FCI flow calibration laboratory

Leading the process

industry with

state-of-the-art

calibration facilities
With constant demand for higher

instrument performance, FCI has

dedicated over 40 years of flow research

and development into the construction

and operation of the process industry’s

leading flow calibration facility.


FCI. The process industry measurement standard.
Fluid Components International is recognized across the process (250 mm). Calibrations for applications with temperature ranges
control industry for precision calibration services and highly from -100 to +1000ºF (-73 to +538ºC) and pressure ranges
reliable flow metering instrumentation. From early designs, from 0 to 1000 psig (0 to 68 atmospheres) are commonly
FCI has focused closely on replicating field conditions in a performed for many fluid services.
controlled laboratory environment.
This assures the highest installed AVAL sizing software
accuracy and minimum metering and instrument
uncertainty. Today, we operate a selection services
state-of-the-art flow calibration FCI has developed sizing and
laboratory with versatility to quickly specification software to make
transition through a wide range of selection of thermal mass and
calibration fluids, line sizes, and Coriolis mass flow instruments
process connections. We proudly quick and reliable. Since process
feature extensive automation and parameters, fluid compositions
data collection capabilities that and installation constraints can
offer unmatched proprietary limit ideal performance, FCI AVAL
calibration solutions that balance and ASAP software will ensure
throughput efficiency with trace- customers are fully aware of pro-
ability and reliability. cess conditions that can potentially
produce installed uncertainty.
NIST AVAL application evaluation soft-
ISO 9001:2000 and ware is a proprietary FCI service
AS9100 Certification designed from years of fluid testing
FCI calibrations are performed and installation experience. Using
utilizing only NIST National AVAL software, FCI models an
Institute for Standards and installation with straight run
Technology traceable equipment variations and obstructions. All
and instrumentation. We pride fluid process conditions are input
ourselves on meeting MIL-STD-45662A, ANSI/NCSL Z-540 to run a suitability evaluation. When a complete evaluation is
requirements and for continuously maintaining performed, an output report clearly defines expected “installed”
ISO 9001:2000 and AS9100 certification. accuracy or offers recommendations for improvement. Different
from other sizing software, FCI relates laboratory calibrations
Always extending flow lab facilities to meet new field conditions, to true installed field performance. ASAP configuration software
FCI features gas flow calibration capabilities ranging as low as matches specific process conditions and makes appropriate
0.001 SCFM (.00017 NCMH) to ranges that exceed 5000 SCFM configuration recommendations for each flow meter application
(8500 NCMH) and higher for line sizes in excess of 10 inches prior to purchase.
Reference metering technologies In large line gas applications such as stacks and air ducts,
Due to the wide range of operating conditions, FCI has selected turbine meters are selectively utilized as flow stand reference
and utilizes proven reference flow standard technologies that meters. Turbines are incrementally selected on bypass sections
provide the maximum reliability and flexibility needed for a such that only the linear output range is utilized. Narrowing
particular range of process conditions and fluids. the useable range enables the highest accuracy calibrations.
Turbine meters are combined with dual temperature and
For maximum reliability in both pressure instrumentation to derive
Sonic nozzle primary meters
air and gas applications, sonic and output real-time mass flow.
nozzles or critical venturis (CVs)
are utilized for both low flow and Automation and
small line size applications. When data acquisition
combined with high accuracy database
pressure standards, sonic nozzles FCI’s Test Engineering group is
deliver industry leading accuracy continuously developing new
and repeatability. Since turndown methods for reducing uncertainty,
is limited on sonic nozzles, FCI improving data collection and
utilizes a sonic nozzle sizing rack recall, expanding rangeability,
to perform calibrations for wide developing automation routines
flow range applications up to and designing new calibration
1000:1 turndown. test stands.

FCI utilizes a portable metrology Automation activities are done


cart designed to travel to each with three mandated objectives
production flow stand and to service outlined and managed before each
scheduled metrology activities and improvement is initiated:
calibration cycles. Accordingly, all  1 Automation activities must

flow laboratory reference meters improve calibration accuracy and


are traceable directly to the nozzle stand stability. Software is developed
standards, or for elevated ranges are to optimize settling time and to
traceable to third party laboratory integrate real time measurement
affiliates. In all cases, every flow reference standard is directly and equipment inputs. As automation continues, human error
traceable to NIST. For optimum mass flow measurement in is designed out of the data collection process.  2 Automation
liquid applications, FCI utilizes Coriolis type standards. The must service throughput improvements. Customers no longer
inherent “mass flow” capability of Coriolis technology assures accept long waits for instrument calibrations no matter how
no need for auxiliary process measuring instruments in order complex the application. Equally important, customers can not
to achieve high accuracy mass flow indication. Flow laboratory sacrifice instrument reliability for turnaround.  3 Improved
accuracy of 0.10% of reading is attainable depending on fluids access to critical calibration data, parameters, flow conditions
and process conditions. and instrument variables. Whether a factory service technician is
in the field or whether a customer needs specific instrument details, every Exclusive calibration documents
calibration variable and instrument detail must be immediately and FCI customers are provided certified calibration
reliably accessible. documents that summarize all calibration
parameters, indicate actual instrument
Electronic calibration library performance during system to system validation,
At FCI, all instrument calibration records are maintained in FCI’s and provide detailed setup information useful for
electronic calibration library.  Every flow meter has a calibration finger troubleshooting or field calibration adjustments.
print that can be accessed either by performing an instrument inquiry This data is additionally accessible via FCI’s
using software contained in the meter or by calling up a serial number website and can be viewed and printed by
address on the world wide web and accessing complete calibration files. registered customers.

Certified calibrations Calibration


summary and
Computer generated and electronically stored calibration documents output chart
describe specific instrument details and can be easily sorted by instrument Records:
serial number, tag number or customer purchase order. Process condi- › Application conditions
tions, calibration fluid, line size and other relevant calibration specifics are › Instrument ID
detailed for each calibration. A certified flow curve table is provided that
› Computer generated
flow curve look up
table
matches instrument current outputs with scaled units of flow.

All NIST traceable equipment utilized during instrument calibration is


Calibration
identified along with the calibration history on all reference equipment.
certificate
ISO calibration procedures followed during the instrument calibration are Records:
listed and certified by the performing calibration technician. › Applicable NIST
traceable equipment
› Calibration procedures

Benefits to customers › Technician
certified stamp
Automation initiatives have continued to deliver the following
benefits for FCI customers:
› FCI is able to offer high value “actual fluid and field condition”
calibrations at prices comparable to fluid equivalency calibrations
Delta R
offered by others. data sheet
› Technicians simultaneously operate multiple calibration stands Records and
enables:
to provide faster instrument throughput and timely calibration › Raw and
turnaround. processed outputs
› Calibration
› Automated processes result in consistent calibrations free verification
of human error. › Tolerance limits
› Storing all data in a database means quicker access and facilitates
process analysis tools.
› Improved customer service response times.
Accurate data acquisition systems then automatically collected over the selected flow range.
All FCI data collection instruments feature NIST traceable Individual programs at each step in the calibration cycle add
calibrations and are systematically checked to verify calibration information to the calibration record much like a virtual data
cycle compliance. FCI flow stands use independent meters for conveyer belt. Following data collection, the flow stand performs
data collection and signal processing. This ensures that incoming a curve fit linearization routine then downloads coefficients into
signals are isolated from any potential PC produced noise effects. the calibrated instrument memory. After the data download phase
In addition, all instruments are is completed, a NIST compliant
Calibration cycle compliance
installed with careful isolation system-to-system flow check is
from line induced or radiant type performed. As the newly calibrated
electrical and RFI interferences. All flow meter passes the final verifi-
measured signals are wired to multi cation flow check-points, all
channeled data collection devices. calibration data is then sent
All devices interface digitally with electronically to a retrievable
the data acquisition systems to database and calibration
eliminate transfer or conversion documents are generated.
errors. The result is uncompromised
reliability in data generation. VORTAB ®
solves straight
Fail safe calibrations run deficiencies
Automated data collection provides Ideally, both process flow meas-
all inputs for real time calculation urement and  actual flow meter
of pressure, temperature and mass calibration should be performed
flow. If any of the signals or under conditions where pipe
corresponding digital calculations straight run is available. Actual
fall outside of the limits retrieved installations with inadequate pipe
from the database, calibration is diameters will usually experience
stopped until corrected. The second flow profile disturbances that
function of the software is to auto- directly result in degraded instru-
matically find a target flow rate ment accuracy. Often piping
and efficiently bring the stand to practices are dictated more by
stable equilibrium. limited plant space than by instrumentation needs. As a result,
to assure that calibration laboratory accuracy is transferred to the
Virtual data conveyer belt field installation, flow conditioning may be recommended. To
After installing the flow meter into the flow stand, a qualified test eliminate inaccuracies associated with velocity profile distortions,
technician electronically retrieves test parameters and flow stand FCI has partnered with VORTAB flow conditioners to isolate flow
setup instructions to begin the calibration process. The flow irregularities and to provide a swirl free symmetric flow profile
stand directly communicates with the instrument being that is reliably transferred from the lab to the field. FCI also
calibrated to download initial settings for the application. Data is features laminar, flat, turbulent and custom profile calibrations.
Gas sonic nozzle stand

Proprietary software equals reliability and repeatability

Customer application data and flow meter Automated software guarantees that all The FCI gas sonic nozzle stand is operated
performance results are easily queried from flow stand equipment is up to date and remotely from a separate control room.
the calibration database. Analysis of this NIST traceable. Computerized process control allows fully
information leads to constantly improving automated calibrations and efficient product
products and test facilities. development testing.
Pressurized gas loop stand

Air sonic nozzle stand Air high temperature stand


European Calibration Center Liquid hydrocarbon stand

FCI offers complete calibration solutions to service European


customer needs for quick turnaround and high reliability.
FCI’s Service and Calibration Center located in our Tilburg
(Netherlands) facility features local European access to traceable
flow stations and is ideal for validation testing, recalibration
services and cycle audits.

Web enabled trouble shooting


Database management is a constantly improving area. With the
application versatility available on the world wide web, FCI can
setup individual instrument addresses and can perform remote
flow meter diagnostic routines directly from the factory. For flow
meters installed in the field without web addresses, access to
calibration records is available to password protected customers for
parameter “read only” access thus making field data collection
and performance comparisons available around the clock.

Field calibration services


Where extreme or challenging installation conditions exist,
FCI provides reliable “in situ” calibration services with highly
qualified field calibration engineers. For customers who are
Sanitary liquid stand
mandated to perform annual calibration validations, FCI
features mobile audit and certification services designed to
meet customer compliance cycles.

Productivity equals customer benefits


Through stand automation and flexible fixturing, throughput
at each station is continually optimized. Automated data entry
maximizes processes speed and accuracy. On-site liquid
and gas reserves means minimal waiting for either basic fluids
or complex mixed gas calibration compositions. Self checking
calibration stands prevent out of tolerance calibrations. By
combining application evaluation services with accurate, reliable
throughput, FCI continues industry leadership with state-of-the-
art calibration services and facilities. 
Six additional flow stands available in Europe. *Applications may be done using fluid properties equivalency.

Applications We are unique among flow service providers because of our


The FCI calibration laboratory features an evolving flow ability to perform in a wide range of gases, gas mixtures,
measurement facility. This one of a kind flow laboratory services liquids, temperatures, pressures, line sizes, and process
the widest range of fixed station and modular calibration rigs. connections. Above are samples of common applications
Each calibration bench is designed to dynamically service a serviced in our facility. Many other fluid service capabilities are
wide range of calibration parameters. available along with special calibration setups.
Bypass gas loop

Air Medium Temperature

10″ wind tunnel “D” and “E”

Gas flow facility

Contact FCI for other custom configurations.

Flow calibration test stand parameters that ensures scheduled traceability verification. Lock-out software
FCI maintains and operates a wide range of calibration test has been developed to pre-test every calibration stand with a
stands in  order to perform calibrations that closely match a start up interrogation  that validates current flow stand
diverse range of customer applications. Each instrument and compliance and proper stand operation. Operators may only
all calibration equipment utilized in the laboratory or located perform calibrations when all test stand instrumentation is
on the calibration stand is placed under a metrology program within its traceable calibration cycle.
Air flow facility

Headquartered in San Marcos, California, FCI is near San Diego’s


high tech and biomedical industrial centers with excellent airport,
freeway and sea transportation to support its global customer base.

Locally Represented By: Visit FCI online at www.FluidComponents.com | FCI is ISO 9001:2000 and AS9100 Certified
FCI World Headquarters
1755 La Costa Meadows Drive | San Marcos, California 92078 USA
Phone: 760-744-6950 Toll Free (US): 800-854-1993 Fax: 760-736-6250

FCI Europe
Persephonestraat 3-01 | 5047 TT Tilburg, The Netherlands | Phone: 31-13-5159989 Fax: 31-13-5799036

FCI Measurement and Control Technology (Beijing) Co., LTD | www.fluidcomponents.cn


Room 107, Xianfeng Building II, No.7 Kaituo Road, Shangdi IT Industry Base, Haidian District | Beijing 100085, P. R. China
Phone: 86-10-82782381 Fax: 86-10-58851152

1210 2K Doc. No. 02MK011508D


© Copyright 2006 by Fluid Components International LLC. All rights reserved. Manufactured in accordance with one or more of the following patents: US Patent Numbers: 6,340,243, 6,843,110
FCI is a registered trademark of Fluid Components International LLC. Information subject to change without notice.

You might also like