SSK DÜZCE HOSPITAL MEDICAL GAS SUPPLY SYSTEM DESIGN

SSK DÜZCE HOSPITAL

CENTRAL MEDICAL GAS


SUPPLY SYSTEM

DESIGN CALCULATIONS

DESIGN BY MEDIKAR of TURKEY PAGE NO:1


22.01.2022
SSK DÜZCE HOSPITAL MEDICAL GAS SUPPLY SYSTEM DESIGN
DESIGN BY MEDIKAR INC. OF TURKEY
COMPLYING WITH HEALTH TECHNICAL MEMORANDUM (HTM) 2022

A- DEMAND GAS FLOW RATE CALCULATIONS:


2ND FLOOR:

C1-B1 BLOCK-PATIENT ROOMS

2 Beds Patient Room 12


1 Bed Patient Room 1
No of O2 Outlet n= 25
No of Vacuum Outlet n= 25

Oxygen Demand Flow [QO2]:


HTM 2022 Table 6.0 Formula A.1
Q=10+(n-1)x6/3 n= 25
[QO2] = 10+(25-1)x6/3= 58 lpm
[QO2] = 58 lpm

Vacuum Demand Flow [QVAC]:


HTM 2022 Table 13.0 Formula A.1 and paragraf 4.71
Total No of Vacuum Outlet in the ward ,If less than 40
Vacuum Flowrate will be calculated as 40 lpm.
[Qv] = 40 lpm n= 25

2ND FLOOR PATIENT ROOMS C1-B1- [QO2] = 58 lpm [Qvac ] = 40 lpm.

C2-B2 BLOCK-PATIENT ROOMS

2 Beds Patient Room 12


1 Bed Patient Room 1

No of O2 Outlet n= 25
No of Vacuum Outlet n= 25

Oxygen Demand Flow [QO2]:


HTM 2022 Table 6.0 Formula A.1
Q=10+(n-1)x6/3 n= 25
[QO2] = 10+(25-1)x6/3= 58 lpm
[QO2] = 58 lpm

Vacuum Demand Flow [QVAC]:


HTM 2022 Table 13.0 Formula A.1 and paragraf 4.71
Total No of Vacuum Outlet in the ward ,If less than 40
Vacuum Flowrate will be calculated as 40 lpm.
[Qv] = 40 lpm n= 25

2ND FLOOR PATIENT ROOMS C2-B2- [QO2] = 58 lpm [Qvac ] = 40 lpm.

DESIGN BY MEDIKAR of TURKEY PAGE NO:2


22.01.2022
SSK DÜZCE HOSPITAL MEDICAL GAS SUPPLY SYSTEM DESIGN

2ND FLOOR PATIENT ROOMS [QO2] = 116 lpm [Qvac ] = 80 lpm.

1ST FLOOR

C2 BLOCK-PATIENT ROOMS

2 Beds Patient Room 12


1 Bed Patient Room 1
No of O2 Outlet n= 25
No of Vacuum Outlet n= 25

Oxygen Demand Flow [QO2]:


HTM 2022 Table 6.0 Formula A.1
Q=10+(n-1)x6/3 n= 25
[QO2] = 10+(25-1)x6/3= 58 lpm
[QO2] = 58 lpm

Vacuum Demand Flow [QVAC]:


HTM 2022 Table 13.0 Formula A.1 and paragraf 4.71
Total No of Vacuum Outlet in the ward ,If less than 40
Vacuum Flowrate will be calculated as 40 lpm.
[Qv] = 40 lpm n= 25

1ST FLOOR PATIENT ROOMS C2 [QO2] = 58 lpm [Qvac ] = 40 lpm.

C1 BLOCK-PATIENT ROOMS

2 Beds Patient Room 9


1 Bed Patient Room 1
No of O2 Outlet n= 19
No of Vacuum Outlet n= 19

Oxygen Demand Flow [QO2]:


HTM 2022 Table 6.0 Formula A.1
Q=10+(n-1)x6/3 n= 19
[QO2] = 10+(19-1)x6/3= 46 lpm
[QO2] = 46 lpm

Vacuum Demand Flow [QVAC]:


HTM 2022 Table 13.0 Formula A.1 and paragraf 4.71
Total No of Vacuum Outlet in the ward ,If less than 40
Vacuum Flowrate will be calculated as 40 lpm.

[Qv] = 40 lpm n= 19

1ST FLOOR PATIENT ROOMS C1 [QO2] = 46 lpm [Qvac ] = 40 lpm.

DESIGN BY MEDIKAR of TURKEY PAGE NO:3


22.01.2022
SSK DÜZCE HOSPITAL MEDICAL GAS SUPPLY SYSTEM DESIGN
1ST FLOOR PATIENT ROOMS [QO2] = 104 lpm [Qvac ]= 80 lpm.

C1 BLOCK-NEONATAL UNIT
11 Bed Incubator Room 1
4 Bed Premature Baby Room 1
1 Bed Baby Milk Room 1
No of Beds nB=16

No of O2 Outlet n= 16
No of Medical Air 4 bar Outlet n= 16
No of Vacuum Outlet n= 16

Oxygen Demand Flow [QO2]

HTM 2022 Table 6.0 Formula C.4


Q=10+(n-1)6
[QO2] = 10+(16-1)x6= 100 lpm n= 16;
[QO2] = 100 lpm

Medical Air 4 Bar Demand Flow[QAir4]:


HTM 2022 Table 10.0 Formula C.4
Q=40*nB
[QAir4] = 40*16=640 lpm nB= 16;
[QAir4] = 640 lpm

Vacuum Demand Flow [Qv]:


HTM 2022 Table 13.0 Formula B.6
Q=40+(nB-1)40/4
[QV] = 40+(16-1)*40/4= 190 lpm nB= 16;
[QV] = 190 lpm

PREMATURE BABY ROOM:


[QO2] = 100 lpm
[QAir4] = 640 lpm
[QV] = 190 lpm

INTENSIVE CARE UNIT:

6 Bed ICU 1
No of ICU Beds nB= 6

No of O2 Outlet n= 12
No of Medical Air 4 bar Outlet n= 12
No of Vacuum Outlet n= 12

Oxygen Demand Flow [QO2]

HTM 2022 Table 6.0 Formula A.3

DESIGN BY MEDIKAR of TURKEY PAGE NO:4


22.01.2022
SSK DÜZCE HOSPITAL MEDICAL GAS SUPPLY SYSTEM DESIGN
Q=10+(nB-1)6
[QO2] = 10+(6-1)x6= 40 lpm nB= 6;
[QO2] = 40 lpm

Medical Air 4 Bar Demand Flow [QAir4]:


HTM 2022 Table 10.0 Formula A.3
Q=80+(nB-1)x80/2
[QAir4] = 80+(6-1)x80/2=280 lpm nB= 6;
[QAir4] = 280 lpm

Vacuum Demand Flow [Qv]:


HTM 2022 Table 13.0 Formula B.6
Q=40+(nB-1)40/4
[QV] = 40+(6-1)*40/4= 90 lpm nB= 6;
[QV] = 90 lpm

1st Floor ICU [QO2] = 40 lpm [QAir4] = 280 lpm [QV] = 90 lpm

OPERATION THEATERS
No of Operation Theaters = 5 T = 9
No of Delivery Room = 2
No of Small O.R. = 2

No of O2 Outlet n= 18
No of N2O Outlet n= 18
No of Medical Air 4 bar Outlet n= 18
No of Medical Air 7 bar Outlet n= 12
No of Vacuum Outlet n= 18

Oxygen Demand Flow [QO2]:


HTM 2022 Table 6.0 Formula D.1
Q=100+20(T-1)
[QO2] = 100+20(9-1)= 260 lpm T = 9;
[QO2] = 260 lpm

Nitrous Oxide Demand Flow [QN2o]


HTM 2022 Table 7.0 Formula A.1
Q=15+(nB-1)x6 nB = 9;
[QN2o] = 15+(9-1)x6 = 63 lpm
[QN2o] = 63 lpm

Medical Air 4 Bar Demand Flow [QAir4]:


HTM 2022 Table 10.0 Formula E.1
Q=40+(T-1)x40/4
[QAir4]: = 40+(9-1)x40/4 = 120 lpm T = 9;
[QAir4]: = 120 lpm

Medical Air 7 Bar Demand Flow [QAir7]:


HTM 2022 Table 12.0 Formula A.1
Q=350+(T-1)x350/4
[QAir7]: = 350+(6-1)x350/4 = 787,5 lpm T = 6;
DESIGN BY MEDIKAR of TURKEY PAGE NO:5
22.01.2022
SSK DÜZCE HOSPITAL MEDICAL GAS SUPPLY SYSTEM DESIGN
[QAir7]: = 787,5 lpm

Vacuum Demand Flow [Qv]:


HTM 2022 Table 13.0 Formula B.3
Q=(120x2)+(S-2)x120/2
[QV] =(120x2)+(9-2)x120/2= 660 lpm S = 9;
[QV] = 660 lpm

OPERATION THEATERS
[QO2] = 260 lpm
[QN2O] = 63 lpm.
[QAir4] = 120 lpm.
[QAir7] = 787,5 lpm.
[QV] = 660 lpm.

O.R PREPARATION ROOM

4 Bed O.R Preparation Room = 1

No of O2 Outlet n= 4
No of Medical Air 4 bar Outlet n= 4
No of Vacuum Outlet n= 4

Oxygen Demand Flow [QO2]:


HTM 2022 Table 6.0 Formula B.3
Q=10+(nB-1)6
[QO2] = 10+(4-1)x6= 28 lpm nB= 4;
[QO2] = 28 lpm

Medical Air 4 Bar Demand Flow [QAir4]:


HTM 2022 Table 10.0 Formula E.1
Q=40+(T-1)x40/4
[QAir4]: = 40+(4-1)x40/4 = 70 lpm T = 4 ;
[QAir4]: = 70 lpm

Vacuum Demand Flow [Qv]:


HTM 2022 Table 13.0 Formula B.6
Q=40+(nB-1)40/4
[QV] = 40+(4-1)*40/4= 70 lpm nB= 4;
[QV] = 70 lpm

O.R PREPARATION [QO2] = 28 lpm


[QAir4] = 70 lpm. [QV] = 70 lpm.

DESIGN BY MEDIKAR of TURKEY PAGE NO:6


22.01.2022
SSK DÜZCE HOSPITAL MEDICAL GAS SUPPLY SYSTEM DESIGN

LABOUR + RECOVERY ROOM

1 Bed Pain Room 3


2 Bed Pain Room 4
2 Bed Recovery Room 1
4 Bed Recovery Room 1

Total No of Beds nB = 17

No of O2 Outlet n= 17
No of Medical Air 4 bar Outlet n= 17
No of Vacuum Outlet n= 17

Oxygen Demand Flow [QO2]

HTM 2022 Table 6.0 Formula B.3


Q=10+(nB-1)6
[QO2] = 10+(17-1)x6= 106 lpm nB= 17;
[QO2] = 106 lpm

Medical Air 4 Bar Demand Flow [QAir4]:


HTM 2022 Table 10.0 Formula E.3
Q=40+(nB-1)x10/4
[QAir4] = 40+(17-1)x10/4=80 lpm nB= 17;
[QAir4] = 80 lpm

Vacuum Demand Flow [Qv]:


HTM 2022 Table 13.0 Formula B.5
Q=40+(nB-1)40/4
[QV] = 40+(17-1)*40/4= 200 lpm nB= 17;
[QV] = 200 lpm

LABOUR+RECOVERY ROOM [QO2] = 106 lpm


[QAir4] = 80 lpm
[QV] = 200 lpm

GROUND FLOOR

OPERATION ROOM
No of OR 2 T = 2

No of O2 Outlet n= 2
No of N2O Outlet n= 2
No of Medical Air 4 bar Outlet n= 2
No of Vacuum Outlet n= 2

DESIGN BY MEDIKAR of TURKEY PAGE NO:7


22.01.2022
SSK DÜZCE HOSPITAL MEDICAL GAS SUPPLY SYSTEM DESIGN

Oxygen Demand Flow [QO2]:


HTM 2022 Table 6.0 Formula D.1
Q=100+20(T-1)
[QO2] = 100+20(2-1)= 120 lpm T = 2;
[QO2] = 120 lpm

Nitrous Oxide Demand Flow [QN2o]


HTM 2022 Table 7.0 Formula A.1
Q=15+(nB-1)x6 nB = 2;
[QN2o] = 15+(2-1)x6 = 21 lpm
[QN2o] = 21 lpm

Medical Air 4 Bar Demand Flow [QAir4]:


HTM 2022 Table 10.0 Formula E.1
Q=40+(T-1)x40/4
[QAir4]: = 40+(2-1)x40/4 = 50lpm T =2;
[QAir4]: = 50 lpm

Vacuum Demand Flow [Qv]:


HTM 2022 Table 13.0 Formula B.3
Q=(120x2)+(S-2)x120/2
[QV] =(120x2)+(2-2)x120/2= 240 lpm S = 2;
[QV] = 240 lpm

OPERATION ROOM
[QO2] = 120 lpm
[QN2O] = 21 lpm.
[QAir4] = 50 lpm.
[QV] = 240 lpm.

GROUND FLOOR B

2 Bed Reanimation Room = 1


1 Bed Examination Room = 3
4 Bed Observation Room = 1
5 Bed Observation Room = 1
1 Bed Treatment Room = 1
1 Bed Plaster Room = 1

No of O2 Outlet n= 16
No of N2O Outlet n= 1
No of Medical Air 4 bar Outlet n= 2
No of Medical Air 7 bar Outlet n= 1
No of Vacuum Outlet n= 16

DESIGN BY MEDIKAR of TURKEY PAGE NO:8


22.01.2022
SSK DÜZCE HOSPITAL MEDICAL GAS SUPPLY SYSTEM DESIGN

Oxygen Demand Flow [QO2]

HTM 2022 Table 6.0 Formula B.3


Q=10+(nB-1)6
[QO2] = 10+(16-1)x6= 100 lpm nB= 16;
[QO2] = 100 lpm

Nitrous Oxide Demand Flow [QN2o]


HTM 2022 Table 7.0 Formula A.1
Q=15+(nB-1)x6 nB=1;
[QN2o] = 15+(1-1)x6 = 15 lpm
[QN2o] = 15 lpm

Medical Air 4 Bar Demand Flow [QAir4]:


HTM 2022 Table 10.0 Formula E.3
Q=40+(nB-1)x10/4
[QAir4] = 40+(2-1)x10/4=42,5 lpm nB= 2;
[QAir4] = 42,5 lpm

Medical Air 7 Bar Demand Flow [QAir7]:


HTM 2022 Table 12.0 Formula A.1
Q=350+(T-1)x350/4
[QAir7]: = 350+(1-1)x350/4 = 350 lpm T = 1;
[QAir7]: = 350 lpm

Vacuum Demand Flow [Qv]:


HTM 2022 Table 13.0 Formula A.1 and para 4.71
Total No of Vacuum Outlet in the ward ,If less than 40
Vacuum Flowrate will be calculated as 40 lpm.
[Qv] = 40 lpm n=16;

GROUND FLOOR B [QO2] = 100 lpm QN2O] = 15 lpm [QAir4] = 42,5 lpm
[QAir7] = 350 lpm [QV] = 40 lpm

GROUND FLOOR C

1 Bed Isolation Room = 1


10 Bed Observation Room = 1

No of O2 Outlet n= 11
No of Vacuum Outlet n= 11

Oxygen Demand Flow [QO2]:


HTM 2022 Table 6.0 Formula A.1
Q=10+(n-1)x6/3 n= 11
[QO2] = 10+(11-1)x6/3= 30 lpm
[QO2] = 30 lpm
DESIGN BY MEDIKAR of TURKEY PAGE NO:9
22.01.2022
SSK DÜZCE HOSPITAL MEDICAL GAS SUPPLY SYSTEM DESIGN

Vacuum Demand Flow [QVAC]:


HTM 2022 Table 13.0 Formula A.1 and para 4.71
Total No of Vacuum Outlet in the ward ,If less than 40
Vacuum Flowrate will be calculated as 40 lpm.
[Qv] = 40 lpm n=11 ;

GROUND FLOOR C [QO2] = 30 lpm [Qvac ] = 40 lpm.

BASEMENT

1 CT SCAN Room = 1

No of O2 Outlet n= 1
No of Medical Air 4 bar Outlet n= 1
No of Vacuum Outlet n= 1

Oxygen Demand Flow [QO2]

HTM 2022 Table 6.0 Formula B.3


Q=10+(nB-1)6
[QO2] = 10+(1-1)x6= 10 lpm nB= 1;
[QO2] = 10 lpm

Medical Air 4 Bar Demand Flow [QAir4]:

HTM 2022 Table 10.0 Formula E.3


Q=40+(nB-1)x10/4
[QAir4] = 40+(1-1)x10/4=40 lpm nB= 1;
[QAir4] = 40 lpm

Vacuum Demand Flow [Qv]:


HTM 2022 Table 13.0 Formula A.1 and para 4.71
Total No of Vacuum Outlet in the ward ,If less than 40
Vacuum Flowrate will be calculated as 40 lpm.
[Qv] = 40 lpm n=1;

BASEMENT
[QO2] = 10 lpm
[QAir4] = 40 lpm
[QV] = 40 lpm

DESIGN BY MEDIKAR of TURKEY PAGE NO:10


22.01.2022