‫ مقارنة ألهم الشركات‬

Dreager
Hamilton
GE Healthcare (Datex-Ohmeda)
Simens
VIASYS Healthcare
DRAEGER Evila GE HEALTHCARE HAMILTON
2 dura (DATEX- OHMEDA_ GALILEO
Centiva/5 GOLD

FDA CLEARANCE Yes Yes Yes


CE MARK (MDD) No Yes Yes
PATIENT TYPE Adult, pediatric, Adult, pediatric, Adult, pediatric,
neonatal optional neonatal neonatal
CONTROLS
Tidal volume, mL 3_2000 20_2000 10_2000
Inspiratory flow, L/min 6_180 2_99 0_180
Inspiratory pressure, 0_80 1_59 0_100
cm H2O
Respiratory rate, 0_150 4_100 0.5_120
breaths/min
Inspiratory time, sec 0.1_30 0.06_13.6 0.2_59.9
Expiratory time, esc 1.1_30 0.2_13.6 0.2_59.9
IE ratio 1:300 TO 300:1 1:9 TO 1:4 1:9.9 TO 4:1
Inspiratory hold/plateau Yes Yes Yes
Expiratory hold Yes NO Yes
FiO2 21_100 21_100 21_100
Manual breath Yes NO Yes
PEEP/CPAP, cm H2O 0-35 2_35 0-50
Pressure support, 0_80 0_59 0_100
cmH2O
Nebulizre Yes Yes Yes
Trigger mechanism Flow(pressure) Flow Pressure(Flow)
Bias/base flow range, Not specified 3_30 1_30

1
L/min Yes/yes Yes/yes Yes/yes
Pressure Yes NO Yes
slop/ramp,L/min
adjustment Yes Yes Yes
Sigh
100%O2

OPERATING MODES
Assist/control
Volume breaths Yes Yes Yes
Pressure breaths Yes Bilevel Yes
SIMV
Volume breaths Yes Yes Yes
Pressure breaths Yes Bilevel Yes
Pressure support Yes Yes Yes
Spontaneous/CPAP Yes Yes Yes
Pressure support
Apena-backup vent Yes Yes Yes
Combination mode Yes Bilevel Yes
Responsive valve Yes Yes Yes
Bilevel/APRV Yes Yes Yes
MONITORED PARAMETERS

PRESSURE Yes Yes Yes


PIP Yes Yes Yes
MAP Yes Yes Yes
PEEP
VOLUME Yes Yes Yes
Tidal Yes Yes Yes
Minute Yes Yes Yes
Spontaneous minute Yes Yes Yes
FiO2 Yes Yes Yes
Respiratory rate Yes No Yes
Inspiratory time Yes No Yes
Expiratory time Yes No Yes
IE
PATIENT ALARMS
FiO2 Yes Yes Yes
Low minute volume Yes Yes Yes
Low Inspiratory Yes Yes Yes
Pressure
High pressure Yes Yes Yes
Loss of PEEP Yes Yes Yes

2
Apena Yes Yes Yes
High continuous Yes Yes Yes
Pressure/occlusion
Inverse IE Yes Yes Yes
High respiratoty rate Yes Yes Yes
High minute volume Yes Yes Yes
High PEEP Yes NO Not specified
Breathing circuit Yes Yes Yes
disconnect
EQUIPMENT ALARMS
Gas supply failure Yes Yes Yes
Power failure Yes Yes Yes
Vent inoperative Yes Yes Yes
Low battery Yes Yes Yes
Self-diagnostics YES Yes Yes
INTERFACING RS232 RS232 RS232
Output ports Optional Yes, nurse call Optional
Remote alarm/display Yes NO Yes
Analog output
Report generation Yes Yes Yes
Display Yes Yes NO
Hardcopy Yes NO NO
Archival disk Yes NO Not specificed
Network
PNEUMATIC POWER
Compressor Optional Optional Optional
Compressor gases O2, air O2, air O2, air
2.7-6 bar,40-87 40-80 psi,280-600
Pressure ranges psi kpa 29-86 psi
Turbine/piston Not specified Not specified Not specificed
INTERNAL BATTERY
Type(number) Not specified Sealed lead aci Sealed lead aci
Operating time, hr 2 0.5 1
Rechargeable Yes Yes Yes
Rechargeable time, hr Not specified 12 6
EXTERNAL BATTERY
Type(number) Not specified Sealed lead aci Not specified
Operating time, hr Not specified 1 Not specified
Rechargeable Not specified Yes Not specified
Rechargeable time, hr Not specified 10 Not specified