You are on page 1of 12

Infrared Physics & Technology 52 (2009) 97–108

Contents lists available at ScienceDirect

Infrared Physics & Technology


journal homepage: www.elsevier.com/locate/infrared

Infrared thermography on ocular surface temperature: A review


Jen-Hong Tan a, E.Y.K. Ng a,*, U. Rajendra Acharya b, C. Chee c
a
School of Mechanical and Aerospace Engineering, College of Engineering, Nanyang Technological University, 50, Nanyang Avenue, 639798, Singapore
b
Department of Electronics and Computer Engineering, Ngee Ann Polytechnic, Singapore
c
Medical and Surgical Retinal Centre Department of Ophthalmology, National University Hospital, 119074, Singapore

a r t i c l e i n f o a b s t r a c t

Article history: Body temperature is a good indicator of human health. Thermal imaging system (thermography) is a non-
Received 2 December 2008 invasive imaging procedure used to record the thermal patterns using Infrared (IR) camera. It provides
Available online 18 May 2009 visual and qualitative documentation of temperature changes in the vascular tissues, and is beginning
to play an important role in the field of ophthalmology. This paper deals with the working principle,
Keywords: use and advantages of IR thermography in the field of ophthalmology. Different algorithms to acquire
Eye the ocular surface temperature (OST), that can be used for the diagnosis of ocular diseases are discussed.
Infrared
! 2009 Elsevier B.V. All rights reserved.
Thermogram
Temperature
Cornea
Ocular

Contents

1. Introduction . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 97
2. The principles of infrared thermography on human eye . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 98
2.1. The physics of IR radiation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 99
2.2. The IR thermography system . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 99
2.3. The measured OST. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 100
3. Methodologies in the study of OST . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 100
3.1. Manual measures in the acquisition of OST . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 100
3.2. Semi-automated method in the acquisition of OST . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 101
3.3. Automated method in the acquisition of OST . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 102
4. Application of IR thermography to ocular studies . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 103
4.1. Studies of ocular physiologies . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 103
4.2. Studies of ocular diseases and surgery . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 104
4.3. Studies of non-ocular diseases . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 105
5. Discussion . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 105
References . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 107

1. Introduction techniques. Galileo invented the thermoscope in 17th century and


then it evolved into modern mercury-in-glass thermometer, radi-
Temperature has been used to investigate the physiological and ometers for middle-ear temperature and disposable sterile ther-
pathological changes in human body since 400 B.C. using different mocouple which is widely used these days. Liquid crystals such
as cholesteric esters, found in 1877 had the property of changing
color with temperatures, were used to display the distribution of
* Corresponding author. Address: Adjunct NUH Scientist, Office of Biomedical temperature on skin topographically. Such detectors were inex-
Research, National University Hospital of Singapore. Tel.: +65 67904455; fax: +65
pensive, had relatively short life span and may alter surface tem-
67911859.
E-mail address: mykng@ntu.edu.sg (E.Y.K. Ng). perature due to large area of contact. Schlerian photography

1350-4495/$ - see front matter ! 2009 Elsevier B.V. All rights reserved.
doi:10.1016/j.infrared.2009.05.002
98 J.-H. Tan et al. / Infrared Physics & Technology 52 (2009) 97–108

enabled researchers to see the convection currents surrounding the techniques are capable of providing accurate description of ana-
body, but is limited to the study of heat transfer around insulated tomical features and help to diagnose the ocular diseases better.
clothing [1]. Except Schlerian photography, most of the methods Infrared thermography is renowned for its ability to detect the
developed in past had limitations of either requiring contact with pathological and physiological changes in the eye which are ob-
subject or incapable of displaying distribution of temperature. scured or unreachable under anatomical examination. A typical
Infrared (IR) thermography is a non-contact and non-intrusive ocular thermogram of normal eye is shown in Fig. 1. It has been
temperature measuring technique, with an advantage of no alter- used to study the inflammation of human lacrimal drainage system
ation in the surface temperature and capable of displaying real- [19], dry eye [20], carotid artery stenosis [21], glaucoma [22], uni-
time surface temperature distribution. It was first introduced by lateral exophthalmos [23], Tolosa–Hunt syndrome [24], and oph-
Lawson in 1956 to modern medicine and discovered the associa- thalmic post-herpetic neuralgia [25]. It was also used to diagnose
tion of elevated skin temperature with breast carcinoma [2] and la- retinoblastoma in children [26] and vascular neuritis [27] of the
ter investigated the feasibility and potential of using IR optic nerve. The OST can be used in the diagnosis of different ocular
thermography as a tool to study breast lesions [3]. This technology diseases.
has revolutionized the field of temperature measurement in the The invasive methods of measuring eye temperature require di-
last 50 years and is widely employed nowadays. rect contact with human cornea. Among the invasive measuring
In the field of thermo-fluid dynamics IR thermography was ap- techniques, needle probe was mainly employed. During measure-
plied to measure convective heat fluxes, and for the comprehen- ment, needle probe acts as a cooling fin when inserted into the
sion of fluid dynamics phenomena on the flow field behaviour eye [28] and error inevitably exists if the penetration depth is be-
over complicated body shapes [4]. The process of ice nucleation low 40 mm. Topical anesthesia is often required, and this instilled
and ice propagation in flowers of fruit trees and other frost sensi- solution often lowers OST. In addition, the penetration of needle
tive plants were studied using IR thermography in agriculture can be traumatic, which often induces further blood flow in eye
[5]. It was used for the measurement of size, depth and thermal and thus alters OST. Hence, this invasive method, is not comfort-
resistance of materials and components [4], and also for environ- able to the subjects and the reported discrepancies in temperature
mental monitoring such as sea, river pollution, information about in some cases can be up to 6 "C [28].
indoor climate [6], inspection of plants and assistance in the reduc- Infrared thermometry and thermography can measure OST
tion of maintenance cost of mechanical equipment [4]. In poly- without causing trauma on subjects. However, they are unable to
graph testing, a standard security procedure favoured by US measure the intraocular temperature. These techniques remotely
government, IR thermography was applied to perform facial image measure the emitted IR radiation and acquire temperature data
analysis [7]. of a specific surface. During measurement there will be no alter-
In medical field, IR thermography has been used to assist in ation in the surface temperature and also the data collected is of
decision making in open heart surgery due to its ability to provide higher precision. With this technology researchers are able to
real-time information [8]. It was also used for the management of study OST with greater ease, and accuracy.
neuropathic pain [9,10] and the assessment of patient response to
chiropractic care by measuring the temperature gradient in clinical 2. The principles of infrared thermography on human eye
setting [11]. So far, IR system has been used to diagnose breast can-
cer [12,13], rheumatism [1], skin lesion [14], fever [15], impotence In general, IR thermography refers to the recording of tempera-
[16] and thyroid gland disease [17]. Currently, it has been applied ture, or the distribution of temperature utilizing infrared radiation
to ophthalmology to diagnose eye diseases [18]. emitted from a body surface, forming an image called thermogram.
The ocular anterior anatomy and physiology nowadays can be The 2-D thermogram presents the distribution of temperature dis-
studied using a number of ophthalmic imaging techniques: slit tinctly unlike IR thermometry, which gives a single temperature
lamp biomicroscopy, confocal microscopy, corneal topography, value. IR thermography also differs with IR photography, where
optical coherence tomography, computerized tomography, ultra- in the latter records infrared radiation reflected back from objects
sonic biomicroscopy, and magnetic resonance imaging. These in the presence of some external infrared energy sources. The

Fig. 1. Typical thermogram of normal eye.


J.-H. Tan et al. / Infrared Physics & Technology 52 (2009) 97–108 99

principle of IR thermography on eye involves the physics of IR radi-


ation and measurement of OST. These concepts are discussed in the
following sections.

2.1. The physics of IR radiation

Any object, when its temperature is above absolute zero, emits


electromagnetic radiation, called thermal radiation. In physics, the
amount of electromagnetic waves radiated by an object is quanti-
fied by the term spectral radiance. It describes the amount of elec-
tromagnetic waves emitted from a particular surface on the basis
of per unit solid angle. The idea of black body radiation forms
the basis of the physics of thermal radiation, though behaviours
of many objects deviate from the ideal black body extensively.
When a black body is in equilibrium at temperature T in an
environment, it absorbs and emits a specific continuous spectrum
Fig. 2. IR camera (VarioTHERM# head II, http://www.infratec.de/fileadmin/down-
of wavelengths and intensities. The profile of the emitted spectrum loads/pdf/Flyer_Variotherm_head_en_mail.pdf).
by a black body depends only on the temperature at equilibrium in
an environment and therefore the temperature of a black body is
principle of an IR thermography system. An IR thermography sys-
directly related to the wavelengths of the light that it emits.
tem in general consists of a camera (as shown in Fig. 2), with some
According to Planck’s law, the spectral radiance, I (in watts per ste-
detachable optics, and a personal computer that controls the cam-
radian per square meter) of electromagnetic radiation at all wave-
era. In the camera there is an IR detector, which absorbs the IR
lengths, k, at temperature T from a black body as a function of
shining by the lenses and converts the IR signal into electrical volt-
wavelength is given by
age or current for processing. The way a desired field of view is
2
2phc 1 projected and recorded on IR detector within the camera, is deter-
Iðk; TÞ ¼ W cm$2 lm$1 ð1Þ
k5
e
hc
kkT $1 mined by the imaging system.
Generally, there are two main sorts of imaging system available
where k is the Boltzmann constant (1.381 % 10$23 J/K), h is the to the IR thermography: scanner (scanning array) and focal plane
Planck constant (6.626 % 10$34 Js), c is the speed of light array. The focal plane array is superior to scanning array in almost
(2.998 % 108 m/s) and T is the absolute temperature. Differentiating every aspect in terms of performance. Scanning array consists of
Planck’s law with respect to k to look for the maximum radiation linear arrays, rastered across the desired field of view using one
intensity, the Wien’s displacement law is obtained detector, two lenses, one horizontal and one vertical deflection
2898 lK mirrors, to construct a 2-D image as illustrated in Fig. 3a. Its work-
kmax ¼ ð2Þ ing principle is analogous to a person looking at a view through
T
narrow slit, rastering both his/her head and slit in the direction
which mathematically illustrates a common observation: as the tem-
perpendicular to the slit to build up 2-D image. Focal plane array
perature of an object rises, the color of light emitted varies from infra-
images a desired field of view without scanning, and it works sim-
red to red to orange. Given human body temperature of 37 "C, the
ilar to a typical camera, in which the film captures the 2-D image
emitted light peaks at 9.35 lm which falls into the infrared region.
directly projected by the lens at image plane (Fig. 3b).
Consider a small flat body radiating outward into a half-sphere
IR detectors of IR thermography consist of an array of elements
under an temperature-equilibrium environment, one can deduce
reactive to IR. In the broader field of IR thermology, there are two
Stefan–Boltzmann’s law from Eq. (1) and get
types of IR detectors: thermal and photonic. Thermal detectors
j ¼ rT 4 ð3Þ such as microbolometer, pyroelectric detectors and Golay cells de-
which states that for a black body the total energy radiated per unit tect heat generated by incidence of IR radiation and are not wave-
surface area in a unit time is directly proportional to the fourth length dependent. They are commonly used in bolometers such as
power of the absolute temperature of the black body. thermocouple and thermopile. They show slow response time and
In reality, real objects are rarely a black body, though in certain low detection capability without the need for cooling. The photonic
spectral intervals they may behave closed to the ideal black body. detectors like quantum well infrared photodetector (QWIP) and
In most cases real object emits only part of the radiation a black mercury–cadmium–telluride, are wavelength dependent, and can
body emits at the same temperature and same wavelength. Denote detect the temperature of a remote object’s surface by measuring
the amount of radiation emitted by real object as Ek and Ebk for the the amount of IR incidence of a specific range. Unlike thermal
amount of radiation emitted by black body, the emissivity is de- detectors, they offer quicker response time and better detection
fined as performance, and therefore are favourite as IR detector for focal
plane array imaging system.
Ek
e¼ ð4Þ Most of the photonic detectors operate only at cryogenic tem-
Ebk perature, otherwise the detected signal would be swamped by
and Eq. (3) can be rewritten for gray body with the knowledge of thermal noise. The IR thermography utilizing imaging system of fo-
emissivity as cal plane array can be of cooled or uncooled type. Most of these
detectors are cooled evaporatively, either by liquid helium, liquid
j ¼ erT 4 ð5Þ
nitrogen, thermal-electric cooler, or Stirling cycle refrigerator.
Though not common, in the latest technologies microbolometer
2.2. The IR thermography system is used in focal plane array, which have lower cost and require
no cooling.
Planck’s law, Wien’s displacement law, Stefan–Boltzmann’s law Presently in the medical field IR thermography captures ther-
and the concept of emissivity is the physics behind the working mogram at wavelength range of 8–12 lm, which falls in the region
100 J.-H. Tan et al. / Infrared Physics & Technology 52 (2009) 97–108

Fig. 3. (a) Scanning array: 1. detector 2. lens 3. horizontal deflection mirror 4. vertical deflection mirror 5. lens 6. object 7. measuring spot. (b) Focal plane array: 1. object with
measuring spots 2. lens 3. detectors (http://www.infratec.de).

of long-wave IR (LWIR). They are further subdivided into a few viewing on any part of the cornea and sclera during measurement
bands [29,30]: near IR (NIR), from 0.7 to 1 lm, short-wave IR with respect to the thermography is within p=4, so that the error
(SWIR), from 1 to 3 lm, mid-wave IR (MWIR), from 3 to 5 lm, induced due to the variation of emissivity at different angle of
long-wave IR (LWIR), from 7 to 14 lm, very long-wave IR (VLWIR), viewing is negligible. Furthermore, as the angle of view becomes
from 12 to 30 lm. NIR and SWIR are used in fiber optics telecom- greater, the amount of reflected thermal radiation from anatomical
munications and long-distance communications, respectively; surface increases and the error in temperature measured grows.
guided missiles technology making use of IR heat is used in MWIR. Such resultant error is found to be negligible for the case where an-
The LWIR is the region of ‘‘thermal imaging” which does not re- gle of view falls within p=4. Therefore, the OST measured can be
quire external thermal source to obtain a passive isotherm of the comfortably taken as temperature of tear film even after consider-
outside world. The IR thermography used for measurement of ations of error incurred by angle of view and reflected thermal
OST obtains thermogram in LWIR. radiation.

2.3. The measured OST


3. Methodologies in the study of OST
Previous studies [31,32] have illustrated that the absorption
bands of ocular tissues are similar to water, which was opaque to IR thermography captures temperatures and its variation over
far infrared. Water is found to have an emissivity of 1, and it be- ocular surface. Generally, it is either displayed in gray-scale or
haves like black body radiator on IR spectral above 3 lm. The emit- RGB color palette. RGB thermogram provides a better visual repre-
ted spectrum falls in between 1 lm and 30 lm with a maximum sentation of temperatures for direct inspection, which relies on
distance at 9 lm for ocular tissues, under normal circumstances users experience and judgment during the study of OST. Gray-scale
(32 "C). Consequently, the spectrum radiated by any ocular tissues IR thermogram is useful in the quantitative analysis of the OST.
will be fully absorbed by ocular tissues anterior to it. In other OST is the temperature of a specific location acquired on corneal
words, the spectrum emitted by vitreous will be absorbed by lens surface or some defined region [19,23,40–43]. Usually, it is studied
that is anterior to it. by comparisons on several regions in normal and diseased eyes.
The transmission in between cornea and tear film is slightly dif-
ferent. Though water behaves similar to black body on IR spectrum 3.1. Manual measures in the acquisition of OST
above 3 lm, for thin water layer the degree of absorption depends
on the thickness of layer. The transmittance of IR on spectrum Efron et al. have estimated the geometric centre of cornea, and
8 lm to 13 lm is about 30% when the layer thickness is 10 lm. measured surface temperature at every 0.5 mm increments on
When the layer thickness was 40 lm or above, the transmittance either side of corneal surface horizontally [44]. In total there are
approximates to zero [33,34]; and tear film thickness was found 11 points running across the anterior surface, as illustrated in
to be around 40 lm [35]. Consequently, the spectrum radiated by Fig. 4. The method enables the study of temperature profile of
cornea was also wholly absorbed by tear film. The temperature OST, using a regression polynomial
of tear film was measured by thermography during the evaluation
DT ¼ ax2 þ bx þ c ð6Þ
of OST [34,36,37]. This study was performed using finite element
2
model of heat transfer in human eye [38]. where a = 0.01 "C/mm , b = 0.003 "C/mm, c = 0.01 "C, and x is the
A recent study [39] verified the above idea and showed that, distance from the geometric centre of cornea in mm.
OST was mainly related to tear film stability, rather than other Morgan et al. have [21,36,45] approached the acquisition of OST
parameters such as central corneal thickness, corneal curvature by different method. Five 10 % 10 boxes were placed in five differ-
or depth of anterior chamber. ent anatomical locations along horizontal meridian running across
In clinical thermography, studies have indicated that an angle of the estimated centre of cornea as shown in Fig. 5. One was placed
viewing of curved anatomical surface beyond 90" will lead to a at the estimated centre of cornea, two at the limbal position, and
reduction in measured temperature of 4 "C or more, due to the var- another two at the nasal and temporal conjunctiva (2 mm from
iation of emissivity at different angle of viewing. For the measure- the limbus). The area of each box was approximately 1 mm2. Mean
ment of OST by IR thermography, it is assumed that, the angle of OST was used for statistical analysis. In one of the studies [45],
J.-H. Tan et al. / Infrared Physics & Technology 52 (2009) 97–108 101

Fig. 4. Methodology by Efron et al. in the acquisition of OST [44]. Fig. 7. Methodology by Sodi et al. in the acquisition of OST [48].

Fig. 5. Methodology by Morgan et al. in the acquisition of OST [21,36,45].

radial temperature difference (RTD) was proposed to represent the Fig. 8. Methodology by Purslow et al. in the acquisition of OST [54].
variation in temperature across the cornea. It is the temperature
difference, in between the average temperature value at the centre
of cornea, and the mean temperature value of the sites of two lim- for subsequent analysis: central, superior, inferior, nasal, and tem-
bal positions. poral [54]. Tan et al. [55] have estimated the geometric centre of
There were other similar measures to study OST. Instead of cornea, and OST were measured by a total of 20 points across the
boxes, points were placed at five different anatomical locations anterior eye: geometric centre of cornea, three points inferiorly
along the horizontal meridian. Five points were placed on centre and four points superiorly at 2 mm separation, six points nasally
of cornea, internal and external canthus, half-way from the inter- and temporally at 2 mm separation, respectively [55] (Fig. 9).
nal canthus and nasal limbus, half-way from the temporal limbus Chang et al. [56] have studied Graves’ ophthalmopathy by acquir-
and external canthus (Fig. 6) [46,47]. In another scheme [48] five ing local temperatures of lateral orbit (reference point), upper eye-
points equally placed along a horizontal line running through cen- lid, caruncle, medial conjunctiva, lateral conjunctiva, lower eyelid,
tre of cornea, connecting medial and lateral canthi and those points and cornea, as shown in Fig. 10.
were placed (Fig. 7).
In some studies either a squared 10 % 10 pixels box [42,49], or a 3.2. Semi-automated method in the acquisition of OST
squared 20 % 20 pixels box [50] with an actual area of 3.3 mm2, or
an area of 4 mm2 [20] or an encircled region of 4.4 mm diameter A semi-auto method was developed to acquire OST on thermo-
[51], or a small circle [52] at the centre of cornea were used to gram using standard procedure [57], as shown in Fig. 11. Thermo-
study OST. gram of eye (in OEM data format) was converted to gray-scale jpeg
A circular region was defined to estimate the centre of cornea file. Then the image was manually cropped to consist only of eye,
and the radius of the circle was either a fixed value [39] or was ac- and resized to a standard size of 400 % 200 pixels. An algorithm
quired by subjective judgment [25,53]. Purslow et al. have re- to detect the circular cornea was proposed, with corneal radius
corded temperature data from 23 points across the anterior eye, being one-fourth of the length of the entire eye. Temperature pro-
as illustrated in Fig. 8, and this data was grouped into five regions file in the middle of the cornea was plotted, as illustrated in Fig. 12.

Fig. 6. Methodology by Galassi et al. in the acquisition of OST [46,47]. Fig. 9. Methodology by Tan et al. in the acquisition of OST [55].
102 J.-H. Tan et al. / Infrared Physics & Technology 52 (2009) 97–108

Fig. 12. Pixel profile across the cornea (corresponds to temperature profile of
cornea).

Fig. 10. Methodology by Chang et al. in the acquisition of OST [56].

[64], these deformable template matching techniques are local


The variation in temperature over the cornea was used to study optimizers and require good initial position or clean image to avoid
temperature deviation along cornea (TDC) incorrect localization [65]. In other words, for some applications
X user is required to select location for placement of initial contour,
TDC ¼ f ðx; yÞ $ f ðx; y þ 1Þ ð7Þ also the size or the shape of contour, in order to get the correct
y
localization. There were cases where user has to monitor the
expansion or convergence of snake.
3.3. Automated method in the acquisition of OST The eye in ocular thermogram is not fixed at the centre of the
image, and may appear anywhere in the image. For some eyes
An automated method to acquire OST was developed [58,59], as which have smaller size, they occupy just some part of the image
illustrated in Fig. 13. In that method, localization of eye and cornea instead of most of the spaces in thermogram. Given these circum-
was achieved by snake algorithm and target tracing function [58] stances, local optimizers are not a viable option if any user initial-
without any manual intervention. Genetic algorithm was utilized ization is to be avoided.
to perform search for minimum on target tracing function. In the automated method, the problem is overcome by using a
Snake [60] is an active contour consisting of a series of points target tracing function. Target tracing function evaluates few con-
(dubbed snake points), moving under forces of gradient vector flow verged snakes, and from them only a few, or probably just one of
(GVF) [61] to lock onto nearby edges, and thereby delineates the them correctly localize the eye. The converged snake which gives
shape of eye if its initial contour is of appropriate shape and placed target tracing function the lowest value, is the snake that accu-
in some suitable locations. GVF, an intense force field pushing rately localizes the eye. In other words, the algorithm performs a
snake contour to reach its minimum in an energy functional (pro- search for minimum on target tracing function to get the right
posed by Kass et al.), is derived from edge map. In the automated snake to correctly localize the eye. Genetic algorithm is utilized
method, the algorithm acquires edge map by applying Gaussian to perform this search process.
blur on the ocular thermogram (gray-scale image). The radius and the centre of cornea are derived from the final
For a snake locking into the desired feature, it requires the resultant snake points. The centre of cornea is the same as centroid
snake to be initially placed closed to the feature of interest, and of the snake, and radius of cornea is acquired on the ground of the
the shape of the snake must be carefully selected. In fact, even in snake points with the use of a formula derived [66]. Fig. 13 shows
some more sophisticated methods such as spline-based shape result of the automatic detection of eye and cornea using genetic
matching [62], diffusion snakes [63], and active shape models snake algorithm.

Fig. 11. Methodology by Acharya et al. in the acquisition of OST [57].


J.-H. Tan et al. / Infrared Physics & Technology 52 (2009) 97–108 103

Fig. 13. Methodology by Tan et al. in the acquisition of OST [58,59].

4. Application of IR thermography to ocular studies and diseased eye. Some characteristic patterns of asymmetry in
orbito-ocular thermogram was proposed for normal and patholog-
In ophthalmology, IR thermography has been applied to study ical eyes [74].
ocular physiologies, ocular diseases and surgery. Several correla- Efron et al. have used a wide-field color-coded infrared imaging
tions between OST and a number of physiological and pathological device, to observe the variation in temperature across the ocular
changes in eye have been studied with greater ease through IR surface and the temporal stability of the central cornea tempera-
thermal imaging. The OST is not only affected by physiological ture [44]. They have observed the presence of ellipsoidal iso-
and pathological changes in eye, but also some of the external therms, with a major horizontal axis for most cases, concentric
factors. on a temperature apex that was slightly inferior to the geometric
Room temperature was shown to influence OST [39,67]; a rise centre of the cornea in most of the OST [44]. Limbus was found
in 1 "C room temperature may lead to an increase of 0.15 "C to to be 0.45 "C warmer than geometric centre of cornea, and the rate
0.2 "C in OST [68,69]. The OST was reported to decrease with an in- of corneal cooling after a blink was positively correlated to the
crease in the air flow [70], therefore uniform room temperature amount of time an eye can remain open.
and humidity are required to minimize the discrepancies in studies Another study similar to Efron et al. was conducted to show that
of OST. limbus was 0.23–0.43 "C warmer than the geometric centre of cor-
It was reported that, OST increases when an eye is infected with nea [55]. All subjects recruited were Chinese, illustrating ‘bowl-
anterior uveitis [36,71], and indicates a negative correlation for car- shaped’ horizontal OST profile and ‘spoon-shaped’ vertical OST
otid artery stenosis [19,67,72]. The degree of hyperemia in bulbar profile [55]. It was shown that Chinese eyes have lower tear vol-
conjunctiva (examined in terms of grade of redness using ume and tear stability compared to Caucasian eyes [75–78]. These
McMonnies scale) were positively correlated to OST [73]. It is in findings suggested brown eyes in general have lower temperature
agreement among most of the investigators that, a large inter-ocular at limbus compared to blue eyes [55].
temperature difference indicates the presence of eye disease. OST was shown to be decreasing by $0.01 "C per year through-
out life, and this rate of reduction increases after middle age [79]. It
4.1. Studies of ocular physiologies was revealed that, 95% normal subjects showed an inter-ocular
temperature difference of less than 0.62 "C and more than 0.62 "C
Mapstone described thermographic patterns in normal, ische- for disease subjects [36].
mic and hyperemic eyes using Bofor IR camera system [40]. Human lacrimal drainage system and the effect of non-contact
Wachtmeister also conducted a similar study, investigating both corneal esthesiometer (NCCA) air stimulus were also investigated
normal and diseased eyes [41]. He found out that, the affected [42]. The anatomy and patency of human lacrimal drainage system
eye was warmer than the normal eye for anterior and posterior were usually examined using radionucleotide lacrimal scanning
diseases. and dacryocystography. But, the subjects were exposed to radia-
Since then a number of researchers looked into the application tions due to these techniques. Raflo et al. have employed IR ther-
of IR thermography as a diagnostic tool, to detect eye diseases mography together with lacrimal irrigation, for the purpose of
using thermal asymmetry [74]. They have compared thermogram visualizing tear ducts in control and patients with obstructive
of both sides at the orbito-ocular region and concluded that, ocular epiphora [80].
disease will be present, if a difference of 0.5 "C exists in between Corneal innervation has a number of specialized nerve types:
the left and right eye. However, this method was shown to be mechano-sensory, polymodal, mechano-heat and ‘cold’ neurons.
incorrect [74]. In one study involving 96 normal subjects, half of Traditional methods of assessing corneal nerve function, such as
them exhibited asymmetry at orbito-ocular region [74]. Hence using nylon thread of fine metal wire cause trauma to cornea.
symmetry alone, may not be a good tool to differentiate normal NCCA was proposed as a new method to stimulate corneal nerves
104 J.-H. Tan et al. / Infrared Physics & Technology 52 (2009) 97–108

utilizing controlled air pulse, though the mode of stimulation was lagens denature and a complete denaturation was reached at 40 "C
unclear to researchers. [82]. The studies were conducted to investigate whether the tem-
Murphy et al. have conducted a study to investigate the mode of perature for denaturation of corneal collagen on corneal surface
stimulation by NCCA and concluded that, the rate of change in cor- was routinely reached throughout the PRK operation [49]. OST
neal surface temperature was due to the principal mode of stimu- was found to reach the threshold at which the corneal collagen
lation by NCCA [42]. In other words, during stimulation both A @ denatures during treatment and there was no correlation reported
(mechano-sensory) and C (temperature) fibers are likely to re- in between any surgical parameters and temperature changes.
spond to air-pulse stimulus signals from C fibers. It was reported that subjects with dry eye showed greater OST
Dynamic IR imaging was used in the NCCA stimulation study, and radial temperature difference than normal subjects [45]. Dur-
and the corneal surface temperature was measured in a sequence ing cooling of ocular surface, the dry eye subjects showed faster
of 32 images with 0.25 s in between each image of subjects after rate of cooling than normal eye [20]. It was in agreement with
exposure to air-pulse stimulus [42]. In another recent study, dy- other investigators that dry eye patients have higher evaporation
namic IR imaging was utilized to investigate the relationship be- rate [83,84]. The effect of conjunctival hyperemia associated with
tween OST and some physical parameters of anterior eye, such as dry eyes seems to outweigh the cooling effect given by the in-
corneal topography, corneal thickness, bulbar hyperemia, and tear creased evaporation, hence dry eyes were found to have higher
film stability [39]. Initial OST after blinking was found to be corre- OST [45].
lated with body temperature and tear film stability, as assessed by However, in another studies it was shown that the centre part
non-invasive breakup time. of corneal temperature of dry eyes was lower than normal eyes,
though mean OST was not evaluated [53]. Similar studies con-
4.2. Studies of ocular diseases and surgery ducted by research group in Japan [50,85], showed that central cor-
neal temperature in dry eye patients were higher than normal
Exophthalmos is a bulging or protruding eyeball. It is a medical control groups. Such differences were attributed to the different
condition that is often seen in severe thyroid eye disease (Graves’ population of dry eye patients recruited in studies [53].
ophthalmopathy), an auto-immune inflammatory disorder which Besides, Craig et al. have also indicated that, the temperature
affects the orbit of the eye, in patients with or without thyroid dis- variation factor and mean osmolality were higher in dry eye groups
order. A number of different cases such as left endocrinal exoph- than control subjects [53]. Temperature variation factor was found
thalmos, metastasis of left orbit were subjectively assessed with to be inversely correlated to central corneal temperature. Hence,
IR thermography [23]. The number of subjects in each case was the ocular surface having higher variation in temperature tends
too small to perform the statistical analysis [23]. to have lower central corneal temperature.
In a later study, IR thermal imaging was utilized to determine Corneal surface temperature decreases exponentially after eye
the inflammatory state and follow-up effect of methylprednisolone opens, and it asymptotically approaches a constant value after
pulse therapy in patients with Graves’ ophthalmopathy [56]. Local some time [50]. This can be modeled by the following formula:
temperatures such as lateral orbit (which is defined as a reference
TðtÞ ¼ ðT 0 $ T 1 Þe$kt þ T 1 ð8Þ
point), upper eyelid, caruncle, medial conjunctiva, lateral conjunc-
tiva, lower eyelid and cornea were measured in patients with In which T is the corneal surface temperature after the eye opens for
Graves’ ophthalmopathy and normal control subjects. Clinical a period of t. T 0 is the temperature immediately the eye opens; T 1 is
activity score and local temperatures in some 11 patients were also the corneal surface temperature at equilibrium; k is the tempera-
measured before and after methylprednisolone pulse therapy. The ture coefficient. By measuring temperature values within a box of
study showed that, temperature difference between lateral orbit 20 % 20 pixels, or 3.3 mm2, placed at the centre of cornea over a
and other target area such as caruncle, medial conjunctiva, lateral period of 30 s, it was found that for normal blinking the mean k va-
conjunctiva and lower eyelid of Graves’ ophthalmopathy patients lue in dry eye patients (5.6 ± 2.9 per s) was significantly less than in
were significantly higher than those of normal subjects [56]. Posi- normal control subjects (9.3 ± 5.0 per s) [50]. It was suggested that k
tive correlation (correlation coefficient = 0.8, n = 22) was found in value might reflect tear film stability [34].
between the difference in the sum of the temperatures before A later studies employed principle similar to the one mentioned
and after treatment, and the change in clinical activity score. The above to diagnose dry eye using IR thermography. Twenty six nor-
IR thermal imaging was thought to be helpful in evaluating the fol- mal and 82 dry eye patients were recruited in that studies. An
low-up effect of methylprednisolone pulse therapy [56]. encircled region having diameter of 4.4 mm (22 pixels) was posi-
In some other studies, the effect of corneal temperature on the tioned at the centre of eye to acquire OST [51]. The study showed
insudation of lipoprotein and its mobility within corneal tissue that, the diagnosis of dry eye patients was accurate with a sensitiv-
were studied with IR thermography [43]. They found that, the local ity of 79%, specificity of 75% with 0.841 area under ROC (receiver
capillary permeability was increased when the regional differences operating characteristics) curve 0.42 "C/s [51].
in temperature was greater, bringing lipoprotein insudation prefer- Furthermore, dry eye was studied in conjunction with the treat-
entially into this warmer corneal region and making it less mobile ment of acupuncture and IR thermography to determine the effect
and relatively inert [43]. of acupuncture [86]. It was reported that acupuncture affects the
Studies on ocular diseases and surgery were conducted, with temperature of the pre-corneal tear film, and lowers the tempera-
the aid of IR thermography on post-herpetic neuralgia (PHN), ture at the middle of cornea.
photorefractive keratectomy (PRK) and dry eye. In patients with Some studies have suggested a correlation in between OST and
Herpes Zoster Ophthalmicus (HZO), it was shown that the affected ocular blood flow. In monkeys, intraocular pressure (IOP) was
eye was warmer than the fellow eye, giving a large inter-ocular dif- found inversely related to ocular perfusion pressure and ocular
ference in temperature [81]. However, for patients with estab- temperature [87]. Corneal temperature was determined to be pos-
lished PHN the affected side was colder than the other side [81]. itively correlated to the ipsilateral values of end diastolic velocity
Another study indicated that for the affected side in PHN the entire (in left and right eyes, respectively), and it also positively correlates
ocular surface was cool with a loss of isotherms [25] and reduction to the resistivity index of left and right eyes, respectively; inter-
in corneal sensitivity of more than 50% [23]. ocular difference in corneal temperature was positively correlated
A rise in the temperature was reported during the PRK treat- with the difference in end diastolic velocity, and was found to be
ment [49]. Studies have shown that, at 38.7 "C 50% of corneal col- negatively correlated with resistivity index [88]. The above
J.-H. Tan et al. / Infrared Physics & Technology 52 (2009) 97–108 105

study was done with IR thermometer instead of IR thermographer OST (OST of right eye minus that of left eye). They have concluded
[88]. that the circle of Willis and the other anastomoses within the brain
Galassi et al. have evaluated the OST of patients with primary are not able to compensate for the reduction in blood flow and
open-angle glaucoma (POAG) and control group through IR ther- hence led to a cooler eye on the affected side [21].
mography [46]. They have investigated the correlation between
OST, IOP and retrobulbar hemodynamics in conjunction with the 5. Discussion
use of color Doppler imaging (CDI). Temperatures of five anatomi-
cal points (internal and external canthus, half-way from the inter- IR thermography is an efficient tool not only to capture temper-
nal canthus and nasal limbus, centre of the cornea, half-way from atures of corneal surface, but also to detect and visualize any subtle
the temporal limbus and external canthus) of POAG patients were changes on the OST. Eye is a delicate organ and highly susceptible
found to be lower than of healthy control group. These tempera- to external variations. The non-intrusiveness of this technique has
tures were significantly correlated to resistivity index [46]. Their encouraged many researchers to use this method to study ocular
results highlight the influence of retrobulbar hemodynamics on physiology, instead of other imaging techniques and methods for
OST [46]. In other study, higher OST was observed for central reti- measuring the OST.
nal vein occlusion (CRVO) compared to normal subjects and lower Zeiss pioneered the use of thermometry in the measurement of
OST was shown for ischemic CRVO eyes compared to non-ischemic ocular temperature [93]. However, Mapstone [32] had first intro-
ones [48]. duced IR thermal imaging into the field of ophthalmology. Since
The effects of two glaucoma surgeries, namely deep sclerec- its inception, several researchers have looked into thermal asym-
tomy and trabeculectomy, on bulbar hemodynamics and corneal metry or anomalies in thermal distribution, both on the eyes and
surface temperature were investigated [47]. Corneal surface tem- region surrounding the eyes. They have tried to establish diagnos-
perature was acquired [46]. Three months after operation, the cor- tic criteria from these asymmetry and anomalies. Due to the poor
neal surface temperature increased in both type of surgeries, and a resolution of IR thermal imaging at that time, the eye on thermo-
negative correlation between postoperative changes in ophthalmic gram was not clearly discernible. Hence, it was not possible to pro-
artery resistivity index and corneal surface temperature was ob- pose diagnostic criteria based on these asymmetry and anomalies.
served in both types of surgeries [47]. They have suggested that Due to the rapid advancement in the IR thermography, more de-
IR thermography was useful in the evaluation of vascular outcome tailed thermal distribution of the anterior eye can be obtained.
of glaucoma surgery [47]. The better technology has prompted researchers to focus their
The cataract patients with an anterior capsulotomy, had corneal studies on the thermographic pattern of the anterior eye, instead
temperature significantly higher not only just after the surgery, but of the entire orbito-ocular region. The horizontal temperature pro-
also at the 30th-day after the operation [89]. The length of surgery file passing through the geometric centre of eye was illustrated,
lasted for more than 40 min, and IR thermometer was used in the and more recently, the vertical temperature profile is also
study. No follow-up investigation with IR thermographer was done described.
to elucidate further detail. Finite element analysis on human eye has shown that, the cen-
IR thermography was employed to analyze and compare differ- tre of the corneal surface has the lowest temperature [94]. How-
ent cataract surgery procedures based on the phacoemulsification ever, experimental studies showed that the coolest point on the
system. In one of the studies, phacoemulsification probes from cornea is slightly inferior to the geometric centre of cornea
Alcon Legacy AdvanTec, Bausch & Lomb Millennium and AMO sov- [36,44,55], and this observation was evident in most of the normal
ereign WhiteStar were placed in air, and testing condition was set subjects [44]. Researchers suggested that the presence of upper
in a way such that a corneal burn might be produced during cata- eyelid, which is a source of heat, shifted the temperature apex
ract surgery [90]. It was found that, Millennium and Sovereign (the coolest point) inferiorly instead of being at the centre of
WhiteStar generated more heat and therefore led to higher tem- cornea.
perature than Legacy AdvanTec phacoemulsification system [90]. Besides, the shape of isotherms predicted by finite element
Three different cataract surgeries performed in vivo with phac- analysis was not the same as the experimental studies. The pres-
oemulsification (phacoemulsification with the traditional Sover- ence of eyelid has made the isotherms as an elliptical shape, in-
eign system without WhiteStar technology, phacoemulsification stead of a circular shape [94].
with the Sovereign WhiteStar system, and phacoemulsification So far thermographic pattern of normal subjects’ OST have been
through micro incision cataract surgery with the bimanual Sover- reported. But for other ocular diseases the corresponding thermo-
eign WhiteStar system) were compared using IR thermography graphic pattern has not been reported. Subjective assessment was
[91]. Among these three techniques phacoemulsification through the only method employed on studies of OST of diseased eye con-
micro incision cataract surgery with the bimanual Sovereign ducted during 1970s and 1980s; for later studies, mean tempera-
WhiteStar system was determined to have the lowest thermal im- tures on one or several sites (in this context, a site refers to a
pact on eye [91]. In another study, intraoperative thermal levels at point or an area) over the anterior eye were utilized to study ocular
wound site were studied, with Sovereign WhiteStar system and diseases.
Legacy Advantec and NeoSoniX system. Legacy AdvanTec and From the viewpoint of image processing, when one acquires
NeoSoniX system produced grater mean temperature change at mean temperature, standard deviation or median over a region
wound site and higher mean peak temperatures in patients [92]. or an area on ocular thermogram, he/she is in fact acquiring some
These studies showed that IR thermography has facilitated effec- first-order spatial statistics of texture analysis on the thermal im-
tive analysis on cataract surgery with phacoemulsification. age. As an example, to obtain the mean temperature in the above
case, the average pixel intensity over the region is calculated and
4.3. Studies of non-ocular diseases converted to the corresponding temperature. The average pixel
intensity is the first-order spatial statistics.
Carotid artery stenosis (CAS) was investigated, using color- Often researchers employed mean temperature as a measure to
coded infrared ocular thermography [21]. They have shown that, get a ‘‘middle” or ‘‘expected” temperature value of the site of inter-
the OST was negatively correlated with the degree of CAS and est, and made comparison with the corresponding value obtained
the relative difference in CAS (CAS of right eye minus that of left from normal subjects. Texture analysis was not actually considered
eye) was also negatively correlated with the relative difference in in their studies. But in coming future such technique may prove to
106
Table 1
Summary of methods used to estimate OST.

Studies Types of Methods Main findings


OST
acquisition
Efron et al. Manual 11 Points running across the anterior surface Average difference in temperature between the GCC and the limbus is 0.45 "C
[44]
Morgan et al. Manual Five 10 % 10 boxes placed in five different anatomical locations along horizontal In [21], a significant negative correlation was found between OST and the degree of CAS [21]
[21,36,45] meridian running across the estimated centre of cornea In [36], 95% normal subjects were found to have inter-ocular temperature differences between $0.57 and
0.62 "C [36]
In [45], dry eye patients were found to have greater mean OST than normal subjects, and RTD was also
found to be greater in dry eye group [45]
Galassi et al. Manual Five points placed on centre of cornea, internal and external canthus, half-way from the For all five points temperature of patients of POAG are lower than temperature of healthy control group
[46,47] internal canthus and nasal limbus, half-way from the temporal limbus and external [46]

J.-H. Tan et al. / Infrared Physics & Technology 52 (2009) 97–108


canthus A negative correlation between postoperative changes in ophthalmic artery resistivity index and corneal
surface temperature was observed 3 months after deep sclerectomy and trabeculectomy surgeries.
Sodi et al. Manual Five points equally placed along a horizontal line running through centre of cornea, In central retinal vein occlusion (CRVO) eyes and in fellow, non-affected eyes, OST values were lower than
[48] connecting medial and lateral canthi in controls [48]
Murphy et al. Manual A squared 10 % 10 pixels box placed at the centre of cornea In [42], using the NCCA air pulse, the principal mode of corneal nerve stimulation was the rate of
[42], temperature change of the corneal surface [42]
Betney et al. In [49], no correlation was demonstrated between any surgical parameters and changes either in absolute
[49] or relative temperature [49]
Mori et al. Manual A squared 20 % 20 pixels box placed at the centre of cornea The rate of decline in corneal temperature for normal blinking in patients with dry eye was significantly
[50] less than that in normal subjects [50]
Chiang et al. Manual An encircled region of 4.4 mm diameter (22 pixels) Diagnostic of dry eye using infrared thermal imager system has reached a sensitivity of 79.3%, a specificity
[51] of 75 [51]
Ng et al. [52] Manual A small circle placed at the centre of cornea OST was found to decrease with age at a rate of 0.0383 "C per year [52]
Cardona et al. Manual A circular region centred at the corneal centre In [25], patients with HZO in acute stage were found to show high inter-ocular temperature differences,
[25] with the affected side being warmer than the other side [25]
Purslow and Manual In [39], strong positive correlations were found in between initial OST and both body (forehead)
Wolffsohn temperature and tear film stability [39]
[39]
Craig et al. Manual In [53], dry eye patients were found to have lower central cornea temperature and higher mean
[53] osmolality [53]
Purslow et al. Manual 23 Points placed across the anterior eye OST is greater eye with hydrogel and greater still with silicone hydrogel contact lenses in situ, regardless
[54] of modality of wear [54]
Tan et al. [55] Manual 20 Points placed across the anterior eye, lined up in the shape of ‘‘+” Vertical OST profile was noted to be a ‘spoon-shaped’; for Chinese the average difference in temperature
between the GCC and the limbus is 0.23–0.43 "C [55]
Chang et al. Manual Acquire local temperatures of lateral orbit, upper eyelid, caruncle, medial conjunctiva, Temperature difference between lateral orbit and three other target areas (caruncle, medial conjunctiva,
[56] lateral conjunctiva, lower eyelid and cornea lateral conjunctiva and lower eyelid) of GO patients were found significantly higher than those of normal
subjects [56]
Acharya et al. Semi-auto Image was manually cropped to consist only of eye, the cornea was then detected by Average OST and temperature deviation across the cornea decreases with age [57]
[57] algorithm developed
Tan et al. [58] Automated The eye was localized by genetic snake algorithm, and the corneal diameter and location The algorithm was able to localize 90% of IR ocular thermogram [58]
were derived from the resultant snake points
J.-H. Tan et al. / Infrared Physics & Technology 52 (2009) 97–108 107

be an alternative approach to effectively study thermographic pat- References


tern of normal and diseased eye, other than descriptively or quan-
titatively accounting the pattern of isotherms. [1] E.F.J. Ring, Progress in the measurement of human body temperature, IEEE
Engineering in Medicine Biology Magazine 17 (1998) 19–24.
The study of OST on IR thermogram needs more analysis tools [2] R.N. Lawson, Implications of surface temperature in the diagnosis of breast
and efforts to deliver better diagnostic value. It is known that eye cancer, Canadian Medical Association Journal 75 (1956) 309–310.
thermogram lacks corneal boundary, which is clearly visible in [3] R.N. Lawson, Thermography: a new tool in the investigation of breast lesions,
Canadian Services Medical Journal 13 (1957) 517.
optical image. This makes the quantification of corneal tempera- [4] C. Meola, G.M. Carlomagno, Recent advances in the use of infrared
ture (or OST) a difficult task to achieve due to the difficulty in cor- thermography, Measurement Science and Technology 15 (2004) R27–R58.
neal localization, and this problem is approached with numerous [5] M. Wisniewski, S. Lindow, E. Ashworth, Observations of ice nucleation and
propagation in plants using infrared video thermography, Plant Physiology 113
methods as introduced in Section 3 and summarized in Table 1.
(1997) 327–334.
For manual acquisition of OST, the centre of cornea is located by [6] M. Cehlin, B. Moshfegh, M. Sandberg, Visualization and measuring of air
subjective judgment, by placing few boxes or points at other loca- temperatures based on infrared thermography, in: Proceedings of the 7th
tions with respect to the centre [20,42,49,50,52]. In some other International Conference on Air Distribution in Rooms ROOMVENT, Reading,
UK, 2000.
studies, a circular region was defined at the centre of cornea. These [7] I. Pavlidis, J. Levine, Thermal image analysis for polygraph testing, IEEE
methods do not indicate the actual corneal temperature and accu- Engineering in Medicine and Biology Magazine 21 (2002) 56–64.
racy depends on the number and position of the boxes. [8] W. Ruddock. Advanced infrared resources. <http://www.infraredthermography.
com>, 2008 (accessed 5.05.09).
Semi-automated method is an improved method. It provides a [9] H. Hooshmand, M. Hashmi, E.M. Phillips, Infrared thermal imaging as a tool in pain
set of rules and procedures to define and locate cornea, though management an 11 year study: I, Thermology International 11 (2001) 53–65.
in some cases there is some slight error in accuracy with regard [10] H. Hooshmand, M. Hashmi, E.M. Phillips, Infrared thermal imaging as a tool in
pain management an 11 year study: II Clinical applications, Thermology
to the location of corneal region. However, human intervention is International 11 (2001) 119–129.
required before the start of the proposed algorithm to crop the [11] G. Stillwagon, L.K. Stillwagon, Vertebral subluxation correction and its affect
eye exactly in the proper location. on thermographic readings: a description of the advent of the visi-therm as
applied to chiropractic patient assessment, Journal of Vertebral Subluxation
The snake algorithm coupled with target tracing function [58] Research 2 (1998) 1–4.
localizes the cornea and eye automatically. This recent method [12] E.Y.K. Ng, Y. Chen, Segmentation of breast thermogram: improved boundary
can be used to evaluate the OST and hence the eye diseases. And detection with modified snake algorithm, Journal of Mechanics in Medicine
and Biology 6 (2006) 123–136.
the acquired temperature profile (similar to Fig. 12) can be used
[13] E.Y.K. Ng, A review of thermography as promising non-invasive detection
as tool to diagnose the ocular diseases. modality for breast tumor, International Journal of Thermal Sciences 48 (2008)
Eye is a delicate organ, and ocular IR thermography brings no 849–859.
harm to subjects’ eye under study or diagnosis. But it is unable [14] T.K. Kalili, B.M. Gratt, Electronic thermography for the assessment of acute
temporomandibular joint pain, Compendium of Continuing Education in
to provide thermal information and description of anatomical fea- Dentistry 10 (1996) 979–983.
tures beneath the anterior eye. Unlike ultrasound imaging, IR ther- [15] E.Y.K. Ng, U.R. Acharya, A review of remote-sensing infrared thermography for
mography cannot provide description (or image) on the posterior indoor mass blind fever screening in containing an epidemic, IEEE Engineering
in Medicine and Biology Magazine 28 (2008) 76–83.
segment of eye, nor can it accurately determine the size and loca- [16] A. Merla, G.L. Romani, A. Tangherlini, S.D. Romualdo, M. Proietti, E. Rosato, A.
tion of uveal melanoma which computed tomography is capable of. Aversa, F. Salsano, Penile cutaneous temperature in systemic sclerosis: a
However, IR thermal imaging does not emit ionizing radiation as thermal imaging study, International Journal of Immunopathology and
Pharmacology 20 (2007) 139–144.
computed tomography does, or generate high-frequency acoustic [17] A. Helmy, M. Holdmann, M. Rizkalla, Application of thermography for non-
wave, expose part of body to powerful magnetic field and infrared invasive diagnosis of thyroid gland disease, IEEE Transactions on Biomedical
light. It passively receives IR radiation emitted by subjects, and Engineering 55 (2008) 1168–1175.
[18] P.B. Stefanie. Thermotopography shows ’enormous promise’ for diagnosis and
does not cause any discomfort either through contact with instru- treatment of eye diseases. <http://www.escrs.org/eurotimes/March2003/
ment or expose the eye to any electromagnetic and acoustic wave. thermo.asp>, 2009 (accessed 5.05.09).
IR thermal imaging is able to detect pathological changes in [19] T. Rosenstock, P. Chart, J.J. Hurwitz, Inflammation of the lacrimal drainage
system-assessment by thermography, Ophthalmic Surgery 14 (1983) 229–
eyes which are obscured under examination by other imaging
237.
techniques and methods. It was reported that, for mild inflamma- [20] P.B. Morgan, A.B. Tullo, N. Efron, Ocular surface cooling in dry eye: a pilot
tion it may not be observed under examination of slit lamp biomi- study, Journal of British Contact Lens Association 19 (1996) 7–10.
croscope since the changes were too subtle but sufficient to cause a [21] P.B. Morgan, J.V. Smyth, A.B. Tullo, N. Efron, Ocular temperature in carotid
artery stenosis, Optometry and Vision Science 72 (1999) 850–854.
measurable increase in OST [55]. [22] S. Tokuoka, M. Nakajima, J. Nishikawa, H. Kuroda, I. Azuma, Cold response of
The IR thermal imaging is non-invasiveness and is an excellent skin and eye temperature in glaucoma, Folia Ophthalmology Japan 41 (1990)
diagnostic tool. As an example, the diagnosis of dry eye syndrome 1159–1165.
[23] P. Bourjat, M. Gautherie, Unilateral exophthalmos investigated by infrared
mainly based on measurement of tear film production, such as thermography, Modern Problems in Ophthalmology 14 (1975) 278–285.
Schirmer test, tear breakup time method, and Rose Bengal stain- [24] J. Hannerz, Pain characteristics of painful ophthalmoplegia (the Tolosa–Hunt
ing method. However, Schirmer test requires contact with eye syndrome), Cephalalgia 5 (1985) 103–106.
[25] G. Cardona, P.B. Morgan, N. Efron, A.B. Tullo, Ocular and skin temperature in
and the use of topical anesthesia, tear breakup time method ophthalmic postherpetic neuralgia, Pain Clinic 9 (1996) 145–150.
requires patients to hold their eye for a long time, and often it [26] I.B. Bogdasarov, O.P. Lenskaia, B.M. Belkina, Thermography in the diagnosis of
is not easy to do so, and Rose Bengal staining method is an inva- retinoblastoma in children, Meditsinskaia Radiologiia 30 (1985) 19–21.
[27] A.I. Eremenko, Thermography in the diagnosis of vascular neuritis of the optic
sive method, patients can feel uncomfortable. The diagnosis of dry nerve, Oftalmologicheskii Zhurnal 4 (1990) 235–239.
eye syndrome using IR thermal imaging has none of the above [28] R.F. Rosenbluth, I. Fatt, Temperature measurements in the eye, Experimental
problems. Eye Research 25 (1977) 325–341.
[29] J.L. Miller, Principles of Infrared Technology: A Practical Guide to the State of
IR Thermography can be extensively used as an effective non-
the Art, Van Nostrand Reinhold, 1994.
invasive ocular temperature measuring tool. The advancement in [30] J.L. Miller, E. Friedman, Photonics Rules of Thumb: Optics, Electro-Optics, Fiber
this area offers a great potential to study and measure OST with Optics and Lasers, McGraw-Hill, 2004.
greater ease and precision. This enables researchers to look into [31] H. Hartridge, A.V. Hill, The transmission of infra-red rays by the media of the
eye and the transmission of radiant energy by Crookes and other glasses,
various aspects in the field of ophthalmology, and understand ocu- Proceedings of the Royal Society 89 (1915) 58–76.
lar physiology and diseases. However, to establish IR thermogra- [32] R. Mapstone, Measurement of Corneal Temperature, Experimental Eye
phy as a tool to diagnose ocular diseases requires more robust Research 7 (1968) 237–243.
[33] E.K. Plyler, N. Aquista, IR absorption of liquid water from 2 to 42 microns,
automatic cornea localization and the use of sophisticated tools, Journal of the Optical Society of America 44 (1954) 505–508.
such as image analysis, and statistical analysis.
108 J.-H. Tan et al. / Infrared Physics & Technology 52 (2009) 97–108

[34] H. Hamano, S. Minami, Y. Sugimori, Experiments in thermometry of the [63] D. Cremers, C. SchnörrJ, J. Weickert, Diffusion-snakes: combining statistical
anterior portion of the eye wearing a contact lens by means of infra-red shape knowledge and image information in a variational framework, in:
thermometer, Contact Lens and Anterior Eye 13 (1969) 12–22. Workshop on Variational and Level-set Methods in Computer Vision,
[35] J.I. Prydal, P. Artal, H. Woon, F.W. Campbell, Study of human precorneal tear Vancouver, BC, Canada, 2001, pp. 137–144.
film thickness and structure using laser interferometry, Investigative [64] T. Cootes, C.J. Taylor, D.H. Cooper, J. Graham, Active shape models-their
Ophthalmology and Visual Science 33 (1992) 2006–2011. training and application, Computer Vision Image Understanding 61 (1995) 38–
[36] P.B. Morgan, M.P. Soh, N. Efron, A.B. Tullo, Potential Applications of Ocular 59.
Thermography, Optometry Vision Science 70 (1993) 568–576. [65] K. Schindler, D. Suter, Object detection by global contour shape, Pattern
[37] I. Fatt, J. Chaston, Temperature of a contact lens on the eye, International Recognition 41 (2008) 3736–3748.
Contact Lens Clinic 7 (1980) 195–198. [66] A.L. Yuille, D.S. Cohen, P.W. Hallinan, Feature Extraction from Faces Using
[38] J.A. Scott, A finite element model of heat transport in the human eye, Physics in Deformable Templates, in: Proceedings of CVPR ’89, IEEE Computer Society
Medicine and Biology 33 (1988) 227–241. Conference, San Diego, CA, USA, 1989, pp. 104–109.
[39] C. Purslow, J.S. Wolffsohn, The relation between physical properties of the [67] P. Rysä, J. Savaranta, Corneal temperature in man and rabbit. Observations
anterior eye and ocular surface temperature, Optometry and Vision Science 84 made using an infra-red camera and a cold chamber, Acta Ophthalmologica 52
(2007) 197–201. (1974) 810–816.
[40] R. Mapstone, Ocular thermography, British Journal of Ophthalmology 54 [68] R. Mapstone, Determinants of corneal temperature, British Journal of
(1970) 751–754. Ophthalmology 52 (1968) 729–741.
[41] L. Wachtmeister, Thermography in the diagnosis of diseases of the eye and the [69] P.B. Morgan, Ocular thermography in health and disease, Ph.D. Thesis,
appraisal of therapeutic effects: a preliminary report, Acta Ophthalmologica 48 University of Manchester, Manchester, UK, 1994.
(1970) 945–958. [70] R.D. Freeman, I. Fatt, Environmental influences on ocular temperature,
[42] P.J. Murphy, P.B. Morgan, S. Patel, J. Marshall, Corneal surface temperature Investigative Ophthalmology and Visual Science 12 (1973) 596–602.
change as the mode of stimulation of the non-contact corneal aesthesiometer, [71] R. Mapstone, Corneal thermal patterns in anterior uveitis, British Journal of
Cornea 18 (1999) 333–342. Ophthalmology 52 (1968) 917–921.
[43] A.R. Fielder, A.F. Winder, G.A.K. Sheraidah, E.D. Cooke, Problems with corneal [72] I. Horven, Corneal temperature in normal subjects and arterial occlusive
arcus, Transactions of the Ophthalmological Societies of the United Kingdom disease, Acta Ophthalmologica (Copenhagen) 53 (1975) 863–874.
101 (1981) 22–26. [73] N. Efron, N. Brennan, J. Hore, et al., Temperature of the hyperaemic bulbar
[44] N. Efron, G. Young, N. Brennan, Ocular surface temperature, Current Eye conjunctiva, Current Eye Research 7 (1988) 615–618.
Research 8 (1989) 901–906. [74] J. Alio, M. Padron, Normal variations in the thermographic pattern of the
[45] P.B. Morgan, A.B. Tullo, N. Efron, Infrared thermography of the tear film in dry orbitoocular regionocular region, Diagnostic Imaging 51 (1982) 93–98.
eye, Eye 9 (1995) 615–618. [75] P. Cho, B. Brown, I. Chan, R. Conway, M. Yap, Reliability of the tear break-up
[46] F. Galassi, B. Giambene, A. Corvi, G. Falaschi, Evaluation of ocular surface time technique of assessing tear stability and the locations of the tear break-up
temperature and retrobulbar haemodynamics by infrared thermography and in Hong Kong Chinese, Optometry and vision science 69 (1992) 879–885.
colour Doppler imaging in patients with glaucoma, British Journal of [76] P. Cho, B. Brown, Review of the TBUT technique and a closer look at the TBUT
Ophthalmology 91 (2007) 878–881. of HK-Chinese, Optometry and vision science 70 (1993) 30–38.
[47] F. Galassi, B. Giambene, A. Corvi, G. Falaschi, U. Menchini, Retrobulbar [77] P. Cho, M. Yap, Age, gender, and tear break-up time, Optometry and vision
hemodynamics and corneal surface temperature in glaucoma surgery, science 70 (1993) 828–831.
International Ophthalmology 28 (2008) 399–405. [78] S. Patel, S. Laidlaw, L. Mathewson, L. McCallum, C. Nicholson, Iris colour and
[48] A.A. Sodi, B.A.D. Giambene, G.B. Falaschi, R.C. Caputo, B.B. Innocenti, A.B. Corvi, the influence of local anaesthetics on precorneal tear film stability, Acta
U.A. Menchini, Ocular surface temperature in central retinal vein occlusion: Ophthalmol (Kbh) 69 (1991) 387–392.
preliminary data, European Journal of Ophthalmology 17 (2007) 755–759. [79] P.B. Morgan, M.P. Soh, N. Efron, Corneal surface temperature decrease with
[49] S. Betney, P.B. Morgan, S.J. Doyle, N. Efron, Corneal temperature changes age, Contact Lens and Anterior Eye 22 (1999) 11–13.
during photorefractive keratectomy, Cornea 16 (1997) 158–161. [80] G.T. Raflo, P. Chart, J.J. Hurwitz, Thermographic evaluation of the human
[50] A. Mori, Y. Oguchi, Y. Okusawa, M. Ono, H. Fujishima, K. Tsubota, Use of high- lacrimal drainage system, Ophthalmic Surgery 13 (1982) 119–124.
speed, high-resolution thermography to evaluate the tear film layer, American [81] A.B. Tullo, G. Cardona, P.B. Morgan, N. Efron, Ocular and facial thermography in
Journal of Ophthalmology 124 (1997) 729–735. herpes zoster ophthalmicus and post-herpetic neuralgia, Investigative
[51] H.K. Chiang, C.Y. Chen, H.Y. Cheng, K.-H. Chen, D.O. Chang, Development of Ophthalmology and Visual Science 37 (1996) S49 (ID code: 11628).
infrared thermal imager for dry eye diagnosis, in: Proceedings of SPIE – The [82] M. Lewis, M. Dubin, V. Aandahl, Physical properties of bovine corneal collagen,
International Society for Optical Engineering, San Diego, CA, USA, 2006. Experimental Eye Research 6 (1967) 57–69.
[52] E.Y.K. Ng, J.H. Tan, E.H. Ooi, C. Chee, U.R. Acharya, Variations of Ocular Surface [83] W.D. Mathers, G. Binarao, M. Petroll, Ocular water evaporation and the dry
Temperature with Different Age Groups, Image Modeling of Human Eye, eye: a new measuring device, Cornea 12 (1993) 335–340.
Artech House, 2008. [84] M. Rolando, M. Refojo, K. Kenyon, Increased tear evaporation in eyes with
[53] J.P. Craig, I. Singh, A. Tomlinson, et al., The role of tear physiology in ocular keratoconjunctivitis sicca, Archives of Ophthalmology 101 (1983) 557–558.
surface temperature, Eye 14 (2000) 635–641. [85] H. Fujishima, I. Toda, M. Yamada, N. Sato, K. Tsubota, Corneal temperature in
[54] C. Purslow, J.S. Wolffsohn, S. Santodomingo-Rubido, The effect of contact lens patients with dry eye evaluated by infrared radiation thermometry, British
wear on dynamic ocular surface temperature, Contact Lens and Anterior Eye Journal of Ophthalmology 80 (1996) 29–32.
28 (2005) 29–36. [86] J. Nepp, K. Tsubota, E. Goto, J. Schauersberger, G. Schild, K. Jandrasits, C. Abela,
[55] L. Tan, Z.-Q. Cai, N.-S. Lai, Accuracy and sensitivity of the dynamic ocular A. Wedrich, The effect of acupuncture on the temperature of the ocular surface
thermography and inter-subjects ocular surface temperature (OST) in Chinese in conjunctivitis sicca measured by non-contact thermography: preliminary
young adults, Contact Lens and Anterior Eye 32 (2009) 78–83. results, Advances in Experimental Medicine and Biology 506 (2002) 723–726.
[56] T.-C. Chang, Y.-L. Hsiao, S.-L. Liao, Application of digital infrared thermal [87] C.R. Auker, L.M. Parver, S. Doyle, et al., Choroidal blood flow. I. Ocular tissue
imaging in determining inflammatory state and follow-up effect of temperature as a measure of flow, Archives of Ophthalmology 100 (1982)
methylprednisolone pulse therapy in patients with Graves’ ophthalmopathy, 1323–1326.
Graefe’s Archive for Clinical and Experimental Ophthalmology 246 (2008) 45– [88] K. Gugleta, S. Orgül, J. Flammer, Is corneal temperature correlated with blood-
49. flow velocity in the ophthalmic artery?, Current Eye Research 19 (1999) 496–
[57] U.R. Acharya, E.Y.K. Ng, C.Y. Gerk, J.H. Tan, Analysis of normal human eye with 501
different age groups using infrared images, Journal of Medical System 33 [89] H. Fujishima, I. Toda, Y. Yagi, K. Tsubota, Quantitative evaluation of
(2008) 207–213. postsurgical inflammation by infrared radiation thermometer and laser flare-
[58] J.H. Tan, E.Y.K. Ng, A.U. Rajendra, Automated detection of eye and cornea on cell meter, Journal of Cataract and Refractive Surgery 20 (1994) 451–454.
infrared thermogram using snake and target tracing function coupled with [90] M.D. Olson, K.M. Miller, In-air thermal imaging comparison of Legacy
genetic algorithm. Quantitative InfraRed Thermography International Journal, Advantec, millennium, and Sovereign Whitestar phacoemulsification
in press. systems, Journal of Cataract and Refractive Surgery 31 (2005) 1640–1647.
[59] J.H. Tan, E.Y.K. Ng, U.R. Acharya, Detection of eye and cornea on IR thermogram [91] A. Corvi, B. Innocenti, R. Mencucci, Thermography used for analysis and
using genetic snake algorithm, in: 9th International Conference on comparison of different cataract surgery procedures based on
Quantitative Infrared Thermography, Krakow, Poland, 2008, pp. 143–150. phacoemulsification, Physiological Measurement 27 (2006) 371–384.
[60] M. Kass, A. Witkin, D. Terzopoulos, Snakes: active contour models, [92] A.D. Rose, V. Kanade, Thermal imaging study comparing phacoemulsification
International Journal of Computer Vision 1 (1988) 321–331. with the Sovereign with WhiteStar system to the Legacy with AdvanTec and
[61] C. Xu, J.L. Prince, Snakes, shapes, and gradient vector flow, IEEE Transactions on NeoSoniX system, American Journal of Ophthalmology 141 (2006) 322–326.
Image Processing 7 (1998) 359–369. [93] E. Zeiss, Über Wärmestrahlungsmessungen an der lebenden menschlichen
[62] A. Del Bimbo, P. Pala, Visual image retrieval by elastic matching of user Hornhaut, Arch Augenheilkd 102 (1930) 523–550.
sketches, IEEE Transactions on Pattern Analysis and Machine Intelligence 19 [94] E.Y.K. Ng, E.H. Ooi, Ocular surface temperature: a 3D FEM prediction using
(1997) 121–132. bioheat equation, Computers in Biology and Medicine 37 (2007) 829–835.

You might also like