Professional Documents
Culture Documents
● Objectives.—To determine the causes of excessive test test performed was a significant factor and was indepen-
turnaround time (TAT) and to identify methods of improve- dent of location: Chemistry–Multiple Test appeared most
ment by studying reasons for those tests reported in excess frequently (;40%), followed closely by Hematology–Com-
of 70 minutes from the time the test was ordered (ie, out- plete Blood Count (;20%) and Chemistry–Single Test
liers). (;18%). Factors of outlier TAT components for intensive
Design.—Self-directed data-gathering of stat outlier TAT care unit specimens were identified using statistical mod-
events from intensive care units and emergency depart- eling and included hour of day, type of health care per-
ments, with descriptive parameters associated with each sonnel collecting specimen, performing the test in a stat
event and additional descriptive parameters associated laboratory, and reason for delay. Outlier rates were not
with the participant. associated with any identified factors. The practice param-
Participants.—Laboratories enrolled in the 1996 College eters of laboratories with outlier rates in the lowest 10th
of American Pathologists Q-Probes program.
percentile significantly differed from those with rates in the
Main Outcome Measures.—Components associated with
top 10th percentile in test request computerization, report
outlier TAT events and outlier TAT rates.
Results.—Four hundred ninety-six hospital laboratories methods, and ordering methods.
returned data on 218 551 stat tests, of which 10.6% had Conclusions.—We observed that outlier analysis yields
TATs in excess of 70 minutes. Ten percent of stat emergen- new information, such as type of test and reason for delay,
cy department tests and 14.7% of stat intensive care unit concerning test delays when compared with TAT determi-
tests were outliers. Major areas in which delays occurred nation alone. Laboratories experiencing stat test TAT prob-
were test ordering, 29.9%; within-laboratory (analytic) lems should use this tool as an adjunct to routine TAT mon-
phase, 28.2%; collection of the specimen, 27.4%; postan- itoring for identifying unique causes of delay.
alytic phase, 1.9%; and undetermined, 12.5%. The type of (Arch Pathol Lab Med. 1999;123:607–614)
Table 7. Overall Stat Test Turnaround Time (TAT) Outlier Rates by Patient Care Location
No. of Total No. of Stat Test With Excessive
Location Institutions Stat Tests TAT, n (%)
Emergency department 484 189 148 18 924 (10.0)
Intensive care unit 380 29 403 4317 (14.7)
Total 496 218 551 23 241 (10.6)
significant (x2, P , .001). In the order-to-collection delay terval. The hours of order, collection, receipt, and verifi-
category (Table 3), Hematology–Complete Blood Count cation were significant factors for all dependent variables
(CBC) appeared as the test 5% more frequently than the (Figure 2). The person collecting the specimen (Table 8),
overall average for the category, whereas Hematology– the broad reason for delay, and type of test (Table 1) were
Other Than CBC appeared 5% less frequently than the also significant factors. Day of the week was a significant
overall average. Similarly, for the collection-to-receipt in- factor only for the time intervals of order to verification
terval (Table 4), Immunology/Serology–Single Test ap- and collection to verification. Testing performed in a stat
peared 5% less frequently than the average, and no test laboratory was a significant factor for all time intervals
occurred 5% more frequently than the average. For delays except receipt to verification.
occurring in the analytic phase (Table 5), Chemistry–Sin- The distribution of outlier rates for those hospitals in
gle Test and Immunology/Serology–Single Test appeared the lowest 10th percentile (highest performers), highest
5% more frequently than the average, and Hematology– 10th percentile (lowest performers), and those between the
CBC and Other appeared 5% less frequently than the av- highest and the lowest 10th percentiles is shown in Table
erage. Postanalytic (Table 6) and Undetermined appeared 10. Modeling of outlier rates using MANOVA produced
too infrequently (3.3% and 1.1%, respectively) by test type excellent correlation coefficients (r 2: ED 5 0.64, ICU 5
to classify in detail. 0.74). Although the correlation coefficients were excellent,
For outlier specimens obtained in the ED, MANOVA the probabilities found for the models were not statisti-
analysis resulted in statistically significant models with r 2 cally significant (P , .05). The extent of computerization
# 0.10 for all independent variables. Time of day (order- was the only significant factor found in the ICU model. A
ing, collection, and receipt hour) (Figure 2) and broad de- similar highest-lowest analysis of the ED outlier rate data
lay category factors (Table 9) were significant for all in- found laboratory monitoring of TATs and how the speci-
dependent variables. Day of the week was a significant mens were transported to be significant factors. Table 11
factor for collection-to-receipt and receipt-to-verification shows the number of participants in each group (1st–10th,
times only. Performing the test in a stat laboratory and the 10th–90th, and 90th–100th percentiles) for each significant
type of test (Table 1) were significant factors for all time practice parameter.
intervals except ordering to collection.
Analysis of outlier factors for ICU specimens by MAN- COMMENT
OVA also produced statistically significant models for all This study attempted to find common causes of outlier
dependent variables. The correlation coefficient for the de- TAT events (arbitrarily defined as an event having a total
pendent variables ranged from 0.20 for the order-to-col- TAT from the time of test order to the time of test result
lection time interval to 0.38 for the collection-to-receipt in- verification in excess of 70 minutes) resulting from stat
610 Arch Pathol Lab Med—Vol 123, July 1999 Turnaround Time Outliers—Steindel & Novis
Table 8. Practice Parameters by Ordering Location studies generally produces a sufficient number of data
points to allow the use of modeling.2,4 Modeling also re-
Emergency Intensive
quires homogeneity within the process, such that the mod-
Department, Care
Practice Parameter % Unit, % el describes the data adequately (and by inference the pro-
cess), as measured by the degree of correlation (r 2). One
Testing done in a stat laboratory difficulty in modeling data is that modeling is sensitive to
n 19 488 4380 diversity among the participants with respect to the se-
Yes 14.8 26.9
No 85.2 73.1
lected factors, as observed by the low correlation coeffi-
cients and few significant parameters. Our study popu-
Test type lation was quite diverse, but the final models identified a
n 19 528 4375 few common significant practice parameters with a high
Chemistry degree of statistical confidence.
Single test 18.0 21.0 Intensive care units were found to have a 4.7% higher
Multiple tests 45.7 34.9 outlier rate than EDs (Table 7). These rates were not found
Hematology to be associated with any practice variables from either
Complete blood count 21.2 20.3 location. Lack of common associations for differences in
Not complete blood count 1.5 1.3 outlier rates indicates the institutional uniqueness of these
Coagulation events, a recognized attribute of health care quality as-
Single Test 4.3 10.8 surance practices. Indeed, the history of present health
Multiple Tests 7.1 10.7 care quality assurance documents many instances in
Immunology/serology which laboratories monitor their performance, compare
Single test 1.1 0.4
that performance with benchmark levels, and then develop
Multiple tests 0.1 0.1 their own methods of improvement.17,18
Our attempt to find significant reasons why one group
Other 1.0 0.5
of participants had extremely low outlier rates while oth-
Collection Personnel ers had extremely high rates produced models describing
n 19 428 4299 the data well (r 2 . 0.6), but the probability of fit for the
Laboratory personnel 38.3 29.1 models was not statistically significant. For these data sets,
Physician/house officer 2.5 1.5
Medical student 0.0 0.0 the number of participants in the highest and lowest
Nursing services 50.1 63.2 groups was too small (Table 10) to confidently attribute
Other 3.3 0.6 practice variables to indications of improved performance.
Do not know 5.0 5.6 Although the statistical fit of the models was not adequate,
Day of the week we did observe significant factors relating to practice pa-
n 19 538 4380 rameters that were consistent with those from previous
Sunday 15.9 16.4 studies (Table 11). More of the lowest 10th percentile per-
Monday 14.8 15.3 formers from the ICU received requests from a laboratory
Tuesday 15.0 17.1 information system than from a hospital information sys-
Wednesday 14.1 15.3 tem, a pattern found previously in ED4,5 and routine test3
Thursday 14.4 12.9
Friday 13.2 11.9 TAT studies. From the ED, that same group did not mon-
Saturday 12.8 11.0 itor TAT and fewer monitored it manually, both factors
found important in maintaining or improving ED TAT.5
The factor not found in prior studies was the order trans-
mission method, where more of those among laboratories
testing. Based on prior studies,4,5 this definition was pre- in the highest 10th percentile used the telephone for ED
dicted to yield an outlier rate of about 10%. Our actual stat test orders. From these statistically identified factors,
results yielded an outlier rate of 10.6% (Table 7), indicating we can conclude that to control outlier rates, it is impor-
that the underlying distribution of TATs found in 1990 and tant for the laboratory to receive and act on an order in a
1993 had not changed significantly at the time this study timely fashion. It must then control the TAT process
was conducted (1996). through monitoring.
This study also investigated the differences in the rate Most individual outlier events were due to causes un-
of outlier events among participants. As with all Q-Probes related to the actual test determination (Figure 1). Person-
studies, conclusions are confounded by self-reported and nel problems (primarily staff shortages) were a major
unverified data. Individual laboratories will interpret the cause of delays and occurred predominantly in the test
supplied instructions in unique fashions that are unantic- ordering (37.8%), collection (51.4%), and analytic (33.7%)
ipated by the study designers. In addition, each laboratory phases of the process. Problems that related to performing
has unique solutions to providing optimum TATs based tests were responsible for only 10.9% of the overall delays.
on their own observations. To overcome these obstacles in Ross and Boone19 first reported on the low percentage of
study design and to provide unbiased identification of fac- errors (,20%) associated with the analytic phase of the
tors significant in outlier events, we used statistical mod- total testing process; other studies have confirmed this
eling of the data. finding.20,21 Our outlier analysis only reinforces the recur-
Such modeling is successful in identifying common fac- ring theme that many laboratory errors occur before the
tors causing outliers only when sufficient data points for specimen arrives.
the statistical tools, as measured by the model’s P values, Many outlier events (12.5%) did not have an identified
are available. Past experience with TAT data indicated that reason for delay and within the major delay categories, Un-
the large number of laboratories participating in Q-Probes determined accounted for approximately 20% of responses.
Arch Pathol Lab Med—Vol 123, July 1999 Turnaround Time Outliers—Steindel & Novis 611
Table 9. Practice Parameters by Broad Delay Category
Broad Delay Category, %
Receipt:
Ordering of Collection of Receipt: Analytic Postanalytic
Practice Parameter Test Specimen Phase Phase Undetermined
Test performed in a stat laboratory
n 5918 5490 11 038 751 264
Yes 14.7 24.8 14.8 8.7 36.4
No 85.3 75.2 85.2 91.3 63.6
Test type
n 5493 5513 11 046 770 264
Chemistry
Single test 16.1 15.0 21.3 19.2 28.4
Multiple tests 43.1 45.1 44.1 43.1 33.3
Hematology
Complete blood count 25.2 23.9 17.2 20.0 17.4
Not complete blood count 1.1 25.7 1.5 2.9 2.3
Coagulation
Single test 5.7 23.9 5.2 3.9 10.2
Multiple tests 6.7 7.3 8.6 8.0 6.1
Immunology/serology
Single test 0.8 0.6 1.3 0.8 1.9
Multiple tests 0.1 0.1 0.1 0.1 0.0
Other 1.0 0.8 0.7 2.0 0.4
Collection personnel
n 5927 5483 11 001 770 179
Laboratory personnel 42.0 18.8 41.3 52.9 27.4
Physician/house officer 0.2 4.2 2.5 5.4 0.0
Medical student 0.0 0.0 0.0 0.0 0.0
Nursing services 53.2 64.5 48.8 38.7 33.0
Other 3.0 2.4 2.9 0.4 16.2
Do not know 1.7 10.1 4.5 2.6 22.9
Knowledge of where outliers arise is the first step in de- mands during periods of high workloads. Analyzing work
termining how to correct them. For example, about 38% of assignments during these times may resolve apparent staff
the outlier delays were attributed to the 4 categories relating shortages without the need for additional staff; for example,
to possible personnel shortages (Table 1). Staff shortages cross-training or redistribution of work tasks, such as pre-
may be related to an inability of existing staff to meet de- ventive maintenance, may alleviate such staffing problems.
612 Arch Pathol Lab Med—Vol 123, July 1999 Turnaround Time Outliers—Steindel & Novis
Table 10. Frequency Distribution of Outlier Rates as tified using the good ICU models were similar to those
Percent of Testing from the weak ED models. Two common factors were as-
sociated with time, namely, day of the week and time of
Emergency Intensive
Department Care Unit day that events occurred. It is possible that institution ad-
mission patterns tend to cluster testing naturally around
Lowest 10th percentile (least percentage of outliers) certain days (Table 8), primarily weekdays, and certain
n 44 22 times (Figure 2), primarily during the daylight hours.
Maximum 2.0 2.0
75th percentile 1.6 1.8
These observations may be useful in helping to arrange
50th percentile (median) 1.0 1.0 staff assignments to cope with personnel shortages. Two
25th percentile 0.5 0.6 other common significant factors resulting in outliers were
Minimum 0.2 0.2 type of health care worker collecting the specimen and
10th to 90th percentile whether the test was performed in a stat laboratory. These
n 372 196 factors were previously found to be significantly related
Maximum 26.9 40.0 to overall ED TAT.5 The last 2 significant factors identified
75th percentile 13.1 22.1 were the broad reason for delay and the test type. Delays
50th percentile (median) 8.5 11.7 that occurred with chemistry tests (Table 9) were most
25th percentile 4.8 6.1 likely due to the time-consuming specimen processing
Minimum 2.1 2.4
that is generally required.4
Highest 10th percentile (greatest percentage of outliers) The most frequently delayed test varied with the broad
n 47 25 delay categories. In the analytic phase, Chemistry–Single
Maximum 187.5* 100.0 Test and Immunology/Serology–Single Test were delayed
75th percentile 46.2 75.0
50th percentile (median) 32.4 50.0
most frequently, perhaps due to analytic processing time,
25th percentile 29.5 42.9 a conclusion supported by noting that Hematology–CBC,
Minimum 28.0 40.4 which has a short analytic processing time, rarely had de-
* Data were self-reported and unverified, resulting in logical incon- lays attributable to this delay group. The other 2 broad
sistencies, such as an emergency department outlier rate greater than delay categories that had significant test groups were in
100% reported by one participant. Four participants reported emergen- the preanalytic phases, order-to-collection time and collec-
cy department outlier rates of 100%. tion-to-receipt time. There is no apparent reason that the
tests observed, Hematology–Other Than CBC and Im-
munology/Serology–Single Test, should stand out.
Table 11. Participant Responses to Significant Factors The homogeneity of individual ICU outlier data allowed
from Highest/Lowest Outlier Rate Analysis (No. of us to make observations concerning each TAT interval.
Participants) Some factors were not significant in all TAT intervals. Day
90th– of the week was significant only in the initial order-to-
0–10th 100th verification and overall collection-to-verification phases,
Percentile 10th–90th Percentile
but its effect was sufficiently large to also appear as a
Factor (Highest) Percentile (Lowest)
factor when all data were combined, perhaps owing to
Intensive care unit: extent of computerization patient and workload distributions. This factor may not be
Requests by hospital amenable to laboratory control. Testing in a stat laboratory
information systems 15 146 18 appeared as a statistically significant factor in all phases
Requests by laboratory
information systems 6 6 36
except the receipt-to-verification phase. It has been hy-
No computer 0 13 1 pothesized that stat laboratories are not equipped to han-
dle high-volume workloads and delays occur when these
Emergency department: turnaround time monitoring
tests form a queue.5 The delays attributed to stat labora-
Monitor by computer 28 270 28
Monitor manually 8 38 6
tory use found in this study do not occur in the testing
No monitor 8 62 13 phase, but rather in those phases that would have a queue,
supporting the earlier observation. It appears that once a
Emergency department: stat order transmission method
laboratory enters the analytic phase, testing is timely.
Messenger 0 1 1
Telephone 16 48 1
Our study found that outlier investigation is a useful
Intercom 1 2 0 tool for general stat TAT monitoring. Information concern-
Fax 0 2 0 ing the causes of extreme TATs was similar to those of
Computer 22 206 24 overall TATs, but some new factors did emerge. Lack of
Pager strong statistical models, however, indicates the local na-
Care area to ture of many factors influencing outlier events. We rec-
phlebotomist 1 13 0 ommend that laboratories add outlier investigation to their
Laboratory area to TAT quality assurance programs both on a periodic basis
phlebotomist 0 2 0 and when they suspect developing TAT problems.
A system to monitor outlier TATs is easy to establish.
The first step is to set your criteria for outliers. This study
Causes of ED outlier events were too diverse to produce assumed a 10% outlier rate and used a criterion of 70 min-
good modeling (low r 2). Intensive care unit outlier events utes from order to reporting time. Each laboratory should
were better related among participants, and strong models determine the distribution of outlier TATs in its own in-
could be developed for each of the TAT components, al- stitution and set a rate at the TAT seen when a sufficient
lowing identification of statistically significant practice pa- volume of outlier specimens (recommendation, 10%) is ob-
rameters. Statistically significant practice parameters iden- served. When investigating outliers, which may occur as
Arch Pathol Lab Med—Vol 123, July 1999 Turnaround Time Outliers—Steindel & Novis 613
a continual process or in response to a specific event, de- ratory tests: a College of American Pathologists Q-Probes study of 653 institutions.
Clin Chim Acta. 1996;251:25–40.
termine the cause of delay for all specimens exceeding the 4. Howanitz PJ, Steindel SJ, Cembrowski GS, Long TA. Emergency department
established outlier criteria so that there are as few un- stat test turnaround times: a College of American Pathologists’ Q-Probes study for
knowns as possible. This study indicates most causes will potassium and hemoglobin. Arch Pathol Lab Med. 1992;116:122–128.
5. Steindel SJ, Howanitz PJ. Changes in emergency department turnaround
be unique to an individual laboratory, but several common time performance from 1990 to 1993: a comparison of two College of American
causes of delay were observed. If an institution has certain Pathologists Q-probes studies. Arch Pathol Lab Med. 1997;121:1031–1041.
6. Novis D, Steindel SJ. Small Hospital Emergency Department Turnaround
admission or ordering patterns centered around times of Time: Data Analysis and Critique. Q-Probes 93-12. Northfield, Ill: College of
day or day of week, the laboratory may see outliers fol- American Pathologists; 1993.
lowing those patterns. Adjusting staffing schedules may 7. Steindel SJ. Timeliness of clinical laboratory tests: a discussion based on five
College of American Pathologists Q-Probe studies. Arch Pathol Lab Med. 1995;
help that situation. Certain test types may be delayed 119:918–923.
more often than others. If chemistry tests are frequently 8. Valenstein PN, Emancipator K. Sensitivity, specificity, and reproducibility of
delayed, consider using methods that do not require spec- four measures of laboratory turnaround time. Am J Clin Pathol. 1989;91:452–
457.
imen processing or optimize the specimen processing 9. Grant EL, Leavenworth RS. Statisical Quality Control. New York, NY: Mc-
area. If delays are in the preanalytic phase, which is the Graw-Hill Inc; 1980.
10. United States v Barr Laboratories, Inc, Civil Action No. 92–1744. United
most common source of delays, study the collection and States v Barr Laboratories, 812 F Supp 458; 1998. 1993 US Dist Lexis 1932, US
transport processes used. Even if nonlaboratory personnel District Court for the District of New Jersey, as amended 30 March 1993.
perform these tasks, the laboratory should control them; 11. Thayer AM. Process safety is no accident. Chem Engineering News. 1998;
76:15–17.
lack of laboratory control in the preanalytic phase is a 12. Dale JC, Howanitz PJ. Patient Satisfaction and Phlebotomy Service: Data
common cause of delay.22 Finally, start to monitor TATs Analysis and Critique. Northfield, Ill: College of American Pathologists; 1994.
and react to outliers before others notice significant prob- 13. Novis DA, Zarbo RJ, Saladino AJ. Interinstitutional comparison of surgical
biopsy diagnosis turnaround time: a College of American Pathologists Q-Probes
lems. Continual investigation of outliers will help achieve study of 5384 surgical biopsies in 157 small hospitals. Arch Pathol Lab Med.
this last goal. 1998;122:951–956.
14. Novis D, Steindel SJ. Stat Test Turnaround Time Outliers: Data Analysis and
This study did not demonstrate clear reasons for outli- Critique. Q-Probes 96-02. Northfield, Ill: College of American Pathologists; 1997.
ers, as first hypothesized. It does demonstrate, however, 15. GLM-VARCOMP. In: SAS/STAT User’s Guide. Version 6, vol 2. Cary, NC:
that investigation of outliers adds new details not found SAS Institute Inc; 1990:891–996.
16. Neter J, Wasserman W, Kutner MH. Applied Linear Regression Methods.
when monitoring overall TATs. These details include such Boston, Mass: Richard D Irwin Inc; 1989.
items as test type, day of the week, and time of day and 17. Bigelow B, Arndt M. Total quality management: field of dreams? Health
Care Manage Rev. 1995;20:15–25.
are most likely specific to each institution. Studying out- 18. Luce JM, Bindman AB, Lee PR. A brief history of health care quality as-
liers can help laboratories find TAT bottlenecks in the sessment and improvement in the United States. West J Med. 1994;160:263–268.
same way the process is used in many other service in- 19. Ross JW, Boone DJ. Assessing the effect of mistakes in the total testing
process on the quality of patient care [abstract 102]. In: Martin ML, Addision BV,
dustries. Wagner WM, Essien JDK, eds. 1989 Institute of Critical Issues in Health Labo-
ratory Practice. Minneapolis, Minn: Dupont Press; 1991:173–173.
References 20. Boone DJ, Steindel SJ, Herron R, et al. Transfusion medicine monitoring
practices: a study of the College of American Pathologists/Centers for Disease
1. Valenstein P. Laboratory turnaround time. Am J Clin Pathol. 1996;105:676– Control and Prevention Outcomes Working Group. Arch Pathol Lab Med. 1995;
688. 119:999–1006.
2. Howanitz PJ, Steindel SJ. Intralaboratory performance and laboratorians’ ex- 21. Plebani M, Carraro P. Mistakes in a stat laboratory: types and frequency.
pectations for stat turnaround times: a College of American Pathologists Q-Probes Clin Chem. 1997;43:1348–1351.
study of four cerebrospinal fluid determinations. Arch Pathol Lab Med. 1991;115: 22. Steindel SJ, Howanitz PJ. Emergency Department Turnaround Time: Data
977–983. Analysis and Critique. 93-04. Northfield, Ill: College of American Pathologists;
3. Steindel SJ, Jones BA, Howanitz PJ. Timeliness of automated routine labo- 1993.
614 Arch Pathol Lab Med—Vol 123, July 1999 Turnaround Time Outliers—Steindel & Novis