Information: -
Manufacturer: HITACHI
Urgency
medium
Units concerned
All c60000 configurations with SW version 02-05 where the 564-0003 alarm occurs.
Summary
In case 564-0003 alarm occurs after the software upgrading to 02-05, please replace the current sensor
plate with the extended one.
Action
In case SW version 02-05 is installed and the alarm 564-0003 (feeder alarm) occurs the current sensor
plate has to be exchanged with the new one.
Note: The old sensor plate has to be stored carefully for a possible re-modification.
Reasons of Change
The movement of Rack ID reading at Sampler unit is changed in Ver. 02-05 software: the shift distance
after detection stop for Rack ID is changed from 2.0 mm to 2.6 mm. This change causes less margin of
stable detection status after the shift, which may result in inadequate detection during next motion.
(Alarm code 564-0003)
Details
Refer to the attachment.
Interchangeability
The extended plate is also available for 02-07 software.
Page: 1/4
Caution
No special caution is necessary
Parts required
Material Number Parts name P/N Remark
05194318001 Plate Sensor L2 727-8639 incl. ratchet
Attachement
PCP SHUTTER ASSY on Sampler unit
8.5 mm
Page: 2/4
P/N 727-8639 PCP SHUTTER 2 ASSY
For 02-05 or 02-06
In case 564-0003 alarm occurs
9.1 mm
<Recommended tool>
M3 elbow screw driver (Ratchet type or bar type. Ratchet type is recommended.)
<Necessary>
L1 D: less than 11 mm
L2 (If not, unable to access the screws)
L1, L3: more than 20 mm
(If not, unable to access the screw)
<Recommended>
L2, L3: less than 50 mm
(If not, difficult to push the screw)
L3
<Exchange procedure>
1. Power OFF the instrument before exchange.
2. Cut two wire binders in advance for easy work, and remove C-4 sensor.
C-4 sensor
Page: 3/4
3. Loosen two screws to exchange the plate through the hole of C-4 sensor.
4. Exchange the sensor plate and tighten the screws as they were.
Know How: Far side first, near side second for easy understanding of screw hole.
5. Make sure that the sensor plate doesn’t touch all sensors when you move the carriage in full-
stroke. If touch, go back to procedure 3.
Page: 4/4