Troubleshooting Guide
Iconos R200
© Siemens, 2009
07030930
07030518
07030500
© Siemens, 2009
- For internal use only - All documents may only
be used by authorized personnel for rendering
services on Siemens Healthcare Products. Any
document in electronic form may be printed
once. Copy and distribution of electronic docu-
ments and hardcopies is prohibited. Offenders
will be liable for damages. All other rights are re-
served.
2009
2 Copyright / Version / Disclaimer
Copyright
1Copyright / Version / Disclaimer
“© Siemens, 2009“ refers to the copyright of a Siemens entity such as Siemens Aktienge-
sellschaft - Germany, Siemens Shenzhen Magnetic Resonance Ltd. - China, Siemens
Shanghai Medical Equipment Ltd. - China, Siemens Medical Solutions USA Inc. - USA
and/or Siemens Healthcare Diagnostics Inc. - USA.
Document Version
Siemens reserves the right to change its products and services at any time.
In addition, manuals are subject to change without notice. The hardcopy documents corre-
spond to the version at the time of system delivery and/or printout. Versions to hardcopy
documentation are not automatically distributed. Please contact your local Siemens office
to order current version or refer to our website http://www.healthcare.siemens.com.
Disclaimer
Siemens provides this documentation “as is“ without the assumption of any liability under
any theory of law.
The service of equipment described herein is to be performed by qualified personnel who
are employed by Siemens or one of its affiliates or who are otherwise authorized by Sie-
mens or one of its affiliates to provide such services.
Assemblers and other persons who are not employed by or otherwise directly affiliated with
or authorized by Siemens or one of its affiliates are not entitled to use this documentation
without prior written authority.
Meaning 0
Reason, Cause: 0
The saved position of the column or the spotfilm device when switching on the NV-RAM on
the D320/XCU board does not match the one that is calculated from the potentiometer val-
ues after switching on the system.
Possible causes:
1. In 99.9% of all cases, the cause is unclear and can be found in almost every error log
- without a HW defect being responsible for it,
- without there being a misadjustment.
Action 0
1. Perform adjustment and linearization of the potentiometers; see "Spotfilm Device and
Column Longitudinal Drive" in the ICONOS Online Help:
¹ Spotfilm Device and Column Longitudinal Drive
Meaning 0
Reason, Cause: 0
The BUC has not signaled a position for too long a period:
Action 0
The BUC (SW) is monitored by the SCU with the new VE00B system SW.
If the SCU senses irregularities (see Reason, Cause), the safety circuit is switched off by
the Master drive. It will no longer be possible to move the unit.
The system will function again properly after a system switch-off/on.
Meaning 0
Reason, Cause: 0
The BUC has indicated that it has moved into the target position, however the actual posi-
tion is too far away from it:
Action 0
If the SCU senses irregularities (see Reason, Cause), the drive is switched off. After
acknowledging the error, the drive can be operated again.
The system will function again properly after acknowledging the error.
Meaning 0
Reason, Cause: 0
Action 0
If the SCU senses irregularities (see Reason, Cause), the drive is switched off. After
acknowledging the error, the drive can be operated again.
The system will function again properly after acknowledging the error.
Meaning 0
Reason, Cause: 0
Action 0
If the SCU senses irregularities (see Reason, Cause), the drive can be operated again only
after acknowledging the error.
The system will function again properly after acknowledging the error.
Meaning 0
Reason, Cause: 0
The BUC drive reaches positions that do not match the previously calculated values:
In addition, the last and the current positions are signaled line numbers with Error Numbers
580 and 581.
Action 0
The BUC (SW) is monitored by the SCU with the new VE00B system SW.
If the SCU senses irregularities (see Reason, Cause), the drive is switched off. After
acknowledging the error, the drive can be operated again.
The system will function again properly after acknowledging the error.
The affected drive can be determined from the additional information, see table.
BUCT_DRV on line: 0 Tilt
1 Tabletop longitudinal
2 Lift
NOTE If an error occurs during combined lift and tilt down move-
ment, drive 0 (tilt) is always displayed.
The drive can be determined only by the position (Error Num-
bers 580 and 581).
Meaning 0
Reason, Cause: 0
Action 0
The BUC (SW) is monitored by the SCU with the new VE00B system SW.
If the SCU senses irregularities (see Reason, Cause), the safety circuit is switched off by
the Master drive. It will no longer be possible to move the unit.
The system will function again properly after a system switch-off/on.
Meaning 0
Reason, Cause: 0
Action 0
The BUC (SW) is monitored by the SCU with the new VE00B system SW.
If the SCU senses irregularities (see Reason, Cause), the safety circuit is switched off by
the Master drive. It will no longer be possible to move the unit.
The system will function again properly after a system switch-off/on.
The affected drive can be determined from the additional information, see table.
BUCT_DRV on line: 0 Tilt
1 Tabletop longitudinal
2 Lift
NOTE If an error occurs during combined lift and tilt down move-
ment, drive 0 (tilt) is always displayed.
The drive can be determined only by the position (Error Num-
bers 580 and 581).
Meaning 0
Reason, Cause: 0
Action 0
The BUC (SW) is monitored by the SCU with the new VE00B system SW.
If the SCU senses irregularities (see Reason, Cause), the safety circuit is switched off by
the Master drive. It will no longer be possible to move the unit.
The system will function again properly after a system switch-off/on.
The affected drive can be determined from the additional information, see table.
BUCT_DRV on line: 0 Tilt
1 Tabletop longitudinal
2 Lift
NOTE If an error occurs during combined lift and tilt down move-
ment, drive 0 (tilt) is always displayed.
The drive can be determined only by the position (Error Num-
bers 580 and 581).
Meaning 0
Reason, Cause: 0
Action 0
The BUC (SW) is monitored by the SCU with the new VE00B system SW.
If the SCU senses irregularities (see Reason, Cause), the safety circuit is switched off by
the Master drive. It will no longer be possible to move the unit.
The system will function again properly after a system switch-off/on.
The affected drive can be determined from the additional information, see table.
BUCT_DRV on line: 0 Tilt
1 Tabletop longitudinal
2 Lift
NOTE If an error occurs during combined lift and tilt down move-
ment, drive 0 (tilt) is always displayed.
The drive can be determined only by the position (Error Num-
bers 580 and 581).
Meaning 0
Reason, Cause: 0
Action 0
The BUC (SW) is monitored by the SCU with the new VE00B system SW.
If the SCU senses irregularities (see Reason, Cause), the safety circuit is switched off by
the Master drive. It will no longer be possible to move the unit.
The system will function again properly after a system switch-off/on.
Meaning 0
Reason, Cause: 0
NOTE Depending on the error before or after Error 580, the meaning
of the value in Error 580 will change.
For example:
1. 032 R0574 N SCU: incorrect position data; Iconos R200 in BUCT_DRV module on line
0
2. 032 R0580 N SCU: BUC info channel 1; Iconos R200 in BUCT_DRV module on line
+8850
Explanation of the Example
1. BUCT_DRV on line 0
- 0: Tilt
- 1: Tabletop longitudinal
- 2: Lift
¹ Tilt drive
2. BUCT_DRV on line +8850
- Tilt resolution 0.01°
¹ The last position of the tilt drive was +88.5°
Meaning of Error:
• The last position of the drive
¹ Error 574
• The time since the stop (msec)
¹ Error 575
• The limit position of the drive
¹ Error 576
• The target position of the drive
¹ Error 577
• The last position of the drive
¹ Error 578
Meaning 0
Reason, Cause: 0
NOTE Depending on the error before or after Error 581, the meaning
of the value in Error 581 will change.
For example:
1. 032 R0574 N SCU: incorrect position data; Iconos R200 in BUCT_DRV module on line
0
2. 032 R0581 N SCU: BUC info channel 1; Iconos R200 in BUCT_DRV module on line
+8850
Explanation of the Example
1. BUCT_DRV on line 0
- 0: Tilt
- 1: Tabletop longitudinal
- 2: Lift
¹ Tilt drive
2. BUCT_DRV on line +8850
- Tilt resolution 0.01°
¹ The last position of the tilt drive was +88.5°
Meaning of Error:
• The actual position of the drive
¹ Error 574
¹ Error 576
¹ Error 577
¹ Error 578
Meaning 0
Bucky wall stand (BWS): CAN bus malfunction. Many telegrams had to be repeated.
Reason, Cause: 0
A CAN controller needed several attempts to establish communication to the SCU or vice
versa. Here the problem is not necessarily the CAN controller for the Bucky wall stand!
Error registered in XCS error log but not displayed on remote control consoles. Error has
no consequence, but might occur in XCS error log frequently.
Error entered into XCS error log during system boot only. Error not logged during operation.
Iconos R200 from system serial number 5100 and Vertix PRO/TOP with catapult bucky tray
16 bit (P/N 7732849 / 7723831) affected exclusively.
Action 0
Solution in work. Replacement of parts does not help solving the feature.
• The minimum insertion force "F" for the 30 x 24 cm /12 x 10 inch cassette should not
exceed 25 N in all unit positions.
¹ A change with "Adjustment/Calibration/Spotfilm device/Cassette tensioning:
Close speed" from 50% to 90% in the SSW results in a change of the insertion
force of approx. 20N.
- Use all long holes; if needed, shorten the plastic part of the switch-off elements.
¹ Measure from the lowest point of the I.I. (light distributor) to the end of the
switch-off element. In addition, make sure that if a switch-off element is manually
actuated, a minimum distance (Z) to the I.I. (light distributor) is maintained (the
minimum distance (Z) is the actuated switch to the I.I.).
From Version 02 to 03