000 h. NOTE 9—When an elevated temperature HDB is validated by this
5.6.1.5 To validate the long-term hydrostatic strength standard method, all lower temperature HDBs are considered validated for
(LTHS) on a given pipe lot, use the twelve data points from that material.
Conditions I and II and the value of the LTHS at 100 000 h for 5.7.2 Alternative Method—Determination of Elevated Tem-
23°C (73°F), as determined in 5.2. Using all these points, perature HDB When Brittle/slit failures Occur Before 10,000
calculate the A, B, and C coefficients for the following hours—If the Standard Method outlined in 5.7.1 is not appro-
three-coefficient rate process extrapolation equation: priate for the material and test data, then use this Alternate
B C log S Method to determine the HDB.
log t 5 A 1 T 1 T (4)
5.7.2.1 Develop data in accordance with 5.2 for the tem-
perature at which an HDB is desired. Using only the ductile
where:
failures, determine the linear regression equation. The failure
t = time, h,
T = absolute temperature, °K (K = C + 273), point data must be spread over at least two log decades and
S = hoop stress, psi, and meet the LCL requirements of 5.7.1.1. The stress intercept at
A, B, C = constants. 100 00 h using this equation is the “ductile” LTHS.
5.6.1.6 Using this model, calculate the mean estimated 5.7.2.2 To determine the brittle/slit failure performance,
failure time for Condition III. When the average time (log solve for the three coefficients of the rate process equation
basis) for the six specimens tested at Condition III has reached using Steps 1 to 4 of Procedure I in 5.6.1, or another
this time, the extrapolation to 100 000 h to obtain the LTHS at recognized rate process method protocol. All failures must be
23°C 73°F) has been validated. (Examples are shown in in the brittle/slit mode. Data developed under 5.6.1 to validate
Appendix X9.) a 73°F HDB can be used to solve for the three-coefficient
5.6.2 Procedure II— The LTHS is validated when either of equation as long as all specimens at the three conditions were
the following is met: tested to failure and resulted in brittle/slit type failures. Use the
5.6.2.1 Stress-rupture tests run in accordance with the pro-
cedures of this test method at 80°C or higher temperature yield TABLE 2 Validation of 100°F (38°C) HDB
all failures in the ductile mode, when run in accordance with HDB to be 193°F (90°C) Test Temperature / 176°F (80°C) Test Temperature
the following program: 12 points total; 4 points in the range of Validated (psi)
Stress (psi) Time (h) Stress (psi) Time (h)
10 to 1000 h; 2 points in the range of 1000 to 4000 h; and one
1600 745 300 835 1000
point over 6000 h tests at a stress at least 85 % of the long-term 1250 580 300 655 1000
hydrostatic strength of the polyethylene material, or, 1000 465 300 520 1000
5.6.2.2 Six specimens which are tested at a stress of not 800 370 300 420 1000
630 295 300 330 1000
more than 100 psi below a reference stress where all failures 500 230 300 260 1000
are ductile, have gone at least 6000 h without failure at 80°C or

4
D 2837
TABLE 3 Validation of 120°F (49°C) HDB 5.8 Pressure Rating—Calculate the pressure rating for each
HDB to be 193°F(90°C) Test Temperature / 176°F(80°C) Test Temperature diameter and wall thickness of pipe from the hydrostatic design
Validated (psi)
Stress (psi) Time (h) Stress (psi) Time (h) stress (hydrostatic design basis 3 service factor) for the
1600 845 1100 950 3400 specific material in the pipe by means of the equations in 3.1.3.
1250 660 1100 740 3400
1000 530 1100 595 3400 6. Report
800 425 1100 475 3400
630 335 1100 375 3400 6.1 The report shall include the following:
500 264 1100 300 3400 6.1.1 Complete identification of the sample, including ma-
terial type, source, manufacturer’s name and code number, and
previous significant history, if any,
TABLE 4 Validation of 140°F (60°C) HDB
6.1.2 Pipe dimensions including nominal size, average and
HDB to be 193°F(90°C) Test Temperature / 176°F(80°C) Test Temperature
Validated (psi)
minimum wall thickness, and average outside diameter,
Stress (psi) Time (h) Stress (psi) Time (h) 6.1.3 Test temperature,
1250 750 3800 845 11300 6.1.4 Test environment inside and outside of the pipe,
1000 600 3800 675 11300
800 480 3800 540 11300
6.1.5 A table of the stresses in pounds-force per square inch
630 380 3800 425 11300 and the time-to-failure in hours for all the specimens tested
500 300 3800 340 11300 (specimens that are designated as failures after they have been
400 240 3800 270 11300
under stress for more than 10 000 h shall be indicated),
6.1.6 The estimated long-term hydrostatic strength (Note
TABLE 5 Validation of 160°F (71°C) HDB
11),
6.1.7 The estimated stress at 50 years,
HDB to be 193°F(90°C) Test Temperature / 176°F(80°C) Test Temperature
Validated (psi) 6.1.8 A table of the percent circumferential expansion
Stress (psi) Time (h) Stress (psi) Time (h)
versus time data and the estimated stress at 5.00 % expansion.
1250 850 12600 960 37500 This item need not be reported if previous test results show that
1000 680 12600 770 37500
800 545 12600 615 37500 the stress calculated for 5 % expansion is significantly greater
630 430 12600 480 37500 than that reported in 6.1.6 or 6.1.7.
500 340 12600 385 37500 6.1.9 The hydrostatic design basis
400 275 12600 305 37500
6.1.10 The nature of the failures in accordance with 3.4,
6.1.11 Any unusual behavior observed in the tests,
failure points at the three conditions to solve for the three 6.1.12 If the material is polyethylene, the results of the
unknown coefficients. Using this brittle/slit failure model, validation in accordance with 5.6,
calculate the stress intercept value at 100 000 h for the 6.1.13 Dates of test, and
temperature at which an HDB is desired. This resulting stress 6.1.14 Name of laboratory and supervisor of the tests.
intercept is the “brittle/slit” LTHS.
NOTE 11—The outside environment of the pipe test specimen shall be
5.7.2.3 The LTHS used to determine the HDB category as placed after the values reported.
per Table 1 shall be the lower value of the ductile failure LTHS
from 5.7.2.1 or this brittle/slit failure LTHS. 7. Precision and Bias
4
NOTE 10—The ISO TR/9080 four coefficient model may be used if it 7.1 No statement is made about either the precision or the
has a better statistical fit to the data. bias of Test Method D 2837 for measuring the hydrostatic
design basis since the result merely states whether there is
4
For additional information contact the Plastic Pipe Institute Hydrostatic Stress conformance to the criteria for success specified in the proce-
Board Chairman, 1801 K St., NW, Suite 600 K, Washington, DC 20006. dure.

5
D 2837
APPENDIXES

(Nonmandatory Information)

X1. LEAST SQUARES CALCULATIONS FOR LONG-TERM HYDROSTATIC STRENGTH

X1.1 The following symbols are used (Note X1.1): b 5 W/U (X1.5)
and
N = number of points on the cycles to failure versus stress a 5 H – bF (X1.6)
plot If b is positive, the data are unsuitable for evaluating the
f = logarithm of failure stress, psi, material.
F = arithmetic average of all f values, X1.1.3 Substitute these values of a and b into the equation:
h = logarithm of failure time, h
H = arithmetic average of all h values. h 5 a 1 bf (X1.7)
The equation of the straight line is: X1.1.4 Arbitrarily select three convenient values for f and
h 5 a 1 bf (X1.1) calculate h for each. The values of f should not be chosen too
close to one another. Plot these three pairs of values for f and
X1.1.1 Compute the three quantities: h. If these three points do not lie on a straight line, there is a
U 5 (f2 2 @~ (f!2/N# ~ or (f 2 2 NF2! (X1.2) mistake in the calculations.
2 2 2 2 X1.1.5 A sample calculation made in accordance with
V 5 (h 2 @~ (h ! /N# ~or (h 2 NH ! (X1.3)
Appendix X1 is given in Appendix X6.
W 5 ( f h – @~ (f! ~ (h!/N# ~or (f h – NFH! (X1.4)
NOTE X1.1—All logarithms are to the base 10. Use 5-place tables for
X1.1.2 Calculate a and b as follows: calculations. A sample calculation is given in Appendix X6.

X2. CALCULATIONS OF LOWER CONFIDENCE LIMIT

X2.1 Let f100 000 represent the value of stress corresponding NOTE X2.2—97.5 % of the expected failures at 100 000 h will be above
to 100 000 h failure-time. Then: this stress.

f100 000 5 ~5 2 a!/b (X2.1)


TABLE X2.1 Calculations
X2.2 The lower confidence value of stress at 100 000 h is
Degrees of Students Degrees of Students Degrees of Students
given by the following calculations: Freedom, “t”A Freedom, “t”A Freedom, “t”A
X2.2.1 Calculate D 5 5 – H. N−2 N−2 N−2
X2.2.2 Calculate the variance, 1 12.7062 46 2.0129 91 1.9864
2 4.3027 47 2.0117 92 1.9861
s2 5 @1/~N 2 2!#@V – ~W2/U!# (X2.2) 3 3.1824 48 2.0106 93 1.9858
4 2.7764 49 2.0096 94 1.9855
and its square root, s, the standard deviation. 5 2.5706 50 2.0086 95 1.9853
X2.2.3 Substitute the value, t, of Student’s t distribution,
6 2.4469 51 2.0076 96 1.9850
from Appendix X4 corresponding to N − 2 degrees of freedom 7 2.3646 52 2.0066 97 1.9847
at the two-sided 5 % level of significance (Note X2.1). See also 8 2.3060 53 2.0057 98 1.9845
Table X2.1. 9 2.2622 54 2.0049 99 1.9842
10 2.2281 55 2.0040 100 1.9840
X2.2.4 Calculate the quantity:
11 2.2010 56 2.0032 102 1.9835
M 5 b2 – ~t2s2/U! (X2.3)
12 2.1788 57 2.0025 104 1.9830
If M is negative or zero, the slope of log cycles versus stress 13 2.1604 58 2.0017 106 1.9826
14 2.1448 59 2.0010 108 1.9822
is not significantly different from zero. In this case, the lower 15 2.1315 60 2.0003 110 1.9818
confidence limit cannot be calculated, and the data are unreli-
able for the evaluation of the material. The calculations below 16 2.1199 61 1.9996 112 1.9814
17 2.1098 62 1.9990 114 1.9810
should be carried out only when the value of M is positive. 18 2.1009 63 1.9983 116 1.9806
X2.2.5 Calculate the quantity: 19 2.0930 64 1.9977 118 1.9803
20 2.0860 65 1.9971 120 1.9799
L 5 @bD – ts=~D 2 /U! 1 ~M/N!# /M (X2.4)
21 2.0796 66 1.9966 122 1.9796
(See Appendix X5.) 22 2.0739 67 1.9960 124 1.9703
X2.2.6 The lower confidence limit of f100 000 is equal to L + 23 2.0687 68 1.9955 126 1.9790
F (Note X2.2). 24 2.0639 69 1.9949 128 1.9787
25 2.0595 70 1.9944 130 1.9784
NOTE X2.1—For instance, Statistical Methods for Chemists by W. J.
26 2.0555 71 1.9939 132 1.9781
Youden, Page 119, Wiley, (1951) New York.

6
D 2837

TABLE X2.1 Continued TABLE X2.1 Continued


Degrees of Students Degrees of Students Degrees of Students Degrees of Students Degrees of Students Degrees of Students
Freedom, “t”A Freedom, “t”A Freedom, “t”A Freedom, “t”A Freedom, “t”A Freedom, “t”A
N−2 N−2 N−2 N−2 N−2 N−2
27 2.0518 72 1.9935 134 1.9778 38 2.0244 83 1.9890 400 1.9659
28 2.0484 73 1.9930 136 1.9776 39 2.0227 84 1.9886 500 1.9647
29 2.0452 74 1.9925 138 1.9773 40 2.0211 85 1.9883 600 1.9639
30 2.0423 75 1.9921 140 1.9771
41 2.0195 86 1.9879 700 1.9634
31 2.0395 76 1.9917 142 1.9768 42 2.0181 87 1.9876 800 1.9629
32 2.0369 77 1.9913 144 1.9766 43 2.0167 88 1.9873 900 1.9626
33 2.0345 78 1.9908 146 1.9763 44 2.0154 89 1.9870 1000 1.9623
34 2.0322 79 1.9905 148 1.9761 45 2.0141 90 1.9867 ` 1.9600
35 2.0301 80 1.9901 150 1.9759 A
Two-sided 0.05 level of significance. The values in this table are taken from the
tables on pages 28–30 of “The Handbook of Statistical Tables,’’ by D. B. Owen,
36 2.0281 81 1.9897 200 1.9719
Addison-Wesley Publishing Co., Reading, MA, 1962, by permission of the author,
37 2.0262 82 1.9893 300 1.9679
publishers and the United States Atomic Energy Commission.

X3. LEAST SQUARES CALCULATIONS FOR CIRCUMFERENTIAL EXPANSION-TIME PLOT

X3.1 The following symbols are used (Note X3.1): b8 5 W8/U8 (X3.4)

and
N8 = number of points on the time versus circumferential a8 5 C b8G (X3.5)
expansion plot,
c = logarithm of circumferential expansion in percent, If b8 is negative, the data are unsuitable for evaluating the
C = arithmetic average of all c values, material.
g = logarithm of time, h, and X3.2.3 Substitute the value, a’ and b’ into the equation:
G = arithmetic average of all g values.
c 5 a’1 b’g (X3.6)

X3.2 The equation of the straight line is: X3.2.4 Arbitrarily select three convenient values of g and
c 5 a8 1 b8g (X3.1) calculate c for each. The values of g should not be chosen too
close to one another. Plot these three pairs of values for g and
X3.2.1 Compute the two quantities:
c. If these three points do not lie on a straight line there is a
U8 5 (g2 2 @~ (g!2/N8# ~or (g2 2 N8G2! (X3.2) mistake in the calculations.
W8 5 (cg – @~ (c!~ (g!/N8# ~or (cg – N8CG! (X3.3)
NOTE X3.1—All logarithms are to the base 10. Use 4-place tables for
X3.2.2 Calculate a’ and b’ as follows: calculations.

X4. LEAST SQUARES CALCULATIONS FOR CIRCUMFERENTIAL EXPANSION-STRESS PLOT

X4.1 The following symbols are used (Note X1.1): X4.2.2 Calculate a9 and b9 as follows:
b9 5 W9/U9 (X4.4)
N9 = number of points on the circumferential expansion- and
stress plot,
a9 5 C – b9R (X4.5)
c = logarithm of circumferential expansion in percent,
C = arithmetic average of all c values, If b9 is negative, the data are unsuitable for evaluating the
r = logarithm of stress, psi, and material.
R = arithmetic average of all r values.
X4.2.3 Substitute the value, a9 and b9 into the equation:
c 5 a9 1 b9r. (X4.6)
X4.2 The equation of the straight line is:
c 5 a9 1 b9r (X4.1) X4.2.4 Arbitrarily select three convenient values for r and
X4.2.1 Compute the two quantities: calculate c for each. The values of r should not be chosen too
close to one another. Plot these three pairs of values for c and
U9 5 (r2 2 @~ (r!2/N 9# ~or (r2 2 N 9R 2! (X4.2) r. If these points do not lie on a straight line there is a mistake
W9 5 (cr 2 @~ (c!~ (r!/N 9# ~or (cr – N9CR ! (X4.3) in the calculations.

7
D 2837

X5. DERIVATION OF FORMULAS

X5.1 The basic equation is: The limits of this interval are given by:
h 5 a 1 bf 1 error (X5.1) @z – E~z!# 2 5 t2 V~z! (X5.14)
which can also be written:
h – H 5 b~f – F! 1 error (X5.2) which, in view of Eq X5.3 and Eq X5.9, becomes:
Consider an assigned value for h (for example h = 5,
corresponding to a failure time of 100 000 h). Denote it by h0. z2 5 t2 s2 F ~ f0 – F ! 2
U
1
1 N G (X5.15)
The problem is to evaluate the uncertainty of the corresponding
value of f 0. The value f0 is evaluated by the equation:
Introducing Eq X5.2, Eq X5.13 can be written:

F G
b~f0 – F! 5 h0 – H (X5.3)
~f0 2 F!2 1
Let @b~f0 – F! – ~h0 2 H!# 2 5 t2s2 U 1N (X5.16)
z 5 b~f0 – F! – ~h0 – H! (X5.4)
Then the expected value of z is zero (because of Eq X5.3): Writing
E~z! 5 0 (X5.5) L 5 f0 – F (X5.17)
and the variance z, V (z), is given by: D 5 h0 – H (X5.18)
2
V~z! 5 ~f0 2 F! V~b! 1 V~H! (X5.6)
By least squares theory we know that: and solving Eq X5.15 for L , we obtain:
V~H! 5 s2/N (X5.7) bD 1 ts=@b22 ~t2 s2/U!/N# 1 ~D2/U!
L5 (X5.19)
and t2 s2
b2 – U
V~b! 5 s2/~f – F!2 2 s 2/U (X5.8)
where s is the variance of the error in the determination of
2
Let
any single h value.
Introducing Eq X5.5 and Eq X5.6 into Eq X5.4 gives: M 5 b2 2 ~t2 s2/U! (X5.20)

V~ z! 5 s 2 F ~ f0 2 F!2 1
U 1N G (X5.9) Then, the lower limit for L is given by:
The estimate for s is:2
bD – ts=~M/N! 1 ~D2/ U!
Llower limit 5 M (X5.21)
s 2 5 @1/~N 2 2!#@V – ~W 2 /U!# (X5.10)

and is evaluated with (N − 2) degrees of freedom. Conse- Consequently, in view of Eq X5.15, the lower limit for f0 is
quently, an estimate for V(z) is given by: given by:
1
F W2
V~z! 5 N – 2 V – U · GF
~f0 2 F!2 1
U 1 N 5 s2 G F
~f 0 2 F!2
U 1N
1
G f 0, lower limit 5 L lower limit 1 F. (X5.22)

(X5.11)
(h = 93.77592 (f = 117.91991
and the estimated standard deviation of z is: ((h)2 = 8793.9231718 ((f )2 = 13905.1051744

Œ
H = 2.930498 F = 3.684997
~f0 2 F!2 1 ( h2 = 300.349955 (f 2 = 434.562639
sz 5 s U 1 N N = 32 (fh = 344.769246

Step 1:
The quantity (z – E (z)/sz) has Student’s t − distribution with U = 434.562638 − 13905.1051744/32
U = 0.028102
(N − 2) degrees of freedom. Let t denote the critical value of V = 300.349955 − 8793.9231718/32
Student’s t, for (N − 2) degrees of freedom and for the chosen V = 25.539856
level of significance. Then the following inequity holds with W = 344.7693246 − (117.91991 3 93.77592)/32
W = 0.794755
probability equal to the applicable confidence coefficient.
Step 2:
~1 2 level of significance!: 2t # @z – E ~z!#/sz # 1 t (X5.12)
b = 0.794755/0.028102 = −28.28108
a = 2.930498 − (−28.28108 3 3.684997)
which is equivalent to: a = 107.14619

@~z – E~z!# 2 /V~z! # t 2 (X5.13) Step 3:


h = 107.14619 − 28.28108 f

8
D 2837
Step 4: Period Stress
Selected Calculated 100 000 h 4091 psi
Stress, Time, 50 years 3883 psi
psi f h h A computer program calculation to eight decimals gives:
h = 107.13634
4000 3.60206 5.27604 188.820 f = 28.27840
4000 3.64345 4.10549 12749 Stress at 100 000 h = 4091 psi
4800 3.68124 3.03675 1088 Stress at 50 years = 3883 psi
A plot of stress versus time on log-log graph paper or of f versus h on regular
graph paper shows that the three points lie on a straight line. Thus, the calcula-
tions are correct.

Step 5:
Calculate stress at 100 000 h and 50 years from the equation in Step 3.

X6. SAMPLE CALCULATION ACCORDING TO APPENDIX X1

X6.1 Table X6.1 shows a sample calculation according to


Appendix X1.

TABLE X6.1 Sample Calculation

Data Time, Stress, Log Log Stress, h2 f2 fh


Point h psi Time, f
h
1 9 5500 0.95424 3.74036 0.910574 13.990293 3.569201
2 13 5500 1.11394 3.74036 1.240862 13.990293 4.166537
3 17 5500 1.23045 3.74036 1.514007 13.990293 4.602326
4 17 5500 1.23045 3.74036 1.514007 13.990293 4.602326

5 104 5200 2.01703 3.71600 4.068410 13.808656 7.495283


6 142 5200 2.15229 3.71600 4.632352 13.808656 7.997910
7 204 5200 2.30963 3.71600 5.334391 13.808656 8.582585
8 209 5200 2.32015 3.71600 5.383096 13.808656 8.621677

9 272 5000 2.43457 3.69897 5.927131 13.682379 9.005401


10 446 5000 2.64933 3.69897 7.018949 13.682379 9.799792
11 466 5000 2.66839 3.69897 7.120305 13.682379 9.870295
12 589 4800 2.77012 3.68124 7.673565 13.551528 10.197477

13 669 4700 2.82543 3.67210 7.983055 13.484318 10.375262


14 684 5000 2.83506 3.69897 8.037565 13.582379 10.486802
15 878 4600 2.94349 3.66276 8.664133 13.415811 10.781297
16 1299 4800 3.11361 3.68124 9.694567 13.551528 11.461946

17 1301 4700 3.11428 3.67210 9.698740 13.484318 11.435948


18 1430 4800 3.15534 3.68124 9.956171 13.551528 11.615564
19 1710 4800 3.23300 3.68124 10.452289 13.551528 11.901449
20 2103 4800 3.32284 3.68124 11.041266 13.551528 12.22172

21 2220 4500 3.34635 3.65321 11.198058 13.345943 12.224919


22 2230 4400 3.34830 3.64345 11.211113 13.274728†A 12.199364
23 3816 4700 3.58161 3.67210 12.827930 13.484318 13.152030
24 4110 4700 3.61384 3.67210 13.059840 13.484318 13.270382

25 4173 4600 3.62045 3.66276 13.107658 13.415811 13.260839


26 5184 4400 3.71466 3.64345 13.798699 13.274728†A 13.534178
27 8900 4600 3.94939 3.66276 15.597681 13.415811 14.465668
28 8900 4600 3.94939 3.66276 15.597681 13.415811 14.465668

29 10900 4500 4.03743 3.65321 16.300841 13.345943 14.749580


30 10920 4500 4.03822 3.65321 16.307221 13.345943 14.752466
31 12340 4500 4.09132 3.65321 16.738899 13.345943 14.946451
32 12340 4500 4.09132 3.65321 16.738899 13.345943 14.946451
A
† Editorially corrected.

X7. EXAMPLES FOR 5.3.2

X7.1 A PVC compound has a long-term hydrostatic The expansion for 4110 psi at 100 000 h is 2.1 %. Because
strength of 4110 psi, a 50-year strength of 3950 psi, and a stress 3950/4110 = 0.961, the selection is between 4110 and 6060 psi.
of 6060 psi is required to give 5 % expansion at 100 000 h. Therefore, the hydrostatic design basis is 4000 psi.

9
D 2837
X7.2 A PVC compound has a long-term hydrostatic is required to give 5 % expansion at 100 000 h. Because
strength of 4320 psi, a 50-year strength of 3310 psi, and a stress 600/810 = 0.74, the selection is among 810, 600, and 560 psi.
of 4400 psi is required to give 5 % expansion at 100 000 h. Therefore, the hydrostatic design basis is 500 psi.
Because 3310/4320 = 0.77, the selection is among 4320, 3310,
and 4400 psi. Therefore, the hydrostatic design basis is 3150 X7.4 An ABS compound has a long-term hydrostatic
psi. strength of 3320 psi, a 50-year strength of 3020 psi, and a stress
of 4870 psi is required to give a 5 % expansion at 100 000 h.
X7.3 A PE compound has a long-term hydrostatic strength Because 3020/3320 = 0.91, the selection is between 3320 and
of 810 psi, a 50-year strength of 600 psi, and a stress of 560 psi 4870 psi. Therefore, the hydrostatic design basis is 3150 psi.

X8. USE OF DATA HAVING FAILURE TIMES LESS THAN 10 h

X8.1 To determine if data having failure times less than 10 Example 1, PVC 1120:
h should be used or excluded when calculating the stress Data Point, h S at 105 LCL, LCL,
h, psi psi Ratio
characteristics of the material, calculate the stress value at Data over 10 only 4073 3807 0.935
100 000 h, the LCL, and the ratio of these two stresses, using
only data having failure times greater than 10 h. By adding 1.9 4050 3887 0.960
0.60 4120 3980 0.966
points, one at a time, having progressively shorter times to 0.40 4105 3989 0.972
failure, and calculating the ratio each time, the acceptability of 0.16 4081 3978 0.975
each point can be determined by inspection (see examples). 0.02 4065 3978 0.979
0.006 4046 3968 0.981
X8.2 The LCL ratio for the PVC data improves as more Example 2, PE 2305:
points are added: this means that all the data shown is part of
the same regression line; all data may be used in the calculation Data over 10 only 1179 1081 0.917
of the stress characteristics for the material. However, for the 7.4 1172 1073 0.916
PE, the LCL ratio remains more or less constant until the 4.9 1153 1062 0.921
0.17-h data point is added, at which time the LCL ratio drops 2.6 1118 1011 0.904
0.98 1104 1014 0.918
markedly. Therefore this latter point should be excluded from 0.17 1003 861 0.858
the stress calculations.

X9. EXAMPLES FOR 5.6.1 (PROCEDURE I)

X9.1 Example Calculation No. 1—For a PE Material, the S = 500 psi hoop stress (approximately 100 psig for SDR
Test Method D 2837 calculated LTHS at 23°C (73°F) (296°K) 11 pipe)
is 1605 psi. To validate this LTHS, pipe specimens must be At this condition, the following slit failure mode data were
tested at three conditions. obtained:
X9.1.1 Tests at Condition I: Failure Time (h) Log Failure Time

815 2.9111
1250 3.0969
T = 90°C (194°F) (363°K) 1930 3.2855
S = 600 psi hoop stress (approximately 120 psig for SDR 2250 3.3521
11 pipe) 2651 3.4234
At this condition, the following slit failure mode data were 3785 3.5780
Average of log failure times = 3.2745
obtained: Average failure time (log basis) = 1882 h
Failure Time (h) Log Failure Time
X9.1.3 Calculate A, B, C—To calculate the three constants
97 1.9867 A, B, and C, we must solve the following three simultaneous
148 2.1553
218 2.3384 equations:
256 2.4082
B C log 600
357 2.5526 log 220 5 A 1 363 1 363 (X9.1)
408 2.6106
Average of log failure times = 2.3420
B C log 500
Average failure time (log basis) = 220 h log 1882 5 A 1 363 1 363 (X9.2)
(220 is the anti-log of 2.3420)
B C log 1605
X9.1.2 Tests at Condition II: log 100 000 5 A 1 296 1 (X9.3)
296
A 5 236.6843
T = 90°C (194°F) (363°K)
B 5 26054.1

10
D 2837
C 5 24276.85 T = 80°C (353°K) (176°F)
X9.1.4 Tests at Condition III—Minimum Time Requirement S = 450 psi hoop stress (approximately 90 psig for SDF 11
for Validation: pipe)
T = 80°C (353°K) (176°F) t = 779, 821, 864, 956, 1201, 1560 h
S = 600 psi X9.2.3 Calculate A, B, C—Determine the average failure
26054.1 4276.85 log 600 time (log basis) for conditions I and II. Remember that the
log t 5 236.6843 1 353 2 353 (X9.4) average must be determined on a log basis. Solve three
t 5 2800 h
simultaneous equations.
B C log 660
When the average time (log basis) for six specimens tested at log 191 5 A 1 353 1 353 (X9.5)
Condition III exceeds 2800 h, the Test Method D 2837 extrapo-
lation to 1605 psi has been validated for this PE pipe lot. B C log 450
log 998 5 A 1 353 1 353 (X9.6)

X9.2 Example Calculation No. 2—For a PE material, the B C log 1365


log 100 000 5 A 1 296 1 296 (X9.7)
Test Method D 2837-calculated LTHS at 23°C is 1365 psi. To
validate this LTHS, pipe specimens must be on test at three A 5 220.2812
conditions. B 5 12265.1
X9.2.1 Tests at Condition I: C 5 21524.04

X9.2.4 Tests at Condition III—Minimum Time Requirement


T = 80°C (353°K) (176°F) for Validation:
S = 660 psi hoop stress (approximately 132 psig for SDR
T = 60°C (333°K) (140°F)
11 pipe)
t = 136, 148, 182, 215, 216, 287 h S = 660 psi hoop stress
X9.2.2 Tests at Condition II: 12265.1 1524.04 log 660
log t 5 220.2812 1 333 2 333 (X9.8)

t 5 4325 h

The American Society for Testing and Materials takes no position respecting the validity of any patent rights asserted in connection
with any item mentioned in this standard. Users of this standard are expressly advised that determination of the validity of any such
patent rights, and the risk of infringement of such rights, are entirely their own responsibility.

This standard is subject to revision at any time by the responsible technical committee and must be reviewed every five years and
if not revised, either reapproved or withdrawn. Your comments are invited either for revision of this standard or for additional standards
and should be addressed to ASTM Headquarters. Your comments will receive careful consideration at a meeting of the responsible
technical committee, which you may attend. If you feel that your comments have not received a fair hearing you should make your
views known to the ASTM Committee on Standards, at the address shown below.

This standard is copyrighted by ASTM, 100 Barr Harbor Drive, PO Box C700, West Conshohocken, PA 19428-2959, United States.
Individual reprints (single or multiple copies) of this standard may be obtained by contacting ASTM at the above address or at
610-832-9585 (phone), 610-832-9555 (fax), or service@astm.org (e-mail); or through the ASTM website (www.astm.org).

11

You might also like