Professional Documents
Culture Documents
G SAFETY WARNINGS
Neck strap
slots
0º/180º
Lock
(Model 220)
Test button
LRCD220 controls
Earth connection
On Off
L - PE Voltage between L- PE
greater than 25 V
OPTION 3
2 wire test lead
(not supplied as standard item)
Loop Testing
Non-tripping loop test [No Trip]
To make a No-trip loop test.
1) Select the NO-TRIP test range on the lower range knob of all
testers.
2) Connect the mains plug test lead or 3 wire test lead to the
instrument as in option 1 or 2 above.
3) Insert the plug into a wall socket or, for the 3 wire test lead,
connect the RED lead to PHASE, the BLACK lead to NEUTRAL
and the GREEN lead to EARTH.
4) Check the L-PE and L-N indicators are lit.
5) Supply voltage is displayed.
6) Press the TEST key.
7) After a test period of 12 seconds the measured loop value is
displayed.
Note: The noise symbol indicates a noisy supply. Repeat the
loop test if displayed after test has completed.
RCD Testing
RCD type selection
1) Set the top RCD selection knob to the desired range.
Options are:
10 mA, 30 mA, 100 mA, 300 mA, 500 mA, 1000 mA (Model 220
only)
NOTE: If ‘trp’ is displayed the RCD has tripped and failed the 1/2I
test.
1) Repeat the previous test for 1/2I, but with the lower selection
knob set to I. The instrument should display the RCD trip time
in milliseconds.
If >400 ms is displayed, the RCD has failed to trip, or if it has
tripped, it did so outside the 400 ms test times.
2) Repeat the previous test for I, but with the lower selection knob
set to 5I. The instrument should display the RCD trip time in
milliseconds
If >40 ms is displayed, the RCD has failed to trip, or if it has
tripped, it did so outside the 40 ms test times.
0° or 180° testing
Both the [I] and [5I] tests should be performed for 0° and 180°.
1) Connect the test leads as described for the 1/2I RCD test
previously.
2) Select the appropriate RCD rated current on the top range
knob.
I = I test
IIIII = 5 I test
Example shows 5I on 0º
Error Messages
Batteries
Replace with new batteries. Do not mix old and new batteries
Megger Limited
Archcliffe Road, Dover
Kent CT17 9EN England
T +44 (0)1 304 502101
F +44 (0)1 304 207342
E uksales@megger.com
Megger
4271 Bronze Way, Dallas,
Texas 75237-1019 USA
T +1 800 723 2861 (USA ONLY)
T +1 214 333 3201
F +1 214 331 7399
E ussales@megger.com
Megger
Z.A. Du Buisson de la Couldre
23 rue Eugène Henaff
78190 TRAPPES France
T +33 (0)1 30.16.08.90
F +33 (0)1 34.61.23.77
E infos@megger.com
Megger Limited
110 Milner Avenue Unit 1
Scarborough Ontario M1S 3R2
Canada
T +1 416 298 9688 (Canada only)
T +1 416 298 6770