REPORT OF ANALYSIS

Reference No. : ADMATEL 2202-2998

Customer : Lagro High School

Sample Label : Chalk Sample

Analysis Requested : X-Ray Fluorescence

Date Received : February 9, 2022

Date Tested : February 9, 2022

I. Test Description

As-received chalk sample, labeled as Chalk Sample, was submitted for element content
analysis using Bruker S1 TITAN 800 handheld X-Ray Fluorescence (HXRF) as shown in Figure
1. It was conducted using the XRF parameters listed in the table below.

Table 1. Parameters used for XRF analysis.

Parameters Parameter Type/ Value


Application Alloys2 (Metals)
Test total duration 35 secs.
Measurement units % conc.
Replicates 3
Maximum voltage 40 keV

Page 1 of 3
This report shall not be reproduced in any form, except in full, without written approval of ADMATEL
Republic of the Philippines
Department of Science and Technology
INDUSTRIAL TECHNOLOGY DEVELOPMENT INSTITUTE
ADVANCED DEVICE AND MATERIALS TESTING LABORATORY
DOST Cpd., General Santos Ave., Bicutan, Taguig City
Tel.: (02) 837-0461 Email: services@admatel.com
http://www.admatel.com/

Figure 1. Photograph of the as-received Chalk Sample for XRF analysis.

II. XRF Examination Results

The summarized results of the XRF test for as-received Chalk Sample are presented in Table
2. The results displayed are averaged from three (3) measurements indicating the elemental
concentrations and standard deviations in percent weight (% wt.) – sorted from highest to lowest
concentration.

Table 2. Summarized XRF results for Chalk Sample.


Element Conc. [% wt.] Stddev [% wt.]
Calcium Ca 33.863 0.130
Aluminum Al 3.002 1.636
Silicone Oxide SiO2 1.565 0.785
Iron Fe 0.183 0.025
Chlorine Cl 0.140 0.054
Silver Ag 0.058 0.009
Strontium Sr 0.023 0.001
Copper Cu 0.007 0.003
Barium Ba < LOD 0.009
*LOD - Limit of Detection = below 0.001 %wt.

Based from the HXRF test results, the Chalk Sample is composed primarily of 33.863%
calcium (Ca), 3.002% aluminum (Al), 1.565% silicon oxide (SiO2), and trace amounts of other
metallic elements.

Page 2 of 3
This report shall not be reproduced in any form, except in full, without written approval of ADMATEL
Republic of the Philippines
Department of Science and Technology
INDUSTRIAL TECHNOLOGY DEVELOPMENT INSTITUTE
ADVANCED DEVICE AND MATERIALS TESTING LABORATORY
DOST Cpd., General Santos Ave., Bicutan, Taguig City
Tel.: (02) 837-0461 Email: services@admatel.com
http://www.admatel.com/

VALIDITY OF THE REPORT: The test results are those obtained at the time of the test
and pertain only to the sample/s received by ADMATEL.

Jess Gabriel V. Belleza


Analyst
02/09/2022
Date:_________________

Cesar J. Bandoquillo III


Laboratory Head
02/09/2022
Date:_______________

Issued under the authority of:

Araceli M. Monsada, Dr.-Engg


Laboratory Manager
02/09/2022
Date:_________________

Form: AL-21-F25
Issue: December 14, 2020
Revision: 03

Page 3 of 3
This report shall not be reproduced in any form, except in full, without written approval of ADMATEL

You might also like