You are on page 1of 67

ADVANCED TECHNIQUES - IMRT

CONCEPT & PRINCIPLES


PLANNING IN ECLIPSE

VARIAN Lena Wittgren November 2015


ONCOLOGY
SYSTEMS
1 | VARIAN ONCOLOGY SYSTEMS

EC13.6-PPT-04-A
Inverse treatment planning

Basic idea

conventional IMRT

2 | VARIAN ONCOLOGY SYSTEMS

THIS DOCUMENT AND ITS CONTENTS ARE VARIAN’S CONFIDENTIAL AND PROPRIETARY INFORMATION AND ARE DISCLOSED SOLELY FOR THE INFORMATION OF IMMEDIATE RECIPIENT ONLY. DO NOT DISSEMINATE. EC13.6-PPT-04-A
Inverse treatment planning

• Intensity Modulated Radiation Therapy – IMRT


• Ability to modulate radiation intensity within one
treatment field

• Concave shaped dose


• Steep gradients
• Low dose bath

3 | VARIAN ONCOLOGY SYSTEMS

THIS DOCUMENT AND ITS CONTENTS ARE VARIAN’S CONFIDENTIAL AND PROPRIETARY INFORMATION AND ARE DISCLOSED SOLELY FOR THE INFORMATION OF IMMEDIATE RECIPIENT ONLY. DO NOT DISSEMINATE. EC13.6-PPT-04-A
Inverse treatment planning

• IMRT
• Definition of fixed Beam Angles
• Optimization of fluence distributions

4 | VARIAN ONCOLOGY SYSTEMS

THIS DOCUMENT AND ITS CONTENTS ARE VARIAN’S CONFIDENTIAL AND PROPRIETARY INFORMATION AND ARE DISCLOSED SOLELY FOR THE INFORMATION OF IMMEDIATE RECIPIENT ONLY. DO NOT DISSEMINATE. EC13.6-PPT-04-A
Inverse treatment planning

• Treatment planning process


• Clinical goals translated into physical constraints that
are specified mathematically
• Planning system able to change beam parameters
leading to the desired dose distribution

5 | VARIAN ONCOLOGY SYSTEMS

THIS DOCUMENT AND ITS CONTENTS ARE VARIAN’S CONFIDENTIAL AND PROPRIETARY INFORMATION AND ARE DISCLOSED SOLELY FOR THE INFORMATION OF IMMEDIATE RECIPIENT ONLY. DO NOT DISSEMINATE. EC13.6-PPT-04-A
Inverse treatment planning

• Voxel Model inside the Optimizer


• Image is converted into a voxel matrix with fixed
resolution
• All structures share the same voxel space
• Dose information is available everywhere

Image
slice

• Each structure’s volumetric


contribution per voxel is
calculated

6 | VARIAN ONCOLOGY SYSTEMS

THIS DOCUMENT AND ITS CONTENTS ARE VARIAN’S CONFIDENTIAL AND PROPRIETARY INFORMATION AND ARE DISCLOSED SOLELY FOR THE INFORMATION OF IMMEDIATE RECIPIENT ONLY. DO NOT DISSEMINATE. EC13.6-PPT-04-A
Inverse treatment planning

• Inverse Plan Optimization


• Each field divided into beamlets (2.5 x 2.5 mm)
• Beamlets are grouped to match leaf width

7 | VARIAN ONCOLOGY SYSTEMS

THIS DOCUMENT AND ITS CONTENTS ARE VARIAN’S CONFIDENTIAL AND PROPRIETARY INFORMATION AND ARE DISCLOSED SOLELY FOR THE INFORMATION OF IMMEDIATE RECIPIENT ONLY. DO NOT DISSEMINATE. EC13.6-PPT-04-A
Inverse treatment planning

• Target masking
OAR 1

OAR 2
PTV 2 PTV 1

Fluence 0
Fluence 1

8 | VARIAN ONCOLOGY SYSTEMS

THIS DOCUMENT AND ITS CONTENTS ARE VARIAN’S CONFIDENTIAL AND PROPRIETARY INFORMATION AND ARE DISCLOSED SOLELY FOR THE INFORMATION OF IMMEDIATE RECIPIENT ONLY. DO NOT DISSEMINATE. EC13.6-PPT-04-A
Inverse treatment planning

Optimization – Objective Function


• Measure of “goodness” of a solution
• Objective function Fobj for a selected set of
fluence matrices (at the end of iteration j) is:

Nt
Fobj ( j )   (d i  p) 2
i 1

9 | VARIAN ONCOLOGY SYSTEMS

THIS DOCUMENT AND ITS CONTENTS ARE VARIAN’S CONFIDENTIAL AND PROPRIETARY INFORMATION AND ARE DISCLOSED SOLELY FOR THE INFORMATION OF IMMEDIATE RECIPIENT ONLY. DO NOT DISSEMINATE. EC13.6-PPT-04-A
Inverse treatment planning

• Optimal solution is reached, when the value of


objective function reaches its minimum
• Minimizing objective function is equivalent to
setting its derivative to zero

Fobj ( j )  MIN

Fobj ( j )  0

10 | VARIAN ONCOLOGY SYSTEMS

THIS DOCUMENT AND ITS CONTENTS ARE VARIAN’S CONFIDENTIAL AND PROPRIETARY INFORMATION AND ARE DISCLOSED SOLELY FOR THE INFORMATION OF IMMEDIATE RECIPIENT ONLY. DO NOT DISSEMINATE. EC13.6-PPT-04-A
Inverse treatment planning

• Objective function Fobj combines all objectives


into a single function

Fobj  F
allPTVs
PTV   FCritical  FNTO 
allCS
F
Fields
SmoothXY

11 | VARIAN ONCOLOGY SYSTEMS

THIS DOCUMENT AND ITS CONTENTS ARE VARIAN’S CONFIDENTIAL AND PROPRIETARY INFORMATION AND ARE DISCLOSED SOLELY FOR THE INFORMATION OF IMMEDIATE RECIPIENT ONLY. DO NOT DISSEMINATE. EC13.6-PPT-04-A
Intensity Modulated Radiation Therapy - IMRT
Beamlet
• Result of Optimization –
Non-uniform fluence Map

12 | VARIAN ONCOLOGY SYSTEMS

THIS DOCUMENT AND ITS CONTENTS ARE VARIAN’S CONFIDENTIAL AND PROPRIETARY INFORMATION AND ARE DISCLOSED SOLELY FOR THE INFORMATION OF IMMEDIATE RECIPIENT ONLY. DO NOT DISSEMINATE. EC13.6-PPT-04-A
Varian MLC Design

• Carriage (bank)
• Part of MLC which
carries leaves
• Leaf
• Part of MLC used
as final beam
limiting device

13 | VARIAN ONCOLOGY SYSTEMS

THIS DOCUMENT AND ITS CONTENTS ARE VARIAN’S CONFIDENTIAL AND PROPRIETARY INFORMATION AND ARE DISCLOSED SOLELY FOR THE INFORMATION OF IMMEDIATE RECIPIENT ONLY. DO NOT DISSEMINATE. EC13.6-PPT-04-A
Varian MLC Design

• Rounded Leaf Edge & Interlocking Leaf Design


• Interlocking Mechanism

14 | VARIAN ONCOLOGY SYSTEMS

THIS DOCUMENT AND ITS CONTENTS ARE VARIAN’S CONFIDENTIAL AND PROPRIETARY INFORMATION AND ARE DISCLOSED SOLELY FOR THE INFORMATION OF IMMEDIATE RECIPIENT ONLY. DO NOT DISSEMINATE. EC13.6-PPT-04-A
Varian MLC Design

• Tongue and Groove Effect

15 | VARIAN ONCOLOGY SYSTEMS

THIS DOCUMENT AND ITS CONTENTS ARE VARIAN’S CONFIDENTIAL AND PROPRIETARY INFORMATION AND ARE DISCLOSED SOLELY FOR THE INFORMATION OF IMMEDIATE RECIPIENT ONLY. DO NOT DISSEMINATE. EC13.6-PPT-04-A
Varian MLC Design

• Leaf Interdigitation
• Ability of MLC to position two opposing neighbor
leaves next to each other

Carriage A Carriage B

16 | VARIAN ONCOLOGY SYSTEMS

THIS DOCUMENT AND ITS CONTENTS ARE VARIAN’S CONFIDENTIAL AND PROPRIETARY INFORMATION AND ARE DISCLOSED SOLELY FOR THE INFORMATION OF IMMEDIATE RECIPIENT ONLY. DO NOT DISSEMINATE. EC13.6-PPT-04-A
Varian MLC Design

• MLC Transmission
• Amount of radiation transmitted through leaves fully
blocking a beam
• Consist of
• Inter-leaf transmission
• Intra-leaf transmission

Arnfield, Med.Phys. 27 (10), 2000

17 | VARIAN ONCOLOGY SYSTEMS

THIS DOCUMENT AND ITS CONTENTS ARE VARIAN’S CONFIDENTIAL AND PROPRIETARY INFORMATION AND ARE DISCLOSED SOLELY FOR THE INFORMATION OF IMMEDIATE RECIPIENT ONLY. DO NOT DISSEMINATE. EC13.6-PPT-04-A
Varian MLC Design

• Dosimetric Leaf Gap


• Accounts for extra transmission through the rounded
leaf edge
• Modeled as an apparent gap between two closed straight
edge leaves

18 | VARIAN ONCOLOGY SYSTEMS

THIS DOCUMENT AND ITS CONTENTS ARE VARIAN’S CONFIDENTIAL AND PROPRIETARY INFORMATION AND ARE DISCLOSED SOLELY FOR THE INFORMATION OF IMMEDIATE RECIPIENT ONLY. DO NOT DISSEMINATE. EC13.6-PPT-04-A
Varian MLC Design

• Minimum Dose/Arc Dynamic Leaf Gap


• Minimal tip to tip distance which needs to be
maintained for any moving leaf pair during dynamic
mode of MLC

19 | VARIAN ONCOLOGY SYSTEMS

THIS DOCUMENT AND ITS CONTENTS ARE VARIAN’S CONFIDENTIAL AND PROPRIETARY INFORMATION AND ARE DISCLOSED SOLELY FOR THE INFORMATION OF IMMEDIATE RECIPIENT ONLY. DO NOT DISSEMINATE. EC13.6-PPT-04-A
Varian MLC Design

• MLC Leaf Speed


• Speed of the leaf at the level of isocenter.
• Maximum limit for Varian MLC is 3 cm/s.
• Model 2.5 cm/s in LMC.
• Allows for adjustment of any leaf during treatment

20 | VARIAN ONCOLOGY SYSTEMS

THIS DOCUMENT AND ITS CONTENTS ARE VARIAN’S CONFIDENTIAL AND PROPRIETARY INFORMATION AND ARE DISCLOSED SOLELY FOR THE INFORMATION OF IMMEDIATE RECIPIENT ONLY. DO NOT DISSEMINATE. EC13.6-PPT-04-A
Varian MLC Design

• Dose/Arc Dynamic Leaf Tolerance – C-Series


• Maximum allowed difference between the planned and
actual leaf positions
Planned leaf
position

Actual leaf
position

Beam On Beam On

 Dynamic Leaf Tolerance

21 | VARIAN ONCOLOGY SYSTEMS

THIS DOCUMENT AND ITS CONTENTS ARE VARIAN’S CONFIDENTIAL AND PROPRIETARY INFORMATION AND ARE DISCLOSED SOLELY FOR THE INFORMATION OF IMMEDIATE RECIPIENT ONLY. DO NOT DISSEMINATE. EC13.6-PPT-04-A
Varian MLC Design

• Jaw Over Travel


• The maximum distance
the collimator jaw can 10cm Y
extend over the central
axis
2cm x

22 | VARIAN ONCOLOGY SYSTEMS

THIS DOCUMENT AND ITS CONTENTS ARE VARIAN’S CONFIDENTIAL AND PROPRIETARY INFORMATION AND ARE DISCLOSED SOLELY FOR THE INFORMATION OF IMMEDIATE RECIPIENT ONLY. DO NOT DISSEMINATE. EC13.6-PPT-04-A
Varian MLC Design

• Leaf Over-travel
• The maximum distance over the beam CAX to which
an MLC leaf can travel
CAX

23 | VARIAN ONCOLOGY SYSTEMS

THIS DOCUMENT AND ITS CONTENTS ARE VARIAN’S CONFIDENTIAL AND PROPRIETARY INFORMATION AND ARE DISCLOSED SOLELY FOR THE INFORMATION OF IMMEDIATE RECIPIENT ONLY. DO NOT DISSEMINATE. EC13.6-PPT-04-A
Varian MLC Design

• MLC Leaf Span


• Maximum distance from tip of the most retracted leaf to
the tip of the most extended leaf

24 | VARIAN ONCOLOGY SYSTEMS

THIS DOCUMENT AND ITS CONTENTS ARE VARIAN’S CONFIDENTIAL AND PROPRIETARY INFORMATION AND ARE DISCLOSED SOLELY FOR THE INFORMATION OF IMMEDIATE RECIPIENT ONLY. DO NOT DISSEMINATE. EC13.6-PPT-04-A
Large Field IMRT

• Delivery of large
IMRT fields
• Treatment field
divided into
subfields with a
width smaller than
the leaf span
• Beam put on hold
while the carriages
move

25 | VARIAN ONCOLOGY SYSTEMS

THIS DOCUMENT AND ITS CONTENTS ARE VARIAN’S CONFIDENTIAL AND PROPRIETARY INFORMATION AND ARE DISCLOSED SOLELY FOR THE INFORMATION OF IMMEDIATE RECIPIENT ONLY. DO NOT DISSEMINATE. EC13.6-PPT-04-A
Multiple Carriage Delivery

• Used for large


modulated fields
• Carriages cannot
move while the
beam is on
• Treatment field split
into subfields with a
width smaller than
the leaf span

26 | VARIAN ONCOLOGY SYSTEMS

THIS DOCUMENT AND ITS CONTENTS ARE VARIAN’S CONFIDENTIAL AND PROPRIETARY INFORMATION AND ARE DISCLOSED SOLELY FOR THE INFORMATION OF IMMEDIATE RECIPIENT ONLY. DO NOT DISSEMINATE. EC13.6-PPT-04-A
Intensity Modulated Radiation Therapy - IMRT

• Delivery Methods

• Multiple Static Segments

• Dynamic mode of MLC (dMLC)


• Multi-leaf collimator is capable of changing the shape while the
machine is in “beam on” state

27 | VARIAN ONCOLOGY SYSTEMS

THIS DOCUMENT AND ITS CONTENTS ARE VARIAN’S CONFIDENTIAL AND PROPRIETARY INFORMATION AND ARE DISCLOSED SOLELY FOR THE INFORMATION OF IMMEDIATE RECIPIENT ONLY. DO NOT DISSEMINATE. EC13.6-PPT-04-A
Multiple Static Segment (MSS) Delivery

• An IMRT delivery method


in which the modulated
beam is delivered in a
sequence of static MLC
positions
• “Step and Shoot”
• This delivery method is
known also as “Stop and
Shoot”, “Step and Shoot” or
“Segmental MLC”

28 | VARIAN ONCOLOGY SYSTEMS

THIS DOCUMENT AND ITS CONTENTS ARE VARIAN’S CONFIDENTIAL AND PROPRIETARY INFORMATION AND ARE DISCLOSED SOLELY FOR THE INFORMATION OF IMMEDIATE RECIPIENT ONLY. DO NOT DISSEMINATE. EC13.6-PPT-04-A
Sliding Window Delivery

• An IMRT delivery method in which the shape of MLC


is changed during the beam is on
• This delivery method is known also as “Sweeping
gap”, “Moving Window”

29 | VARIAN ONCOLOGY SYSTEMS

THIS DOCUMENT AND ITS CONTENTS ARE VARIAN’S CONFIDENTIAL AND PROPRIETARY INFORMATION AND ARE DISCLOSED SOLELY FOR THE INFORMATION OF IMMEDIATE RECIPIENT ONLY. DO NOT DISSEMINATE. EC13.6-PPT-04-A
Optimal Fluence

• Result of the mathematical optimization


• Independent of the delivery method

30 | VARIAN ONCOLOGY SYSTEMS

THIS DOCUMENT AND ITS CONTENTS ARE VARIAN’S CONFIDENTIAL AND PROPRIETARY INFORMATION AND ARE DISCLOSED SOLELY FOR THE INFORMATION OF IMMEDIATE RECIPIENT ONLY. DO NOT DISSEMINATE. EC13.6-PPT-04-A
Leaf Motion Calculator (LMC)

• Software module that converts the optimal


fluence into Actual Fluence
• Leaf Sequencing File (STT file) deliverable on
treatment unit by MLC device
• Takes into account DMLC characteristics
• Supports Sliding Window and MSS for Varian
• Supports MSS for third party linac

• Calculates MU (internally only)

31 | VARIAN ONCOLOGY SYSTEMS

THIS DOCUMENT AND ITS CONTENTS ARE VARIAN’S CONFIDENTIAL AND PROPRIETARY INFORMATION AND ARE DISCLOSED SOLELY FOR THE INFORMATION OF IMMEDIATE RECIPIENT ONLY. DO NOT DISSEMINATE. EC13.6-PPT-04-A
Actual Fluence

• Derived from the optimal fluence by Leaf Motion


Calculator
• Depends on the delivery technique
• Accounts for physical, mechanical and dosimetric
characteristics of MLC

actual fluence - sliding window actual fluence – step&shoot

32 | VARIAN ONCOLOGY SYSTEMS

THIS DOCUMENT AND ITS CONTENTS ARE VARIAN’S CONFIDENTIAL AND PROPRIETARY INFORMATION AND ARE DISCLOSED SOLELY FOR THE INFORMATION OF IMMEDIATE RECIPIENT ONLY. DO NOT DISSEMINATE. EC13.6-PPT-04-A
Process
Step Fluence

IMRT Planning • Constraints Optimal fluence


• Fluence Optimization

Converter “Sequencer” – LMC Actual fluence


• Leaf Motion-Calculation
• Machine parameters

Dose Calculation Algorithms Calculated fluence


• AAA
• Acuros XB
Pre-Treatment QA
Delivery “Sliding window” Real Fluence

33 | VARIAN ONCOLOGY SYSTEMS

THIS DOCUMENT AND ITS CONTENTS ARE VARIAN’S CONFIDENTIAL AND PROPRIETARY INFORMATION AND ARE DISCLOSED SOLELY FOR THE INFORMATION OF IMMEDIATE RECIPIENT ONLY. DO NOT DISSEMINATE. EC13.6-PPT-04-A
Contouring

• Review and post-process all structures


• Voxels delineated outside intended target volumes
may cause hotspots
• Smooth contours preferred
• High density objects (ex skin-markers from CT)
• Artefacts
• BODY
• Superficial PTV:s

34 | VARIAN ONCOLOGY SYSTEMS

THIS DOCUMENT AND ITS CONTENTS ARE VARIAN’S CONFIDENTIAL AND PROPRIETARY INFORMATION AND ARE DISCLOSED SOLELY FOR THE INFORMATION OF IMMEDIATE RECIPIENT ONLY. DO NOT DISSEMINATE. EC13.6-PPT-04-A
Contouring – artefacts

35 | VARIAN ONCOLOGY SYSTEMS

THIS DOCUMENT AND ITS CONTENTS ARE VARIAN’S CONFIDENTIAL AND PROPRIETARY INFORMATION AND ARE DISCLOSED SOLELY FOR THE INFORMATION OF IMMEDIATE RECIPIENT ONLY. DO NOT DISSEMINATE. EC13.6-PPT-04-A
Contouring

• To better control the dose, additional structures


might be utilized for optimization
• Overlapping structures might need separation
• Simultaneous boost => delta PTV
• Consider margins between structures used for
optimization
• Consider use of “Dummy structures” ex “Optimization
PTV”

36 | VARIAN ONCOLOGY SYSTEMS

THIS DOCUMENT AND ITS CONTENTS ARE VARIAN’S CONFIDENTIAL AND PROPRIETARY INFORMATION AND ARE DISCLOSED SOLELY FOR THE INFORMATION OF IMMEDIATE RECIPIENT ONLY. DO NOT DISSEMINATE. EC13.6-PPT-04-A
Contouring – overlapping structures

37 | VARIAN ONCOLOGY SYSTEMS

THIS DOCUMENT AND ITS CONTENTS ARE VARIAN’S CONFIDENTIAL AND PROPRIETARY INFORMATION AND ARE DISCLOSED SOLELY FOR THE INFORMATION OF IMMEDIATE RECIPIENT ONLY. DO NOT DISSEMINATE. EC13.6-PPT-04-A
Contouring – overlapping structures

• Target coverage is most important


OAR NEW = OAR – Target

• Critical structure sparing is most important


PTV NEW = PTV – OAR

• Always keep original structures

38 | VARIAN ONCOLOGY SYSTEMS

THIS DOCUMENT AND ITS CONTENTS ARE VARIAN’S CONFIDENTIAL AND PROPRIETARY INFORMATION AND ARE DISCLOSED SOLELY FOR THE INFORMATION OF IMMEDIATE RECIPIENT ONLY. DO NOT DISSEMINATE. EC13.6-PPT-04-A
Contouring – simultaneous boost

• When optimizing several target volumes to


different dose levels – create delta-volumes to
avoid ambiguity in dose constraints

39 | VARIAN ONCOLOGY SYSTEMS

THIS DOCUMENT AND ITS CONTENTS ARE VARIAN’S CONFIDENTIAL AND PROPRIETARY INFORMATION AND ARE DISCLOSED SOLELY FOR THE INFORMATION OF IMMEDIATE RECIPIENT ONLY. DO NOT DISSEMINATE. EC13.6-PPT-04-A
Contouring – Organ motion

• Inter and Intra-fraction motion


• Addressed by margins for conventional treatment
• Important for IMRT as there may be interplay between
anatomy and MLC motion
Dose

Planned
Delivered
dose

40 | VARIAN ONCOLOGY SYSTEMS

THIS DOCUMENT AND ITS CONTENTS ARE VARIAN’S CONFIDENTIAL AND PROPRIETARY INFORMATION AND ARE DISCLOSED SOLELY FOR THE INFORMATION OF IMMEDIATE RECIPIENT ONLY. DO NOT DISSEMINATE. EC13.6-PPT-04-A
Contouring – Organ motion

Interplay Between Anatomy and MLC Leaves


Moving target leads to differences between planned
and delivered dose distribution
Dose

Planned dose
Delivered dose

Position
41 | VARIAN ONCOLOGY SYSTEMS

THIS DOCUMENT AND ITS CONTENTS ARE VARIAN’S CONFIDENTIAL AND PROPRIETARY INFORMATION AND ARE DISCLOSED SOLELY FOR THE INFORMATION OF IMMEDIATE RECIPIENT ONLY. DO NOT DISSEMINATE. EC13.6-PPT-04-A
IMRT Field setup

• Manual Setup of Fields / Template

42 | VARIAN ONCOLOGY SYSTEMS

THIS DOCUMENT AND ITS CONTENTS ARE VARIAN’S CONFIDENTIAL AND PROPRIETARY INFORMATION AND ARE DISCLOSED SOLELY FOR THE INFORMATION OF IMMEDIATE RECIPIENT ONLY. DO NOT DISSEMINATE. EC13.6-PPT-04-A
IMRT Field setup

Energy
• Photons, 6 MV (predominantly used)

Dose rate
• Often 400 MU/min (“output vs leaf speed)

Beam directions (Gantry)


• Check BEV for Target and OAR
• Avoid opposing fields

43 | VARIAN ONCOLOGY SYSTEMS

THIS DOCUMENT AND ITS CONTENTS ARE VARIAN’S CONFIDENTIAL AND PROPRIETARY INFORMATION AND ARE DISCLOSED SOLELY FOR THE INFORMATION OF IMMEDIATE RECIPIENT ONLY. DO NOT DISSEMINATE. EC13.6-PPT-04-A
IMRT Field setup

Number of fields
• Often odd number, 5-9 fields

Beam directions (Collimator)


• Check BEV for OAR
Use Jaws (not MLC) for
critical OAR:S

44 | VARIAN ONCOLOGY SYSTEMS

THIS DOCUMENT AND ITS CONTENTS ARE VARIAN’S CONFIDENTIAL AND PROPRIETARY INFORMATION AND ARE DISCLOSED SOLELY FOR THE INFORMATION OF IMMEDIATE RECIPIENT ONLY. DO NOT DISSEMINATE. EC13.6-PPT-04-A
IMRT Field setup

• Consider field size in “leaf moving”-direction (X)


to avoid splitting fields into multiple carriage
groups (~15 cm)

45 | VARIAN ONCOLOGY SYSTEMS

THIS DOCUMENT AND ITS CONTENTS ARE VARIAN’S CONFIDENTIAL AND PROPRIETARY INFORMATION AND ARE DISCLOSED SOLELY FOR THE INFORMATION OF IMMEDIATE RECIPIENT ONLY. DO NOT DISSEMINATE. EC13.6-PPT-04-A
Terminology

• Clinical objectives (Dose limits) are defined by the


Radiation Oncologist:
• Acceptable target coverage
• Acceptable max/mean doses to critical structures/ normal tissue

• Physical constraints entered into optimization are


generally different
• The objectives and priorities is a tool to get an appropriate dose
distribution and DVH

46 | VARIAN ONCOLOGY SYSTEMS

THIS DOCUMENT AND ITS CONTENTS ARE VARIAN’S CONFIDENTIAL AND PROPRIETARY INFORMATION AND ARE DISCLOSED SOLELY FOR THE INFORMATION OF IMMEDIATE RECIPIENT ONLY. DO NOT DISSEMINATE. EC13.6-PPT-04-A
IMRT Optimization

Relative weight in
V (%)

cost function
Initial
condition

Target

D-V constraints in
cost function
Organ at risk

D (Gy)

Acknowledgement A Fogliata Humanitas, Milano

47 | VARIAN ONCOLOGY SYSTEMS

THIS DOCUMENT AND ITS CONTENTS ARE VARIAN’S CONFIDENTIAL AND PROPRIETARY INFORMATION AND ARE DISCLOSED SOLELY FOR THE INFORMATION OF IMMEDIATE RECIPIENT ONLY. DO NOT DISSEMINATE. EC13.6-PPT-04-A
IMRT Optimization in Eclipse

• Check that correct algorithms are selected


• Start IMRT fluence Optimizer

48 | VARIAN ONCOLOGY SYSTEMS

THIS DOCUMENT AND ITS CONTENTS ARE VARIAN’S CONFIDENTIAL AND PROPRIETARY INFORMATION AND ARE DISCLOSED SOLELY FOR THE INFORMATION OF IMMEDIATE RECIPIENT ONLY. DO NOT DISSEMINATE. EC13.6-PPT-04-A
IMRT optimization

User selection in Calculation Model defines which UI is


opened.

49 | VARIAN ONCOLOGY SYSTEMS

THIS DOCUMENT AND ITS CONTENTS ARE VARIAN’S CONFIDENTIAL AND PROPRIETARY INFORMATION AND ARE DISCLOSED SOLELY FOR THE INFORMATION OF IMMEDIATE RECIPIENT ONLY. DO NOT DISSEMINATE. EC13.6-PPT-04-A
Optimization from Objectives

 How do we tell the system what we want?


• Contoured Structures
• Upper objectives
• Max doses to structures,
based upon the volume
• Lower objectives
• Min doses to targets
• Priorities
• Importance

 Objectives control
 Dose distribution
 Fluence

50 | VARIAN ONCOLOGY SYSTEMS

THIS DOCUMENT AND ITS CONTENTS ARE VARIAN’S CONFIDENTIAL AND PROPRIETARY INFORMATION AND ARE DISCLOSED SOLELY FOR THE INFORMATION OF IMMEDIATE RECIPIENT ONLY. DO NOT DISSEMINATE. EC13.6-PPT-04-A
Dose Volume Objectives

• Upper objective
• Maximum allowed dose
• Penalty issued if the delivered dose is higher than an
objective

51 | VARIAN ONCOLOGY SYSTEMS

THIS DOCUMENT AND ITS CONTENTS ARE VARIAN’S CONFIDENTIAL AND PROPRIETARY INFORMATION AND ARE DISCLOSED SOLELY FOR THE INFORMATION OF IMMEDIATE RECIPIENT ONLY. DO NOT DISSEMINATE. EC13.6-PPT-04-A
Dose Volume Objectives

• Lower objective
• Minimum dose request
• Structure(s) with a lower objective
are considered target structure
• Penalty issued if delivered
dose is lower than objective

52 | VARIAN ONCOLOGY SYSTEMS

THIS DOCUMENT AND ITS CONTENTS ARE VARIAN’S CONFIDENTIAL AND PROPRIETARY INFORMATION AND ARE DISCLOSED SOLELY FOR THE INFORMATION OF IMMEDIATE RECIPIENT ONLY. DO NOT DISSEMINATE. EC13.6-PPT-04-A
Dose Volume Objectives

• Mean objective
• Used as “upper objective” (OAR)
• Penalty issued if the delivered dose is higher than an
objective

53 | VARIAN ONCOLOGY SYSTEMS

THIS DOCUMENT AND ITS CONTENTS ARE VARIAN’S CONFIDENTIAL AND PROPRIETARY INFORMATION AND ARE DISCLOSED SOLELY FOR THE INFORMATION OF IMMEDIATE RECIPIENT ONLY. DO NOT DISSEMINATE. EC13.6-PPT-04-A
Normal Tissue Objective (NTO)

• Controls dose outside the target


structures
• Used to control hot spots
• Obtain sharp gradient around PTV
• Defined in the form of dose fall-off
as a function of distance from PTV
• Manual
• Automatic – penalty issued for points
having higher dose than average at a
given distance

54 | VARIAN ONCOLOGY SYSTEMS

THIS DOCUMENT AND ITS CONTENTS ARE VARIAN’S CONFIDENTIAL AND PROPRIETARY INFORMATION AND ARE DISCLOSED SOLELY FOR THE INFORMATION OF IMMEDIATE RECIPIENT ONLY. DO NOT DISSEMINATE. EC13.6-PPT-04-A
X Smoothing and Y Smoothing

• Smooth X
• Direction of leaf movement for Varian MLC
• Has higher impact on number of MU
• Smooth Y
• Does not have very high impact on number of MU
• Helps to minimize tongue and groove effect

55 | VARIAN ONCOLOGY SYSTEMS

EC13.6-PPT-04-A
IMRT optimization

56 | VARIAN ONCOLOGY SYSTEMS

THIS DOCUMENT AND ITS CONTENTS ARE VARIAN’S CONFIDENTIAL AND PROPRIETARY INFORMATION AND ARE DISCLOSED SOLELY FOR THE INFORMATION OF IMMEDIATE RECIPIENT ONLY. DO NOT DISSEMINATE. EC13.6-PPT-04-A
IMRT optimization

57 | VARIAN ONCOLOGY SYSTEMS

THIS DOCUMENT AND ITS CONTENTS ARE VARIAN’S CONFIDENTIAL AND PROPRIETARY INFORMATION AND ARE DISCLOSED SOLELY FOR THE INFORMATION OF IMMEDIATE RECIPIENT ONLY. DO NOT DISSEMINATE. EC13.6-PPT-04-A
IMRT optimization

58 | VARIAN ONCOLOGY SYSTEMS

THIS DOCUMENT AND ITS CONTENTS ARE VARIAN’S CONFIDENTIAL AND PROPRIETARY INFORMATION AND ARE DISCLOSED SOLELY FOR THE INFORMATION OF IMMEDIATE RECIPIENT ONLY. DO NOT DISSEMINATE. EC13.6-PPT-04-A
Intermediate Dose Calculation

• Automatic intermediate dose allows Optimizer to perform a 3D dose


calculation when Overall Objective Function becomes static
• IMRT optimization continues when Intermediate Dose is completed
using calculated dose as reference
• Manual Intermediate Dose Calculation is also available

59 | VARIAN ONCOLOGY SYSTEMS

THIS DOCUMENT AND ITS CONTENTS ARE VARIAN’S CONFIDENTIAL AND PROPRIETARY INFORMATION AND ARE DISCLOSED SOLELY FOR THE INFORMATION OF IMMEDIATE RECIPIENT ONLY. DO NOT DISSEMINATE. EC13.6-PPT-04-A
IMRT Post-processing

• Edit fluence

60 | VARIAN ONCOLOGY SYSTEMS

THIS DOCUMENT AND ITS CONTENTS ARE VARIAN’S CONFIDENTIAL AND PROPRIETARY INFORMATION AND ARE DISCLOSED SOLELY FOR THE INFORMATION OF IMMEDIATE RECIPIENT ONLY. DO NOT DISSEMINATE. EC13.6-PPT-04-A
Absorbed Dose calculation
Dose Prescription:
- Dose / fraction
- Amount of fractions
- Total Dose

Define algorithm for dose calculation:


- Photons: AAA, AcurosXB

61 | VARIAN ONCOLOGY SYSTEMS

THIS DOCUMENT AND ITS CONTENTS ARE VARIAN’S CONFIDENTIAL AND PROPRIETARY INFORMATION AND ARE DISCLOSED SOLELY FOR THE INFORMATION OF IMMEDIATE RECIPIENT ONLY. DO NOT DISSEMINATE. EC13.6-PPT-04-A
Rerun LMC after Renormalizing

• Leaf sequence based


on MU
• Rerun after
normalization or
changes to dose
prescription
• Always rerun for non-
Varian treatment
machines

62 | VARIAN ONCOLOGY SYSTEMS

THIS DOCUMENT AND ITS CONTENTS ARE VARIAN’S CONFIDENTIAL AND PROPRIETARY INFORMATION AND ARE DISCLOSED SOLELY FOR THE INFORMATION OF IMMEDIATE RECIPIENT ONLY. DO NOT DISSEMINATE. EC13.6-PPT-04-A
Absorbed Dose calculation

Define algorithm for dose calculation:


- Photons: AAA
Anisotropic Analytical Algorithm

3D pencil beam convolution / superposition IPS


algorithm using separate Monte Carlo
derived modeling for primary photons,
scattered extra-focal photons and electrons
scattered from the beam limiting devices.

MPS
Increasing tissue heterogeneity modeling and
accuracy of the scattered dose calculation.

63 | VARIAN ONCOLOGY SYSTEMS

THIS DOCUMENT AND ITS CONTENTS ARE VARIAN’S CONFIDENTIAL AND PROPRIETARY INFORMATION AND ARE DISCLOSED SOLELY FOR THE INFORMATION OF IMMEDIATE RECIPIENT ONLY. DO NOT DISSEMINATE. EC13.6-PPT-04-A
Absorbed Dose calculation

Define algorithm for dose calculation:


- Photons: AcurosXB
Accuracy comparable to Monte Carlo
calculations.

Very efficient for


- Large attenuation areas
- High number of fields and arc calculations.

Calculation based on cross sections. Conversion


of HU-units not just to electron density but to a
specific material.

Assignment of biological material to existing HU


is performed automatically based on CT
calibration and Acuros material library.

64 | VARIAN ONCOLOGY SYSTEMS

THIS DOCUMENT AND ITS CONTENTS ARE VARIAN’S CONFIDENTIAL AND PROPRIETARY INFORMATION AND ARE DISCLOSED SOLELY FOR THE INFORMATION OF IMMEDIATE RECIPIENT ONLY. DO NOT DISSEMINATE. EC13.6-PPT-04-A
Absorbed Dose calculation

65 | VARIAN ONCOLOGY SYSTEMS

THIS DOCUMENT AND ITS CONTENTS ARE VARIAN’S CONFIDENTIAL AND PROPRIETARY INFORMATION AND ARE DISCLOSED SOLELY FOR THE INFORMATION OF IMMEDIATE RECIPIENT ONLY. DO NOT DISSEMINATE. EC13.6-PPT-04-A
Thank you !

66 | VARIAN ONCOLOGY SYSTEMS

EC13.6-PPT-04-A
Disclaimer
Notice
• This information is an educational aid only and is intended solely as a supplemental
document. This provides information concerning the use of the Eclipse Treatment
Planning System. For complete information concerning components, safety
instructions, installation, maintenance, troubleshooting, etc., refer to the applicable
Varian product Operator’s Manual. This information does not replace the Varian
Operator’s Manual.
• Carefully read all instructions prior to use. Observe all contraindications, warnings and
precautions noted in these instructions. Failure to do so may result in patient and/or
user complications.
• This information is subject to change without notice and does not represent a
commitment on the part of Varian. Varian is not liable for errors contained in this
information or for incidental or consequential damages in connection with furnishing or
use of this material. This contains proprietary information protected by copyright. No
part of this may be reproduced, translated, or transmitted without the express written
permission of Varian Medical Systems, Inc.

67 | VARIAN ONCOLOGY SYSTEMS

EC13.6-PPT-04-A

You might also like