M Maintenance and Safety

ULS 0132 SM 01/03052 /M/1.1.0

1. Planned maintenance

The Maintenance and Safety has to be done once a year according


to the procedure described in the Maintenance and Safety LIT 160/
170 protocoll and send a copy to HMT Customer Support.

After each service operation you need to check the ULS132


according to the procedure described in the Test and Release
ULS0132 protocoll and send a copy to HMT Customer Support.

This is to document the good condition of the device after service


operation for both service technican and customer.

Note
Use T&R document in this chapter as COPY MASTER.

M-0
ULS 0132PA10 T&R TEST and RELEASE Option LIT 01X0/TRC0215
Note:
This T&R Protokoll has to be filled out by a HMT authorized Service Engineer for all Service activities.
SR #
CS # /
Data of Service Call

Service Organisation: Date of Customer Call: Time of Call:


Service Representative: Date of Service: Date Service
Customer / Site: completed:
Serial # LithoTron: Serial # SD360/: Unit down/ operational
Serial # Ultrasound: # Unison 400: Priority:
Original Problem:

Corrective maintenance:

Further Actions necessary /Remarks:

Part exchange Part description: HMT part number: Serial # Part In: Serial # Part out:

Test Procedure o.K. Fault Fault


Additional checks for correct alignment of the Ultrasound-option clear
ULS 132 to LithoTron F2-Focus

Refer to the radiation protection standards before X-raying!

1. Usage of the F2-focus phantom LithoTron without x-ray


• Adapt the F2-focus phantom to the therapy head
• Mount the Ultrasound phantom to the transducer holder
• Start the US-device SD360/Unison400 and go to the Litho-mode
• Select at the LithoTron user panel the ULS-mode
• Start the ULS-system
• Change the coordinate-cross at the US-monitor to 0,720 (CA3540)
• Set the front tip of the Ultrasound phantom to the cone of the F2-focus phantom
• Check the coordinates at the user panel display for x,y,z
<>

<>= (=±2,0mm) = (=±2,0mm) +/ - = (=±2,0mm)

2. Usage of ultrasound customer phantom with x-ray (Box with metal tip and water inside)
• Follow the instructions in the User Manual for precision checks (Chapter Maintenance)
• Move table that all three coordinates x,y,z at the display user panel are smaller than ±0.5mm
• X-ray in PA and CC position
• Measure the distance between the tip and the center of the x-ray cross on the X-ray monitor
Distance tip to center cross:
Calculation see chapter C
distance
PA-position (mm) : ( Value < 3mm)
CC-position (mm) : ( Value < 3mm)

Authorized Service Engineer Customer

Date /Signature

ULS 0132PA10 T&R Released: 29.02.00/ NG approved/ jf Page 1 / 1